AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetabl/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transvaginal.

The clinical environments where the EPIQ 5, EPIQ 7, Affiniti 50 Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical point-of-care for diagnosis of patients.

Device Description

The proposed EPIO and Affiniti Diagnostic Ultrasound Systems, which includes EPIO 5. EPIQ 7, Affiniti 50 and Affiniti 70 systems, are general purpose, software controlled, diagnostic ultrasound systems. Their function is to acquire ultrasound data and to display the data in various modes of operation.

The devices consist of two parts: the system console and the transducers. The system console contains the user interface, a display, system electronics and optional peripherals (ECG, printers). In addition to the physical knobs and buttons of the main control panel, the user interface consists of a touch screen with soft key controls. EPIO also has a QWERTY keyboard.

The removable transducers are connected to the system using a standard technology, multipin connectors. The proposed EPIQ and Affiniti systems use standard transducer technology, and supports phased, linear, curved linear array, TEE, motorized 3D curved linear arrays as well as non-imaging (pencil) probes.

Clinical data storage consists of a local repository as well as off-line image storage via the network, DVR, DVD, and USB storage devices. The images are stored in industry-standard formats (e.g. JPEG. AVI, DICOM) and are intended to be readable using industry-standard hardware and software. On-line review of the images is available. Secure access tools are provided to restrict and log access to the clinical data repository according to HIPAA.

The system circuitry generates an electronic voltage pulse, which is transmitted to the transducer. In the transducer, a piezo electric array converts the electronic pulse into an ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body tissues. The Doppler functions of the system process the Doppler shift frequencies from the echoes of moving targets such as blood to detect and graphically display the Doppler shifts of these tissues as flow.

The proposed EPIQ and Affiniti systems give the operator the ability to measure anatomical structures and offer analysis packages that provide information used by competent healthcare professionals to make a diagnosis. The proposed EPIQ and Affiniti systems enable image guided navigation and image fusion via the optional PercuNav feature (K121498).

AI/ML Overview

The document describes the Philips EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems, and Affiniti 50 and Affiniti 70 Diagnostic Ultrasound Systems. It primarily details their indications for use and compares their technological characteristics to a previously cleared predicate device (Philips EPIQ Diagnostic Ultrasound System K132304).

Based on the provided text, a "study that proves the device meets the acceptance criteria" in terms of clinical performance or specific statistical metrics is not explicitly described. The document explicitly states: "Clinical data was not required to demonstrate safety and effectiveness of the proposed EPIQ and Affiniti Diagnostic Ultrasound Systems since the proposed EPIQ or Affiniti system introduces no new indications for use, modes or features that have not been previously cleared with the predicate device EPIQ system (K132304). The clinical safety and effectiveness of ultrasound systems with these characteristics are well accepted for both predicate and subject devices."

Therefore, the acceptance criteria are demonstrated through substantial equivalence to a predicate device, and compliance with recognized safety and performance standards, rather than a de novo clinical study for this specific submission.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria (from Standards) and Reported Device Performance (Compliance Statement)

Acceptance Criteria (from Standards)Reported Device Performance
Acoustic Output Limits:
Ispta.3 ≤ 720 MW/cm2Ispta.3 ≤ 720 MW/cm2 (Compliant)
MI < 1.9MI < 1.9 (Compliant)
TI < 6.0TI < 6.0 (Compliant)
Safety and Performance Standards:
IEC 60601-2-37 Ed 2.0 (Acoustic Output Display)Complies with IEC 60601-2-37 Ed 2.0
IEC 62359, Ed 2.0 (Thermal and Mechanical Indices)Complies with IEC 62359, Ed 2.0
FDA Ultrasound Specific Guidance (Sept 9, 2008)Complies with FDA ultrasound specific guidance
IEC 60601-1 (Basic Safety and Essential Performance)Compliant to IEC 60601-1:2005 + A1:2012
IEC 60601-1-2 (EMC)Compliant to IEC 60601-1-2:2007
IEC 60601-1-6 (Usability)Compliant to IEC 60601-1-6:2010
ISO 10993 (Biological Evaluation of Medical Devices)Compliant to ISO 10993
Quality Assurance (Risk Analysis, Product Specs, Design Reviews, V&V)Applied to system design and development

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: Not applicable. No new clinical test set data from human subjects was used for this submission to demonstrate safety and effectiveness.
  • Data Provenance: Not applicable for a new clinical test set. The submission relies on the established safety and effectiveness of the predicate device and compliance with recognized standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

  • Not applicable. As no new clinical study was required, there was no independent ground truth labeling process with experts described for a test set. The predicate device's clinical safety and effectiveness are considered "well accepted."

4. Adjudication Method for the Test Set:

  • Not applicable. There was no new clinical test set requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, ... and its effect size:

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done as this submission states "Clinical data was not required". The device is considered substantially equivalent to a predicate.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable in the context of device performance claims. The devices are general-purpose diagnostic ultrasound systems used by human operators. The document mentions "Integration of OLAB Heart Model software" and "PercuNav feature" as integrated software components, but their standalone performance is not detailed in this summary.

7. The Type of Ground Truth Used:

  • Established Equivalence to Predicate Device: The primary "ground truth" or basis for acceptance is the substantial equivalence to the Philips EPIQ Diagnostic Ultrasound System (K132304), whose safety and effectiveness are considered "well accepted."
  • Compliance with Recognized Standards: Compliance with various IEC and ISO standards for acoustic output, electrical safety, EMC, usability, and biological evaluation also serves as a "ground truth" for non-clinical performance aspects.

