K Number
K192249

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-12-17

(120 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The InMode System with Tone Applicator is used in EMS mode for:
Prevention or retardation of disuse atrophy
Maintaining or increasing range of motion
Muscle re-education
Relaxation of muscle spasms
Increasing local blood circulation
Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis

And in TENS mode for: Symptomatic relief and management of chronic, intractable pain Post-surgical acute pain Post-traumatic acute pain

Device Description

The InMode System in combination with Tone Applicator (manufactured by InMode Ltd.), is a versatile device intended to employ EMS (Electrical Muscle Stimulation) and TENS (Transcutaneous Electrical Nerve Stimulation) technologies for various medical applications. The InMode System with Tone Applicator consists of an AC/DC power supply unit, controller and user interface including an LCD touch screen. The delivery of the electrical energy is controlled by a Start/Stop button positioned on the front panel.

AI/ML Overview

This document describes the InMode System with Tone Applicator, a powered muscle stimulator. The information provided is primarily focused on demonstrating substantial equivalence to a predicate device (Vectra Neo Clinical Therapy System, K132284), rather than presenting a standalone study with acceptance criteria for device performance in the clinical sense (e.g., sensitivity, specificity for a diagnostic device).

Therefore, this response will reframe the "acceptance criteria" as the performance standards and safety criteria the device was tested against for regulatory clearance. The "study that proves the device meets the acceptance criteria" refers to the non-clinical (bench) performance data and compliance with recognized standards.


1. A table of acceptance criteria and the reported device performance

Since this is a submission for a powered muscle stimulator, the "acceptance criteria" revolve around demonstrating safety, electrical performance, and biocompatibility, as well as functional equivalence to the predicate device. The "reported device performance" are the results of the non-clinical tests and conformance to standards.

Acceptance Criteria (Performance/Safety Standard)Reported Device Performance
Electrical Safety & Essential Performance
IEC 60601-1 (General Requirements for Basic Safety and Essential Performance)Device complies with the standard. Specific safety parameters: Patient Leakage Current - Normal Condition (<100uA), Single Fault Condition (<300uA). Method of Line Current Isolation: Independent transformer isolated. Electrical Type: Type BF. Automatic Overload Trip, No-Load Trip, Shut Off, and Patient Override Control: Yes.
IEC 60601-1-2 (Electromagnetic Compatibility - Requirements and Tests)Device complies with the standard.
IEC 60601-2-10 (Particular Requirements for Nerve and Muscle Stimulators)Device complies with the standard.
Usability
IEC 60601-1-6 (Usability)Device complies with the standard.
Biocompatibility of User-Contacting Materials
ISO 10993-1 (Biological evaluation of medical devices)Cytotoxicity Study: Test article extract showed no evidence of causing cell lysis or toxicity (Conclusion: Non-toxic).ISO Acute Systemic Toxicity Study: Test article extracts showed no evidence of causing delayed dermal contact sensitization in the guinea pig (Conclusion: Non-toxic).ISO Guinea Pig Maximization Sensitization Test: No mortality or evidence of systemic toxicity (Conclusion: No irritation/sensitization).
Output Waveform ValidationBench testing demonstrated validation of output waveform.
Basic Unit Characteristics ValidationBench testing demonstrated validation of basic unit characteristics.
Output Specifications ValidationBench testing demonstrated validation of output specifications. See detailed specifications in the document for EMS and TENS modes (e.g., Waveform: Symmetrical Biphasic, Pulse Shape: Rectangular, Max Output Voltage/Current, Pulse Width, Frequency, Net Charge, Max Phase Charge, Max Current/Power Density, Burst Mode parameters, On/Off time, Treatment Time). All conform to design requirements.
Mechanical Safety TestingTesting performed according to IEC 60601-1 standard.
Software Validation TestingTesting performed.
Equivalence to Predicate Device (K132284)The subject device and predicate device utilize the same technology, for the same indication for use, and with similar design specifications. Performance specifications are in range or equal to those of the predicate device. Minor technical differences do not alter safety/effectiveness.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size:
    • For biocompatibility testing, the sample sizes are not explicitly stated in the provided text but are typically small animal groups (e.g., mice, guinea pigs) as per ISO guidelines.
    • For bench testing, the "sample size" refers to the device itself being tested, which is typically one or a few units to ensure manufacturing consistency.
  • Data Provenance: The data is from non-clinical (bench) performance testing and biocompatibility assessments conducted by the manufacturer, InMode Ltd., based in Israel. There is no indication of country of origin for the data other than the manufacturer's location. The studies are prospective in the sense that they are specifically performed for this regulatory submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable as the provided document describes a submission for a powered muscle stimulator based on non-clinical (bench) testing and substantial equivalence. There is no clinical study involving human or image data requiring expert interpretation or ground truth establishment for the device's performance characteristics. The "ground truth" for the technical and safety performance tests are the established standards (e.g., IEC, ISO) and the device's design specifications.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable. The document does not describe a clinical study or a test set requiring adjudication by experts.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. The device is a powered muscle stimulator, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted or is relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This section is not applicable. The device is a physical medical device (muscle stimulator), not an algorithm or AI system. Its performance evaluation is based on its physical and electrical output characteristics and safety, not on algorithmic standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the device's performance is established by:

