Search Filters

Search Results

Found 13 results

510(k) Data Aggregation

    K Number
    K250850
    Device Name
    Nanox.ARC X
    Date Cleared
    2025-04-16

    (27 days)

    Product Code
    Regulation Number
    892.1740
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    IZF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nanox.ARC X is a stationary X-ray system intended to produce tomographic images for general use including human musculoskeletal system, pulmonary, intra abdominal, and paranasal sinus indications, adjunctive to conventional radiography, on adult patients.

    This device is intended to be used in professional healthcare facilities or radiological environments, such as hospitals, clinics, imaging centers, and other medical practices by trained radiographers, radiologists, physicists.

    Digital Tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep. Applications can be performed with the patient in prone, supine, and lateral positions.

    This device is not intended for mammographic, angiographic, cardiac, intra-cranial, interventional, or fluoroscopic applications. This device is not intended for imaging pediatric or neonatal patients.

    Device Description

    Nanox.ARC X is a stationary, floor-mounted, stand-alone digital tomosynthesis system intended to produce tomographic images for general use including human musculoskeletal system, pulmonary, intra-abdominal, and paranasal sinus indications, from a single tomographic sweep. It serves as an adjunct to conventional radiography, for adult patients in recumbent positions. The system is intended for use in professional healthcare settings such as hospitals, clinics, and imaging centers by trained radiographers, radiologists, and physicists

    The Nanox.ARC X includes a secured, dedicated off-the-shelf handheld operator console, a multisource, tiltable arc gantry with five identical tubes, a motorized patient table, and a flat panel detector of a scintillator-photodetector type. The image reconstruction service and DICOMization services can be hosted either locally or as part of the secured Nanox.CLOUD, according to customer preference. Nanox.CLOUD also hosts a protocol database service package.

    The Nanox.ARC X X-ray tubes are operated sequentially, one at a time, generating multiple low-dose images acquired from different angles, during a single sweep, dividing the overall power requirements among the tubes. The sweep is performed over a motorized patient table. Patients can be placed in prone, supine, and lateral positions.

    The acquired projection imaging data is anonymized and automatically reconstructed to form tomographic slices of the imaged object, with each slice parallel to the table plane. The Tomosynthesis image result reduces the effect of overlying structures and provides depth information on structures of interest. The resultant images are re-identified and sent using the DICOM protocol.

    AI/ML Overview

    Here's an analysis of the provided FDA 510(k) clearance letter for Nanox.ARC X, focusing on the acceptance criteria and the study that proves the device meets those criteria.

    Key Observation: The provided document is a 510(k) Clearance Letter. These letters primarily address the "substantial equivalence" of a new device to a predicate device, rather than providing detailed clinical efficacy trial results as would be found in a Premarket Approval (PMA) application or a de novo classification request. This type of clearance often relies heavily on non-clinical bench testing and technological comparisons to demonstrate that the new device is as safe and effective as a legally marketed predicate.

    Therefore, the information regarding in-depth clinical studies (like MRMC studies, specific ground truth methods, or detailed acceptance criteria for diagnostic accuracy) is limited or absent in this document because it's not typically required for a 510(k) clearance based on substantial equivalence to an existing device with similar technological characteristics. The focus is on demonstrating that the modifications to the predicate device (Nanox.ARC) do not negatively impact its safety or effectiveness.


    Acceptance Criteria and Device Performance Assessment

    Based on the provided document, the "acceptance criteria" are primarily framed around demonstrating that the modified device (Nanox.ARC X) is as safe and effective as its predicate (Nanox.ARC), despite minor technological changes. The proof relies heavily on non-clinical bench testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of a 510(k) summary focused on substantial equivalence and technological comparison, the "acceptance criteria" are inferred from the types of non-clinical tests performed to ensure the new device functions as intended and is as safe and effective as the predicate. The "reported device performance" are the general conclusions drawn from these tests.

    Acceptance Criterion (Inferred from testing performed)Reported Device Performance
    System Electrical QualificationFunctioned as intended.
    System PerformanceFunctioned as intended.
    Longevity and ConsistencyFunctioned as intended.
    Tube Longevity and ReliabilityFunctioned as intended.
    Functional VerificationFunctioned as intended.
    Motion ControlFunctioned as intended.
    Dimensional and Mechanical PropertiesFunctioned as intended.
    Image QualityFunctioned as intended.
    Tube Comparison CEI and Nanox KoreaFunctioned as intended.
    Human Factors SummaryFunctioned as intended.
    Phantom ValidationFunctioned as intended.
    Weight ConsiderationsFunctioned as intended.
    TransportationFunctioned as intended.
    Software Verification and ValidationFunctioned as intended.
    Overall Safety and EffectivenessSimilar to predicate device.

    Note: The level of detail provided in a 510(k) letter doesn't include specific quantitative metrics for each test, only a general statement that the system "functioned as intended" and overall safety/effectiveness are similar to the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated in terms of patient data. The testing described primarily involves bench testing, phantom studies, and system-level verification and validation. There is no indication of a clinical test set involving human patients as one might expect for a diagnostic accuracy study.
    • Data Provenance: Not applicable in the context of clinical patient data for this 510(k) pathway, as no clinical tests were performed. The "data" comes from the results of the various non-clinical bench tests.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not applicable. Since no clinical tests were performed on human patients and no diagnostic accuracy claims are being established through reader studies, there was no need for expert ground truth establishment for a clinical test set.
    • Qualifications of Experts: N/A.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. No clinical test set requiring expert adjudication was conducted.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was explicitly NOT done. The document states: "No clinical tests were performed for the subject device." This type of study would be a clinical test.
    • Effect Size of Human Readers Improvement: Not applicable, as no MRMC study was done.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Standalone Performance: The document does not describe a standalone diagnostic accuracy study of an AI algorithm. The device is a tomographic X-ray system, not an AI diagnostic algorithm, although it does include "image reconstruction service" and "DICOMization services." These are intrinsic functionalities of the imaging system itself, not separate AI components whose standalone diagnostic performance would be evaluated. The "Software Verification and Validation" likely covers the functional correctness of these reconstruction algorithms.

    7. The Type of Ground Truth Used for the Test Set

    • Type of Ground Truth: Not applicable for a clinical test set. The "ground truth" for the non-clinical tests would be the established engineering specifications, phantom measurements, and functional requirements against which the device's performance was measured (e.g., a known phantom structure for image quality, or expected electrical parameters for qualification).

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This 510(k) is for a hardware device (X-ray system) with associated software for image reconstruction. It is not an AI/ML algorithm that undergoes a distinct "training" phase on a specific dataset for diagnostic interpretation. The image reconstruction algorithms are typically deterministic or based on established physics and signal processing, not on deep learning models trained on large image datasets.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable, as there isn't a "training set" in the context of an AI/ML diagnostic algorithm for which ground truth would be established. The "ground truth" for the development of image reconstruction algorithms would be based on mathematical models, physical principles of X-ray interaction, and calibrated phantom data to optimize image quality and accuracy.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242395
    Device Name
    Nanox.ARC
    Date Cleared
    2024-12-04

    (113 days)

    Product Code
    Regulation Number
    892.1740
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    IZF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nanox.ARC is a stationary X-ray system intended to produce tomographic images for general use including human musculoskeletal system, pulmonary, intra-abdominal, and paranasal sinus indications, adjunctive to conventional radiography, on adult patients.

