K Number
K190412

Validate with FDA (Live)

Date Cleared
2019-03-21

(28 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CLUNGENE® Multi-Drug Test Dip Card is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Opiates, Oxycodone, Secobarbital, Methadone, Buprenorphine, Methylenedioxymethamphetamine, Tricyclic Antidepressants, EDDP and Propoxyphene in human urine at the cutoff concentrations of:

Drug(Identifier)CalibratorCut-off level
Amphetamined-Amphetamine1000 ng/mL
OxazepamOxazepam300 ng/mL
CocaineBenzoylecgonine300 ng/mL
Marijuana11-Nor-△9-Tetrahydrocannabinol-9-COOH50 ng/mL
Methamphetamined-Methamphetamine1000 ng/mL
OpiatesMorphine2000 ng/mL
OxycodoneOxycodone100 ng/mL
SecobarbitalSecobarbital300 ng/mL
MethadoneMethadone300 ng/mL
BuprenorphineBuprenorphine10 ng/mL
MethylenedioxymethamphetamineD,L-Methylenedioxymethamphetamine500 ng/mL
PhencyclidinePhencyclidine25 ng/mL
Tricyclic AntidepressantsNortriptyline1000 ng/mL
EDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300 ng/mL
Propoxyphened-Propoxyphene300 ng/mL

Configuration of the CLUNGENE® Multi-Drug Test Dip Card can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Nortriptyline, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GCMS or LC/MS is the preferred confirmatory method.

For in vitro diagnostic use only.

CLUNGENE® Multi-Drug Test Easy Cup is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Opiates, Oxycodone, Secobarbital, Methadone, Buprenorphine, Phencyclidine, Methylenedioxymethamphetamine, Tricyclic Antidepressants, EDDP and Propoxyphene in human urine at the cutoff concentrations of:

Drug(Identifier)CalibratorCut-off level
Amphetamined-Amphetamine1000 ng/mL
OxazepamOxazepam300 ng/mL
CocaineBenzoylecgonine300 ng/mL
Marijuana11-Nor-△9-Tetrahydrocannabinol-9-COOH50 ng/mL
Methamphetamined-Methamphetamine1000 ng/mL
OpiatesMorphine2000 ng/mL
OxycodoneOxycodone100 ng/mL
SecobarbitalSecobarbital300 ng/mL
MethadoneMethadone300 ng/mL
BuprenorphineBuprenorphine10 ng/mL
MethylenedioxymethamphetamineD,L-Methylenedioxymethamphetamine500 ng/mL
PhencyclidinePhencyclidine25 ng/mL
Tricyclic AntidepressantsNortriptyline1000 ng/mL
EDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300 ng/mL
Propoxyphened-Propoxyphene300 ng/mL

Configuration of the CLUNGENE® Multi-Drug Test Easy Cup can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Nortriptyline, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

For in vitro diagnostic use only.

Device Description

The CLUNGENE Multi-Drug Test Dip Card and CLUNGENE Multi-Drug Test Easy Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Methadone, Buprenorphine, Phencyclidine, Methylenedioxymethamphetamine, Tricyclic Antidepressants, EDDP and Propoxyphene (target analytes) in human urine. The products are single-use in vitro diagnostic devices. The CLUNGENE Multi-Drug Test Dip Card kit contains a Dip Card device, a package insert and a urine cup for sample collection. The CLUNGENE Multi-Drug Test Easy Cup kit contains a Cup device, a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

AI/ML Overview

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" as a single, consolidated table with pass/fail metrics. However, the "Performance Characteristics" section details various studies with implied performance targets. The key performance demonstrated is in the precision studies and lay-user studies for detecting drug analytes in urine at specific cut-off levels.

Implied Acceptance Criteria (based on presented data patterns):