  • Compliance with recognized consensus standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-10, and ISO 10993-1. These standards define acceptable performance and safety limits.
  • Device design requirements: The device's output specifications (e.g., voltage, current, frequency, pulse width). Bench tests verify that the device's actual output matches these specified requirements.

8. The sample size for the training set

This section is not applicable. As a physical medical device, the InMode System with Tone Applicator does not utilize machine learning or AI, and therefore, no "training set" of data is used in its development or validation.

9. How the ground truth for the training set was established

This section is not applicable for the same reasons as #8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 17, 2019

InMode Ltd. % Amit Goren Regulatory Manager A. Stein - Regulatory Affairs Consulting Ltd. 20 Hata'as Str., Suite 102 Kfar Saba, 4442520 Israel

Re: K192249

Trade/Device Name: InMode System with Tone Applicator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, GZJ Dated: September 16, 2019 Received: September 18, 2019

Dear Amit Goren:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Vivek Pinto, Ph.D. Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192249

Device Name InMode System with Tone Applicator

Indications for Use (Describe)

The InMode System with Tone Applicator is used in EMS mode for:

Prevention or retardation of disuse atrophy

Maintaining or increasing range of motion

Muscle re-education

Relaxation of muscle spasms

Increasing local blood circulation

Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis

And in TENS mode for: Symptomatic relief and management of chronic, intractable pain Post-surgical acute pain Post-traumatic acute pain

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

THE INMODE SYSTEM WITH TONE APPLICATOR

510(k) Number K192249

Applicant Name:

Company Name:InMode Ltd.
Address:Tabor Building, Shaar YokneamYokneam 20692IsraelTel: +972-4-9097470Fax: +972-4-9097471E-mail: amit@asteinrac.com
Contact Person:
Official Correspondent:Amit Goren
Company Name:A. Stein - Regulatory Affairs Consulting Lt
Address:20 Hata'as Str., Suite 102Kfar Saba 4442520 IsraelTel: +972-9-7670002Fax: +972-9-7668534E-mail: amit@asteinrac.com
Date Prepared:December 11, 2019
Trade Name:The InMode System with Tone Applicator
Classification Name:Powered Muscle Stimulator
Common Name:Powered Muscle Stimulator
Regulation Number:890.5850 & 882.5890
Product Code:IPF & GZJ
Classification:Class II Medical Device

Predicate Device:

The InMode System with Tone Applicator is substantially equivalent to the following predicate device.

PredicateManufacturer510(k) No.
Vectra Neo Clinical Therapy SystemDJO, LLCK132284

{4}------------------------------------------------

Device Description:

The InMode System in combination with Tone Applicator (manufactured by InMode Ltd.), is a versatile device intended to employ EMS (Electrical Muscle Stimulation) and TENS (Transcutaneous Electrical Nerve Stimulation) technologies for various medical applications. The InMode System with Tone Applicator consists of an AC/DC power supply unit, controller and user interface including an LCD touch screen. The delivery of the electrical energy is controlled by a Start/Stop button positioned on the front panel.

The System support the following components:

  • . LCD display touch screen
  • . Audio loudspeaker
  • 48V AC/DC power supply .
  • . Controller
  • . Fans

The System operates while connected to the Tone Applicator.

