K Number
K190076

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-09-27

(254 days)

Product Code
Regulation Number
866.6060
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Xpert BCR-ABL Ultra test is an in vitro diagnostic test for the quantitation of BCR-ABL1 and ABL1 mRNA transcripts in peripheral blood specimens of diagnosed t(9;22) positive Chronic Myeloid Leukemia (CML) patients expressing BCR-ABLI fusion transcripts type e13a2 and/or e14a2. The test utilizes automated, quantitative, real-time reverse transcription polymerase chain reaction (RT-qPCR). The Xpert BCR-ABL Ultra test is intended to measure BCR-ABL1 to ABL1 percent ratios on the International Scale (IS), and also expressed as a log molecular reduction (MR value) from a baseline of 100% (IS), in t(9;22) positive CML patients during of treatment with Tyrosine Kinase Inhibitors (TKIs).

The test does not differentiate between e13a2/b2a2 or e14a2/b3a2 fusion transcripts and does not monitor other rare fusion transcripts resulting from t(9;22). This test is not intended for the diagnosis of CML.

The Xpert BCR-ABL Ultra test is intended for use only on the Cepheid GeneXpert® Dx System and the GeneXpert Infinity System.

Device Description

The Xpert BCR-ABL Ultra test is an automated in vitro diagnostic test for quantifying the amount of BCR-ABLI (BCR-ABL, hereafter) mRNA transcript as a ratio of BCR-ABL/ABL per the International Scale (IS).

The test is performed on the Cepheid GeneXpert® Dx System and GeneXpert Infinity System (referred to as the GeneXpert systems). The GeneXpert systems require the use of single-use, disposable cartridges that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained and specimens never come into contact with working parts of the instrument modules, cross-contamination between samples is minimized. The GeneXpert systems have 1 to 80 randomly accessible modules, depending upon the instrument, that are each capable of performing separate sample preparation and real-time PCR and RT-PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), and a proprietary I-CORE® thermocycler for performing real-time PCR and RT-PCR and detection.

The Xpert BCR-ABL Ultra test includes reagents to detect BCR-ABL fusion genes resulting from two major breakpoints, translocation e13a2/b2a2 and e14a2/b3a2 and the ABL transcript as an endogenous control in peripheral blood specimens. The amount of BCR-ABL transcript in the patient sample is reported as a percent ratio of BCR-ABL/ABL on the International Scale (IS), and also expressed as a log molecular reduction (MR value) from a baseline of 100% (IS), using the GeneXpert software.

There are two controls included in each Xpert BCR-ABL Ultra test, which are the ABL Endogenous Control and the Probe Check Control (PCC). The ABL Endogenous Control normalizes the BCR-ABL target and ensures that sufficient sample is used in the test. The PCC verifies reagent rehydration. PCR tube filling, and that all reaction components, including probes and dyes, are present and functional in the cartridge.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Acceptance Criteria and Device Performance for Xpert® BCR-ABL Ultra

Note: The provided document is a 510(k) Summary, which typically focuses on demonstrating substantial equivalence to a predicate device rather than explicitly stating pre-defined "acceptance criteria" in a quantitative, pass/fail manner. For this analysis, I will infer relevant performance metrics from the non-clinical and clinical studies presented as evidence of substantial equivalence. The "Acceptance Criteria" column will represent the demonstrated performance that FDA found acceptable for classification, and the "Reported Device Performance" will be the specific results from the studies.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Inferred)Reported Device Performance (Xpert® BCR-ABL Ultra)
Linearity/Dynamic Range: - Linear regression R² for e13a2/b2a2 breakpoint. - Linear regression R² for e14a2/b3a2 breakpoint. - Maximum Standard Deviation (SD) within reportable range.- e13a2/b2a2: R² = 0.98304 - e14a2/b3a2: R² = 0.9788 - Maximum SD = 0.26 (within reportable range of MR0.26 to MR4.52)
Analytical Sensitivity (LoD): - LoD for e13a2/b2a2 breakpoint (target 95% positivity). - LoD for e14a2/b3a2 breakpoint (target 95% positivity). - Overall LoD claimed for both breakpoints.- e13a2/b2a2: 0.0030% (IS)/MR4.52 (95.74% positivity at this level) - e14a2/b3a2: 0.0029% (IS)/MR4.55 (96.04% positivity at this level) - Overall LoD: 0.0030% (IS)/MR4.52
Analytical Sensitivity (LoQ): - Equivalence to LoD. - MR Standard Deviation for e13a2/b2a2 and e14a2/b3a2 within defined limits (e.g., < 0.36).- LoQ determined to be equal to LoD: 0.0030% (IS)/MR4.52. - e13a2/b2a2 MR SD range: 0.27-0.34 - e14a2/b3a2 MR SD range: 0.29-0.31 (both within <0.36)
Analytical Sensitivity (LoB): - No measurable BCR-ABL values in healthy donor specimens.- 0.00% (IS)
Analytical Specificity: - No BCR-ABL signal in non-CML specimens. - Specificity for p210 BCR-ABL fusion transcript.- 100% analytical specificity for non-CML EDTA blood specimens. - Correct positive results for CML K562 and BV173 cell lines. - AR230 cell line (p230 e19a2 breakpoint) mostly negative, with one positive below LoD at higher concentration.
Interfering Substances: - % (IS)/MR ratio within 3-fold difference compared to control.- No clinically significant inhibitory effects; ratios within 3-fold range for tested substances (Bilirubin, Cholesterol, Triglycerides, Heparin, EDTA).
Carry-Over Contamination: - All negative samples following high positive samples reported correctly as NEGATIVE.- All 20 negative samples following high positive samples reported correctly as NEGATIVE [Sufficient ABL transcript].
Traceability to WHO Panel: - Slope of measured vs. published MR values close to unity (e.g., 0.96-1.1). - Intercept close to 0 (e.g., -0.03 to -0.06).- Slope: 0.96 to 1.1 - Intercept: -0.03 to -0.06
Clinical Performance (Comparison to Predicate): - Successful result rate. - Deming regression slope and intercept compared to predicate. - Bias at MMR (MR3). - Majority of results within 2SD of Bland-Altman mean difference.- Initial non-determinate rate: 2.9%; Final non-determinate rate: 0.5%. - Deming regression: Slope = 1.0266, Intercept = 0.0600. - Predicted bias at MR3: MR0.1244 (95% CI: 0.0969 - 0.1519). - 96.6% (142/147) of results within 2SD range (-0.5372 and 0.8044).
Reproducibility: - Total SD for MR1, MR2, MR3 levels. - Maximum Total SD for samples near LoD and MR4.- Total SD for MR1, MR2, MR3: ≤ 0.15. - Maximum Total SD for samples near LoD and MR4: 0.33.

2. Sample Size Used for the Test Set and Data Provenance

The primary clinical performance study used:

  • Total specimens initially enrolled: 266
  • Specimens available for analysis: 208 (after exclusions)
  • Specimens within quantitative reportable range for both assays: 147
    • Frozen specimens: 117
    • Fresh, prospectively collected specimens: 30

Data Provenance:

  • Country of Origin: U.S. (evaluated at four institutions in the U.S. for the clinical study).
  • Retrospective/Prospective: The study included both:
    • Prospective: Fresh, prospectively collected EDTA whole blood specimens from CML patients.
    • Retrospective/Archived: Leftover specimens stored as frozen lysates (from the same patient population).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document states that the Xpert BCR-ABL Ultra test performance was compared to an "FDA-cleared molecular assay" (the predicate device). This implies the ground truth for the clinical study was established by the predicate device's results.

  • Number of Experts: Not applicable in the sense of human experts reviewing diagnostic images or pathology slides. The ground truth was based on the quantitative results from the predicate molecular assay.
  • Qualifications of Experts: Not applicable, as the "ground truth" was an analytical measurement from another device, not human expert consensus.

4. Adjudication Method for the Test Set

Not applicable. The study compares the new device's quantitative results directly against those of an existing FDA-cleared molecular assay. There is no mention of a human adjudication process for discrepancies between the two assays; rather, statistical methods (Deming regression, Bland-Altman) are used to assess agreement and bias.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. An MRMC study is typically performed for diagnostic imaging devices where human readers interpret cases, and AI assistance might improve their performance. This device is an in vitro diagnostic (IVD) based on RT-qPCR, where the result is generated automatically by the instrument, not interpreted by human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant or applicable for this type of device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the studies described are standalone performance evaluations of the Xpert BCR-ABL Ultra system. The system performs automated sample preparation, RT-qPCR, and result interpretation without human intervention in the result generation process. The "Non-Clinical Studies" (Linearity, LoD, LoQ, Specificity, Interfering Substances, Carry-Over, Traceability) and the "Clinical Performance" study (comparison to predicate device) all evaluate the device's performance as an automated, standalone system.

7. The Type of Ground Truth Used

The ground truth used for the clinical performance evaluation was the quantitative results obtained from an FDA-cleared molecular assay (the predicate device). The predicate device measures BCR-ABL1 and ABL1 mRNA transcripts using similar RT-qPCR principles.

8. The Sample Size for the Training Set

The document does not explicitly state the sample size for a "training set" for the Xpert BCR-ABL Ultra assay. This is common for IVD assays based on established PCR technology, where the assay design is based on molecular biology principles and analytical validation, rather than machine learning models requiring large datasets for "training." The method is "Reverse transcription, quantitative, polymerase chain reaction (RT-qPCR) based nucleic acid amplification," which is a deterministic analytical method, not an AI/ML algorithm that is "trained" on data in the traditional sense.

9. How the Ground Truth for the Training Set Was Established

As noted in point 8, the concept of a "training set" and "ground truth for training" in the context of machine learning (AI/ML) does not directly apply to this device. The development of an IVD assay like Xpert BCR-ABL Ultra involves:

  • Assay Design: Primers and probes are designed based on known genetic sequences (BCR-ABL1 and ABL1 mRNA transcripts).
  • Analytical Validation: Extensive non-clinical studies (linearity, LoD, LoQ, specificity, interference, etc.) are performed using characterized samples (e.g., CML patient specimens, spiked controls, cell lines, WHO reference panels) to establish the assay's performance characteristics. This is a process of characterizing the assay, not "training" it.

The "ground truth" for these analytical validation steps would be:

  • Known concentrations in serially diluted samples.
  • Characterized cell lines with specific breakpoints.
  • Reference materials (e.g., WHO International Genetic Reference Panel) with assigned values.
  • Negative controls (healthy donor blood, non-CML leukemic specimens).

These ground truths are established through laboratory preparation, internationally recognized standards, and biological characterization, rather than human expert consensus on a dataset.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which stands for U.S. Food & Drug Administration. The FDA logo is in blue.

