K Number
K190632

Validate with FDA (Live)

Date Cleared
2019-10-24

(226 days)

Product Code
Regulation Number
878.4460
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination

Device Description

The proposed device, Latex Surgeon's Gloves Powder Free with protein content labeling claim of 50 µg/dm² or less per glove of extractable protein is a sterile and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound fromcontamination. The proposed device is made of' natural rubber latex, as per standard ASTM D35 77-09(2015). The classification is: Type I - gloves compounded primarily from natural rubber latex". The proposed device is Powder Free Latex Surgeon's Gloves, and variants of different sizes. All variants share the same color, creamy, white. The proposed device is sterilized either using Ethylene Oxide Sterilization or Gamma irradiation method to achieve the Sterility Assurance Level (SAL) of 10-6 and place in a sterility maintenance package to ensure a shelf life of 3 years.

AI/ML Overview

This document is a 510(k) Premarket Notification for Latex Surgeon's Gloves Powder Free with protein content labeling claim of 50 µg/dm² or less per glove of extractable protein.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

CriteriaSpecification as per ASTM D3577-09 StandardAverage Value of Subject Device (Reported Performance)Device Complied?
Length
Size 5½Min 265mm282mmYes
Size 6Min 265mm282mmYes
Size 6½Min 265mm282mmYes
Size 7Min 265mm282mmYes
Size 7½Min 265mm282mmYes
Size 8Min 265mm282mmYes
Size 8½Min 265mm282mmYes
Size 9Min 265mm282mmYes
Width
Size 5½70 ± 6mm71mmYes
Size 676 ± 6mm78mmYes
Size 6½83 ± 6mm82mmYes
Size 789 ± 6mm90mmYes
Size 7½95 ± 6mm95mmYes
Size 8102 ± 6mm100mmYes
Size 8½108 ± 6mm106mmYes
Size 9114 ± 6mm112mmYes
Finger Thickness (All sizes)Min 0.10mm0.20mmYes
Palm Thickness (All sizes)Min 0.10mm0.16mmYes
Cuff Thickness (All sizes)Min 0.10mm0.13mmYes
Tensile Strength Before aging (All sizes)24Mpa minimum26.5MpaYes
Tensile Strength After aging (All sizes)18Mpa minimum21.0MpaYes
Ultimate Elongation Before aging (All sizes)750% minimum860%Yes
Ultimate Elongation After aging (All sizes)560% minimum725%Yes
Stress at 500% before aging (All sizes)5.5 MPa Max2.6 MpaYes
Pinhole AQL Before aging (All sizes)Max 1.51.0Yes
Pinhole AQL After aging (All sizes)Max 1.51.0Yes
Powder Residue (Powder-Free Glove)< 2 mg/Glove(Not explicitly stated in table, but compliance with ASTM D6124-06 implies meeting this)Yes
Protein Content50 µg/dm² or less per glove(Not explicitly stated in table, but compliance with ASTM D5712-15 and ASTM D6499-16 implies meeting this)Yes
Biocompatibility (Irritation, Sensitization, Acute Systemic Toxicity)Non-irritant, Non-Sensitizer and Non-Toxic(Not explicitly stated in table, but compliance with ISO 10993-10 implies meeting this)Yes
Sterility Assurance Level (SAL)10⁻⁶(Implied by sterilization methods; not in table)Yes

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample size for test set: Not explicitly stated for each test (length, width, thickness, tensile strength, elongation, stress, pinhole). However, "Average Value of Subject Device" generally indicates testing was done on multiple samples.
  • Data provenance:
    • Country of origin: Testing was conducted by Lenora Glove Pvt. Ltd., located in Coimbatore, Tamil Nadu, India.
    • Retrospective or prospective: The document describes "bench tests" performed to verify the device met specifications, which indicates prospective testing of the manufactured product.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This device is not an AI/diagnostic device that typically involves expert ground truth establishment in the same way. The "ground truth" for the performance characteristics comes from established international and national standards (e.g., ASTM D3577-09, ASTM D5151-06, ISO 10993-10). These standards define the acceptable range for physical and biological properties of latex surgeon's gloves. Therefore, there were no "experts" in the context of medical image interpretation for this type of device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As described above, the acceptance criteria are based on objective measurements against established engineering and medical device standards, not subjective interpretations requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a physical medical device (surgical glove), not an AI-assisted diagnostic tool or an imaging product. MRMC studies are not relevant here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device is defined by the acceptance criteria established in recognized consensus standards. These standards (e.g., ASTM, ISO) outline specific physical, mechanical, and biological properties that surgical gloves must meet for safety and effectiveness.

8. The sample size for the training set

Not applicable. This is a physical medical device, not an AI/machine learning model that requires a training set.

9. How the ground truth for the training set was established

Not applicable. No training set was used for this physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

October 24, 2019

Lenora Glove PVT LTD. % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Drive Ste 114 Aurora, Illinois 60504

Re: K190632

Trade/Device Name: Latex Surgeon's Gloves Powder Free with protein content labeling claim of 50ug/dm2 or less per glove of extractable protein Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I Product Code: KGO Dated: September 18, 2019 Received: September 24, 2019

Dear Manoj Zacharias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190632

Device Name

Latex Surgeon's Gloves Powder Free with protein content labeling claim of 50 µg/dm² or less per glove of extractable protein

Indications for Use (Describe)

A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

LENORA GLOVE PVT.LTD. SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVE POWDER FREE

510K SUMMARY as required by: 21CFR § 807.92

K190632

A. APPLICANT INFORMATION

1.Submitter NameLENORA GLOVE PVT. LTD.
Date SubmittedSeptember 18, 2019
AddressPlot No:15/104-1, Rottigoundanur,Thirumalayampalayam P.O., Coimbatore,Tamilnadu State, India-641 105
Phone+91 422 2656 443, 2656 941
Fax1 (815) 986-2632
E-mailqa@lenoraglove.com
Contact PersonANTONY KURIYAN
DesignationManaging Director
Contact Number+91 974700797
Contact Emailqa@lenoraglove.com

B. US AGENT & CONTACT PERSON INFORMATION

US agent & contact person nameManoj Zacharias
AddressLiberty Management Group Ltd.75 Executive Dr. STE 114, Aurora, IL-60504, USA.
Phone(630) 270-2921
Fax(815) 986-2632
E-mailmanoj@libertymanagement.us

C. DEVICE IDENTIFICATION

Common NameSurgeon's Glove
Device NameLatex Surgeon's Glove powder free withprotein content labeling claim of 50 µg/dm²or less per glove of extractable protein
Product proprietary or tradenameLatex Surgeon's Glove powder free withprotein content labeling claim of 50 µg/dm²or less per glove of extractable protein
Classification nameSurgeon's Glove
Device Classification1
Product CodeKGO
Regulation Number21 CFR 878.4460
Review PanelGen & Plastic Surgery

{4}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVE POWDER FREE

D.PREDICATE DEVICE INFORMATION

Name of devices510k Number510K Owner
Predicate device“PRISTEEN” Latex Surgeon'sGlove powder free withprotein content labeling claimof 50 $µg/dm^2$ or less perglove of extractable proteinK172942Beta Health Care productsPvt.Ltd,Plot No 21B,Cochin Special EconomicZone,Kakkanad, kerala,India-682037
Reference device“Medismart+” Latex Surgeon'sGlove powder free withprotein content labeling claimof 50 $µg/dm^2$ or less perglove of extractable proteinK151114St.Marys Rubbers Pvt.Ltd,Koovappally P.O,Kanjirappally, KottayamDistrict, Kerala State,India-686518

E. DESCRIPTION OF THE DEVICE

The proposed device, Latex Surgeon's Gloves Powder Free with protein content labeling claim of 50 µg/dm² or less per glove of extractable protein is a sterile and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound fromcontamination.

The proposed device is made of' natural rubber latex, as per standard ASTM D35 77-09(2015).

The classification is: Type I - gloves compounded primarily from natural rubber latex".

The proposed device is Powder Free Latex Surgeon's Gloves, and variants of different sizes.

All variants share the same color, creamy, white.

The proposed device is sterilized either using Ethylene Oxide Sterilization or Gamma irradiation method to achieve the Sterility Assurance Level (SAL) of 10 6and place in a sterility maintenance package to ensure a shelf life of 3 years.

F. INDICATION FOR USE

A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

{5}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVE POWDER FREE

G. TECHNOLOGICAL COMPARISON

a. General Characteristics Comparison

CharacteristicSubject deviceK190632Predicate deviceK172942Reference deviceK151114Comparison
Product CodeKGOKGOKGOSame
Regulation No.21 CFR 878.446021 CFR 878.446021 CFR 878.4460Same
Class111Same
Intended Use forLatex Surgeon'sGloves PowderFree, with proteincontent labelingclaim of50µg/dm² or lessper glove ofextractableprotein.A latex surgeon's glove is adevice made of naturalrubber intended to be wornby operating room personnelto protect a surgical woundfrom contamination.A latex surgeonSurgeons glove isa device made ofnatural rubberintended to beworn byoperating roompersonnel toprotect a surgicalwound fromcontaminationA latex surgeonSurgeons glove isa device made ofnatural rubberintended to beworn byoperating roompersonnel toprotect a surgicalwound fromcontaminationSimilar
Powdered orPowder freePowder freePowder freePowder freeSame
Method ofpowder freePolymer coatsamesameSame
Classification perASTMD3577-99Type I - gloves compoundedprimarily from naturalrubber latexType I - glovescompoundedprimarily fromnatural rubberlatexType I - glovescompoundedprimarily fromnatural rubberlatexSame
SterilizationETO/as well as Radiation,SAL- 10-6ETO/as well asRadiation,SAL- 10-6ETO/as well asRadiation,SAL- 10-6Same
Label andLabelingMeet FDA's RequirementsMeet FDA'sRequirementsMeet FDA'sRequirementsSame
Special labelclaimProtein content labelingclaim of 50 µg/dm² or lessper glove of extractableprotein for Latex Surgeon'sGloves Powder Free.SameSameSame
Type of useOver the counter useOver the counteruseOver the counteruseSame

{6}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVE POWDER FREE

CharacteristicsAcceptance criteria of the standard
Subject deviceK190632Predicate deviceK172942Reference deviceK151114Comparison
DimensionsASTM D3577-09((Reapproved 2015))ASTM D3577ASTM D3577Similar
PhysicalPropertiesASTM D3577-09(Reapproved 2015)ASTM D3577ASTM D3577Similar
Freedom fromHolesASTM D3577 andAST MD5151-06,(Reapproved 2015)ASTM D3577 andAST M D5151ASTM D3577 andAST M D5151Similar
Powder residuefor powder freegloveASTM D3577 and ASTMD6124-06, (Reapproved2017)Powder content< 2 mg/GloveASTM D3577 andASTM D6124,Powder content< 2 mg/GloveASTM D3577 andASTM D6124,Powder content< 2 mg/GloveSimilar
Protein ContentASTM D3577, ASTMD5712-15 and ASTMD6499-16ASTM D3577,ASTM D5712 andASTM D6499ASTM D3577,ASTM D5712 andASTM D6499Similar
Biocompatibility(Irritation,Sensitization,Acute SystemicToxicity)ISO 10993-10Non- irritant,Non-Sensitizer and Non-ToxicISO 10993-10ISO 10993-10Similar

Technological Characteristics Comparison b.

H. Summary of Non-Clinical Testing

Bench tests were conducted to verify that the proposed device met all design specifications as the predicate devices. The standards used in this submission and the test results demonstrated that the proposed device complies with the following standards:

  • . ASTM D3577-09 (2015):- Standard Specification for Rubber Surgical Gloves.
  • ASTM D 5151-06(2015):-Standard Test Method for Detection of Holes in . Medical Gloves.
  • ASTM D6124-06(2017):-Standard Test Method for Residual Powder on . Medical Gloves.

