The DC-70/ DC-70T / DC-70 Pro/ DC-70 Exp/ DC-75/ DC-78/ DC-70S Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-esoph(Cardiac), musculo-skeletal (conventional, superficial), adult and pediatric cardiac, peripheral vessel and urology exams.
DC-70/DC-70T /DC-70 Pro/DC-70 Exp/DC-75/DC-78/DC-70S is a software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound data in B-Mode, M-Mode, PW-Mode, CW-Mode, Color-Mode , Power/Dirpower Mode, THI, iScape,4D, Smart3D,TDI, Color M, Elastography, Biopsy Guidance, Contrast imaging or the combined mode (i.e. B/M-Mode, B/PW-mode, B/PW/Color). This system is a Track 3 device that employs an array of probes that include linear array and convex array with a frequency range of approximately 2.0 MHz to 10.0 MHz.
This document describes the premarket notification (510(k)) for the Mindray DC-70/DC-70T/DC-70 Pro/DC-70 Exp/DC-75/DC-78/DC-70S Diagnostic Ultrasound System. It is a submission to the FDA for substantial equivalence to legally marketed predicate devices.
The information provided focuses on the technical specifications, intended uses, and compliance with safety standards, rather than a study proving the device meets specific acceptance criteria related to AI performance or diagnostic accuracy in the way described in the prompt's request. Crucially, the document explicitly states that no clinical studies were performed.
Therefore, I cannot fulfill all parts of your request as the provided text does not contain data on AI performance, human reader improvement with AI assistance, or specific acceptance criteria met through clinical trials. It primarily addresses the substantial equivalence of an ultrasound imaging system in terms of its features, safety, and technological characteristics compared to predicate devices.
However, I can extract the information relevant to the device's stated performance and the basis for its acceptance from a regulatory perspective, focusing on the equivalence to existing devices and compliance with safety standards.
Here's an attempt to answer using the provided text, while noting the limitations:
Acceptance Criteria and Device Performance (based on provided text):
The document does not define specific "acceptance criteria" for diagnostic accuracy or AI performance in the traditional sense, nor does it report "device performance" in terms of metrics like sensitivity, specificity, or AUC, as it's a 510(k) submission primarily focused on substantial equivalence (safety and efficacy similar to existing devices). Instead, the acceptance is based on demonstrating the device's technical characteristics, safety, and intended uses are comparable to already cleared predicate devices and comply with relevant standards.
Table of "Acceptance Criteria" (interpreted as demonstrated compliance/equivalence) and "Reported Device Performance" (interpreted as features/capabilities):
| Acceptance Criteria (from a regulatory perspective for 510(k)) | Reported Device "Performance" / Characteristics |
|---|---|
| Intended Use Equivalence: Device intended uses are substantially equivalent to legally marketed predicate devices. | The DC-70/DC-70T/DC-70 Pro/DC-70 Exp/DC-75/DC-78/DC-70S Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients, and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-vaginal, Trans-esoph (Cardiac), musculo-skeletal (conventional, superficial), adult and pediatric cardiac, peripheral vessel, and urology exams. This is stated to be the same as the predicate device DC-70 (K150204). (Minor differences in specific transducer indications are noted but cleared based on "P=previously cleared by FDA" or "N=new indication" status, implying equivalence or new but standard indications supported by other predicate devices). |
| Acoustic Power Levels: Acoustic output levels are within FDA limits and comparable to predicates. | The acoustic power levels of the subject device are below the limits of FDA, and are the same as the predicate device DC-70 (K150204). |
| Electrical and Physical Safety: Device complies with recognized safety standards. | The device is designed in compliance with the FDA recognized electrical and physical safety standard, and is the same as the predicate device DC-70 (K150204). It has been tested and evaluated under standards including AAMI/ANSI ES60601-1, AAMI/ANSI/IEC 60601-1-2, IEC 60601-2-37, IEC 62304, IEC 62366, IEC 60601-1-6, ISO14971, UD 2, UD 3, and ISO 10993-1. |
| Mechanical Safety & Biocompatibility: Device features are safe for use. | Evaluated for biocompatibility, cleaning and disinfection effectiveness, and mechanical safety. (Specific results are not detailed but compliance to standards like ISO 10993-1 is stated). |
