K Number
K162895

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-01-18

(93 days)

Product Code
Regulation Number
862.1215
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Immunoassay for the in vitro quantitative determination of cardiac troponin T (cTnT) in lithium heparin plasma. The immunoassay is intended to aid in the diagnosis of myocardial infarction. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas system analyzers. CalSet Troponin T Gen 5 STAT is used for calibrating the quantitative Elecsys Troponin T Gen 5 STAT immunoassay on the cobas system analyzers. PreciControl Troponin is used for quality control of the Elecsys Troponin I and Elecsys Troponin I STAT immunoassays on the Elecsys and cobas e immunoassay analyzers. PreciControl Troponin is also used for quality control of the Elecsys Troponin T Gen 5 STAT immunoassay on the cobas system analyzers. The Elecsys Troponin T Gen 5 CalCheck 5 is an assayed control for use in the calibration verification and for use in the verification of the assay range established by the Elecsys Troponin T Gen 5 reagent on the cobas system analyzers.

Device Description

(1) The Elecsys Troponin T Gen 5 STAT Immunoassay is a two-step sandwich immunoassay on the cobas e 411analyzer and a one-step process on the cobas e 601 analyzer. The assay uses streptavidin-coated.

AI/ML Overview

This document describes the analytical and clinical performance of the Elecsys Troponin T Gen 5 STAT Assay and its associated calibrators and quality controls. It demonstrates the device's substantial equivalence to previously marketed devices.

Here's a breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally implied by the "Results" sections showing that the device's performance metrics either

  • met specific numerical targets (e.g., CV% for precision, ng/L for LoQ), or
  • were within a defined percentage recovery or range (e.g., ±10% for interferences, 90-110% for control recovery, 80-120% for CalCheck recovery).

Here’s a table summarizing the performance based on the provided text, where acceptance criteria are explicitly stated or clearly implied by the successful results.

Performance CharacteristicAcceptance Criteria (Explicit or Implied)Reported Device Performance (Elecsys TnT Gen 5 STAT)
PrecisionCLSI Guideline EP5-A2 met. Accuracy goal: CV of 10% (Intermediate Precision). LoQ: CV of 20% (Intermediate Precision).Precision met goals on both cobas e 411 and cobas e 601.
Troponin T value at 10% CVCV of 10% achievable as per CLSI EP17-A2.cobas e 411: 10.4 ng/L (Lot 170511), 6.70 ng/L (Lot 173678) at 10% CV. cobas e 601: 4.76 ng/L (Lot 170511), 3.85 ng/L (Lot 173678) at 10% CV.
Limit of Quantitation (LoQ)20% CV (intermediate precision) at ≤ 6 ng/L.cobas e 411: 5.5 ng/L (Lot 170511), 3.6 ng/L (Lot 173678) at 20% CV. cobas e 601: 2.5 ng/L (Lot 170511), 2.0 ng/L (Lot 173678) at 20% CV. LoQ labeled as 6 ng/L.
LinearityFor cobas e 411, deviation from linearity not more than 6.3%. For cobas e 601, deviation from linearity not more than 12.4%.Linearity confirmed in the overall range from 3.19 ng/L to 10,439 ng/L (meets specifications set by the study data). Labeled range: 6 – 10000 ng/L.
High Dose Hook EffectNot explicitly stated as a numerical criterion, but "No hook effect was seen".No hook effect seen up to 100,000 ng/L.
Endogenous InterferencesRecoveries within ± 10% of values in samples not containing interferents.Met for Bilirubin, Biotin, Lipemia, Rheumatoid Factors, Hemoglobin, Human Serum Albumin, Cholesterol (up to specified concentrations).
HAMA InterferenceNot explicitly stated as a numerical criterion for acceptance, but "Interference of approximately 10% was seen with HAMA concentrations > 322 µg/L" is reported.Interference of approximately 10% seen with HAMA concentrations > 322 µg/L.
Analytical Specificity/Cross ReactivityRecoveries within ± 10% of values in samples not containing cross-reactants.Met for Skeletal muscle TnT, Skeletal muscle TnI, Cardiac TnI, Human TnC (up to specified concentrations).
Exogenous Interference (Drugs)Recoveries within ± 10% of values in samples not containing the drugs.Met for 16 common pharmaceutical compounds and 18 cardiac drugs (up to specified concentrations).
Sample StabilityAll stressed samples recovered within ± 10% of fresh samples.Stable for 24 hours at 2-8 °C, 12 months at -20 °C (±5°C). Freeze only once.
Calibration Stability (Onboard)7 days.8 days.
Calibration Stability (Lot)12 weeks.13 weeks.
Reagent Stability (First Opening)12 weeks.13 weeks.
Reagent Stability (On-board)4 weeks.5 weeks.
Reagent Stability (Stress)3 weeks.3 weeks.
Reagent Stability (Shelf Life)Not explicitly stated beyond "n/a", but aiming for 18 months.19 months. (Claimed shelf life: 18 months)
Elecsys Troponin T CalSet: StabilityConcentrations < 14 ng/L: Recovery ± 1.4 ng/L. Concentrations ≥ 14 ng/L: Recovery 100 ± 10%.On board stability for CalSet on cobas e 411: up to 5 hours. cobas e 601: use only once. Stable at -20°C (±5°C) for three months.
Elecsys Troponin T CalSet: Real-Time StabilityPreciControl Troponin 1: 90-110%. PreciControl Troponin 2: 90-110%. Internal Control samples: 90-110%.Stable for at least 19 months. Claimed shelf-life: 18 months.
Elecsys PreciControl Troponin: In-Use StabilityConcentrations < 14 ng/L: Recovery ± 1.4 ng/L. Concentrations > 14 ng/L: Recovery 100 ± 10%.On board stability: up to 5 hours at 20-25°C. At 2-8°C: 4 days. At -20°C (±5°C): 3 months.
Elecsys PreciControl Troponin: Real-Time StabilityPreciControl Troponin 1: 90-110%. PreciControl Troponin 2: 90-110%.Shelf life claim: 18 months.
Elecsys PreciControl Troponin: ReconstitutionRecovery 74-126% for PC Troponin 1, 81-119% for PC Troponin 2 (vs. 60 min).Data supports complete reconstitution after 60 minutes.
Troponin T Gen 5 CalCheck 5: Open-Vial StabilityCalCheck Level 1: ≤ 6 ng/L. L2-5: 90-110% recovery of reference value.Stable up to 4 hours at 20-25°C.
Troponin T Gen 5 CalCheck 5: Real-Time StabilityCalCheck Level 1: < 6 ng/L. L2-5: 80-120% recovery of reference value.Labeled shelf life claim: 18 months.

