K Number
K151114

Validate with FDA (Live)

Date Cleared
2015-12-11

(228 days)

Product Code
Regulation Number
878.4460
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A latex surgeon's gloves is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Device Description

The proposed device, Latex Surgeon's Gloves Powdered and Latex Surgeon's Gloves Powder Free, Polymer coated with protein content labeling claim of 50 µg/dm² or less per glove of extractable protein is a sterilized and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination. The proposed device is made of natural rubber latex, as per standard ASTM D3577-09. The rubber surgical gloves classification is: Type I - gloves compounded primarily from natural rubber latex. The proposed device includes Powdered and Powder Free Latex Surgeon's Gloves, and the come in different sizes ( 5½, 6, 6½, 7, 7½, 8, 8½, 9 ). All gloves share the same color, creamy white and all have beaded cuff and textured at the finger tips and palm. The proposed device is sterilized either using Ethylene Oxide Sterilization or Gamma irradiation method to achieve the Sterility Assurance Level (SAL) of 10th and place in a sterility maintenance package to ensure a shelf life of 3 years.

AI/ML Overview

This appears to be a 510(k) summary for a medical device (latex surgeon's gloves), not a study specifically designed to assess an AI/ML device or its performance against acceptance criteria in the way typically expected for such technologies. Therefore, many of the requested elements (like sample size for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance) are not applicable or not detailed in this document.

However, I can extract the acceptance criteria and demonstrated performance for the physical and biological characteristics of the surgeon's gloves, as described in the "Product Comparison in Compliance with ASTM D 3577-09 Standard" and "Technological Characteristics Comparison" sections.

Here's the information formatted according to your request, with notes indicating where the requested AI-specific information is not present in this document:

1. Table of Acceptance Criteria and Reported Device Performance

SL.NOCRITERIASPECIFICATION AS PER ASTM D3577-09 STANDARD / Acceptance CriteriaAVERAGE VALUE OF SUBJECT DEVICE / Reported Device Performance
1Length
Size 5½Min 265mm276 mm
Size 6Min 265mm278mm
Size 6½Min 265mm281mm
Size 7Min 265mm281mm
Size 7½Min 265mm281mm
Size 8Min 265mm283mm
Size 8½Min 265mm283mm
Size 9Min 265mm284mm
2Width
Size 5½70+/-6mm74mm
Size 676+/-6mm79mm
Size 6½83+/-6mm87mm
Size 789+/-6mm93mm
Size 7½95+/-6mm98mm
Size 8102+/-6mm106mm
Size 8½108+/-6mm112mm
Size 9114+/-6mm118mm
3Finger Thickness (All sizes)Min 0.10mm0.18mm
4Palm Thickness (All sizes)Min 0.10mm0.16mm
5Cuff Thickness (All sizes)Min 0.10mm0.12mm
6Tensile Strength
Before aging (All sizes)24Mpa minimum28.5Mpa
After aging @ 70°±2C for 166±2 hr (All sizes)18Mpa minimum24.2Mpa
7Ultimate Elongation
Before aging (All sizes)750% minimum934%
After aging @ 70°±2C for 166±2 hr (All sizes)560% minimum758%
8Stress at 500% before aging (All sizes)5.5 MPa Max3 Mpa
9Pinhole AQL
Before aging (All sizes)Max 1.51.0
After aging @ 70°C for 7 days (All sizes)Max 1.51.0
10Freedom from HolesComplies with ASTM D3577 and ASTM D5151Yes (device met acceptance criteria)
11Powder Content for powdered gloveASTM D3577 and ASTM D6124 (Max 15 mg/dm²)10 mg/dm² on average (device met acceptance criteria)
12Powder Content for powder free gloveASTM D3577 and ASTM D6124 (Powder content < 2 mg/Glove)Yes (device met acceptance criteria - implicitly < 2 mg/Glove, or else the text would specify)
13Protein Content (for Latex Surgeon's Gloves Powder Free, Polymer coated)ASTM D3577, ASTM D5712-10 and ASTM D6499 (Protein Content less than 50 µg/dm²)Less than 50 µg/dm² (device met acceptance criteria)
14Biocompatibility - Skin sensitizationISO 10993-10: Non-sensitizer under the conditions of the studiesNon-Sensitizer (device met acceptance criteria)
15Biocompatibility - Skin irritationISO 10993-10: Non-irritant under the conditions of the studiesNon-Irritant (device met acceptance criteria)
16SterilizationETO SAL- 10⁻⁶ OR Radiation SAL- 10⁻⁶ (meets acceptance level for sterilization)ETO SAL- 10⁻⁶ OR Radiation SAL- 10⁻⁶ (device met acceptance criteria)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Set Sample Size: The document does not specify exact sample sizes for each test conducted (e.g., how many gloves were tested for length, pinholes, tensile strength, etc.). It refers to meeting the standards like ASTM D3577-09, which would define the statistical sampling plans.
  • Data Provenance: The submitting company is ST.MARY'S RUBBERS PVT. LTD. from Kottayam District, Kerala State, India. The tests were presumably conducted internally or by a contracted lab, but the specific location of data generation for these bench tests is not explicitly stated. The nature of these tests (physical and chemical properties) means there isn't "retrospective" or "prospective" data in the clinical sense. It's a characterization of manufactured product batches.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • This information is not applicable for this type of device and study. The "ground truth" for the physical and chemical properties of the gloves is established by objective measurements against international standards (ASTM, ISO), not by expert consensus or interpretation of medical images/data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • This information is not applicable for this type of device and study. Adjudication methods are typically used in clinical or image-based studies where expert interpretation can vary. The tests here are objective measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This information is not applicable for this type of device and study. This is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This information is not applicable for this type of device and study. This is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for these tests is defined by the quantitative specifications and methodologies outlined in the referenced international standards (ASTM D3577-09, ASTM D5151-2011, ASTM D6124-2011, ASTM D5712-10, ASTM D6499-12, ASTM F1929-2004, ISO 10993-10:2010, ISO 1137-2:2006, ISO 10993-7:2008, ISO 11135-1:2007). These standards provide the validated methods and acceptable limits for measuring physical, mechanical, and biological properties.