| Technological Characteristics Equivalence: Key features and operating modes are comparable to predicates. | The system is a software-controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound data in B-Mode, M-Mode, PW-Mode, CW-Mode, Color-Mode, Power/Dirpower Mode, THI, iScape, 4D, Smart3D, TDI, Color M, Elastography, Biopsy Guidance, Contrast imaging, or combined modes (i.e., B/M-Mode, B/PW-mode, B/PW/Color). It is a Track 3 device employing array probes with a frequency range of approximately 2.0 MHz to 10.0 MHz. New features include Contrast Imaging, Smart FLC, and Tissue Tracking Elastography. It has similar probes as the predicated device. |
| Quality System Compliance: Manufacturing processes adhere to quality standards. | The design, development, and quality process of the manufacturer conforms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems. |
Study Details (based on provided text):
-
Sample size used for the test set and the data provenance: Not applicable. The document explicitly states "Clinical Studies: Not applicable. The subject of this submission... does not require clinical studies to support substantial equivalence." The evaluation relies on non-clinical (engineering, safety, and performance) testing and comparison to predicate devices, not diagnostic performance on a test set of patient data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical studies or test sets with ground truth were used or are described.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable, as no clinical studies or test sets were used.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document is for a diagnostic ultrasound system, not an AI or CAD (Computer-Aided Diagnosis) device providing analytical results that would assist human readers in a diagnostic task. The "Smart3D" and "iScape" features mentioned are imaging modes/enhancements, not necessarily AI for diagnostic assistance. There is no mention of such a study or effect size.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI diagnostic algorithm; it's an ultrasound imaging device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable, as no clinical studies or test sets for diagnostic accuracy were performed or are detailed.
-
The sample size for the training set: Not applicable. This is not an AI/ML device that underwent training on a dataset for a diagnostic task.
-
How the ground truth for the training set was established: Not applicable.
In summary, this 510(k) submission primarily relies on demonstrating substantial equivalence to existing, legally marketed ultrasound devices through non-clinical testing (e.g., electrical safety, acoustic output, biocompatibility, software validation) and detailed comparison of technological characteristics and intended uses. It does not involve AI performance evaluation or clinical studies that would generate data for the specific "acceptance criteria" and "study proving device meets acceptance criteria" as outlined in your prompt for AI/diagnostic accuracy studies.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three heads, depicted in a simple, line-art style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. April 20, 2017 YANG ZHAOHUI ENGINEER OF TECHNICAL REGULATION MINDRAY BUILDING, KEJI 12TH ROAD SOUTH HI-TECH INDUSTRIAL PARK NASHAN, SHENZHEN 518057 CN
Re: K163690
Trade/Device Name: DC-70/DC-70T/DC-70 Pro/DC-70 Exp/DC-75/DC-78/DC-70S Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX, LLZ Dated: March 30, 2017 Received: April 4, 2017
Dear Mr. Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D.'Hara
For
Robert Ochs Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name
DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System
Indications for Use (Describe)
The DC-70/ DC-70T / DC-70 Pro/ DC-70 Exp/ DC-78/ DC-78/ DC-70S Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-esoph(Cardiac), musculo-skeletal (conventional, superficial), adult and pediatric cardiac, peripheral vessel and urology exams.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
Diagnostic Ultrasound Indications For Use Format
DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System System: Transducer:
шиниси N/A
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: ﺴﻤﻴﻨﻴ
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | P | Note 1,2,3, 4,6,7 | ||
| Abdominal | P | P | P | P | P | P | P | Note 1,2,3, 4,6,7,9 | |
| Intra-operative (Specify*) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | P | P | P | P | P | P | Note 1,2, 4,6,7 | ||
| Small Organ (Specify**) | P | P | P | P | P | P | Note 1,2, 4,7,8 | ||
| FetalImaging &Other | Neonatal Cephalic | P | P | P | P | P | P | P | Note 1,2,4,6,7 |
| Adult Cephalic | P | P | P | P | P | P | P | Note 1,2,4,6,7 | |
