K Number
K190054

Validate with FDA (Live)

Date Cleared
2019-10-21

(283 days)

Product Code
Regulation Number
880.5860
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Becton Dickinson insulin syringes are intended for subcutaneous injection of U-100 insulins

Device Description

The subject BD Insulin Syringes are a plastic syringe designed for subcutaneous injection of a desired dose of U 100 insulin. The BD Insulin Syringe consists of a graduated barrel, plunger rod and needle/hub assembly. The BD Insulin Syringes are sterile, single use, and non-pyrogenic. These devices operate on the principles of a piston syringe. The BD Insulin Syringe with 6mm length cannula have a sub-brand name of BD Veo™ Insulin Syringe.

AI/ML Overview

This document describes the premarket notification (510(k)) for the BD Insulin Syringe (K190054). The information provided focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a study proving a device meets specific acceptance criteria in the context of diagnostic performance or clinical outcomes.

However, based on the non-clinical testing section, we can infer the acceptance criteria for various performance aspects and that the device met these criteria through testing.

Here's an interpretation of the requested information, focusing on the available details from the document:

1. A table of acceptance criteria and the reported device performance

Since this is a 510(k) submission for an insulin syringe, the "acceptance criteria" discussed are largely related to functional performance and biocompatibility, as defined by international standards. The document states that the "Results of testing demonstrated the BD Insulin Syringes met requirements for its intended use" for all listed tests. Specific numerical acceptance criteria are not provided in this summary, but rather the standards against which compliance was demonstrated.

Acceptance Criteria (Standards Met)Reported Device Performance
ISO 8537:2016 Sterile single-use syringes, with or without needle, for insulin (for functional performance tests)Met all requirements for intended use
ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices (for functional performance tests)Met all requirements for intended use
ISO 10993:2019 Biological evaluation of medical devices (for biocompatibility tests)Met all requirements as safe and effective
ASTM F2148: 2013 Standard Practice for Evaluation of Delayed Contact Hypersensitivity Using the Murine Local Lymph Node Assay (LLNA)Met all requirements for intended use
ISO 11137-1:2006/Amd 2013(E) Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devicesMet all requirements for intended use
ISO 11137-2:2013(E) Sterilization of health care products -Radiation - Part 2: Establishing the sterilization doseMet all requirements for intended use
ISO 11137-3: 2017(E) Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects of development, validation and routine controlMet all requirements for intended use
ISO 11737-1:2018(E) Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on productsMet all requirements for intended use

Specific functional performance tests mentioned under ISO 8537 and ISO 9626 (for which the device met requirements):

  • Air Leakage
  • Cannula Pull Force
  • Hub Pull Force
  • Liquid Leakage
  • Needle Break-Off Force
  • Penetration Force
  • Scale Print Permanency
  • Shield Pull Force
  • Syringe Filling Force
  • Syringe Injection Force (Force to Expel Water)
  • Volumetric Accuracy
  • Dead Space
  • Corrosion Resistance

Specific biocompatibility tests mentioned under ISO 10993 (for which the device met requirements):

  • Cytotoxicity
  • Sensitization (GP Max)
  • Sensitization, Intracutaneous Reactivity
  • Irritation
  • Acute Systemic Toxicity
  • Material Mediated Pyrogenicity
  • Subacute/Subchronic Toxicity
  • Hemocompatibility
  • Genotoxicity (Bacterial and Mammalian)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the exact sample sizes used for each non-clinical performance and biocompatibility test. It mentions that "required testing was conducted to validate the cumulative modifications." The data provenance (e.g., country of origin) is not detailed, but it would typically be internal testing conducted by Becton Dickinson. These are prospective tests conducted on manufactured devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable to this type of device submission. The "ground truth" for a medical syringe is its physical and chemical properties and functional performance, which are evaluated against international standards by laboratory testing, not by expert consensus on clinical images or diagnoses.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not a clinical study involving human readers or interpretation. The performance is objectively measured against established criteria in laboratory settings.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/imaging device or a diagnostic device, and no MRMC study was performed or required.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For this device, the "ground truth" is defined by the technical specifications and performance requirements outlined in the referenced international standards (e.g., ISO 8537, ISO 9626, ISO 10993). Compliance with these standards, as measured by objective laboratory tests, serves as the basis for demonstrating safety and effectiveness.

8. The sample size for the training set

Not applicable. This is a physical medical device, not an AI or machine learning model that requires a training set.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 21, 2019 Becton Dickinson and Company Meriam Youssef Senior Manager Regulatory Affairs 1 Becton Drive Franklin Lakes, New Jersey 07417

Re: K190054

Trade/Device Name: BD Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: September 19, 2019 Received: September 23, 2019

Dear Meriam Youssef:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190054

Device Name BD Insulin Syringe

Indications for Use (Describe)Becton Dickinson insulin syringes are intended for subcutaneous injection of U-100 insulins
-----------------------------------------------------------------------------------------------------------------------------
Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K190054 510(k) Summary

Submitted By:Meriam YoussefSr. Manager Regulatory Affairs, BD Medical- Diabetes Care1 Becton DriveFranklin Lakes, NJ 07417Tel: 201 847 6557Fax: 201 847 5307
Date Prepared:September 19, 2019
Device Name:Trade Name:BD Insulin Syringe™
Common Name:Syringe, Piston
Regulation Name:Piston syringe
Regulation:21 CFR 880.5860
Product Code:FMF

Legally marketed predicate device to which substantial equivalence is being claimed:

  • BD Ultra-Fine™ Insulin Syringe, BD Ultra-Fine™ II Insulin Syringe, BD Insulin Syringes with the . BD Ultra-Fine™ needle and BD Insulin Syringes with Micro-Fine™ IV Needle (K170386)

Device Description:

The subject BD Insulin Syringes are a plastic syringe designed for subcutaneous injection of a desired dose of U 100 insulin. The BD Insulin Syringe consists of a graduated barrel, plunger rod and needle/hub assembly. The BD Insulin Syringes are sterile, single use, and non-pyrogenic. These devices operate on the principles of a piston syringe. The BD Insulin Syringe with 6mm length cannula have a sub-brand name of BD Veo™ Insulin Syringe.

Syringe size (mL)BevelCannula length (mm)Gauge (G)
0.3mL312.7mm29G
0.3mL312.7mm30G
0.3mL38mm29G
0.3mL38mm31G
0.3mL36mm31G
0.5mL312.7mm28G
0.5mL312.7mm29G
0.5mL312.7mm30G
0.5mL38mm31G
0.5mL36mm31G
1mL316mm27G
1mL312.7mm28G
1mL312.7mm29G
1mL312.7mm30G
1mL38mm30G
1mL38mm31G
1mL36mm31G
1mL316mm27G

The BD Insulin Syringes are offered in the following sizes:

{4}------------------------------------------------

Indications for Use:

Becton Dickinson insulin syringes are intended for subcutaneous injection of U-100 insulins.

Differences in Indications for use: The indications for use are the same as the predicate.

Substantial Equivalence Discussion:

Predicate DeviceComparison
General InformationFeatureSubject Device:BD Insulin SyringeBD Ultra-Fine™ InsulinSyringe, BD Ultra-Fine™II Insulin Syringe, BDInsulin Syringes with theBD Ultra-Fine™ needleand BD Insulin Syringeswith Micro-Fine™ IVNeedle
510(k) NumberK190054K170386N/A
Device ClassificationSyringe PistonSameSame
Regulation Number880.586SameSame
Product CodeFMFSameSame
Indications of UseBecton Dickinson insulinsyringes are intended forsubcutaneous injection ofU 100 insulinsBecton Dickinson insulinsyringes are intended forsubcutaneous injection ofU-100 insulins.Same
Specific Drug UseU-100 InsulinSameSame
Capacity1mL,0.5mL,0.3mL1mLThe subject deviceincludes a 0.5mL and0.3ml model.
Bevel33Dimensionalmodifications toprimary bevel ofcannula
Cannula Gauge Size(s)27G, 28G, 29G, 30G, 31G27G, 28G, 29G, 30G, 31GSame
Nozzle Tip TypeSnap fit needle hub for0.3mL and 0.5mL designsfor permanentlyattached needlesIntegral barrel for 1mLdesign forpermanently attachedIntegral barrel for 1mLdesign forpermanently attachedneedlesIntroduction of one gluewell design for 1mLbarrel configuration
Scale Markingsneedles1 unit increments and ½unit increments (0.3mL)1 unit increments (0.5mL)2 units increments (1mL)2 units increments(1mL)Addition of ½ unit scalemark increments
General InformationFeatureSubject Device:Predicate DeviceComparison
General InformationFeatureSubject Device:BD Insulin SyringeBD Ultra-Fine™ InsulinSyringe, BD Ultra-Fine™II Insulin Syringe, BDInsulin Syringes with theBD Ultra-Fine™ needleand BD Insulin Syringeswith Micro-Fine™ IVNeedle
Packaging ConfigurationPolybag for self-containedsyringe and blister pack(multiple blister packs perperforated sheet) offeringsPolybag for self-containedsyringe and blister pack(individual blister pack)offeringsAddition of retentionring to 1mLself-contained insulinsyringeMultiple blister packsper perforated sheet
Cannula Length Size(s)6mm, 8mm, 12.7mm,16mm6mm, 8mm, 12.7mm,16mmSame
Cannula (Needle) material304 Stainless Steel304 Stainless SteelSame
Cannula (Needle) shieldmaterial and Cap (forplunger rod) materialPolyethylenePolyethyleneModification to the co-monomer ofpolyethylene material
Cannula cover colorOrangeOrangeSame
LubricantMedical Grade SiliconeMedical Grade SiliconeModification tolubricant formulationcarrier solvent
Cannula bondingadhesive (UV cured)UV Cured AdhesiveUV Cured AdhesiveNew 0.3mL & 0.5mLUV Cured Adhesive
Scale marking inkSolvent based Markemblack colored inkSolvent based Markemblack colored inkSame
Plunger rod materialPolystyrenePolystyreneRemove colorantadditive from theplunger rod component
Plunger tip (stopper)materialRubber stopperRubber stopperSame
Barrel and Hub MaterialPolypropylenePolypropyleneNew materialformulation ofpolypropylene
Single Use OnlyYESYESSame
Non-PyrogenicYESYESSame
Sterile (10-6)YESYESSame

{5}------------------------------------------------

{6}------------------------------------------------

Discussion:

Apart from the above mentioned modifications to the design and materials, there are no technological differences between the subject and the predicate devices. The design modifications were assessed through performance testing per ISO 8537 and ISO 9626. The material changes were evaluated per ISO 10993-1.

The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness.

Non-Clinical Testing:

Following BD's Quality System processes, required testing was conducted to validate the cumulative modifications made to the subject devices.

Substantial Equivalence is being supported with full performance testing the current BD Insulin Syringe devices currently marketed.

Bench functional performance

  • Testing was conducted according to ISO 8537:2016 Sterile single-use syringes, with or without needle, ● for insulin, and ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices. This testing included Air Leakage, Cannula Pull Force, Hub Pull Force, Hub Pull Force, Liquid Leakage, Needle Break-Off Force, Penetration Force, Scale Print Permanency, Shield Pull Force, Syringe Filling Force, Syringe Injection Force (Force to Expel Water), Volumetric Accuracy, Dead Space and Corrosion Resistance.

Biocompatibility testing

  • Testing was conducted according to relevant sections of ISO 10993:2019 Biological evaluation of ● medical devices. This testing included Cytotoxicity, Sensitization (GP Max), Sensitization, Intracutaneous Reactivity, Irritation, Acute Systemic Toxicity, Material Mediated Pyrogenicity, Subacute/Subchronic Toxicity, Hemocompatibility, and Genotoxicity (Bacterial and Mammalian). The medical device is classified as: externally communicating, blood path indirect, with prolonged contact duration.

Additional Testing:

  • ASTM F2148: 2013 Standard Practice for Evaluation of Delayed Contact Hypersensitivity Using the Murine Local Lymph Node Assay (LLNA), ISO 11137-1:2006/Amd 2013(E) Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. ISO 11137-2:2013(E) Sterilization of health care products -Radiation - Part 2: Establishing the sterilization dose, ISO 11137-3: 2017(E) Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects of development, validation and routine control, and ISO 11737-1:2018(E) Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on products .
    Results of testing demonstrated the BD Insulin Syringes met requirements for its intended use and is as safe and effective as its predicate device.

Clinical Testing:

Not Applicable

Conclusion:

The modifications to the design, dimensions, and materials of the subject device met the standards. The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness.

The BD Insulin Syringes are substantially equivalent to the BD Ultra-Fine™ Insulin Syringe, BD Ultra-Fine™ II Insulin Syringe, BD Insulin Syringes with the BD Ultra-Fine™ needle and BD Insulin Syringes with Micro-Fine™ IV Needle cleared under K170386 with respect to the indications for use, target populations, treatment method, use environment and technological characteristics.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).