K Number
K182603

Validate with FDA (Live)

Date Cleared
2018-11-14

(54 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Z6 Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediativ, small organ(breast, thyroid, testes), neonatal and adult cephalic, transrectal, trans-vaginal, musculo-skeletal(conventional, superficial), adult and pediatric cardiac, peripheral vessel , intraoperative(abdominal, thoracic, and vascular) and urology exams.

Device Description

Z6 Diagnostic Ultrasound System is a general purpose, portable, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound data in B-Mode, M-Mode, PW-Mode, CW-Mode, Color-Mode, Power/Dirpower Mode, 4D, Smart3D,TDI, Color M, Elastography, Biopsy Guidance, Contrast imaging (Contrast agent for Liver) or the combined mode (i.e. B/M-Mode, B/PW-mode, B/PW/Color). This system is a Track 3 device that employs an array of probes that include linear array and convex array.

AI/ML Overview

This document is a 510(k) premarket notification for the Z6 Diagnostic Ultrasound System, which is a general-purpose ultrasonic diagnostic system. The document states that the device is substantially equivalent to legally marketed predicate devices and does not require clinical studies to support substantial equivalence. Therefore, the information typically associated with acceptance criteria and studies proving the device meets these criteria for a novel AI-driven medical device, such as detailed performance metrics, sample sizes for test sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, or standalone algorithm performance, are not explicitly provided or applicable in the context of this specific 510(k) submission.

This 510(k) is primarily focused on demonstrating "substantial equivalence" to existing, legally marketed ultrasound devices (predicates). This means the device is shown to have the same intended use, similar technological characteristics, and be as safe and effective as a predicate device. The performance validation relies on non-clinical tests (acoustic output, biocompatibility, electrical/mechanical safety, etc.) and comparisons to previously cleared devices rather than new clinical efficacy studies.

Based on the provided text, here's what can be extracted regarding the device's validation and "acceptance criteria" in the context of a 510(k):

Acceptance Criteria (Implied by 510(k) Process for Substantial Equivalence):

The acceptance criteria for a 510(k) submission that relies on substantial equivalence are not explicit quantitative performance metrics like those for a novel AI algorithm. Instead, they are met by demonstrating that the new device:

  • Has the same intended use as a predicate device.
  • Has the same technological characteristics as a predicate device, or if it has different technological characteristics, that those differences do not raise new questions of safety or effectiveness AND the device is as safe and effective as the predicate device.
  • Complies with applicable recognized safety standards.

The "Study" Proving the Device Meets Accepted Criteria (as described in the 510(k)):

The "study" here is the overall premarket notification submission process, which includes non-clinical testing and a detailed comparison to predicate devices, rather than a single clinical trial.

1. Table of Acceptance Criteria and Reported Device Performance:

Since this is a 510(k) for substantial equivalence and not a De Novo or PMA for a novel AI, there isn't a table of specific performance acceptance criteria like sensitivity, specificity, or AUC for an AI. Instead, the "performance" is demonstrated through adherence to safety standards and equivalence to predicate devices across various modes and applications.

