K Number
K182744

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-02-01

(126 days)

Product Code
Regulation Number
864.8625
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HemoTrol® WB is an assayed quality control material for professional use to verify the performance characteristics of the HemoCue ® Hb 301 and the HemoCue ® Hb 801 System. HemoTrol® WB is intended for the quantitative determination of hemoglobin.

Device Description

HemoTrol® WB is an assayed hemoglobin quality control material intended for professional use in the verification of the precision and accuracy of the HemoCue® 301 and HemoCue® 801 systems. HemoTrol® WB contains bovine red blood cells with hemoglobin lysates in MetHb and additional stabilizers. For daily quality control, three physiological relevant levels are available.

HemoTrol® WB solutions are filled in reclosable plastic primary containers. Each bottle contains 1.1 g of HemoTrol® WB solution. The primary containers are equipped with colored polypropylene caps. Cap color depends on the concentration of hemoglobin (Low: red cap; High: blue cap). Two (2) bottles of the same level are placed in a plastic blister and packed in a product box together with the combined instructions for use (IFU) and value sheet. Both the primary containers and product box are labeled.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (HemoTrol® WB) and focuses on establishing substantial equivalence to a predicate device. It describes performance data related to stability testing of the quality control material itself, rather than a clinical study evaluating the performance of an AI-powered diagnostic device in a patient population.

Therefore, the information required to answer your query about acceptance criteria and study proving a device meets acceptance criteria, specifically for an AI/diagnostic device, is not present in the provided text. The document describes:

  1. Device: HemoTrol® WB (a quality control material for hemoglobin measurement, not an AI diagnostic device).
  2. Purpose: To demonstrate substantial equivalence to a predicate quality control device (Eurotrol HemoTrol).
  3. Performance Data: Focuses on stability of the quality control material (open-vial and closed-vial stability), not diagnostic performance metrics like sensitivity, specificity, or reader improvement in an MRMC study.

Here's a breakdown of why I cannot fulfill your request based on this document:

  • Acceptance Criteria for AI/Diagnostic Performance: The text does not mention any acceptance criteria for sensitivity, specificity, AUC, or other typical metrics for an AI or diagnostic device. The "acceptance criteria" presented here implicitly relate to demonstrating stability within acceptable ranges for a quality control material.
  • Study Proving Acceptance Criteria: The studies described ("Open Stability" and "Closed Stability") are designed to show how long the quality control material remains stable, not to evaluate the performance of an AI or diagnostic device against a ground truth in a clinical setting.
  • Sample Size (Test Set): Not applicable in the context of an AI/diagnostic test set. For stability testing, the sample size was "n=6" for open stability and "n=8" for closed stability.
  • Data Provenance: Not relevant for these stability tests. The "samples used were produced according the regular production process of HemoTrol® WB."
  • Experts for Ground Truth: Not applicable. Ground truth for stability is measurement against a known reference method for hemoglobin on specific analyzers (HemoCue® Hb 301 and HemoCue® Hb 801 systems).
  • Adjudication Method: Not applicable.
  • Multi-Reader Multi-Case (MRMC) Study: Not performed, as this is a quality control material, not a diagnostic device for human interpretation.
  • Standalone (Algorithm Only) Performance: Not applicable.
  • Type of Ground Truth: For stability, the ground truth is the measured hemoglobin value on the specified HemoCue systems.
  • Training Set Sample Size: Not applicable, as this is not an AI/machine learning device requiring a training set.
  • Ground Truth for Training Set: Not applicable.

To answer your detailed questions, you would need a document that describes the clinical performance evaluation of an AI-powered diagnostic device in a patient population.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 1, 2019

Eurotrol B.V. Mylene de Boer QA/RA Officer Keplerlaan 20 Ede, Gelderland, 6716BS, NL

Re: K182744

Trade/Device Name: HemoTrol WB - Low, HemoTrol WB - Normal and HemoTrol WB - High Regulation Number: 21 CFR 864.8625 Regulation Name: Hematology quality control mixture Regulatory Class: Class II Product Code: GGM Dated: September 26, 2018 Received: September 28, 2018

Dear Mylene de Boer:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Leonthena R. Carrington -S

Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182744

Device Name

HemoTrol® WB - Low, HemoTrol® WB - Normal and HemoTrol® WB - High

Indications for Use (Describe)

HemoTrol® WB is an assayed quality control material for professional use to verify the performance characteristics of the HemoCue ® Hb 301 and the HemoCue ® Hb 801 System. HemoTrol® WB is intended for the quantitative determination of hemoglobin.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the word "Confidential" in a light gray color. The word is slanted diagonally across the image. The background is a plain white color.

Image /page/3/Picture/1 description: The image shows the Eurotrol logo. The logo consists of a red circle with a white center, surrounded by two gray ovals. Below the logo is the word "Eurotrol" in a bold, sans-serif font. Below the word "Eurotrol" is the phrase "Your Global Reference Point for Quality Control" in a smaller, sans-serif font.

510(K) Summary HemoTrol WB

September 2018

This summary of the 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92.

Submitter

Eurotrol B.V.

Keplerlaan 20

6716 BS, Ede, the Netherlands

T: +31 318 695777

Primary Contact: Ms. Mylène de Boer, mdeboer@eurotrol.com

Device Information

Device nameREF
HemoTrol® WB – Low08718734960942
HemoTrol® WB – Normal08718734960959
HemoTrol® WB – High08718734960966

Classification

Class II (21 CFR 864,8625) GGM - control, Hemoglobin

Predicate Device

Eurotrol HemoTrol

510(k) number: K964052

Device Description

HemoTrol® WB is an assayed hemoglobin quality control material intended for professional use in the verification of the precision and accuracy of the HemoCue® 301 and HemoCue® 801 systems. HemoTrol® WB contains bovine red blood cells with hemoglobin lysates in MetHb and additional stabilizers. For daily quality control, three physiological relevant levels are available.

HemoTrol® WB solutions are filled in reclosable plastic primary containers. Each bottle contains 1.1 g of HemoTrol® WB solution. The primary containers are equipped with colored polypropylene caps. Cap color depends on the concentration of hemoglobin (Low: red cap; High: blue cap). Two (2) bottles of the same level are placed in a plastic blister and packed in a product box together with the combined instructions for use (IFU) and value sheet. Both the primary containers and product box are labeled.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Eurotrol logo. The logo consists of a red circle with a white center, surrounded by two gray ovals. Below the logo is the word "Eurotrol" in a bold, sans-serif font. Below the word "Eurotrol" is the phrase "Your Global Reference Point for Quality Control" in a smaller, sans-serif font.

Intended Use

New Device

Confidential

HemoTrol® WB is an assayed quality control material for professional use to verify the performance characteristics of the HemoCue ® Hb 301 and the HemoCue ® Hb 801 System. HemoTrol® WB is intended for the quantitative determination of hemoglobin.

Predicate Device

Eurotrol HemoTrol is an assayed hemoglobin control intended for professional use in the verification of the precision and accuracy of the HemoCue® B-Hemoglobin and the HemoCue Hb 201 systems.

The indication for use for HemoTrol® WB is not identical to the predicate device; however, the differences do not alter the intended diagnostic use of the device, nor do they affect the safety or effectiveness of the device compared to the predicate device. Both Hemotro® WB and the predicate device are assayed quality control materials for professional use to verify the performance characteristics of a hemoglobin analyzer but are designed for a different hemoglobin analyzer.

ParameterSimilaritiesDifferences
AnalyteHemoglobinNone
Analyte ConcentrationBoth devices cover the range of 95-160 g/LPredicate device also includes:80-94 g/L
Color of liquidBloodlikeNone
ContentsPurified bovine hemolisate and preservatives.Predicate device does notcontain stabilized red blood cells
Filling volume1 mlNone
Intended UseQuality Control material forprofessional use to verify theperformance characteristics of aspecific Hemoglobin analyzer.Different target Hemoglobinanalyzer
Number of Levels3 levels: Low, Normal, HighNone
Open vail stability31 daysNone
Primary ContainerReclosable plastic primary containerswith polypropylene capsNone
Principle of OperationsThe absorbance is measured by usinga dual wavelength spectrophotometrictechnology.Measurements are performed at506 nm and 880 nm. Predicatedevice at 570 nm and 880 nm.
ProcedureAllow the vial to stand for 15 minutes atroom temperatureMix the vial before sampling.Do not fill the cuvette from the vial.Dispense a drop of the control materialonto a hydrophobic surface.Fill the cuvette according to themanufacturer's instructions.Wipe any excess material from the vialand the cap with a clean tissue.Recap the vial tightly.None

Predicate Device Comparison

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the Eurotrol logo. The logo consists of a red circle with a white center, surrounded by two gray ovals. Below the logo is the word "Eurotrol" in a serif font. The logo is simple and modern, and the colors are eye-catching.

Your Global Reference Point for Quality Control

ParameterSimilaritiesDifferences
Secondary PackagingTwo (2) bottles of the same level areplaced in a plastic blister and packed ina product box together with thecombined instructions for use (IFU)and value sheet.None
Storage temperature2 - 8 degrees CelsiusNone
Where usedPOC and laboratoryNone

Performance Data

Open Stability:

onfidential

According to the guidelines described in "CLSI-EP25A-2009" at least on batch of a quality control was monitored to determine the in-use stability. One batch of each level was monitored at four (4) time points (n=6) for in total 35 days at 2-8ºC. Based on this study, HemoTrol® WB is stable for 31 days when stored at 2-8ºC.

Closed Stability:

According to guidelines in "CLSI-EP25A-2009" at least 3 batches of a quality control were monitored to determine the shelf life. Three (3) batches of the three HemoTro® WB levels were monitored at 8 time points for up to 127 days (n=8) at 2-8℃. The samples used were produced according the regular production process of HemoTrol® WB. The total hemoglobin levels were measured on the HemoCue® Hb 301 and HemoCue® Hb 801 systems. Based on this study, HemoTrol® WB is stable for 108 days when stored at 2-8ºC.

Conclusion

Based on the demonstrated intended use, performance characteristics and comparison of technical characteristics HemoTrol® WB is deemed substantially equivalent to the predicate device: Eurotrol HemoTrol (K964052)

§ 864.8625 Hematology quality control mixture.

(a)
Identification. A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).(b)
Classification. Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.