2. Sample Size for the Test Set and Data Provenance

  • Analytical Performance Test Sets:

    • Precision:
      • Sample size: Multiple human plasma samples (5 distinct pools) and 2 PreciControl levels were tested. For each, measurements were done in single determinations in four separate aliquots (2 runs per day) for 21 operating days (n=84 replicates for each sample type).
      • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be laboratory-based analytical studies (retrospective in nature of samples, but prospective in terms of the testing design).
    • LoQ/Functional Sensitivity:
      • Sample size: Ten Li-Heparin plasma pools and two controls tested. Data collected with two lots over 21 days, two runs per day with two replicates per run. (Total 84 measurements for each sample).
      • Data Provenance: Not explicitly stated. Likely laboratory-based.
    • Linearity:
      • Sample size: One high analyte plasma sample (spiked with recombinant TnT) diluted to 21 concentrations (19 dilutions). Measured in three-fold determination within a single run.
      • Data Provenance: Not explicitly stated. Likely laboratory-based.
    • High Dose Hook Effect:
      • Sample size: Two samples spiked to high TnT concentrations, with dilution series.
      • Data Provenance: Not explicitly stated. Likely laboratory-based.
    • Endogenous Interferences:
      • Sample size: Three native or spiked plasma samples (one low, one medium, one high concentration of Troponin T) for each of 7 interfering substances. Dilution series tested.
      • Data Provenance: Not explicitly stated. Likely laboratory-based.
    • HAMA Interference:
      • Sample size: Samples with three different levels of TnT spiked with HAMA. Tested in duplicate.
      • Data Provenance: The HAMA serum used was tested in an "external laboratory".
    • Analytical Specificity/Cross Reactivity:
      • Sample size: Three concentrations of troponin spiked with potential cross-reacting compounds. Tested in duplicate.
      • Data Provenance: Not explicitly stated. Likely laboratory-based.
    • Exogenous Interference (Drugs):
      • Sample size: 16 common pharmaceutical compounds and 18 cardiac drugs spiked into plasma samples. Tested in 3-fold determination.
      • Data Provenance: Not explicitly stated. Likely laboratory-based.
    • Sample Stability:
      • Study 1 (2-8°C): Nine Li-Heparin Plasma samples.
      • Study 2 (-20°C): Ten Li-Heparin Plasma samples.
      • Data Provenance: Not explicitly stated. Likely laboratory-based.
    • Calibration Stability:
      • Onboard: Three plasma samples and two control levels. Each tested with two-fold determination.
      • Lot: Three human plasma samples and two control levels. Each tested with two-fold determination.
      • Data Provenance: Not explicitly stated. Likely laboratory-based.
    • Reagent Stability: Done in four studies (first opening, on-board, stress, shelf life) with specific testing protocols.
      • Data Provenance: Not explicitly stated. Likely laboratory-based.
  • Clinical Performance (Diagnostic Sensitivity and Specificity) Test Sets:

    • APACE Population:
      • Sample Size:
        • Total enrolled: 718 subjects.
        • Baseline test result: 718 subjects.
        • Second test result (3 hour time point): 515 subjects.
        • Test result (6 hour time point): 310 subjects.
      • Data Provenance: International, multicentric prospective trial (APACE study, ClinicalTrials.gov number NCT00470587.6). Countries of origin are not specified beyond "international".
    • Second Multicenter Study:
      • Sample Size:
        • Total enrolled: 1679 subjects presenting with chest pain.
        • Evaluated on cobas e 411: 1679 subjects.
        • Evaluated on cobas e 601: 1675 subjects.
        • Adjudicated MIs: 173 subjects.
      • Data Provenance: Multicenter study. Countries of origin are not specified. Subjects were prospectively enrolled.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Clinical Performance (APACE Population):
    • Number of Experts: An "independent adjudication committee" was used. The document specifies that this committee "included cardiologists" and, "In the case of a disagreement, a third independent cardiologist was used as the tie breaker." This implies at least two cardiologists for initial adjudication and a third for tie-breaking, so a minimum of 3 experts involved in ground truth establishment for each case.
    • Qualifications: "Cardiologists." No specific experience levels (e.g., "10 years of experience") are mentioned, but their specialization is stated.
  • Clinical Performance (Second Multicenter Study):
    • Number of Experts: An "independent adjudication committee" was used, which "included cardiologists and emergency medicine physicians." Similar to APACE, it implies multiple experts, but the exact number or tie-breaking mechanism is not detailed as explicitly as for APACE.
    • Qualifications: "Cardiologists and emergency medicine physicians." No specific experience levels are mentioned.

4. Adjudication Method for the Test Set

  • Clinical Performance (APACE Population): "Diagnosis of MI was done through an independent adjudication committee which included cardiologists. This included 60 day follow-up information on each subject. In the case of a disagreement, a third independent cardiologist was used as the tie breaker." This is a "2+1" or similar consensus-based adjudication method.
  • Clinical Performance (Second Multicenter Study): "Final diagnoses were determined by an independent adjudication committee which included cardiologists and emergency medicine physicians using the universal guidelines." While an independent committee is stated, the specific "tie-breaker" or "consensus" method is not as explicitly detailed as for APACE. No indication of a lack of adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document focuses on the performance of the immunoassay itself, both analytically (precision, linearity, interference, etc.) and clinically (diagnostic sensitivity and specificity) in aid of myocardial infarction diagnosis. It does not describe a study comparing human readers with and without AI assistance or any other form of AI. This is a submission for an in vitro diagnostic (IVD) assay, not an AI/ML-driven diagnostic device.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

This question is not directly applicable as the device is a laboratory immunoassay, which naturally performs "standalone" as an analytical instrument. The "performance" described is the standalone performance of the assay. There is no AI algorithm component mentioned or evaluated as a separate entity that would require a human-in-the-loop for its performance to be assessed.

7. The Type of Ground Truth Used

  • Clinical Performance:
    • For both clinical studies (APACE and the second multicenter study), the ground truth for Myocardial Infarction (MI) diagnosis was established by an independent adjudication committee composed of medical experts (cardiologists and/or emergency medicine physicians). This adjudication was based on "diagnostic criteria described in the ACC/ESC/AHA guidelines reference including ECG changes, symptoms characteristic for ischemia and elevations of cardiac troponin," and "60 day follow-up information on each subject" for the APACE study. This falls under expert consensus guided by clinical outcomes data and established guidelines.
  • Analytical Performance:
    • For analytical studies (precision, linearity, interference, stability, etc.), the ground truth relies on reference methods, spiked samples with known concentrations, or comparison to fresh/unstressed samples, which are standard practices for IVD assay validation.

8. The Sample Size for the Training Set

This information is not applicable/provided. This submission is for an in vitro diagnostic (IVD) assay, not a machine learning or AI model. Therefore, there is no "training set" in the context of an algorithm that learns from data. The studies described are validation studies for the assay's analytical and clinical performance.

9. How the Ground Truth for the Training Set was Established

This information is not applicable/provided for the same reason as above (no training set for an AI/ML algorithm).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles a caduceus, a symbol often associated with medicine and healthcare.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 18, 2017

ROCHE DIAGNOSTICS JANE E. PHILLIPS, Ph.D. SENIOR REGULATORY PROGRAM MANAGER 9115 HAGUE ROAD INDIANAPOLIS, IN 46250

Re: K162895

Trade/Device Name: Elecsys Troponin T Gen 5 STAT Assay, Elecsys Troponin T Gen 5 STAT CalSet, Elecsys PreciControl Troponin, Elecsys Troponin T Gen 5 CalCheck 5 Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: Class II Product Code: MMI, JIT, JJX, JJY Dated: October 20, 2016 Received: October 21, 2016

Dear Dr. Jane Phillips:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known) K162895

Device Name Elecsys Troponin T Gen 5 STAT

Indications for Use (Describe)

Immunoassay for the in vitro quantitative determination of cardiac troponin T (cTnT) in lithium heparin plasma. The immunoassay is intended to aid in the diagnosis of myocardial infarction.