Acceptance Criteria (Implied by 510(k) Process)Reported Device "Performance" / Compliance
Intended Use Equivalence: Same intended use as predicate devices.The Z6 system has the same intended uses as the predicate device Z6 (K122010). It is applicable for adults, pregnant women, pediatric patients, and neonates across various anatomical regions (fetal, abdominal, pediatric, small organ, neonatal/adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal, cardiac, peripheral vessel, and urology exams).
Technological Equivalence / No New Safety/Effectiveness Questions: Same or comparable technological characteristics as predicates, and any differences don't raise new safety/effectiveness concerns.Z6 has the same technological characteristics as predicate devices. New added features (ECG, 4D, Smart3D, Contrast imaging, Elastography, Free Xros X, TDI, Color M, iNeedle, Medsight, iWorks, iLive) are deemed "substantially equivalent" to features found in other predicate Mindray devices (DC-30, M7, DC-70).
Acoustic Output Safety: Acoustic power levels are below FDA limits.The acoustic power levels of Z6 are below the limits of FDA, same as predicate device Z6(K122010).
Electrical and Physical Safety: Compliance with recognized safety standards.Z6 is designed in compliance with FDA recognized electrical and physical safety standards (e.g., AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 14971).
Biocompatibility: Patient contact materials meet ISO standards.Patient contact materials of new transducers and needle-guided brackets (NGB-001, NGB-002) are tested under ISO 10993-1.
Cleaning and Disinfection Effectiveness: (Implied by general safety requirements)Evaluated for cleaning and disinfection effectiveness.
Usability: (Implied by general safety requirements)Complies with IEC 60601-1-6 (usability standard).

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not applicable in the context of this 510(k). The regulatory pathway for this device (a general-purpose ultrasound system with additional features found in other cleared devices) does not typically require a "test set" of patient data for performance evaluation in the same manner a novel AI algorithm would. Instead, substantial equivalence is primarily demonstrated through non-clinical testing and comparison to existing, cleared devices.
  • Data Provenance: Not applicable, as patient data test sets are not described. The validation is based on engineering testing and comparison to predicate device specifications.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

  • Not applicable. There is no mention of a formal "ground truth" establishment by experts using a test set of patient data for the Z6 Diagnostic Ultrasound System in this 510(k). The submission focuses on device specifications, safety standards, and equivalence to predicate devices.

4. Adjudication Method for the Test Set:

  • Not applicable. No test set requiring adjudication is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size of Human Readers Improvement with AI vs. Without AI Assistance:

  • No. This 510(k) does not describe an MRMC study. The device is a diagnostic ultrasound system, not an AI-assisted interpretation tool in the manner that would typically necessitate such a study (e.g., for an AI CAD system). The "AI" features mentioned (Smart3D, iNeedle, etc.) are likely integrated functionalities that enhance image acquisition or guidance, not interpretative AI requiring MRMC.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

  • Not applicable for the overall device. The device is an ultrasound system with integrated features. While some features (like "Smart3D" or "iNeedle") might involve internal algorithms, their performance is demonstrated as part of the overall device's functionality and comparison to similar features in predicate devices, not as standalone AI outputs that require independent assessment against a ground truth.

7. The Type of Ground Truth Used:

  • Not explicitly defined as a "ground truth" for clinical performance. For a traditional ultrasound system seeking 510(k) clearance via substantial equivalence, the "ground truth" for regulatory purposes often refers to established engineering standards, existing predicate device performance, and the inherent safety/efficacy of ultrasound technology itself, rather than a clinical dataset with adjudicated diagnoses.

8. The Sample Size for the Training Set:

  • Not applicable. This submission is for a diagnostic ultrasound system, not a machine learning model developed with a training set of patient data and medical images.

9. How the Ground Truth for the Training Set was Established:

  • Not applicable. As there is no mention of a training set for an AI model, there is no ground truth establishment for it.

In summary: This 510(k) submission for the Z6 Diagnostic Ultrasound System relies on demonstrating substantial equivalence to existing, legally marketed ultrasound devices. It does so by showing that the device has the same intended use, comparable technological characteristics (with new features demonstrated as equivalent to those found in other cleared predicate devices), and complies with relevant safety and performance standards through non-clinical testing. It does not involve the type of clinical performance studies, test sets, or AI model validation methods typically associated with novel AI-driven devices seeking a different regulatory pathway or greater emphasis on clinical efficacy.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

November 14, 2018

Shenzhen Mindray Bio-medical Electronics Co., LTD Jiang Xiaoyong Engineer of Technical Regulation Keji 12th Road South, Hi-tech Industrial Park Shenzhen, Guangdong 518057 CHINA

