K Number
K191935

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-04-09

(265 days)

Product Code
Regulation Number
870.4100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nautilus(tm) Smart ECMO Module with integrated heat exchanger is intended to provide assisted long-term extracorporeal circulation and physiologic gas exchange (oxygenation and carbon dioxide removal) of the patient's blood for up to 48 hours in adult and pediatric adolescent patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The integrated heat exchanger is intended to heat or cool the blood as needed during use. Integrated fluid path pressure, temperature, and oxygen saturation monitoring is achieved by built-in sensor modules and display.

The Nautilus(tm) ECMO Oxygenator with integrated heat exchanger is intended to provide assisted long-term extracorporeal circulation and physiologic gas exchange (oxygenation and carbon dioxide removal) of the patient's blood for up to 48 hours in adult and pediatric adolescent patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The integrated heat exchanger is intended to heat or cool the blood as needed during use.

Device Description

The Nautilus™ ECMO oxygenators are diffusion membrane oxygenators used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Blood enters the device and passes through both the heat exchange membrane, where temperature is adjusted, and the gas transfer membrane, where oxygen is added and carbon dioxide is removed.

The Nautilus™ Smart ECMO Module device contains integrated sensors with an electronic touch screen display. The following measured parameters are visible on the electronic display: inlet pressure, inlet oxygen saturation, outlet pressure, outlet oxygen saturation, and outlet blood temperature. The difference between the inlet and outlet pressure, delta pressure, is calculated and also displayed on the screen. The touchscreen display allows users to set alarm limits for all measured parameters. The Nautilus™ Smart ECMO Module will alarm visually and audibly when the limits are exceeded.

The devices are single-use, nontoxic, non-pyrogenic, and not made from natural latex rubber materials.

AI/ML Overview

The provided text describes performance evaluations for the MC3 Nautilus Smart ECMO Module and Nautilus ECMO Oxygenator. However, it does not explicitly state "acceptance criteria" in a tabulated format and then directly link each criterion to "reported device performance" in a separate column. Instead, it presents various performance evaluations and their findings.

Based on the information provided, here's a structured summary attempting to address your request as closely as possible:

1. Table of acceptance criteria and the reported device performance

While explicit acceptance criteria are not tabulated with specific numerical targets, the document implies that the device is considered acceptable if its performance is "comparable or better" than a state-of-the-art device and meets the "special controls" and standards.

Acceptance Criteria (Implied)Reported Device Performance
Gas Exchange (Oxygen and Carbon Dioxide Transfer)Long-term Bench Gas Exchange testing (ISO 7199:2016): "Nautilus had comparable or better oxygen and carbon dioxide transfer rates... as a heparin coated state-of-the-art device, over 14-days of use."
Pressure DropLong-term Bench Gas Exchange testing (ISO 7199:2016): "...pressure drop performance as a heparin coated state-of-the-art device, over 14-days of use."
Clotting Prevention / ThrombogenicityIn-Vivo Study in Sheep (96 hours): "After 4 days, there were no clots in any location in the Nautilus device..." Clinical information (Real World Evidence): "There were no reports of circuit clotting..."
Integrity / Mechanical FailureNon-clinical Performance: "...mechanical integrity... durability, reliability..." demonstrated substantial equivalence. Clinical information (Real World Evidence): "There were no reports of... mechanical failure."
HemolysisPerformance Evaluations: "Hemolysis" was evaluated. Clinical information (Real World Evidence): "Fifteen percent of patients had reports of hemolysis complications, of which none was directly attributed to the Nautilus™. One hemolysis complication was reported to have occurred after 27 days of operation, another was reported prior to Nautilus™ use while operating on a different oxygenator, and one was attributed to patient condition (sepsis) and pump clotting." (This implies hemolysis, when it occurred, was generally not attributed to the device itself.)
Biocompatibility"The subject device is demonstrated to be biocompatible for prolonged use in circulating blood in accordance with ISO 10993-1:2009 and in accordance with GLP (21 CFR 58)."
Sterility and Shelf-life"Testing demonstrates the sterility of the subject device as provided and that it maintains its sterility, integrity, durability, and reliability over the stated shelf-life of the device."
Sensor Accuracy (for Smart ECMO Module)Smart Module Testing: "Sensor Accuracy" was evaluated. Reference Device (Sensor Module): "The sensor module functions are comparable to the Maquet CARDIOHELP System with regard to sensor accuracy..."
Electrical Safety/EMC (for Smart ECMO Module)Smart Module Testing: "Electrical Safety/EMC" was evaluated. Non-clinical Performance: "...electromagnetic compatibility..." demonstrated substantial equivalence.
Water Ingress Protection (for Smart ECMO Module)Smart Module Testing: "Water Ingress Protection" was evaluated.
Adverse Events / Complications (e.g., infections, CNS events)Clinical information (Real World Evidence): - "The most common complication reported was circuit exchange in twenty percent of patients..." - "Ten percent of patients had reports of blood infections on ECMO, which is similar to the rate of infections (11%) previously reported by others." - "One patient suffered a CNS infarction event (5%), and another patient, a CNS hemorrhage."
Compatibility with other circuit devicesTechnological Characteristics: "The subject device is designed to be compatible with other extracorporeal circuit devices and accessories."