    This device is intended to be used in professional healthcare facilities or radiological environments, such as hospitals, clinics, imaging centers, and other medical practices by trained radiographers, radiologists, and physicists.

    Digital Tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep. Applications can be performed with the patient in prone, supine, and lateral positions.

    This device is not intended for mammographic, cardiac, intra-cranial, interventional, or fluoroscopic applications. This device is not intended for imaging pediatric or neonatal patients.

    Device Description

    Nanox.ARC is a stationary, floor-mounted, stand-alone digital tomosynthesis system intended to produce tomographic images for general use including human musculoskeletal system, pulmonary, intra-abdominal, and paranasal sinus indications, from a single tomographic sweep. It serves as an adjunct to conventional radiography, for adult patients in recumbent positions. The system is intended for use in professional healthcare settings such as hospitals, clinics, and imaging centers by trained radiologists, and physicists.

    The Nanox.ARC includes a secured, dedicated off-the-shelf handheld operator console, a multisource, tiltable arc gantry with five identical tubes, a motorized patient table, and a flat panel detector type. The image reconstruction service and DICOMization services can be hosted either locally or as part of the secured Nanox.CLOUD, according to customer preference.

    Nanox.CLOUD also hosts a protocol database service package.

    The Nanox.ARC X-ray tubes are operated sequentially, one at a time, generating multiple low-dose images acquired from different angles, during a single sweep, dividing the overall power requirements among the tubes. The sweep is performed over a motorized patient table. Patients can be placed in prone, supine, and lateral positions.

    The acquired projection imaging data is anonymized and automatically reconstructed to form tomographic slices of the imaged object, with each slice parallel to the table plane. The Tomosynthesis reduces the effect of overlying structures and provides depth information on structures of interest. The resultant images are re-identified and sent using the DICOM protocol.

    AI/ML Overview

    The provided text is a 510(k) summary for the Nanox.ARC device. It mentions a "Clinical Sample Data evaluation" and confirms that the device can "generate diagnostic-quality images for the expanded Indications for Use," but it does not provide specific details on acceptance criteria or the study design and results as requested in the prompt.

    Therefore, I cannot provide a table of acceptance criteria, reported performance, sample sizes (for test/training), ground truth details, expert qualifications, or adjudication methods directly from the provided text. The document states that "The non-clinical performance testing conducted on the predicate device submitted under K222934 remain applicable to the subject device," implying that some of the detailed testing justification might reside in the predicate device's 510(k) submission (K222934).

    However, I can extract the information that is present:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated in terms of quantitative metrics (e.g., sensitivity, specificity, image quality scores).
    • Reported Device Performance: "Nanox.ARC System functioned as intended" and "generate diagnostic-quality images for the expanded Indications for Use. This includes the evaluation of complex and abnormalities of various sizes and shapes."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size (test set): Not specified. The document mentions "clinical sample data" but not the number of cases.
    • Data Provenance: Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not specified.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not specified.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • An MRMC study is not mentioned. The device is described as an imaging system intended to produce tomographic images, with "adjunctive to conventional radiography." This phrasing suggests human interpretation of the images produced by the device, but not necessarily an AI-assisted interpretation workflow and its comparative effectiveness.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • The document implies the device generates images for human interpretation ("trained radiographers, radiologists, and physicists"). A standalone algorithm performance (without human-in-the-loop) is not discussed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not specified. The document uses terms like "diagnostic-quality images" and "evaluation of complex and abnormalities," which generally implies comparison against established diagnostic standards, likely expert-interpreted images or clinical findings, but the specific type of ground truth (e.g., expert consensus, pathology, follow-up) is not detailed.

    8. The sample size for the training set

    • Not specified.

    9. How the ground truth for the training set was established

    • Not specified.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222934
    Device Name
    Nanox.ARC
    Date Cleared
    2023-04-28

    (214 days)

    Product Code
    Regulation Number
    892.1740
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    IZF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nanox.ARC is a stationary X-ray system intended to produce tomographic images of the human musculoskeletal system adjunctive to conventional radiography, on adult patients. This device is intended to be used in professional healthcare facilities or radiological environments, such as hospitals, clinics, imaging centers, and other medical practices by trained radiographers, radiologists, and physicists. Digital Tomosynthesize tomographic slices from a single tomographic sweep. Applications can be performed with the patient in prone, supine, and lateral positions. This device is not intended for mammographic, cardiac, pulmonary, intra-abdominal, intra-cranial, intra-cranial, interventional, or fluoroscopic applications. This device is not intended for imaging pediatric or neonatal patients.

    Device Description

    Nanox.ARC is a tomographic and solid-state X-ray system (product codes IZF and MQB) intended to produce tomographic images of the human musculoskeletal system from a single tomographic sweep, as an adjunct to conventional radiography, on adult patients.

    Nanox.ARC is a floor-mounted tomographic system that consists of a user control console, a multisource, tiltable arc gantry with five alternately-switched tubes, a motorized patient table, a flatpanel detector of a scintillator-photodetector type, and Protocols database and Image processing software packages.

    Nanox.ARC utilizes several small-sized X-ray tubes that are independently and electronically switched, thereby dividing the overall power requirements over multiple tubes. Nanox.ARC utilizes a tilting imaging ring with five X-ray tubes, operated sequentially, one at a time, used to generate multiple low-dose X-ray projection images acquired from different angles during a single spherical (non-linear) sweep. The sweep is performed over a motorized patient table. Patients can be placed in prone, supine, and lateral positions.

    The acquired projection imaging data is automatically reconstructed to form tomographic slices of the imaged object, with each slice parallel to the table plane. The Tomosynthesis image result reduces the effect of overlying structures and provides depth information on structures of interest. The image reconstruction service, as well as the system's protocol database and DICOMization services, can be hosted either locally or as part of the Nanox.CLOUD, according to customer preference. The resultant images are sent using the DICOM protocol.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    The document doesn't explicitly list specific quantitative acceptance criteria in a table format with separate reported device performance values for each criterion. Instead, it states that "Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device."

    The "Table 2: Non-clinical Performance Data" lists various tests performed and reports a "PASS" for each, indicating that the device met the acceptance criteria for those specific tests.