Following are The InMode System with Tone Applicator specifications:

Main Line Frequency (nominal):50-60 Hz
Input Voltage (nominal):100-240 VAC
Input Current (rms)2A
Dimension:
Console35cm W x 35cm D x 100cm H
[W x H x D][18.2" W x 18.2" D x 40" H]
ApplicatorTone Applicator
[L x D]12cm L x 10cm D [4.7" L x 4" D]
Weight Console:20 Kg (44 lbs)
Tone Applicator Weight:0.22 Kg [0.5 lbs.]
Platform modulesConverts AC input voltage (90-264Vac) to
AC/DC power supply48Vdc, 300W
WaveformSymmetrical Biphasic
ShapeRectangular
Intensity (output Voltage)Up to 50 intensity level (=54 Vpeak)
Pulse Width
Tone Applicator20 to 400 $ \mu $ S
Frequency
Tone Applicator3 to 200 Hz

{5}------------------------------------------------

Intended Use/Indication for Use:

The InMode System with Tone Applicator is used in EMS mode for:

  • 트 Prevention or retardation of disuse atrophy
  • Maintaining or increasing range of motion ■
  • Muscle re-education ■
  • Relaxation of muscle spasms
  • Increasing local blood circulation
  • Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis

And in TENS mode for:

  • Symptomatic relief and management of chronic, intractable pain 트
  • 트 Post-surgical acute pain
  • Post-traumatic acute pain

{6}------------------------------------------------

Performance Standards:

The InMode System with Tone Applicator has been tested and complies with the following FDA recognized consensus standards:

  • · IEC 60601-1: Medical Electrical Equipment Part 1: General Requirements For Basic Safety And Essential Performance ES60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005 Edition 3rd, MOD)
  • · Appendix 6-Clause 3.2.2.1 and Clause 3.2.2.2 from ANSI/AAMI NS4:2013/(R)2017
  • · IEC 60601-1-2: Medical electrical equipment; Part 1-2: Collateral Standard: Electromagnetic compatibility - Requirements and tests, Edition 4.0 (2014). Environment of intended uses: Professional Healthcare Facility Environment
  • · IEC 60601-2-10 Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators (IEC 60601-2-10 Edition 2.1 2016-04) IEC 60601-2-10: 2012, AMD1:2016 for use in conjunction with IEC 60601-1:2005/AMD1:2016
  • · IEC 60601-1-6 Medical Electrical Equipment Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability (60601-1-6 Edition 3.1 2013-10, AMD1:2013)

Non-Clinical (Bench) Performance Data:

Bench testing was conducted to demonstrate that the InMode System with Tone Applicator performs as expected under anticipated conditions of use and to verify that the device performance meets the device design requirements. The device was tested for validation of output waveform, basic unit characteristics, and output specifications.

The bench testing results demonstrated that the device performs as expected under anticipated conditions of use

The biocompatibility of the Tone Applicator outer components (handle and electrodes)

{7}------------------------------------------------

was justified using a biocompatibility assessment performed on a family of related products and by scientific evidence. The following biocompatibility tests were performed as part of the biocompatibility assessment:

TestTest SummaryConclusions
Cytotoxicity Study Using the ISOElution MethodThe test article extractshowed no evidence ofcausing cell lysis ortoxicityNon-toxic
ISO Acute Systemic ToxicityStudy in MiceThe test article extractsshowed no evidence ofcausing delayed dermalcontact sensitization in theguineapig.Non-toxic
ISO Guinea Pig MaximizationSensitization TestThere was no mortality orevidence of systemictoxicity from the extractsinjected into miceNo irritation/sensitization

Pre-Clinical (Animal) Performance Data:

Non-Applicable.

Clinical Performance Data:

Non-Applicable.

Substantial Equivalence:

The below table summarizes the main comparison aspects between the InMode System with Tone Applicators and the proposed predicate device.