TestAcceptance Criteria (implied from data)Reported Device Performance
Precision (EDDP, Opiates, Propoxyphene) - Dip Card & Easy CupAll samples at -100%, -75%, -50%, -25% Cut-off should be Negative. All samples at +25%, +50%, +75%, +100% Cut-off should be Positive. Samples at Cut-off should show a mix of positive and negative results, indicating accurate determination around the cutoff.Met: For all three analytes and both device types, samples at -100% to -25% Cut-off were 100% negative (50-/0+ for each lot). Samples at +25% to +100% Cut-off were 100% positive (50+/0- for each lot). At the Cut-off concentration, results were mixed (e.g., Dip Card EDDP Lot 1: 27-/23+), demonstrating ability to differentiate around the cutoff.
InterferenceNo interference at a concentration of 100 µg/mL for listed compounds.Met: No differences observed between device types; comprehensive list of non-interfering compounds provided (e.g., Acetaminophen, Aspirin, Albumin, Hemoglobin).
Specificity (Cross-Reactivity)List of compounds and their cross-reactivity percentages. Implied: High specificity for the target analyte (100% at cutoff), quantifiable cross-reactivity for structurally related compounds.Met: High specificity shown for target analytes (e.g., EDDP 100% at 300 ng/mL, Morphine 100% at 2000 ng/mL, d-Propoxyphene 100% at 300 ng/mL). Cross-reactivity for related compounds is quantified at higher concentrations (e.g., Disopyramide 0.4% for EDDP, Thebaine 10% for Opiate).
Effect of Urine Specific Gravity and pHAll samples at or above +25% Cut-Off should be Positive. All samples at or below -25% Cut-Off should be Negative.Met: "Results were all positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off." No differences observed between device types.
Lay-User Study (Overall Performance)High percentage of correct results across all drug analytes and concentration levels. Instructions easily followed.Met: The table for the lay-user study shows high percentages of correct results across all drugs and concentration levels, frequently 100% for clear negatives and positives, and 95% for those near the +/-25% cutoff. All lay users indicated that instructions were easily followed.
Lay-User Study (Reading Level)Reading Grade Level should be appropriate for lay users.Met: Flesch-Kincaid reading analysis showed a Grade Level of 7, which is generally considered appropriate for a broad lay audience.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Set Sample Size:

    • Precision Studies: For EDDP, Opiates, and Propoxyphene, each concentration level (-100% to +100% cut-off) was tested with 50 samples for each of three lots, making a total of 450 tests per analyte per lot (50 samples * 9 concentrations) * 3 lots = 1350 tests per analyte. Since there are three analytes and two device types (Dip Card and Easy Cup), this would be 1350 tests * 3 analytes * 2 devices = 8100 individual tests for these three new analytes in the precision study.
    • Method Comparison Studies: For EDDP, Opiates, and Propoxyphene, 80 unaltered clinical samples (40 negative and 40 positive) were used for each drug. This means 80 samples * 3 drugs * 2 device types = 480 samples.
    • Lay-user Study: 310 lay persons were involved for each device format (Dip Card and Easy Cup). The total number of individual tests performed by lay users is not explicitly stated as a single number but would be 310 tests * 15 analytes * 2 device formats = 9300 tests (based on the "entire panel" evaluation). For each drug, a varying number of samples were prepared across different concentration levels, totaling 360 samples for Amphetamine, for example.
    • Interference and Specificity Studies: Samples were "spiked into negative urine" and tested using three lots of each device. Numerical sample sizes are not provided for these, but rather the list of compounds and their effects.
    • Effect of Urine Specific Gravity and pH: Urine samples were spiked with drugs at +/-25% Cut-Off levels and tested using three lots of each device. Numerical sample sizes are not provided.
  • Data Provenance: The document does not explicitly state the country of origin of the data for the experimental studies. However, the submitter, Hangzhou Clongene Biotech Co.,Ltd., is located in China, which might suggest the studies were conducted there. The clinical samples for the method comparison were "unaltered clinical samples." The lay-user study was performed "at three intended user sites."

  • Retrospective or Prospective: Not explicitly stated, but the precision, interference, specificity, and pH/specific gravity studies appear to be prospective, controlled laboratory studies. The method comparison using "unaltered clinical samples" suggests a prospective collection or at least a retrospective analysis of prospectively collected samples with LC/MS confirmation. The lay-user study appears to be prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Number of Experts: For the method comparison studies, the "in-house" tests were performed by "three laboratory assistants for each device." Their qualifications are not specified beyond being "laboratory assistants."
  • Qualifications of Experts: Not specified.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. The ground truth for the test set for the method comparison studies was established by LC/MS results, not by human expert adjudication of the device results themselves. The "Viewer" results (Viewer A, B, C) in the method comparison tables refer to the readings by the three laboratory assistants using the multi-drug test devices, which were then compared against the LC/MS ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • A Multi-Reader, Multi-Case (MRMC) comparative effectiveness study was not explicitly stated as being performed in the context of comparing human readers with and without AI assistance.
  • The document describes a "Comparison Studies" section comparing the device's performance to LC/MS results, with three "viewers" (laboratory assistants) reading the device results. This is a form of multi-reader study, but it is a standalone performance assessment against a gold standard (LC/MS), not a comparative effectiveness study of human readers with and without AI.
  • Therefore, an effect size of how much human readers improve with AI vs. without AI assistance is not provided.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • The devices (CLUNGENE Multi-Drug Test Dip Card and Easy Cup) are described as "competitive binding, lateral flow immunochromatographic assay" and are read visually. These are not AI-driven devices, and their performance inherently involves "human-in-the-loop" visual interpretation.
  • Therefore, a standalone (algorithm only) performance study was not done, as the device itself is not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The ground truth for all performance studies (e.g., Precision, Specificity, Method Comparison, Lay-user study) was established using Liquid Chromatography-Mass Spectrometry (LC/MS). For the precision study, samples were "prepared by spiking drug in negative urine samples" and "confirmed by LC/MS." For the method comparison study, "unaltered clinical samples" were "compared to LC/MS results."

8. The sample size for the training set

  • This information is not applicable. The CLUNGENE devices are immunoassay-based test kits, not machine learning or AI algorithms that require a "training set" in the computational sense. The "training" for the device refers to its manufacturing and validation processes, not data-driven algorithmic training.

9. How the ground truth for the training set was established

  • This information is not applicable as the device is not an AI/ML algorithm requiring a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 21, 2019

Hangzhou Clongene Biotech Co.,Ltd. % Joe Shia, Manager LSI International 504 E Diamond Ave., Suite I Gaithersburg, MD 20877

Re: K190412

Trade/Device Name: CLUNGENE Multi-Drug Test Dip Card CLUNGENE Multi-Drug Test Easy Cup Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Codes: NGL, PTG, QBF Dated: February 15, 2019 Received: February 21, 2019

Dear Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K190412

Device Name

CLUNGENE® Multi-Drug Test Dip Card CLUNGENE® Multi-Drug Test Easy Cup

Indications for Use (Describe)

CLUNGENE® Multi-Drug Test Dip Card is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Opiates, Oxycodone, Secobarbital, Methadone, Buprenorphine, Methylenedioxymethamphetamine, Tricyclic Antidepressants, EDDP and Propoxyphene in human urine at the cutoff concentrations of:

Drug(Identifier)CalibratorCut-off level
Amphetamined-Amphetamine1000 ng/mL
OxazepamOxazepam300 ng/mL
CocaineBenzoylecgonine300 ng/mL
Marijuana11-Nor-△9-Tetrahydrocannabinol-9-COOH50 ng/mL
Methamphetamined-Methamphetamine1000 ng/mL
OpiatesMorphine2000 ng/mL
OxycodoneOxycodone100 ng/mL
SecobarbitalSecobarbital300 ng/mL
MethadoneMethadone300 ng/mL
BuprenorphineBuprenorphine10 ng/mL
MethylenedioxymethamphetamineD,L-Methylenedioxymethamphetamine500 ng/mL
PhencyclidinePhencyclidine25 ng/mL
Tricyclic AntidepressantsNortriptyline1000 ng/mL
EDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300 ng/mL
Propoxyphened-Propoxyphene300 ng/mL

Configuration of the CLUNGENE® Multi-Drug Test Dip Card can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Nortriptyline, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GCMS or LC/MS is the preferred confirmatory method.

For in vitro diagnostic use only.

CLUNGENE® Multi-Drug Test Easy Cup is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Opiates, Oxycodone, Secobarbital, Methadone, Buprenorphine, Phencyclidine, Methylenedioxymethamphetamine, Tricyclic Antidepressants, EDDP and Propoxyphene in human urine at the cutoff concentrations of:

Drug(Identifier)CalibratorCut-off level
Amphetamined-Amphetamine1000 ng/mL
OxazepamOxazepam300 ng/mL
CocaineBenzoylecgonine300 ng/mL
Marijuana11-Nor-△9-Tetrahydrocannabinol-9-COOH50 ng/mL

{3}------------------------------------------------

Methamphetamined-Methamphetamine1000 ng/mL
OpiatesMorphine2000 ng/mL
OxycodoneOxycodone100 ng/mL
SecobarbitalSecobarbital300 ng/mL
MethadoneMethadone300 ng/mL
BuprenorphineBuprenorphine10 ng/mL
MethylenedioxymethamphetamineD,L-Methylenedioxymethamphetamine500 ng/mL
PhencyclidinePhencyclidine25 ng/mL
Tricyclic AntidepressantsNortriptyline1000 ng/mL
EDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300 ng/mL
Propoxyphened-Propoxyphene300 ng/mL

Configuration of the CLUNGENE® Multi-Drug Test Easy Cup can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Nortriptyline, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

For in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K190412 510(k) SUMMARY

The purpose of this submission is to add analytes 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine, Opiates and Propoxyphene to previously cleared device (K181790). These three new analytes were evaluated in this submission. For other analytes, please refer to K153741 for Methamphetamine, Morphine and Cannabinoids, and to K161251 for Amphetamine, Cocaine and Oxazepam, and to K180255 for Secobarbital, Methadone and Oxycodone, and K181790 for Methylenedioxymethamphetamine, Buprenorphine, Phencyclidine and Nortriptyline. In this submission the performance results are presented for the three new analytes, but the lay user study was conducted using the entire panel.

    1. Date: March 19, 2019 2. Submitter: Hangzhou Clongene Biotech Co., Ltd. 20 Longquan Road Hangzhou 311121, China 3. Contact person: Joe Shia LSI International Inc 504E Diamond Ave., Suite J Gaithersburg. MD 20877 Telephone: 240-505-7880 Email: shiajl@yahoo.com.
    1. Device Name: CLUNGENE Multi-Drug Test Dip Card CLUNGENE Multi-Drug Test Easy Cup
Classification:Class 2
Product CodeClassificationRegulation SectionPanel
NGLMorphineII21 CFR § 862.3650, Opiate Test SystemToxicology (91)
PTG2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidineII21 CFR § 862.3620, Methadone Test SystemToxicology (91)
QBFPropoxypheneII21 CFR, 862.3700 Propoxyphene Test SystemToxicology (91)
    1. Predicate Devices: K180349
      The Assure Tech Panel Dip Tests/AssureTech Quick Cup Tests
    1. Intended Use
      CLUNGENE® Multi-Drug Test Dip Card is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Opiates Oxycodone, Secobarbital, Methadone, Buprenorphine, Phencyclidine, Methylenedioxymethamphetamine, Tricyclic Antidepressants, EDDP and Propoxyphene in human urine at the cutoff concentrations of:

{5}------------------------------------------------

DrugCalibratorCut-off
Amphetamined-Amphetamine1000 ng/mL
SecobarbitalSecobarbital300 ng/mL
OxazepamOxazepam300 ng/mL
CocaineBenzoylecgonine300 ng/mL
Methamphetamined-Methamphetamine1000 ng/mL
OpiatesMorphine2000 ng/mL
MethadoneMethadone300 ng/mL
OxycodoneOxycodone100 ng/mL
Marijuana11-Nor-△9-Tetrahydrocannabinol-9 COOH50 ng/mL
BuprenorphineBuprenorphine10 ng/mL
MethylenedioxymethamphetamineD,L-Methylenedioxymethamphetamine500 ng/mL
PhencyclidinePhencyclidine25 ng/mL
Tricyclic AntidepressantsNortriptyline1000 ng/mL
EDDP(2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine)300 ng/mL
Propoxyphened-Propoxyphene300 ng/mL

Configuration of the CLUNGENE® Multi-Drug Test Dip Card can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Nortriptyline, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. For in vitro diagnostic use only.

CLUNGENE® Multi-Drug Test Easy Cup is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Opiates, Oxycodone, Secobarbital, Methadone, Buprenorphine, Phencyclidine, Methylenedioxymethamphetamine, Tricyclic Antidepressants, EDDP and Propoxyphene in human urine at the cutoff concentrations of:

DrugCalibratorCut-off
Amphetamined-Amphetamine1000 ng/mL
SecobarbitalSecobarbital300 ng/mL
OxazepamOxazepam300 ng/mL
CocaineBenzoylecgonine300 ng/mL
Methamphetamined-Methamphetamine1000 ng/mL
OpiatesMorphine2000 ng/mL
MethadoneMethadone300 ng/mL
OxycodoneOxycodone100 ng/mL
Marijuana11-Nor-△9-Tetrahydrocannabinol-9 COOH50 ng/mL
BuprenorphineBuprenorphine10 ng/mL
MethylenedioxymethamphetamineD,L-Methylenedioxymethamphetamine500 ng/mL

{6}------------------------------------------------

PhencyclidinePhencyclidine25 ng/mL
Tricyclic AntidepressantsNortriptyline1000 ng/mL
EDDP(2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine)300 ng/mL
Propoxyphened-Propoxyphene300 ng/mL

Configuration of the CLUNGENE® Multi-Drug Test Easy Cup can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine. Oxazepam. Secobarbital, Nortriptyline, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. For in vitro diagnostic use only.

    1. Device Description
      The CLUNGENE Multi-Drug Test Dip Card and CLUNGENE Multi-Drug Test Easy Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Methadone, Buprenorphine, Phencyclidine, Methylenedioxymethamphetamine, Tricyclic Antidepressants, EDDP and Propoxyphene (target analytes) in human urine. The products are single-use in vitro diagnostic devices. The CLUNGENE Multi-Drug Test Dip Card kit contains a Dip Card device, a package insert and a urine cup for sample collection. The CLUNGENE Multi-Drug Test Easy Cup kit contains a Cup device, a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.
    1. Substantial Equivalence Information
      A summary comparison of features of the CLUNGENE Multi-Drug Test Dip Card and CLUNGENE Multi-Drug Test Easy Cup and the predicate devices is provided in following tables.

Table 1: Features Comparison of CLUNGENE Multi-Drug Test Dip Card and the Predicate Devices

ItemDevicePredicate - K180349
Indication(s)for UseFor the qualitative determination of drugs ofabuse in human urine.Same (but the number ofdrugs detected is different)

{7}------------------------------------------------

Calibrator andCut-Off ValuesAmphetamine (AMP): 1,000 ng/mlOxazepam (BZO):300 ng/mlCocaine(COC): 300 ng/ml11-Nor-Δ 9-Tetrahydrocannabinol-9-COOH(THC):50 ng/mlMethamphetamine (MET): 1,000 ng/mlMorphine (MOR): 2000ng/mLOxycodone(OXY) : 100 ng/mlSecobarbital (BAR): 300 ng/mlMethadone (MTD): 300 ng/mlBuprenorphine (BUP): 10 ng/mlD,L-Methylenedioxymethamphetamine(MDMA): 500 ng/mlPhencyclidine (PCP): 25 ng/mlNortriptyline (TCA): 1000 ng/ml2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP): 300 ng/mlPropoxyphene (PPX): 300 ng/mlSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Intended UseFor over-the-counterSame
ConfigurationsDip CardSame
Table 2: Features Comparison of CLUNGENE Multi-Drug Test Easy Cup Tests and the
Predicate Devices
ItemDevicePredicate - K180349
Indication(s)for UseFor the qualitative determination ofdrugs of abuse in human urine.Same (but the number ofdrugs detected is different)
Calibrator and Cut-OffValuesAmphetamine (AMP): 1,000 ng/mlOxazepam (BZO):300 ng/mlCocaine(COC): 300 ng/ml11-Nor-△9-Tetrahydrocannabinol-9-COOH(THC):50 ng/mlMethamphetamine (MET): 1,000 ng/mlMorphine (MOR): 300ng/mLOxycodone(OXY) : 100 ng/mlSecobarbital (BAR): 300 ng/mlMethadone (MTD): 300 ng/mlBuprenorphine (BUP): 10 ng/mlSame

{8}------------------------------------------------

D,L-Methylenedioxymethamphetamine(MDMA): 500 ng/ml
Phencyclidine (PCP): 25 ng/ml
Nortriptyline (TCA): 1000 ng/ml
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP): 300 ng/ml
Propoxyphene (PPX): 300 ng/ml
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodyimmunochemistry.Same
Type of TestQualitative
Specimen TypeHuman Urine
Intended UseFor over-the-counter
ConfigurationsCupDip Card

9. Test Principle

The CLUNGENE Multi-Drug Test Dip Card, and CLUNGENE Multi-Drug Test Easy Cup are rapid tests for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Methadone, Buprenorphine, Phencyclidine, Methylenedioxymethamphetamine, Tricyclic Antidepressants, 2-ethylidene-1,5dimethyl-3.3-diphenylpyrrolidine and Propoxyphene in urine samples. The tests are lateral flow chromatographic immunoassays. During testing, a urine specimen migrates upward by capillary action. If target drugs present in the urine specimen are below the cut-off concentration, it will not saturate the binding sites of its specific monoclonal mouse antibody coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The will not form in the test line region if the target drug level exceeds its cutoff-concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the tests have been performed properly.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, cut off, +25% cut off, +50% cut off , +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative urine samples. Each drug concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device in a randomized order. The results obtained are summarized in the following tables for 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine, Opiates and Propoxyphene. The data for Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Oxycodone, Secobarbital, Methadone, Buprenorphine,

{9}------------------------------------------------

Phencyclidine, Methylenedioxymethamphetamine and Tricyclic Antidepressants were reported in K181790.

CLUNGENE Multi-Drug Test Dip Card
Concentration byLC/MS (ng/mL)LotNumber-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
071.50143.83217.20298.13353.33461.00515.53618.63
Lot 150-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
EDDP
CLUNGENE Multi-Drug
CLUNGENE Multi-Drug Test Easy Cup
Concentration byLC/MS (ng/mL)-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
LotNumber071.50143.83217.20298.13353.33461.00515.53618.63
Lot 150-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-

Opiates

CLUNGENE Multi-Drug Test Dip Card

Concentration byLC/MS (ng/mL)Lot-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Number0492.7985.71394.01941.82433.22876.93519.63873.9
Lot 150-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-

CLUNGENE Multi-Drug Test Easy Cup

Concentration byLC/MS (ng/mL)-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
LotNumber0492.7985.71394.01941.82433.22876.93519.63873.9
Lot 150-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-

Propoxyphene CLUNGENE Multi-Drug Test Dip Card

Concentration byLC/MS (ng/mL)-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
LotNumber075.48153.63222.68296.13372.90464.60501.28588.43
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-

{10}------------------------------------------------

Concentration byLC/MS (ng/mL)Lot-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Number075.48153.63222.68296.13372.90464.60501.28588.43
Lot 350-/0+50-/0+50-10+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
CLUNGENE Multi-Drug Test Easy Cup
Concentration byLC/MS (ng/mL)Lot-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Number075.48153.63222.68296.13372.90464.60501.28588.43
Lot 150-/0+50-10+50-10+50-10+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-10+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

The following cut-off values are verified.

Drug (Identifier)Cut-off level
2-ethylidene-1,5-dimethyl-3, 3-diphenylpyrrolidine (EDDP)300 ng/mL
Opiates (OPI)2000 ng/mL
Propoxyphene (PPX)300 ng/mL
  • b. Linearity
    Not applicable.

  • c. Stability
    The devices are stable at 4-30 ℃ for 24 months based on the accelerated stability study at 45 °C. Real time stability studies are ongoing.

  • d. Interference
    Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentrations at 25% below and 25% above Cut-Off levels. These urine samples were tested using three lots of each device. Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed between the CLUNGENE Multi-Drug Test Dip Card and CLUNGENE Multi-Drug Test Easy Cup.

Acetaminophenβ-EstradiolOxalic acid
AcetophenetidinErythromycinOxolinic acid
N-AcetylprocainamideFenoprofenOxymetazoline
Acetylsalicylic acidFurosemidePapaverine
AlbuminGentisic acidPenicillin G
AminopyrineHemoglobinPerphenazine
AmoxicillinHydralazinePhenelzine
AmpicillinHydrochlorothiazidePrednisone
ApomorphineHydrocortisone(±)-Propranolol
Ascorbic acidO-Hydroxyhippuric acidPseudoephedrine
Aspartame3-HydroxytyramineQuinine

{11}------------------------------------------------

AtropineIbuprofenRanitidine
Benzilic acidIsoproterenolSalicylic acid
Benzoic acidIsoxsuprineSerotonin (5- Hydroxytyramine)
BilirubinKetamineSulfamethazine
Chloral hydrateKetoprofenSulindac
ChloramphenicolLabetalolTetrahydrocortisone 3-(β- Dglucuronide)
ChlorothiazideLoperamideTetrahydrocortisone 3-acetate
ChlorpromazineMeperidineTetrahydrozoline
CholesterolMeprobamateThiamine
ClonidineMethoxyphenamineThioridazine
CortisoneNalidixic acidTriamterene
(-)-CotinineNaloxoneTrifluoperazine
CreatinineNaltrexoneTrimethoprim
DeoxycorticosteroneNaproxenDL-Tryptophan
DextromethorphanNiacinamideTyramine
DiclofenacNifedipineDL-Tyrosine
DiflunisalNorethindroneUric acid
DigoxinNoscapineVerapamil
Diphenhydramine(±)-OctopamineZomepirac
Ecgonine methyl ester

e. Specificity

To test specificity, drug metabolites and other structurally related compounds that are likely to cross-react in urine samples were spiked into negative urine and were tested using three lots of each device. The lowest concentration that caused a positive result for each compound are listed below for 2-ethylidene-1, 5-dimethyl-3, 3diphenylpyrrolidine, Opiates and Propoxyphene. The data for Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Oxycodone, Secobarbital, Methadone, Buprenorphine, Phencyclidine, Methylenedioxymethamphetamine and Tricyclic Antidepressants were reported in K181790. There were no differences observed between the CLUNGENE Multi-Drug Test Dip Card and CLUNGENE Multi-

Drug Test Easy Cup.
EDDP(Cut-off=300 ng/mL)ResultPositive at (ng/ml)% Cross-Reactivity
EDDP (2-ethylidene-1,5-Dimethyl-3,3-diphenylpyrrolidine)300100%
Disopyramide750000.4%
Doxylamine750000.4%
LAAM (Levo-alpha-acetylmethadol)>100000<0.3%
Alpha Methadol>100000<0.3%
Methadone>100000<0.3%
EMDP (2-Ethyl-5-methyl-3,3-diphenylpyrroline)>100000<0.3%
OpiateResult%
(Cut-off=2000 ng/mL)Positive at(ng/ml)Cross-Reactivity
Morphine2,000100%
O6-Acetylmorphine2,000100%
Codeine2,000100%

{12}------------------------------------------------

Propoxypheneﺍ(1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1An Andrews AndResult0/01s
Thebaine20,00010%
Procaine50,0004%
Oxycodone100,0002%
Levorphanol5,00040%
Hydromorphone10,00020%
Heroin2,000100%
EthylMorphine20,00010%
r robov) Premo(Cut-off=300 ng/mL11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11Positive at(ng/ml)/ UCross-Reactivity
d-Propoxyphene300100%
Norpropoxyphene300100%

f. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above Cut-Off levels. These samples were tested using three lots of each device. Results were all positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off. There were no differences observed between the CLUNGENE Multi-Drug Test Dip Card and CLUNGENE Multi-Drug Test Easy Cup.

2. Comparison Studies

Method comparison studies for the CLUNGENE Multi-Drug Test Dip Card and the CLUNGENE Multi-Drug Test Easy Cup were performed in-house with three laboratory assistants for each device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples for each drug. The samples were blind labeled and compared to LC/MS results. The results are presented in the tables below for 2-ethylidene-1, 5-diphenylpyrrolidine, Opiates and Propoxyphene. The data for Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Oxycodone, Secobarbital, Methadone, Buprenorphine, Phencyclidine, Methylenedioxymethamphetamine and Tricyclic Antidepressants were reported in K181790.

CLUNGENEMulti-DrugTest Dip CardNegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0011920
Negative10151410
ViewerBPositive0012020
Negative10151400
ViewerCPositive0001920
Negative10151510

EDDP

Discordant Results

ViewerSample NumberLC/MS ResultDip CardViewer Results
Viewer AEDDP10285.53Positive

{13}------------------------------------------------

Viewer BEDDP98284.15Positive
Viewer AEDDP39318.03Negative
Viewer CEDDP39318.03Negative
CLUNGENEMulti-DrugTest EasyCupNegativeLow Negative byLC/MS(less than-50%)Near Cutoff Negative byLC/MS(Between-50% andcutoff)Near Cutoff Positive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0011920
Negative10151410
ViewerBPositive0001920
Negative10151510
ViewerCPositive0011920
Negative10151410

Discordant Results

ViewerSample NumberLC/MS ResultEasy CupViewer Results
Viewer AEDDP98284.15Positive
Viewer CEDDP10285.53Positive
Viewer AEDDP39318.03Negative
Viewer BEDDP07330.83Negative
Viewer CEDDP39318.03Negative

Opiates

CLUNGENEMulti-DrugTest Dip CardNegativeLow Negative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0021920
ViewerANegative10151310
ViewerBPositive0011820
ViewerBNegative10151420
ViewerCPositive0011920
ViewerCNegative10151410

Discordant Results

ViewerSample NumberLC/MS ResultDip CardViewer Results
Viewer AOPI351978.6Positive
Viewer AOPI511924.8Positive
Viewer BOPI511924.8Positive
Viewer COPI351978.6Positive

{14}------------------------------------------------

Viewer AOPI662010.4Negative
Viewer BOPI662010.4Negative
Viewer BOPI572087.1Negative
Viewer COPI572087.1Negative
CLUNGENEMulti-DrugTest EasyCupNegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0021820
Negative10151320
ViewerBPositive0011920
Negative10151410
ViewerCPositive0021820
Negative10151320

Discordant Results

ViewerSample NumberLC/MS ResultEasy CupViewer Results
Viewer AOPI351978.6Positive
Viewer AOPI511924.8Positive
Viewer BOPI511924.8Positive
Viewer COPI511924.8Positive
Viewer COPI351978.6Positive
Viewer AOPI562061.2Negative
Viewer AOPI662010.4Negative
Viewer BOPI572087.1Negative
Viewer COPI562061.2Negative
Viewer COPI662010.4Negative

Propoxyphene

CLUNGENEMulti-DrugTest Dip CardNegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0011920
Negative10151410
ViewerBPositive0011920
Negative10151410
ViewerCPositive0001920
Negative10151510

{15}------------------------------------------------

ViewerSample NumberLC/MS ResultDip CardViewer Results
Viewer APPX22292.00Positive
Viewer BPPX22292.00Positive
Viewer APPX59322.25Negative
Viewer BPPX62334.73Negative
Viewer CPPX59322.25Negative
CLUNGENEMulti-DrugTest EasyCupNegativeLow Negative byLC/MS(less than-50%)Near Cutoff Negative byLC/MS(Between-50% andcutoff)Near Cutoff Positive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0011920
Negative10151410
ViewerBPositive0001820
Negative10151520
ViewerCPositive0011920
Negative10151410

Discordant Results

ViewerSample NumberLC/MS ResultEasy CupViewer Results
Viewer APPX22292.00Positive
Viewer CPPX22292.00Positive
Viewer APPX59322.25Negative
Viewer BPPX59322.25Negative
Viewer CPPX59322.25Negative
Viewer BPPX62334.73Negative

Lay-user study

A lay user study was performed at three intended user sites with 310 lay persons for each device format. The lay users had diverse educational and professional backgrounds and ranged in age from 18 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. Each device was tested. Typical Results are shown below.