    Table of Acceptance Criteria (Implied) and Reported Device Performance:

    Acceptance Criterion (Implied by Test Description)Reported Device Performance
    System Electrical QualificationPASS
    System Performance (Motion resolution & accuracy)PASS
    System Longevity & ConsistencyPASS
    Tube Longevity and ReliabilityPASS
    Functional VerificationPASS
    Motion Control stabilityPASS
    Detector and image acquisition functionalityPASS
    Usability Summative (Safety, effectiveness, no failures)PASS
    Transportation safetyPASS
    Dimensional and Mechanical PropertiesPASS
    Image QualityPASS
    Phantom Validation (Diagnostic quality vs. predicate)PASS
    Software verification and validationPASS
    Compliance to 21 CFR 1020.30 and 1020.31PASS
    Electrical Safety & EMC (IEC 60601-1, IEC 60601-1-2)PASS
    Radiation Safety (IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54)PASS
    Biocompatibility (ISO 10993-1)PASS

    Study Details:

    1. Sample size used for the test set and the data provenance:

      • Clinical Sample Evaluation (for image quality): Nine (9) Digital Tomosynthesis image cases were acquired from healthy adult human subjects (patients).
      • Phantom Performance Exams: Twelve (12) Digital Tomosynthesis phantom performance exams (total cases = 9 human + 12 phantom = 21 cases).
      • Data Provenance: From a clinical study conducted at Shamir Medical Center in Israel. The study appears to be prospective as it states "image cases were acquired from healthy adult human subjects (patients) from a clinical study conducted at Shamir Medical Center in Israel."
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: One (1)
      • Qualifications: An ABR-certified radiologist.
    3. Adjudication method for the test set:

      • Adjudication Method: Not explicitly stated, but with only one radiologist reviewing, there was no multi-expert adjudication mentioned (e.g., 2+1, 3+1). If only one expert makes the determination, it's effectively "none" in terms of reaching a consensus among multiple experts.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • MRMC Study: No, an MRMC comparative effectiveness study was not conducted. The clinical sample evaluation involved a single ABR-certified radiologist evaluating the diagnostic quality of the Nanox.ARC images themselves, "against a reference comparison which was the standard of care radiographies." This was a direct comparison of images, not a study on human reader performance with or without AI assistance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Standalone Performance: Yes, the described "Bench Testing" and "Non-clinical Performance Data" table largely represent standalone algorithm and system performance without human intervention in the diagnostic interpretation loop. The "Image Quality" and "Phantom Validation" tests also assessed the device's output directly. The clinical sample evaluation by the radiologist was to evaluate the diagnostic quality of the images produced by the device, effectively assessing the device's standalone output for clinical utility.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Type of Ground Truth: For the clinical sample evaluation, the diagnostic quality of the Nanox.ARC images was evaluated by an ABR-certified radiologist "against a reference comparison which was the standard of care radiographies." This implies the "ground truth" was essentially the interpretive diagnostic quality determined by a single expert, compared to standard of care imaging. For the phantom studies, the ground truth would be based on the known physical properties and measurements within the phantoms.
    7. The sample size for the training set:

      • Training Set Sample Size: The document does not provide any information regarding the sample size used for the training set of the Nanox.ARC system's image reconstruction or processing algorithms.
    8. How the ground truth for the training set was established:

      • Training Set Ground Truth: The document does not provide any information on how ground truth was established for the training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221949
    Date Cleared
    2023-01-26

    (205 days)

    Product Code
    Regulation Number
    892.1740
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    IZF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ortho Device is intended to generate tomosynthesis images of human anatomy for diagnostic purposes of the hand, elbow and foot in patients of all ages.

    The imaging will provide the physician visualized information about anatomical structures to facilitate assessment in orthopedic cases such as:
    • Fractures of bones in finger, metacarpus or wrist
    • Fractures of foot, ankle or elbow joint
    • Arthritis

    Device Description

    The Ortho Device is a 3D tomographic X-ray device intended to be used to produce radiological images of a specific cross-sectional plane of the body. The device is comprised of a Flat Panel X-ray source combined with a digital detector within a mounting frame, a control unit and a workstation. It is intended to offer 3D imaging of orthopedic structures by using a panel of X-ray sources that construct a 3D tomosynthesis image with the associated reconstruction software from individual images; it is also possible to create synthetic 2D images of the desired anatomy.

    The Ortho Device is a portable system that can be mounted on a stand for tabletop applications or on a trolley cart for added mobility with motorized vertical positioning. The C-Arm and Control Unit components are both designed to be carriable by a single person. To allow for the ideal positioning of the anatomy (hand and weight-bearing foot images) in the beam path and to achieve the desired plane of view, the Ortho Device C-Arm can be manually rotated by up to 90°. The central beam is aligned perpendicularly to the image receptor.

    The "Ortho Device" was created to fill a diagnostic niche in orthopedic medicine for cost effective and portable imaging for patients and is used, amongst other applications, for 3D-radiographic diagnostic imaging of hand, elbow and foot in orthopedic and radiological practices as well as in emergency departments of hospitals. The Ortho Device results are detailed multi-slice 3D images of patients that allow radiologist interpretation of clinical image data and by this support medical professionals decisionmaking on human anatomy.

    The Ortho Device system is designed to meet the requirements in accordance with relevant sections of 21CFR 1020.30-1020.31.

    AI/ML Overview

    The provided text is a K221949 510(k) summary for the ADAPTIX 3D Orthopedic Imaging System ("Ortho Device"). It does not contain information about acceptance criteria, detailed study designs, or reader study results with explicit performance metrics. The document primarily focuses on demonstrating substantial equivalence to predicate devices through technical comparisons and non-clinical testing.

    Therefore, I cannot fulfill your request for:

    • A table of acceptance criteria and reported device performance.
    • Sample size used for the test set and data provenance.
    • Number of experts and their qualifications for ground truth establishment.
    • Adjudication method for the test set.
    • MRMC comparative effectiveness study results or effect sizes.
    • Standalone performance details.
    • Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for the test set.
    • Sample size for the training set.
    • How ground truth for the training set was established.

    However, based on the section "9. Non-Clinical Performance Data," I can extract the following relevant information regarding performance evaluation, albeit without specific quantitative acceptance criteria or detailed study methodologies:

    The study that "proves the device meets the acceptance criteria" in this context refers to a series of non-clinical tests summarized in Section 9. While "acceptance criteria" for specific performance metrics are not explicitly stated with quantitative thresholds, the document implies that these criteria were met by stating "Passed" for each test.

    1. A table of acceptance criteria and the reported device performance:

    Since explicit quantitative acceptance criteria are not provided, the table below lists the performance aspects tested and the reported outcome.