8. The sample size for the training set

  • This information is not applicable for this type of device and study. This is not an AI/ML device, and there is no "training set" in the machine learning sense. The manufacturing process is controlled to meet the specifications, and testing is performed on manufactured lots.

9. How the ground truth for the training set was established

  • This information is not applicable for this type of device and study. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure, composed of three interconnected profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 11, 2015

ST.MARY'S RUBBERS PVT. LTD c/o Mr. Manoj Zacharias Liberty Management Group Ltd 2871 Coastal Dr. Aurora, IL 60503

Re: K151114

Trade/Device Name: Medismart ( Latex Surgeon's Gloves powdered ) and Medismart + (Latex Surgeon's Gloves Powder Free, Polymer coated with protein content labeling claim of 50 ug/dm2 or less per glove of extractable protein ) Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's glove Regulatory Class: I Product Code: KGO Dated: November 7, 2015 Received: November 13, 2015

Dear Mr. Manoj Zacharias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151114

Device Name

Medismart (Latex Surgeon's Gloves Powdered)

Indications for Use (Describe)

A latex surgeon's gloves is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K151114

Device Name

Medismart + (Latex Surgeon's Gloves Powder Free, Polymer coated with protein content labeling claim of 50 µg/dm² or less per glove of extractable protein.)

Indications for Use (Describe)

A latex surgeon's gloves is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVES POWDERED & LATEX SURGEON'S GLOVES POWDER FREE (POLYMER COATED)

PAGE NO:-27-1 SECTION NO:27

510K SUMMARY as required by: 21CFR § 807.92 A. APPLICANT INFORMATION

K151114

1.Submitter NameST.MARY'S RUBBERS PVT. LTD.
Date SubmittedDecember 10, 2015
AddressKoovappally P.O, Kanjirappally, Kottayam District,Kerala State, India-686518
Phone+91 4828 251533 , 239857,324570
Fax+91 4828 251353
E-mailinfo@stmarysrubbers.com
Contact PersonS.Srinivasan
DesignationQA Manager
Contact Number+91 9497244270
Contact Emailqa@stmarysrubbers.com