The studies collectively show that the Nautilus device meets performance expectations by being comparable to or better than predicate/reference devices and established clinical norms, and by demonstrating safety and efficacy in various testing environments.


2. Sample size used for the test set and the data provenance

  • Long-term Bench Gas Exchange testing: The sample size isn't explicitly stated as a number of devices, but it refers to "MC3 Nautilus oxygenators from accelerated age populations" and "competitive heparin coated oxygenators."
    • Provenance: In vitro, conducted according to ISO 7199:2016, using bovine blood/plasma.
  • In-Vivo Study in Sheep:
    • Sample size: A total of 15 sheep. Specifically, 10 sheep were used to evaluate the Nautilus device (at either 2 or 5 L/min), and 5 control animals were studied with a state-of-the-art heparin-coated device (at 2 L/min).
    • Provenance: Prospective animal study.
  • Clinical information (Real World Evidence):
    • Sample size: Not explicitly stated as a number of patients, but data were collected from patients at 7 centers in 4 countries. The average duration of ECMO support was 378.5 hours (15.8 days), with a range of 17 to 1271 hours (52.9 days). Given these durations, it implies a significant number of patient-hours of use.
    • Provenance: Retrospective, voluntarily entered into the ELSO registry from initial real-world clinical experience. Data from 7 centers in 4 countries.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the text. The studies described are performance-based (bench, in-vivo) and real-world data collection, rather than studies requiring expert adjudication of device performance in a diagnostic sense.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the text. The studies described (bench, animal, real-world registry data) do not appear to involve a human adjudication process for establishing a "ground truth" in the way a diagnostic algorithm might be evaluated.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study, nor any evaluation of human readers improving with AI assistance. The device is an ECMO module/oxygenator, not an AI-assisted diagnostic tool.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not directly applicable. The device (ECMO module/oxygenator) is a medical device that physically interacts with a patient's blood, not an algorithm. The "Smart Module" has integrated sensors and an electronic display, which would operate in a "standalone" algorithmic sense for its sensor functions, but its primary function as an ECMO device is not an algorithm. The sensor accuracy was evaluated as part of "Smart Module Testing."


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Long-term Bench Gas Exchange testing: Ground truth for gas exchange and pressure drop was established by measurements according to the ISO 7199:2016 standard using bovine blood, providing objective scientific measurements.
  • In-Vivo Study in Sheep: Ground truth for clotting was established by direct observation ("no clots in any location") and confirmation of oxygenator functionality in the animal model.
  • Clinical information (Real World Evidence): Ground truth for complications and device performance in a clinical setting was derived from clinical observation and patient outcomes data voluntarily entered into the ELSO registry by clinicians at the participating centers.

8. The sample size for the training set

This information is not applicable/provided. The device is a physical medical device, not a machine learning algorithm that requires a "training set" for its operation in the way an AI diagnostic tool would.