% of Cut-offNumberofsamplesConcentration byLC/MS(ng/mL)Lay person resultsThepercentageof correctresults(%)
DrugsNo. ofPositiveNo. ofNegative
AMP-100% Cut-off200020100
-75% Cut-off20250020100
-50% Cut-off1705000170100
-25% Cut-off2075011995
+25% Cut-off20125019195
+50% Cut-off401500400100
+75% Cut-off201750200100
BAR-100% Cut-off200020100
-75% Cut-off2075020100
-50% Cut-off1701500170100
-25% Cut-off20225020100
+25% Cut-off2037518290
+50% Cut-off40450400100
+75% Cut-off20525200100
-100% Cut-off200020100
COC-75% Cut-off2075020100
-50% Cut-off1701500170100
-25% Cut-off2022511995
+25% Cut-off2037519195
+50% Cut-off40450400100
+75% Cut-off20525200100
-100% Cut-off200020100
-75% Cut-off2075020100
BZO-50% Cut-off1701500170100
-25% Cut-off20225020100
+25% Cut-off2037519195
+50% Cut-off40450400100
+75% Cut-off20525200100
-100% Cut-off200020100
-75% Cut-off2075020100
-50% Cut-off1701500170100
MET-25% Cut-off20225020100
+25% Cut-off2037519195
+50% Cut-off40450400100
+75% Cut-off20525200100
-100% Cut-off200020100
-75% Cut-off2075020100
-50% Cut-off1701500170100
-25% Cut-off20225020100
MTD+25% Cut-off2037519195
+50% Cut-off40450400100
+75% Cut-off20525200100
-100% Cut-off200020100
-75% Cut-off2075020100
-50% Cut-off1701500170100
-25% Cut-off20225020100
+25% Cut-off2037519195
OPI+50% Cut-off40450400100
+75% Cut-off20525200100
-100% Cut-off200020100
-75% Cut-off2075020100
-50% Cut-off1701500170100
-25% Cut-off20225020100
+25% Cut-off2037519195
+50% Cut-off40450400100
OXY+75% Cut-off20525200100
-100% Cut-off200020100
-75% Cut-off2075020100
-50% Cut-off1701500170100
-25% Cut-off20225020100
+25% Cut-off2037519195
+50% Cut-off40450400100
+75% Cut-off20525200100
THC+75% Cut-off20175200100
-100% Cut-off200020100
-75% Cut-off2012.5020100
-50% Cut-off170250170100
-25% Cut-off2037.5020100
+25% Cut-off2062.519195
+50% Cut-off4075400100
+75% Cut-off2087.5200100
TCA-100% Cut-off200020100
-75% Cut-off20250020100
-50% Cut-off1705000170100
-25% Cut-off2075011995
+25% Cut-off20125019195
+50% Cut-off401500400100
+75% Cut-off201750200100
-100% Cut-off200020100
BUP-75% Cut-off202.5020100
-50% Cut-off17050170100
-25% Cut-off207.511995
+25% Cut-off2012.519195
+50% Cut-off4015400100
+75% Cut-off2017.5200100
-100% Cut-off200020100
PCP-75% Cut-off206.25020100
-50% Cut-off17012.50170100
-25% Cut-off2018.7511995
+25% Cut-off2031.25200100
+50% Cut-off4037.5400100
+75% Cut-off2043.75200100
-100% Cut-off200020100
-75% Cut-off20125020100
MDMA-50% Cut-off1702500170100
-25% Cut-off20375020100
+25% Cut-off2062519195
+50% Cut-off40750400100
+75% Cut-off20875200100
-100% Cut-off200020100
-75% Cut-off2075020100
-50% Cut-off1701500170100
EDDP-25% Cut-off2022511995
+25% Cut-off2037519195
+50% Cut-off40450400100
+75% Cut-off20525200100
-100% Cut-off200020100
-75% Cut-off2075020100
-50% Cut-off1701500170100
-25% Cut-off20225020100
PPX+25% Cut-off2037519195
+50% Cut-off40450400100
-100% Cut-off200020100
-75% Cut-off2075020100
-50% Cut-off1701500170100
-25% Cut-off20225020100
+25% Cut-off2037519195
+50% Cut-off40450400100

CLUNGENE Multi-Drug Test

{16}------------------------------------------------

{17}------------------------------------------------

{18}------------------------------------------------

(ﺎﺗﺎ ﺗﺎEA Eل ڪاريCarlos Concession Company of Children
--------------------------------------------------------------------------

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

  1. Clinical Studies

Not applicable.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison, and lay-user studies of the devices, it's concluded that the CLUNGENE Multi-Drug Test Dip Card and CLUNGENE Multi-Drug Test Easy Cup are substantially equivalent to the predicate.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).