    Performance Aspect TestedReported Device Performance/Outcome
    In vitro Cytotoxicity (per ISO 10993-5)Passed
    Irritation and skin Sensitization (per ISO 10993-10)Passed
    Systemic toxicity (per ISO 10993-11)Passed
    Electrical safety (per IEC 60601-1)Passed
    Electromagnetic Disturbance (EMD) (per IEC 60601-1-2)Passed
    Radiation protection (per IEC 60601-1-3)Passed
    Medical Electrical Equipment Usability (per IEC 60601-1-6)Passed
    Safety and essential performance of X-ray tube assemblies (per IEC 60601-2-28 and IEC 60601-2-54)Passed
    Particular electrical testing performance req. for Radiation dose documentation (per IEC 61910-1)Passed
    Digital Imaging and Communications in Medicine (DICOM) (per NEMA PS 3.1)Passed
    Transportation Testing (per ASTM D4169)Passed
    Image quality (spatial and contrast resolution, homogeneity, linearity)Passed
    Ability of device to image all intended body parts (fingers, metacarpus/wrist, elbow, foot, ankle)Evaluated and confirmed by radiologists
    Ability of device to provide imaging data for assessment of bone fracture and arthritisEvaluated and confirmed by radiologists
    Software verification and validation (functional level, system compatibility, risk analysis per IEC 62304/FDA Guidance)Completed for Moderate Level of Concern software
    Risk Management (per EN ISO 14971)All requirements met, risks reduced

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
    The document mentions "sample clinical images" being evaluated by radiologists but does not specify the sample size, data provenance, or whether the study was retrospective or prospective for these clinical image evaluations. For other non-clinical tests (e.g., toxicity, electrical safety), the "sample size" would refer to the number of device units or components tested, which is not stated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
    The document states "Evaluation of sample clinical images by radiologists". It does not specify the number of radiologists, their qualifications, or their experience levels.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    The document does not describe any adjudication method used for the evaluation of clinical images.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    The document does not mention an MRMC comparative effectiveness study or any evaluation of human reader performance with or without AI assistance. The device is an imaging system, not explicitly described as having AI for interpretation in this summary.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    The document describes the device as an "imaging system" that "results are detailed multi-slice 3D images of patients that allow radiologist interpretation of clinical image data and by this support medical professionals decision-making." This implies the device provides images for human interpretation, and there is no mention of an algorithm performing standalone diagnoses.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):
    For the evaluation of clinical images, "Evaluation of sample clinical images by radiologists to demonstrate that the device is able to image all intended body parts" and "to help clinician for the assessment of bone fracture and arthritis" implies that the ground truth for these evaluations was based on expert assessment/consensus (i.e., the radiologists' judgment). No pathology or outcomes data is mentioned as ground truth.

    8. The sample size for the training set:
    Not applicable/not provided. This document describes a medical imaging device, not a machine learning algorithm that requires a separate training set. The "software verification and validation testing" mentioned refers to the device's operational software, not an AI training process.

    9. How the ground truth for the training set was established:
    Not applicable/not provided, as there is no mention of a training set for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213081
    Date Cleared
    2021-10-26

    (33 days)

    Product Code
    Regulation Number
    892.1740
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    IZF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CLISIS SYSTEMS, Discovery RF180 is indicated for performing general radiography, fluoroscopy and angiography procedures.

    Applications and techniques:

    • Gastroenterology
    • Skeleton
    • Thorax and lungs
    • Paediatrics
    • Urology and gynecology
    • Emergency/traumatology
    • Digital angiography
    • Linear tomography
    • Auto Image Paste (Stitching)
    • Tomosynthesis
    Device Description

    Not Found

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter for the CLISIS SYSTEMS, Discovery RF180, a tomographic x-ray system. It states the device is substantially equivalent to legally marketed predicate devices and outlines the indications for use.

    However, the provided text DOES NOT contain any information regarding acceptance criteria or the study that proves the device meets those criteria. It is a regulatory clearance document, not a detailed technical report or clinical study summary.

    Therefore, I cannot fulfill your request for:

    1. A table of acceptance criteria and the reported device performance
    2. Sample size used for the test set and the data provenance
    3. Number of experts used to establish the ground truth
    4. Adjudication method
    5. MRMC comparative effectiveness study details
    6. Standalone performance
    7. Type of ground truth used
    8. Training set sample size
    9. How ground truth for the training set was established

    The document focuses on the regulatory approval process and the intended uses of the device, not the technical details of its performance or the studies conducted to validate that performance. To obtain that information, one would typically need to refer to separate study reports, a detailed 510(k) summary (if available publicly with more technical details), or an Instructions for Use (IFU) document if it contains performance specifications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193262
    Manufacturer
    Date Cleared
    2020-03-09

    (104 days)

    Product Code
    Regulation Number
    892.1740
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    IZF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DR 600 system is a General Radiography X-ray imaging system used in hospitals, clinics and medical practices by radiographers, radiologists and physicists to make, process and view static X-ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.

    In addition, the system provides the Agfa tomosynthesis option, which is intended to acquire tomographic slices of human anatomy and to be used with Agfa DR X-ray systems. Digital tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep.

    Applications can be performed with the patient in the sitting, standing or lying position.

    This system is not intended for mammography applications.

    Device Description

    The DR 600 with Tomosynthesis is a tomographic and solid state x-ray system (product codes IZF and MQB) intended to capture tomographic slices and static images of the human body. The DR 600 with Tomosynthesis is a ceiling mounted tomographic and general radiographic system that consists of a tube and operator console with a motorized patient table and/or wall stand. The DR 600 with Tomosynthesis uses Agfa's NX workstation with MUSICA2 ™ image processing and flat-panel detectors of the scintillator-photodetector type (Cesium Iodide - CsI or Gadolinium Oxysulfide - GOS). It is capable of replacing other direct radiography, tomography, image intensifying tubes and TV cameras, including computed radiography systems with conventional or phosphorous film cassettes.

    AI/ML Overview

    The provided text describes the Agfa DR 600 with Tomosynthesis device and its K193262 510(k) submission. However, it does not contain specific acceptance criteria or a detailed clinical study demonstrating the device's meeting of these criteria. The document focuses on showing substantial equivalence to predicate devices primarily through technological characteristics and bench testing, not through comparative clinical effectiveness studies with explicit acceptance criteria for diagnostic performance.

    Therefore, many of the requested details about acceptance criteria, clinical study design, sample sizes, ground truth establishment, expert qualifications, and MRMC studies are not present in the provided text. The document explicitly states: "No clinical trials were performed in the device. No animal or clinical studies were performed in the development of the new device."

    Based on the available information, here's what can be extracted and what is missing:


    Acceptance Criteria and Device Performance (as inferred from the document's approach to substantial equivalence):

    Since no specific acceptance criteria for diagnostic performance (e.g., sensitivity, specificity, AUC) are presented, the "acceptance criteria" for this 510(k) appear to be primarily focused on demonstrating substantial equivalence to predicate devices through technical specifications, image quality evaluations (bench testing), and compliance with various electrical safety, EMC, and software standards.

    Acceptance Criteria (Inferred from Document)Reported Device Performance
    1. Technological Characteristics are Identical/Equivalent to Predicates:
    - Communications (DICOM)
    - Flat Panel Detectors (type, material, sizes, pixel size, dynamic range)
    - Operator Workstation (Agfa NX)
    - Image Processing (MUSICA DTS, MUSICA2, MUSICA3/3+)
    - Operating System (Windows 7, 8, 8.1, 10)
    - Display System (Separately cleared medical display)
    - Power Supply
    - Generators
    

    2. Indication for Use statement is consistent/identical to predicates.

    3. Performance/Functionality as Intended:
    - Confirmed functions and performs as intended.
    - Supports a tomographic workflow and Smart Dr visualization (including adult and pediatric patients).

    4. Image Quality Equivalent to Predicate:
    - For both adult and pediatric patients.

    5. Software Validation:
    - Verification and validation plans confirmed.
    - Risk assessment shows no unacceptable risks.

    6. Electrical Safety and EMC Compliance:
    - Adherence to specified IEC standards (60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54).
    - Compliance with FDA Subchapter J mandated performance standard 21 CFR 1020.30 and 1020.31.

    7. Quality Management System Compliance:
    - Adherence to ISO 13485:2015, ISO 14971:2012, ACR/NEMA PS3.1-3.20 (DICOM). | 1. Technological Characteristics:
    - Communications: Same as both predicates (DICOM).
    - Flat Panel Detectors: Same as both predicates (Flat Panel Detectors, GOS/CsI Scintillator, various sizes 17x17, 14x17, 10x10 in., 148 µm pixel size in primary predicate, 139 µm in other, 16 bit dynamic range in primary predicate, 14 bit in other). The new device shares characteristics with both, indicating equivalence.
    - Operator Workstation: Same as both predicates (Agfa NX).
    - Image Processing: MUSICA DTS, MUSICA2, MUSICA3/3+. The addition of tomographic image processing is identical to the DR 800 (K183275) primary predicate device.
    - Operating System: Same as predicate K183275 (Windows 7, 8, 8.1, 10).
    - Display System: Same as both predicates (Separately cleared medical display K051901).
    - Power Supply: Same as predicate K152639 (50-60 Hz, 380/400/415/440/480V + 10%).
    - Generators: Same as predicate K183275 (Choice of three models: 50, 65KW, 80 KW).
    - Overall: "Principles of operation and technological characteristics of the new and predicate devices are the same."

    2. Indication for Use: "The DR 600 system is a General Radiography X-ray imaging system... In addition, the system provides the Agfa tomosynthesis option... Digital tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep." This is stated to be "virtually identical" to K152639 with the tomosynthesis addition from K183275.

    3. Performance/Functionality:
    - "Technical and acceptance testing was completed on the DR 600 in order to confirm the medical device functions and performs as intended. All deviations or variances are documented... All design input requirements have been tested and passed."
    - "Functionality evaluations were conducted with three qualified radiographers. Usability and TBD. The results of these tests fell within the acceptance criteria for the DR 600; therefore, the DR 600 supports a tomographic workflow and Smart Dr visualization including adult and pediatric patients."

    4. Image Quality: "Image quality bench tests were conducted in support of this 510(k) submission in which anthropomorphic adult and pediatric images taken with the DR 600 and the primary predicate device, DR 800 (K183275) were compared to ensure substantial equivalency. The test results indicated the image processing of the DR 600 passed the acceptance criteria and was equal to the image processing for the primary predicate, DR 800 (K183275) device for both adult and pediatric patients."

    5. Software Validation: "Verification and validation testing confirmed the device meets performance, safety, usability and security requirements... For the NX22 (NX Nomad) software there are a total of 342 risks in the broadly acceptable region and 27 risks in the ALARP region with only one of these risks identified. Zero risks were identified in the Not Acceptable Region."

    6. Electrical Safety and EMC Compliance: Document states compliance with all listed IEC standards and FDA performance standards.

    7. Quality Management System Compliance: Document states adherence to all listed ISO and other standards. |


    Study Details (Based on the provided text):

    1. Sample sizes used for the test set and the data provenance:

      • Test Set: No specific numerical sample size is mentioned for image quality evaluations beyond "anthropomorphic adult and pediatric images." The document mentions "functionality evaluations were conducted with three qualified radiographers," but this refers to human user testing of workflow and usability, not diagnostic image performance.
      • Data Provenance: Not explicitly stated, but likely retrospective as it refers to comparisons of images taken with the new device and a predicate device. The comparison of anthropomorphic phantom images suggests a controlled laboratory setting.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document states "Laboratory data and image quality evaluations conducted with internal and independent specialists confirm that performance is equivalent to the predicates." It also mentions "clinical image quality evaluations for adults and pediatric patients" and "functionality evaluations were conducted with three qualified radiographers."
      • Number of Experts: At least "three qualified radiographers" for functionality, and "internal and independent specialists" for image quality, but exact numbers or specific qualifications (e.g., years of experience, board certification) are not detailed.
      • Qualifications: "Qualified radiographers" and "internal and independent specialists."
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • No adjudication method is described for the image quality evaluations or other performance tests. The comparison to predicates implies direct visual or quantitative comparison by specialists.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. The document explicitly states: "No clinical trials were performed in the device. No animal or clinical studies were performed in the development of the new device."
      • Therefore, no effect size of human readers improving with AI assistance is provided as this type of study was not conducted. The device in question is an imaging system, not an AI-based diagnostic assistance tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The "performance data including clinical image quality evaluations for adults and pediatric patients" involved the system's output. The "image processing of the DR 600 passed the acceptance criteria and was equal to the image processing for the primary predicate, DR 800 (K183275) device for both adult and pediatric patients." This implies an evaluation of the algorithm's output (image quality) without necessarily focusing on a human-in-the-loop diagnostic task. So, in essence, standalone image quality performance was evaluated against a predicate.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for image quality evaluation appears to be comparison to a cleared predicate device's image quality, as judged by "internal and independent specialists," using anthropomorphic phantoms. There is no mention of pathological confirmation or patient outcomes for establishing ground truth, as it was not a clinical trial.
    7. The sample size for the training set:

      • This device is an X-ray system with image processing, not a machine learning/AI algorithm that requires a "training set" in the traditional sense of AI development. The software capabilities (MUSICA DTS, MUSICA2, MUSICA3/3+) are described as being identical to previously cleared versions in predicate devices. Therefore, a "training set" for a new AI model is not applicable here.
    8. How the ground truth for the training set was established:

      • Not applicable, as no new AI model training set is described. The image processing algorithms are identical to those previously cleared.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183275
    Manufacturer
    Date Cleared
    2019-02-01

    (70 days)

    Product Code
    Regulation Number
    892.1740
    Why did this record match?
    Product Code :

    IZF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures:

    • Positioning fluoroscopy procedures
    • Gastro-intestinal examinations
    • Urogenital tract examinations
    • Angiography
      It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts.
      In addition, the system provides the Agfa Tomosynthesis option, which is intended to acquire tomographic slices of human anatomy and to be used with Agfa DR X-Ray systems. Tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep.
      The DR 800 is not intended for mammography applications.
    Device Description

    Agfa's DR 800 with Tomosynthesis a tomographic and fluoroscopic x-ray system (product codes IZF and JAA) intended to capture tomographic slices of the human body. The DR 800 is a floormounted radiographic, fluoroscopic and tomographic system that consists of a tube and operator console with a motorized tilting patient table and bucky with optional wall stand, FLFS overlay and ceiling suspension. The new device uses Agfa's NX workstation with MUSICA image processing and flat-panel detectors for digital and wide dynamic range capture. It is capable of replacing other direct radiography, tomography, image intensifying tubes and TV cameras, including computed radiography systems with conventional or phosphorous film cassettes.

    AI/ML Overview

    The Agfa DR 800 with Tomosynthesis underwent bench testing and software verification and validation to demonstrate substantial equivalence to its predicate devices, the GE Medical System's Discover XR656 with VolumeRAD (K132261) and Agfa's previous version of the DR 800 with MUSICA Dynamic (K180589). The primary focus of the testing for this submission was on the new Digital TomoSynthesis (DTS) software and its performance in generating tomographic slices.

    Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria and the Reported Device Performance

    Performance MetricAcceptance CriteriaReported Device Performance
    Technical & Acceptance TestingAll deviations or variances are documented, addressed in CR&T (Corrective and Remedial Actions) documentation, and verified. All mitigations have been tested and passed. All design input requirements have been tested and passed. All planned verification activities have been successfully completed.Verification and validation testing confirmed the device meets performance, safety, usability, and security requirements. Pediatric indications were also taken into account. Results were verified and validated. Technical and acceptance testing was completed on the DR 800 with Tomosynthesis to confirm the medical device functions and performs as intended. All deviations or variances are documented in a defect database and addressed in the CR&T documentation and verified. All mitigations have been tested and passed. All design input requirements have been tested and passed. All planned verification activities have been successfully completed.
    Usability & Functionality EvaluationThe results of these tests fell within the acceptance criteria for the DR 800 X-ray system.Usability and functionality evaluations were conducted with qualified independent radiographers and internal experts. The results of these tests fell within the acceptance criteria for the DR 800 X-ray system; therefore, the DR 800 supports a radiographic, fluoroscopic, and tomosynthesis workflow including dynamic and static imaging, continuous and rapid sequence exams, tomographic slices calibration, and positioning.
    Image Quality Validation (Adults)The reconstruction software of the image processing for Digital TomoSynthesis (DTS) of the DR 800 X-ray system passed the acceptance criteria. DTS images were suitable for diagnosis.Image Quality Validation testing was conducted using anthropomorphic phantoms and evaluated by qualified independent radiographers and internal experts. The image quality validation included testing a full range of applications for the DR 800 X-ray system with Tomosynthesis compared to reference images from the primary predicate GE Discovery XR656 with VolumeRAD (K132261) using anonymized adult phantoms. The test results indicated that the reconstruction software of the image processing for Digital TomoSynthesis (DTS) of the DR 800 X-ray system passed the acceptance criteria and that the DR 800 with Tomosynthesis is capable of making DTS studies for adult patients. The test results showed MUSICA Digital TomoSynthesis (DTS) images were suitable for diagnosis for adult patients.
    Image Quality Validation (Pediatric)The reconstruction software of the image processing for Digital TomoSynthesis (DTS) of the DR 800 X-ray system passed the acceptance criteria. Both 5x the dose and 10x the dose images were clinically sufficient and within the intended use, and DTS images were suitable for diagnosis for pediatric patients.Image Quality Validation testing was conducted using anthropomorphic phantoms and evaluated by qualified independent radiographers and internal experts. The image quality validation included testing using anonymized pediatric phantoms. The pediatric phantom image quality validation testing analyzed five tomographic slices at 5x the dose and five tomographic slices at 10x the dose. Both the 5x the dose and 10x the dose images are clinically sufficient and within the intended use. The test results indicated that the reconstruction software of the image processing for Digital TomoSynthesis (DTS) of the DR 800 X-ray system passed the acceptance criteria and that the DR 800 with Tomosynthesis is capable of making DTS studies for pediatric patients. The test results showed MUSICA Digital TomoSynthesis (DTS) images were suitable for diagnosis for pediatric patients.
    Software Risk AssessmentNo risks identified in the Not Acceptable Region. The device is assumed to be safe, and the benefits of the device outweigh the residual risk.During the final risk analysis meeting, the risk management team concluded that the medical risk is no greater than with conventional x-ray film previously released to the field. For the NX4.x.21 (NX Mentor) there are a total of 322 risks in the broadly acceptable region and 27 risks in the ALARP (As Low As Reasonably Practicable) region with only eight of these risks identified. Zero risks were identified in the Not Acceptable Region.
    Electrical Safety & EMC TestingCompliance with various IEC 60601 standards and FDA Subchapter J.The DR 800 with Tomosynthesis is compliant to the FDA Subchapter J mandated performance standard 21 CFR 1020.30 - 1020.32. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, and IEC 60601-2-54 was confirmed.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated as a numerical count of cases/images. The testing involved "anthropomorphic phantoms" for image quality evaluation, including both adult and pediatric phantoms. The pediatric phantom testing analyzed "five tomographic slices at 5x the dose and five tomographic slices at 10x the dose."
    • Data Provenance: The data provenance is from bench testing using anonymized anthropomorphic phantoms. This indicates that the data is prospective in the sense that the phantoms were specifically used for this testing, but it is not from human patients. The country of origin of the data is not specified, but the manufacturer is Agfa N.V. (Belgium).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Number of Experts: "Qualified independent radiographers and internal experts" were used for usability, functionality, and image quality evaluations. The exact number of each group is not specified.
    • Qualifications of Experts: They are described as "qualified independent radiographers and internal experts" and "qualified radiologists" (in the "Descriptive characteristics and performance data including image quality evaluations by qualified radiologists are adequate to ensure equivalence" section). Specific details like years of experience or subspecialty are not provided.

    4. Adjudication Method for the Test Set

    • The document implies that the "qualified independent radiographers and internal experts" evaluated the images and that the results "fell within the acceptance criteria" or "passed the acceptance criteria," suggesting a consensus or individual assessment against predefined criteria. However, a specific adjudication method (e.g., 2+1, 3+1) is not explicitly stated.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No MRMC comparative effectiveness study was done. The study design described is a bench test comparison of the device against a predicate device's reference images using phantoms, with evaluation by human experts, rather than an assessment of human reader performance with or without AI assistance. The device itself is an imaging system, not an AI-powered diagnostic tool for interpretation assistance in the sense of comparing human performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, a standalone evaluation of the algorithm's output was done, as part of the image quality validation. The "reconstruction software of the image processing for Digital TomoSynthesis (DTS)" was evaluated to ensure the generated tomographic images were suitable for diagnosis. This is an assessment of the algorithm's output (the tomographic slices) without direct human intervention in the image generation process, beyond setting up the acquisition parameters.

    7. The Type of Ground Truth Used

    • The ground truth for the image quality evaluation was based on comparison to reference images from the primary predicate device (GE Discovery XR656 with VolumeRAD - K132261) using anthropomorphic phantoms, and expert assessment by "qualified independent radiographers and internal experts" confirming the images were "suitable for diagnosis" and "clinically sufficient." It is not pathology, or outcomes data.

