K Number
K153741

Validate with FDA (Live)

Date Cleared
2016-03-18

(81 days)

Product Code
Regulation Number
862.3870
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Clungene Methamphetamine Tests are immunochromatographic assays for the qualitative determination of dmethamphetamine in human urine at cut-off concentration of 1000 ng/mL. The calibrator is d-methamphetamine. The tests are available in a Cassette format, a Easy Cup format, a Split Key Cup format and a Dip Card format. The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling. For in vitro diagnostic use only.

Clungene Marijuana Tests are immunochromatographic assays for the qualitative determination of 11-Nor-△9-Tetrahydrocannabinol-9-COOH in human urine at cut-off concentration of 50 ng/mL. The calibrator is 11-Nor-△9-Tetrahydrocannabinol-9-COOH. The tests are available in a Cassette format, a Easy Cup format, a Split Key Cup format and a Dip Card format. The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, is provided in the package labeling. For in vitro diagnostic use only.

Clungene Morphine Tests are immunochromatographic assays for the qualitative determination of Morphine in human urine at cut-off concentration of 300 ng/mL. The callbrator is Morphine. The tests are available in a Cassette format, a Easy Cup format, a Split Key Cup format and a Dip Card format. The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. This test is intended for over the counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling. For in vitro diagnostic use only.

Device Description

The CLUNGENE Methamphetamine Tests, CLUNGENE Morphine Tests, and CLUNGENE Marijuana Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of d-Methamphetamine, Morphine and 11-Nor-A9-Tetrahydrocannabinol-9-COOH (target analytes) in human urine. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Clungene Methamphetamine, Morphine, and Marijuana Tests, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" in a separate section with specific numerical thresholds for accuracy, sensitivity, or specificity that the device must meet for approval. Instead, it describes performance characteristics and then presents the results of those studies. The implicit acceptance criteria appear to be the demonstrated performance of the device showing appropriate qualitative detection relative to the cut-off concentrations when compared to GC/MS, and user readability for OTC use.

Performance Characteristics Summary (Implicit Acceptance Criteria and Reported Performance)

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
PrecisionConsistent results across different lots and concentrations around cut-off.Across 3 lots and 9 different concentrations (-100% to +100% cut-off), the devices consistently showed 100% agreement for concentrations far from the cut-off. For samples at the cut-off, there was variability in positive/negative calls (e.g., for Methamphetamine Cassette, Lot 1: 22-/28+, Lot 2: 29-/21+, Lot 3: 29-/21+ at cut-off). This implies appropriate performance around the threshold.
Cut-off AccuracySamples below -25% cut-off should be negative, and samples above +25% cut-off should be positive.All samples at and below -25% Cut-Off were negative. All samples at and above +25% Cut-Off were positive for Methamphetamine, Morphine, and Marijuana.
InterferenceNo interference from common physiological/pathological substances or drugs at specified concentrations.Numerous substances (listed in the document) showed no interference at 100µg/mL.
Specificity (Cross-reactivity)Limited cross-reactivity with structurally similar compounds; identified cross-reactive compounds should be listed with their reactivity percentage.Detailed tables provided showing cross-reactivity percentages for various related compounds (e.g., Methamphetamine: MDEA 5%, D/L-Methamphetamine 33%; Morphine: O6-Acetylmorphine 100%, Codeine 100%; Marijuana: 11-Hydroxy-Δ9-Tetrahydrocannabinol 1%).
Effect of Urine Specific Gravity & pHNo significant impact on results within specified ranges (pH 4-9, SG 1.000-1.035).All samples within these ranges showed expected positive/negative results when spiked at +/- 25% cut-off.
Method Comparison (Professional User)High agreement with GC/MS results, especially for samples far from the cut-off. Discordant results should primarily occur near the cut-off.Methamphetamine, Morphine, Marijuana: High agreement for negative, low negative, high positive samples. Discordant results were primarily observed for samples very close to the cut-off (e.g., GC/MS result 936 ng/mL for Methamphetamine vs. 1000 ng/mL cut-off).
Lay-user Study (Accuracy)High percentage of correct results by lay users, particularly for samples clearly below or above the cut-off.For concentrations -100%, -75%, -50% cut-off, lay users showed 100% correct negative results. For +50%, +75% cut-off, lay users showed 100% correct positive results. For -25% and +25% cut-off, correctness ranged from 90-95%.
Lay-user Study (Usability)Instructions should be easily understood by lay users.All lay users indicated the device instructions can be easily followed. Flesch-Kincaid Grade Level of 7.