Re: K182603

Trade/Device Name: Z6 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: August 31, 2018 Received: September 21, 2018

Dear Jiang Xiaoyong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Hole 2. Nils

for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K182603

Device Name Z6 Diagnostic Ultrasound System

Indications for Use (Describe)

The Z6 Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediativ, small organ(breast, thyroid, testes), neonatal and adult cephalic, transrectal, trans-vaginal, musculo-skeletal(conventional, superficial), adult and pediatric cardiac, peripheral vessel , intraoperative(abdominal, thoracic, and vascular) and urology exams.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)


{3}------------------------------------------------

Diagnostic Ultrasound Indications For Use Format

System: Z6 Diagnostic Ultrasound System

Transducer: N

Transducer: N/A
Intenueu
Liamsducer: Diagr Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
FetalPPPPPPPNote 1,2,5,6,7
AbdominalPPPPPPPNote 1,2,3,4,5,6,7,9
Intra-operative (Specify*)PPPPPPPNote 1,2,3,6
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1,2,3,4,5,6
Small Organ (Specify**)PPPPPPPNote 1,2,3,6,8
Neonatal CephalicPPPPPPPNote 1,2,3,4,5,6
FetalImaging &OtherAdult CephalicPPPPPPPNote 1,2,4,5,6
Trans-rectalPPPPPPPNote 1,2,5,6
Trans-vaginalPPPPPPPNote 1,2,5,6
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPPNote 1,2,3,4,5,6,8
Musculo-skeletal(Superficial)PPPPPPPNote 1,2,3,6,8
Intravascular
Cardiac AdultPPPPPPPNote 1,2,4,5,6
Cardiac PediatricPPPPPPPNote 1,2,4,5,6
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPPNote 1,2,3,5,6
Other (Specify***)PPPPPPPNote 1,2,5,6
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Transducer:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
FetalPPPPPPPNote 1,2,5,6
AbdominalPPPPPPPNote 1,2,5,6,9
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1,2,5,6
Small Organ (Specify**)
Neonatal Cephalic
FetalImaging &OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1,2,5,6
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1,2,5,6
Other (Specify***)
N=new indication; P=previously cleared by FDA(K122010); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{4}------------------------------------------------

Transducer:

{5}------------------------------------------------

System:Z6 Diagnostic Ultrasound System
------------------------------------------

Transducer: 6C2P Internutu
Haasta Diagn

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal
AbdominalPPPPPPNote 1,2,5,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2,5,6
Small Organ (Specify**)
Neonatal CephalicPPPPPPNote 1,2,5,6
FetalImaging &OtherAdult CephalicPPPPPPNote 1,2,5,6
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K122010); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Transducer:6CVIPDiagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
FetalPPPPPPNote 1,2,5,6
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
FetalImaging &OtherAdult Cephalic
Trans-rectalPPPPPPNote 1,2,5,6
Trans-vaginalPPPPPPNote 1,2,5,6
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)PPPPPPNote 1,2,5,6
N=new indication; P=previously cleared by FDA(K122010);E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW+B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{6}------------------------------------------------

Transducer: 6CV1P
Diagnos

{7}------------------------------------------------

System:Z6 Diagnostic Ultrasound System
------------------------------------------

Transducer: 7L4P IIIvinusu Diagn

includedDiagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
----------------------------------------------------------------------------------------------
Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal
AbdominalPPPPPPNote 1,2,3,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2,3,6
Small Organ (Specify**)PPPPPPNote 1,2,3,6,8
Neonatal CephalicPPPPPPNote 1,2,3,6
FetalImaging &OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1,2,3,6,8
Musculo-skeletal(Superficial)PPPPPPNote 1,2,3,6,8
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1,2,3,6
Other (Specify***)

{8}------------------------------------------------

System:Z6 Diagnostic Ultrasound System
------------------------------------------