B. US AGENT & CONTACT PERSON INFORMATION

17US agent & contact person nameManoj Zacharias
18AddressLiberty Management Group Ltd.2871, Coastal Dr. Aurora, IL-60503, USA.
19Phone(630) 270-2921
20Fax(815) 986-2632
21E-mailmanoj@libertymanagement.us

C. DEVICE IDENTIFICATION

Common NameSurgeon's Gloves
Device proprietary or trade namea. Medismart (Latex Surgeon's Gloves powdered ),b. Medismart + (Latex Surgeon's Gloves Powder Free, Polymer coated with protein content labeling claim of 50 $\mu$ g/dm² or less per glove of extractable protein )
Classification nameSurgeon's Gloves
Device ClassificationI
Product CodeKGO
Regulation Number21 CFR 878.4460
Review PanelGen & Plastic Surgery

{5}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVES POWDERED & LATEX SURGEON'S GLOVES POWDER FREE (POLYMER COATED)

SECTION NO:27 PAGE NO:-27-2

D.PREDICATE DEVICE INFORMATION

Name of device510k Number510K Owner
SURGTEX Latex Surgeon's Glovespowdered andSURGTEX Latex Surgeon's Glovespowder free-polymer coated with proteincontent labeling claim of 50 $[mu]g/dm^2$ orless per glove of extractable proteinK130450Purna Bina SDN BHDPlot 5, Jalan Mahsuri, 7.5km,Jalan Mersing,Kluang Industrial AreaKluang, Johor 86000
Latex Surgeon's Gloves (powdered andPowder free)K130301Elimedical Devices (Eujian) Inc.1E06, Wuping Industrial Park, Wuping,Fujian 364300, China

E. DESCRIPTION OF THE DEVICE

The proposed device, Latex Surgeon's Gloves Powdered and Latex Surgeon's Gloves Powder Free, Polymer coated with protein content labeling claim of 50 µg/dm² or less per glove of extractable protein is a sterilized and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination.

The proposed device is made of natural rubber latex, as per standard ASTM D3577-09

The rubber surgical gloves classification is: Type I - gloves compounded primarily from natural rubber latex.

The proposed device includes Powdered and Powder Free Latex Surgeon's Gloves, and the come in different sizes ( 5½, 6, 6½, 7, 7½, 8, 8½, 9 )

All gloves share the same color, creamy white and all have beaded cuff and textured at the finger tips and palm.

The proposed device is sterilized either using Ethylene Oxide Sterilization or Gamma irradiation method to achieve the Sterility Assurance Level (SAL) of 10th and place in a sterility maintenance package to ensure a shelf life of 3 years.

{6}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVES POWDERED & LATEX SURGEON'S GLOVES POWDER FREE (POLYMER COATED)

SECTION NO:27 PAGE NO:-27-3

F.INTENDED USE STATEMENT:

A latex surgeon's gloves is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

G. NON-CLINICAL TESTING

Bench tests were conducted to verify that the proposed device met all design specifications and complies with the following standards-

ASTM D3577-09:- Standard Specification for Rubber Surgical Gloves.

ASTM D5151-2011:- Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6124-2011:- Standard Test Method for Residual Powder on Medical Gloves.

ASTM D5712-10:- Standard Test Method for the Analysis of Aqueous Extractable Protein in Natural Rubber and Its Products Using the Modified Lowry Method.

ASTM D6499-12:- Standard Test Method for the Immunological Measurement of Antigenic Protein in Natural Rubber and Its Products.

ASTM F1929-2004:- Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration.

ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

1SO 1137-2: 2006. Sterilization of healthcare products-Radiation-Part2: Establishing the sterilization dose.