The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas system analyzers.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{3}------------------------------------------------

510(k) Number (if known) K162895

Device Name Elecsys CalSet Troponin T Gen 5 STAT

Indications for Use (Describe)

CalSet Troponin T Gen 5 STAT is used for calibrating the quantitative Elecsys Troponin T Gen 5 STAT immunoassay on the cobas system analyzers.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{4}------------------------------------------------

510(k) Number (if known) K162895

Device Name Elecsys PreciControl Troponin

Indications for Use (Describe)

PreciControl Troponin is used for quality control of the Elecsys Troponin I STAT immunoassays on the Elecsys and cobas e immunoassay analyzers.

PreciControl Troponin is also used for quality control of the Elecsys Troponin T Gen 5 STAT immunoassay on the cobas system analyzers.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{5}------------------------------------------------

510(k) Number (if known) K162895

Device Name Elecsys CalCheck Troponin T Gen 5

Indications for Use (Describe)

The Elecsys Troponin T Gen 5 CalCheck 5 is an assayed control for use in the callbration verification and for use in the verification of the assay range established by the Elecsys Troponin T Gen 5 reagent on the cobas system analyzers.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{6}------------------------------------------------

510(k) Summary

Device NameProprietary name:(1) Elecsys Troponin T Gen 5 STAT Assay(2) Elecsys Troponin T Gen 5 STAT CalSet(3) Elecsys PreciControl Troponin(4) Elecsys Troponin Gen 5 CalCheck 5
Common name:(1) Troponin T Gen 5 STAT(2) CalSet Troponin T Gen 5 STAT(3) PreciControl Troponin(4) CalCheck Troponin T Gen 5
OwnerRoche Diagnostics9115 Hague RoadIndianapolis, IN 46250Phone: 317-521-2000Fax: 317-521-2423
ContactJane Ellen Phillips, PhD
DateJanuary 13th, 2017
ClassificationThe FDA has classified the Elecsys Troponin T Gen 5 STAT(Immunoassay method, troponin subunit) and the Troponin T Gen 5CalSet (Calibrator, Secondary) as Class II devices.
The FDA has classified the multi-analyte and single-analytecontrols, all kinds (assayed and unassayed) as Class I devices(PreciControl Troponin and Troponin T Gen 5 CalCheck 5). Aspart of the test system, they are considered as Class II devices andincluded in this submission.
PanelProductCodeClassification NameRegulationCitation
Clinical ChemistryMMIImmunoassay method,troponin subunit862.1215
Clinical ChemistryJITCalibrator, Secondary862.1150
Clinical ChemistryJJYQuality control material(assayed and unassayed)862.1660
Clinical ChemistryJJXSingle (specified) analytecontrols (assayed andunassayed)862.1660

{7}------------------------------------------------

SubstantialEquivalenceThe Elecsys Troponin T Gen 5 STAT Test System issubstantially equivalent to other devices legally marketedin the United States.
(1) Elecsys Troponin T Gen 5 STAT Immunoassay isequivalent to Elecsys Troponin T, 4th generation STATImmunoassay, Roche Diagnostics (K051752).
(2) Elecsys Troponin T Gen 5 STAT CalSet is equivalentto the Elecsys Troponin T CalSet (K961500).
(3) Elecsys PreciControl Troponin is equivalent to theElecsys PreciControl Troponin T (K082699).
(4) Elecsys Troponin T Gen 5 CalCheck 5 is equivalent tothe Elecsys CA 15-3 II CalCheck 5 (K122242).
IntendedUse/Indications forUseImmunoassay for the in vitro quantitative determinationof cardiac troponin T (cTnT) in lithium heparin plasma.The immunoassay is intended to aid in the diagnosis ofmyocardial infarction. The electrochemiluminescence immunoassay "ECLIA"is intended for use on the cobas system analyzers. CalSet Troponin T Gen 5 STAT is used for calibratingthe quantitative Elecsys Troponin T Gen 5 STATimmunoassay on the cobas system analyzers. PreciControl Troponin is used for quality control of theElecsys Troponin I and Elecsys Troponin I STATimmunoassays on the Elecsys and cobas eimmunoassay analyzers.PreciControl Troponin is also used for quality control of theElecsys Troponin T Gen 5 STAT immunoassay on thecobas system analyzers. The Elecsys Troponin T Gen 5 CalCheck 5 is an assayedcontrol for use in the calibration verification and for use inthe verification of the assay range established by the ElecsysTroponin T Gen 5 reagent on the cobas system analyzers.
(1) The Elecsys Troponin T Gen 5 STAT Immunoassay isa two-step sandwich immunoassay on the cobas e411analyzer and a one-step process on the cobas e 601analyzer. The assay uses streptavidin-coated

{8}------------------------------------------------

Substantial Equivalence – Comparison

The following tables compare the Elecsys Troponin T Gen 5 STAT Immunoassay, Elecsys Troponin T Gen 5 STAT CalSet, PreciControl Troponin and Elecsys Troponin T Gen 5 CalCheck5 with their predicate devices.

Comparison of Immunoassays-Elecsys Troponin T STAT Gen 5 and Gen 4 Immunoassays

Immunoassay Comparison
FeatureElecsys Troponin T Gen 5 STATImmunoassayPredicate Device: Elecsys 4thGeneration Troponin T STATImmunoassay (K051752)
General Immunoassay Features
Intended Use/Indicationsfor UseImmunoassay for the in vitroquantitative determination of cardiactroponin T (cTnT) in lithium heparinplasma. The immunoassay isintended to aid in the diagnosis ofmyocardial infarction.The electrochemiluminescenceimmunoassay “ECLIA” is intendedfor use on the cobas system analyzers.Immunoassay for the in vitroquantitative determination of cTnT inhuman serum and plasma. TheElecsys cTnT immunoassay can beused as an aid in the differentialdiagnosis of acute coronary syndrometo identify necrosis, e.g. acutemyocardial infarction. The test isfurther indicated for the riskstratification of patients presentingwith acute coronary syndrome and forcardiac risk in patients with chronicrenal failure. The test may also beuseful for the selection of moreintensive therapy and intervention inpatients with elevated levels ofcardiac cTnT.The electrochemiluminescenceimmunoassay “ECLIA” is intendedfor use on Elecsys and cobas eimmunoassay analyzers.
ImmunoassayProtocolSandwich immunoassaySame
DetectionProtocolElectrochemiluminescentImmunoassaySame

{9}------------------------------------------------

Analytical Performance

Precision

Precision of the TnT Gen 5 STAT assay was evaluated on the cobas e 411 and cobas e 601 analyzer.

Methods:

Within-run precision (repeatability) and total imprecision (intermediate precision) were determined according to the CLSI Guideline EP5-A2.

The protocol consisted of testing different sets of plasma samples and controls in single determinations in four separate aliquots (divided in two runs per day) for 21 operating days (n=84). The measurements were performed on the cobas e 411 and cobas e 601 with three reagent lots, performing rackpack calibration according to instruction for use.