CharacteristicSubject DevicePredicate
510(k) file No.K192249K132284
Device NameInMode System withTone ApplicatorVectra NeoClinical TherapySystem
ManufacturerInMode Ltd.DJO, LLC
Class, Product Code.Class IIIPF GZJClass IIIPF, GZJIMG, HCC, GZI
CharacteristicSubject DevicePredicate
510(k) file No.K192249K132284
Device NameInMode System withTone ApplicatorVectra NeoClinical TherapySystem
ManufacturerInMode Ltd.DJO, LLCLIH, ILY
Design:The InMode System withTone Applicator consistsof an AC/DC powersupply unit, controller anduser interface including anLCD touch screen. Thedelivery of the electricalenergy is controlled by aStart/Stop buttonpositioned on the frontpanel.The System support thefollowing components:• LCD display touchscreen• Audio loudspeaker• 48V AC/DC powersupply• Controller• FansThe System operateswhile connected to theTone Applicator.The Vectra NeoVlinical TherapySystem consists ofan AC/DC powersupply unit,controller and userinterface includingcolor display. Thedelivery of theelectrical energy iscontrolled by aStart/Stop buttonpositioned on thefront panel.
Mechanism of ActionMuscle contraction byelectrical pulsing.Idem
ComponentsConsoleThe InMode Systemconsists of the followingcomponents:• Console, including apower supply unit,controller and userinterface including anLCD touch screen.• Tone Applicatorconnected to theconsole via a cable.Vectra Neo ClinicalTherapy Systemconsists of thefollowingcomponents:• Console,including a powersupply unit,controller anduser interfaceincluding an LCDtouch screen.• Several types ofapplicators fordifferent
CharacteristicSubject DevicePredicate
510(k) file No.K192249K132284
Device NameInMode System withTone ApplicatorVectra NeoClinical TherapySystem
ManufacturerInMode Ltd.DJO, LLC
indications (EMS,Ultrasound, EMG& Laser)connected to theconsole via acable.
DimensionConsole[W x H x D]35cm W x 35cm Dx 100cm H[18.2" W x 18.2" D x40" H]not publiclyavailable
Applicator[L x D]Tone Applicator12cm L x 10cm D[4.7" L x 4" D]
Weight Console20.0 Kg [44 lbs.]not publiclyavailable
Weight applicatorTone: 0.22 Kg [0.5 lbs.]
PerformanceSpecifications:ComponentsConsoleMain Line Frequency(nominal)50-60HzInput Voltage (nominal)100-240VACInput Current (rms)2Anot publiclyavailable
Method of Line Current IsolationIndependent transformerisolatednot publiclyavailable
Electrical TypeType BFnot publiclyavailable
Patient Leakage Current - NormalCondition (μΑ)<100uA patient leakagenot publiclyavailable
Patient Leakage Current - SingleFault Condition (μΑ)<300uA line leakagenot publiclyavailable
Number of Output Modes2not publiclyavailable
Number of Output Channels2not publiclyavailable
CharacteristicSubject DevicePredicate
510(k) file No.Device NameK192249InMode System withTone ApplicatorK132284Vectra NeoClinical TherapySystem
ManufacturerInMode Ltd.DJO, LLC
Synchronous or AlternatingSee Output SpecificationsBelownot publiclyavailable
Method of Channel IsolationThrough transformers andisolatorsnot publiclyavailable
Regulated Current or RegulatedVoltage (output signals only)Regulated voltage on allchannelsWith current limitnot publiclyavailable
Software/Firmware/MicroprocessorControlYesnot publiclyavailable
Automatic Overload TripYesnot publiclyavailable
Automatic No-Load TripYesnot publiclyavailable
Automatic Shut OffYes, On/off switchnot publiclyavailable
Patient Override ControlYesnot publiclyavailable
Indicator DisplayYesnot publiclyavailable
On/Off StatusYesnot publiclyavailable
BatteryNo batterynot publiclyavailable
Voltage/Current levelYes, voltage levelsnot publiclyavailable
Timer Range (Minutes)0-60 [minutes]not publiclyavailable
Compliance with 21 CFR 890.5850(IPF)Yesnot publiclyavailable
Compliance with 21 CFR 882.5890(GZJ)Yesnot publiclyavailable
Applicator NameTone ApplicatorVectra Neo ClinicalTherapy SystemApplicator
CharacteristicSubject DevicePredicate
510(k) file No.K192249K132284
Device NameInMode System withTone ApplicatorVectra NeoClinical TherapySystem
ManufacturerInMode Ltd.DJO, LLC
EMS Output Mode
Output Specifications:WaveformSymmetrical BiphasicWaveformnot publiclyavailable
Pulse ShapeRectangularnot publiclyavailable
Maximum Output Voltage (± 10%)56V @500Ωnot publiclyavailable
56V @2 kΩnot publiclyavailable
56V @10kΩnot publiclyavailable
Maximum Output Current (± 10%)112 mA @ 500 Ωnot publiclyavailable
28 mA @ 2 kΩnot publiclyavailable
5.6 mA @ 10 kΩnot publiclyavailable
Pulse Width (μsec.) - The outputactive positive pulse width20 to 400 [µs]not publiclyavailable
Frequency (Hz)3 to 200 [Hz]not publiclyavailable
Net Charge @ 500 ohms[µC/pulse]0 [μC] @ 500Ωnot publiclyavailable
Maximum Phase Charge [µC]44.8 [µC] @ 500Ωnot publiclyavailable
Maximum Current Density[mA/cm2]1 [mA/cm2]Surface = 12cm²not publiclyavailable
Maximum Power Density[mW/cm2]55[mW/cm2] @500Ωnot publiclyavailable
Burst Mode (i.e., pulse trains)Yes:not publiclyavailable
a. Pulses per bursta. 3 - 200
b. Bursts per secondb. 1
c. Burst duration (seconds)c. 1-60 sec
d. Duty Cycle [Line (b) x Line (c)]d. Time on / off
On Time (sec.)1 - 60 [sec]not publiclyavailable
Off Time (sec.)1 - 60 [sec]not publiclyavailable
Treatment Time (min) - the timelimit that will put the system inSTOP stateUp to 60 min.not publiclyavailable
CharacteristicSubject DevicePredicate
510(k) file No.K192249K132284
Device NameInMode System withTone ApplicatorVectra NeoClinical TherapySystem
ManufacturerInMode Ltd.DJO, LLC
Level - The output intensity level1 to 50 (5-54v)
TENS Output Mode
Output Specifications:WaveformSymmetrical BiphasicWaveformnot publiclyavailable
Pulse ShapeRectangularnot publiclyavailable
Maximum Output Voltage (± 10%)36V @500Ωnot publiclyavailable
36V @2 kΩnot publiclyavailable
36V @10kΩnot publiclyavailable
Maximum Output Current (± 10%)72 mA @ 500 Ωnot publiclyavailable
18 mA @ 2 kΩnot publiclyavailable
3.6 mA @ 10 kΩnot publiclyavailable
Pulse Width (μsec.) - The outputactive positive pulse width20 to 400 [μς]not publiclyavailable
Frequency (Hz)3 to 200 [Hz]not publiclyavailable
Net Charge @ 500 ohms[µC/pulse]0 [μC] @ 500Ωnot publiclyavailable
Maximum Phase Charge [µC]28.8 [μC] @ 500Ωnot publiclyavailable
Maximum Current Density[mA/cm2]0.65 [mA/cm2]Surface = 12cm²not publiclyavailable
Maximum Power Density[mW/cm2]22.7 [mW/cm2] @500Ωnot publiclyavailable
Burst Mode (i.e., pulse trains)Yes:not publiclyavailable
a. Pulses per burst3 - 200
b. Bursts per second1
c. Burst duration (seconds)1-60 sec
d. Duty Cycle [Line (b) x Line (c)]Time on / off
On Time (sec.)1 - 60 [sec]not publiclyavailable
Off Time (sec.)1 - 60 [sec]not publiclyavailable
CharacteristicSubject DevicePredicate
510(k) file No.K192249K132284
Device NameInMode System withTone ApplicatorVectra NeoClinical TherapySystem
ManufacturerInMode Ltd.DJO, LLC
Treatment Time (min) - the timelimit that will put the system inSTOP stateUp to 60 min.not publiclyavailable
Level - The output intensity level1 to 50 (5-54v)not publiclyavailable
Electrode area12 cm2not publiclyavailable
Housing MaterialPC Makrolon 2458not publiclyavailable

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

The subject device and predicate device utilize the same technology, for the same indication for use, and with similar design specifications. The device emits electrical signals with similar power and current densities, pulse characteristics, and bear similar system components to its predicate device such as; user interface, and hardware components. All of the subject device performance specifications are in range or equal to those of its predicate device. The minor differences in technical specifications should not alter the device safety and effectiveness. Furthermore, the subject device had underwent the required performance testing and validation testing and demonstrates its conformance with device design requirements and with applicable standards. The safety features and compliance with safety standards of the subject device are similar to the safety features and compliance with safety standards of the predicate device. All user-contacting materials were tested for biocompatibility and found to comply with the ISO 10993-1 standard. Furthermore, the design and development phases of the subject device were validated throughout a set of performance tests, including software validation testing, electrical and mechanical safety testing according to IEC 60601-1 standard, electromagnetic compatibility testing according to IEC 60601-1-2 standard, safety and essential performance of nerve and muscle stimulators testing according to IEC 60601-2-10 standard, and bench performance tests. All in all, these performance tests demonstrated that the device specifications meet the system requirements and do not raise new safety or effectiveness concerns.

{14}------------------------------------------------

Consequently, it can be concluded that the InMode System with the Tone Applicator is substantially equivalent to its predicate device and can be sold in the US market.

Conclusions:

Based on the comparison to the predicate device and on the non-clinical performance testing results demonstrating that The InMode System with Tone Applicator is as safe and effective as the predicate device, it can be concluded that The InMode System with Tone Applicator is substantially equivalent to the predicate device; the Vectra Neo Clinical Therapy System cleared under 510(k) K132284, and therefore, may be legally marketed in the USA.

N/A