K Number
K182838

Validate with FDA (Live)

Date Cleared
2018-11-08

(30 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Geistlich Derma-Gide™ is intended for the management of wounds including:
· partial and full thickness wounds

  • · pressure ulcers
  • venous ulcers
  • · diabetic ulcers
  • · chronic vascular ulcers
  • · surgical wounds (donor sites/grafts, post Moh's surgery, post laser surgery, podiatric, wound dehiscence)
  • · trauma skin wounds (abrasions, laceration, second degree burns, skin tears)
Device Description

Geistlich Derma-Gide™ is an acellular advanced wound care device. The Geistlich Derma-Gide™ is derived from porcine tissue (mostly collagen Type 1) and is processed using proprietary technologies to remove bacteria and inactivate viruses. Geistlich Derma-Gide"" features a bilayer structure. The upper dense compact collagen layer protects the wound and the open healing process. This structure has a smooth texture with appropriate pull out strength properties to allow suturing. The second lower layer consists of a thick, porous spongeous collagen scaffold.

AI/ML Overview

This document is a 510(k) Premarket Notification from the FDA for a medical device called Geistlich Derma-Gide™. It is an equivalence claim to an existing device, adding new sizes and shapes. Therefore, it does not contain a study to prove acceptance criteria for a new device's performance.

Here's a breakdown based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Not applicable. This document is for a special 510(k) submission, indicating minor changes to an already cleared device. The "acceptance criteria" here relate to demonstrating substantial equivalence to a predicate device, rather than performance criteria for a novel device. The document explicitly states: "New performance testing was not necessary to support the additional sizes/shapes or the updates to the IFU."

CriterionReported Performance
Indications for UseIdentical to Predicate
Device DescriptionIdentical to Predicate (except for added sizes/shapes)
MaterialIdentical to Predicate (Porcine)
ThicknessIdentical to Predicate (2.5 - 5 mm)
SterilityIdentical to Predicate (Gamma-sterilized)
Single UseIdentical to Predicate (Yes)
BiocompatibilityExisting information from K171842 applies
Sterilization Assurance LevelExisting information from K171842 applies
Shelf LifeExisting information from K171842 applies
Safety and Effectiveness IssuesNo new questions of safety or effectiveness raised by additional sizes/shapes or IFU updates (Claimed)

2. Sample size used for the test set and the data provenance:

Not applicable. No new performance testing with a "test set" was conducted for this submission. The basis for equivalence relies on the predicate device's existing data and the argument that the changes (new sizes/shapes) do not alter the fundamental characteristics or performance of the device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

Not applicable. No new test set requiring expert-established ground truth was part of this submission. The evaluation is focused on regulatory equivalence.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable. See point 3.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a wound dressing, not an AI-powered diagnostic tool. The submission does not involve AI or human reader performance studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This device is a physical wound dressing, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

Not applicable in the context of new device performance testing. The "ground truth" for this submission is the established safety and effectiveness of the predicate device (Geistlich Wound Matrix, K171842), based on its original clearance. The current submission argues that the new sizes and shapes maintain this established safety and effectiveness.

8. The sample size for the training set:

Not applicable. No machine learning or AI models are involved.

9. How the ground truth for the training set was established:

Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 8, 2018

Geistlich Pharma AG % Samie Allen Consultant, Devices Biologics Consulting Group, Inc. 1555 King Street, Suite 300 Alexandria, Virginia 22314

Re: K182838

Trade/Device Name: Geistlich Derma-Gide Regulatory Class: Unclassified Product Code: KGN Dated: October 5, 2018 Received: October 9, 2018

Dear Samie Allen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Cynthia Chang -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182838

Device Name Geistlich Derma-Gide™

Indications for Use (Describe)

The Geistlich Derma-Gide™ is intended for the management of wounds including:

· partial and full thickness wounds

  • · pressure ulcers
  • venous ulcers
  • · diabetic ulcers
  • · chronic vascular ulcers
  • · surgical wounds (donor sites/grafts, post Moh's surgery, post laser surgery, podiatric, wound dehiscence)
  • · trauma skin wounds (abrasions, laceration, second degree burns, skin tears)

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the 510(k) Summary is provided below.

SUBMITTER 1.

Geistlich Pharma AG Bahnhofstrasse 40 CH-6110 Wolhusen Switzerland

Contact Person: Marco Steiner Phone: +41 41 492 67 64 Email: marco.steiner@geistlich.ch

Prepared By: Samie Allen, Biologics Consulting Group, sallen@biologicsconsulting.com Date Prepared: November 7, 2018

2. DEVICE

Name of Device: Geistlich Derma-Gide™ Common Name: Collagen Wound Dressing Classification Regulation/Class: unclassified Product Code: KGN Panel: General and Plastic Surgery

PREDICATE DEVICE 3.

Predicate Device: Geistlich Wound Matrix (subsequently renamed to Geistlich Derma-Gide™) (K171842)

DEVICE DESCRIPTION 4.

Geistlich Derma-Gide™ is an acellular advanced wound care device. The Geistlich Derma-Gide™ is derived from porcine tissue (mostly collagen Type 1) and is processed using proprietary technologies to remove bacteria and inactivate viruses. Geistlich Derma-Gide"" features a bilayer structure. The upper dense compact collagen layer protects the wound and the open healing process. This structure has a smooth texture with appropriate pull out strength properties to allow suturing. The second lower layer consists of a thick, porous spongeous collagen scaffold.

The sizes cleared under K171842 include:

  • Rectangular: 1.5 x 2.0 cm; 2.0 x 3.0 cm; and 3.0 x 4.0 cm; the thickness ranges from ● 2.5-5 mm

{4}------------------------------------------------

The new sizes in this Special 510(k) include:

  • Rectangular: 1.0 x 1.5 cm; 2.0 x 2.0 cm; 2.0 x 4.0 cm; the thickness ranges from 2.5-5 ● mm
  • Round (disk) shape: 10, 12, 14, 16, 18, 20, 22, and 24 mm diameters; the thickness . ranges from 2.5-5 mm.

Fixation by sutures is possible. The device can be trimmed as needed. The device is terminally sterilized (by gamma irradiation) in its packaging. The device is intended to be used by licensed surgeons and will be supplied sterile for single use.

INDICATION FOR USE ട്.

The Geistlich Derma-Gide™ is intended for the management of wounds including:

  • partial and full thickness wounds ●
  • pressure ulcers
  • venous ulcers ●
  • diabetic ulcers
  • chronic vascular ulcers
  • surgical wounds (donor sites/grafts, post Moh's surgery, post laser surgery, podiatric, wound dehiscence)
  • trauma skin wounds (abrasions, laceration, second degree burns, skin tears) ●

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS 6.

The subject Geistlich Derma-Gide™ has the same technological characteristics as the predicate Geistlich Wound Matrix (K171842). They have the same thickness (2.5-5 mm), source material, manufacturing, sterilization, and packaging. The only difference is that 3 additional rectangular sizes and 8 round (disk) sizes are being added to the sizes cleared under K171842. Fixation by sutures is possible for all sizes. All devices can be trimmed as needed.

All of the new sizes are smaller than the 3.0 x 4.0 cm device size cleared under K171842. However, the new range of sizes or the additional round-shaped sizes do not raise any new questions of safety or effectiveness based on the intended use of the devices. The additional sizes and shapes are for user convenience and minimizing waste.

Table 1 provides a summary comparison between the subject and predicate devices.

{5}------------------------------------------------

Subject DevicePredicate Device
510(k) NumberTBDK171842
ApplicantGeistlich Pharma AGGeistlich Pharma AG
Device NameGeistlich Derma-GideGeistlich Wound Matrix (tradenamesubsequently changed to GeistlichDerma-Gide)
ClassificationRegulation, ClassUnclassifiedsame
Product CodeKGNsame
Indications for Useintended for the management of woundsincluding:• partial and full thickness wounds• pressure ulcers• venous ulcers• diabetic ulcers• chronic vascular ulcers• surgical wounds (donor sites/grafts, postMoh's surgery, post laser surgery,podiatric, wound dehiscence)• trauma skin wounds (abrasions, laceration,second degree burns, skin tears)same
Device DescriptionThe Geistlich Derma-Gide is derived fromporcine tissue (mostly collagen Type 1) and isprocessed using proprietary technologies toremove bacteria and inactivate viruses.Geistlich Derma-Gide features a bilayerstructure. The upper dense compact layerprotects the wound and the open healingprocess. This structure has a smooth texturewith appropriate pull out strength properties toallow suturing. The second lower layerconsists of a thick, porous spongeous scaffold.same
MaterialPorcinesame
Thickness2.5 - 5 mmsame
Range of sizesNew Sizes:Rectangular: 1.0 x 1.5 cm; 2.0 x 2.0 cm; 2.0 x4.0 cmRound (disk) shape: 10, 12, 14, 16, 18, 20, 22,and 24 mm diametersRectangular: 1.5 x 2.0 cm; 2.0 x 3.0 cm; and3.0 x 4.0 cm
Single UseYessame
Subject DevicePredicate Device
SterileGamma-sterilizedsame

Table 1: Device Comparison Table

{6}------------------------------------------------

7. PERFORMANCE DATA

Because there is no change to the device thickness, source material, manufacturing, sterilization, or packaging compared to the predicate K171842, the existing biocompatibility, sterilization, and shelf life information from K171842 fully applies.

New performance testing was not necessary to support the additional sizes/shapes or the updates to the IFU.

CONCLUSIONS 8.

The subject Geistlich Derma-Gide™ is identical in indications to the predicate K171842. The design changes involve 3 additional rectangular sizes and 8 new round (disk) shaped versions of the device. The additional design configurations are intended to simply provide ready-to-use shapes which require less trimming to the wound size to reduce waste. Overall, the convenience for the user is improved. The new design configurations raise no new questions of safety or effectiveness.

There are no changes in product thickness, source material, manufacturing, sterilization, or packaging.

None of the changes to the IFU raise a new question of safety and effectiveness.

Accordingly, the Geistlich Derma-Gide™ is substantially equivalent to the predicate Geistlich Wound Matrix cleared under K171842.

N/A