September 27, 2019

Cepheid Sudhakar Marla, Ph.D. Senior Director, Regulatory Affairs 904 Caribbean Drive Sunnyvale, California 94089

Re: K190076

Trade/Device Name: Xpert BCR-ABL Ultra, GeneXpert Dx System, GeneXpert Infinity-48s and GeneXpert Infinity-80 Systems Regulation Number: 21 CFR 866.6060 Regulation Name: BCR-ABL quantitation test Regulatory Class: Class II Product Code: OYX, OOI Dated: August 15, 2019 Received: August 19, 2019

Dear Sudhakar Marla:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

gencies. You must comply with all the Act’s
provisions (21 CFR Part 807) labeling (21 CFR Part

Page 2

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Reena Philip, Ph.D. Director Division of Molecular Genetics and Pathology OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190076

Device Name Xpert® BCR-ABL Ultra

Indications for Use (Describe)

The Xpert BCR-ABL Ultra test is an in vitro diagnostic test for the quantitation of BCR-ABL1 and ABL1 mRNA transcripts in peripheral blood specimens of diagnosed t(9;22) positive Chronic Myeloid Leukemia (CML) patients expressing BCR-ABLI fusion transcripts type e13a2 and/or e14a2. The test utilizes automated, quantitative, real-time reverse transcription polymerase chain reaction (RT-qPCR). The Xpert BCR-ABL Ultra test is intended to measure BCR-ABL1 to ABL1 percent ratios on the International Scale (IS), and also expressed as a log molecular reduction (MR value) from a baseline of 100% (IS), in t(9;22) positive CML patients during of treatment with Tyrosine Kinase Inhibitors (TKIs).

The test does not differentiate between e13a2/62a2 or e14a2/63a2 fusion transcripts and does not monitor other rare fusion transcripts resulting from t(9;22). This test is not intended for the diagnosis of CML.

The Xpert BCR-ABL Ultra test is intended for use only on the Cepheid GeneXpert® Dx System and the GeneXpert Infinity System.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

As required by 21 CFR Section 807.92(c).

Submitted by:Cepheid904 Caribbean DriveSunnyvale, CA 90489Phone number: (408) 548-8946
Contact:Sudhakar Marla, Ph.D.
Date of Preparation:September 24, 2019
Device:
Trade name:Xpert® BCR-ABL Ultra
Common name:Xpert BCR-ABL Ultra
Type of Test:Reverse transcription, quantitative, polymerase chain reaction(RT-qPCR) based nucleic acid amplification
Regulation number,Classification name,Product code:21 CFR 866.6060, BCR-ABL quantitation test, OYX21 CFR 862.2570, Instrumentation for clinical multiplex testsystems, OOI
ClassificationAdvisory PanelPathology (88)
Prescription UseYes
Predicate DeviceAsuragen QuantideX qPCR BCR-ABL IS Kit[DEN160003]

{4}------------------------------------------------

Device Intended Use:

The Xpert BCR-ABL Ultra test is an in vitro diagnostic test for the quantitation of BCR-ABL1 and ABL1 mRNA transcripts in peripheral blood specimens of diagnosed t(9:22) positive Chronic Myeloid Leukemia (CML) patients expressing BCR-ABLI fusion transcripts type e13a2 and/or e14a2. The test utilizes automated, quantitative, real-time reverse transcription polymerase chain reaction (RT-qPCR). The Xpert BCR-ABL Ultra test is intended to measure BCR-ABL1 to ABL1 percent ratios on the International Scale (IS), and also expressed as a log molecular reduction (MR value) from a baseline of 100% (IS), in t(9:22) positive CML patients during monitoring of treatment with Tyrosine Kinase Inhibitors (TKIs).

The test does not differentiate between e13a2/b2a2 or e14a2/b3a2 fusion transcripts and does not monitor other rare fusion transcripts resulting from t(9;22). This test is not intended for the diagnosis of CML.

The Xpert BCR-ABL Ultra test is intended for use only on the Cepheid GeneXpert® Dx System and GeneXpert Infinity System.

Special conditions for use statement(s):

For in vitro diagnostic use only

For Prescription use only

Device Description:

The Xpert BCR-ABL Ultra test is an automated in vitro diagnostic test for quantifying the amount of BCR-ABLI (BCR-ABL, hereafter) mRNA transcript as a ratio of BCR-ABL/ABL per the International Scale (IS).

The test is performed on the Cepheid GeneXpert® Dx System and GeneXpert Infinity System (referred to as the GeneXpert systems). The GeneXpert systems require the use of single-use, disposable cartridges that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained and specimens never come into contact with working parts of the instrument modules, cross-contamination between samples is minimized. The GeneXpert systems have 1 to 80 randomly accessible modules, depending upon the instrument, that are each capable of performing separate sample preparation and real-time PCR and RT-PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), and a proprietary I-CORE® thermocycler for performing real-time PCR and RT-PCR and detection.

{5}------------------------------------------------

The Xpert BCR-ABL Ultra test includes reagents to detect BCR-ABL fusion genes resulting from two major breakpoints, translocation e13a2/b2a2 and e14a2/b3a2 and the ABL transcript as an endogenous control in peripheral blood specimens. The amount of BCR-ABL transcript in the patient sample is reported as a percent ratio of BCR-ABL/ABL on the International Scale (IS), and also expressed as a log molecular reduction (MR value) from a baseline of 100% (IS), using the GeneXpert software.

There are two controls included in each Xpert BCR-ABL Ultra test, which are the ABL Endogenous Control and the Probe Check Control (PCC). The ABL Endogenous Control normalizes the BCR-ABL target and ensures that sufficient sample is used in the test. The PCC verifies reagent rehydration. PCR tube filling, and that all reaction components, including probes and dyes, are present and functional in the cartridge.

A description of the reagents provided in the kit is described below in Table 1.

ItemDescriptionUse
Proteinase KSerine proteaseDigests proteins andinactivates nucleases in EDTAwhole blood specimen duringthe sample preparation andnucleic acid purification steps
Lysis ReagentGuanidinium chloridebuffered solutionDenaturant used to lyse cells,release of nucleic acids anddecrease nuclease activityduring the sample preparationstep
Wash ReagentGuanidinium thiocyanate andethanol solutionReagent used to removecellular contaminants duringnucleic acid binding step
Xpert BCR-ABL UltraCartridgeswith Integrated ReactionTubesSingle-use test cartridges thathouse, buffered solutions(rinse and elution),lyophilized beads containingreverse transcriptase, DNApolymerase, primers andprobesReagents used to perform theon-board nucleic acidisolation, purification and real-time RT-qPCR.
CD/SoftwareCompact discContains assay definition file(ADF), instruction to importADF into GeneXpert softwareand information for use

Table 1. Reagents in the Xpert BCR-ABL Ultra Test Kit

{6}------------------------------------------------

The peripheral blood specimens are collected in EDTA blood tubes following the user institution's standard procedures and can be stored for up to 72 hours prior to use when stored at 4℃. The test sample, from a peripheral blood specimen, is prepared in a sample preparation tube provided with the kit, with reagents that are provided with the kit (proteinase K and lysis reagent) plus user-supplied ethanol, following instructions in the package insert. After sample preparation, the prepared sample and one reagent provided with the kit (Wash Reagent) are pipetted to separate chambers of the Xpert BCR-ABL Ultra cartridge. The user initiates a test from the system user interface and places the cartridge into the GeneXpert instrument platform, which performs hands-off additional sample preparation and nested, real-time, quantitative reverse transcription polymerase chain reaction (RT-qPCR) for detection of RNA. In this platform, the additional sample preparation, amplification, and real-time detection are all fully-automated and completely integrated. The time to result for the Xpert BCR-ABL Ultra test including offline sample preparation and real-time RT-qPCR is approximately 2.25 hours.

Substantial Equivalence:

The Xpert BCR-ABL Ultra is substantially equivalent to the Asuragen QuantideX qPCR BCR-ABL IS Kit [DEN160003]. The performance of the Xpert BCR-ABL Ultra test was evaluated in a multi-site clinical study using whole blood specimens and was compared to a FDA-cleared molecular assay. The results from the clinical study demonstrated that the performance of Xpert BCR-ABL Ultra is substantially equivalent to the predicate device.

Table 2 shows the similarities and differences between the Xpert BCR-ABL Ultra test and the predicate device.

{7}------------------------------------------------

Similarities
ItemDevicePredicate
Intended UseFor the quantitation of BCR-ABL1 and ABL mRNAtranscripts in peripheral bloodspecimens of diagnosed t(9;22)positive Chronic MyeloidLeukemia (CML) patientsexpressing BCR ABL1 fusiontranscripts type e13a2 and/ore14a2.Same
TargetPopulation/IndicationCMP-positive patients duringmonitoring of treatment withTyrosine Kinase Inhibitors(TKIs)Same
Test LimitationThe test does not differentiatebetween e13a2/b2a2 ore14a2/b3a2 fusion transcripts anddoes not monitor other rare fusiontranscripts resulting from t(9;22).This test is not intended for thediagnosis of CML.Same
MeasurandBCR-ABL1 fusion transcripts(e13a2/b2a2 and/or e14a2/b3a2)and the ABL1 endogenouscontrol mRNASame
Measurement typeQuantitativeSame
Principle of AssayReverse transcription,quantitative, polymerase chainSame
Similarities
ItemDevicePredicate
reaction (RT-qPCR) basednucleic acid amplification
Traceability Standard1st World Health Organization(WHO) International GeneticReference Panel for quantitationof BCR-ABL translocation byRQ-PCRSame
UnitsBoth % (IS) and MolecularResponse (MR)Same
Specimen TypeWhole Blood (EDTA)Same

Table 2: Comparison of Similarities and Differences of Xpert BCR-ABL Ultra with the Predicate Device

{8}------------------------------------------------

Differences
ItemDevicePredicate
Extraction and AssayPreparationSingle Use cartridgeExtraction steps may bemanual followed bymanual assaypreparation
InstrumentCepheid GeneXpert®DxGeneXpert Infinity-48s,and GeneXpert Infinity-80Applied Biosystems7500 Fast Dx Real TimePCR Instrument
Input RangeRNA is isolated fromwhite blood cell countRNA input range of 1 to5µg

{9}------------------------------------------------

Differences
ItemDevicePredicate
(WBCC) range$1.5 x 10^5$ to $3.0 x 10^7$cells/mL
ControlsProbe Check ControlThree controls RNAHigh (MR 1.5), mRNALow (MR 3.5), and RNANegative. Each providedin separate tubes
CalibratorsNone provided. Each lotof Xpert BCR-ABLUltra is calibrated tosecondary standards thatwere calibrated to theWorld HealthOrganization (WHO)international geneticreference panel forquantitation of BCR-ABL transcript.Four levels formulatedto MR1.0, 2.0, 3.0, 4.0traceable to the WorldHealth Organization(WHO) internationalgenetic reference panelfor quantitation ofBCR-ABL transcript.

The Xpert BCR-ABL Ultra test has the same general intended use as the predicate device and has the same technological characteristics as the predicate device. The differences between the Xpert BCR-ABL Ultra test and the predicate device do not raise different questions of safety and effectiveness. The clinical study demonstrates that the Xpert BCR-ABL Ultra test is acceptable for its intended use and is substantially equivalent to the predicate device described above.

{10}------------------------------------------------

Non-Clinical Studies:

Linearity/Dynamic Range

Linearity was evaluated independently for each of the two major breakpoints, e13a2/b2a2 and e14a2/b3a2, using CML clinical specimens that were specific for a high level of either the e13a2/b2a2 or e14a2/b3a2 breakpoint. Lysate from each high level of BCR-ABL transcript CML specimen was diluted in a background lysate prepared from CML negative clinical specimen to target ranges of ~50% (IS)/MR0.30 to 0.000625% (IS)/MR5.20. The panel members, including the negative level, were tested on two assay kit lots in replicates of 4 per kit lot.

Testing and statistical analyses were conducted in accordance with CLSI EP06-A. Linear regression analyses were performed for first, second and third order polynomials. The results for each breakpoint were considered linear if the polynomial regression coefficients were insignificant (p-values > 0.05). The linear regression curves for both transcripts are shown in Figure 1 and Figure 2 below.

Image /page/10/Figure/4 description: The figure shows linear regression curves for breakpoint transcript e13a2/b2a2. The x-axis represents the expected MR, while the y-axis represents the output MR and output BCR-ABL % (IS). The plot includes data points and three regression models: simple linear regression, second-order polynomial, and third-order polynomial. Equations for each model are displayed, such as y = -0.05833 + 0.99501x with R^2 = 0.98304 for the simple linear regression model.

Figure 1: Linear Regression Curves for Breakpoint Transcript

{11}------------------------------------------------

Image /page/11/Figure/0 description: The image is a scatter plot comparing "Output_MR" to "e14a2/b3a2 Expected_MR". The plot includes data points labeled as "Output MR" and three trendlines: "Simple Linear Regression Model", "Second-Order Polynomial Model", and "Third-Order Polynomial Model". Equations for each trendline are displayed, such as "y = 0.03647 + 1.03153x, R2 = 0.97876" for the linear regression model.

Figure 2: Linear Regression Curves for Breakpoint Transcript e14a2/b3a2

The estimated regression intercepts, slopes and R2 values from the linear model are shown in Table 3.

BreakpointInterceptSlopeR2
e13a2/b2a2-0.058330.995010.98304
e14a2/b3a20.036471.031530.9788

Table 3: Regression Coefficients from Linear Model

{12}------------------------------------------------

Collectively, the data support an observation of linearity from at least 55% (IS)/MR 0.26 to ~0.0019% (IS)/MR4.75 with a maximum SD of 0.26. The reportable range spans from the limits of linearity at 55% (IS)/MR0.26 to the LoD/LOQ at 0.0030% (IS)/MR4.52.

Analytical Sensitivity

The limit of detection (LoD) was estimated for both e13a2/b2a2 and e14a2/b3a2 breakpoints by testing serial dilutions of High CML positive specimens [>10% (IS)/MR1] as well as testing Low CML positive specimens [<0.1% (IS)/MR3]. Data for each breakpoint across dilutions and specimens were separately compiled and the LoD was estimated by using probit regression analysis. The resulting analysis yielded an estimated LoD of 0.0035% (IS)/MR4.45 for the e13a2/b2a2 breakpoint and 0.0030% (IS)/MR4.52 for the e14a2/b3a2 breakpoint.

The LoD was verified by adapting the non-parametric method described in the CLSI guidance document, EP17-A2 (Table 4). Two unique CML positive specimens representing each breakpoint were diluted to a targeted 0.0030% (IS)/MR4.52 level. For e13a2/b2a2, 94 replicates were tested by 2 operators across 4 test kit lots over 4 days. For e14a2/b3a2, 101 replicates were tested by 2 operators across 4 test kit lots over 7 days.

BreakpointPositives/Replicates% of PositivesMedian% (IS)/MR
e13a2/b2a290/9495.74%0.0030% (IS)/MR4.52
e14a2/b3a297/10196.04%0.0029% (IS)/MR4.55

Table 4: Verified Limit of Detection in % (ZS)/MR

Since the Xpert BCR-ABL Ultra test does not distinguish between the two breakpoints. e13a2/b2a2 and e14a2/b3a2, the higher of the two is claimed as the assay LoD. Thus, the overall Xpert BCR-ABL Ultra LoD for both e13a2/b2a2 and e14a2/b3a2 is 0.0030% (IS)/MR4.52.

The limit of quantitation (LoQ) was estimated with the data obtained from the LoD studies. The mean and standard deviation for the % (IS) values and MR values were calculated for replicates at levels equal to the LoD, 0.0030% (IS)/MR4.52, or greater with positivity greater or equal to 95%. The LoQ of the test is constrained by the LoD of the test; therefore, the LoQ was determined to be equal to the LoD, 0.0030% (IS)/MR4.52. The results were also evaluated against the acceptance criteria for standard deviation (SD) < 0.36. The MR standard deviation for both e13a2/b2a2 (observed SD range MR0.27-MR0.34) and e14a2/b3a2 (observed SD range MR0.29-MR0.31) were within the acceptance criteria.

The limit of blank (LoB) was determined with 50 presumptive non-CML, normal healthy

{13}------------------------------------------------

donor blood specimens, drawn into EDTA tubes. No measurable BCR-ABL values were observed for any of the tests. Thus, the overall LoB was determined to be 0.00% (IS).

Analytical Specificity

The analytical and clinical specificity of Xpert BCR-ABL Ultra was evaluated for exclusivity by analyzing EDTA whole blood specimens drawn from fifty (50) healthy donors (non-CML) and twenty (20) leukemic specimens (AML/ALL). Breakpoint specificity was determined by testing normal healthy donor EDTA blood spiked with five (5) different leukemia cell lines representing 3 different types of leukemia (CML, ALL and APL) and 5 disease-breakpoints: K562 (CML/e14a2/b3a2) and BV173 (CML/e13a2/b2a2) served as positive controls; SUP-B15 (ALL/e1a2), AR230 (CML/e19a2) and NB4 (APL/PML-RARA) were evaluated for specificity.

No BCR-ABL signal was detected by Xpert BCR-ABL Ultra in any of the healthy non-CML specimens or AML/ALL leukemic specimens evaluated in this study.

Among the leukemia cell lines tested, CML cell lines (K562 and BV173) with p210 major breakpoints yielded the expected positive results. The CML cell line (AR230) with the p230 e19a2 breakpoint reported POSITIVE [Below LoD: >MR4.52/<0.0030% (IS)] for 1 of 4 replicates tested at the targeted 10% (IS)/MR1.00 level based on the number of K562 cell. The positive result for the AR230 cell line was for a target level 3.52 logs above assay LoD and was not observed at the lower levels of 1% (IS)/MR2.00 and 0.1% (IS)/MR3.00.

Xpert BCR-ABL Ultra is specific to the p210 BCR-ABL fusion transcript associated with CML and has an analytical specificity of 100% for non-CML EDTA blood specimens.

Potentially Interfering Substances Study

This study evaluated five substances that may be present in EDTA whole blood specimens with the potential to interfere with the performance of the Xpert BCR-ABL Ultra test. The compounds and levels tested (Table 5) were based on guidance from the CLSI document EP07-A2. Interferents were tested in the background of CML clinical EDTA whole blood specimens representing three levels with five specimens per level: >1% (IS)/<MR2, 0.1-1% (IS)/MR3-MR2, and <0.1% (IS)/>MR3). Test controls consisted of CML clinical specimens in EDTA whole blood at the respective BCR-ABL transcript level without the interfering substance. Each CML specimen was tested in the absence and presence of the five individual interferents at 4 replicates per condition.

A substance was considered non-interfering if in its presence the % mean (IS)/MR ratio observed was within 3-fold difference when compared to the control.

No clinically significant inhibitory effects on the Xpert BCR-ABL Ultra test were observed with any of the interfering substances evaluated in this study. Although some variability and statistically significant differences (p-value <0.05) in some tested

{14}------------------------------------------------

conditions were observed, the reported % (IS)/MR ratios for test and control conditions were within the acceptable 3-fold range.

Interfering SubstancesConcentration Tested
Unconjugated Bilirubin20 mg/dL
Cholesterol, Total500 mg/dL
Triglycerides, Total (Lipids)1800 mg/dL
Heparin3500 U/L
EDTA (short draw)750 mg/dL (5X)

Table 5: Potentially Interfering Substances Tested Using the 4 Dod ADD 114

Carry-Over Contamination

A study was conducted to demonstrate that single-use, self-contained GeneXpert cartridges prevent carry-over contamination from cartridges run sequentially in the same module. To demonstrate this, negative samples were run following very high positive samples in the same GeneXpert module. This study consisted of processing a NEGATIVE EDTA normal sample (CML-negative blood) in the same GeneXpert module immediately following a high POSITIVE sample (simulated CML positive blood) with 4.5 x 105 cells/mL of K562 cells spiked into CML-negative blood to vield ~10% (IS)/MR1.00. This testing sequence was repeated five times on each of the four GeneXpert modules. All twenty BCR-ABL positive samples were correctly reported as POSITIVE [#.##% (IS) and MR#.##], while all twenty BCR-ABL negative samples were correctly reported as NEGATIVE [Sufficient ABL transcript].

Traceability to WHO Panel

Traceability to the 1st World Health Organization (WHO) International Genetic Reference Panel for quantitation of BCR-ABL translocation by RQ-PCR (NIBSC code: 09/138) was demonstrated by measuring the WHO Reference Panel with 3 lots of the Xpert BCR-ABL Ultra test and comparing the measured values to the values published in the Reference Panel's Instructions for Use. Each of the 4 Reference Panel members was tested with a minimum of 10 replicates per assay kit lot. The measured MR values for each level of the WHO Primary panel were calculated by regression to each lot of the Xpert BCR-ABL Ultra test (i.e., the WHO panel members were treated as clinical samples and fit to the linear regression model of the assay's standard curve). Furthermore, the measured MR values were compared to the published MR values through an additional regression analysis to determine slope and intercept values (Figure 3). The slope of the line was close to unity (0.96 to 1.1) and the intercept was calculated to be close to 0 (-0.03 to -0.06).

{15}------------------------------------------------

Image /page/15/Figure/0 description: The image contains three scatter plots, labeled as Lot 1, Lot 2, and Lot 3. Each plot shows a linear relationship between measured values and concentrations, with data points for MR4.0 (0.01%), MR3.0 (0.1%), MR2.0 (1.0%), and MR1.0 (10%). The plots include regression equations and R-squared values, with Lot 1 having y = -0.06 + 1.0x and R2 = 0.0992, Lot 2 having y = -0.04 + 0.96x and R2 = 0.0992, and Lot 3 having y = -0.03 + 1.1x and R2 = 0.0994.

Published %(/S)/MR Values for WHC

Figure 3. Measured vs Published Values for WHO Primary Reference Panel, Lot-to-Lot.

Xpert BCR-ABL Ultra test kit-generated MR values (y-axis) are plotted against the MR values published in the WHO Primary Reference Panel's Instruction for Use (x-axis). The three lots are represented by (black) data points. Regression analyses and confidence intervals are based upon data for each lot separately.

Clinical Studies

Clinical Performance

The clinical performance of the Xpert BCR-ABL Ultra test was evaluated at four institutions in the U.S as part of a multi-site clinical study. Three additional institutions served as specimen collection only sites. The study was conducted using fresh, prospectively collected EDTA whole blood specimens from patients with CML at any stage of disease, following initial diagnosis, with or without prior exposure to Tyrosine Kinase Inhibitor therapy or other CML treatment. In addition, the study included left over specimens stored as frozen lysates which were prepared from EDTA whole blood from the same patient population. The Xpert BCR-ABL Ultra test performance was compared to a FDA-cleared molecular assay which detects and quantifies the mRNA transcripts for the p210 translocation types (e13a2/b2a2 or e14a2/b3a2) and uses the ABL as the endogenous control mRNA transcript.

A total of 266 eligible specimens were initially enrolled in the study, from which 57 were excluded due to use of an obsoleted procedure for extraction method (27), subject did not complete blood draw (8), shipping or testing delay (6), insufficient volume for testing (6), comparator test failed (6) or testing with incorrect Xpert BCR-ABL Ultra assay definition file (4) leaving 209 specimens which were tested.

Of 209 specimens, 97.1% (203/209) of the Xpert BCR-ABL Ultra results were successful on the first attempt giving an initial non-determinate rate of 2.9% (6/209) and 99.5%

{16}------------------------------------------------

(208/209) were successful upon retesting giving a final non-determinate rate of 0.5% (1/209).

Of the 208 specimens available for analysis, 150 (72.1%) were frozen specimens and 58 (27.9%) were fresh, prospectively collected specimens, for which demographic information was available. Among the fresh specimens, 24 (41.4%) were collected from female subjects and 34 (58.6%) from male subjects. The mean subject age for those providing fresh specimens was 60.5 years (range 28-85 years).

Of the 208 results that were available for analysis, 147 had results that were within the quantitative reportable range for both assays [0.0030% - 55% (IS)/MR4.52 - MR0.26 for Xpert BCR-ABL Ultra and 0.002% - 50% (IS)/MR4.72 - MR0.30 for the Comparator Assay]: 117 of which were from frozen left-over lysates and 30 of which were fresh prospectively collected specimens. The performance of the Xpert BCR-ABL Ultra test versus the Comparator Assay was evaluated using a Deming regression to determine the slope and intercept. Figure 4 shows the Deming regression and linear regression analysis of the 147 assay results (MR values).

Image /page/16/Figure/3 description: This scatter plot shows the correlation between two different assays, Xpert BCR-ABL Ultra MR and Comparator Assay MR. The x-axis represents the Comparator Assay MR, while the y-axis represents the Xpert BCR-ABL Ultra MR. The plot includes two regression lines: a Deming regression (Y=0.0600+1.0266X) and a linear regression (Y=0.2504+0.9580X), with an R-value of 0.9348 and an R-squared value of 0.8738.

Figure 4: Deming and Linear Regression Analyses

{17}------------------------------------------------

The slope and intercept from the Deming regression were 1.0266 and 0.0600 respectively. From these results, the predicted bias at the MMR (MR3) was calculated to be MR0.1244 (95% confidence interval of 0.0969 - 0.1519).

A Bland-Altman difference analysis was also performed using the 147 quantitative results that were within the reportable range for both the Xpert BCR-ABL Ultra test and Comparator Assay. The Bland-Altman graph (Figure 5) shows the upper and lower 2SD of the mean difference that was observed. The trend line of the bias across the MR range is also shown.

Image /page/17/Figure/2 description: This image is a scatter plot that compares two different methods of measuring MR. The x-axis represents the mean of the two methods, while the y-axis represents the difference between the two methods. The plot includes a regression line, as well as horizontal lines at 0, 0.8044, and -0.5372. The data points are scattered around the regression line, with some points falling outside the horizontal lines.

Figure 5: Xpert BCR-ABL Ultra Test MR vs Comparator Assay BCR-ABL MR Bland-Altman Difference Analysis

The mean difference (bias) was calculated to be 0.1336 with a SD of 0.3354. The majority (96.6%, 142/147) of the results were within the 2SD range (between -0.5372 and 0.8044).

Precision and Reproducibility Study

The precision and reproducibility of the Xpert BCR-ABL Ultra test was evaluated in a multisite study in accordance with CLSI EP05-A3, "Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline" and CLSI EP15-A3, "User

{18}------------------------------------------------

Verification of Performance for Precision and Trueness, Approved Guideline".

A panel of eleven samples was prepared which included the following: One sample negative for BCR-ABL, two samples near the limit of detection (LoD) and eight samples at molecular response (MR) levels 1-4, using the two targets detected by the Xpert BCR-ABL Ultra test: e13a2/b2a2 and e14a2/b3a2. The sample panel was made by diluting a bulk lysate of high %BCR-ABL/ABL specimens from patients with CML into pooled whole blood collected from healthy donors to obtain the desired level.

Table 6 shows the eleven samples included in this study.

SampleNo.Description% (IS)
1MR1.0 e13a2/b2a2BCR-ABL at ~ 10% (IS)
2MR1.0 e14a2/b3a2BCR-ABL at ~ 10% (IS)
3MR2.0 e13a2/b2a2BCR-ABL at ~ 1% (IS)
4MR2.0 e14a2/b3a2BCR-ABL at ~1% (IS)
5MR3.0 e13a2/b2a2BCR-ABL at ~ 0.1% (IS)
6MR3.0 e14a2/b3a2BCR-ABL at ~0.1% (IS)
7MR4.0 e13a2/b2a2BCR-ABL at ~ 0.01% (IS)
8MR4.0 e14a2/b3a2BCR-ABL at ~0.01% (IS)
9Near LoD e13a2/b2a2BCR-ABL at ~ 0.005% (IS)
10Near LoD e14a2/b3a2BCR-ABL at ~ 0.005% (IS)
11NegativeBCR-ABL Not Detected

Table 6: Reproducibility Panel for Xpert BCR-ABL Ultra

Each of the eleven panel members was tested in duplicate two times per day on four different days by each of three different operators at three different sites. Three lots of Xpert BCR ABL kits were used and each operator performed testing with one lot (3 sites x 3 lots x 1 operator/lot x 4 days x 2 runs/operator x 2 replicates/run = 144 replicates/panel member).

The quantitative results were analyzed by Analysis of Variance (ANOVA) and the major components of variance were identified.

The ANOVA analysis for each panel member are shown in Table 7.

{19}------------------------------------------------

SampleNMean(MR)Site/InstrumentSDOperator/LotSDDaySDWithin-runSDTotalSDa
Target MR1.0e13a2/b2a21440.9600.050.010.060.08
Target MR1.0e14a2/b3a21440.9900.0600.080.1
Target MR2.0e13a2/b2a21432.0400.060.020.100.11
Target MR2.0e14a2/b3a21442.090.030.070.020.100.13
Target MR3.0e13a2/b2a21442.890.060.040.030.100.12
Target MR3.0e14a2/b3a21443.120.060.0800.110.15
Target MR4.0e13a2/b2a2143b3.670.030.0200.150.15
Target MR4.0e14a2/b3a21443.910.050.080.040.140.17
Target nearMR 4.0e13a2/b2a2140c4.360.040.0400.330.33
Target nearMR 4.0e14a2/b3a2143d4.220.030.0800.170.19

Table 7: Reproducibility Study: Results from Analysis of Variance

a. The Xpert BCR-ABL Ultra test performed on the GeneXpert Instrument Systems integrates sample purification and nucleic acid amplification. The overall variability of the test observed in this study (expressed as Total SD) includes variability contributed by both the on-board sample preparation and RT-qPCR steps.

b. One replicate meeting the outlier requirements at the 99% level per CLSI EP15-A3 was removed from the analysis.

c. 4 samples out of the 144 test results yielded a NEGATIVE result

d. 1 sample out of the 144 test results yielded a NEGATIVE result

The observed total standard deviation for samples at MR1, MR2 and MR3 was ≤ 0.15. The maximum observed total standard deviation for samples near the LoD and MR4 was 0.33.

Conclusions

The results of the nonclinical analytical and clinical performance studies summarized above demonstrate that the Xpert BCR-ABL Ultra is substantially equivalent to the predicate device.

§ 866.6060 BCR-ABL quantitation test.

(a)
Identification. A BCR-ABL quantitation test is identified as a reverse transcription-quantitative polymerase chain reaction (RT-qPCR) test for the quantitation of BCR-ABL1 expressed on the International Scale (IS) and control transcripts in total RNA from whole blood of diagnosed t(9;22) positive chronic myeloid leukemia (CML) patients during monitoring of treatment with tyrosine kinase inhibitors. This test is not intended for the diagnosis of CML.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indication for use must indicate the variant(s) for which the assay was designed and validated, for example BCR-ABL e13a2 and/or e14a2.
(ii) A detailed description of all components in the test, including the following:
(A) A detailed description of the test components, all required reagents, instrumentation and equipment, including illustrations or photographs of non-standard equipment or methods;
(B) Detailed documentation of the device software including, but not limited to, standalone software applications and hardware-based devices that incorporate software;
(C) Methodology and protocols for control procedures for the assay to allow reporting on the International Scale;
(D) A description of the result outputs, analytical sensitivity of the assay, and the range of values that will be reported; and
(E) A description of appropriate internal and external controls that are recommended or provided. The description must identify those control elements that are incorporated into the testing procedure.
(iii) Information that demonstrates the performance characteristics of the test, including:
(A) For indications for use based on a threshold established in a predicate device of this generic type, device performance data from either a method comparison study to the predicate device or through a clinical study demonstrating clinical validity using well-characterized prospectively or retrospectively obtained clinical specimens, as appropriate, representative of the intended use population;
(B) For indications for use based on a threshold not established in a predicate device of this generic type, device performance data from a clinical study demonstrating clinical validity using well-characterized prospectively or retrospectively obtained clinical specimens, as appropriate, representative of the intended use population;
(C) Device reproducibility data generated, using a minimum of three sites, of which at least two sites must be external sites, with two operators at each site. Each site must conduct a minimum of three runs per operator over non-consecutive days evaluating a minimum of five different BCR-ABL concentrations that span and are well distributed over the measuring range and include MR3 (0.1 percent IS). Results shall be reported as the standard deviation and percentage coefficient of variation for each level tested. Prespecified acceptance criteria must be provided and followed;
(D) Device precision data using clinical samples to evaluate the within-lot, between-lot, within-run, between run, and total variation;
(E) Device linearity data using a dilution panel created from clinical samples;
(F) Device analytic sensitivity data, including limit of blank, limit of detection, and limit of quantification;
(G) Device specificity data, including interference and cross-contamination; and
(H) Device stability data, including real-time stability of samples under various storage times, temperatures, and freeze-thaw conditions.
(iv) Identification of risk mitigation elements used by your device, including a detailed description of all additional procedures, methods, and practices incorporated into the instructions for use that mitigate risks associated with testing using your device.
(2) Your 21 CFR 809.10 compliant labeling must include the following:
(i) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) complaint labeling must include an indication for use statement that reads “This test is not intended for the diagnosis of CML”; and
(ii) A detailed description of the performance studies conducted to comply with paragraph (b)(1)(iii) of this section and a summary of the results.
(3) Your device output must include results on the International Scale (IS) and your assay must include multipoint calibration controls traceable to a relevant international reference panel (
e.g., the World Health Organization International Genetic Reference Panel for quantitation of BCR-ABL mRNA).