510(k) Data Aggregation
(265 days)
The Nautilus(tm) Smart ECMO Module with integrated heat exchanger is intended to provide assisted long-term extracorporeal circulation and physiologic gas exchange (oxygenation and carbon dioxide removal) of the patient's blood for up to 48 hours in adult and pediatric adolescent patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The integrated heat exchanger is intended to heat or cool the blood as needed during use. Integrated fluid path pressure, temperature, and oxygen saturation monitoring is achieved by built-in sensor modules and display.
The Nautilus(tm) ECMO Oxygenator with integrated heat exchanger is intended to provide assisted long-term extracorporeal circulation and physiologic gas exchange (oxygenation and carbon dioxide removal) of the patient's blood for up to 48 hours in adult and pediatric adolescent patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The integrated heat exchanger is intended to heat or cool the blood as needed during use.
The Nautilus™ ECMO oxygenators are diffusion membrane oxygenators used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Blood enters the device and passes through both the heat exchange membrane, where temperature is adjusted, and the gas transfer membrane, where oxygen is added and carbon dioxide is removed.
The Nautilus™ Smart ECMO Module device contains integrated sensors with an electronic touch screen display. The following measured parameters are visible on the electronic display: inlet pressure, inlet oxygen saturation, outlet pressure, outlet oxygen saturation, and outlet blood temperature. The difference between the inlet and outlet pressure, delta pressure, is calculated and also displayed on the screen. The touchscreen display allows users to set alarm limits for all measured parameters. The Nautilus™ Smart ECMO Module will alarm visually and audibly when the limits are exceeded.
The devices are single-use, nontoxic, non-pyrogenic, and not made from natural latex rubber materials.
The provided text describes performance evaluations for the MC3 Nautilus Smart ECMO Module and Nautilus ECMO Oxygenator. However, it does not explicitly state "acceptance criteria" in a tabulated format and then directly link each criterion to "reported device performance" in a separate column. Instead, it presents various performance evaluations and their findings.
Based on the information provided, here's a structured summary attempting to address your request as closely as possible:
1. Table of acceptance criteria and the reported device performance
While explicit acceptance criteria are not tabulated with specific numerical targets, the document implies that the device is considered acceptable if its performance is "comparable or better" than a state-of-the-art device and meets the "special controls" and standards.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Gas Exchange (Oxygen and Carbon Dioxide Transfer) | Long-term Bench Gas Exchange testing (ISO 7199:2016): "Nautilus had comparable or better oxygen and carbon dioxide transfer rates... as a heparin coated state-of-the-art device, over 14-days of use." |
Pressure Drop | Long-term Bench Gas Exchange testing (ISO 7199:2016): "...pressure drop performance as a heparin coated state-of-the-art device, over 14-days of use." |
Clotting Prevention / Thrombogenicity | In-Vivo Study in Sheep (96 hours): "After 4 days, there were no clots in any location in the Nautilus device..." |
Clinical information (Real World Evidence): "There were no reports of circuit clotting..." | |
Integrity / Mechanical Failure | Non-clinical Performance: "...mechanical integrity... durability, reliability..." demonstrated substantial equivalence. |
Clinical information (Real World Evidence): "There were no reports of... mechanical failure." | |
Hemolysis | Performance Evaluations: "Hemolysis" was evaluated. |
Clinical information (Real World Evidence): "Fifteen percent of patients had reports of hemolysis complications, of which none was directly attributed to the Nautilus™. One hemolysis complication was reported to have occurred after 27 days of operation, another was reported prior to Nautilus™ use while operating on a different oxygenator, and one was attributed to patient condition (sepsis) and pump clotting." (This implies hemolysis, when it occurred, was generally not attributed to the device itself.) | |
Biocompatibility | "The subject device is demonstrated to be biocompatible for prolonged use in circulating blood in accordance with ISO 10993-1:2009 and in accordance with GLP (21 CFR 58)." |
Sterility and Shelf-life | "Testing demonstrates the sterility of the subject device as provided and that it maintains its sterility, integrity, durability, and reliability over the stated shelf-life of the device." |
Sensor Accuracy (for Smart ECMO Module) | Smart Module Testing: "Sensor Accuracy" was evaluated. |
Reference Device (Sensor Module): "The sensor module functions are comparable to the Maquet CARDIOHELP System with regard to sensor accuracy..." | |
Electrical Safety/EMC (for Smart ECMO Module) | Smart Module Testing: "Electrical Safety/EMC" was evaluated. |
Non-clinical Performance: "...electromagnetic compatibility..." demonstrated substantial equivalence. | |
Water Ingress Protection (for Smart ECMO Module) | Smart Module Testing: "Water Ingress Protection" was evaluated. |
Adverse Events / Complications (e.g., infections, CNS events) | Clinical information (Real World Evidence): |
- "The most common complication reported was circuit exchange in twenty percent of patients..."
- "Ten percent of patients had reports of blood infections on ECMO, which is similar to the rate of infections (11%) previously reported by others."
- "One patient suffered a CNS infarction event (5%), and another patient, a CNS hemorrhage." |
| Compatibility with other circuit devices | Technological Characteristics: "The subject device is designed to be compatible with other extracorporeal circuit devices and accessories." |
The studies collectively show that the Nautilus device meets performance expectations by being comparable to or better than predicate/reference devices and established clinical norms, and by demonstrating safety and efficacy in various testing environments.
2. Sample size used for the test set and the data provenance
- Long-term Bench Gas Exchange testing: The sample size isn't explicitly stated as a number of devices, but it refers to "MC3 Nautilus oxygenators from accelerated age populations" and "competitive heparin coated oxygenators."
- Provenance: In vitro, conducted according to ISO 7199:2016, using bovine blood/plasma.
- In-Vivo Study in Sheep:
- Sample size: A total of 15 sheep. Specifically, 10 sheep were used to evaluate the Nautilus device (at either 2 or 5 L/min), and 5 control animals were studied with a state-of-the-art heparin-coated device (at 2 L/min).
- Provenance: Prospective animal study.
- Clinical information (Real World Evidence):
- Sample size: Not explicitly stated as a number of patients, but data were collected from patients at 7 centers in 4 countries. The average duration of ECMO support was 378.5 hours (15.8 days), with a range of 17 to 1271 hours (52.9 days). Given these durations, it implies a significant number of patient-hours of use.
- Provenance: Retrospective, voluntarily entered into the ELSO registry from initial real-world clinical experience. Data from 7 centers in 4 countries.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the text. The studies described are performance-based (bench, in-vivo) and real-world data collection, rather than studies requiring expert adjudication of device performance in a diagnostic sense.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided in the text. The studies described (bench, animal, real-world registry data) do not appear to involve a human adjudication process for establishing a "ground truth" in the way a diagnostic algorithm might be evaluated.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study, nor any evaluation of human readers improving with AI assistance. The device is an ECMO module/oxygenator, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not directly applicable. The device (ECMO module/oxygenator) is a medical device that physically interacts with a patient's blood, not an algorithm. The "Smart Module" has integrated sensors and an electronic display, which would operate in a "standalone" algorithmic sense for its sensor functions, but its primary function as an ECMO device is not an algorithm. The sensor accuracy was evaluated as part of "Smart Module Testing."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Long-term Bench Gas Exchange testing: Ground truth for gas exchange and pressure drop was established by measurements according to the ISO 7199:2016 standard using bovine blood, providing objective scientific measurements.
- In-Vivo Study in Sheep: Ground truth for clotting was established by direct observation ("no clots in any location") and confirmation of oxygenator functionality in the animal model.
- Clinical information (Real World Evidence): Ground truth for complications and device performance in a clinical setting was derived from clinical observation and patient outcomes data voluntarily entered into the ELSO registry by clinicians at the participating centers.
8. The sample size for the training set
This information is not applicable/provided. The device is a physical medical device, not a machine learning algorithm that requires a "training set" for its operation in the way an AI diagnostic tool would.
9. How the ground truth for the training set was established
This information is not applicable/provided, as the concept of a "training set" does not apply to this type of device.