8. The Sample Size for the Training Set:

  • Not applicable. This submission describes hardware and integrated software for an ultrasound system, not a machine learning algorithm requiring a separate training set. The software components mentioned (OLAB Heart Model, PercuNav) were previously cleared, implying their training and validation were addressed in prior submissions.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable for this submission directly, as it does not detail the training of a new algorithm. For historical or integrated software components, the ground truth would have been established during their respective development and clearance processes, but this is not specified here.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all capital letters.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 6, 2016

Philips Ultrasound, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K160807 Trade/Device Name: EPIO 5 Diagnostic Ultrasound System EPIQ 7 Diagnostic Ultrasound System Affiniti 50 Diagnostic Ultrasound System Affiniti 70 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: March 22, 2016 Received: March 24, 2016

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D.'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K160807

Device Name

EPIQ 5 Diagnostic Ultrasound System, EPIQ 7 Diagnostic Ultrasound System, Affiniti 50 Diagnostic Ultrasound System, Affiniti 70 Diagnostic Ultrasound System

Indications for Use (Describe)

Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric,

Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetabl/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transvaginal.

The clinical environments where the EPIQ 5, EPIQ 7, Affiniti 50 Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical point-of-care for diagnosis of patients.

Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect

and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

510(k) No: _

EPIQ 5 and EPIQ 7 Ultrasound Systems System:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13,17
AbdominalPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,15,16,17,
Intra-operative (Cardiac)PPPPPNote: 1,2,3Note: 5,8,9,10,11,12,13
Intra-operative (Vascular)PPPPNote: 1,2,3Note: 5,8,9,10,11,12,13
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,15,17
Neonatal CephalicPPPPNote: 1,2,3Note: 5,8,9,10,11,12,13,17
Adult CephalicPPPPNote: 1,2,3Note: 8,9,10,11,12,13,17
Trans-rectalPPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,15,17
Trans-vaginalPPPPNote: 1,2,3Note:5,6,7,8,9,10,11,12,13,15,17
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note:5,6,7,8,9,10,11,12,13,15,17
Other: UrologyPPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,15,17
CardiacCardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Cardiac PediatricPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Other (Fetal Echo)PPPPPNote: 1,2,3,4Note: 5,6,8,10,12,13,14
PeripheralPeripheral vesselPPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
VesselCerebral vascularPPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic Imaging
Note 9: Color Power Angio (CPA)
Previous submission: K132304, EPIQ

{4}------------------------------------------------

510(k) No: __

3D9-3v

EPIQ 5 and EPIQ 7 Ultrasound Systems

Transducer:

System:

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
BMPWDCWDColorDoppler**Combined(Spec.)Other(Spec.)
General(Track I only)Specific(Tracks I & III)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note:5,6,8,10,11,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote: 1,2,3Note: 5,6,7,8,10,11,12,13, 15
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNP
Other: UrologyPPPP
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication; P= previously cleared (; E= previously cleared by earlier letter to file

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic Imaging
Note 9: Color Power Angio (CPA)
Previous Submission: K043535, HD 11

{5}------------------------------------------------

510(k) No:

System: EPIQ 5 and EPIQ 7 Ultrasound Systems

C5-1

Transducer:

Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13
AbdominalPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,16,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note: 5,6,7,8, 910,11,12,13,17
Other: UrologyPPPPNote: 1,2,3Note: 5,6,,8, 9,10,11,12,13,17
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,6,10,12,13
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Cerebral vascular
N= new indication P= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 8: Panoramic ImagingNote 16: ElastPQ (for Liver)Note 17: PercuNav
Previous submission: K132304, EPIQ
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Small Organ (breast, thyroid,testicle)
Neonatal CephalicPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
VesselCerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13

{6}------------------------------------------------

510(k) No: EPIQ 5 and EPIQ 7 Ultrasound Systems System: Transducer: C8-5

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304, EPIΩ

{7}------------------------------------------------

510(k) No: _

System: EPIQ 5 and EPIQ 7 Ultrasound Systems

Transducer: C9-2

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OBPPPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13,17
AbdominalPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
PediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)NNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Other: UrologyPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,6,10,12,13
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Cerebral Vascular
N= new indicationP= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous submission: K132304, EPIQ
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmicPPPPNote: 1,2,3Note:5,6,7,8,9,10,12,13,17
Fetal Imaging& OtherFetal / OBPPPPNote: 1,2,3Note:5,6,7,8,9,10,11,12,13, 15,17
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote: 1,2,3Note:5,6,7,8,9,10,11,12,13, 15,17
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note:5,6,7,8,9,10,11,12,13,15,17
Other: UrologyPPPPNote: 1,2,3Note:5,6,7,8,9,10,11,12,13,15,17
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,6,8,10,12,13
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indication P= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304, EPIQ
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificColorCombined*Other
(Track I only)(Tracks I & III)BMPWDCWDDoppler*(Spec.)(Spec.)
OphthalmicOphthalmicP
Fetal / OBPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13,17
Trans-vaginalPPPPNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13,17
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note:5,6,7,8,9,10,11,12,13,17
Other: UrologyPPPPNote: 1,2,3Note:5,6,7,8,9,10,11,12,13,17
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,8,9,10,12,13
PeripheralPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
VesselCerebral Vascular
N= new indicationP= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304, EPIQ
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultP
CardiacCardiac PediatricP
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWD
Note 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304. EPIQ

{8}------------------------------------------------

510(k) No: _

System: EPIQ 5 and EPIQ 7 Ultrasound Systems

Transducer: C10-3v

{9}------------------------------------------------

510(k) No:

System: EPIQ 5 and EPIQ 7 Ultrasound Systems C10-4ec

Transducer: Intended Use:

{10}------------------------------------------------

510(k) No: _

EPIQ 5 and EPIQ 7 Ultrasound Systems System:

Transducer: D2cwc

{11}------------------------------------------------

510(k) No: _

System: EPIQ 5 and EPIQ 7 Ultrasound Systems

Transducer: D2tcd

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral VascularP
N= new indication P= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304, EPIQ

{12}------------------------------------------------

510(k) No: _

EPIQ 5 and EPIQ 7 Ultrasound Systems System:

Transducer: D5cwc

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselP
Cerebral VascularP
N= new indication P= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude Doppler
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 9: Color Power Angio (CPA)
Note 2: Combined modes include: B+M+ColorNote 10: Harmonic Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 11: Contrast Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 12: 3D/4D Imaging
Note 5: SonoCTNote 13: XRES
Note 6: Imaging for guidance of biopsyNote 14: TDI
Note 7: Infertility monitoring of follicle developmentNote 15: Elastography
Note 8: Panoramic ImagingNote 16: ElastPQ (for Liver)
Note 17: PercuNav
Previous Submission: K132304, EPIQ
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Neonatal CephalicPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Cerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
N= new indication P= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;Note 10: Harmonic Imaging
B+M
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;Note 12: 3D/4D ImagingB+Amplitude+PWD
Note 4: Combined modes include: B+CWD; B+Color+CWD;
B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13
AbdominalPPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,15,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 5,6,8,10,11,12,13,17
Small Organ (breast, thyroid, testicle)PPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,15,17
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,17
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,17
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,17
Cerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 6: Imaging for guidance of biopsyNote 14: TDINote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304. EPIQ
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)PPPPNote: 1,2,3Note: 5,8,9,10,11,12,13
Intra-operative (Vascular)PPPPNote: 1,2,3Note: 5,8,9,10,11,12,13
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 5,8,9,10,11,12,13
Small Organ (breast, thyroid, testicle)PPPPNote: 1,2,3Note: 5,8,10,11,12,13
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,8,9,10,12,13
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,8,9,10,12,13
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note: 5,8,9,10,11,12,13
Cerebral VascularPPPPNote: 1,2,3Note: 5,8,9,10,11,12,13
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304, EPIQ

{13}------------------------------------------------

510(k) No: System: EPIQ 5 and EPIQ 7 Ultrasound Systems Transducer: L12-3

{14}------------------------------------------------

510(k) No: _

System: EPIQ 5 and EPIQ 7 Ultrasound Systems

Transducer: L12-5 50

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

{15}------------------------------------------------

510(k) No: _

EPIQ 5 and EPIQ 7 Ultrasound Systems System:

Transducer: L15-7io

{16}------------------------------------------------

510(k) No: _

System: EPIQ 5 and EPIQ 7 Ultrasound Systems

Transducer: L18-5

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
AbdominalPPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,15,17
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Cerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304. EPIQ

{17}------------------------------------------------

510(k) No: _

EPIQ 5 and EPIQ 7 Ultrasound Systems System:

Transducer: S5-1

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 6,8,9,10,12,13, 17
AbdominalPPPPNote: 1,2,3Note: 6,8,9,10,11,12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 6,8,9,10,11,12,13,17
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult CephalicPPPPNote: 1,2,3Note: 8,9,10,11,12,13, 17
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14, 17
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,11,12,13,14, 17
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPPNote: 1,2,3,4Note: 10,13,14
PeripheralPeripheral vesselPPPPNote: 1,2,3Note: 6,8,9,10,11,12,13,17
VesselCerebral Vascular
N= new indication P= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304, EPIQ

{18}------------------------------------------------

510(k) No: _

System: EPIQ 5 and EPIQ 7 Ultrasound Systems

Transducer: S7-3t

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultNNNNNNote: 1,2,3,4Note: 10,11,12,13,14
Cardiac PediatricNNNNNNote: 1,2,3,4Note: 10,11,12,13,14
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indication P= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304,EPIQ

{19}------------------------------------------------

510(k) No: _

EPIQ 5 and EPIQ 7 Ultrasound Systems System:

Transducer: ട്ട-3

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OBPPPPNote: 1,2,3Note: 7,8,10,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
PediatricPPPPNote: 1,2,3Note: 7,8,10,11,12,13
Small Organ (breast, thyroid,testicle)
Neonatal CephalicPPPPNote: 1,2,3Note: 8,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Cardiac PediatricPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indication P= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav

{20}------------------------------------------------

510(k) No: _

System: EPIQ 5 and EPIQ 7 Ultrasound Systems

Transducer: S12-4

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 8,10,11,12,13
Small Organ (breast, thyroid,testicle)
Neonatal CephalicPPPPNote: 1,2,3Note: 8,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 11,12,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 11,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indication P= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304, EPIQ

{21}------------------------------------------------

510(k) No: J

System: EPIQ 5 and EPIQ 7 Ultrasound Systems V6-2

Transducer:

Intended Use:

Mode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
AbdominalPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,9,10,11,13
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTB+Amplitude+CWDNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304, EPIQ

{22}------------------------------------------------

System: Transducer: Intended Use: EPIQ 5 and EPIQ 7 Ultrasound Systems VL13-5

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler**Combined(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBNNNNNNote: 1,2,3Note: 5,6,8,10,11,12,13
AbdominalNNNNNNote: 1,2,3Note: 5,6,8,10,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNNote: 1,2,3Note: 5,6,8,10,11,12,13
Small Organ (breast, thyroid,testicle)NNNNNNote: 1,2,3Note: 5,6,8,10,11,12,13
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)NNNNNNote: 1,2,3Note: 5,6,8,10,11,12,13
Musculo-skel. (Superficial)NNNNNNote: 1,2,3Note: 5,6,8,10,11,12,13
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselNNNNNNote: 1,2,3Note: 5,6,8,10,11,12,13
VesselCerebral VascularNNNNNNote: 1,2,3Note: 5,6,8,10,11,12,13