    8. The Sample Size for the Training Set

    • The document does not explicitly state the sample size for the training set for the MUSICA DTS software. It mentions that "The image processing algorithms in the new device are similar to those previously cleared in the DR 800 with MUSICA Dynamic (K180589) and other devices in Agfa's radiography portfolio today... The addition of the tomographic image processing is similar to the predicate device (K132261)." This suggests leveraging existing, previously trained algorithms or development methodologies, rather than describing a specific new training dataset for this submission.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not provided in the document. As noted above, the submission emphasizes similarity to existing, cleared technologies, rather than detailing the unique training of a novel algorithm from scratch.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132261
    Date Cleared
    2013-11-18

    (122 days)

    Product Code
    Regulation Number
    892.1740
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    IZF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Discovery XR656 is intended to generate digital radiographic imoges of the skull, spinal column, chest, abdomen, extremities, and other body parts in patients of all ages. Applications con be performed with the patient sitting, standing, or lying in the prone or supine position and is intended for use in all routine radiography exams.

    When the VolumeRAD option is included on the system, the system can generate tomographic images of human anatomy including the skull, spinal column, chest, abdomen, extremities, and other body parts in patients of all ages.

    When the VolumeRAD option is used for potients undergoing thorocic imaging, it is indicated for the detection of lung nodules. VolumeRad generates diagnostic images of the chest that aid the radiologist in achieving superior detectability of lung nodules versus posterior-anterior and left lateral views of the chest, at o comparable radiation level.

    The device is not intended for mammographic applications.

    Device Description

    The Discovery™ XR656 with VolumeRAD extended claims remains a radiographic X-ray system capable of generating radiographic and tomographic images of human anatomy.

    The Discovery XR656 is designed to handle radiographic applications using GE's flat-panel wireless digital detector. The digital detector is comprised of amorphous silicon and cesium iodide scintillator. The resulting digital image can be sent through a DICOM network for applications such as printing, viewing and storage.

    The Discovery XR656 Digital Radiographic Imaging system consists of a Wallstand, elevating table, overhead Tube support, X-ray tube, collimator, system controller, X-ray generator, and wireless/tethered digital detector. Various configurations are available to meet radiographic requirements. The Optima XR640 version of this product does not include advanced applications and uses the tethered version of the digital detector.

    The Discovery XR656 offers a wide range of advanced clinical applications including GE's VolumeRAD™ (Digital Tomosynthesis), Dual Energy Subtraction, Tissue Equalization and Auto Image Paste.

    This 510(k) is to expand our marketing claims for lung nodule detection using the VolumeRAD application. There is no change in design, manufacturing, materials or energy source of the already cleared device. All the abilities required to achieve the results of the extended claims during the clinical trial were available at the initial release of the system.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the GE Healthcare Discovery XR656 with VolumeRAD, seeking to expand its marketing claims for lung nodule detection. The relevant information to address your request is extracted and organized below.

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission describes the goal of the clinical study, which implicitly sets the acceptance criteria: the device should increase physician accuracy in detecting lung nodules compared to conventional X-ray. The reported performance is that this goal was met.

    Acceptance Criteria (Implicit Goal)Reported Device Performance
    Increase physician accuracy in the detection of lung nodules (3-20mm)"A Multi-centered clinical study was performed to demonstrate that the use of VolumeRAD increases physician accuracy in the detection of lung nodules 3-20mm in diameter, when compared to conventional two-view CxR (posterior-anterior (PA) and left lateral (LAT)), at an average effective dose less than 0.1mSv."
    "With the extended VolumeRAD claims, the Discovery XR656 with VolumeRAD has demonstrated effectiveness beyond the original limited claims of the VolumeRAD feature."
    Comparable radiation level"at a comparable radiation level."

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "A Multi-centered clinical study was performed..." However, it does not specify the sample size used for the test set.

    The data provenance is a "Multi-centered clinical study," implying prospective data collection across multiple sites. The document does not explicitly state the country of origin of the data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts used to establish ground truth or their qualifications. It only refers to "physician accuracy."

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method used for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study and Effect Size

    Yes, a comparative effectiveness study was done. The study aimed to demonstrate that "the use of VolumeRAD increases physician accuracy in the detection of lung nodules... when compared to conventional two-view CxR (posterior-anterior (PA) and left lateral (LAT))."

    The document explicitly states that the study demonstrated an "increase[d] physician accuracy," implying a positive effect size. However, the exact effect size of how much human readers improve with AI (VolumeRAD) vs. without AI assistance (conventional CxR) is not quantified in the provided text (e.g., as a percentage increase in AUC, sensitivity, or specificity).

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The study described is focused on "physician accuracy," indicating a "human-in-the-loop" scenario where radiologists use the VolumeRAD system. A standalone (algorithm only without human-in-the-loop performance) study is not described in the provided text.

    7. Type of Ground Truth Used

    The document does not explicitly state the type of ground truth used. Given the focus on "lung nodule detection" and "physician accuracy," it is highly probable that the ground truth involved expert consensus, pathology, and/or follow-up imaging to confirm the presence or absence of lung nodules. However, this is an inference, not explicitly stated.

    8. Sample Size for the Training Set

    The document does not mention a training set sample size. The submission is for an expanded marketing claim for an existing device ("There is no change in design, manufacturing, materials or energy source of the already cleared device. All the abilities required to achieve the results of the extended claims during the clinical trial were available at the initial release of the system."). This suggests that the VolumeRAD feature was already developed, and this 510(k) is about validating its clinical utility for a specific indication (lung nodule detection), rather than the development of a new AI model requiring a separate training set description in this context.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is described in the provided text (see point 8), how its ground truth was established is not applicable/described here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121499
    Date Cleared
    2012-11-20

    (183 days)

    Product Code
    Regulation Number
    892.1740
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    IZF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fujifilm's Tomosynthesis option is intended to acquire tomographic images of human anatomy and to be used with Fujifilm's DR X-ray systems. Tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep. It is not intended for mammographic applications.

    Device Description

    Tomosynthesis is an advanced radiographic application that produces individual coronal "slice" images through an anatomical region of interest (ROI). To produce these slices, multiple projection radiographic images are acquired in rapid succession as the X-ray tube sweeps and rotates across the ROI. Once acquired, these projection images are subject to image processing that registers and reconstructs them into individual tomographic slices.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Regulatory ComplianceConformance to applicable voluntary standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-4, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, IEC 60825-1, UL 60601-1, DICOM.The proposed device conforms to the listed voluntary standards.
    Device PerformanceSatisfactory results from bench testing.Bench testing to evaluate Tomosynthesis performance was completed, and the results were satisfactory.
    Software FunctionalitySatisfactory results from software verification and validation testing.Software verification and validation testing demonstrated satisfactory results.
    Image QualityDiagnostic image quality substantially equivalent to legally marketed devices, including the existing linear tomography capability of the AcSelerate system.Image quality evaluation demonstrated the proposed Tomosynthesis' diagnostic image quality is substantially equivalent to legally marketed devices, including the existing linear tomography capability of the AcSelerate system.
    Substantial EquivalenceSimilar Indications for Use, functional, and technical requirements as the predicate device (K051967), and similar X-ray system and generator specifications as the previously-cleared FDR AcSelerate System (K093427).The proposed device has very similar Indications for Use, functional and technical requirements as the predicate device, K051967. Most X-ray system and X-ray generator specifications remain the same as the previously-cleared FDR AcSelerate System in K093427.

