K Number
K040835

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-06-15

(76 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PeriPatch Sheet is intended for use as a surgical patch material for: cardiac and vascular reconstruction and repair, soft tissue deficiency repair and reinforcing the suture line during general surgical procedures.

Device Description

The PeriPatch™ Sheet is a quadrilateral shaped xenograft patch made from a sheet of glutaraldehyde fixed bovine pericardium selected for even thickness. It is designed to repair the body's natural organs and functions like natural tissue. It is intended for intra-cardiac repair procedures. It is identical to other marketed bovine pericardial patches. Extensive quality control procedures assure a consistent, high quality product for clinical use.

AI/ML Overview

This document describes a 510(k) premarket notification for the "PeriPatch™ Sheet," a processed bovine pericardial patch. The submission claims substantial equivalence to previously cleared devices.

Here's the breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text does not explicitly state "acceptance criteria" in a formalized, quantifiable manner for biological performance. Instead, it demonstrates substantial equivalence to predicate devices through comparisons of physical, chemical, and manufacturing characteristics. The reported device performance is presented as direct comparisons to these predicates.

CharacteristicPM Devices Inc. PeriPatch (Reported Performance)Predicate Device (e.g., BioVascular, Extracorporeal)Significance (as stated in document)
MaterialBovine PericardiumBovine PericardiumSE (Substantially Equivalent)
Cross-linkingGlutaraldehyde cross-linkingGlutaraldehyde cross-linkingSE
ShapeFlat, square & RectangularFlat, square & RectangularSE
Sizes (cm)1.5x8, 1.5x16, 4x4, 4x6, 6x8, 10x16Range of sizes, with overlaps with PeriPatchSE
PackagingSealed, sterile containerSealed, sterile containerSE
Thickness (mm)0.58 ± 0.170.25 (BioVascular), 0.5 ± 0.25 (BioVascular), 0.35 (Extracorporeal)SE
Weight (g/m²)1172 ± 311 g/m²1080 ± 330 g/m² (BioVascular), 4360 ± 1600 g (Extracorporeal)SE
Melting Point (°C)86.2°C83.6°C (Extracorporeal)SE
Tensile Elongation (g)969 ± 1141121 ± 102 (BioVascular), 1280 ± 108 (Extracorporeal)SE
Tensile Strength (g)46.1 ± 5.4Unknown (BioVascular), 31.2 ± 6.0 (Extracorporeal)SE
Burst Strength (mmHg)8200 / 1597033 / 136 (BioVascular)SE
Sterilization Agent0.2% Glutaraldehyde / PBSPropylene oxide in water (BioVascular, Glycar)SE
Rinsing ConditionsTwo 2 min. rinsesOne 3 min. Rinse (BioVascular), Three 2 min. rinse (Extracorporeal)SE
Sterility MethodLiquid AlcoholicLiquid Alcoholic (BioVascular, Extracorporeal)SE

2. Sample Size Used for the Test Set and Data Provenance

The document does not describe a "test set" in the context of a clinical study or a specific, independent dataset used to evaluate the device's performance against defined acceptance criteria. Instead, it relies on comparisons of its physical and performance characteristics to predicate devices. The data provenance is derived from the manufacturing and characterization of the PeriPatch™ Sheet itself and "literature" and "validations" cited in Appendix A and D respectively, which are not provided in this excerpt. This is a retrospective comparison to existing, legally marketed devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. No independent "test set" with a ground truth established by experts is described. The document focuses on demonstrating substantial equivalence to predicate devices through technical comparisons rather than an independent clinical validation.

4. Adjudication Method for the Test Set

Not applicable. No "test set" or adjudication method is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Reader Improvement

No. A MRMC comparative effectiveness study was not done. This is a premarket notification for a medical device (a xenograft patch), not an AI-assisted diagnostic tool.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No. This question applies to AI/ML software, not to a physical medical device like the PeriPatch™ Sheet.

7. The Type of Ground Truth Used

The "ground truth" in this context is the established performance and safety profiles of the predicate devices, as detailed in their 510(k) clearances and scientific literature. The PeriPatch™ Sheet aims to demonstrate that its characteristics are similar enough to these established devices to be considered substantially equivalent. It is based on engineering and material property comparisons rather than clinical outcomes or pathology data from a new study.

8. The Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI/ML system. There is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable. There is no "training set" for this type of device.

{0}------------------------------------------------

K040835 510(K) SUMMARY SECTION 4

  1. Submitters Name, Address etc.:

JUN 1 5 2004

PM Devices Inc. 2135 – 13700 Mayfield Place Richmond, British Columbia V6V 2E4, CANADA Ph: 604.270-4344 Fx: 604.270-4384 www.pmdevices.com Contact: Britta Dombovari Date: March 2004

PeriPatch™ Sheet 2. Name of Devices: Trade Name:

Common Name: Processed bovine pericardial patch

Classification Name: Currently: Intracardiac Patch or Pledget -Class II-Product Code: DXZ Expanded Indications: Mesh, surgical, polymeric –Class Il-Product Code: FTM

    1. Predicate Devices: Legally marketed devices which PM Devices Inc. claims substantial equivalence:
Predicate DeviceManufacture510(k) #Class
PeriPatch™ SheetPM Devices Inc.K031948II
Supple Peri-Guard® PericardiumBioVascular Inc.K983162II
Glycar Pericardial PatchGlycar Inc.K963967II
Vascu-GuardBioVascular Inc.K942010II
Supple Peri-GuardBioVascular IncK921895II
Hancock Pericardial PatchExtrcorpealK830863II

All of the above previously cleared products are composed of processed bovine pericardium and are all used as a mesh material for surgical repair of pericardial structures and soft tissue deficiencies.

4. Device Description:

4a. How the Device Works

PeriPatch™ Sheet

The PeriPatch™ Sheet is a quadrilateral shaped xenograft patch made from a sheet of glutaraldehyde fixed bovine pericardium selected for even thickness. It is designed to repair the body's natural organs and functions like natural tissue. It is intended for intra-cardiac repair procedures. A picture of a patch can be seen in Figure 1 - PeriPatch Sheet, and the engineering drawing can be found in Appendix C, Engineering Drawings. It is identical to other marketed bovine Extensive quality control procedures pericardial patches. assure a consistent, high quality product for clinical use.

Image /page/0/Picture/15 description: In the image, a translucent rectangular object is being held by a pair of tweezers. The object appears to be thin and flexible, possibly made of plastic or a similar material. The tweezers are holding the object near one of its ends. The background is a dark, textured surface, which contrasts with the lighter color of the object and the tweezers.

Figure 1 - PeriPatch Sheet

The PeriPatch Sheet is available in 6 sizes (Table 4.1 – page 4-2), but can also be trimmed to specific size depending on the procedure.

4

{1}------------------------------------------------

$$\sim / \Psi$$

Table 4.1 PeriPatch Sheet Model Numbers & Dimensions

Comments of Concession Commend Commend Comments of Concession Come of Concession Come of Concession Come of Concession Come of Concession of Concession of Concession of Conce1.5P81A . A . AA------------------------------------------------------------------------------------------------------------------------------------------------------------------------------10P16
Model1.5P164P44P6e68------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
September 1998 - 1998 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 199Size (cm)1.5 × 81.5 x 16------------------------------------------------------------------------------------------------and and and and in the commend of the commend of the commend of the comment of4×4------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Status and Address & Address4 × 66×810 x 16.

4b. Scientific Concepts

PeriPatch Sheet

The PeriPatch Sheet is manufactured from glutaraldehyde fixed bovine pericardium, this is the The I Chi aten Oneono mananate devices. Physical configuration of the PeriPatch sheet is similar to predicate patches, see Table 4.2 below:

Table 4.2: Comparison of PeriPatch Sheet sizes available to current predicate devices.

PeriPatch SheetModels (DXZ/FTM) *1.5P81.5P164P44P66P810P16
PeriPatch SheetSizes (DXZ/FTM)1.5cmx8cm1.5cmx16cm4cmx4cm4cmx6cm6cmx8cm10cmx16cm
Bio-Vascular SupplePeri-guard Patch**N/AN/A4cmx4cmN/A6cmx8cm10cmx16cm
Bio-Vascular Vascu-Guard***1.5cmx8cmN/AN/AN/AN/AN/A
  • PeriPatch Sheet (DXZ) – K031948

** Bio-Vascular Supple Peri-guard Patch - K981895

*** Bio-Vascular Vascu-Guard – K942010

The treatment and processing (relevant for the PeriPatch Sheet) for cross-linking bovine pericardial tissue with glutaraldehyde is well described in the literature (Appendix A, Literature), and similar to those used in predicate devices, and has been validated (Appendix D, Validations). Sterilization is performed using a liquid alcoholic sterilant which is similar to the predicates and validated to be effective (Appendix D, Validations). The finished devices are packaged and labeled in a similar manner as the predicates (Section 6, Proposed Labeling; Appendix E, Predicate Device Labeling).

The PeriPatch Sheet is considered to be similar to the PeriPatch Sheet and Bio-Vascular predicates because:

  • Same raw material Bovine Pericardium +
  • Same intended medical use .
  • Operates using the same fundamental scientific technology
  • Similar shapes & sizes .
  • Similar method of processing .
  • Similar method of sterilization .
  • Similar packaging and labelling

4c. Physical & Performance Characteristics

The PeriPatch Sheet is designed to repair the body's natural organs and function like natural tissue.

4d. Safety & Effectiveness

The devices are designed and manufactured in such a way that, when used under the conditions and the purposes intended, they will not compromise the clinical condition or the

{2}------------------------------------------------

safety of patients, or the safety and health of users or, where applicable, other persons, sancity of patients, or the carety as associated with their use constitute acceptable risks provided that any hiske we the patient and are compatible with a high level of protection of health and safety.

The safety and effectiveness of bovine pericardial patches used for reconstruction and repair r rie salely and encoments of sture (Appendix A, Literature). They have been in use for are well one of and have proven to be effective in achieving the desired result and well tolerated by host tissue.

5. Intended use of the Device

Below is a list of the diseases or conditions that the device will treat, prevent, cure or mitigate and a description:

The PeriPatch Sheet has been approved for the following intended use:

  • Surgical patch material for cardiac and vascular reconstruction and repair.
    The intended expanded indications for use would include:

  • ... soft tissue deficiency repair and reinforcing the suture line during general surgical procedures.

6. Technological Characteristics

The devices have the same technological characteristics as the predicate devices identified in Section 4-3. A comparison of the PeriPatch Sheet to the predicate devices can be found in Table 4.3 on the next page. As shown in the table, the applicant device is substantially equivalent to the predicates technological characteristics.

{3}------------------------------------------------

PM DEVICES INC. – PERIPATCH™ SHEET 510(K) PREMARKET NOTIFICATION

able 4.3 - Similarities of the PeriPatch Sheet to the predicate devices

NumberUsePM Devices Inc.PeriPatch (DXZ & FTM)PM Devices Inc.PeriPatch (DXZ) K031948BioVascularK983162Glycar Inc.K963967BioVascularK942010BioVascularK921895ExtracorporealK830863Significance
The PeriPatch Sheet isintended for use as a surgicalpatch material for: cardiacand vascular reconstructionand repair, soft tissuedeficiency repair andreinforcing the suture lineduring general surgicalprocedures.The PeriPatch Sheet is intendedfor use as a surgical patchmaterial for: cardiac andvascular reconstruction andrepair.Repair of pericardialstructures, soft tissuedeficiencies, defectsof the abdominal andthoracic wall, gastricbanding, muscle flapreinforcement, rectaland vaginal pro-lapse, reconstructionof the pelvic floor, andhernias.Cardiac re-construction &repair, vascularpatching,pericardial closurePeripheralvascular repair/re-constructionFor closure of thepatient's pericardium& Peri-Strip sleeveconfiguration - reinforcestaple lines duringsegmental resectionsPericardial closurePeriPatchindications aresame as(combined)K031948 andK983162.
Bovine PericardiumSameSameSameSameSameSameSameSE
Glutaraldehyde cross-linkingSameSameSameSameSameSameSameSE
Flat, square & RectangularSameSameSameNot determined attime of submissionSameRectangular & Peri-GuardSleeve, thin strips ofbovine pericardiumsutured to a polyethylenebacking.SameSE
Size (cm)1.5x81.5x164x44x66x810x16Same4x46x88x1410x16Not determined attime of submission1.5x81.5x91.5x102x82x92x102.5x82.5x92.5x104x46x88x1410x16Peri-Guard Sleeves areavailable in sizes to fitcommon staplers5x12SE
Sealed, sterile containerSameSameSameSameSameSameSameSE
Thickness (mm)0.58±0.17Same0.25Unknown0.5±0.250.250.35SE
Weight (g/m)1172±311 g/m2SameUnknownUnknown1080±330 g/m2Unknown4360±1600 gSE
Melting Point (°C)86.2°CSameUnknownUnknownUnknownUnknown83.6°CSE
Tensile Elongation (g)969±114SameUnknownUnknown1121±102Unknown1280±108SE
Tensile Strength (g)46.1±5.4SameUnknownUnknownUnknownUnknown31.2±6.0SE
Burst Strength (mmHg)8200 / 159SameUnknownUnknown7033 / 136UnknownUnknownSE
Sterilization0.2% Glutaralde-hyde /phosphate buffered solution(PBS)SamePropylene oxide inwaterPropylene oxidein waterPropylene oxidein waterPropylene oxide in waterSameSE
Rinsing ConditionsTwo 2 min. rinsesSameOne 3 min. RinseUnknownOne 3 min. rinseOne 3 min. RinseThree 2 min. rinseSE
Sterility MethodLiquid AlcoholicSameSameUnknownSameSameSameSE

|<040835

c/4

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three stylized waves or lines, which may represent the flow of information or resources. The image is black and white.

Public Health Service

JUN 1 5 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Britta Dombovari Regulatory Affairs/Quality Assurance PM Devices, Inc. 2135-13700 Mayfield Place Richmond, British Columbia V6V 2E4 Canada

Re: K040835

Trade/Device Name: PeriPatch™ Sheet Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTM Dated: March 29, 2004 Received: March 31, 2004

Dear Ms. Dombovari:

We have reviewed your Section 510(k) premarket notification of intent to market the device i we nave teviewed your bection of the device is substantially equivalent (for the indications felerenced above and nave determined and and and and one inces marketed in interstate for use stated in the encrosule) to regars actual date of the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Medical Device Americ commerce prior to May 20, 1770, the enastance with the provisions of the Federal Food, Drug, devices that have been recuire approval of a premarket approval application (PMA). and Cosmelle Act (Act) market the device, subject to the general controls provisions of the Act. The r ou may, merciole, market the act include requirements for annual registration, listing of general controls provisions of use ice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (sociation major regulations affecting your device can
may be subject to such additional controls. Existing major regulations of the contract hilly be subject to suell additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Obas neements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised hiat i Dri s issuated on a business with other requirements of the Act that I DA has made a colorimiation administered by other Federal agencies. You must of any Federal statutes and regulations damading, but not limited to: registration and listing (21 comply with an the Ace STequirements, were 801); good manufacturing practice requirements as set and CFR Part 807), adocing (21 OFF Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Ms. Britta Dombovari

This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin manieting your antial equivalence of your device to a legally premarket notification. THC FDA mixing of substantial equate and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice ion your ac not and subser note the regulation entitled, contact the Office of Complanes as (Set notification" (21CFR Part 807.97). You may obtain Misorallung by release to promance no promance notice ander the Act from the Division of Small other general information on your response at its toll-free number (800) 638-2041 or Manufacturers, International and Soless http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Muriam C. Provost

(Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K040835

Device Name: PeriPatchTM Sheet - Processed Bovine Pericardial Pati

Indications For Use:

The Por Use.
The Por Parch sheet is intended for intended for works of a construction The Perforth 5neet 15 Orders and vasular re-constitution.
Patch material for ; cardiac and vasular and soft fissne deficiency repair and repair surgical 1 general the mitche line during re inforcing procedures .

Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please Do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Muriam C. Provost

(Division Sign-Off Division of General, Restorative. and Neurological Devices

510(k) Number_K 040835

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.