ISO 10993-7:2008:- Biological evaluation of medical devices -- Ethylene oxide sterilization residuals

ISO 11135-1:2007:- Sterilization of healthcare products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices

{7}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVES POWDERED & LATEX SURGEON'S GLOVES POWDER FREE (POLYMER COATED)

PAGE NO:-27-4 SECTION NO:27

H. PRODUCT COMPARISON IN COMPLIANCE WITH ASTM D 3577-09 STANDARD

SIZES AVAILABLE: - 5½, 6, 6½, 7, 7½, 8, 8½, 9

SL.NOCRITERIASPECIFICATION ASPER ASTM D3577-09STANDARDAVERAGE VALUE OFSUBJECT DEVICEWHETHER SUBJECTDEVICE COMPLIED WITHTHE ASTM D3577 -09STANDARD
1Length
Size 5½Min 265mm276 mmYes
Size 6Min 265mm278mmYes
Size 6½Min 265mm281mmYes
Size 7Min 265mm281mmYes
Size 7½Min 265mm281mmYes
Size 8Min 265mm283mmYes
Size 8½Min 265mm283mmYes
Size 9Min 265mm284mmYes
2Width
Size 5½70+/-6mm74mmYes
Size 676+/-6mm79mmYes
Size 6½83+/-6mm87mmYes
Size 789+/-6mm93mmYes
Size 7½95+/-6mm98mmYes
Size 8102+/-6mm106mmYes
Size 8½108+/-6mm112mmYes
Size 9114+/-6mm118mmYes
3Finger Thickness(All sizes)Min 0.10mm0.18mmYes
4Palm Thickness(All sizes)Min 0.10mm0.16mmYes
5Cuff Thickness(All sizes)Min 0.10mm0.12mmYes
SL.NOCRITERIASPECIFICATION ASPER ASTM D3577-09STANDARDAVERAGE VALUE OFSUBJECT DEVICEWHETHER SUBJECTDEVICE COMPLIED WITHTHE ASTM D3577 -09STANDARD
6Tensile Strength
Before aging(All sizes)24Mpa minimum28.5MpaYes
After aging@70°±2C for166±2 hr(All sizes)18Mpa minimum24.2MpaYes
7Ultimate Elongation
Before aging(All sizes)750% minimum934%Yes
After aging@70°±2C for166±2 hr(All sizes)560% minimum758%Yes
8Stress at 500%before aging(All sizes)5.5 MPa Max3 MpaYes
9Pinhole AQL
Before aging(All sizes)Max 1.51.0Yes
After aging@70°C for 7 days(All sizes)Max 1.51.0Yes

{8}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVES POWDERED & LATEX SURGEON'S GLOVES POWDER FREE (POLYMER COATED)

SECTION NO:27 PAGE NO:-27-5

SECTION NO:27PAGE NO:-27-6
------------------------------

{9}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVES POWDERED & LATEX SURGEON'S GLOVES POWDER FREE (POLYMER COATED)

I. SUBSTANTIALLY EQUIVALENT COMPARISON

General Characteristics Comparison a.

CharacteristicSubject devicePredicate deviceK130450Predicate deviceK130301SubstantiallyEquivalent (SE)or Not (NSE)
Product CodeKGOKGOKGOSE
Regulation No.21 CFR 878.446021 CFR 878.446021 CFR 878.4460SE
Class111SE
Intended Use forLatex Surgeon'sGloves PowderFree, Polymercoated withprotein contentlabeling claim of50 µg/dm² or lessper glove ofextractableprotein.A latex surgeon's gloves is a devicemade of natural rubber intended tobe worn by operating roompersonnel to protect a surgicalwound from contamination.This Surgeonsglove is a devicemade of naturalrubber intendedto be worn byoperating roompersonnel toprotect a surgicalwound fromcontamination.-Substantiallyequivalent
Intended UseLatex Surgeon'sGloves PowderedA latex surgeon's gloves is a devicemade of natural rubber intended tobe worn by operating roompersonnel to protect a surgicalwound from contamination.This Surgeonsglove is a devicemade of naturalrubber intendedto be worn byoperating roompersonnel toprotect a surgicalwound from contamination.A latex surgeonsglove is a devicemade of naturalrubber intendedto be worn byoperating roompersonnel toprotect a surgicalwound from contaminationSubstantiallyequivalent
Powdered orPowder freePowdered, and powered freePowdered, andpowered freePowdered, andpowered freeSE
Classification perASTMD3577-99Type I - gloves compoundedprimarily from natural rubber latexType I - glovescompoundedprimarily fromnatural rubberlatexType I - glovescompoundedprimarily fromnatural rubberlatexSE
SterilizationETO SAL- 10-6 ORRadiation SAL- 10-6Radiation SAL:10-6Radiation SAL:10-6SE
Label andLabelingMeet FDA's RequirementsMeet FDA'sRequirementsMeet FDA'sRequirementsSE
Special labelclaimProtein content labeling claim of 50µg/dm²or less per glove ofextractable protein for LatexSurgeon's Gloves Powder Free,Polymer coatedSame-SE
Type of useOver the counter useOver the counteruseOver the counteruseSE