    2. Sample size used for the test set and the data provenance:

    The document mentions "Various bench testing" and "software verification and validation testing" but does not specify the sample size of cases or tests used in these evaluations. It also does not provide details on the data provenance (e.g., country of origin, retrospective or prospective nature) for the images or data used in the image quality evaluation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The document does not specify the number of experts used or their qualifications for establishing ground truth in any of the evaluations. The comparison is made against "legally marketed devices" and the "existing linear tomography capability," implying a comparative assessment rather than an independent expert ground truth establishment for a specific test set.

    4. Adjudication method for the test set:

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for establishing ground truth or evaluating the test results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No information indicates that a multi-reader multi-case (MRMC) comparative effectiveness study was conducted, nor is there any mention of AI assistance or its effect size on human readers. The Tomosynthesis option itself is a new imaging modality, not an AI-assisted interpretation tool mentioned in this document.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device described is an X-ray system component that produces images, not an algorithm that interprets images. Therefore, the concept of a "standalone (algorithm only without human-in-the-loop performance)" study as typically applied to AI interpretation algorithms is not directly applicable here. The "image quality evaluation" assesses the quality of the output images themselves, which would then be interpreted by humans.

    7. The type of ground truth used:

    For the image quality evaluation, the implicit ground truth appears to be established by comparison to the image quality of "legally marketed devices" and the "existing linear tomography capability" of the predicate system. This suggests a comparative evaluation against established diagnostic quality standards rather than an independent gold standard like pathology or long-term outcomes data for each case. The "satisfactory" outcomes for bench testing and software verification likely rely on meeting predefined engineering and performance specifications.

    8. The sample size for the training set:

    The document is a 510(k) summary for a hardware/software upgrade to an X-ray system, not a machine learning algorithm that requires a training set in the typical sense. Therefore, information about a "training set" is not relevant and not provided.

    9. How the ground truth for the training set was established:

    As this is not a machine learning device with a training set, this question is not applicable to the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013620
    Date Cleared
    2002-01-22

    (78 days)

    Product Code
    Regulation Number
    892.1740
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    IZF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medstone, Radiographic Linear Tomography Option is designed as an optional additional device to the Medstone UroPro 2000 table, intended for use in general radiology. It is designed to enable the production of diagnostic radiologic images of a specific linear cross-sectional plane of the body, eliminating unwanted anatomy and detail by motion blurring of other planes.

    This device is designed to provide and control the sweep speed and angle of a tomography function to enable the production of a properly imaged X-ray film to accomplish diagnostic imaging of a patient.

    The table and tomography option are intended to be used by trained professionals, schooled in proper radiology procedures, techniques, positioning and safety.

    Device Description

    The Medstone, Radiographic Linear Tomography Option is designed as an 'add-on' optional device to enable the production of radiologic images of a specific linear cross-sectional plane of the body by blurring or eliminating detail from other planes. A high degree of emphasis for the design of this device was placed on error detection and reliability as well as usability and safety.

    AI/ML Overview

    The provided 510(k) summary for the Medstone Linear Tomography Option does not include the typical acceptance criteria and a detailed study proving the device meets those criteria, as one might expect for a modern AI/software as a medical device (SaMD).

    Instead, this submission primarily relies on substantial equivalence to a predicate device (Pausch Corporation Microtom, Tomographic Option) based on non-clinical performance data and an absence of clinical performance data being deemed necessary.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a linear tomography option rather than an AI/SaMD, the "acceptance criteria" are more about functional equivalence and safety rather than performance metrics like sensitivity, specificity, or AUC.

    Acceptance Criteria (Inferred from Substantial Equivalence)Reported Device Performance (Medstone Linear Tomography Option)
    Functional Equivalence:
    - Tomographic Movement TypeLINEAR
    - Input/Line Voltage110 VAC 1-phase
    - Input/Line Frequency50/60 Hz
    - Patient positioning and image production capabilitiesEnables production of diagnostic radiologic images of a specific linear cross-sectional plane.
    Performance Parameters (within acceptable deviation):
    - Fulcrum Range0 mm to 250 mm
    - Tomographic Angles20°, 10°, 5°
    - Exposure Times (controlled by X-Ray Generator)0.2 to 1.6 secs.
    - Operational performance ('sweep time' per 'sweep angle')Within 20% differential of predicate, equivalent with respect to imaging output.
    Safety and Effectiveness:
    - Error detection and reliabilityHigh degree of emphasis placed on these.
    - Usability and safetyHigh degree of emphasis placed on these.
    - Absence of new questions of efficacy or substantial riskNo new questions of efficacy or substantial risk raised.

    Note: The reported performance is primarily a comparative statement against the predicate device, asserting "complete equivalence" in safety and effectiveness.

    2. Sample Size for Test Set and Data Provenance

    • Not Applicable. The submission states: "It has been concluded that clinical performance data for the Medstone Linear Tomographic, device was not needed for this 510(k) process. The determination of substantial equivalence is, therefore, not based on an assessment of clinical performance for the purpose of this 510(k) process."
    • There is no mention of a "test set" in the context of clinical data or image analysis. Performance was assessed via non-clinical comparisons and engineering specifications.

    3. Number of Experts and Qualifications for Ground Truth

    • Not Applicable. As no clinical performance study was conducted, there was no need for expert review to establish ground truth from clinical data. The assessment was based on engineering and functional comparisons.

    4. Adjudication Method for Test Set

    • Not Applicable. No clinical test set or adjudication process was required or mentioned.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. No MRMC study was conducted or mentioned. The device is a hardware option for an X-ray table, not an AI-assisted diagnostic tool.

    6. Standalone (Algorithm Only) Performance Study

    • Not Applicable. This device is a mechanical/electronic add-on option for an X-ray table, not a standalone algorithm or software. It facilitates image acquisition, not interpretation.

    7. Type of Ground Truth Used

    • Technical Specifications and Functional Equivalence. The "ground truth" for this submission was implicitly defined by the technical specifications and operational performance of the predicate device. The Medstone device had to demonstrate comparable technical parameters and functional output (e.g., sweep speed, angles, fulcrum range, image output).

    8. Sample Size for Training Set

    • Not Applicable. As there is no AI/machine learning component, there is no "training set."

    9. How Ground Truth for Training Set Was Established

    • Not Applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2