{7}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVE POWDER FREE

  • ASTM D5712-15:-Standard Test Method for the Analysis of Aqueous Extractable Protein in Natural Rubber and Its Products Using the Modified Lowry Method.
  • . ASTM D6499-18:-Standard Test Method for the Immunological Measurement of Antigenic Protein in Natural Rubber and Its Products.
  • . ASTM F 1929-2015:-Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration.
  • . ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
  • ISO 11137-1-2006/ (R) 2010 validation of sterilization process. ●
  • ISO 11137-2:2013, Sterilization of health care products - radiation - part 2: establishing the sterilization dose
  • 1S01137-2:2013: Sterilization of healthcare products-Radiation-Part2: Establishing the sterilization dose.
  • ISO 10993-7:2008:-Biological evaluation of medical devices --: Ethylene oxide ● sterilization residuals.
  • . ISO 11135-1:2014 :- Sterilization of healthcare products -- Ethylene oxide -Part 1: Requirements for development, validation and routine control of a sterilization process for medicaldevices.

A summary of the non-clinical testing results for the subject device is provided below. The standards used in this submission and the test results demonstrate that the proposed device meets the acceptance criteria for the referenced standards.

CRITERIASPECIFICATION ASAVERAGE VALUEWHETHER
PER ASTMD3577-09OF SUBJECTSUBJECT DEVICE
STANDARDDEVICECOMPLIED WITH
THE ASTMD3577 -
09 STANDARD
1Length ASTM D3577-09
Size 5'/2Min 265mm282 mmYes
Size 6Min 265mm282mmYes
Size 6'/2Min 265mm282mmYes
Size 7Min 265mm282mmYes

SIZES AVAILABLE: - 5'/2, 6, 6'/2, 7, 7'/2, 8, 8'/2, 9

{8}------------------------------------------------

LENORA GLOVE PVT.LTD.
SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'SGLOVE POWDER FREE
Size 7'/2Min 265mm282mmYes
Size 8Min 265mm282mmYes
Size 8'/2Min 265mm282mmYes
Size 9Min 265mm282mmYes
2Width ASTM D3577-09
Size 5'/270+/-6mm71mmYes
Size 676+/-6mm78mmYes
Size 6'/283+/-6mm82mmYes
Size 789+/-6mm90mmYes
Size 7'/295+/-6mm95mmYes
Size 8102+/-6mm100mmYes
Size 8'/2108+/-6mm106mmYes
Size 9114+/-6mm112mmYes
3FingerThickness(All sizes)Min 0.10mm0.20mmYes
4Palm Thickness(All sizes)Min 0.10mm0.16mmYes
5Cuff Thickness(All sizes)Min 0.10mm0.13mmYes

{9}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR LATEX SURGEON'S GLOVE POWDER FREE

CRITERIASPECIFICATION ASPER ASTMD3577-09STANDARDAVERAGEVALUE OFSUBJECTDEVICEWHETHER SUBJECTDEVICE COMPLIEDWITH THEASTMD3577 -09STANDARD
6Tensile Strength ASTM D3577-09
Before aging(All sizes)24Mpa minimum26.5MpaYes
After aging@70°±2C for166±2 hr(All sizes)18Mpa minimum21.0MpaYes
7Ultimate Elongation ASTM D3577-09
Before aging(All sizes)750% minimum860%Yes
After aging@70°±2C for166±2 hr(All sizes)560% minimum725%Yes
8Stress at 500%before ageing(All sizes)5.5 MPa Max2.6 MpaYes
9Pinhole AQL ASTM D 5151-06(2015)
Before aging(All sizes)Max 1.51.0Yes
After aging@70°C for 7 days(All sizes)Max 1.51.0Yes

I. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).