Transducer: 7L5P
Diagn

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal
AbdominalPPPPPPNote 1,2,3,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2,3,6
Small Organ (Specify**)PPPPPPNote 1,2,3,6
Neonatal CephalicPPPPPPNote 1,2,3,6
FetalImaging &OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1,2,3,6
Musculo-skeletal(Superficial)PPPPPPNote 1,2,3,6
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1,2,3,6
Other (Specify***)
N=new indication; P=previously cleared by FDA(K122010); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)

{9}------------------------------------------------

System:Z6 Diagnostic Ultrasound System
------------------------------------------

Transducer: - L14-6P
Diagno

11966Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
-------------------------------------------------------------------------------------------
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
General(Track 1Clinical ApplicationMode of OperationOther (specify)
Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)
OphthalmicOphthalmic
Fetal
AbdominalPPPPPPNote 1,2,3,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2,3,6
Small Organ (Specify**)PPPPPPNote 1,2,3,6
Neonatal CephalicPPPPPPNote 1,2,3,6
Adult Cephalic
FetalImaging &OtherTrans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1,2,3,6
Musculo-skeletal(Superficial)PPPPPPNote 1,2,3,6
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1,2,3,6
Other (Specify***)
N=new indication; P=previously cleared by FDA(K122010); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power+B、PW+Color+B、Power + PW+B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Clinical ApplicationMode of Operation
General(Track 1)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
FetalImaging &OtherTrans-rectalPPPPPPNote 1,2,5,6
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)PPPPPPPNote 1,2,5,6
N=new indication; P=previously cleared by FDA(K122010); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal
AbdominalPPPPPPPNote 1,2,4,5,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1,2,4,5,6
Small Organ (Specify**)
Neonatal CephalicPPPPPPPNote 1,2,4,5,6
FetalImaging &OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac AdultPPPPPPPNote 1,2,4,5,6
Cardiac PediatricPPPPPPPNote 1,2,4,5,6
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K122010); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)

{10}------------------------------------------------

System: Z6 Diagnostic Ultrasound System

Transducer: CB10-4P

{11}------------------------------------------------

Transducer: 2P2P Internusu
Diagn

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

{12}------------------------------------------------

System:Z6 Diagnostic Ultrasound System
------------------------------------------

Transducer: Diagnostic

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
FetalPPPPPPNote 1,2,5,6
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
FetalImaging &OtherAdult Cephalic
Trans-rectalPPPPPPNote 1,2,5,6
Trans-vaginalPPPPPPNote 1,2,5,6
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)PPPPPPNote 1,2,5,6
N=new indication; P=previously cleared by FDA(K122010); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+ B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.)
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal
AbdominalPPPPPPPNote 1,2,3,6
Intra-operative (Specify*)PPPPPPPNote 1,2,3,6
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1,2,3,6
Small Organ (Specify**)PPPPPPPNote 1,2,3,6
Neonatal CephalicPPPPPPPNote 1,2,3,6
Adult Cephalic
FetalImaging &OtherTrans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPPNote 1,2,3,6
Musculo-skeletal(Superficial)PPPPPPPNote 1,2,3,6
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPPNote 1,2,3,6
Other (Specify***)
Transducer:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
FetalImaging &OtherAdult Cephalic
Trans-rectalPPPPPPNote 1,2,5,6
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)PPPPPPNote 1,2,5,6
N=new indication; P=previously cleared by FDA(K122010); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power+B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal
AbdominalNNNNNNNNote 1,2,4,5,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNNNote 1,2,4,5,6
Small Organ (Specify**)
Neonatal CephalicNNNNNNNNote 1,2,4,5,6
Adult CephalicNNNNNNNNote 1,2,4,5,6
FetalImaging &OtherTrans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)NNNNNNNNote 1,2,4,5,6
Musculo-skeletal(Superficial)
Intravascular
Cardiac AdultNNNNNNNNote 1,2,4,5,6
Cardiac PediatricNNNNNNNNote 1,2,4,5,6
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K172970); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW+B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{13}------------------------------------------------

Transducer: 7LT4P
Diagno

{14}------------------------------------------------

Transducer: 6LE7P Diagnos

{15}------------------------------------------------

Transducer:

{16}------------------------------------------------

Transducer:35C50P
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
FetalPPPPPPPNote 1,2,5,6
AbdominalPPPPPPPNote 1,2,5,6,9
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1,2,5,6
Small Organ (Specify**)
Neonatal Cephalic
FetalImaging &OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPPNote 1,2,5,6
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPPNote 1,2,5,6
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173369); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+ B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
** Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Transducer: 35C50P
Diagnostic

{17}------------------------------------------------

Transducer:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal
AbdominalPPPPPPPNote 1,2,3,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1,2,3,6
Small Organ (Specify**)PPPPPPPNote 1,2,3,6,8
Neonatal CephalicPPPPPPPNote 1,2,3,6
Adult Cephalic
FetalImaging &OtherTrans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPPNote 1,2,3,6,8
Musculo-skeletal(Superficial)PPPPPPPNote 1,2,3,6,8
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPPNote 1,2,3,6
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173369); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+ B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Transducer:D6-2P
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track 1Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
FetalPPPPPPNote 1,2,5,7
AbdominalPPPPPPNote 1,2,5,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
FetalImaging &OtherTrans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173369); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note5: Color M
Note6: Smart3D
Note7: 4D(Real-time 3D)
Note8: Strain Elastography
Note 9: Contrast imaging (Contrast agent for Liver)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Transducer:

{18}------------------------------------------------

Transducer:
Imenucu D6-2P

{19}------------------------------------------------

510(K) SUMMARY

K182603

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92(c).

1. Submitter:

Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China Tel: +86 755 8188 6238 Fax: +86 755 2658 2680

Contact Person:

Jiang Xiaoyong Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China

Date Prepared: September 15, 2018

2. Device Name: Z6 Diagnostic Ultrasound System

Classification

Regulatory Class: II Review Category: Tier II 21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System (IYN) 21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO) 21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX)

3. Device Description:

Z6 Diagnostic Ultrasound System is a general purpose, portable, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound data in B-Mode, M-Mode, PW-Mode, CW-Mode, Color-Mode, Power/Dirpower Mode, 4D,

{20}------------------------------------------------

Smart3D,TDI, Color M, Elastography, Biopsy Guidance, Contrast imaging (Contrast agent for Liver) or the combined mode (i.e. B/M-Mode, B/PW-mode, B/PW/Color). This system is a Track 3 device that employs an array of probes that include linear array and convex array.

4. Intended Use:

The Z6 Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, intraoperative(abdominal, thoracic, and vascular), pediatric, small organ(breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal(conventional, superficial), adult and pediatric cardiac, peripheral vessel and urology exams.

5. Summary of Modifications

This submission device is a modification to Z6 Diagnostic Ultrasound System previously cleared in K122010.

● New Added Transducers:

P7-3P 75L38P 35C50P D6-2P

New Added Needle-Guided Bracket: .

NGB-001

NGB-002

Main Added Features: .

Contrast Imaging (Contrast agent for Liver)

TDI 4D Smart 3D

Color M

ECG

{21}------------------------------------------------

Free Xros M iNeedle Medsight iWorks iLive ExFov Elastography to 75L38P and 7L4P Probe

6. Comparison with Predicate Devices:

Z6 Diagnostic Ultrasound System is comparable with and substantially equivalent to these predicate devices:

Predicate DeviceManufacturerModel510(k) ControlNumber
1. Primary predicate deviceMindrayZ6K122010
2. Reference deviceMindrayDC-30K173369
3. Reference deviceMindrayM7K172970
4. Reference deviceMindrayDC-70K163690

Z6 has the same technological characteristics, is comparable in key safety and effectiveness features, and has the same intended uses and basic operating modes as the predicate devices. All systems transmit ultrasonic energy into patients and perform post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures and flow, and calculations.