| Trans-rectal | P | P | P | P | P | P | Note 1,2,3,4,6,7 | ||
| Trans-vaginal | P | P | P | P | P | P | Note 1,2,3,4,6,7,8 | ||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal (Conventional) | P | P | P | P | P | P | Note 1,2,4,6,7,8 | ||
| Musculo-skeletal (Superficial) | P | P | P | P | P | P | Note 1,2,4,7,8 | ||
| Intravascular | |||||||||
| Cardiac Adult | P | P | P | P | P | P | P | Note 1,2,4,5,6,7 | |
| Cardiac Pediatric | P | P | P | P | P | P | P | Note 1,2,4,5,6,7 | |
| Cardiac | Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | N | N | N | N | N | N | N | Note 1, 5,6 | |
| Intra-cardiac | |||||||||
| Peripheralvessel | Peripheral vessel | P | P | P | P | P | P | Note 1,2,4,6,7,8 | |
| Other (Specify***) | P | P | P | P | P | P | Note 1,2,4,6,7 | ||
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | |||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | |||||||||
| *Intraoperative includes abdominal, thoracic, and vascular | |||||||||
| **Small organ-breast, thyroid, testes. | |||||||||
| ***Other use includes Urology. | |||||||||
| Note 1: Tissue Harmonic Imaging. | |||||||||
| Note 2: Smart3D | |||||||||
| Note 3:4D(Real-time 3D) | |||||||||
| Note 4: iScape | |||||||||
| Note5: TDI | |||||||||
| Note6: Color M | |||||||||
| Note7: Biopsy Guidance | |||||||||
| Note8: Elastography | |||||||||
| Note 9: Contrast imaging | |||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | |||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
{4}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | C5-2E |
| Clinical Application | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | P | Note 1, 2, 4,6,7 | ||
| Abdominal | P | P | P | P | P | P | Note 1, 2, 4,6,7,9 | ||
| Intra-operative (Specify*) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | P | P | P | P | P | P | Note 1, 2, 4,6,7 | ||
| Small Organ (Specify**) | |||||||||
| FetalImaging &Other | Neonatal Cephalic | ||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal (Conventional) | P | P | P | P | P | P | Note 1, 2, 4,6,7 | ||
| Musculo-skeletal (Superficial) | |||||||||
| Intravascular | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Cardiac | Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Peripheralvessel | Peripheral vessel | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |
| Other (Specify***) |
1 Parisation
Hitchuru
Hearthern Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{5}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | C11-3E |
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(Track 1) | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| Small Organ (Specify**) | ||||||||||
| FetalImaging &Other | Neonatal Cephalic | P | P | P | P | P | P | Note 1, 2, 4,6,7 | ||
| Adult Cephalic | ||||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | ||||||||||
| Musculo-skeletal (Superficial) | ||||||||||
| Intravascular | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| Cardiac | Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheralvessel | Peripheral vessel | P | P | P | P | P | P | Note 1, 2, 4,6,7 | ||
| Other (Specify***) | ||||||||||
| N=new indication; P=previously cleared by FDA(K150204); E=added under Appendix E | ||||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power+B、PW+Color+B、Power + PW +B. | ||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | ||||||||||
| **Small organ-breast, thyroid, testes. | ||||||||||
| ***Other use includes Urology. | ||||||||||
| Note 1: Tissue Harmonic Imaging. | ||||||||||
| Note 2: Smart3D | ||||||||||
| Note 3:4D(Real-time 3D) | ||||||||||
| Note 4: iScape | ||||||||||
| Note5: TDI | ||||||||||
| Note6: Color M | ||||||||||
| Note7: Biopsy Guidance | ||||||||||
| Note8: Elastography | ||||||||||
| Note 9: Contrast imaging | ||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | ||||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) | ||||||||||
| Clinical Application | Mode of Operation | |||||||||
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | |||||||||
| FetalImaging &Other | Fetal | P | P | P | P | P | P | Note 1, 2, 4,6,7 | ||
| Abdominal | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| Pediatric | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Small Organ (Specify**) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Cardiac | Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal (Conventional) | ||||||||||
| Musculo-skeletal (Superficial) | ||||||||||
| Intravascular | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular (Cardiac) | ||||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheralvessel | Peripheral vessel | |||||||||
| Other (Specify***) | ||||||||||
