K Number
K191879

Validate with FDA (Live)

Date Cleared
2019-12-20

(158 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is a permanently installed diagnostic x-ray system for general radiographic x-ray imaging including tomography. This device also supports digital tomosynthesis. The tomography and digital tomosynthesis features are not to be used for imaging pediatric patients.

Device Description

Carestream Digital Tomosynthesis (DT) is a limited "sweep" imaging technique that generates multiple two-dimensional (2D) coronal slices (i.e. planes) from a series of low dose x-ray images of the same anatomy taken at the same exposure but at different angles. During a tomosynthesis acquisition the detector will be stationary while the tube head travels (sweeps) in a straight path (i.e. focal spot travel path). For each exposure, the tube will be angled toward the center of the detector. The Carestream Digital Tomosynthesis contains 3 options: Sweep angle option is to provide the desired slice thickness. The number of images per degree of sweep angle. The Projection Image Resolution allows for the selection of speed of capture versus image resolution.

AI/ML Overview

The Carestream Digital Tomosynthesis (DT) system was evaluated through non-clinical (bench) testing and a clinical reader study to demonstrate its diagnostic image quality and equivalence to predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device PerformanceComments
Diagnostic Image QualityMean RadLex Rating (4-point scale) for scout and Digital Tomosynthesis exams.3.7156The RadLex scale ranged from 1 (non-diagnostic) to 4 (exemplary). All ratings were above non-diagnostic.
Equivalence to Predicate"Equivalent or better in diagnostic quality compared to images obtained using commercially available predicate and reference devices."AchievedStatistical test results demonstrated equivalence or superiority.
Non-clinical PerformanceConformance to specifications, intended workflow, related performance, overall function, verification and validation of requirements for intended use, and reliability of system software.MetPredefined acceptance criteria were met, demonstrating the device is as safe, effective, and performs as well as or better than the predicate device.

2. Sample Size and Data Provenance

  • Test Set Sample Size:
    • Clinical Images: 17 Digital Tomosynthesis image cases from adult human subjects (patients). Each case included a thoracic digital radiograph (PA and lateral chest exposure) and a DT exam (scout PA chest image and DT exposures).
    • Phantom Images: 11 Digital Tomosynthesis phantom exams and corresponding Linear Tomography exams.
  • Data Provenance: Clinical study conducted at Toronto General Hospital located in Toronto, Ontario, Canada (prospective). Phantom studies were also conducted.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Seven (7) board certified radiologists.
  • Qualifications: "general varying reading experience." (No further specific details on years of experience were provided in the text).

4. Adjudication Method for the Test Set

The text indicates that seven radiologists performed an evaluation, but it does not specify an adjudication method (e.g., 2+1, 3+1 consensus). It only states they used a "graduated 4-point RadLex rating scale" and the mean rating was calculated from their assessments.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A clinical reader study was performed. The study involved seven radiologists evaluating images from the investigational device, a reference comparison (standard of care PA and lateral chest x-rays), and the predicate device (Linear Tomography phantom studies). The "statistical test results demonstrate the Carestream Digital Tomosynthesis delivers quality imaging performance that is equivalent or better in diagnostic quality compared to images obtained using commercially available predicate and reference devices."

  • Effect Size: The document does not provide a specific quantitative effect size of how much human readers improved with AI (Digital Tomosynthesis) vs. without AI assistance. It states the DT system was found to be "equivalent or better" in diagnostic quality.

6. Standalone (Algorithm Only) Performance

The document describes the "Carestream Digital Tomosynthesis reconstruction software leverages algorithms that are the same in principle to those applied in computed tomography (CT), such as filtered back projection or iterative reconstruction etc." While it implies algorithm processing, the overall evaluation was of the imaging system producing the images for radiologist interpretation. The reader study assessed the diagnostic image quality facilitated by the DT feature. It is not explicitly stated whether a standalone algorithm-only performance assessment without human-in-the-loop was conducted. The focus was on the diagnostic utility of the images produced by the device.

7. Type of Ground Truth Used for the Test Set

The ground truth for the clinical cases was based on the diagnostic image quality ratings by board-certified radiologists using a RadLex scale. For the phantom studies, the ground truth would inherently be known from the phantom's construction and expected imaging characteristics, used for comparison with Linear Tomography.

8. Sample Size for the Training Set

The document does not provide information on the sample size used for the training set of any AI or reconstruction algorithms.