N= new indication; P= previously cleared

Additional Comments:

  • New

**Color Doppler includes Color Amplitude Doppler

Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+M;

Note 2: Combined modes include: B+M+Color

Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWD

  • Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWD
    Note 5: SonoCT

Note 6: Imaging for guidance of biopsy

  • Note 7: Infertility monitoring of follicle development
  • Note 8: Extended Field of View (EFOV), AKA Panoramic Imaging includes SonoCT imaging
  • Note 9: EFOV including Amplitude Doppler

Note 10: Harmonic Imaging

Note 11: Contrast Imaging

  • Note 12: 3D Imaging
    Note 13: XRES

Note 14: TDI

N= new indication P= previously cleared Additional Comments:

Note 15: Elastography

{23}------------------------------------------------

510(k) No: _

System:

X5-1

Transducer:

Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,2,3Note: 8,9,10,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 8,9,10,11,12,13
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult CephalicPPPPNote: 1,2,3Note: 8,9,10,11,12,13
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indication P= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304. EPIO

{24}------------------------------------------------

510(k) No: J

System: EPIQ 7 Ultrasound System

X6-1

Transducer:

Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 6,8,9,10,12,13
AbdominalPPPPNote: 1,2,3Note: 6,8,9,10,11,12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note:6,8,9,10,11,12,13,17
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note: 6,8,9,10,11,12,13
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note: 6,8,9,10,11,12,13,17
Cerebral VascularPPPPNote: 1,2,3Note: 6,8,9,10,11,12,13,17
N= new indication      P= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304, EPIQ

{25}------------------------------------------------

510(k) No: _

System: EPIQ 7 Ultrasound System Transducer: X7-2 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler**Combined(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OBNNNNNote: 1,2,3Note: 6,7,8,10,11,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
PediatricNNNNNote: 1,2,3Note: 6,7,8,10,11,12,13
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultNNNNNNote: 1,2,3,4Note: 10,11,13,14
Cardiac PediatricNNNNNNote: 1,2,3,4Note: 10,11,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)NNNNNNote: 1,2,3,4Note: 10,11,13,14
PeripheralVesselPeripheral vesselNote: 1,2,3
Cerebral Vascular

N= new indication; P= previously cleared

Additional Comments:

  • New
  • **Color Doppler includes Color Amplitude Doppler
  • Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+M;
  • Note 2: Combined modes include: B+M+Color
  • Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWD
  • Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWD
  • Note 5: SonoCT
  • Note 6: Imaging for guidance of biopsy
  • Note 7: Infertility monitoring of follicle development
  • Note 8: Extended Field of View (EFOV), AKA Panoramic Imaging includes SonoCT imaging
  • Note 9: EFOV including Amplitude Doppler
  • Note 10: Harmonic Imaging
  • Note 11: Contrast Imaging
  • Note 12: 3D Imaging
  • Note 13: XRES
  • Note 14: TDI
  • Note 15: Elastography

{26}------------------------------------------------

510(k) No:

System: EPIQ 5 and EPIQ 7 Ultrasound Systems

Transducer:

X7-2t Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indication P= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude;B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD;B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav

Previous Submission: K132304, EPIQ

{27}------------------------------------------------

System: Affiniti 70 and Affiniti 50 Ultrasound Systems

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBNNNNNote: 1,2,3Note: 5,6,7,8,9,10,12,13,17
AbdominalNNNNNote: 1,2,3Note: 5,6, 8,9,10,11,12,13,15,16,17
Intra-operative (Cardiac)NNNNNote: 1,2,3Note: 5,8,10,11,12,13
Intra-operative (Vascular)NNNNNote: 1,2,3Note: 5,8,10,11,12,13
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Small Organ (breast, thyroid,testicle)NNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13,15,17
Neonatal CephalicNNNNNote: 1,2,3Note: 5,8,9,10,11,12,13
Adult CephalicNNNNNote: 1,2,3Note: 8,10,11,12,13
Trans-rectalNNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Trans-vaginalNNNNNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13,15,17
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)NNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Musculo-skel. (Superficial)NNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Intra-luminal
Other: GYNNNNNNote: 5,6,7,8,9,10,11,12,13, 15,17
Other: UrologyNNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13, 15,17
Cardiac AdultNNNNNNote: 1,2,3,4Note: 10,11,12,13,14
CardiacCardiac PediatricNNNNNNote: 1,2,3,4Note: 10,11,12,13,14
Trans-esophageal (Cardiac)NNNNNNote: 1,2,3,4Note: 10,11,12,13,14
Other (Fetal Echo)NNNNNNote: 1,2,3,4Note: 5,6,8,10,12,13,14
PeripheralVesselPeripheral vesselNNNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Cerebral VascularNNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17 PercuNav

{28}------------------------------------------------

System:Affiniti 70 Ultrasound System
----------------------------------------

C5-1

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OBPMPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13,17
AbdominalPPPPNote: 1,2,3Note: 5,6, 8,9,10,11,12,13,15,16,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPMPPNote: 1,2,3Note: 5,6,7,8, 9,10,11,12,13,17
Other: UrologyPMPPNote: 1,2,3Note: 5,6,8, 9,10,11,12,13,17
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,6,10,12,13
PeripheralPeripheral Vessel
VesselCerebral Vascular

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote:17 PercuNav
Previous submission: K132304, EPIQ

{29}------------------------------------------------

Intended Use: Diagnostic ultrasound Imaging or fluid flow analysis of the human body as follows:Mode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBNNNNNote: 1,2,3Note: 5,6,7,8,9,10,12,13
AbdominalNNNNNote: 1,2,3Note: 5,6, 8,9,10,11,12,13,15
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNNNNNNote: 1,2,3Note: 5,6,7,8, 9,10,11,12,13
Other: UrologyNNNNNote: 1,2,3Note: 5,6,8, 9,10,11,12,13
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)NNNNNote: 1,2,3Note: 5,6,10,12,13
PeripheralVesselPeripheral Vessel
Cerebral Vascular

Affiniti 70 and Affiniti 50 Ultrasound Systems System:

C6-2

Transducer:

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging

{30}------------------------------------------------

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,2,3Note: 5,6, 8,9,,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 5,6,8,9,11,12,13
Small Organ (breast, thyroid,testicle)
Neonatal CephalicPPPPNote: 1,2,3Note: 5,6,8,9,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,11,12,13
VesselCerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,11,12,13

Affiniti 70 and Affiniti 50 Ultrasound Systems System:

C8-5

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

E = added via Appendix E

N= new indication

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous Submission: K132304, EPIQ

{31}------------------------------------------------

Affiniti 70 Ultrasound System System:

Transducer:

C9-2 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
AbdominalPPPPNote: 1,2,3Note: 5,6, 8,9,10,11,12,13.17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Small Organ (breast, thyroid,
testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5.6,10,12,13
PeripheralPeripheral vessel
VesselCerebral Vascular

E = added via Appendix E

N= new indication

P= previously cleared (K132304)

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Previous Submission: K132304, EPIQ

{32}------------------------------------------------

Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote: 1,2,3Note: 5,6,7, 8,10,11,12,13, 15
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13,15
Other: UrologyPPPPNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13,15
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,6,8,10,12,13
PeripheralVesselPeripheral vessel
Cerebral Vascular

Affiniti 70 Ultrasound System System: Transducer: C10-3v

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous Submission: K132304, EPIQ

{33}------------------------------------------------

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote: 1,2,3Note: 5,6,7, 8,10,11,12,13
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13
Other: UrologyPPPPNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,6,8,10,12,13
PeripheralPeripheral vessel
VesselCerebral Vascular

Affiniti 70 and Affiniti 50 Ultrasound Systems System:

3D9-3v

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous Submission: K043535, HD 11

{34}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote: 1,2,3Note: 5,6,7,8,10,11,12,13, 15
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13,15
Other: UrologyNNNNNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13,15
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)NNNNNote: 1,2,3Note: 5,6,8,10,12,13
PeripheralVesselPeripheral vessel
Cerebral Vascular

System: Affiniti 70 and Affiniti 50 Ultrasound Systems

C9-4v

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K120321, ClearVue 350/550

{35}------------------------------------------------

Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote: 1,2,3Note: 5,6,7, 8,9,10,11,12,13,17
Trans-vaginalPPPPNote: 1,2,3Note: 5,6,7, 8,9,10,11,12,13,17
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: UrologyPPPPNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13,17
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,6,8,10,11,12,13
PeripheralVesselPeripheral Vessel
Cerebral Vascular

Affiniti 70 and Affiniti 50 Ultrasound Systems System: C10-4ec

Transducer:

E = added via Appendix E N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote:17 PercuNav
Previous submission: K132304, EPIQ

{36}------------------------------------------------

Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: UrologyNNNNNote: 5,6,8,9,10,11,12,13
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vessel
VesselCerebral Vascular

Affiniti 70 and Affiniti 50 Ultrasound Systems System: Transducer: BP10-5ec

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K081661, HD 15

{37}------------------------------------------------

Intended Use:Clinical ApplicationDiagnostic ultrasound imaging or fluid flow analysis of the human body as follows:Mode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
Ophthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultP
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

System: Affiniti 70 and Affiniti 50 Ultrasound Systems

Transducer: Intended Use:

D2cwc
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K132304, EPIQ

{38}------------------------------------------------

Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

System: Affiniti 70 and Affiniti 50 Ultrasound Systems

D2tcd

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K132304, EPIQ

{39}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselP
Cerebral VascularP

Affiniti 70 and Affiniti 50 Ultrasound Systems System:

D5cwc

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K132304, EPIQ

{40}------------------------------------------------

System: Affiniti 70 Ultrasound System

L12-3

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OBNNNNNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13
AbdominalPPPPNote: 1,2,3Note: 5,6, 8,9,10,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
PediatricNNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Small Organ (breast, thyroid, testicle)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Neonatal CephalicPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Cerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K132304, EPIQ

{41}------------------------------------------------

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBNNNNNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13
AbdominalNNNNNote: 1,2,3Note: 5,6, 8,9,10,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Neonatal CephalicPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Cerebral VascularNNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13

System: Affiniti 70 and Affiniti 50 Ultrasound Systems

L12-4

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K120321, ClearVue 350/550

{42}------------------------------------------------

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13,17
AbdominalPPPPNote: 1,2,3Note: 5,6, 8,9,10,11,12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,15,17
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
VesselCerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17

System: Affiniti 70 and Affiniti 50 Ultrasound Systems

Transducer:

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote:17 PercuNav
Previous submission: K132304, EPIQ

{43}------------------------------------------------

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
AbdominalNNNNNote: 1,2,3Note: 5,6, 8,9,10,11,12,13
Intra-operative (Cardiac)PPPPNote: 1,2,3Note: 5,8,9,10,11,12,13
Intra-operative (Vascular)PPPPNote: 1,2,3Note: 5,8,9,10,11,12,13
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10,12,13
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10,12,13
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
VesselCerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13

Affiniti 70 and Affiniti 50 Ultrasound Systems System: L15-7io

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

E = added via appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K132304, EPIQ

{44}------------------------------------------------

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
Intended Use:General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OB
AbdominalPPPPNote: 1,2,3Note: 5,6, 8,9,10,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
PediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note: 5,6,8,10,11,12,13
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
VesselCerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13

System: Affiniti 70 and Affiniti 50 Ultrasound Systems

L18-5

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

E = added via Appendix E

N= new indication

P= previously cleared (K132304)

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K132304, EPIQ

{45}------------------------------------------------

System:Affiniti 70 Ultrasound System
----------------------------------------

S5-1

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,2,3Note: 5,6, 8,9,10,11,12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,
testicle)
Neonatal Cephalic
Adult CephalicPPPPNote: 1,2,3Note: 8,9,10,11,12,13,17
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Cardiac PediatricPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vessel
VesselCerebral Vascular

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote:17 PercuNav
Previous submission: K132304, EPIQ

{46}------------------------------------------------

System: Affiniti 70 and Affiniti 50 Ultrasound Systems

Transducer: S4-2

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,2,3Note: 5,6, 8,9,10,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,
testicle)
Neonatal Cephalic
Adult CephalicPPPPNote: 1,2,3Note: 8,9,10,11,12,13
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vessel
VesselCerebral Vascular

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES2.0
Note 5: SonoCTNote 14: TDI3.0
Note 6: Imaging for guidance of biopsyNote 15: Elastography4.0
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)5.0
Note 8: Panoramic Imaging6.0
Previous submission: K043535, HD11

{47}------------------------------------------------

System: Affiniti 70 and Affiniti 50 Ultrasound Systems

S7-3t

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultNNNNNNote: 1,2,3,4Note: 10,11,12,13,14
Cardiac PediatricNNNNNNote: 1,2,3,4Note: 10,11,12,13,14
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K132304, EPIQ

{48}------------------------------------------------

Intended Use:Clinical ApplicationDiagnostic ultrasound imaging or fluid flow analysis of the human body as follows:Mode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
Ophthalmic
Fetal / OBPPPPNote: 1,2,3Note: 7,8,9,10,12,13
AbdominalNNNNNote: 1,2,3Note: 8,9,10,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3
Small Organ (breast, thyroid, testicle)
Neonatal CephalicPPPPNote: 1,2,3Note: 8,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Cardiac PediatricPPPPPNote: 1,2,3,4
Trans-esophageal (Cardiac)
Other (Fetal Echo)NNNNNNote: 1,2,3,4Note: 10,11,12,13
PeripheralVesselPeripheral vessel
Cerebral Vascular

System: Affiniti 70 and Affiniti 50 Ultrasound Systems

Transducer:
Intended Use:

S8-3
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K132304, EPIQ

{49}------------------------------------------------

Affiniti 70/50 Ultrasound System System: S12-4

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OB
AbdominalNNNNNote: 1,2,3Note: 8,9,10,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
PediatricPPPPNote: 1,2,3Note: 8,9,10,11,12,13
Small Organ (breast, thyroid, testicle)
Neonatal CephalicPPPPNote: 1,2,3Note: 8,9,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K132304, EPIQ

{50}------------------------------------------------

Intended Use:Clinical ApplicationDiagnostic ultrasound imaging or fluid flow analysis of the human body as follows:Mode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OBPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,9,10,11,13
PeripheralVesselPeripheral vessel
Cerebral Vascular

System: Affiniti 70 and Affiniti 50 Ultrasound Systems Transducer: V6-2

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

E = added via Appendix E

N= new indication

P= previously cleared (K132304)

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD;B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K132304, EPIQ

{51}------------------------------------------------

Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultNNNNNNote: 1,2,3,4Note: 10,11,12,13,14
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

Affiniti 70 and Affiniti 50 Ultrasound Systems System: X7-2t

Transducer:

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic Imaging
Previous submission: K132304, EPIQ

{52}------------------------------------------------

Intended Use:Clinical ApplicationDiagnostic ultrasound imaging or fluid flow analysis of the human body as follows:Mode of Operation (*includes simultaneous B-mode)
ColorCombined*
General(Track I only)Specific(Tracks I & III)BMPWDCWDDoppler*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)NNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselNNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13
Cerebral VascularNNNNNote: 1,2,3Note: 5,6,8,9,10,11,12,13

System: Affiniti 70 and Affiniti 50 Ultrasound Systems

E = added via Appendix E

N= new indication

P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 9: Color Power Angio (CPA)
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging (for LVO applications)
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic imaging

{53}------------------------------------------------

510(k) Summary of Safety and Effectiveness EPIQ 5, EPIQ 7, Affiniti 50 and Affiniti 70 Diagnostic Ultrasound Systems

This summary of safety and effectiveness information is submitted in accordance with 21CFR §807.92

1. Submitter's name, address, telephone number, contact person.

Philips Ultrasound, Inc. 22100 Bothell Everett Hwy Bothell, WA 98021-8431

Contact person: Peng Cui, Regulatory Affairs Specialist Email: peng.cui@philips.com Tel: (978) 659-7966 Fax: (978) 487-4450

Date prepared: March 21, 2016

2. Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:

Common/usual name:Diagnostic ultrasound system and transducers
Proprietary name:EPIQ 7 Diagnostic Ultrasound System
EPIQ 5 Diagnostic Ultrasound System
Affiniti 50 Diagnostic Ultrasound System
Affiniti 70 Diagnostic Ultrasound System

Within this notification,

  • the reference to EPIQ represents both the EPIQ 7 and the EPIQ 5,
  • the reference to Affiniti represents both the Affiniti 70 and the Affiniti 50. .

These devices are classified as follows:

Classification Name21 CFR SectionProduct Code
Ultrasonic Pulsed Doppler Imaging System892.1550IYN
Ultrasonic Pulsed Echo Imaging System892.1560IYO
Diagnostic Ultrasound Transducer892.1570ITX

As stated in 21 CFR, parts 892.1550, 892.1560, and 892.1570, each of these generic types of devices has been classified as Class II.