| N=new indication; P=previously cleared by FDA(K150204); E=added under Appendix E | ||||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | ||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | ||||||||||
| **Small organ-breast, thyroid, testes. | ||||||||||
| ***Other use includes Urology. | ||||||||||
| Note 1: Tissue Harmonic Imaging. | ||||||||||
| Note 2: Smart3D | ||||||||||
| Note 3: 4D(Real-time 3D) | ||||||||||
| Note 4: iScape | ||||||||||
| Note5: TDI | ||||||||||
| Note6: Color M | ||||||||||
| Note7: Biopsy Guidance | ||||||||||
| Note8: Elastography | ||||||||||
| Note 9: Contrast imaging | ||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | ||||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) | ||||||||||
| Clinical Application | Mode of Operation | |||||||||
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | P | P | P | P | P | P | Note 1, 2, 4,7 | |||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | P | P | P | P | P | P | Note 1, 2, 4,7 | |||
| Small Organ (Specify**) | P | P | P | P | P | P | Note 1,2, 4,7,8 | |||
| FetalImaging &Other | Neonatal Cephalic | |||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | P | P | P | P | P | P | Note 1, 2, 4,7 | |||
| Musculo-skeletal (Superficial) | P | P | P | P | P | P | Note 1, 2, 4,7 | |||
| Intravascular | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Cardiac | Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheralvessel | Peripheral vessel | P | P | P | P | P | P | Note 1, 2, 4,7 | ||
| Other (Specify***) | ||||||||||
| Clinical Application | Mode of Operation | |||||||||
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | |||||||||
| FetalImaging &Other | Fetal | |||||||||
| Abdominal | P | P | P | P | P | P | Note 1,2, 4,7 | |||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | P | P | P | P | P | P | Note 1,2, 4,7 | |||
| Small Organ (Specify**) | P | P | P | P | P | P | Note 1,2, 4,7,8 | |||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | P | P | P | P | P | P | Note 1,2, 4,7 | |||
| Musculo-skeletal (Superficial) | P | P | P | P | P | P | Note 1,2, 4,7 | |||
| Intravascular | ||||||||||
| Cardiac | Cardiac Adult | |||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular (Cardiac) | ||||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheralvessel | Peripheral vessel | P | P | P | P | P | P | Note 1,2, 4,7 | ||
| Other (Specify***) | ||||||||||
| N=new indication; P=previously cleared by FDA(K150204); | E=added under Appendix E | |||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | ||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | ||||||||||
| **Small organ-breast, thyroid, testes. | ||||||||||
| ***Other use includes Urology. | ||||||||||
| Note 1: Tissue Harmonic Imaging. | ||||||||||
| Note 2: Smart3D | ||||||||||
| Note 3:4D(Real-time 3D) | ||||||||||
| Note 4: iScape | ||||||||||
| Note5: TDI | ||||||||||
| Note6: Color M | ||||||||||
| Note7: Biopsy Guidance | ||||||||||
| Note8: Elastography | ||||||||||
| Note 9: Contrast imaging | ||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | ||||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
I Tanback
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{6}------------------------------------------------
DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System System: Transducer: C C7-3E
Intenueu
Home Diagno
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{7}------------------------------------------------
DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System System: Transducer: L12-3E
Intendeu
Hagnos
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{8}------------------------------------------------
DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System System: Transducer:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{9}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | L14-6WE |
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | |||||||||
| FetalImaging &Other | Fetal | |||||||||
| Abdominal | P | P | P | P | P | P | Note 1,2, 4,7 | |||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | P | P | P | P | P | P | Note 1,2, 4,7 | |||
| Small Organ (Specify**) | P | P | P | P | P | P | Note 1,2, 4,7,8 | |||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | P | P | P | P | P | P | Note 1,2, 4,7 | |||
| Musculo-skeletal (Superficial) | P | P | P | P | P | P | Note 1,2, 4,7 | |||
| Intravascular | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Cardiac | Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheralvessel | Peripheral vessel | P | P | P | P | P | P | Note 1,2, 4,7 | ||
| Other (Specify***) | ||||||||||
| N=new indication; P=previously cleared by FDA(K150204); E=added under Appendix E | ||||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | ||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | ||||||||||
| **Small organ-breast, thyroid, testes. | ||||||||||
| ***Other use includes Urology. | ||||||||||
| Note 1: Tissue Harmonic Imaging. | ||||||||||
| Note 2: Smart3D | ||||||||||
| Note 3:4D(Real-time 3D) | ||||||||||
| Note 4: iScape | ||||||||||
| Note5: TDI | ||||||||||
| Note6: Color M | ||||||||||
| Note7: Biopsy Guidance | ||||||||||
| Note8: Elastography | ||||||||||
| Clinical Application | Mode of Operation | |||||||||
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | |||||||||
| FetalImaging &Other | Fetal | P | P | P | P | P | P | Note 1, 2, 4,6,7 | ||
| Abdominal | ||||||||||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| Small Organ (Specify**) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| Trans-vaginal | P | P | P | P | P | N | Note 1, 2, 4,6,7,8 | |||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | ||||||||||
| Musculo-skeletal (Superficial) | ||||||||||
| Intravascular | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Cardiac | Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheralvessel | Peripheral vessel | |||||||||
| Other (Specify***) | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| N=new indication; P=previously cleared by FDA(K150204); | E=added under Appendix E | |||||||||
| Additional comments: Combined modes--B+M、PW+B、Color+B、Power + B、PW+Color+B、Power + PW +B. | ||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | ||||||||||
| ** Small organ-breast, thyroid, testes. | ||||||||||
| ***Other use includes Urology. | ||||||||||
| Note 1: Tissue Harmonic Imaging. | ||||||||||
| Note 2: Smart3D | ||||||||||
| Note 3:4D(Real-time 3D) | ||||||||||
| Note 4: iScape | ||||||||||
| Note5: TDI | ||||||||||
| Note6: Color M | ||||||||||
| Note7: Biopsy Guidance | ||||||||||
| Note8: Elastography | ||||||||||
| Note 9: Contrast imaging | ||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | ||||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) | ||||||||||
| Mode of Operation | ||||||||||
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| Abdominal | ||||||||||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| FetalImaging &Other | Small Organ (Specify**) | |||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| Trans-vaginal | P | P | P | P | P | P | Note 1, 2, 4,6,7,8 | |||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | ||||||||||
| Musculo-skeletal (Superficial) | ||||||||||
| Intravascular | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Cardiac | Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheralvessel | Peripheral vessel | |||||||||
| Other (Specify***) | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| N=new indication; P=previously cleared by FDA(K150204); E=added under Appendix E | ||||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | ||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | ||||||||||
| **Small organ-breast, thyroid, testes. | ||||||||||
| ***Other use includes Urology. | ||||||||||
| Note 1: Tissue Harmonic Imaging. | ||||||||||
| Note 2: Smart3D | ||||||||||
| Note 3:4D(Real-time 3D) | ||||||||||
| Note 4: iScape | ||||||||||
| Note5: TDI | ||||||||||
| Note6: Color M | ||||||||||
| Note7: Biopsy Guidance | ||||||||||
| Note8: Elastography | ||||||||||
| Note 9: Contrast imaging | ||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | ||||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
miendeu Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: ﺴﺘﻠ
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{10}------------------------------------------------
DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System System: Transducer: V11-3E
Intendeu
Haansducer: Diagnost
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{11}------------------------------------------------
DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System System: Transducer: V11-3BE
Intenueu
Haardeu Diagnostic
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{12}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | V11-3WE |
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| Abdominal | ||||||||||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| FetalImaging &Other | Small Organ (Specify**) | |||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| Trans-vaginal | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | ||||||||||
| Musculo-skeletal (Superficial) | ||||||||||
| Intravascular | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Cardiac | Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheralvessel | Peripheral vessel | |||||||||
| Other (Specify***) | P | P | P | P | P | P | Note 1, 2, 4,6,7 | |||
| N=new indication; P=previously cleared by FDA(K150204); E=added under Appendix E | ||||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | ||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | ||||||||||