9. How the ground truth for the training set was established

This information is not applicable/provided, as the concept of a "training set" does not apply to this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box. Underneath that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 9, 2020

MC3 Incorporated Martha Rumford Director RA/QA 2555 Bishop Circle W Dexter, Michigan 48130

Re: K191935

Trade/Device Name: Nautilus Smart ECMO Module and Nautilus ECMO Oxygenator Regulation Number: 21 CFR 870.4100 Regulation Name: Extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary failure Regulatory Class: Class II Product Code: BYS Dated: March 10, 2020 Received: March 11, 2020

Dear Martha Rumford:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Nicole Ibrahim, Ph.D. Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191935

Device Name MC3 Nautilus(tm) Smart ECMO Module

Indications for Use (Describe)

The Nautilus(tm) Smart ECMO Module with integrated heat exchanger is intended to provide assisted long-term extracorporeal circulation and physiologic gas exchange (oxygenation and carbon dioxide removal) of the patient's blood for up to 48 hours in adult and pediatric adolescent patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The integrated heat exchanger is intended to heat or cool the blood as needed during use. Integrated fluid path pressure, temperature, and oxygen saturation monitoring is achieved by built-in sensor modules and display.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K191935

Device Name MC3 Nautilus(tm) ECMO Oxygenator

Indications for Use (Describe)

The Nautilus(tm) ECMO Oxygenator with integrated heat exchanger is intended to provide assisted long-term extracorporeal circulation and physiologic gas exchange (oxygenation and carbon dioxide removal) of the patient's blood for up to 48 hours in adult and pediatric adolescent patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The integrated heat exchanger is intended to heat or cool the blood as needed during use.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for MC3 Cardiopulmonary. The logo consists of a stylized symbol on the left, resembling three curved shapes arranged in a circular pattern. To the right of the symbol, the letters "MC3" are displayed in a bold, sans-serif font. Below "MC3", the word "Cardiopulmonary" is written in a smaller, sans-serif font.

510(k) Summary

Sponsor Information:
------------------------
Owner/Applicant/Submitter:MC3 Incorporated2555 Bishop Circle WestDexter, MI 481301-734-995-9089Registration number: 3011468686
Contact Person:Martha RumfordDirector of RA/QA
Date Prepared:4/5/2020

Device Names/Classification:

Device Trade Name:Nautilus™ Smart ECMO ModuleNautilus™ ECMO Oxygenator
Device Common Name:Oxygenator, Long Term Support Greater Than 6 hours
Regulation Name:Extracorporeal circuit and accessories for long-termrespiratory/cardiopulmonary failure
Regulation Number:21 CFR 870.4100
Product Code:BYS
Predicate:81 FR 7451, Feb. 12, 2016
Reference Devices:Maquet Quadrox iD Oxygenator (K150267)CARDIOHELP System, Model CARDIOHELP-i(K133598)

Indications for Use:

The Nautilus™ ECMO Oxygenator with integrated heat exchanger is intended to provide assisted longterm extracorporeal circulation and physiologic gas exchange (oxygenation and carbon dioxide removal) of the patient's blood for up to 48 hours in adult and pediatric adolescent patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The integrated heat exchanger is intended to heat or cool the blood as needed during use.

The Nautilus™ Smart ECMO Module with integrated heat exchanger is intended to provide assisted longterm extracorporeal circulation and physiologic gas exchange (oxygenation and carbon dioxide removal) of the patient's blood for up to 48 hours in adult and pediatric adolescent patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The integrated heat exchanger is intended to heat or cool the blood as needed during use. Integrated fluid path pressure, temperature, and oxygen saturation monitoring is achieved by built-in sensor modules and display.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for MC3 Cardiopulmonary. The logo consists of a stylized symbol on the left, resembling three curved shapes arranged in a circular pattern. To the right of the symbol are the letters "MC3" in a bold, sans-serif font. Below "MC3" is the word "Cardiopulmonary" in a smaller, lighter font.

Device Description:

The Nautilus™ ECMO oxygenators are diffusion membrane oxygenators used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Blood enters the device and passes through both the heat exchange membrane, where temperature is adjusted, and the gas transfer membrane, where oxygen is added and carbon dioxide is removed.

The Nautilus™ Smart ECMO Module device contains integrated sensors with an electronic touch screen display. The following measured parameters are visible on the electronic display: inlet pressure, inlet oxygen saturation, outlet pressure, outlet oxygen saturation, and outlet blood temperature. The difference between the inlet and outlet pressure, delta pressure, is calculated and also displayed on the screen. The touchscreen display allows users to set alarm limits for all measured parameters. The Nautilus™ Smart ECMO Module will alarm visually and audibly when the limits are exceeded.

The devices are single-use, nontoxic, non-pyrogenic, and not made from natural latex rubber materials.

Performance Evaluations:

Evaluation and testing was executed to demonstrate the substantial equivalence of the subject devices. Performance assessments for substantial equivalence were accomplished through bench and in vivo studies that included the following evaluations:

  • Gas exchange
  • Heat exchange ●
  • Prime volume
  • . Coating coverage
  • . Low-flow clotting
  • . Hemolysis
  • Integrity
  • Operating Conditions
  • In-vivo study

Smart Module Testing also included:

  • Electrical Safety/EMC
  • . Sensor Accuracy
  • . Water Ingress Protection

Substantial Equivalence:

Substantial equivalence analysis includes both comparison to clinically-relevant Reference Devices and comparison to the Special Controls of FDA's final order, 81 FR 7451, Feb. 12, 2016 (Predicate). This is due to the fact that there has been no oxygenator device cleared under regulation 21 CFR 870.4100 to date, although ECMO has been well established clinically using currently marketed, state-of-the-art oxygenators.

Reference Device (Oxygenator): The subject oxygenator is comparable to the Maquet Quadrox iD Oxygenator in design, principles of operation, materials of construction, performance, and fundamental scientific technology.

Reference Device (Sensor Module): The sensor module functions are comparable to the Maquet CARDIOHELP System with regard to sensor accuracy and standards compliance.

Predicate: The Nautilus™ ECMO oxygenators meet all special controls required by 21 CFR 870.4100.

Special Controls met are:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for MC3 Cardiopulmonary. The logo consists of a stylized symbol on the left, resembling three interconnected crescent shapes. To the right of the symbol is the text "MC3" in bold, sans-serif font. Below "MC3" is the word "Cardiopulmonary" in a smaller font size.

  • . Technological Characteristics: Geometry and design parameters are consistent with the device's intended use in extracorporeal life support procedures. The subject device is designed to be compatible with other extracorporeal circuit devices and accessories.
  • . Biocompatibility: The subject device is demonstrated to be biocompatible for prolonged use in circulating blood in accordance with ISO 10993-1:2009 and in accordance with GLP (21 CFR 58).
  • Sterility and Shelf-life: Testing demonstrates the sterility of the subject device as provided and that it maintains its sterility, integrity, durability, and reliability over the stated shelf-life of the device.
  • . Non-clinical Performance: Substantial equivalence is demonstrated by performance characteristics on the bench, mechanical integrity, electromagnetic compatibility, software, durability, reliability, and accuracy.
  • . In vivo Evaluation: In vivo evaluation demonstrates the subject device's performance over a long-term duration of use in a biologic test system. In addition to in-vivo study, a summary was prepared that described the initial real-world clinical experience of the device with the first records of clinical ECMO cases entered sequentially into the ELSO registry.
  • . Labeling: The Instructions for Use includes a detailed summary of the non-clinical evaluations pertinent to the device's use in an extracorporeal circuit. Adequate instructions are included with respect to installation, circuit setup, maintenance during a procedure, changeout, adverse effects, and performance characteristics relevant to compatibility among different devices and accessories in the circuit.

ECMO Performance Evidence

● Long-term Bench Gas Exchange testing

Gas transfer testing of the MC3 Nautilus oxygenator device and competitive heparin coated oxygenators was executed for the comparative evaluation of oxygen and carbon dioxide transfer, and blood path pressure drop performance over an extended duration of use. Testing was performed in accordance with the ISO standard for blood-gas exchangers (ISO 7199:2016). using bovine blood at determined time points over a full range of operating variables, simulating a use duration of 14 days by circulating bovine plasma between test points . MC3 Nautilus oxygenators from accelerated age populations were tested. Nautilus had comparable or better oxygen and carbon dioxide transfer rates and pressure drop performance as a heparin coated state-of-the art device, over 14-days of use.

