K Number
K182648

Validate with FDA (Live)

Date Cleared
2019-01-29

(127 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Muscular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), OB/Gyn and Urology.

Device Description

This SonoScape X5 Series Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

The X5 Series system utilizes the ultrasound echo characteristics, transmits ultrasonic energy into patient body, sweeps in a certain direction, processes the signals according to the delay time and the echo strength, and images the organs by using the electronic circuits and backend controller to process, then analyzes the distance and the status of organs; and at the same time, this system utilizes Doppler and auto-correlation technology to image the blood flow and add the color-coding information to the grayscale image of B mode, then displays the image in real time. The probes provided with this system are electrical-acoustical and acoustical-electrical transducers. The probes firstly convert the electric excitation signal to the acoustic signal and transmit the signal into the patient body, then converts the echo signals from the patient body to electric signal. The echo signal is processed and converted by DSC to image signal to output to the LCD display.

This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array.

This system consists of keyboard control panel, power supply module, color LCD monitor and optional probes.

This system (X5 Series Digital Color Doppler Ultrasound System) are additional models (X5 Exp, X5 Pro, X6, X6 Exp, X6 Pro), additional transducers (C1-6, 12L-B, 9L-A, 18L-A, 13L-A, L746, 10L-I, 1012, 6V7, 6V3, 6V3A, C361, VC6-2, C542, C322, 12LT-A, 12LI-A, 6CT-A, 6CI-A, BCC9-5, S1-5, 2P1, CWD5.0, CWD2.0, MPTEE and MPTEE min probes), additional functions (Elastography, Contrast imaging) and updated intended use (Trans-esoph.(Cardiac)) to legally marketed SonoScape X5 (K160258).

AI/ML Overview

The provided document describes the Sonoscape Medical Corp.'s X5 Series Digital Color Doppler Ultrasound System (K182648) and its substantial equivalence to predicate devices, rather than detailing a study that proves the device meets specific performance acceptance criteria for a diagnostic AI algorithm.

The document primarily focuses on demonstrating substantial equivalence to legally marketed predicate devices by comparing intended use, technical characteristics, and probes, as well as adherence to recognized safety and performance standards. No information regarding an AI algorithm's performance or study details (like sample size for test sets, ground truth establishment, expert qualifications, or MRMC studies) is present.

Therefore, I cannot extract the requested information regarding acceptance criteria and the study that proves an AI device meets these criteria, as the provided text pertains to a general diagnostic ultrasound system and its transducers, not an AI-powered diagnostic tool.

The "Performance test" section under "Non-Clinical Tests" lists compliance with electrical safety, EMC, acoustic testing, and software verification/validation standards, indicating the device meets these standards. However, these are general device safety and performance criteria, not specific diagnostic performance criteria for an AI algorithm.


Disclaimer: The provided document is a 510(k) summary for a medical device and does not contain details about the performance of an AI algorithm or a study testing such an algorithm. The information below is based solely on the content available in the provided text, which primarily addresses the substantial equivalence of a physical ultrasound system to predicate devices. As such, many of the requested fields are not applicable or cannot be extracted from this document because it does not describe an AI medical device or its validation study.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 29, 2019

Sonoscape Medical Corp. Toki Wu Regulatory Affairs Manager 4/f. 5/f. 8/f. 9/f & 10/f Yizhe building, Yuquan road, Nanshan SHENZHEN, 518051 GUANGDONG CHINA

Re: K182648

Trade/Device Name: X5 Series Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN Dated: December 14, 2018 Received: December 27, 2018

Dear Toki Wu:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Hol. 2. mils

for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K182648

Device Name

X5 Series Digital Color Doppler Ultrasound System

Indications for Use (Describe)

The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Muscular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), OB/Gyn and Urology.

Type of Use (Select one or both, as applicable)

× Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

{3}------------------------------------------------

X5 Series Digital Color Doppler Ultrasound System System: (including: X5 Exp, X5, X5 Pro, X6 Exp, X6, X6 Pro) Diagnostic Ultrasound Pulsed Echo System Diagnostic Ultrasound Pulsed Doppler Imaging System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2
AbdominalPPPPPNote 1Notes 2,8
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2
Small Organ (specify)PPPPPNote 1Notes 2,6,7
Neonatal CephalicPPPPPPNote 1Notes 2,3
FetalAdult CephalicPPPPPPNote 1Notes 2,3
Imaging&Trans-rectalPPPPPNote 1Notes 2
OtherTrans-vaginalPPPPPNote 1Notes 2
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2
Other (Urology)PPPPPNote 1Notes 2
Cardiac AdultPPPPPPNote 1Notes 2,3
Cardiac PediatricPPPPPPNote 1Notes 2,3
CardiacIntravascular(Cardiac)
Trans-esoph.(Cardiac)PPPPPPNote 1Notes 2,3
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselPPPNPPNote 1Notes 2
VesselCerebral vascularN

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{4}------------------------------------------------

Transducer: 3C-A Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1
AbdominalPPPPPNote 1
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{5}------------------------------------------------

Transducer: C613 Curved Array

Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
AbdominalPPPPPNote 1Notes 2
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac PediatricPPPPPPNote 1Notes 2
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{6}------------------------------------------------

Transducer: 3P-A Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
AbdominalPPPPPNote 1Notes 2
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3
FetalAdult CephalicPPPPPPNote 1Notes 2,3
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac AdultPPPPPPNote 1Notes 2,3
Cardiac PediatricPPPPPPNote 1Notes 2,3
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{7}------------------------------------------------

Transducer: 7P-B Phased Array

Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
CardiacCardiac PediatricPPPPPPNote 1Notes 2
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{8}------------------------------------------------

Transducer: L741 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,6,7
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselPPPPPNote 1Notes 2
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{9}------------------------------------------------

Transducer: 6V1 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2
Trans-vaginalPPPPPNote 1Notes 2
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{10}------------------------------------------------

Transducer: EC9-5 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2
Trans-vaginalPPPPPNote 1Notes 2
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{11}------------------------------------------------

Transducer: C322 Micro-Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2
AbdominalPPPPPNote 1Notes 2,8
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{12}------------------------------------------------

Transducer: C1-6 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2
AbdominalPPPPPNote 1Notes 2,8
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{13}------------------------------------------------

Transducer: VC6-2 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2
AbdominalPPPPPNote 1Notes 2
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{14}------------------------------------------------

Transducer: 2P1 Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
CardiacCardiac PediatricPPPPPPNote 1Notes 2,3
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

P = previously cleared by FDA; E = added under this appendix N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{15}------------------------------------------------

Transducer: S1-5 Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
AbdominalPPPPPNote 1Notes 2
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3
FetalAdult CephalicPPPPPPNote 1Notes 2,3
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac AdultPPPPPPNote 1Notes 2,3
Cardiac PediatricPPPPPPNote 1Notes 2,3
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{16}------------------------------------------------

Transducer: L746 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
FetalImaging&OtherAbdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,6
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2
CardiacIntravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{17}------------------------------------------------

Transducer: 1012 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,6
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{18}------------------------------------------------

Transducer: 9L-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,6,7
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2
Intravascular
Other (Ob/GYN)
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{19}------------------------------------------------

Transducer: 12L-B Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,6,7
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2
Intravascular
Other (Ob/GYN)
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2
Cerebral vascular

P = previously cleared by FDA; E = added under this appendix N = new indication; Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{20}------------------------------------------------

Transducer: 6V3 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectalPPPPPNote 1Notes 2
OtherTrans-vaginalPPPPPNote 1Notes 2
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{21}------------------------------------------------

Transducer: 6V7 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*Combined
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2
Trans-vaginalPPPPPNote 1Notes 2
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
Peripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{22}------------------------------------------------

Transducer: BCC9-5 Biplane (curved+curved) Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2
Trans-vaginalPPPPPNote 1Notes 2
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{23}------------------------------------------------

Transducer: MPTEE Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)PPPPPPNote 1Notes 2,3
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{24}------------------------------------------------

Transducer: MPTEE mini Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)PPPPPPNote 1Notes 2,3
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{25}------------------------------------------------

Transducer: 13L-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,6
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2
Intravascular
Other (Ob/GYN)
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{26}------------------------------------------------

Transducer: 18L-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2
Cerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{27}------------------------------------------------

Transducer: 10L-I Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2
Cerebral vascular

P = previously cleared by FDA; E = added under this appendix N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{28}------------------------------------------------

Transducer: 6V3A Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginalPPPPPNote 1Notes 2
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; E = added under this appendix N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{29}------------------------------------------------

Transducer: 12LT-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,6
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselPPPPPNote 1Notes 2
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{30}------------------------------------------------

Transducer: 12LI-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{31}------------------------------------------------

Transducer: 6CI-A Curved Array

Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2
AbdominalPPPPPNote 1Notes 2
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{32}------------------------------------------------

Transducer: 6CT-A Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2
AbdominalPPPPPNote 1Notes 2
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

P = previously cleared by FDA; E = added under this appendix N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{33}------------------------------------------------

Transducer: CWD2.0

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac AdultP
Cardiac PediatricP
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{34}------------------------------------------------

Transducer: CWD5.0

Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselN
Cerebral vascularN

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{35}------------------------------------------------

Transducer: C542 Curved Array

Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
AbdominalPPPPPNote 1Notes 2
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac PediatricPPPPPPNote 1Notes 2
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{36}------------------------------------------------

Transducer: C361 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2
AbdominalPPPPPNote 1Notes 2
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2
Other (Urology)
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

P = previously cleared by FDA; E = added under this appendix N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

Note 3: TDI

{37}------------------------------------------------

510(k) Summary K182648

1. Submitter [21 CFR807.92 (a) (1)]

Submitter:SONOSCAPE MEDICAL CORP.
Address:4/f, 5/f, 8/f, 9/f & 10/f, Yizhe Building, Yuquan Road,
Nanshan, Shenzhen 518051, Guangdong, China
Contact Person:Toki Wu
Tel:+86 755 26722890
Fax:+86 755 26722850
Email:ra@sonoscape.net
Date PreparedJanuary 22 2019

2. Device [21 CFR807.92 (a) (2)]

Trade Name:X5 Series Digital Color Doppler Ultrasound System
Model:X5/X5 Exp/X5 Pro/X6/X6 Exp/X6 Pro
Common Name:Diagnostic Ultrasound System and Transducers

Classification Regulatory:

CFR NumberProduct Code
Ultrasonic Pulsed Doppler Imaging System(Primary)892.155090-IYN
Ultrasonic Pulsed Echo Imaging System892.156090-IYO
Diagnostic Ultrasound Transducer892.157090-ITX
Classification Panel:Radiology
Device Class:II

3. Predicate Device(s) [21 CFR 807.92(a) (3)]

The identified predicate device within this submission is as follows:

TypeManufacturerDevice510 (K) Number
Main predicatedeviceSonoscapeMedical Corp.P10 Series Digital ColorDoppler Ultrasound SystemK173058
Reference DeviceSonoscapeMedical Corp.X5 Digital Color DopplerUltrasound SystemK160258
Reference DevicePhilipsUltrasound, Inc.EPIQ 5/EPIQ 7 DiagnosticUltrasound SystemK160807

{38}------------------------------------------------

4. Device Description [21 CFR 807.92(a) (4)1

This SonoScape X5 Series Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

The X5 Series system utilizes the ultrasound echo characteristics, transmits ultrasonic energy into patient body, sweeps in a certain direction, processes the signals according to the delay time and the echo strength, and images the organs by using the electronic circuits and backend controller to process, then analyzes the distance and the status of organs; and at the same time, this system utilizes Doppler and auto-correlation technology to image the blood flow and add the color-coding information to the grayscale image of B mode, then displays the image in real time. The probes provided with this system are electrical-acoustical and acoustical-electrical transducers. The probes firstly convert the electric excitation signal to the acoustic signal and transmit the signal into the patient body, then converts the echo signals from the patient body to electric signal. The echo signal is processed and converted by DSC to image signal to output to the LCD display.

This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array.

This system consists of keyboard control panel, power supply module, color LCD monitor and optional probes.

This system (X5 Series Digital Color Doppler Ultrasound System) are additional models (X5 Exp, X5 Pro, X6, X6 Exp, X6 Pro), additional transducers (C1-6, 12L-B, 9L-A, 18L-A, 13L-A, L746, 10L-I, 1012, 6V7, 6V3, 6V3A, C361, VC6-2, C542, C322, 12LT-A, 12LI-A, 6CT-A, 6CI-A, BCC9-5, S1-5, 2P1, CWD5.0, CWD2.0, MPTEE and MPTEE min probes), additional functions (Elastography, Contrast imaging) and updated intended use (Trans-esoph.(Cardiac)) to legally marketed SonoScape X5 (K160258).

Note: X5 Exp, X5, X5 Pro, X6 Exp, X6, X6 Pro models are considered as a serial products as they are the same except partial functions, including the same design, hardware, software, mechanic construction, power supply board, main board, specification and etc. The differences among them can be seen as the following table.

ModelConfiguration Description
X5 ExpWith full functions and configuration
X5Does not support the velocity measurement compared with X5 Exp
X5 ProDoes not support the slope ratio measurement compared with X5 Exp

Table 1 Comparison table

{39}------------------------------------------------

X6 ExpDoes not support the time ratio measurement compared with X5 Exp
X6Does not support the time ratio, slope ratio measurement compared with X5 Exp
X6 ProDoes not support the time ratio, velocity measurement compared with X5 Exp
Note: The slope ratio, time ratio, velocity and velocity ratio measurement items mentionedabove refer to those of the M-mode basic measurements for Urology.

5. Intended Use [21 CFR 807.92(a) (5)]

The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal. Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), OB/Gyn and Urology.

6. Comparison with the Predicate device [21 CFR 807.92(a) (6)]

X5 Series Digital Color Doppler Ultrasound System is comparable with and substantially equivalent to the predicate device:

TypeManufacturerDevice510 (K) Number
MaindeviceSonoscapeMedical Corp.P10 Series Digital ColorDoppler Ultrasound SystemK173058
Reference DeviceSonoscapeMedical Corp.X5 Digital Color DopplerUltrasound SystemK160258
Reference DevicePhilipsUltrasound, Inc.EPIQ 5/EPIQ 7 DiagnosticUltrasound SystemK160807

X5 Series Digital Color Doppler Ultrasound System has the same intended uses, complies with the same regulation and safety standards, has the consistent acoustic output levels, has similar probes and has the same technical characteristics as the predicate device legally marketed P10 Series (K173058) , X5 (K160258) and EPIQ 5/EPIQ (K160807).

Intended Use Comparison:

Compared with the predicate device P10 Series (K173058), the Subject Device X5 Series has the same intended use.

Table 2 Intended use Comparison

Subject DevicePredicate DeviceRemark
------------------------------------------

{40}------------------------------------------------

SonoScape X5/X5 Exp/X5 Pro/X6/X6Exp/X6 ProLegally marketedSonoScape P10 Exp, P10, P10Pro,P11, M10, P9 (K173058)Same
The Digital Color Doppler UltrasoundSystem is a general-purposeultrasonic imaging instrumentintended for use by a qualifiedphysician for evaluation of Fetal,Abdominal, Pediatric, Small Organ(breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal,Trans-vaginal, PeripheralVascular, Cerebral Vascular,Musculo-skeletal (Conventional andSuperficial), Cardiac (pediatric andadult), Trans-esoph.(Cardiac),OB/Gyn and Urology.The device is a general-purposeultrasonic imaging instrumentintended for use by a qualifiedphysician for evaluation of Fetal,Abdominal, Pediatric, Small Organ(breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal,Trans-vaginal, Peripheral Vascular,Cerebral Vascular, Musculo-skeletal(Conventional and Superficial),Cardiac (pediatric and adult),Trans-esoph.(Cardiac), Laparoscopic,OB/Gyn and Urology.

Technical Characteristics Comparison:

Compared with the predicate device P10 Series (K173058), the Subject Device X5 Series has the similar technical characteristics, including Design, Operation Controls, Display Modes, Operation Modes, Measurement Items, Cine Loop, Operating and Storage Condition. And the differences will not raise new risk and different questions of safety and effectiveness.

Probes Comparison:

Subject device X5 Series has the similar probes as the predicate device SonoScape P10 Series (K173058).

Subject devicePredicate Device
SonoScape X5/X5 Exp/X5Pro/X6/X6 Exp/X6 ProSonoScape P10 Exp, P10,P10Pro, P11, M10, P9 (K173058)Remark
C322 Micro-curved ArrayC322 Micro-curved Array
C361 Micro-curved ArrayC351 Micro-curved Array
3C-A Curved ArrayC361 Micro-curved Array
C1-6 Curved Array3C-A Curved Array
6CT-A Curved ArrayC1-6 Curved Array
6CI-A Curved Array6CT-A Curved Array
6CI-A Curved Array
VC6-2 Curved ArrayVC6-2 Curved Array
C613 curved ArrayC613 curved Array
C542 curved ArrayC542 curved Array
2P1 Phased Array2P1 Phased Array
S1-5 Phased ArrayS1-5 Phased Array
7P-B Phased Array7P-B Phased Array
3P-A Phased Array3P-A Phased Array
L741 Linear ArrayL742 Linear Array
L746 Linear ArrayL741 Linear Array
10I2 Linear ArrayL746 Linear Array
9L-A Linear Array10I2 Linear Array
12L-B Linear Array9L-A Linear Array
13L-A Linear Array12L-B Linear Array
18L-A Linear Array13L-A Linear Array
10L-I Linear Array18L-A Linear Array
12LT-A Linear Array10L-I Linear Array
12LI-A Linear Array12LT-A Linear Array
12LI-A Linear Array
6V1 Micro-curved Array6V1 Micro-curved Array
6V3 Micro-curved Array6V3 Micro-curved Array
6V3A Micro-curved Array6V3A Micro-curved Array
6V7 Micro-curved Array6V7 Micro-curved Array
EC9-5 Micro-curved ArrayEC9-5 Micro-curved Array
BCC9-5 Micro-curved ArrayBCC9-5 Micro-curved Array
BCL10-5 Biplane (Micro-curved +Linear Array)
CWD2.0 CWPWD2.0 Doppler
CWD5.0 CWCWD2.0 CWSE
MPTEE Phased ArrayMPTEE Phased ArrayAnalysis 1
MPTEE mini Phased ArrayMPTEE mini Phased Array
LAP7 Linear Array

Table 3 a) Probes Comparison

{41}------------------------------------------------

{42}------------------------------------------------

SE Analysis 1:

Compared with the predicate device, there is new CW probes (CWD5.0), which is similar with the probe D5cwc cleared with predicate device EPIQ 5/EPIQ 7 Diagnostic Ultrasound System (K160807). The clinical application is the same, the performance or frequency is similar, and the difference of these doesn't affect the safety, effectiveness and clinical use.

The engineering drawings of the additional transducers/probes (CWD5.0), and the further comparison are provided as follows.

Comparison ItemsSubject DeviceSonoScape X5/X5 Exp/X5Pro/X6 /X6Exp/X6 ProPredicate DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K160807)Remark
Probe1CWD5.0D5cwc/
PhotoImage: ProbeImage: Probe/
Probe TypeDopplerDopplerSame
Frequency5.0MHz5.0MHzSame
Indicationfor usePeripheral vessel, CerebralVascularPeripheral vessel,CerebralVascularSame
OperationModeCWCWSame
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA:720mW/cm2 maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
Comparison ItemsSubject DeviceSonoScape X5/X5 Exp/X5Pro/X6 /X6Exp/X6 ProPredicate DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K160807)Remark
Probe2C322C6-2/
PhotoImage: SonoScape X5/X5 Exp/X5 Pro/X6 /X6Exp/X6 Pro ProbeImage: EPIQ 5/EPIQ 7 Diagnostic Ultrasound System Probe/
ProbeTypeMicro-curved ArrayCurved ArraySame
Frequency2.0-7.0MHz2.0-6.0MHzSEAnalysis 2
Indicationfor useFetal,Abdominal,Other(Ob/GYN)General purpose abdominal(adult and pediatric, includingvascular), bowel, obstetrical,gynecological, prostate andinterventional applicationsSame
OperationModeB,THI,M,CFM,PDI,PWB,THI,M,CFM,PDI,PWSame
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging,THI,Contrast ImagingCompound Imaging,CPA,harmonic Imaging,ContrastImaging,3D/4DImaging,XRES, ElastographySEAnalysis 3
Comparison ItemsSubject DeviceSonoScape X5/X5 Exp/X5Pro/X6 /X6Exp/X6 ProPredicate DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K160807)Remark
Probe33C-AC6-2/
PhotoImage: SonoScape X5/X5 Exp/X5 Pro/X6 /X6Exp/X6 Pro probeImage: EPIQ 5/EPIQ 7 Diagnostic Ultrasound System probe/
ProbeTypeCurved ArrayCurved ArraySame
Frequency1.0-7.0MHz2.0-6.0MHzSEAnalysis 2
Indicationfor useFetal,Abdominal,Other(Ob/GYN)General purpose abdominal(adult and pediatric, includingvascular), bowel, obstetrical,gynecological, prostate andinterventional applicationsSame
OperationModeB,THI,M,CFM,PDI,PWB,THI,M,CFM,PDI,PWSame
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging,THI,Contrast ImagingCompound Imaging, CPA,harmonic Imaging,ContrastImaging,3D/4DImaging,XRES, ElastographySEAnalysis 3

Table 3 b) Further Comparison for CWD5.0 Probes

Therefore, they can be considered Substantially Equivalent in safety and effectiveness, and no new risk is raised, so the SE is not affected.

Functional Comparison of probes

Compared with the predicate device (K173058), there is new function of contrast

{43}------------------------------------------------

imaging in probes (C322, 3C-A and C1-6), which is similar with the probe (C6-2 and C5-1) cleared with predicate device EPIQ 5/EPIQ 7 Diagnostic Ultrasound System (K160807). The clinical application is the same, the performance or frequency is similar, and the difference of these doesn't affect the safety, effectiveness and clinical use.

The engineering drawings of the new function transducers(probes) (C322, 3C-A, C1-6), and the further comparison are provided as follows.

Table 3 c) Further Comparison for C322 Probes

{44}------------------------------------------------

Table 3 d) Further Comparison for 3C-A Probes

Table 3 e) Further Comparison for C1-6 Probes

Comparison ItemsSubject DeviceSonoScape X5/X5 Exp/X5Pro/X6 /X6Exp/X6 ProPredicate DeviceRemark
Legally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System

{45}------------------------------------------------

(K160807)
Probe4C1-6C5-1/
PhotoImage: C1-6 probeImage: C5-1 probe/
ProbeTypeMicro-curved ArrayCurved ArraySame
Frequency1.0-8.0MHz1.0-5.0MHzSEAnalysis 2
Indicationfor useFetal, Abdominal,Other(Ob/GYN)General purpose abdominal(adult and pediatric, includingvascular), bowel, obstetrical,gynecological, prostate andinterventional applicationsSame
OperationModeB,THI,M,CFM,PDI,PWB,THI,M,CFM,PDI,PWSame
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, THI,Contrast ImagingCompound Imaging, CPA,harmonic Imaging, ContrastImaging, 3D/4D Imaging,XRES, PercuNavSEAnalysis 3

SE Analysis 2:

Although the frequency is different, the parameters of the subject device are included in the predicated device, and no new risk is raised.

SE Analysis 3:

Although the functions is different, the functions of the subject device are included in the predicated device, and no new risk is raised.

Moreover, compared with predicate device, the subject device (X5 Series) complies with the same regulation and safety standards and has the consistent acoustic output levels.

Summary of the comparison

{46}------------------------------------------------

Compared with the the predicate device legally marketed SonoScape X5 (K160258), the additional transducers, additional functions and updated intended use of the subjective device (X5 Series Digital Color Doppler Ultrasound System) are all analyzed with the predicate device P10 Series (K173058) and EPIQ 5/EPIQ (K160807). The comparison showed that they can be considered Substantially Equivalent in safety and effectiveness. Therefore, there is no new risk raised, and the SE is not affected.

7. Non-Clinical Tests [21 CFR 807.92(b) (1)]

Non-clinical testing to assure compliance with electrical, mechanical, thermal and electromagnetic compatibility safety, acoustic output and biocompatibility were performed and have been found to conform to applicable standards. The X5 Series system has been designed and manufactured to meet the following standards:

IEC 60601-1:2005+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance [08/20/2012];

IEC 60601-1-2:2007, Medical Electrical Equipment -Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests [03/30/2007];

IEC 60601-2-37:2007, Medical Electrical Equipment-Part 2-37: Particular requirements for basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment [08/09/2007];

ISO 10993-5:2009, Biological Evaluation of Medical Devices, Part 5-Tests for in vitro cytotoxicity [06/01/2009];

ISO 10993-10:2010, Biological Evaluation of Medical Devices- Part 10: Tests for irritation and skin sensitization [08/01/2010];

AIUM/NEMA UD 2:2004 (R2009), Acoustic output measurement standard for diagnostic ultrasound equipment [08/21/2009]; and

AIUM/NEMA UD 3:2004 (R2009), Standard for real-time display of thermal and mechanical acoustic output indices on diagnostic ultrasound equipment [08/13/2010].

The new added probes and software feature are also verified through system tests and did not raise any concerns regarding safety or effectiveness. The relevant summarized information are as followed:

Performance testTesting protocols andfail/acceptance criteriaTesting results
Electrical safety testingIEC 60601-1:2005+A1:2012Passed

{47}------------------------------------------------

EMC testingIEC 60601-1-2:2007Passed
Acoustic testingIEC 60601-2-37:2007AIUM/NEMA UD 2:2004 (R2009)Passed
Software Verification andValidationIEC 62304:2006 +A 1:2015Passed

Laboratory tests (including Phantom tests) were conducted to verify that the X5 Series system met all design specifications and the X5 Series system conformed to applicable medical device standards.

8. Clinical Test [21 CFR 807.92(b) (2)]

No clinical testing was required.

9. Substantially Equivalent Conclusions [21 CFR 807.92(b) (3)]

In accordance with the 21 CFR Part 807 and based on the information provided in this premarket notification, SONOSCAPE MEDICAL CORP. concludes that X5 Series Digital Color Doppler Ultrasound System is substantially equivalent to the predicate device with regard to safety and effectiveness.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.