| ** Small organ-breast, thyroid, testes. | ||||||||||
| ***Other use includes Urology. | ||||||||||
| Note 1: Tissue Harmonic Imaging. | ||||||||||
| Note 2: Smart3D | ||||||||||
| Note 3:4D(Real-time 3D) | ||||||||||
| Note 4: iScape | ||||||||||
| Note5: TDI | ||||||||||
| Note6: Color M | ||||||||||
| Note7: Biopsy Guidance | ||||||||||
| Note8: Elastography | ||||||||||
| Note 9: Contrast imaging | ||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | ||||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) | ||||||||||
| Clinical Application | Mode of Operation | Other (specify) | ||||||||
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | ||
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | P | P | P | P | P | P | P | Note 1, 2,4,6,7 | ||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | P | P | P | P | P | P | Note 1, 2,4,6,7 | |||
| Small Organ (Specify**) | ||||||||||
| FetalImaging &Other | Neonatal Cephalic | P | P | P | P | P | P | P | Note 1, 2,4,6,7 | |
| Adult Cephalic | P | P | P | P | P | P | P | Note 1, 2,4,6,7 | ||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | ||||||||||
| Musculo-skeletal (Superficial) | ||||||||||
| Intravascular | ||||||||||
| Cardiac Adult | P | P | P | P | P | P | P | Note 1, 2,4,5,6,7 | ||
| Cardiac Pediatric | P | P | P | P | P | P | P | Note 1, 2,4,5,6,7 | ||
| Cardiac | Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheralvessel | Peripheral vessel | |||||||||
| Other (Specify***) | ||||||||||
| N=new indication; P=previously cleared by FDA(K150204); E=added under Appendix E | ||||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | ||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | ||||||||||
| **Small organ-breast, thyroid, testes. | ||||||||||
| ***Other use includes Urology. | ||||||||||
| Note 1: Tissue Harmonic Imaging. | ||||||||||
| Note 2: Smart3D | ||||||||||
| Note 3:4D(Real-time 3D) | ||||||||||
| Note 4: iScape | ||||||||||
| Note5: TDI | ||||||||||
| Note6: Color M | ||||||||||
| Note7: Biopsy Guidance | ||||||||||
| Note8: Elastography | ||||||||||
| Note 9: Contrast imaging | ||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | ||||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
i Pansado
III entileu Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{13}------------------------------------------------
DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System System: Transducer:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{14}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | P7-3E |
| Clinical Application | Mode of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | ||
| Ophthalmic | Ophthalmic | ||||||||||
| Fetal | |||||||||||
| Abdominal | P | P | P | P | P | P | P | Note 1,2,4,6 | |||
| Intra-operative (Specify*) | |||||||||||
| Intra-operative (Neuro) | |||||||||||
| Laparoscopic | |||||||||||
| Pediatric | P | P | P | P | P | P | Note 1,2,4,6 | ||||
| FetalImaging &Other | Small Organ (Specify**) | ||||||||||
| Neonatal Cephalic | P | P | P | P | P | P | P | Note 1,2,4,6 | |||
| Adult Cephalic | P | P | P | P | P | P | P | Note 1,2,4,6 | |||
| Trans-rectal | |||||||||||
| Trans-vaginal | |||||||||||
| Trans-urethral | |||||||||||
| Trans-esoph. (non-Card.) | |||||||||||
| Musculo-skeletal (Conventional) | |||||||||||
| Musculo-skeletal (Superficial) | |||||||||||
| Intravascular | |||||||||||
| Cardiac Adult | P | P | P | P | P | P | P | Note 1, 2,4,5,6 | |||
| Cardiac Pediatric | P | P | P | P | P | P | P | Note 1, 2,4,5,6 | |||
| Cardiac | Intravascular (Cardiac) | ||||||||||
| Trans-esoph. (Cardiac) | |||||||||||
| Intra-cardiac | |||||||||||
| Peripheralvessel | Peripheral vessel | ||||||||||
| Other (Specify***) | |||||||||||
| N=new indication; P=previously cleared by FDA(K150204); E=added under Appendix E | |||||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | |||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | |||||||||||
| **Small organ-breast, thyroid, testes. | |||||||||||
| ***Other use includes Urology. | |||||||||||
| Note 1: Tissue Harmonic Imaging. | |||||||||||
| Note 2: Smart3D | |||||||||||
| Note 3:4D(Real-time 3D) | |||||||||||
| Note 4: iScape | |||||||||||
| Note5: TDI | |||||||||||
| Note6: Color M | |||||||||||
| Note7: Biopsy Guidance | |||||||||||
| Note8: Elastography | |||||||||||
| Note 9: Contrast imaging | |||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | |||||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) | |||||||||||
| Clinical Application | Mode of Operation | ||||||||||
| General | Ophthalmic | Specific (Track 1 & 3) | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Combined (specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | ||||||||||
| FetalImaging &Other | Fetal | P | P | P | P | P | P | Note 1, 3, 4, 6, 7 | |||
| Abdominal | P | P | P | P | P | P | Note 1, 3, 4, 6, 7 | ||||
| Intra-operative (Specify*) | |||||||||||
| Intra-operative (Neuro) | |||||||||||
| Laparoscopic | |||||||||||
| Pediatric | |||||||||||
| Small Organ (Specify**) | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Trans-rectal | |||||||||||
| Trans-vaginal | |||||||||||
| Trans-urethral | |||||||||||
| Trans-esoph. (non-Card.) | |||||||||||
| Musculo-skeletal (Conventional) | |||||||||||
| Musculo-skeletal (Superficial) | |||||||||||
| Intravascular | |||||||||||
| Cardiac | Cardiac Adult | ||||||||||
| Cardiac Pediatric | |||||||||||
| Intravascular (Cardiac) | |||||||||||
| Trans-esoph. (Cardiac) | |||||||||||
| Intra-cardiac | |||||||||||
| Peripheralvessel | Peripheral vessel | ||||||||||
| Other (Specify***) | |||||||||||
| N=new indication; P=previously cleared by FDA(K150204); E=added under Appendix E | |||||||||||
| Additional comments: Combined modes--B+M、PW+B、Color+B、Power + B、PW+Color+B、Power + PW +B. | |||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | |||||||||||
| **Small organ-breast, thyroid, testes. | |||||||||||
| ***Other use includes Urology. | |||||||||||
| Note 1: Tissue Harmonic Imaging. | |||||||||||
| Note 2: Smart3D | |||||||||||
| Note 3:4D(Real-time 3D) | |||||||||||
| Note 4: iScape | |||||||||||
| Note5: TDI | |||||||||||
| Note6: Color M | |||||||||||
| Note7: Biopsy Guidance | |||||||||||
| Note8: Elastography | |||||||||||
| Note 9: Contrast imaging | |||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | |||||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
Praisado
Illiemaeu Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{15}------------------------------------------------
DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System System: Transducer: D6-2E
Intendeu Diagno
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{16}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | DE10-3E |
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | P | Note 1, 3, 4, 6 | ||
| Abdominal | |||||||||
| Intra-operative (Specify*) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (Specify**) | |||||||||
| Fetal | Neonatal Cephalic | ||||||||
| Imaging & | Adult Cephalic | ||||||||
| Other | Trans-rectal | P | P | P | P | P | P | Note 1, 3, 4, 6 | |
| Trans-vaginal | P | P | P | P | P | P | Note 1, 3, 4, 6 | ||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal (Conventional) | |||||||||
| Musculo-skeletal (Superficial) | |||||||||
| Intravascular | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Cardiac | Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Peripheral | Peripheral vessel | ||||||||
| vessel | Other (Specify***) | ||||||||
| N=new indication; P=previously cleared by FDA(K150204); | E=added under Appendix E | ||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | |||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | |||||||||
| ** Small organ-breast, thyroid, testes. | |||||||||
| ***Other use includes Urology. | |||||||||
| Note 1: Tissue Harmonic Imaging. | |||||||||
| Note 2: Smart3D | |||||||||
| Note 3:4D(Real-time 3D) | |||||||||
| Note 4: iScape | |||||||||
| Note5: TDI | |||||||||
| Note6: Color M | |||||||||
| Note7: Biopsy Guidance | |||||||||
| Note8: Elastography | |||||||||
| Note 9: Contrast imaging | |||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | |||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
rransado
Imendeu Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{17}------------------------------------------------
DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System System: Transducer: CW5s
Intenueu
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General | Specific (Track 1 & 3) | B | M | PWD | CWD | Color | Amplitude Combined | Other (specify) | |
| (Track 1 | Doppler | Doppler | (specify) | ||||||
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative (Specify*) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (Specify**) | |||||||||
| Fetal | Neonatal Cephalic | ||||||||
| Imaging &Other | Adult Cephalic | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal (Conventional) | |||||||||
| Musculo-skeletal (Superficial) | |||||||||
| Intravascular | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Cardiac | Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Peripheral | Peripheral vessel | P | |||||||
| vessel | Other (Specify***) | ||||||||
| N=new indication; P=previously cleared by FDA(K150204); | E=added under Appendix E | ||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | |||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | |||||||||
| ** Small organ-breast, thyroid, testes. | |||||||||
| ***Other use includes Urology. | |||||||||
| Note 1: Tissue Harmonic Imaging. | |||||||||
| Note 2: Smart3D | |||||||||
| Note 3:4D(Real-time 3D) | |||||||||
| Note 4: iScape | |||||||||
| Note5: TDI | |||||||||
| Note6: Color M | |||||||||
| Note7: Biopsy Guidance | |||||||||
| Note8: Elastography | |||||||||
| Note 9: Contrast imaging | |||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | |||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{18}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: L9-3E |
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track 1) | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) |
| Ophthalmic | Ophthalmic | ||||||||
| FetalImaging &Other | Fetal | ||||||||
| Abdominal | N | N | N | N | N | N | Note 1,2, 4,7 | ||
| Intra-operative (Specify*) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | N | N | N | N | N | N | Note 1,2, 4,7 | ||
| Small Organ (Specify**) | N | N | N | N | N | N | Note 1,2, 4,7,8 | ||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal (Conventional) | N | N | N | N | N | N | Note 1,2, 4,7 | ||
| Musculo-skeletal (Superficial) | N | N | N | N | N | N | Note 1,2, 4,7 | ||
| Intravascular | |||||||||
| Cardiac | Cardiac Adult | ||||||||
| Cardiac Pediatric | |||||||||
| Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Peripheralvessel | Peripheral vessel | N | N | N | N | N | N | Note 1,2, 4,7 | |
| Other (Specify***) |
{19}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | DE11-3E |
| Transducer: DE11-3E | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | Mode of Operation | |||||||||
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | N | N | N | N | N | N | Note 1, 3, 4, 6, 7 | |||
| Abdominal | ||||||||||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| Small Organ (Specify**) | ||||||||||
| FetalImaging &Other | Neonatal Cephalic | |||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | N | N | N | N | N | N | Note 1, 3, 4, 6, 7 | |||
| Trans-vaginal | N | N | N | N | N | N | Note 1, 3, 4, 6, 7, | |||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | ||||||||||
| Musculo-skeletal (Superficial) | ||||||||||
| Intravascular | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Cardiac | Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheralvessel | Peripheral vessel | |||||||||
| Other (Specify***) | ||||||||||
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | ||||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | ||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | ||||||||||
| **Small organ-breast, thyroid, testes. | ||||||||||
| ***Other use includes Urology. | ||||||||||
| Note 1: Tissue Harmonic Imaging. | ||||||||||
| Note 2: Smart3D | ||||||||||
| Note 3:4D(Real-time 3D) | ||||||||||
| Note 4: iScape | ||||||||||
| Note5: TDI | ||||||||||
| Note6: Color M | ||||||||||
| Note7: Biopsy Guidance | ||||||||||
| Note8: Elastography | ||||||||||
| Note 9: Contrast imaging | ||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) |
| Intended | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |
|---|---|
| ---------- | ------------------------------------------------------------------------------------ |
{20}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | V11-3HE |
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track 1) | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | CombinedDoppler (specify) | Other (specify) |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | N | N | N | N | N | N | Note 1, 2, 4, 6, 7 | ||
| Abdominal | |||||||||
| Intra-operative (Specify*) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (Specify**) | |||||||||
| FetalImaging &Other | Neonatal Cephalic | ||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | N | N | N | N | N | N | Note 1, 2, 4, 6, 7 | ||
| Trans-vaginal | N | N | N | N | N | N | Note 1, 2, 4, 6, 7, 8 | ||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal (Conventional) | |||||||||
| Musculo-skeletal (Superficial) | |||||||||
| Intravascular | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Cardiac | Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Peripheralvessel | Peripheral vessel | ||||||||
| Other (Specify***) | N | N | N | N | N | N | Note 1, 2, 4, 6, 7 | ||
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | |||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | |||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | |||||||||
| **Small organ-breast, thyroid, testes. | |||||||||
| ***Other use includes Urology. | |||||||||
| Note 1: Tissue Harmonic Imaging. | |||||||||
| Note 2: Smart3D | |||||||||
| Note 3:4D(Real-time 3D) | |||||||||
| Note 4: iScape | |||||||||
| