K Number
K191639

Validate with FDA (Live)

Date Cleared
2020-09-11

(450 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TK Insulin Syringe with Safety Retractable Device is a sterile, single-use, disposable and non-reusable, manual retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse.

The TK Insulin Syringe is a sterile, single-use, disposable, insulin syringe intended for injection of U-100 insulin into the body.

Device Description

The TK Insulin Syringe with/without Safety Retractable Device are similar devices except one has a safety feature and the other is a standard insulin syringe.

The TK Insulin Syringe with/without Safety Retractable Device is an integrated needle and piston syringe with an anti-needle-stick mechanism. The mechanism allows clear visualization of the injection site at all times. The mechanism shows the needle is contained within the syringe barrel. After standard techniques for injection, the plunger is withdrawn and snapped off which renders the needle unusable and prevents accidental needle sticks. The used syringe is then discarded into a sharp's container.

The subject devices are sterile, single-use, disposable and non-reusable, insulin syringe intended for injection of U-100 insulin into the body.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the "TK Insulin Syringe with/without Safety Retractable Device." The submission aims to demonstrate substantial equivalence to a predicate device, the "U&U Insulin Syringe with/without Safety Retractable Device."

Here's the breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided document:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria with corresponding reported device performance values. Instead, it relies on demonstrating conformance to established international and national standards. The implicit acceptance criteria are defined by these standards, and the reported performance is that the device "conforms" to them.

Here's a summary of the standards the device was tested against, which serve as the implicit acceptance criteria:

Acceptance Criteria (Standard Conformance)Reported Device Performance
ISO 8537:2016 (Sterile single-use syringes, with or without needle, for insulin)Conforms
ISO 7864:2016 (Sterile hypodermic needles for single use - Requirements and test method)Conforms
ISO 9626:2016 (Stainless steel needle tubing for manufacture of medical device - Requirements and test method)Conforms
ISO 23908:2011 (Sharps Injury Protection - Requirements and test methods - Sharps protection features)Conforms
Guidance for Industry and FDA Staff: Medical Devices with Sharps Injury Prevention FeaturesConforms
ISO 10993 (Biocompatibility - various parts: 1, 4, 5, 7, 10, 11)Conforms
USP <788> (Particulate Matter for injection)Conforms
21 CFR Part 801 (Labeling requirements)Meets requirements

2. Sample size used for the test set and the data provenance

The document does not specify the sample sizes used for any of the non-clinical tests performed on the subject device (e.g., number of syringes tested for performance standards, number of samples for biocompatibility).

The data provenance for these tests is that they were "performed on the subject device" by the manufacturer, Anhui Tiankang Medical Technology Co.,Ltd. The origin of the data would be associated with the manufacturing and testing facility in Tianchang City, Anhui, China. The studies are by nature prospective, as they involve testing the newly manufactured device against established criteria.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to this type of device submission. The "ground truth" for insulin syringes and needles is established by internationally recognized performance and safety standards (e.g., ISO, USP), not by expert consensus on individual test instances in the way one might evaluate an AI diagnostic tool. The compliance with these standards is objectively measurable through predefined test methods.

4. Adjudication method for the test set

This information is not applicable. Since the evaluation is based on conformance to objective standards with predefined test methods, there is no need for expert adjudication of results in the traditional sense of a clinical or image interpretation study. The interpretation of test results against the specified criteria of the standards would be straightforward.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is an insulin syringe, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance would not be relevant or performed.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable. The device is a physical medical device (an insulin syringe), not a standalone algorithm.

7. The type of ground truth used

The "ground truth" for this device's performance and safety is derived from established international and national standards for medical devices, specifically:

  • Performance Standards: ISO 8537, ISO 7864, ISO 9626, ISO 23908.
  • Biocompatibility Standards: ISO 10993 (various parts).
  • Other Quality Standards: USP <788> (Particulate Matter), 21 CFR Part 801 (Labeling).