9. How Ground Truth for the Training Set Was Established

The document does not provide information on how ground truth was established for the training set. It focuses on the validation of the device performance.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Carestream Health, Inc. c/o Duane Gutowski Manager Regulatory Affairs, Clearance and Surveillance 150 Verona Street ROCHESTER, NY 14608

December 20, 2019

Re: K191879

Trade/Device Name: DRX-Evolution with Carestream Digital Tomosynthesis Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: KPR, IZF Dated: November 26, 2019 Received: November 29, 2019

Dear Duane Gutowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191879

Device Name

DRX-Evolution Plus with Carestream Digital Tomosynthesis

Indications for Use (Describe)

The device is a permanently installed diagnostic x-ray system for general radiographic x-ray imaging including tomography. This device also supports digital tomosynthesis. The tomography and digital tomosynthesis features are not to be used for imaging pediatric patients.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the text "K191879" at the top, followed by the word "Carestream" in a stylized orange font. The word "Carestream" appears to be a logo or brand name. The text is clear and legible against a white background.

"510(k) Summary"

510(k) Owner Name:510(k) Owner Address:Carestream Health, Inc.150 Verona StreetRochester, NY, 14608
510(k) Owner Phone:510(k) Owner Fax:585-627-8760585-627-8802
Contact Person & Info:Duane Gutowski RegulatoryAffairs Managerdusty.gutowski@carestream.com585-627-8760
Date Summary Prepared:December 16, 2019
Device Trade Name:Device Common Name:DRX-Evolution with CarestreamDigital Tomosynthesis,
Classification Name:Primary Device Class:Primary Device Code:Primary Regulation Number:Stationary x-ray system2KPR21 CFR 892.1680
Secondary Product CodeIZF
Predicate Device:DRX-Evolution with Linear TomographyManufactured by: Carestream Health, Inc. Systems510(k) No.: K141837 (3/11/2015)21 CFR 892.1680; KPR

Reference Device: Reference 510(k) Submitter: FUJIFILM Medical Systems U.S.A., Inc., 419 West Avenue, Stamford, Connecticut, 06902 USA.

{4}------------------------------------------------

Reference Trade Name:FUJFILM Tomosynthesis option for FDR AcSelerateStationary X-ray System
Reference 510(k) Number:K121499
Reference Regulation Description:Tomographic x-ray system
Reference Review Panel:Radiology
Reference Product Code:IZF
Reference Regulation Number:21 CFR 892.1740
Reference Device Class:2

Reference Predicate Device Indications for Use – Fujifilm's Tomosynthesis option is intended to acquire images of human anatomy and to be used with Fujifilm's DR X-ray systems. Tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep. It is not intended for mammographic applications.

Reference Predicate Device Description - Tomosynthesis is an advanced radiographic application that produces individual coronal "slice" images through an anatomical region of interest (ROL). To produce these slices multiple projection radiographic images are acquired in rapid succession as the X-ray tube sweeps and rotates across the ROI. Once acquired these projection images are subject to processing that registers and reconstructs them to individual tomographic slices.

We have included the Fuji FDR AcSelerate Stationary X-ray System with tomosynthesis, FDA 510k number K121499, as the reference predicate to support the differences in characteristic effects on the safety and effectiveness of the subject device.

Device Description:

Carestream Digital Tomosynthesis (DT) is a limited "sweep" imaging technique that generates multiple two-dimensional (2D) coronal slices (i.e. planes) from a series of low dose x-ray images of the same anatomy taken at the same exposure but at different angles. During a tomosynthesis acquisition the detector will be stationary while the tube head travels (sweeps) in a straight path (i.e. focal spot travel path). For each exposure, the tube will be angled toward the center of the detector.

The Carestream Digital Tomosynthesis contains 3 options: Sweep angle option is to provide the desired slice thickness. The number of images per degree of sweep angle. The Projection Image Resolution allows for the selection of speed of capture versus image resolution.

{5}------------------------------------------------

Indications for Use / Intended Use:

The Indications for Use for the device, as described in its labeling, are:

"The device is a permanently installed diagnostic x-ray system for general radiographic xray imaging including tomography. This device also supports digital tomosynthesis. The tomography and digital tomosynthesis features are not to be used for imaging pediatric patients."

The intended use for this device, as determined by descriptions and the proposed labeling contained in this submission, is similar to the Indications for Use statement provided above.

The Indications for Use for the subject device is the similar as that for the predicate device and the intended use remains unchanged. Any variation in features or technical specifications have been identified and addressed through testing (described below) to support a substantial equivalence determination.

Comparison of Technological Characteristics:

Based upon information provided within this submission, we believe that the Carestream Digital Tomosynthesis is substantially equivalent to the legally marketed DRX-Evolution Plus with Linear Tomography, the identified predicate device.

Digital tomosynthesis operates based on the same principle as linear tomography except the detector can either be mechanically motorized in synchronization with the x-ray source or simply be stationary during the exam. The digital tomosynthesis implemented by Carestream chooses to use a stationary x-ray detector for the purpose of reducing motion blur and improving spatial resolution.

Both linear tomography and digital tomosynthesis use a sweep angle to isolate the anatomy of interest. By establishing the initial pivot level (fulcral plane) and exposing across the sweep angle the anatomical details in the fulcral plane are persevered, while anatomical clutter above or below the plane will be blurred. Both linear tomography and digital tomosynthesis isolate the anatomy of interest in the fulcral plane from unwanted clutter and improve the contrast to noise ratio of the image details.

Each linear tomography acquisition uses one single continuous x-ray exposure. Many acquisitions are frequently needed in order to fully cover the anatomical area of interest. This would increase the overall exam time and the total radiation dose by the total number of scans. As the overall exam time and the total radiation dose must be limited, the number of tomographic scans is limited.

Digital tomosynthesis uses multiple low-power, pulsed x-ray exposures instead of a single continuous exposure. Correspondingly the x-ray detector in a digital tomosynthesis exam will capture each short x-ray pulse as individual images from different projection angles of the x-ray source. These projections are used to reconstruct a stack of digital tomosynthesis

{6}------------------------------------------------

slices for viewing, which each slice equivalent to a traditional tomogram with the fulcral plane set at the center of the digital tomosynthesis slice.

As the image acquisition is separated from the digital tomosynthesis image volume formation process, a reconstruction process is used to maximize the anatomy contrast to noise ratio in the tomosynthesis slices. The Carestream Digital Tomosynthesis reconstruction software leverages algorithms that are the same in principle to those applied in the computed tomography (CT), such as filtered back projection or iterative reconstruction etc. As a result the anatomical details are greatly improved with digital tomosynthesis when compared to traditional linear tomography.

Discussion of Testing

The performance characteristics and operation / usability of the Carestream Digital Tomosynthesis were evaluated in non-clinical (bench) testing. These studies have demonstrated the intended workflow, related performance, overall function, verification and validation of requirements for intended use, and reliability of the system software requirements. Non-clinical test results have demonstrated that the device conforms to its specifications. Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device.

A clinical reader study was completed to evaluate the imaging performance of the Carestream Digital Tomosynthesis software. The purpose of the study was to demonstrate the diagnostic image quality of the Carestream Digital Tomosynthesis feature. This image quality evaluation was done against a reference comparison which was the standard of care PA and lateral chest x-rays. Additionally, phantom Digital Tomography studies were done including a scout and Digital Tomosynthesis volume from the investigational device and a Linear Tomography exam to serve as the predicate device comparison.

Seventeen (17) Digital Tomosynthesis image cases were acquired from adult human subjects (patients) from a clinical study conducted at Toronto General Hospital located in Toronto, Ontario, Canada. Each Digital Tomosynthesis (DT) case consisted of each study participant having a thoracic digital radiograph (DR) exam (an exam includes a PA and lateral chest exposure) and a DT exam (a scout PA chest image and the DT exposures).

In addition, 11 Digital Tomosynthesis phantom exams and corresponding Linear Tomography exams were captured. The acquired clinical exams met the inclusion/exclusion criteria. X-ray images were acquired using appropriate device dependent exposure for Digital Tomosynthesis acquisitions.

Seven (7) board certified radiologists, of general varying reading experience performed an evaluation of the images/exams captured using investigational, reference and predicate devices.

{7}------------------------------------------------

A graduated 4 point RadLex rating scale based on diagnostic image quality (with a rating of 1 being non-diagnostic to a rating of 4 being exemplary) was used to determine the quality of the scout and digital tomosynthesis exams. The mean RadLex rating for the scout and Digital Tomosynthesis is 3.7156 with none of the ratings being non-diagnostic.

The statistical test results demonstrate the Carestream Digital Tomosynthesis delivers quality imaging performance that is equivalent or better in diagnostic quality compared to images obtained using commercially available predicate and reference devices.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.