In-Vivo Study in Sheep

A 96 hour animal study was conducted to evaluate the safety and performance of Nautilus during extracorporeal membrane oxygenation (ECMO) for long term use. Target ACT during this study was 180-220 sec. The study was conducted on a total of 15 sheep. Ten of the 15 sheep were used specifically to evaluate the Nautilus at either 2 or 5 1pm, and 5 control animals were studied at 2 1/min with a state of the art heparin coated device. After 4 days, there were no clots in any location in the Nautilus device and oxygenator functionality was confirmed.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for MC3 Cardiopulmonary. The logo consists of a stylized clover-like symbol on the left, followed by the text "MC3" in bold, sans-serif font. Below "MC3" is the word "Cardiopulmonary" in a smaller, sans-serif font. The logo is simple and modern, with a focus on the company's name and area of expertise.

Clinical information (Real World Evidence) .

The purpose of the Real World Evidence (RWE) study was to collect and summarize the initial real world performance of the Nautilus™ when used as part of an ECMO/ECLS system in patients. Data were voluntarily entered into the ELSO registry.

Data were collected on patients at 7 centers in 4 countries. The cases reported demonstrate experience with the Nautilus™ in a diverse set of ECMO conditions, including multiple ECMO centers, venovenous and venoarterial support, and a typical variety of primary indications for ECMO support.

The average duration of ECMO support was 378.5 hours (15.8 days), with a range of 17 to 1271 hours (52.9 days). The average Nautilus™ use duration is comparable to the typical length of ECMO support reported in the literature.

The most common complication reported was circuit exchange in twenty percent of patients, of which one exchange was attributed to diminished performance after 12 days; one exchange was due to hemolysis not attributed to the Nautilus™; and two exchanges were electively exchanged.

Fifteen percent of patients had reports of hemolysis complications, of which none was directly attributed to the Nautilus™. One hemolysis complication was reported to have occurred after 27 days of operation, another was reported prior to Nautilus™ use while operating on a different oxygenator, and one was attributed to patient condition (sepsis) and pump clotting.

Ten percent of patients had reports of blood infections on ECMO, which is similar to the rate of infections (11%) previously reported by others.1

One patient suffered a CNS infarction event (5%), and another patient, a CNS hemorrhage. There were no reports of circuit clotting, air in the circuit, or mechanical failure.

Conclusion:

MC3 Incorporated concludes the Nautilus™ Oxygenators are substantially equivalent to the Predicate by the meeting the special controls (81 FR 7451, Feb 12, 2016) and have demonstrated to have performed comparably to clinical Reference devices with respect to intended use, design, materials of construction, principles of operation, performance, and specifications.

1Daniel Brodie, et al "Extracorporeal Life Support for Adults With Respiratory Failure And Related Indications A Review" JAMA August 13, 2019.

§ 870.4100 Extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary failure.

(a)
Identification. An extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary support (>6 hours) is a system of devices and accessories that provides assisted extracorporeal circulation and physiologic gas exchange of the patient's blood in patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The main devices and accessories of the system include, but are not limited to, the console (hardware), software, and disposables, including, but not limited to, an oxygenator, blood pump, heat exchanger, cannulae, tubing, filters, and other accessories (e.g., monitors, detectors, sensors, connectors).(b)
Classification —Class II (special controls). The special controls for this device are:(1) The technological characteristics of the device must ensure that the geometry and design parameters are consistent with the intended use, and that the devices and accessories in the circuit are compatible;
(2) The devices and accessories in the circuit must be demonstrated to be biocompatible;
(3) Sterility and shelf-life testing must demonstrate the sterility of any patient-contacting devices and accessories in the circuit and the shelf life of these devices and accessories;
(4) Non-clinical performance evaluation of the devices and accessories in the circuit must demonstrate substantial equivalence of the performance characteristics on the bench, mechanical integrity, electromagnetic compatibility (where applicable), software, durability, and reliability;
(5) In vivo evaluation of the devices and accessories in the circuit must demonstrate their performance over the intended duration of use, including a detailed summary of the clinical evaluation pertinent to the use of the devices and accessories to demonstrate their effectiveness if a specific indication (patient population and/or condition) is identified; and
(6) Labeling must include a detailed summary of the non-clinical and in vivo evaluations pertinent to use of the devices and accessories in the circuit and adequate instructions with respect to anticoagulation, circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure.