K Number
K183187

Validate with FDA (Live)

Date Cleared
2019-03-15

(116 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The nextaro® Transfer System and nextaro® va is indicated for the transfer and mixing of drugs contained in vials.

Device Description

The nextaro® Transfer System and nextaro® va is practically one system with two variants which are to be used independently.

The nextaro® Transfer System consists of two components already assembled (screwed together) in the delivery condition. Each of the components has a plastic spike which is used to perforate the seals of a solvent vial, or lyophilizated pharmaceutical vial, respectively. The components have a male and female Luer adapter to create an air-tight inner media-carrying system with open ends for fluid transfer. Both components are equipped with a filter element to prevent particles (fragments of vial seals, undissolved pharmaceutical) from being administered into the patient's circulatory system.

To be used as intended, the spike of the blue ("upper") part of the nextaro® Transfer System is used to perforate the seal of the solvent vial. The assembly is flipped vertically and the spike of the white ("lower") part of the device is used to perforate the seal of the pharmaceutical vial. The solvent transfer process starts immediately due to the vacuum in the pharmaceutical vial.

After solvent transfer and reconstitution of the pharmaceutical by gentle swirling, the blue component of the device is removed from the white part by unscrewing, a standard syringe with a male Luer is attached to the exposed female Luer of the device to aspirate the ready-to-use solution.

The nextaro® va is a standalone device. Basically, it is the lower part of the nextaro® Transfer System, but without the thread which the latter has to connect the upper part. After connecting the nextaro® va to a pharmaceutical vial by pushing it downwards, a solvent can be transferred by a syringe via the female Luer. After solvent transfer and reconstitution of the pharmaceutical by gentle swirling, a standard syringe with a male Luer is attached to the female Luer of the device to aspirate the ready-to-use solution.

AI/ML Overview

The provided document is a 510(k) summary for the nextaro® Transfer System and nextaro® va, which are drug transfer devices. It does not describe a study involving human readers or AI. Therefore, I cannot provide details on sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, or MRMC studies.

Here's the information regarding acceptance criteria and device performance based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally established by meeting various international and ASTM standards. The reported device performance is indicated by "Pass" or "Equivalent" to the predicate device, or by specific measured values where stated.

Standard / Test PerformedAcceptance Criteria (Implied by Standard)Reported Device Performance
ISO 22413
FragmentationMeets standard requirements for limiting fragmentation."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended." (Implies meeting the standard's criteria)
Penetration ForceMeets standard requirements for penetration force."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
PiercingMeets standard requirements for piercing."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
Verification of Design Specifications for Transfer Devices with HousingMeets specified design requirements."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
ISO 594-2
Male Conical Fitting / Luer ConnectorMeets standard requirements for Luer fittings (e.g., leakage, security)."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
ISO 8536-4
Flow Rate of Infusion FluidMeets standard requirements for flow rate."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
Fluid Filter (Retention Rate)Retains particles as specified by the standard."A retention rate of > 95 % of particles sized ≥ 15 um has been confirmed by testing." "The performance of the filters (retention) of the subject and predicate devices has been shown to be substantially equivalent." (Note: The filter mesh traps particles ≥ 11 µm, but retention rate is reported for ≥ 15 µm).
Leakage (Aging/ Submersion Test)No leakage after aging and submersion."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
Metal IonsBelow specified limits."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
Particulate ContaminationBelow specified limits."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
Penetration Force (also under ISO 22413)Meets standard requirements."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
Reduction of Oxidizable MatterWithin specified limits."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
Residue on EvaporationWithin specified limits."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
Tensile StrengthMeets standard requirements."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
Titration Acidity or AlkalinityWithin specified limits."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
UV Absorption of Extract SolutionWithin specified limits."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
ISO 11607
Sterile Barrier and Packaging SystemsMeets standard requirements for maintaining sterility of the product until point of use."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
ASTM D3078
Bubble EmissionsNo unacceptable bubble emissions, indicating seal integrity."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
ASTM F 1886
Integrity of SealsSeals maintain integrity."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
ASTM F 1929-12
Seal LeakageNo unacceptable seal leakage."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
ASTM F 1980-07
Accelerated AgingPerformance maintained after accelerated aging, correlating to shelf-life."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended." Also implies devices meet 5-year shelf-life claim.
ASTM F 2054-07, EN 868-5
Seal StrengthSeals meet required strength for packaging."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
ISTA 2A
Transport TestingDevice integrity and functionality maintained after transport simulation."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
USP <85> , USP <161>
Bacterial EndotoxinEndotoxin levels below specified limits."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended."
Biocompatibility (ISO 10993-1)All biocompatibility tests (e.g., cytotoxicity, sensitization, irritation, material mediated pyrogenicity, LAL Testing) pass for patient contact type."Tests were conducted to show that the nextaro® Transfer System and the nextaro® va were evaluated for biocompatibility in accordance with ISO 10993-1." "The biocompatibility tests show that the subject devices are biocompatible and can therefore be regarded as substantially equivalent in regard to biocompatibility." (Pass)
Material Mediated Pyrogenicity, LAL TestingMaterials are non-pyrogenic."Material Mediated Pyrogenicity LAL Testing" (Implies pass, as part of biocompatibility.)
Sterility (ISO 11135:2014)Sterility Assurance Level (SAL) of 10-6 achieved. Residual EtO and ECH within ISO 10993-7 guidelines."The sterility... were validated in accordance with ISO 11135:2014... SAL 10-6." "The maximum levels of ethylene oxide (EtO) and ethylene chlorohydrin (ECH) residuals remaining on the products were within the guidelines for... ISO 10993-7:2008."
N/A - Roll Inhibition (to design specification)Meets specific design requirements for roll inhibition."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended." (Implies meeting the design specification)
N/A - Transfer performance for residual volume (to design specification)Residual volume after transfer is within specified limits."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended." (Implies meeting the design specification)
N/A - Opening torque (to design specification)Opening torque is within specified limits."Testing conducted [to demonstrate] that the nextaro® Transfer System and nextaro® va perform as intended." (Implies meeting the design specification)
Shelf LifeAt least 3 years (predicate) / 5 years (proposed device).Predicate: > 3 years. Proposed Device: 5 years. "The subject devices are substantially equivalent with regard to their shelf-life."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample sizes for the various tests performed. The testing was conducted by sfm medical devices GmbH, which is located in Waechtersbach Hessen, Germany, suggesting the data provenance is likely Germany. The tests appear to be prospective, laboratory-based engineering and performance tests on the manufactured devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable. The device is a physical medical device (fluid transfer set), not an AI/imaging diagnostic device that would require expert-established ground truth for a test set. Its performance relies on objective measurements against engineering and biocompatibility standards.

4. Adjudication method for the test set

This information is not applicable for this type of device and testing. The tests are based on objective pass/fail criteria from recognized standards, not on subjective expert evaluations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This device is a physical drug transfer system, not an AI or imaging system designed to be used by human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, this question is not applicable to the device described. This is a physical, manually operated medical device, not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for evaluating the nextaro® Transfer System and nextaro® va is based on:

  • Compliance with established international and national standards (ISO, ASTM, USP).
  • Meeting pre-defined engineering design specifications (e.g., residual volume, roll inhibition, opening torque).
  • Laboratory testing results (e.g., retention rate, biocompatibility, sterility) against specified acceptance limits derived from these standards.

8. The sample size for the training set

This information is not applicable. The nextaro® Transfer System and nextaro® va are physical devices, not AI/machine learning models that require a "training set."

9. How the ground truth for the training set was established

This information is not applicable, as there is no training set for a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.

March 15, 2019

sfm medical devices GmbH % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313

Re: K183187

Trade/Device Name: nextaro® Transfer System, nextaro® va Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: LHI Dated: February 18, 2019 Received: February 21, 2019

Dear Mark Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

K183187 - Mark Job

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Sapana Patel -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183187

Device Name

The nextaro® Transfer System and nextaro® va

Indications for Use (Describe) The nextaro® Transfer System and nextaro® va is indicated for the transfer and mixing of drugs contained in vials.

Type of Use (Select one or both, as applicable)

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Traditional 510(k) 510k Summary

510K SUMMARY K183187

Submitter's Nameand Addresssfm medical devices GmbHBrueckenstrasse 5Waechtersbach HessenGERMANY 63607
Manufacturersfm medical devices GmbHBrueckenstrasse 5Waechtersbach HessenGERMANY 63607
Contact Person:Phil TrioloPhone: 801-699-9846Facsimile: 801-328-2399
Date Prepared:March 14, 2019
Device:nextaro® Transfer System and nextaro® va
FDA Product Code:LHI
Class:II
Common or Usual Name:I.V. Fluid Transfer Set
Regulation Description:Intravascular Administration Set
Regulation Number:21 CFR 880.5440
Legally Marketed Predicate:Mix2Vial® Transfer Device (K031861)West Pharmaceutical Services, Inc.
Reason for Submission:New Device
Device Description:The nextaro® Transfer System and nextaro® va is indicated for the transfer andmixing of drugs contained in vials. The nextaro® Transfer System and nextaro® va is

practically one system with two variants which are to be used independently.

{4}------------------------------------------------

TechnologicalCharacteristics:The nextaro® Transfer System consists of two components already assembled(screwed together) in the delivery condition. Each of the components has a plasticspike which is used to perforate the seals of a solvent vial, or lyophilizatedpharmaceutical vial, respectively. The components have a male and female Lueradapter to create an air-tight inner media-carrying system with open ends for fluidtransfer. Both components are equipped with a filter element to prevent particles(fragments of vial seals, undissolved pharmaceutical) from being administered intothe patient's circulatory system.
To be used as intended, the spike of the blue ("upper") part of the nextaro®Transfer System is used to perforate the seal of the solvent vial. The assembly isflipped vertically and the spike of the white ("lower") part of the device is used toperforate the seal of the pharmaceutical vial. The solvent transfer process startsimmediately due to the vacuum in the pharmaceutical vial.
After solvent transfer and reconstitution of the pharmaceutical by gentle swirling,the blue component of the device is removed from the white part by unscrewing, astandard syringe with a male Luer is attached to the exposed female Luer of thedevice to aspirate the ready-to-use solution.
The nextaro® va is a standalone device. Basically, it is the lower part of thenextaro® Transfer System, but without the thread which the latter has to connectthe upper part. After connecting the nextaro® va to a pharmaceutical vial bypushing it downwards, a solvent can be transferred by a syringe via the female Luer.After solvent transfer and reconstitution of the pharmaceutical by gentle swirling, astandard syringe with a male Luer is attached to the female Luer of the device toaspirate the ready-to-use solution.

Indications for The nextaro® Transfer System and nextaro® va is indicated for the transfer and Use: mixing of drugs contained in vials.

Areas for ComparisonPredicate DeviceMix2VialK031861Proposed Devicenextaro®Transfer SystemProposed Devicenextaro® vaComparison
Indications for Use
Intended use orindicationsIntended fortransferring andmixing drugscontained in twovialsIndicated for thetransfer and mixingof drugs containedin vialsIndicated for thetransfer and mixingof drugs containedin vialsSee Discussion ofDifferences
Regulatory Information
Device ClassificationNameSet I.V. FluidTransferSet I.V. FluidTransferSet I.V. FluidTransferEquivalent
Regulation Number880.5440880.5440880.5440Equivalent
Regulation DescriptionIntravascularAdministration SetIntravascularAdministration SetIntravascularAdministration SetEquivalent
Product CodeLHILHILHIEquivalent
Areas for ComparisonPredicate DeviceMix2VialK031861Proposed Devicenextaro®Transfer SystemProposed Devicenextaro® vaComparison
Regulatory MedicalSpecialtyGeneral HospitalGeneral HospitalGeneral HospitalEquivalent
Device Class222Equivalent
GMP ExemptNoNoNoEquivalent
Design Features
Uses vacuum pressureYesYesNoSee Discussion ofDifferences
Vial size the product isto be used with20 mm /14 mm20 mm20 mmEquivalentto 20 mm model
Vial adapter / vial accesscomponentYesYesYesEquivalent
In-line liquid filterYesYesYesEquivalent
Mesh opening15 μm11 μm11 μmSee Discussion ofDifferences
Open area9 %5 %5 %See Discussion ofDifferences
Female Luer adapter forconnection to syringesYesYesYesEquivalent
Luer adapter and vialadapter are integratedin one componentYesYesYesEquivalent
Needleless access to vialYesYesYesEquivalent
Sterile, biocompatiblefluid pathYesYesYesEquivalent
Manufactured by plasticinjection moldingYesYesYesEquivalent
Single use, sterileYesYesYesEquivalent
SterilizationGamma / EtOEtOEtOEquivalent to one ofthe methods (EtO)
Sterility Assurance LevelSAL 10-6SAL 10-6SAL 10-6Equivalent
BiocompatibilityPassPassPassEquivalent
Shelf Life> 3 years5 years5 yearsEquivalent
Principles of Operation
Manually operatedYesYesYesEquivalent
Mechanically connectedto a drug vialYesYesYesEquivalent
Transfers solventmanually using a syringeplunger rodNoNoYessimilar
Drug reconstitutionthrough manualagitation of vial whileconnected to deviceYesYesYesEquivalent
Mixed drug is manuallyaspirated into a syringebarrel using the syringeplunger rodYesYesYesEquivalent
Areas for ComparisonPredicate DeviceMix2VialK031861Proposed Devicenextaro®Transfer SystemProposed Devicenextaro® vaComparison
Single UseYesYesYesEquivalent
Materials
Transfer devicePolycarbonatePolypropylene, TPEPolypropyleneSimilar
In-line FilterPolyethyleneterephthalatePolyethyleneterephthalatePolyethyleneterephthalateEquivalent
PackagingBlister Pack: PETGLid: TyvekBlister Pack: PETGLid: TyvekBlister Pack: PETGLid: TyvekEquivalent

Substantial Equivalence Comparison:

{5}------------------------------------------------

{6}------------------------------------------------

Discussion of Differences

  • Difference in While the indications for use for the nextaro® Transfer System and nextaro® va are worded slightly different than that of the predicate device, both the subject devices Indications: and predicate device are used for the same purpose: The transfer of solvents into a drug vial / the aspiration of the ready-to-use solution of the dissolved pharmaceuticals which are contained in vials. The nextaro® Transfer System is intended to be used with two vials (solvent and pharmaceutical agent vial) just as is the predicate device, while the nextaro® va is intended to be used with one vial (pharmaceutical agent vial; the solvent transferred into pharmaceutical agent vial by a syringe with male Luer adapter filled with solvent). There is no critical difference in the intended therapeutic use of the subject and predicate devices.
  • Uses Vacuum The nextaro® Transfer System, the nextaro® va and the Mix2Vial all are used to Pressure: transfer solvent into a vial which contains a freeze-dried pharmaceutical agent (lyophilizate). All of the devices use pressure to transfer the solvent: In the case of the nextaro® Transfer System, the devices use the negative pressure within the pharmaceutical agent vial to transfer the solvent from the solvent vial which was connected to the device beforehand, just as is the case with the predicate device. The nextaro® va also uses pressure to transfer the solvent, but due to the design of the nextaro® va, the solvent must be applied by means of a syringe. However, the direction of the solvent flow is the same for all predicate and subject devices.

There is no critical difference in the use of vacuum pressure or manual pressure (to the plunger of the syringe) to transfer the solvent. The solvent transfer methods of the subject and predicate devices can be regarded as equivalent.

Filter The nextaro® Transfer System, the nextaro® va and the Mix2Vial all are equipped with Elements: filter elements, to retain fragments which could be punched out during perforation of the vial seals from being transferred or aspirated into the ready-to-use solution. The same applies for undissolved pharmaceutical agent particles.

The filter mesh used in the nextaro® and nextaro® va are capable to retain particles of ≥ 15 um size. A retention rate of > 95 % of particles sized ≥ 15 um has been confirmed by testing. The performance of the filters (retention) of the subject and predicate devices has been shown to be substantially equivalent.

{7}------------------------------------------------

  • Shelf life: The predicate device has a shelf-life of > 3 years while the subject devices have a shelflife of 5 years. The packaging materials are equivalent. The subject devices are substantially equivalent with regard to their shelf-life. This does not raise different questions of safety or effectiveness
  • Both the predicate and the nextaro® devices are made of biocompatible thermoplastic Materials: polymer materials. Although the basic materials for the manufacturing of the devices is not the same, the performance tests show that the subject devices are substantially equivalent regarding the performance of the devices. Both the subject device and predicate use the same filter material (PET). The biocompatibility tests show that the subject devices are biocompatible and can therefore be regarded as substantially equivalent in regard to biocompatibility.

Testing

The following tests were conducted to demonstrate that the nextaro® Transfer System and nextaro® va perform as intended.

StandardTest Performed
ISO 22413• Fragmentation• Penetration Force• Piercing• Verification of Design Specifications for Transfer Devices with Housing
ISO 594-2• Male Conical Fitting / Luer Connector
ISO 8536-4• Flow Rate of Infusion Fluid• Fluid Filter (Retention Rate)• Leakage (Aging/ Submersion Test)• Metal Ions• Particulate Contamination• Penetration Force• Reduction of Oxidizable Matter• Residue on Evaporation• Tensile Strength• Titration Acidity or Alkalinity• UV Absorption of Extract Solution

{8}------------------------------------------------

StandardTest Performed
ISO 11607• Sterile Barrier and Packaging Systems
ASTM D3078• Bubble Emissions
ASTM F 1886• Integrity of Seals
ASTM F 1929-12• Seal Leakage
ASTM F 1980-07• Accelerated Aging
ASTM F 2054-07, EN 868-5• Seal Strength
ISTA 2A• Transport Testing
USP <85> , USP <161>• Bacterial Endotoxin
N/A• Roll Inhibition (to design specification)• Transfer performance for residual volume (to design specification)• Opening torque (to design specification)

Biocompatibility

Tests were conducted to show that the nextaro® Transfer System and the nextaro® va were evaluated for biocompatibility in accordance with ISO 10993-1. Filter particulate testing was also conducted (see above table).

Material Mediated Pyrogenicity LAL Testing

Sterility

The sterility of the nextaro® Transfer System and nextaro® va were validated in accordance with ISO 11135:2014 Sterilization of health care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices. The maximum levels of ethylene oxide (EtO) and ethylene chlorohydrin (ECH) residuals remaining on the products were within the guidelines for limited and tolerable contact limit devices according to ISO 10993-7:2008.

Summary / Conclusion

The evaluation of the proposed nextaro® devices has demonstrated through performance testing to be substantially equivalent to the predicate device. Based on the device type, indications, intended use, technological characteristics, physical characteristics, the nextaro® Transfer Device and nextaro® va are substantially equivalent to the predicate device.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.