2. Sample Size Used for the Test Set and Data Provenance

  • Professional User (Method Comparison) Test Set:
    • Sample Size: 80 unaltered clinical urine samples per drug (40 negative, 40 positive).
    • Data Provenance: Retrospective, described as "in-house" studies using "unaltered clinical samples." The country of origin is not explicitly stated, but the manufacturer is based in China.
  • Lay-user Study Test Set:
    • Sample Size:
      • Participants: 1680 lay persons.
      • Samples: 1 device per participant, with one blind-labeled urine sample.
      • For each drug (Methamphetamine, Morphine, Marijuana) and each device format (Cassette, Dip Card, Split-Key Cup, Easy Cup), there were a total of 7 concentrations tested. For each concentration, 20 samples were used.
      • Total samples per drug (e.g., Methamphetamine): 7 concentrations * 20 samples/concentration = 140 samples.
      • Total samples for all 3 drugs across all 4 device formats: 3 drugs * 4 device formats * (7 concentrations * 20 samples/concentration) = 3 * 4 * 140 = 1680 samples. This matches the number of lay persons, implying each lay person tested one sample with one device.
    • Data Provenance: Prospective, as it involved lay persons testing samples. The study was performed "at three intended user sites," but the country of origin for these sites is not specified. Samples were prepared by spiking known drug concentrations into drug-free pooled urine.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Professional User (Method Comparison) Test Set:
    • Number of "Experts": Three "laboratory assistants" were involved in running the visual tests for comparison. Their qualifications are not specified beyond being "laboratory assistants."
    • Ground Truth: The ground truth for the clinical samples was established by Gas Chromatography/Mass Spectrometry (GC/MS) results. This is considered the gold standard for drug quantification.
  • Lay-user Study Test Set:
    • Number of Experts: No explicit "experts" were used to establish the ground truth for reading the device results in this specific study, as the lay users were the test subjects. The ground truth for the sample concentrations was established by GC/MS.

4. Adjudication Method for the Test Set

  • Professional User (Method Comparison) Test Set: No formal adjudication method (like 2+1 or 3+1) is described for the visual interpretation by the three laboratory assistants. Each assistant's interpretation was recorded and compared to the GC/MS ground truth. Discordant results between the device and GC/MS were tabulated for each viewer individually.
  • Lay-user Study Test Set: No adjudication. Each lay user read their own device result.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such MRMC comparative effectiveness study was done. The device is a qualitative immunochromatographic assay for drug detection, not an AI interpretation system.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is a manual, visually interpreted immunochromatographic assay, not an algorithm or AI system. Its performance is inherently "standalone" in that it produces a visual result without external algorithm assistance for interpretation, but it always requires human observation and interpretation.

7. The Type of Ground Truth Used

  • For all studies (Precision, Cut-off, Specificity, Interference, Method Comparison, Lay-user): The primary ground truth for the presence and concentration of drugs in urine samples was Gas Chromatography/Mass Spectrometry (GC/MS). This is considered a highly accurate and quantitative analytical method.

8. The Sample Size for the Training Set

This information is not provided. The document describes performance testing of the finished device. For immunochromatographic assays, there isn't typically a "training set" in the machine learning sense. Instead, development involves R&D to optimize reagents, membrane materials, and assay parameters. The performance studies described serve to validate the finalized design.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as a "training set" in the AI sense is not relevant for this type of device. During the development and optimization of the assay, standard reference materials of known drug concentrations (confirmed by methods like GC/MS) would be used to evaluate and refine the test's linearity, sensitivity, and specificity.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing to the right, arranged in a cascading manner.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 18, 2016

HANGZHOU CLONGENE BIOTECH CO., LTD. C/O JOE SHIA MANAGER 504 E DIAMOND AVE., SUITE I GAITHERSBURG MD 20877

Re: K153741

Trade/Device Name: Clungene Methamphetamine Test, Clungene Morphine Test, Clungene Marijuana Test Regulation Number: 21 CFR 862.3870 Regulation Name: Cannabinoid test system Regulatory Class: II Product Code: LDJ, DJC, DJG Dated: December 19, 2015 Received: December 28, 2015

Dear Mr. Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153741

Device Name Clungene Methamphetamine Tests Clungene Marijuana Tests Clungene Morphine Tests

Indications for Use (Describe)

Clungene Methamphetamine Tests are immunochromatographic assays for the qualitative determination of dmethamphetamine in human urine at cut-off concentration of 1000 ng/mL. The calibrator is d-methamphetamine. The tests are available in a Cassette format, a Easy Cup format, a Split Key Cup format and a Dip Card format. The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling. For in vitro diagnostic use only.

Clungene Marijuana Tests are immunochromatographic assays for the qualitative determination of 11-Nor-△9-Tetrahydrocannabinol-9-COOH in human urine at cut-off concentration of 50 ng/mL. The calibrator is 11-Nor-△9-Tetrahydrocannabinol-9-COOH. The tests are available in a Cassette format, a Easy Cup format, a Split Key Cup format and a Dip Card format.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, is provided in the package labeling. For in vitro diagnostic use only.

Clungene Morphine Tests are immunochromatographic assays for the qualitative determination of Morphine in human urine at cut-off concentration of 300 ng/mL. The callbrator is Morphine. The tests are available in a Cassette format, a Easy Cup format, a Split Key Cup format and a Dip Card format.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

This test is intended for over the counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling. For in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)

|X | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

    1. Date: February 22, 2016
  • Hangzhou Clongene Biotech Co., Ltd. 2. Submitter: 20 Longquan Road Hangzhou 311121, China
    1. Contact person: Zheng Shujian Hangzhou Clongene Biotech Co., Ltd. 20 Longquan Road Hangzhou 311121, China Telephone: 86 571 88262120 Email: frank@clongene.com.
    1. Device Name: CLUNGENE Methamphetamine Tests CLUNGENE Morphine Tests CLUNGENE Marijuana Tests

Classification:

Product CodeCFRPanel
DJC21 CFR, 862.3610 Methamphetamine TestToxicology
LDJ21 CFR, 862.3870 Cannabinoid Test SystemToxicology
DJG21 CFR, 862.3650 Opiate Test SystemToxicology

5. Predicate Devices: K052115

The FIRST CHECK MULTI DRUG CUP Urine Test

    1. Intended Use
      Clungene Methamphetamine Tests are immunochromatographic assays for the qualitative determination of d-methamphetamine in human urine at cut-off concentration of 1000 ng/mL. The calibrator is d-methamphetamine. The tests are available in a Cassette format, an Easy Cup format, a Split Key Cup format and a Dip Card format.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling. For in vitro diagnostic use only.

Clungene Marijuana Tests are immunochromatographic assays for the qualitative determination of 11-Nor-A9-Tetrahydrocannabinol-9-COOH in human urine at cut-off concentration of 50 ng/mL. The calibrator is 11-Nor-A9-Tetrahydrocannabinol-9-COOH. The tests are available in a Cassette format, an Easy Cup format, a Split Key Cup format and a Dip Card format.

The tests provide only preliminary test results. A more specific alternative chemical must

{5}------------------------------------------------

be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling. For in vitro diagnostic use only.

Clungene Morphine Tests are immunochromatographic assays for the qualitative determination of Morphine in human urine at cut-off concentration of 300 ng/mL. The calibrator is Morphine. The tests are available in a Cassette format, an Easy Cup format, a Split Key Cup format and a Dip Card format.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling. For in vitro diagnostic use only.

    1. Device Description
      The CLUNGENE Methamphetamine Tests, CLUNGENE Morphine Tests, and CLUNGENE Marijuana Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of d-Methamphetamine, Morphine and 11-Nor-A9-Tetrahydrocannabinol-9-COOH (target analytes) in human urine. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.
    1. Substantial Equivalence Information
      A summary comparison of features of the CLUNGENE Methamphetamine Tests, CLUNGENE Morphine Tests and CLUNGENE Marijuana Tests and the predicate devices is provided in following tables.
ItemDevicePredicate - K052115
Indication(s)for UseFor the qualitative determination of drugs ofabuse in human urine.Same (but the number ofdrugs detected is different)
Calibratord-MethamphetamineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same

Table 1: Features Comparison of CLUNGENE Methamphetamine Tests and the Predicate Devices

{6}------------------------------------------------

Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Cut-Off Values1000 ng/mLSame
Intended UseFor over-the-counter and prescription uses.Same
ConfigurationsCassette, Dip Card and CupsCup

Table 2: Features Comparison of CLUNGENE Morphine Tests and the Predicate Devices

ItemDevicePredicate - K052115
Indication(s)for UseFor the qualitative determination ofdrugs of abuse in human urine.Same (but the number ofdrugs detected is different)
CalibratorMorphineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Cut-Off Values300 ng/mLSame
Intended UseFor over-the-counter and prescription uses.Same
ConfigurationsCassette, Dip Card and CupsCup

Table 3: Features Comparison of CLUNGENE Marijuana Tests and the Predicate Devices

ItemDevicePredicate - K052115
Indication(s)for UseFor the qualitative determination ofdrugs of abuse in human urine.Same (but the number ofdrugs detected is different)
Calibrator11-Nor-△9-Tetrahydrocannabinol-9-COOHSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Cut-Off Values50 ng/mLSame
Intended UseFor over-the-counter and prescription uses.Same
ConfigurationsCassette, Dip Card and CupsCup

{7}------------------------------------------------

9. Test Principle

The CLUNGENE Methamphetamine Tests, CLUNGENE Morphine Tests, and CLUNGENE Marijuana Tests are rapid tests for the qualitative detection of d-Methamphetamine, Morphine and 11-Nor-A9-Tetrahydrocannabinol-9-COOH in urine samples. The tests are lateral flow chromatographic immunoassays. During testing, a urine specimen migrates upward by capillary action. If target drugs present in the urine specimen are below the cut-off concentration, it will not saturate the binding sites of its specific monoclonal mouse antibody coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The will not form in the test line region if the target drug level exceeds its cutoff-concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the tests have been performed properly.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut off. -75% cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off , +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device in a randomized order. The results obtained are summarized in the following tables.

Cassette
LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+29-/21+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+29-/21+50+/0-50+/0-50+/0-50+/0-
Dip Card
LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+21-/29+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Split-Key Cup
LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+21-/29+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-

CLUNGENE Methamphetamine Tests

Easy Cup

Lot 1

Lot 2

Lot 3

50-/0+

50-/0+

50-/0+

50-/0+

50-10+

50-/0+

50-/0+

50-/0+

50-10+

50-/0+

50-/0+

50-/0+ 20-/30+ 50+/0-

17-/33+

21-/29+

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

{8}------------------------------------------------

Lot Number-100% cut off-75% cut off-50% cut off-25% cutoff+25% cut off+50% cut off+75% cut off+100% cut off
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+17-/33+50+/0-50+/0-50+/0-

CLUNGENE Morphine Tests

Cassette

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-

Dip Card

Lot Number-100% cut off-75% cut off-50% cut off-25% cutoffcut off+25% cut off+50% cut off+75% cut off+100% cut off
Lot 150-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

Split-Key Cup

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+29-/21+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+19-/31+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+32-/18+50+/0-50+/0-50+/0-50+/0-

Easy Cup

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoff+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+31-/19+50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-

CLUNGENE Marijuana Tests

Cassette

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+21-/29+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-

Dip Card

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoff+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+19-/31+50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-

Split-Key Cup

{9}------------------------------------------------

Lot-100%-75%-50%-25%+25%+50%+75%+100%
Numbercut offcut offcut offcutoffcut offcut offcut offcut offcut off
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+31-/19+50+/0-50+/0-50+/0-50+/0-
Easy Cup
Lot-100%-75%-50%-25%+25%+50%+75%+100%
Numbercut offcut offcut offcutoffcut offcut offcut offcut offcut off
Lot 150-/0+50-/0+50-/0+50-/0+30-/20+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+29-/21+50+/0-50+/0-50+/0-50+/0-

b. Linearity

Not applicable.

  • c. Stability
    The devices are stable at 4-30 ℃ for 24 months based on the accelerated stability study at 45 °C and real time stability determination at both 4 °C and 30 °C.

d. Cut-off

A total of 150 samples equally distributed at concentrations of -50% Cut-Off; -25% Cut-Off; Cut-Off; +25% Cut-Off; +50% Cut-Off were tested using three different lots of each device by three different operators. Results were all positive at and above +25% Cut-off and all negative at and below -25% Cut-off for Methamphetamine, Morphine and Marijuana.

The following cut-off values for the candidate devices have been verified.

CalibratorCut-off (ng/mL)
d-Methamphetamine1000
Morphine300
11-Nor-△9-Tetrahydrocannabinol-9-COOH50

e. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentrations at 25% below and 25% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed for different devices.

4-Acetamidophenol(-) Y EphedrinePenicillin-G
AcetophenetidinErythromycinPentazocaine
N-Acetylprocainamideß-EstradiolPentobarbital
Acetylsalicylic acidEstrone-3-sulfatePerphenazine
AminopyrineEthyl-p-aminobenzoatePhencyclidine
AmitriptylineFenfluraminePhenelzine
AmobarbitalFenoprofenPhendimetrazine
AmoxicillinFurosemidePhenobarbital
AmpicillinGentisic acidPhetoin
Ascorbic acidHemoglobinL-Phenylephrine
ApomorphineHydralazineβ-Phenylethlamine
AspartameHydrochlorothiazidePhenylpropanolamine
AtropineHydrocodonePrednisolone
Benzilic acidHydrocortisonePrednisone
Benzoic acidO-Hydroxyhippuric acidProcaine
Benzoylecgonine3-HydroxytyraminePromazine
BilirubinIbuprofenPromethazine
BrompheniramineImipramineD,L-Propranolol
Caffeine(-) IsoproterenolPropiomazine
CannabidiolIsoxsuprineD-Propoxyphene
CannabinolKetamineQuinidine
ChloralhydrateKetoprofenQuinine
ChloramphenicolLabetalolRanitidine
ChlordiazepoxideLevorphanolSalicylic acid
ChlorothiazideLoperamideSecobarbital
(+) ChlorpheniramineMaprotilineSerotonin
ChlorpromazineMeperidineSulfamethazine
ChlorquineMeprobamateSulindac
CholesterolMethadoneTemazepam
ClomipramineMethylphenidateTetracycline
ClonidineMorphine-3-DglucuronideTetrahydrocortisone
Cocaine hydrochlorideNalidixic acidTetrahydrozoline
CodeineNaloxoneΔ9-THC-COOH
CortisoneNaltrexoneThebaine
(-) CotinineNaproxenThiamine
CreatinineNiacinamideThioridazine
DeoxycorticosteroneNifedipineD,L-Thyroxine
DextromethorphanNorcodeinTolbutamine
DiazepamNorethindroneTriamterene
DiclofenacD-NorpropoxypheneTrifluoperazine
DiflunisalNoscapineTrimethoprim
DigoxinD,L-OctopamineTrimipramine
DiphenhydramineOxalic acidTryptamine
DoxylamineOxazepamD, L-Tyrosine
Ecgonine hydrochlorideOxolinic acidUric acid
Ecgonine methylesterOxycodoneVerapamil
(IR,2S)-(-)-EphedrineOxymetazolineZomepirac
L-EphedrinePapaverine

Methamphetamine:

{10}------------------------------------------------

Morphine

{11}------------------------------------------------

Acebutolol(-) Y EphedrineHydroxymethamphetamine
Acetopromazine - d6ErythromycinPapaverine
4-Acetamidophenolβ-EstradiolPenicillin-G
AcetophenetidinEstrone-3-sulfatePentazocaine
N-AcetylprocainamideEthyl-p-aminobenzoatePentobarbital
Acetylsalicylic acidFenoprofenPerphenazine
D,L-AmphetamineFurosemidePhencyclidine
L-AmphetamineGentisic acidPhenelzine
AminopyrineHemoglobinPhenobarbital
AmitryptylineHydralazinePhentermine
AmobarbitalHydrochlorothiazideβ-Phenylethylamine
AmoxicillinHydrocortisoneL-Phenylephrine
AmpicillinO-Hydroxyhippuric acidβ-Phenylethlamine
Ascorbic acid3-HydroxytyraminePhenylpropanolamine
ApomorphineIbuprofenPrednisolone
AspartameImipraminePrednisone
AtropineIproniazidPromazine
Benzilic acid(-) IsoproterenolPromethazine
Benzoic acidIsoxsuprineD,L-Propanolol
BenzoylecgonineKetamineD-Propoxyphene
BenzphetamineKetoprofenD-Pseudoephedrine
BilirubinLabetalolQuinidine
BrompheniramineLoperamideQuinine
CaffeineMaprotilineSalicylic acid
ChloramphenicolMeprobamateSecobarbital
ChlordiazepoxideMethoxyphenamineSerotonin
ChlorothiazideMethylenedioxyamphetamineSulfamethazine
(±) ChlorpheniramineMethylenedioxymethamphetamineSulindac
ChlorpromazineMethadoneTemazepam
ChlorquineMethylphenidateTetracycline
CholesterolMethyprylonTetrahydrozoline
ClomipramineNalorphineTetrahydrocortisone
ClonidineNalidixic acidTetrahydrocortisone-(5-Dglucuronide)
Cocaine hydrochlorideNaloxoneThiamine
CortisoneNaltrexoneThioridazine
(-) CotinineNaproxenD,L-Thyroxine
CreatinineNiacinamideTolbutamine
DeoxycorticosteroneNifedipineTriamterene
DextromethorphanNorcodeinTrifluoperazine

{12}------------------------------------------------

DiazepamNorethindroneTrimethoprim
DiclofenacD-NorpropoxypheneTrimipramine
DiflunisalNoscapineTryptamine
DigoxinD,L-OctopamineD,L-Tyrosine
DiphenhydramineOxalic acidUric acid
DoxylamineOxazepamVerapamil
EcgonineOxolinic acidZomepirac
Ecgonine methylesterOxymetazoline

Marijuana

Marijuana
4-AcetamidophenolEstrone-3-sulfatePenicillin-G
AcetophenetidinEthyl-p-aminobenzoatePentazocine
N-AcetylprocainamideFenoprofenPentobarbital
Acetylsalicylic acidFurosemidePerphenazine
AminopyrineGentisic acidPhencyclidine
AmitryptylineHemoglobinPhenelzine
AmobarbitalHydralazinePhenobarbital
AmoxicillinHydrochlorothiazidePhentermine
AmpicillinHydrocodoneL-Phenylephrine
Ascorbic acidHydrocortisoneβ-Phenylethlamine
D,L-AmphetamineO-Hydroxyhippuric acidβ-Phenyllethylamine
L-Amphetamine3-HydroxytyraminePhenylpropanolamine
ApomorphineIbuprofenPrednisolone
AspartameImipraminePrednisone
AtropineIproniazidProcaine
Benzilic acid(-) IsoproterenolPromazine
Benzoic acidIsoxsuprinePromethazine
BenzoylecgonineKetamineD,L-Propanolol
BenzphetamineKetoprofenD-Propoxyphene
BilirubinLabetalolD-Pseudoephedrine
BrompheniramineLevorphanolQuinidine
CaffeineLoperamideQuinine
ChloralhydrateMaprotilineRanitidine
ChloramphenicolMeprobamateSalicylic acid
ChlordiazepoxideMethadoneSecobarbital
ChlorothiazideMethoxyphenamineSerotonin (5-
(±) Chlorpheniramine(+)3,4-Sulfamethazine
Chlorpromazine(+)3,4-Sulindac
ChlorquineMethylphenidateTemazepam
CholesterolMethyprylonTetracycline
ClomipramineMorphine-3-β-DglucuronideTetrahydrocortisone,3

{13}------------------------------------------------

ClonidineNalorphineTetrahydrocortisone3 (5-Dglucuronide)
Cocaine hydrochlorideNaloxoneTetrahydrozoline
CodeineNalidixic acidThebaine
CortisoneNaltrexoneThiamine
(-) CotinineNaproxenThioridazine
CreatinineNiacinamideD, L-Thyroxine
DeoxycorticosteroneNifedipineTolbutamine
DextromethorphanNorcodeinTriamterene
DiazepamNorethindroneTrifluoperazine
DiclofenacD-NorpropoxypheneTrimethoprim
DiflunisalNoscapineTrimipramine
DigoxinD,L-OctopamineTryptamine
DiphenhydramineOxalic acidD, L-Tryptophan
DoxylamineOxazepamTyramine
Ecgonine hydrochlorideOxolinic acidD, L-Tyrosine
Ecgonine methylesterOxycodoneUric acid
(-) Y EphedrineOxymetazolineVerapamil
Erythromycinp-HydroxymethamphetamineZomepirac
β-EstradiolPapaverine

f. Specificity

To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device. The lowest concentration that caused a positive result for each compound are listed below. There were no differences observed for different devices.

MethamphetamineResult% Cross-
(Cut-off=1000 ng/mL)Positive at (ng/mL)Reactivity
D - Methamphetamine1000100%
(+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDEA)200005%
D/L-Methamphetamine300033%
p-Hydroxymethamphetamine300003.3%
L-Methamphetamine800012.5%
(+/-)3,4-Methylenedioxyamphetamine(MDA)200005%
(+/-)3,4-methylenedioxymethamphetamine(MDMA)Negative at 100000<1%
D-AmphetamineNegative at 100000<1%
L-AmphetamineNegative at 100000<1%
Morphine(Cut-off=300 ng/mL)ResultPositive at (ng/mL)% Cross-Reactivity
Morphine300100%

{14}------------------------------------------------

O6-Acetylmorphine300100%
Codeine300100%
EthylMorphine60005%
Heroin300100%
Hydromorphone300010%
Hydrocodone500000.6%
Levorphanol150020%
Oxycodone300001%
Procaine150002%
Thebaine60005%
Morphine-3-glucuronide>100000<0.3%
MarijuanaResult% Cross-Reactivity
(Cut-off=50 ng/mL)Positive at(ng/mL)
11-Nor-Δ9-Tetrahydrocannabinol-9-COOH50100%
11-Hydroxy-Δ9-Tetrahydrocannabinol50001%
11-Nor-Δ8-Tetrahydrocannabinol-9-COOH50100%
Cannabinol200000.3%
Δ8-Tetrahydrocannabinol100000.5%
Δ9-Tetrahydrocannabinol100000.5%
Cannabidiol200000.3%
11-Nor-Δ9-THC-carboxy glucuronide25002%
(-)-11-nor-9-carboxy-Δ 9-THC25002%

g. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above Cut-Off levels. These samples were tested using three lots of each device. Results were all positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different devices.

2. Comparison Studies

Method comparison studies for the CLUNGENE Methamphetamine Tests, the CLUNGENE Morphine Tests and the CLUNGENE Marijuana Tests were performed in-house with three laboratory assistants for each device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to GC/MS results. The results are presented in the tables below:

Methamphetamine
CassetteNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cutoff)Near Cutoff Positive by GC/MS (Between the cutoff and +50%)High Positive by GC/MS (greater than +50%)

Methamphetamine

{15}------------------------------------------------

Viewer0011821
ANegative14121310
ViewerPositive0011721
BNegative14121320
ViewerPositive0011721
CNegative14121320

Discordant Results of Methamphetamine Cassette

ViewerSample NumberGC/MS ResultCassetteViewer Results
Viewer AMET09936Positive
Viewer BMET60963Positive
Viewer CMET60963Positive
Viewer AMET151010Negative
Viewer BMET151010Negative
Viewer BMET691100Negative
Viewer CMET151010Negative
Viewer CMET691100Negative
PanelDipNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011821
Negative14121310
ViewerBPositive0011821
Negative14121310
ViewerCPositive0011921
Negative14121300

Discordant Results of Methamphetamine Panel Dip

ViewerSample NumberGC/MS ResultPanel DipViewer Results
Viewer AMET60963Positive
Viewer BMET60963Positive
Viewer CMET60963Positive
Viewer AMET691100Negative
Viewer BMET151010Negative
Split-KeyCupNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)

{16}------------------------------------------------

Viewer0021921
APositive0021921
ANegative14121200
BPositive0011821
BNegative14121310
CPositive0011821
CNegative14121310

Discordant Results of Methamphetamine Split-Key Cup

ViewerSample NumberGC/MS ResultSplit-Key CupViewer Results
Viewer AMET05947Positive
Viewer AMET60તે જેવી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડPositive
Viewer BMET60તે જિલ્લામાં આવેલું એક ગામનાં મુખ્યત્વે ખેત-ઉત્પત્તમજૂરી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતોPositive
Viewer CMET60તે ઉત્તરPositive
Viewer BMET151010Negative
Viewer CMET151010Negative
Easy CupNegativeLow Negative by GC/MS(less than -50%)Near Cutoff Negative by GC/MS(Between -50% and cutoff)Near Cutoff Positive by GC/MS(Between the cutoff and +50%)High Positive by GC/MS(greater than +50%)
ViewerAPositive0001621
Negative14121430
ViewerBPositive0001721
Negative14121420
ViewerCPositive0001821
Negative14121410

Discordant Results of Methamphetamine Easy Cup

ViewerSample NumberGC/MS ResultEasy CupViewer Results
Viewer AMET651065Negative
Viewer AMET691100Negative
Viewer AMET151010Negative
Viewer BMET691100Negative
Viewer BMET151010Negative
Viewer CMET691100Negative

Morphine

CassetteNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)

{17}------------------------------------------------

Viewer
APositive0001623
Negative9191210
BPositive0021723
Negative9191000
CPositive0021723
Negative9191000

Discordant Results of Morphine Cassette

ViewerSample NumberGC/MS ResultCassette Viewer Results
Viewer BMOP53274Positive
Viewer BMOP01297Positive
Viewer CMOP53274Positive
Viewer CMOP01297Positive
Viewer AMOP43314Negative
Panel DipNegativeLow Negative by GC/MS(less than -50%)Near Cutoff Negative by GC/MS(Between -50% and cutoff)Near Cutoff Positive by GC/MS(Between the cutoff and +50%)High Positive by GC/MS(greater than +50%)
Viewer A Positive0001623
Viewer A Negative9191210
Viewer B Positive0001623
Viewer B Negative9191210
Viewer C Positive0001723
Viewer C Negative9191200

Discordant Results of Morphine Panel Dip

ViewerSample NumberGC/MS ResultPanel Dip Viewer Results
Viewer AMOP43314Negative
Viewer BMOP43314Negative
Split-KeyCupNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011623
Negative9191110
ViewerBPositive0011723
Negative9191100
ViewerCPositive0001723
Negative9191200

{18}------------------------------------------------

Discordant Results of Morphine Split-Key Cup
ViewerSample NumberGC/MS ResultSplit Cup Viewer Results
Viewer AMOP01297Positive
Viewer BMOP01297Positive
Viewer AMOP43314Negative

Discordant Results of Morphine Split-Ke Cun

EasyNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
Cup
ViewerPositive0021723
ANegative9191000
ViewerPositive0021723
BNegative9191000
ViewerPositive0011623
CNegative9191110

Discordant Results of Morphine Easy Cup

ViewerSample NumberGC/MS ResultEasy CupViewer Results
Viewer AMOP53274Positive
Viewer AMOP01297Positive
Viewer BMOP53274Positive
Viewer BMOP01297Positive
Viewer CMOP01297Positive
Viewer CMOP43314Negative
Marijuana
CassetteMarijuana
NegativeLow Negative by GC/MS(less than -50%)Near Cutoff Negative by GC/MS(Between -50% and cutoff)Near Cutoff Positive by GC/MS(Between the cutoff and +50%)High Positive by GC/MS(greater than +50%)
Viewer Positive0022416
A Negative8181200
Viewer Positive0012316
B Negative8181310
Viewer Positive0022416

{19}------------------------------------------------

CNegative8181200
Discordant Results of Marijuana Cassette
ViewerSample NumberGC/MS ResultCassetteViewer Results
Viewer ATHC0549Positive
Viewer ATHC6747Positive
Viewer BTHC0549Positive
Viewer CTHC0549Positive
Viewer CTHC6747Positive
Viewer BTHC7951Negative
PanelDipNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0002216
Negative8181420
ViewerBPositive0002216
Negative8181420
ViewerCPositive0002016
Negative8181440

Discordant Results of Marijuana Panel Dip

ViewerSample NumberGC/MS ResultPanel DipViewer Results
Viewer ATHC6652Negative
Viewer ATHC7951Negative
Viewer BTHC6652Negative
Viewer BTHC7951Negative
Viewer CTHC0453Negative
Viewer CTHC6652Negative
Viewer CTHC6852Negative
Viewer CTHC7951Negative
Split-KeyCupNegativeLow Negative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0012216
Negative8181320
ViewerBPositive0012316
Negative8181310
ViewerPositive0002316

{20}------------------------------------------------

CNegative8181410
Discordant Results of Marijuana Split-Key Cup
ViewerSample NumberGC/MS ResultSplit CupViewer Results
Viewer ATHC0549Positive
Viewer BTHC0549Positive
Viewer ATHC6852Negative
Viewer ATHC7951Negative
Viewer BTHC6852Negative
Viewer CTHC6852Negative
EasyCupNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0002316
Negative8181410
ViewerBPositive0002116
Negative8181430
ViewerCPositive0002316
Negative8181410

Discordant Results of Marijuana Easy Cup

ViewerSample NumberGC/MS ResultEasy CupViewer Results
Viewer ATHC6652Negative
Viewer BTHC0453Negative
Viewer BTHC6652Negative
Viewer BTHC7951Negative
Viewer CTHC1653Negative

Lay-user study

A lay user study was performed at three intended user sites with 1680 lay persons. The lay users had diverse educational and professional backgrounds and ranged in age from 20 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. Each device was tested.

Comparison between GC/MS and Lay Person Results for Methamphetamine Cassette

Lay person resultsThe

{21}------------------------------------------------

% of CutoffNumberofsamplesd-MethamphetamineConcentration by GC/MS(ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100
-75% Cutoff2025020100
-50% Cutoff20500020100
-25% Cutoff2075011995
+25% Cutoff201,25019195
+50% Cutoff201,500200100
+75% Cutoff201,750200100

Comparison between GC/MS and Lay Person Results for Methamphetamine Dip Card

% of CutoffNumberofsamplesd-MethamphetamineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
-100% Cutoff200020100
-75% Cutoff2025020100
-50% Cutoff20500020100
-25% Cutoff2075011995
+25% Cutoff201,25018290
+50% Cutoff201,500200100
+75% Cutoff201,750200100

Comparison between GC/MS and Lay Person Results for Methamphetamine Split-Key Cup

Numberofsamplesd-MethamphetamineConcentration by GC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100
-75%Cutoff2025020100
-50% Cutoff20500020100
-25% Cutoff2075011995
+25% Cutoff201,25018290
+50% Cutoff201,500200100
+75% Cutoff201,750200100

Comparison between GC/MS and Lay Person Results for Methamphetamine Easy Cup

% of CutoffNumberofsamplesd-MethamphetamineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
-100% Cutoff200No. ofPositiveNo. ofNegative100
-75% Cutoff2025020100
-50% Cutoff20500020100
-25% Cutoff2075021890
+25% Cutoff201,25019195

{22}------------------------------------------------

+50% Cutoff201,500200100
+75% Cutoff201,750200100
% of CutoffNumberofsamplesMorphine Concentrationby GC/MS (ng/mL)Lay person resultsThepercentage ofcorrect results(%)
-100% Cutoff200020100
-75% Cutoff2075020100
-50% Cutoff20150020100
-25% Cutoff2022511995
+25% Cutoff2037519195
+50% Cutoff20450200100
+75% Cutoff20525200100

Comparison between GC/MS and Lay Person Results for Morphine Cassette

Comparison between GC/MS and Lay Person Results for Morphine Dip Card

% of CutoffNumber ofsamplesMorphine Concentrationby GC/MS (ng/mL)Lay person resultsThepercentage ofcorrect results(%)
-100% Cutoff200020100
-75% Cutoff2075020100
-50% Cutoff20150020100
-25% Cutoff2022511995
+25% Cutoff2037519195
+50% Cutoff20450200100
+75% Cutoff20525200100

Comparison between GC/MS and Lay Person Results for Morphine Split-Key Cup

% of CutoffNumberofsamplesMorphine Concentrationby GC/MS (ng/mL)Lay person resultsThepercentage ofcorrect results(%)
-100% Cutoff200No. ofPositiveNo. ofNegative100
-75% Cutoff2075020100
-50% Cutoff20150020100
-25% Cutoff2022521890
+25% Cutoff2037518290
+50% Cutoff20450200100
+75% Cutoff20525200100

Comparison between GC/MS and Lay Person Results for Morphine Easy Cup

Lay person resultsThe
-----------------------------

{23}------------------------------------------------

% of CutoffNumberofsamplesMorphine Concentrationby GC/MS (ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100
-75% Cutoff2075020100
-50% Cutoff20150020100
-25% Cutoff2022521890
+25% Cutoff2037519195
+50% Cutoff20450200100
+75% Cutoff20525200100

Comparison between GC/MS and Lay Person Results for Marijuana Cassette

% of CutoffNumber of samplesMarijuana Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
-100% Cutoff200020100
-75% Cutoff2012.5020100
-50% Cutoff2025020100
-25% Cutoff2037.511995
+25% Cutoff2062.519195
+50% Cutoff2075200100
+75% Cutoff2087.5200100

Comparison between GC/MS and Lay Person Results for Marijuana Dip Card

% of CutoffNumber of samplesMarijuana Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
-100% Cutoff200No. of PositiveNo. of Negative100
-75% Cutoff2012.5020100
-50% Cutoff2025020100
-25% Cutoff2037.502010
+25% Cutoff2062.518290
+50% Cutoff2075200100
+75% Cutoff2087.5200100

Comparison between GC/MS and Lay Person Results for Marijuana Split-Key Cup

% of CutoffNumberofsamplesMarijuana Concentrationby GC/MS (ng/mL)Lay person resultsThe
No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100

{24}------------------------------------------------

-75% Cutoff2012.5020100
-50% Cutoff2025020100
-25% Cutoff2037.511995
+25% Cutoff2062.519195
+50% Cutoff2075200100
+75% Cutoff2087.5200100

Comparison between GC/MS and Lay Person Results for Marijuana Easy Cup

% of CutoffNumber of samplesMarijuana Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
-100% Cutoff200020100
-75% Cutoff2012.5020100
-50% Cutoff2025020100
-25% Cutoff2037.511995
+25% Cutoff2062.5200100
+50% Cutoff2075200100
+75% Cutoff2087.5200100

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

    1. Clinical Studies
      Not applicable.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison, and lay-user studies of the devices, it's concluded that the CLUNGENE Methamphetamine Tests, CLUNGENE Morphine Tests and CLUNGENE Marijuana Tests are substantially equivalent to the predicate.

§ 862.3870 Cannabinoid test system.

(a)
Identification. A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds includedelta -9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.(b)
Classification. Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).