8. The sample size for the training set

This information is not applicable. As a physical medical device, there is no "training set" in the context of machine learning or AI. The design and manufacturing processes are informed by established engineering principles and prior knowledge from similar devices, but not by a data training set.

9. How the ground truth for the training set was established

This information is not applicable for the same reasons as #8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 11, 2020

Anhui Tiankang Medical Technology Co.,Ltd. Bai Baodong RA Manager No.228 Weiyi Road, Economic Development Zone Tianchang, 239300 Cn

Re: K191639

Trade/Device Name: TK Insulin Syringe with/without Safety Retractable Device Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: MEG, FMF Dated: August 13, 2020 Received: August 13, 2020

Dear Bai Baodong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191639

Device Name

TK Insulin Syringe with/without Safety Retractable Device

Indications for Use (Describe)

The TK Insulin Syringe with Safety Retractable Device is a sterile, single-use, disposable and non-reusable, manual retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse.

The TK Insulin Syringe is a sterile, single-use, disposable, insulin syringe intended for injection of U-100 insulin into the body.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary: K191639

Date prepared: 09.11.2020

1. Submitter Name and Address:

Owner Name:Anhui Tiankang Medical Technology Co.,Ltd.
Address:No.228 Weiyi Road, Development Zone,Tianchang City, Anhui,China
Contactor Name:Bai Baodong
TEL:+86-550-7309187
E-mail:zy@tkmedical.com

Manufacturer

Name:Anhui Tiankang Medical Technology Co.,Ltd
Address:No.228 Weiyi Road, Economic DevelopmentTianchang City, Anhui,China

US Agent:

US Agent:James H . Liao
Address:6775 Verde Ridge Rd Rancho Palos Verdes, CA 90275
TEL:310 3758169 Ext
Email:James@Sino2us.Com

2.Submission Devices :

Product Name: TK Insulin Syringe with/without Safety Retractable Device Classification Name: Syringe, Antistick Piston Syringe Classification : II. Product code: MEG and FMF Submission Type: 510(K) Traditional Regulation Number: 21 CFR 880.5860

3.Predicate Device :

K152808

Product Name: U&U Insulin Syringe with/without Safety Retractable Device Classification name: Syringe, Antistick Piston Syringe Classification: II Product Code: MEG and FMF Regulation No. 21 CFR 880.5860 Review Panel :80

{4}------------------------------------------------

4. Indication for Use

The TK Insulin Syringe with Safety Retractable Device is a sterile, single-use, disposable and non-reusable, manual retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse.

The TK Insulin Syringe is a sterile, single-use, disposable and non-reusable, insulin syringe intended for injection of U-100 insulin into the body.

5. Device Description:

The TK Insulin Syringe with/without Safety Retractable Device are similar devices except one has a safety feature and the other is a standard insulin syringe.

The TK Insulin Syringe with/without Safety Retractable Device is an integrated needle and piston syringe with an anti-needle-stick mechanism. The mechanism allows clear visualization of the injection site at all times. The mechanism shows the needle is contained within the syringe barrel. After standard techniques for injection, the plunger is withdrawn and snapped off which renders the needle unusable and prevents accidental needle sticks. The used syringe is then discarded into a sharp's container.

The subject devices are sterile, single-use, disposable and non-reusable, insulin syringe intended for injection of U-100 insulin into the body.

The TK Insulin Syringe with/without Safety Retractable Device will be available in numerous sizes and combinations in the following table.

Ref NumberModel NumberDescriptionSize
TKIS001TKISInsulin Syringe0.3cc/ml Needle 27G × 1/2"
0.3cc/ml Needle 28G × 1/2"
0.3cc/ml Needle 29G × 1/2"
0.3cc/ml Needle 30G × 1/2"
0.3cc/ml Needle 31G × 3/8"
TKIS002TKISInsulin Syringe0.5cc/ml Needle 27G × 1/2"
0.5cc/ml Needle 28G × 1/2"
0.5cc/ml Needle 29G × 1/2"
0.5cc/ml Needle 30G × 1/2"

{5}------------------------------------------------

0.5cc/ml Needle 31G ×3/8"
TKIS003TKISInsulin Syringe1cc/ml Needle 27G × 1/2"
1cc/ml Needle 28G × 1/2"
1cc/ml Needle 29G × 1/2"
1cc/ml Needle 30G × 1/2"
1cc/ml Needle 31G × 3/8"
TKSIS001TKSISInsulin Syringe with SafetyRetractable Device0.5cc/ml Needle 27G × 1/2"
0.5cc/ml Needle 28G × 1/2"
0.5cc/ml Needle 29G × 1/2"
0.5cc/ml Needle 30G × 1/2"
0.5cc/ml Needle 31G × 3/8"
TKSIS002TKSISInsulin Syringe with SafetyRetractable Device1cc/ml Needle 27G × 1/2"
1cc/ml Needle 28G × 1/2"
1cc/ml Needle 29G × 1/2"
1cc/ml Needle 30G × 1/2"
1cc/ml Needle 31G × 3/8"

6. Comparison of technological characteristics with the predicate:

Through comparisons between the submitted devices with the predicate devices as follows Tables. We believe the applicant devices are substantially equivalent with the predicate devices.

The main components of a TK insulin syringe with a safety retractable device are manufactured to the same or similar specifications as the said assertion device. The intended use, basic design, function, and materials of the predicate device are identical to or similar to the predicate device. The difference between the subject device and the predicate does not affect the safety and effectiveness of the subject device because the evaluation is conducted using industry consensus standards and based on the performance requirements of the device. Therefore, different technical features do not introduce new safety and effectiveness issues and are essentially equivalent to predicate devices.

{6}------------------------------------------------

Comparison Table

Element ofComparisonSubmission DevicePredicate Device K152808Comment
Intended UseThe TK Insulin Syringe withSafety Retractable Device is asterile, single-use, disposableand non-reusable, manualretractable safety insulinsyringe intended for injectionof U-100 insulin into the body,while reducing the risk ofsharps injuries and thepotential for insulin syringereuse.The TK Insulin Syringe is asterile, single-use, disposableand non-reusable, insulinsyringe intended for injectionof U-100 insulin into the body.The U&U Insulin Syringe withSafety Retractable Device is asterile, single-use, disposable andnon-reusable, manualretractable safety insulin syringeintended for injection of U-100insulin into the body, whilereducing the risk of sharps injuriesand the potential for insulinsyringe reuse.The U&U Insulin Syringe is asterile, single-use, disposable andnon-reusable, insulin syringeintended for injection of U-100insulin into the body.Same
Principle of OperationNormalNormalSame
Syringe CapacityTK Insulin Syringe:0.3cc/ml Needle 27G × 1/2"0.3cc/ml Needle 28G × 1/2"0.3cc/ml Needle 29G × 1/2"0.3cc/ml Needle 30G × 1/2"0.3cc/ml Needle 31G × 3/8"0.5cc/ml Needle 27G × 1/2"0.5cc/ml Needle 28G × 1/2"0.5cc/ml Needle 29G × 1/2"0.5cc/ml Needle 30G × 1/2"U&U Insulin Syringe0.3cc/ml Needle 27G to 31G0.5cc/ml Needle 27G to 31G1cc/ml Needle 27G to 31GU&U Insulin Syringe with SafetyRetractable Device:1cc/ml only Needle 27G to 31GDifferentBecause thereare manycustomers inour market whoneed 0.5cc/ml,which indicatesthat customershave demandfor thisspecification,we have added0.5cc/ml in thesyringe
0.5cc/ml Needle 31G × 3/8"capacity.Adding 0.5mLdoes not affectthe safety andeffectiveness ofthe subjectdevice and isverified bynon-clinicaltesting listed inSection 7.
1cc/ml Needle 27G × 1/2"
1cc/ml Needle 28G × 1/2"
1cc/ml Needle 29G × 1/2"
1cc/ml Needle 30G × 1/2"
1cc/ml Needle 31G × 3/8"
TK Insulin Syringe withSafety Retractable Device:
0.5cc/ml Needle 27G × 1/2"
0.5cc/ml Needle 28G × 1/2"
0.5cc/ml Needle 29G × 1/2"
0.5cc/ml Needle 30G × 1/2"
0.5cc/ml Needle 31G × 3/8"
1cc/ml Needle 27G × 1/2"
1cc/ml Needle 28G × 1/2"
1cc/ml Needle 29G × 1/2"
1cc/ml Needle 30G × 1/2"
1cc/ml Needle 31G × 3/8"
Nozzle TypeN.AN.ASame
Lubricant for BarrelSilicone OilSilicone OilSame
Barrel TransparencyTransparent and ClearTransparent and ClearSame
Gradations LegibilityLegibleLegibleSame
Materials
BarrelPPPPSame
PlungerPPPP
Same
PistonRubberRubber
Lubricant for BarrelHC-SS36
Needle HubPP
NeedleStainless
Needle SheathPP
Needle Gauge andLengthVarious SizesSame
Lubricant for NeedleSilicone OilSame
TK Insulin Syringewith Safety RetractableDevice: Sharps InjuryPrevention FeaturesManual Retractable Conformsto ISO 23908Same
PerformancesConforms to ISO7864ISO8537Same
BiocompatibilityConforms to ISO10993(Part 1: Evaluation and testing, Part 4: Selection of tests for interactions with blood, Part 5: Tests for in vitro cytotoxicity, Part 7: Ethylene oxide sterilization residuals, Part 10: Tests for irritation and delayed-type hypersensitivity, Part 11: Tests for systemic toxicity, Tests for Bacterial endotoxins, Tests for Pyrogenicity)Conforms to USP <788>:Particulate Matter for injectionConforms to ISO10993(Part 1: Evaluation and testing, Part 4: Selection of tests for interactions with blood, Part 5: Tests for in vitro cytotoxicity, Part 7: Ethylene oxide sterilization residuals, Part 10: Tests for irritation and delayed-type hypersensitivity, Part 11: Tests for systemic toxicity)Conforms to USP <788>:Particulate Matter for injectionSame
LabelingMeet the requirements of21 CFR Part 801Same

{7}------------------------------------------------

{8}------------------------------------------------

7.Non-Clinical Tests performed on the subject device:

The proposed devices were tested per the following standards, to evaluate its performance.

[PMN-510(K) SUBMISSION]

{9}------------------------------------------------

  • ISO 8537: 2016 Third edition, Sterile single-use syringes, with or without ● needle, for insulin.
  • ISO 7864:2016 Fourth edition, Sterile hypodermic needles for single use-● Requirements and test method.

For the TK Insulin Syringe with Safety Retractable Device Only:

  • ISO 9626:2016 Second edition 2016-08-01 Stainless steel needle tubing for manufacture of medical device-Requirements and test method.
  • . ISO 23908:2011 First edition 2011-06-11 Sharps Injury Protection-Requirements and test methods-Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling.
  • Guidance for Industry and FDA Staff: Medical Devices with Sharps Injury . Prevention Features:

Biocompatibility

Conforms to ISO10993. The TK Insulin Syringe with/without Safety Retractable Device is classified as prolonged, external communicating device, indirect blood contact with the following endpoints:

  • . Cytotoxicity
  • . Sensitization
  • Irritation or Intracutaneous reactivity ●
  • . Acute systemic toxicity
  • . Material mediated pyrogenicity
  • Subacute/Subchronic toxicity
  • Indirect hemolysis
  • . USP 788
  • . ISO 10993-1:2018 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process
  • · ISO 10993-4:2017 Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood
  • ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
  • · ISO 10993-7:2008 Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals

{10}------------------------------------------------

  • ISO 10993-10: 2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
  • ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity
  • USP <788>: Particulate Matter for injection

8. Conclusions:

The TK Insulin Syringe with/without Safety Retractable Device is substantially equivalent to the U&U Insulin Syringe with Safety Retractable Device. The materials, performance, and operational features of both the subject device and the predicate devices are substantially equivalent.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).