{10}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVES POWDERED & LATEX SURGEON'S GLOVES POWDER FREE (POLYMER COATED)

PAGE NO:-27-7 SECTION NO:27

b. Technological Characteristics Comparison

CharacteristicsAcceptance criteria of the standardWhether the subjectdevice met theacceptance criteria ofthe standard
Subject devicePredicate deviceK130450Predicate deviceK130301
DimensionsASTM D3577-09(published date02/01/2009)ASTM D3577-09ASTM D3577-09Yes
PhysicalPropertiesASTM D3577-09(published date02/01/2009)ASTM D3577-09ASTM D3577-09Yes
Freedom fromHolesASTM D3577 andASTM D5151,(publisheddate 08/20/2012)ASTM D3577 andASTM D5151ASTM D3577 andASTM D5151Yes
Powder Contentfor powderedgloveASTM D3577 and ASTMD6124,(published date08/20/2012) (Max 15 mg/dm² )Our glove has 10mg/dm²on average.ASTM D3577 andASTM D6124ASTM D3577 andASTM D6124Yes
Powder Contentfor powder freegloveASTM D3577 and ASTMD6124,(published date08/20/2012)Powder content < 2mg/GloveASTM D3577 andASTM D6124,Powder content < 2mg/GloveASTM D3577 andASTM D6124,Powder content < 2mg/GloveYes
Protein ContentASTM D3577, ASTMD5712-10 and *ASTM D6499(published date02/01/2009)Our glove has ProteinContent less than 50µg/dm²ASTM D3577,ASTM D5712-10 andASTM D6499ASTM D3577,ASTM D5712-10and ASTM D6499Yes
BiocompatibilitySkin sensitizationSkin irritationISO 10993-10Non- sensitizer under theconditions of the studies.Non -irritant under theconditions of the studies.ISO 10993-10ISO 10993-10Yes,Non-Sensitizer andNon- irritant

*Conformance with ASTM D 6499 was an extra step to corroborate the claim 50 µg/dm² or less / glove

{11}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVES POWDERED & LATEX SURGEON'S GLOVES POWDER FREE (POLYMER COATED)

SECTION NO:27 PAGE NO:-27-8

Substantial Equivalence Discussion

The difference with predicate device is in the sterilization method. The proposed device is sterilized by ETO or Radiation method. Both give a final SAL of 106 which meets the acceptance level for sterilization.

Clinical data was not required for this submission.

Substantial Equivalence Conclusions

The conclusion drawn from the non-clinical tests demonstrate that the devices Medismart (Latex Surgeon's Gloves Powdered) and Medismart + (Latex Surgeon's Gloves Powder Free, Polymer coated with protein content labeling claim of 50 µg/dm² or less per glove of extractable protein), are as safe and as effective and perform as well as the legally marketed predicate devices SURGTEX Latex Surgeon's Gloves powdered and SURGTEX Latex Surgeon's Gloves powderfree-polymer coated with protein content labeling claim of 50 µg/dm² or less per glove of extractable protein (K130450) and Latex Surgeon's Gloves (powdered and Powder free (K130301); therefore, we conclude that the subject device is substantially equivalent to the predicate devices.

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).