3. Substantially Equivalent Devices

Philips Ultrasound believes the proposed EPIO and Affiniti Ultrasound Systems are substantially equivalent to the following currently marketed devices:

Product510(k)
-----------------

{54}------------------------------------------------

Philips EPIQ Diagnostic Ultrasound System K132304

Below is a comparison to the predicate devices:

StandardFeatureEPIQK132304(Predicate Device)EPIQ 1.5(Proposed Device)Affiniti 1.5( Proposed Device)
Indication forUseAbdominal, Cardiac Adult,Cardiac other (Fetal),Cardiac Pediatric, CerebralVascular, Cephalic (Adult),Cephalic (Neonatal),Fetal/Obstetric,Gynecological,Intraoperative (Vascular),Intraoperative (Cardiac),Musculoskeletal(Conventional),Musculoskeletal(Superficial), Other:Urology, Pediatric,Peripheral Vessel, SmallOrgan (Breast, Thyroid,Testicle), Transesophageal(Cardiac), Transrectal,TransvaginalSameSame
TransducersC5-1 C8-5 C9-2 C10-3v C10-4ec D2cwc D5cwc L12-3 L12-5 50 L15-7io L18-5 S5-1 S7-3t S8-3 S12-4 V6-2 X7-2t D2tcd X5-1 X6-1Same with the followingadditional transducerswith no new indications VL13-5 X7-2 3D9-3v The new transducersintroduce neither newindications for use nornew patient contactmaterial.Same with the followingadditional transducers withno new indications S4-2 3D9-3v C9-4v C6-2 BP10-5ec L12-4 VL13-5 The new transducersintroduce neither newindications for use nor newpatient contact material.Affiniti does not include: X5-1 X6-1
TransducerPhased arraySameSame *
StandardFeatureEPIQK132304(Predicate Device)EPIQ 1.5(Proposed Device)Affiniti 1.5( Proposed Device)
TypesLinear array Curved array Pencil Probes Multi-plane Transesophageal Motorized arrays 3D Matrix Array*Affiniti does not have 3D Matrix Array transducers
TransducerFrequency1.0 – 18.0 MHzSameSame
AcousticOutput Display& FDA LimitsIEC 62359 ISPTA max=720 mw/cm2 MI max =1.9 MI display TI displaySameSame
Imaging Mode2D Echo Imaging M-mode Echo Imaging PW Doppler Imaging CW Doppler Imaging 2D Color Doppler Imaging Color Power Angio (CPA) Color Amplitude Doppler M Mode Color Doppler Imaging Broadband Imaging Harmonic Imaging 2D Tissue Doppler (TDI) Combination modesSameSame
# TransmitChannels128SameSame
# ReceiveChannels256SameSame
510(k) TrackTrack 3SameSame
SystemCharacteristicsBeamformer 128/128 Mobile cart Single LCD monitorSameSame**The Affiniti cart is smaller. Brakes are locking casters on Affiniti while EPIQ has integrated brakes and a control pedal.
Product SafetyCertificationCSA InternationalSameSame
StandardFeatureEPIQK132304(Predicate Device)EPIQ 1.5(Proposed Device)Affiniti 1.5(Proposed Device)
EMCComplianceC22.2 No. 60601-1IEC 60601-1IEC 60601-1-2ANSI/AAMI ES60601-1IEC 60601-2-37SameSame
Patient ContactMaterialsAll patient contactmaterials of the EPIQUltrasound System andtransducers are detailed inK132304.SameNo new patient contactmaterials comparing withthe predicates.SameNo new patient contactmaterials comparing with thepredicates.
AccessoriesVarious hardcopy, recording and printing devices: video thermal printer, report printer, DVD, USB PercuNav Instruments Biopsy guides FootswitchSameSame

{55}------------------------------------------------

{56}------------------------------------------------

4. Device Description

The proposed EPIO and Affiniti Diagnostic Ultrasound Systems, which includes EPIO 5. EPIQ 7, Affiniti 50 and Affiniti 70 systems, are general purpose, software controlled, diagnostic ultrasound systems. Their function is to acquire ultrasound data and to display the data in various modes of operation.

The devices consist of two parts: the system console and the transducers. The system console contains the user interface, a display, system electronics and optional peripherals (ECG, printers). In addition to the physical knobs and buttons of the main control panel, the user interface consists of a touch screen with soft key controls. EPIO also has a OWERTY keyboard.

The removable transducers are connected to the system using a standard technology, multipin connectors. The proposed EPIQ and Affiniti systems use standard transducer technology, and supports phased, linear, curved linear array, TEE, motorized 3D curved linear arrays as well as non-imaging (pencil) probes.

Clinical data storage consists of a local repository as well as off-line image storage via the network, DVR, DVD, and USB storage devices. The images are stored in industry-standard formats (e.g. JPEG. AVI, DICOM) and are intended to be readable using industry-standard hardware and software. On-line review of the images is available. Secure access tools are provided to restrict and log access to the clinical data repository according to HIPAA.

The system circuitry generates an electronic voltage pulse, which is transmitted to the transducer. In the transducer, a piezo electric array converts the electronic pulse into an

{57}------------------------------------------------

ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body tissues. The Doppler functions of the system process the Doppler shift frequencies from the echoes of moving targets such as blood to detect and graphically display the Doppler shifts of these tissues as flow.

The proposed EPIQ and Affiniti systems give the operator the ability to measure anatomical structures and offer analysis packages that provide information used by competent healthcare professionals to make a diagnosis. The proposed EPIQ and Affiniti systems enable image guided navigation and image fusion via the optional PercuNav feature (K121498).

5. Technological Comparison to Predicate Devices

The proposed EPIQ and Affiniti ultrasound systems, and the predicate device EPIQ ultrasound system cleared in K132304 are track 3 systems and employ the same fundamental scientific technology. They are the same in indications for use, materials, type of transducers, optimization, accessories and imaging modes. The primary differences between the proposed EPIQ and Affiniti systems and the predicate device EPIO system are as follows:

  • Integration of OLAB Heart Model software which was cleared via K130159 and ● K132165.
  • . Addition of the new security features to help the healthcare providers be compliant with HIPAA and the regulations imposed by the other governmental and regulatory bodies.
  • . Integration of a service tool called OmniSphere. OmniSphere gives the Ultrasound system the ability to create utilization reports which describe how the system is being used. OmniSphere also allows the Biomedical Engineer within a medical facility to remotely log into a connected system to perform service related operations.
  • Update of the analysis packages to allow the acquisition and export of measurements and . calculations in the fetal echo DICOM template and the small parts DICOM template, to allow the standardization of user defined measurement export , and to include new or newly standardized measurements from standards bodies e.g. ASE, AIUM, ACEP.
  • Addition of the following transducers: VL13-5, X7-2, 3D9-3v, S4-2,C9-4v,C6-2,BP10-● 5ec,L12-4.

6. Indications for Use

Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult). Cephalic (Neonatal). Fetal/Obstetric. Gynecological. Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal.

The clinical environments where the EPIQ and Affiniti Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical point-of-care for diagnosis of patients.

7. Safety Considerations

{58}------------------------------------------------

As track 3 ultrasound devices the proposed EPIQ and Affiniti Ultrasound Systems are designed to comply with the acoustic output display requirements of IEC 60601-2-37 Ed 2.0 (Particular requirements for the basic safety and essential performance of ultrasonic medical and monitoring equipment) and IEC 62359, Ed 2.0 (Ultrasonics - Field characterization -Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields).

The proposed EPIQ and Affiniti Ultrasound Systems comply with the referenced standard as well as the FDA ultrasound specific guidance, Guidance for Industry and FDA Staff -Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (September 9, 2008).

The system acoustic output limits are:

  • . Ispta.3 ≤ 720 MW/cm2
  • MI < 1.9 .
  • . TI < 6.0

The system and transducers are compliant to:

  • . IEC 60601-1: Medical electrical equipment. General requirements for basic safety and essential performance, 2005. Amendment 1, 2012
  • IEC 60601-1-2 Medical Electrical Equipment Part 1-2, General Requirements ● for Basic Safety and Essential Performance - Collateral Standard Electromagnetic Compatibility, 2007
  • IEC 60601-1-6 Medical Electrical Equipment - Part 1-6, General Requirements for Basic Safety and Essential Performance - Usability, 2010
  • IEC 60601-2-37: Medical electrical equipment. Particular requirements for the . basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment, 2007
  • IEC 62359. Ultrasonics – Field characterization – Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, 2010

8. Nonclinical Performance Data

Philips Ultrasound performed the following testing to ensure the safety and effectiveness of the proposed EPIQ and Affiniti systems:

  • . Software Verification and Validation
  • Non-Clinical Performance Data ●
  • Non-Clinical Tests
    • IEC 60601-1: Medical electrical equipment. General requirements for basic . safety and essential performance, 2005, Amendment 1, 2012
    • IEC 60601-1-2 Medical Electrical Equipment Part 1-2, General ● Requirements for Basic Safety and Essential Performance - Collateral Standard Electromagnetic Compatibility, 2007
    • IEC 60601-1-6 Medical Electrical Equipment Part 1-6, General . Requirements for Basic Safety and Essential Performance - Usability, 2010

{59}------------------------------------------------

  • . IEC 60601-2-37: Medical electrical equipment. Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment, 2007
  • ISO 10993: Biological evaluation of medical devices. .
  • Ouality assurance measures applied to the system design and development . include, but were not limited to:
    • Risk Analysis ●
    • Product Specifications ●
    • . Design Reviews
    • Verification and Validation .

9. Clinical Data

Clinical data was not required to demonstrate safety and effectiveness of the proposed EPIQ and Affiniti Diagnostic Ultrasound Systems since the proposed EPIO or Affiniti system introduces no new indications for use, modes or features that have not been previously cleared with the predicate device EPIQ system (K132304). The clinical safety and effectiveness of ultrasound systems with these characteristics are well accepted for both predicate and subject devices.

10. Conclusion

The proposed EPIQ and Affiniti Ultrasound Systems are substantially equivalent in safety and effectiveness to the predicate device identified above:

  • The predicate device and the proposed EPIQ and Affiniti systems are indicated for the diagnostic ultrasonic imaging and fluid flow analysis.
  • The predicate device and the proposed EPIQ and Affiniti systems have the same ● gray-scale and Doppler capabilities.
  • The predicate device and the proposed EPIQ and Affiniti systems use essentially the same technologies for imaging, Doppler functions and signal processing.
  • The predicate device and the proposed EPIQ and Affiniti systems have acoustic output levels within the Track 3 FDA limits.
  • . The predicate device and the proposed EPIQ and Affiniti systems are manufactured under equivalent quality systems.
  • . The predicate device and the proposed EPIQ and Affiniti systems are manufactured of materials with equivalent bio safety. There is no new material or material change in application.
  • . The predicate device and the proposed EPIQ and Affiniti systems are designed and manufactured to the same electrical and physical safety standards.

514 Performance Standards

There are no Sec. 514 performance standards for the proposed EPIQ and Affiniti systems.

Prescription Status

The proposed EPIQ and Affiniti systems are prescription devices. The prescription device statement appears in the labeling.

{60}------------------------------------------------

Sterilization Sites Not applicable. No components are supplied sterile.

Track

The proposed EPIQ and Affiniti systems are Track 3 systems

N/A