Results:

Precision goals were met on both instruments and are summarized below.

cobas e 411 analyzer
RepeatabilityIntermediate precision
Sample (Li heparin plasma)Mean ng/LSD ng/LCV %SD ng/LCV %
Human plasma 17.270.4085.60.74610.3
Human plasma 212.20.3733.10.7175.9
Human plasma 31521.430.92.201.4
Human plasma 4467338.20.81172.5
Human plasma 5934164.50.72622.8
PreciControl TN120.00.4522.30.7403.7
PreciControl TN2173915.50.936.32.1
cobas e 601 analyzer
RepeatabilityIntermediate precision
Sample (Li heparin plasma)Mean ng/LSD ng/LCV %SD ng/LCV %
Human plasma 17.420.2243.00.4736.4
Human plasma 213.50.2521.90.5584.1
Human plasma 31541.230.82.241.5
Human plasma 4483138.00.81242.6

{10}------------------------------------------------

cobas e 601 analyzer
Sample (Li heparin plasma)Mean ng/LRepeatabilityIntermediate precision
SD ng/LCV %SD ng/LCV %
Human plasma 5945562.70.72562.7
PreciControl TN124.20.2701.10.7743.2
PreciControl TN2197113.30.745.02.3

Troponin T value at intermediate precision of CV = 10%

In addition, we report the value where a CV of ± 10% can be achieved in our product insert. This was determined according to CLSI EP17-A2.

The accuracy goal was defined as intermediate precision equal to a CV of 10 % , using the CLSI EP5-A2 20 day protocol to estimate intermediate precision. Ten Li-Heparin plasma pools were prepared across the low-end of the measuring range of the assay. Data were collected with two lots over 21 days, two runs per day with two replicates per run. Estimates of mean and intermediate precision were calculated for each sample.

The functional relationship between CV and concentration is modeled according to the suggestion in EP17 as:

%CV = A*concentrationB

To simplify the fit of the data, the CV and the mean are log-transformed. “Log” here refers to the natural logarithm. After log-transformation a linear regression using least squares can be performed and the model looks like this:

$log(%CV) = A + B * log(concentration)$

With X = log(A)

Results are visually assessed and the goodness of fit of the data is evaluated.

The concentration where 10% CV is achieved is calculated by:

$$\text{concentration} = \exp(\frac{\log(\mathbf{100}) - \mathcal{A}}{\mathbf{E}})$$

{11}------------------------------------------------

Results:

AnalyzerLotABTroponin T [ng/L]intermediate precisionCV [%]
cobas e 4111705114.861-1.09110.410
cobas e 4111736784.407-1.1066.7010
cobas e 6011705113.944-1.0524.7610
cobas e 6011736783.780-1.0973.8510

Determination of Troponin T concentration at CV = 10% on cobas e 411 and cobas e 601

LoQ

LoQ/Functional Sensitivity

Methods:

LoO of the Troponin T Gen 5 STAT assay was determined according to CLSI EP17-A2.

LoO is defined as the lowest amount of analyte that can be quantitatively established with stated accuracy and stated experimental conditions. The LoQ was set as the lowest concentration of analyte which can be quantified with a CV (intermediate precision) of no more than 20%.

Limit of Quantitation as Functional Sensitivity:

Functional sensitivity has been used as a detection capability performance attribute for the TnTGen5 Assay. It represents the measurand concentration associated with a desired within-lab (intermediate) precision, based upon a precision profile experiment in the low-end region of the measuring interval. This performance attribute however, simply represents a limiting form of the limit of quantitation (LoQ) in which the acceptable accuracy goal is based solely upon a precision requirement. This is suggested by CLSI EP5-A2 for troponin.

The accuracy goal was defined as intermediate precision equal to 20 % CV, using the CLSI EP5-A2 20 day protocol to estimate intermediate precision.

Ten Li-Heparin plasma pools were prepared across the low-end region of the measuring interval and in addition two controls were included. Data were collected with two lots over 21 days, two runs per day with two replicates per run. Estimates of the mean and intermediate precision were calculated for each sample for each reagent lot.

Total CV is based on the total variance, calculated as the sum of the variance components from day, run and within run (2122 = 84 measurements). The square root from the total variance divided by the grand mean times 100 is the results for the total CV [%]. The calculation of the variance components is based on a strict hierarchical model with random factors according CLSI EP5-A2.

{12}------------------------------------------------

The functional relationship between CV and concentration is modeled according to the suggestion in EP17 as:

%CV = A*concentration$^B$

To simplify the fit of the data, the CV and the mean are log-transformed. "Log" here refers to the natural logarithm. After log-transformation a linear regression using least squares can be performed and the model looks like this:

log(%CV) = $A$+ $B$ * log(concentration) With $A$ = log(A)

Results are visually assessed and the goodness of fit of the data is evaluated.

The concentration where 20% CV is achieved is calculated by:

log(20) concentration = exp B

Specification: 20% CV (intermediate precision) at ≤ 6 ng/L

Results:

For visual illustration the graphs below show the relationship between the total CV (log transformed) and concentration (log transformed).

Relationship between the Total CV and Concentration on cobas e 411-Lot 170511 (left) and Lot 173678 (right)

Image /page/12/Figure/11 description: The image contains two scatter plots, each displaying the relationship between 'logmean' on the x-axis and 'logcv_interm' on the y-axis. Both plots show a negative correlation, where 'logcv_interm' decreases as 'logmean' increases. A red line of best fit is drawn through the data points in each plot, visually representing the trend. The 'logmean' values range from approximately 0.5 to 2.5, while the 'logcv_interm' values range from about 1.5 to 4.

{13}------------------------------------------------

LoQ Results for cobas e 411

AnalyzerLotABLimit of Quantitation[ng/L]CV[%]
cobas e 4111705114.861-1.0915.520
cobas e 4111736784.407-1.1063.620

Relationship between the Total CV and Concentration on the cobas e 601-Lot 170511 (left) and Lot 173678 (right)

Image /page/13/Figure/3 description: The image contains two scatter plots, each displaying a negative correlation between 'logmean' on the x-axis and 'logcv_interm' on the y-axis. Both plots feature a set of data points scattered around a downward-sloping trendline. The left plot shows 'logmean' ranging from approximately 0.8 to 2.7, while 'logcv_interm' ranges from 1 to 3, and the right plot shows 'logmean' ranging from approximately 0.3 to 2.7, while 'logcv_interm' ranges from 1 to 3.5.

LoQ Results for cobas e 601

AnalyzerLotABLimit of Quantitation [ng/L]CV [%]
cobas e 6011705113.944-1.0522.520
cobas e 6011736783.780-1.0972.020

LoQ was determined to be 5.5 ng/L on cobas e 411 and 2.5 ng/L on the cobas e 601 analyzer and meets the specification. LoQ for the assay will be labeled as 6 ng/L.

Linearity

Methods:

The linearity of the Troponin T Gen 5 STAT assay was assessed on the cobas e 411 and cobas e 601 immunoassay analyzer by diluting one high analyte plasma sample (spiked with recombinant TnT) with native low analyte plasma. 21 concentrations (19

{14}------------------------------------------------

dilutions) throughout the measuring range were prepared. The dilutions were measured in three-fold determination within a single run using one lot.

In the first step, a linearity check was performed with a first order (linear) regression analysis and then with higher order models (quadratic and cubic). Data were analyzed using linear, quadratic and cubic order least square regression analysis according to CLSI protocol EP6-A:

y = a + b1 * x (first order polynomial or linear fit) y = a + b1 * x + b2 * x$^2$ (second order polynomial fit [quadratic]) y = a + b1 * x + b2 * x$^2$ + b3 * x$^3$ (third order polynomial fit[cubic])

Results:

The nonlinear coefficient b2 in the second order model (polynomial fit [quadratic]) is significant at the 5% level. Therefore it is used for the calculation of deviation from linearity. For cobas e 411, the test results did not deviate from linearity by more than 6.3%. For cobas e 601, the test results did not deviate from linearity by more than 12.4%.