  • Subject device Z6 has the same intended uses as the predicated device Z6(K122010) ●
Subject DevicePredicate device
Z6Z6(K122010)
The Z6 Diagnostic Ultrasound System isapplicable for adults, pregnant women,pediatric patients and neonates. It is intendedThe Z6 Diagnostic Ultrasound System isapplicable for adults, pregnant women, pediatricpatients and neonates. It is intended for use in

{22}------------------------------------------------

for use in fetal, abdominal,intraoperative(abdominal, thoracic, andvascular), pediatric, small organ(breast,thyroid, testes), neonatal and adult cephalic,trans-rectal, trans-vaginal,musculo-skeletal(conventional, superficial),adult and pediatric cardiac, peripheral vesseland urology exams.fetal, abdominal, intraoperative(abdominal,thoracic, and vascular), pediatric, smallorgan(breast, thyroid, testes), neonatal and adultcephalic, trans-rectal, trans-vaginal,musculo-skeletal(conventional, superficial),adult and pediatric cardiac, peripheral vesseland urology exams.
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Compared to the predicated devices Z6(K122010):

  • The acoustic power levels of Z6 are below the limits of FDA, which is the same as the predicated device Z6(K122010).
  • Z6 is designed in compliance with the FDA recognized electrical and physical safety standard, which is the same as the predicated device Z6(K122010).
  • Z6 has the same performance specification as the predicate device Z6(K122010) except adding ECG, 4D, Smart 3D, Contrast imaging (Contrast agent for Liver), Elastography, Free Xros X. New added features are substantial equivalent with the predicated device DC-30( K173369) .
  • . Z6 has the same performance specification as the predicate device Z6(K122010) except adding TDI, Color M. New added features are substantial equivalent with the predicated device M7( K172970)
  • . Z6 has the same function specification as the predicate device Z6(K122010) except adding iNeedle, Medsight. New added features are substantial equivalent with the predicated device M7( K172970).
  • . Z6 has the same function specification as the predicate device Z6(K122010) except adding iWoks, iLive. New added features are substantial equivalent with the predicated device DC-70( K163690).
  • The patient contact material of the new added transducers are tested under ISO 10993-1.
  • The patient contact materials of the new added needle-guided bracket are cleared along with NGB-001(K172970) and NGB-002(K172970).
  • The transducers are same except for adding P7-3P, 75L38P, 35C50P, D6-2P. All of

{23}------------------------------------------------

Subject devicePredicated device
Z6
P7-3PP7-3s(M7)
75L38P75L38P(DC-30)
35C50P35C50P(DC-30)
D6-2PD6-2P(DC-30)

the new added probes are substantial equivalent with the predicated device listed below:

7. Non-clinical Tests:

Z6 Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been designed to conform with applicable medical safety standards. This device has been tested and evaluated under the following standards:

  • AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2 Edition 3: 2007: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 3)
  • . IEC 60601-2-37 Edition 2.0 2007: Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
  • . IEC 60601-1-6 Edition 3.1 2013: medical electrical equipment - part 1-6: general requirements for basic safety and essential performance - collateral standard: usability.
  • ISO 14971 Second edition 2007: Medical devices Application of risk management ● to medical devices
  • UD 2: Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment . Revision 3
  • . ISO10993-1 Fourth edition 2009: Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process

These non-clinical tests relied on in this premarket notification submission can support the

{24}------------------------------------------------

determination of substantial equivalence of the subject device.

8. Clinical Studies

Not applicable. The subject of this submission, Z6 Diagnostic Ultrasound System, does not require clinical studies to support substantial equivalence.

Conclusion:

Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines and established methods of patient examination. The design, development and quality process of the manufacturer confirms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards. Therefore, the Z6 Diagnostic Ultrasound System is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.