K Number
K183513

Validate with FDA (Live)

Date Cleared
2019-06-13

(177 days)

Product Code
Regulation Number
882.5910
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DuraSure Biologic Patch is intended for use as a surgical patch material to close dura mater during neurosurgery.

Device Description

The DuraSure consists of one piece of bovine pericardial tissue that has been selected for minimal tissue blemishes. The tissue is treated with a glutaraldehyde process which crosslinks the collagen fibers and minimizes antigenicity. The DuraSure is liquid chemical sterilized and packaged in a plastic jar containing sterile glutaraldehyde storage solution. The DuraSure is designed to repair the body's natural organs.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the DuraSure Biologic Patch, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaDuraSure Biologic Patch PerformancePredicate Device Performance
Tensile Strength≥ 2 MPaMean: 11.9 MPaNot explicitly stated (implied to meet acceptance criteria, but specific value for predicate device not given, only "mean of tensile strength of the predicate device was [value not provided]" in text)
Elongation5% - 50% elongationMean: 25% elongationMean: 32% elongation
Burst Strength≥ 12 PSIMean: 127 PSIMean: 59 PSI
Suture Retention≥ 300 gfMean: 970 gfMean: 978 gf
Thickness0.35 mm - 0.75 mmPassed acceptance criteria0.32 mm - 0.71 mm
BiocompatibilitySatisfactory biocompatibility (implied, by comparison to predicate)Satisfactory biocompatibility resultsEstablished biocompatibility
SterilizationChemically sterilized according to ISO14160: 2011 with 10⁻⁶ SALChemically sterilized according to ISO14160: 2011 with 10⁻⁶ SALChemically sterilized with 10⁻⁶ SAL
Animal Study ConclusionTest article is locally non-toxic and performs equivalently to control in dural repair model.Test article is locally non-toxic; performed equivalently to control.Control article deemed locally non-toxic and performed equivalently to test article.

2. Sample Size Used for the Test Set and Data Provenance

The document focuses on "pre-clinical" testing, which includes both in-vitro (bench) testing and an in-vivo animal study.

  • Bench Testing (Tensile, Elongation, Burst Strength, Suture Retention, Thickness):

    • Sample Size: Not explicitly stated for each test (e.g., "All DuraSure patches passed the acceptance criteria"). However, results are given as "mean" values, implying multiple samples were tested for each characteristic.
    • Data Provenance: Not specified, but generally refers to laboratory testing conducted by the manufacturer or contracted labs. The country of origin is not mentioned. This is retrospective data collected for the 510(k) submission.
  • In-vivo Animal Study:

    • Sample Size: Forty-nine (49) rabbits.
    • Data Provenance: Not explicitly stated, but implies a controlled laboratory setting for animal research. This is prospective data collected during the study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Bench Testing: No experts are typically used to establish ground truth for this type of quantitative functional testing. The "ground truth" is defined by the physical measurement methods and the device specifications.
  • In-vivo Animal Study:
    • Number of Experts: At least one. The document states, "All slides were evaluated by a veterinary pathologist for neuropathological changes in brain tissue, dural integrity, neoduralization and local tissue reaction according to ISO 10993-6 and FDA Guidance."
    • Qualifications: "veterinary pathologist." No further details on years of experience or specific sub-specialties are provided.

4. Adjudication Method for the Test Set

  • Bench Testing: Not applicable. These are objective measurements following established test methods.
  • In-vivo Animal Study: Not explicitly described. It states "All slides were evaluated by a veterinary pathologist." This implies a single expert assessment. If multiple pathologists were involved, no method for resolving disagreements (e.g., 2+1, 3+1) is mentioned.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The studies presented are pre-clinical (bench and animal studies) and do not involve human readers evaluating cases.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

No, this question is not applicable. The DuraSure Biologic Patch is a physical medical device (surgical patch material), not an AI algorithm or software device. Therefore, a standalone performance evaluation of an algorithm is not relevant.

7. The Type of Ground Truth Used

  • Bench Testing: The ground truth is based on objective quantitative measurements obtained using standardized test methods (e.g., Instron for tensile/elongation/suture retention, pressure sensors for burst strength, thickness gauges).
  • In-vivo Animal Study: The ground truth for the animal study was established through histopathological evaluation by a veterinary pathologist, based on internationally recognized standards (ISO 10993-6) and FDA guidance for dural substitute devices. This includes macroscopic and microscopic evaluations, with scoring.

8. The Sample Size for the Training Set

This question is not applicable. The DuraSure Biologic Patch is a physical medical device, not a machine learning model. Therefore, there is no "training set" in the context of AI.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable, as there is no training set for a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 13, 2019

LeMaitre Vascular Inc. Xiang Zhang Vice President of Regulatory Affairs 63 Second Ave Burlington, Massachusetts 01803

Re: K183513

Trade/Device Name: DuraSure Biologic Patch Regulation Number: 21 CFR 882.5910 Regulation Name: Dura Substitute Regulatory Class: Class II Product Code: GXQ Dated: May 8, 2019 Received: May 14, 2019

Dear Xiang Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Matthew Krueger, M.S.E Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183513

Device Name DuraSure Biologic Patch

Indications for Use (Describe)

The DuraSure Biologic Patch is intended for use as a surgical patch material to close dura mater during neurosurgery.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510k Summary K183513

Submitter's Information
Name:LeMaitre Vascular, Inc.
Address:63 Second Avenue, Burlington, MA 01803
Phone:781-425-1706
Fax:781-425-5049
Contact Person:Xiang (Vic) ZhangVP of Regulatory AffairsLeMaitre Vascular, Inc.Email: xzhang@lemaitre.com
Date Prepared:May 8, 2019
Device Name:DuraSure Biologic Patch
Trade Name:DuraSure Biologic Patch
Common Name:Dura Substitute
Regulation Number:21CFR §882.5910
Classification Panel:Neurology
Class:II (2)
Product Code:GXQ
EstablishmentRegistration:1220948
Establishment:63 Second AvenueBurlington, MA 01803
Predicate Device:Dura-Guard Dura Repair Patch K973706
Reference Device:XenoSure Biologic Patch K040835

{4}------------------------------------------------

Device Description:The DuraSure consists of one piece of bovine pericardial tissue that has been selected for minimal tissue blemishes. The tissue is treated with a glutaraldehyde process which crosslinks the collagen fibers and minimizes antigenicity. The DuraSure is liquid chemical sterilized and packaged in a plastic jar containing sterile glutaraldehyde storage solution. The DuraSure is designed to repair the body's natural organs.
Indication for Use:The DuraSure Biologic Patch is intended for use as a surgical patch material to close dura mater during neurosurgery.
Summary of TechnologicalCharacteristics:Comparisons of the DuraSure Biologic Patch with the predicate Dura-Guard Patch show that technological characteristics such as materials, biocompatibility, performance, and sterilization of the proposed device are substantially equivalent to the predicate device. The difference between the proposed device and the predicate device is: The DuraSure biologic patch is stored in glutaraldehyde solution and the Dura-Guard is stored in propylene oxide solution. This difference does not raise additional risk concern because DuraSure biologic patch has satisfactory biocompatibility results and in-vivo animal study results.
Proposed DevicePredicate DeviceComparison
Product Name:DuraSure Biologic PatchProduct Name:Dura-Guard Patch
ManufacturerLeMaitre Vascular Inc.Synovis/Baxter
ClearanceThis submissionK973706
Indications forUseThe XenoSure Dura BiologicPatch is intended for use as asurgical patch material toclose dura mater duringneurosurgery.For use as a durasubstitute for theclosure of dura materduring neurosurgery.Same
Common NameDura SubstituteDura SubstituteSame
ClassificationII (2)II (2)Same
RegulationNumber882.5910882.5910Same
Product CodeGXQGXQSame
MaterialsBovine pericardiumBovine pericardiumSame
SterilityChemical sterilization with 106 SALChemical sterilization with 10-6 SALSame
Single UseYesYesSame
MedicalSpecialistsNeurologyNeurologySame
PackagingJar/LidJar/LidSame
Storage solution0.2% glutaraldehyde1% propylene oxideDifferent
Functional/Safety Testing:The verification activities conducted indicate that DuraSure biologic patch meets the product performance requirements of the device specifications and does not raise any additional safety issues.
Sterilization:DuraSure biologic patch is chemically sterilized according to ISO14160: 2011, "Sterilization of health care products -- Liquid chemical sterilizing agents for single-use medical devices utilizing animal tissues and their derivatives -- Requirements for characterization, development, validation and routine control of a sterilization process for medical devices".
Biocompatibility:The material for DuraSure Biologic Patch is bovine pericardium. It is identical to that in the predicate device and the reference device which have established biocompatibility. This submission is to expand the indication with no change to product design, materials, or manufacturing processes.
Summary of Product Testing:The following tests have been completed to evaluate the safety and performance of the DuraSure Biologic Patch when compared with the predicate device:• Tensile test• Burst strength test• Suture retention test• Thickness measurement
TestTest method summaryResults
TensileUse Instron pull the sample until it fails. Record the ultimate tensile strength in MPa.All DuraSure patches passed the acceptance criteria of ≥ 2 MPa. The mean of tensile strength of the predicate device was
of tensile strength of DuraSurepatch was measured as 11.9MPa.
ElongationUse Instron pull the sample until itfails. Record the elongation at thefailure as percent of the original samplelength.All DuraSure patches passed theacceptance criteria of 5~50%elongation. The mean ofelongation of the predicatedevice was measured as 32%.The mean of elongation ofDuraSure patch was measured as25%.
Burst strengthSample is secured in the testing fixtureas a membrane between two chambers.One side of the sample is slowlypressurized using water. Record thepressure at the time of burst or leak.All DuraSure patches passed theacceptance criteria of ≥ 12 PSI.The mean of burst strength ofthe predicate device wasmeasured as 59 PSI. The meanof burst strength of DuraSurepatch was measured as 127 PSI.
Suture retentionMake sutures on the edge of the patch.Pull the suture using Instron until eitherpatch or suture fail. Record the force atthe failure.All DuraSure patches passed theacceptance criteria of ≥ 300 gf.The mean of suture retention ofthe predicate device wasmeasured as 978 gf. The meanof suture retention of DuraSurepatch was measured as 970 gf.
ThicknessMeasure the thickness of the patchusing a thickness gaugeAll DuraSure patches passedacceptance criteria of 0.350.75mm. The thickness of thepredicate device was measuredbetween 0.320.71 mm.

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

Summary of Pre-clinical Study:The purpose of this animal study was to assess the local tissue response to implantation following 1, 8, and 26-week brain dural repair/neural implantation in rabbits compared to a predicate control to satisfy ISO 10993-6, Annex D, Tests for local effects after implantation and US-FDA Guidance Document for Dura Substitutes Devices, Section XI, 2000.Forty nine (49) rabbits received single dural defects of approximately 0.6 cm x 0.6 cm. Test or control article were sutured in place to cover each dural defect. The bone flap was replaced, the dermal incisions were surgically closed, and animals were recovered and returned to standard housing. Animals were observed closely and weighed weekly to as one measure of general clinical health. After 1, 8, or 26 weeks post-surgery, animals were humanely euthanized and draining lymph nodes and the whole skull with brain was excised and placed in formalin. The formalin-fixed lymph nodes and implant sites were processed by histopathology techniques, using decalcification (as needed) and paraffin embedding. Resulting lymph node slides were stained with H&E. The dura/brain sections were treated with H&E, Fluoro-Jade, Luxol Fast Blue, anti-GFAP (glial fibrillary acidic protein) and anti-Iba-1 immunostaining on slides. All slides were evaluated by a veterinary pathologist for neuropathological changes in brain tissue, dural integrity, neoduralization and local tissue reaction according to ISO 10993-6 and FDA Guidance.The overall conclusion of the control and test article brain dural repair local effect evaluation based on in vivo and terminal evaluations, including an extensive histopathological evaluation of the implantation site, adjoining areas and draining lymph nodes, is that the implantation and dural repair surgery was successful and the test article is locally non-toxic. This applies to evaluations of animals exposed for 1, 8 or 26-weeks. General
Conclusion:clinical evaluations/cageside observations, neurophysiological evaluations and body weights provide no evidence supporting an adverse effect of the surgery, control material or test article. These conclusions are supported by macroscopic evaluations and microscopic evaluations with scoring. Resorption of the test article or control article was minimal at the last exposure time evaluated. Adhesions at or near the repair site were noted, especially at one week exposure, and evidence of hydrocephaly was also noted, especially at 1 and 8 week exposure, however neither were associated with any pathology or adverse effects. From this investigation, it appears that the test and control articles performed equivalently in the in-vivo rabbit model of dural repair. LeMaitre Vascular has demonstrated that DuraSure Biologic Patch is substantially equivalent to the predicate devices based on its intended use and fundamental scientific technology.

{8}------------------------------------------------

§ 882.5910 Dura substitute.

(a)
Identification. A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).(b)
Classification. Class II (performance standards).