Linearity was confirmed in the overall range from 3.19 ng/L to 10,439 ng/L. The labeled measuring range is 6 – 10000 ng/L.

High Dose Hook Effect

Methods:

The high dose hook effect of the Troponin T Gen 5 STAT assay was assessed on the cobas e 411 and cobas e 601 immunoassay analyzer.

Two samples were spiked with analyte to high TnT concentrations. For each sample a dilution series was performed using a low TnT concentration Li-Hep plasma.

The hook concentration reported corresponds to the analyte concentration with a signal corresponding to at least 10% above the highest master calibrator.

Results:

No hook effect was seen up to 100,000 ng/L

Endogenous Interferences

Methods:

The effect on quantitation of analyte in the presence of endogenous interfering substances using the Troponin T Gen 5 STAT was determined on the cobas e 411 and cobas e 601 immunoassay analyzers for the following 7 interfering substances. Intralipid, biotin, bilirubin, hemoglobin, rheumatoid factors, cholesterol and human serum albumin using three native or spiked (for Bilirubin and Lipemia interference) plasma samples (one low, one medium, and one high concentration of Troponin T) to prepare dilution series that were tested with one reagent lot.

{15}------------------------------------------------

Results:

Recoveries of samples containing interferents were within ± 10% of the values measured in samples not containing interferents.

Labeling states the following:

Interfering SubstanceNo interference seen up to:
Bilirubin25 mg/dL
Biotin20 ng/mL
Lipemia (Intralipid®)1500 mg/dL
Rheumatoid Factors900 IU/mL
Hemoglobin0.1 g/dL
Human Serum Albumin7 g/dL
Cholesterol310 mg/dL
HAMA (see below)322 µg/mL

HAMA Interference

The effect of the presence of human anti-mouse-antibodies on the TnT Gen 5 STAT assay was assessed on the cobas e 411 and cobas e 601 analyzers. Samples with three different levels of TnT were spiked with potential interfering HAMA and tested in duplicate. The HAMA serum used for the interference testing has been tested in an external laboratory with a commercial assay yielding a HAMA concentration of 805 ug/L.

Results:

Interference of approximately 10% was seen with HAMA concentrations > 322 µg/L.

Analytical Specificity/Cross Reactivity

Methods:

The analytical specificity of the Troponin T Gen 5 STAT assay was assessed on the cobas e 411 and cobas e 601 analyzers using three concentrations of troponin in plasma samples spiked with the potential cross-reacting compounds listed below. The spiked and non-spiked samples were tested in singlicate on cobas e 411 and cobas e 601 analyzer.

Note: Human cardiac and skeletal TNC are identical proteins.

Results:

Recoveries of samples containing the potential cross-reactants were within ± 10% of the values measured in samples not containing cross-reactants.

{16}------------------------------------------------

InterferingNo interference seen uptoMethod Sheet claim
Substance
Skeletal muscle TnT60000 ng/L10000 ng/L
Skeletal muscle TnI100000 ng/L100000 ng/L
Cardiac TnI100000 ng/L10000 ng/L
Human TnC80000 ng/L80000 ng/L

Summary of Cross-Reactivity Study

Exogenous Interference Drugs

Methods:

16 common pharmaceutical compounds were spiked into plasma samples. The analyte concentrations were approximately 15 to 9800 ng/mL. The drug concentrations tested are according to the recommendation (if available) given in the CLSI guideline EP7-A. When concentrations are not given in the guideline, we used at least 3-times the maximum recommended dosage.

In addition 18 cardiac drugs were assessed by spiking into human plasma samples. The analyte concentrations were approximately 15 and 1900 ng/L.

Plasma samples were divided into aliquots and spiked with the potential interferents. The reference sample without interferent was spiked with the respective amount of solvent only.

Testing was performed in 3-fold determination with one reagent lot in one run on one cobas e 411 and cobas e 601 analyzer.

{17}------------------------------------------------

Active agentConcentration[mg/L]
Acetylcysteine1660
Ampicillin-Na1000
Ascorbic acid300
Cyclosporine5
Cefoxitin2500
Heparin5000 U
Levodopa20
Methyldopa +1.520
Metronidazole200
Phenylbutazone400
Doxycycline50
Acetylsalicylic Acid1000
Rifampicin60
Acetaminophen200
Ibuprofen500
Theophylline100

Concentrations of Cardiac Drugs Tested

Active agentConcentration [mg/L]
Carvedilol37.5
Clopidogrel75
Digoxin0.25
Epinephrine0.5
Insulin1.6
Lidocaine80
Lisinopril10
Methylprednisolone7.5
Metoprolol150
Nifedipine30
Phenprocoumon3
Propafenone300
Reteplase33.3
Simvastatin30
Spironolactone75
Tolbutamide1500
Torasemide15
Verapamil240

Results:

Recoveries of samples containing the drugs were within ± 10% of the values measured in samples not containing the drugs.

{18}------------------------------------------------

Sample Stability

Methods:

The sample stability was performed in two different studies: Study 1. Sample stability at 2-8°C Study 2. Sample stability at -20°C (±5°C)

Study 1 - Sample stability at 2-8°C

Nine samples of Li-Heparin Plasma were aliquoted and measured fresh and after storage at 2-8℃ for 9 and 25 hours. Measurements were performed with threefold determination on the cobas e 601 analyzer and recovery was calculated.

Study 2 - Sample stability at -20℃ (±5℃):

Ten samples of Li-Heparin Plasma were aliquoted and measured fresh and after storage at - 20℃ (±5℃) for 13 months. Measurements were performed with threefold determination on the cobas e 601 analyzer and recovery was calculated.

Results:

All stressed samples recovered within ± 10% of fresh samples.

The following information will be provided in the method sheet: Stable for 24 hours at 2-8 °C, 12 months at -20 °C (±5°C). Freeze only once.

Calibration Stability

Methods:

The calibration stability on the cobas e 411 and cobas e 601 immunoassay analyzer for the Troponin T Gen 5 STAT assay was determined by reading the recovery of controls and plasma samples of a day 1 calibration curve.

Onboard Calibration Frequency- Study 1:

Calibration is recommended every 7 days if kit is not consumed. Three plasma samples and two control levels were tested with reagents opened on day 1 and kept at 22 ℃ (on board) between test points. Each sample was tested with two-fold determination.

Lot calibration - Study 2:

Calibration is recommended every 12 weeks with the same reagent lot if the reagent pack is consumed within 7 days. Three human plasma samples and two control levels were tested with fresh reagents kits of the same lot after up to 13 weeks. Each sample was tested with two-fold determination.

{19}------------------------------------------------

Results:

SpecificationResults
Onboard Calibration FrequencyStudy 17 days8 days
Lot calibrationStudy 212 weeks13 weeks

Calibration frequency will be reported as:

  • After 12 weeks when using the same reagent lot .
  • . After 7 days when using the same reagent kit

Reagent Stability

Methods:

Reagent stability for the Elecsys Troponin T Gen 5 STAT assay was determined on a cobas e 411 and cobas e 601 analyzers.

The reagent stability was performed in four different studies:

Study 1 - Reagent stability after first opening

Study 2 - On-board reagent stability

Study 3 - Stress stability

Study 4 - Shelf life stability

Results:

SpecificationResult
Reagent stability after first opening12 weeks13 weeks
On-board reagent stability4 weeks5 weeks
Stress stability3 weeks3 weeks
Shelf life stabilityn/a19 months

Reagent stability will be reported as:

  • Unopened at 2-8°C: Up to the stated expiration date ●
  • After opening at 2-8°C: 12 weeks ●
  • On the analyzers: 4 weeks ●

{20}------------------------------------------------

Troponin T Gen 5 STAT Clinical Performance

Expected Values: Healthy Reference Cohort

A total of 1301 healthy subjects (656 females, 645 males), were enrolled and included in the testing for determining the 99th percentile upper reference limit of a normal US population (age range 21 to 89 years) using the Elecsys Troponin T Gen 5 STAT assay. in lithium heparin samples on both the cobas e 411 and the cobas e 601 analyzers. The 99th percentile upper reference limits were determined to be:

  • 19 ng/L for both genders (n = 1301)
  • 14 ng/L for females (n = 656)
  • 22 ng/L for males (n = 645)

The Universal Definition of AMI takes into consideration the ESC/ACC/AHA/WHF definition. These guidelines recommend the detection of at least one value above the 99th percentile and a rise and/or fall of cardiac troponin in the clinical setting of acute myocardial ischemia. 1

Diagnostic Sensitivity and Specificity

APACE Population

The Advantageous Predictors of Acute Coronary Syndromes Evaluation (APACE) study is an international, multicentric prospective trial of acute chest pain patients that is currently continuing enrollment (ClinicalTrials.gov number NCT00470587.6). The study sites enrolled all patients who presented to the emergency department with symptoms of chest pain and angina pectoris. Peak of symptoms had to have occurred within the last 12 hours (onset of symptoms reported ranged from 0 to 72 hours). The only exclusion criterion was kidney failure that required dialysis. Diagnosis of MI was done through an independent adjudication committee which included cardiologists. This included 60 day follow-up information on each subject. In the case of a disagreement, a third independent cardiologist was used as the tie breaker. The subjects were diagnosed with acute MI by using the diagnostic criteria described in the ACC/ESC/AHA guidelines reference including ECG changes, symptoms characteristic for ischemia and elevations of cardiac troponin. All 718 subjects had a baseline test result. Five hundred fifteen (515) of the 718 subjects had a second test result at the 3 hour time point and 310 (of the 718) subjects had a test result at the 6 hour time point.

Results:

The clinical performance (clinical sensitivity, clinical specificity, positive predictive value and negative predictive value) of the Elecsys Troponin T Gen 5 STAT assay in the diagnosis of MI in this trial is shown below using a single 99th percentile cutoff 19 ng/L for all patients:

{21}------------------------------------------------

SensitivitySpecificityPPVNPV
Timepoint%95%CI%95%CI%95%CI%95%CI
Baseline93.5(115/123)87.6-97.286.4(514/595)83.4-89.058.7(115/196)51.4-65.698.5(514/522)97.0-99.3
3 hour98.3(59/60)91.1-10085.1(387/455)81.4-88.246.5(59/127)37.6-55.599.7(387/388)98.6-100
6hours100(37/37)90.5-10082.4(225/273)77.4-86.743.5 (37 / 85)32.8-54.7100(225/225)98.4-100

All subjects using 19ng/L cut-off

The clinical performance using sex specific cutoffs 14 ng/L for women and 22 ng/L for males is provided below:

SensitivitySpecificityPPVNPV
Timepoint%95%CI%95%CI%95%CI%95%CI
Baseline97.1(34/35)85.1-99.977.4(164/212)71.1-82.841.5(34/82)30.7-52.999.4(164/165)96.7-100
3 hour100(17/17)80.5-10075.2(124/165)67.8-81.529.3(17/58)18.1-42.7100(124/124)97.1-100
6hours100(15/15)78.2-10072.3(68/94)62.2-81.136.6(15/41)22.1-53.1100(68/68)94.7-100

Females using 14ng/L cut-off

The positive predictive value for females using the lower sex-specific cutoff (14 ng/L) is lower when compared to the higher cutoff of 19 ng/L. When looking at the lower bound of the 95 % CI, up to 69 %, 82 % and 78 % of positive test results for females are non-MI. Troponin results should always be used in conjunction with clinical signs and symptoms.

These observations underline the Universal AMI guideline requirements to use troponin results always in conjunction with at least one of the following criteria: symptoms of ischemia, ECG changes (ST and/or Q wave), left bundle branch block, imaging evidence of viable myocardium loss, wall motion abnormality or intracoronary thrombus to clarify the origin of myocardial injury.

SensitivitySpecificityPPVNPV
Timepoint%95%CI%95%CI%95%CI%95%CI
Baseline90.9(80/88)82.9-96.089.3(342/383)85.8-92.266.1(80/121)57.0-74.597.7(342/350)95.5-99.0
3 hour97.7(42/43)87.7-99.986.9(252/290)82.5-90.652.5(42/80)41.0-63.899.6(252/253)97.8-100
6hours100(22/22)84.6-10086.0(154/179)80.1-90.846.8(22/47)32.1-61.9100(154/154)97.6-100

Males using 22 ng/L cut-off

{22}------------------------------------------------

In a second multicenter study, a total of 1679 subjects presenting emergently with chest pain were enrolled. The trial excluded chest pain subjects with an MI within the last 3 months, subjects with surgery or hospitalization within the last 3 months, subjects with revascularization or percutaneous coronary intervention (PCI) within the last 3 months, subjects with an established acute non-cardiac primary illness and subjects transferred from another hospital or facility. These excluded subjects could be expected to have elevated troponin concentrations that would likely reflect cardiac comorbidities besides MI, and yield positive results; therefore, the specificity and positive predictive value may be overestimated. Within this population, there were 173 adjudicated MIs. 1679 of these subjects were evaluated on the cobas e 411 analyzer and 1675 subjects were evaluated on the cobas e 601 analyzer. Final diagnoses were determined by an independent adjudication committee which included cardiologists and emergency medicine physicians using the universal guidelines.

The clinical performance of the Elecsys Troponin T Gen 5 STAT assay in the diagnosis of MI in this trial is shown below using a single 99th percentile cutoff (i.e., 19 ng/L) for all patients:

Number of myocardial infarctions based on adjudicated diagnosis
cobas e 411 analyzercobas e 601 analyzer
MInon-MITotalMInon-MITotal
N17315061679N17315021675
%10.389.7100%10.389.7100
Clinical performance of single 99th percentile cutoff (19 ng/L) for aid in diagnosis of AMI in both genders
InstrumentTimepointnSensb) %95 % CIc)Specd) %95 % CIPPVe) %95 % CINPVf) %95 % CI
cobas e411analyzerBase- line162886.7 % (80.5 -91.5)87.8 % (86.0 -89.5)44.5 % (39.0 -50.2)98.3 % (97.5 -98.9)
3 hours142994.4 % (89.2 -97.5)87.4 % (85.5 -89.2)45.3 % (39.5 -51.1)99.3 % (98.6 -99.7)
6-9 hours117894.2 % (88.9 -97.5)85.3 % (83.0 -87.4)45.9 % (40.0 -51.9)99.1 % (98.2 -99.6)
12-24hours88792.9 % (86.4 -96.9)81.7 % (78.8 -84.3)42.3 % (36.0 -48.7)98.8 % (97.6 -99.5)
cobas e601analyzerBase- line160086.0 % (79.7 -90.9)88.0 % (86.2 -89.6)44.9 % (39.3 -50.6)98.2 % (97.3 -98.9)
3 hours141594.3 % (89.1 -97.5)86.6 % (84.6 -88.4)43.6 % (37.9 -49.4)99.3 % (98.6 -99.7)
6-9 hours115894.9 % (89.8 -99.9)85.3 % (83.0 -87.4)46.6 % (40.7 -52.6)99.2 % (98.4 -99.9)

{23}------------------------------------------------

Clinical performance of single 99th percentile cutoff (19 ng/L) for aid in diagnosis of AMI in both genders
InstrumentTimepointnSensb) %95 % CIc)Specd) %95 % CIPPVe) %95 % CINPVf) %95 % CI
97.9)87.4)52.6)99.7)
12-24hours87291.9 % (85.2 -96.2)80.6 % (77.6 -83.3)40.8 % (34.6 -47.2)98.6 % (97.3 -99.3)

b) Sensitivity = 100xA/A+C

c) CI = confidence interval

d) Specificity = 100xD/B+D

e) Positive predictive value = 100xA/A+B

f) Negative predictive value = 100xD/D+C

Diagnosis
MINon-MI
cTnT positiveAB
cTnT negativeCD
Clinical performance of gender-specific 99th percentile cutoff (14 ng/L) for aid in diagnosis of AMI infemales
InstrumentTimepointnSens %95 % CISpec %95 % CIPPV %95 % CINPV %95 % CI
cobas e411analyzerBase- line78787.3 (76.5-94.4)87.6 (84.9-89.9)37.9 (30.0-46.4)98.8 (97.6-99.5)
3 hours68792.0 (80.0-97.8)86.2 (83.3-88.8)34.3 (26.3-43.0)99.3 (98.2-99.8)
6-9 hours55391.7 (80.0-97.7)85.1 (81.7-88.1)37.0 (28.3-46.3)99.1 (97.7-99.7)
12-24hours41092.3 (79.1-98.4)79.8 (75.3-83.8)32.4 (23.9-42.0)99.0 (97.1-99.8)
cobas e601analyzerBase- line77185.7 (74.6-93.3)88.1 (85.5-90.4)39.1 (30.9-47.8)98.6 (97.3-99.3)
3 hours68291.8 (80.4-97.7)86.9 (84.0-89.4)35.2 (26.9-44.1)99.3 (98.2-99.8)
6-9 hours53691.3 (79.2-97.6)86.5 (83.2-89.4)38.9 (29.7-48.7)99.1 (97.6-99.7)
12-24hours39992.3 (79.1-98.4)81.4 (77.0-85.3)35.0 (25.8-45.0)99.0 (97.1-99.8)

{24}------------------------------------------------

The positive predictive value for females using the lower sex-specific cutoff (14 ng/L) is lower when compared to the higher cutoff of 19 ng/L. When looking at the lower bound of the 95 % CI, up to 69 %, 82 % and 78 % of positive test results for females are non-MI. Troponin results should always be used in conjunction with clinical signs and symptoms.

These observations underline the Universal AMI guideline requirements to use troponin results always in conjunction with at least one of the following criteria: symptoms of ischemia, ECG changes (ST and/or Q wave), left bundle branch block, imaging evidence of viable myocardium loss, wall motion abnormality or intracoronary thrombus to clarify the origin of myocardial injury.

Clinical performance of gender-specific 99th percentile cutoff (22 ng/L) for aid in diagnosis of AMI in males
InstrumentTimepointnSens %95 % CISpec %95 % CIPPV %95 % CINPV %95 % CI
cobas e411analyzerBase- line84186.3 (78.0-92.3)87.3 (84.7-89.6)48.4 (40.9-55.9)97.9 (96.5-98.8)
3 hours74295.7 (89.2-98.8)85.7 (82.8-88.3)48.6 (41.1-56.1)99.3 (98.2-99.8)
6-9 hours62593.3 (86.1-97.5)82.1 (78.5-85.2)46.7 (39.2-54.2)98.7 (97.1-99.5)
12-24hours47794.5 (86.6-98.5)80.2 (76.0-84.0)46.3 (38.1-54.7)98.8 (96.9-99.7)
cobas e601analyzerBase- line82985.1 (76.7-91.4)87.2 (84.6-89.6)48.0 (40.5-55.6)97.7 (96.2-98.7)
3 hours73395.6 (89.1-98.8)86.3 (83.4-88.9)49.7 (42.1-57.4)99.3 (98.2-99.8)
6-9 hours62293.5 (86.3-97.6)82.3 (78.7-85.4)47.8 (40.3-55.3)98.6 (97.1-99.5)
12-24hours47394.4 (86.4-98.5)80.0 (75.8-83.9)45.9 (37.7-54.3)98.8 (96.9-99.7)

{25}------------------------------------------------

Troponin T Gen 5 CalSet

CalSet Comparison
FeatureElecsys Troponin T Gen 5 STATCalSetPredicate Device:CalSet for the Elecsys 4thGeneration Troponin T STATImmunoassay (K961500)
General Features
Intended Use/Indicationsfor UseCalSet Troponin T Gen 5 STAT isused for calibrating the quantitativeElecsys Troponin T Gen 5 STATimmunoassay on the cobas systemanalyzers.Used for calibrating the quantitativeElecsys Troponin T STATimmunoassay on the Elecsys andcobas e immunoassay analyzers.
LevelsTwoSame
MatrixHuman serumSame
FormatLyophilizedSame

CalSet Stability

Methods:

The stability for the Elecsys Troponin T CalSet was determined by calculating the recovery of stressed calibrators compared to unstressed/freshly reconstituted calibrators.

Study 1: Combined On Board Stability/ Stability after reconstitution:

Stability of reconstituted Elecsys Troponin T Gen 5 STAT CalSet onboard is claimed to be up to 5 hours at 20 – 25°C within two weeks after reconstitution when stored at 2 – 8°C.

CalSets were reconstituted and kept at 2 – 8°C for three weeks. During this time period the opened calibrators were incubated at 25°C (±2°C) for a total of 6 hours.

Study 2: Stability at - 20°C (±5°C)

Stability of Elecsys Troponin T Gen 5 STAT CalSet at -20°C (±5°C) is claimed to be 3 months. Calibrators should be frozen for one time only and also be used only once after thawing.

CalSets were reconstituted and stored at -20°C (±5°C) for four months.

The TnT concentration of six human plasma samples and two controls was determined using the calibration curve generated by the stressed calibrators and compared to the TnT concentration determined by using the calibration curve generated by the freshly reconstituted calibrators (reference).

Acceptance criteria:

Concentrations of $<$ 14 ng/L: Recovery ± 1.4 ng/L Concentrations of ≥ 14 ng/L: Recovery 100 ± 10%

{26}------------------------------------------------

Results:

On board stability for the CalSet on the cobas e 411 analyzer (25°C (±2°C) was determined to be up to 5 hours. For the CalSet on the cobas e 601 analyzers, the labeling states to use only once. CalSets stored at -20°C (±5°C) are stable for three months.

CalSet Real-Time Stability

Methods:

Real-time stability of Troponin T Gen 5 STAT CalSet was evaluated as follows:

In the real-time stability study, the Troponin T Gen 5 STAT CalSet material was stored at 2 - 8°C. The stored assay reagents were tested at time point t = 0 and at specified intervals over the shelf life of the device up to the planned shelf life plus one month. Testing was performed using PreciControl Troponin 1 and PreciControl Troponin 2 as well as of internal manufactured control samples.

Data for the time-points tested are available in duplicate. The average on-test recovery value was calculated as the percent recovery compared to the reference value (assigned value of PreciControl Troponin 1 and PreciControl Troponin 2 and assigned value of internal manufactured control samples).

Sample TypeRecovery of the Reference Value
PreciControl Troponin 190-110%
PreciControl Troponin 290-110%
Internal Control samples(concentrations ≤ 100 ng/L)90-110%
Internal Control samples(concentrations > 100 ng/L)90-110%

Acceptance Criteria of CalSet Real-Time Stability Testing:

Results:

The Troponin T Gen 5 STAT CalSet is stable for at least 19 months. A shelf-life claim of 18 months has been established.

{27}------------------------------------------------

PreciControl Troponin

PreciControl Comparison
FeatureElecsys PreciControl TroponinPredicate Device:Elecsys PreciControl Troponin T(K082699)
General Features
Intended Use/Indicationsfor UsePreciControl Troponin is used forquality control of the ElecsysTroponin I and Elecsys Troponin ISTAT immunoassays on the Elecsysand cobas e immunoassay analyzers.PreciControl Troponin is also used forquality control of the Elecsys Troponin TGen 5 STAT immunoassay on the cobassystem analyzers.Elecsys PreciControl cTnT is used forquality control of the ElecsysTroponin T (CARDIAC T)immunoassay on the Elecsys andcobas e immunoassay analyzers.
LevelsTwoSame
FormatLyophilizedSame
MatrixHuman serumSame

PreciControl In-Use Stability

Methods:

The stability for the PreciControl Troponin was determined by calculating the recovery based on the use of stressed PreciControl Troponin 1 and 2 compared to unstressed freshly reconstituted PreciControl Troponin 1 and 2.

Study 1 - Combined in-use stability: On board stability and stability of reconstituted PreciControl Troponin at 2-8°C

Stability of reconstituted PC Troponin onboard is claimed to be up to 5 hours at 20 – 25°C and is claimed to be 4 days when stored at 2 - 8°C.

PreciControl 1 and 2 were reconstituted and kept at 2 - 8°C for 97 h. During this time period the opened controls were incubated at 25°C (±2°C) for a total of 6 hours (6 x 1hour).

Study 2 - Stability of reconstituted PreciControls stored at - 20°C (±5°C):

PC Troponin 1 and 2 were reconstituted and stored at -20°C (±5°C) for 94 days.

{28}------------------------------------------------

Acceptance criteria:

Concentrations of < 14 ng/L: Recovery ± 1.4 ng/L Concentrations of > 14 ng/L: Recovery 100 ± 10%

Results:

Stability of PreciControl Troponin at -20°C is claimed to be 3 months. Controls should be frozen for one time only and also be used only once after thawing.

PreciControl Real-Time Stability

Methods:

In the real-time stability study, the PreciControl Troponin test material is stored at 2-8°C. The PreciControls were tested in duplicate at specified intervals over the shelf life of the device up to the planned shelf life plus one month (19 months). The average ontest recovery value was calculated as the percent recovery compared to the reference value (Assigned value of PreciControl Troponin 1 and PreciControl Troponin 2

Acceptance Criteria of PreciControl real-time stability testing:

Sample TypeRecovery of the Reference Value
PreciControl Troponin 190 - 110%
PreciControl Troponin 290 - 110%

Results:

The shelf life claim is 18 months.

PreciControl Reconstitution Study

PreciControl Troponin was reconstituted for 60 minutes (reference) and 120 minutes. Samples were evaluated in duplicate on the cobas e 411 analyzer. The average recovery after 120 minutes of reconstitution was calculated as percent recovery compared to the value obtained at 60 minutes of reconstitution (the reference value).

The acceptance criterion is recovery of 74-126% for PreciControl Troponin 1 and 81-119% for PreciControl Troponin 2 of the value obtained for the 60 minute reconstituted material.

The data support the package insert claim that the PreciControl Troponin is completely reconstituted after 60 minutes.

{29}------------------------------------------------

Troponin T Gen 5 CalCheck5

CalCheck Comparison
Elecsys Troponin T Gen 5CalCheck 5Predicate Device:Elecsys CA 15-3 II CalCheck 5(K122242)
General Immunoassay Features
Intended Use/Indications forUseThe Elecsys Troponin T Gen 5 CalCheck5 is an assayed control for use in thecalibration verification and for use in theverification of the assay rangeestablished by the Elecsys Troponin TGen 5 reagent on the cobas systemanalyzers.The Elecsys CA 15-3 II CalCheck 5is an assayed control for use incalibration verification and for use inthe verification of the assay rangeestablished by the Elecsys CA 15-3 IIreagent on the indicated Elecsys andcobas e immunoassay analyzers. Forin vitro diagnostic use only.
LevelsFiveSame
FormatLyophilizedSame
MatrixHuman serumEquine serum

Troponin T Gen 5 CalCheck 5 Stability

Methods:

Open vial stability was performed on the cobas e 411 and cobas e 601 in order to verify the claims for the Troponin T Gen 5 CalCheck 5.

Study 1 - Open-Vial Stability

The on-test and reference materials were tested in duplicate. The on-test material was reconstituted and stored for 5 hours at 25℃ (in an open vial). The reference material was a freshly reconstituted set of CalChecks. The on-test recovery was calculated as a percent of the reference value.

One Troponin T Gen 5 CalCheck 5 lot was evaluated in duplicate.

Acceptance criteria:

CalCheck Level 1: ≤ 6 ng/L CalCheck Level 2-5: recovery of 90-110% of the reference value.

Results:

The data support the package insert claim that reconstituted Troponin T Gen 5 CalCheck 5 is stable up to 4 hours at 20-25°C.

The CalCheck products are not stored on-board the analyzers, therefore no on-board stability claims are made.

Study 2 - Real-time stability

{30}------------------------------------------------

In the real-time stability study, the Troponin T Gen 5 CalCheck 5 test material is stored at 2-8°C. The CalChecks are tested at T=0 and at intervals over the shelf life of the device up to the planned shelf life plus at least one month. The target shelf life claim for the Troponin T Gen 5 CalCheck 5 is 18 months.

For the lot on stability, data for the time-points 0, 13, 19, 25 and 37 months tested in duplicate will be available. The average on-test recovery value will be calculated as percent recovery compared to the T=0 reference value.

Acceptance criteria:

CalCheck Level 1: < 6 ng/L CalCheck Level 2-5: recovery of 80-120% of the reference value.

The data will support the package insert claim that unopened Troponin T Gen 5 CalCheck 5 is stable up to the expiration date when stored at 2-8°C.

Results:

The labeled shelf life claim is 18 months.

Troponin T Gen 5 STAT Test System -- Conclusion

The Elecsys Troponin T Gen 5 STAT Assay, Elecsys Troponin T Gen 5 STAT CalSet, Elecsys PreciControl Troponin and Elecsvs Troponin T Gen 5 CalCheck 5 substantially equivalent to the Elecsys Gen 4 STAT assay, Elecsys Gen 4 STAT CalSet, Elecsys PreciControl Troponin and Elecsys CalCheck 5 products currently on the market. The clinical and nonclinical data provided in this submission demonstrate that the safety and efficacy of these devices are equivalent to their predicates.

§ 862.1215 Creatine phosphokinase/creatine kinase or isoenzymes test system.

(a)
Identification. A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.(b)
Classification. Class II.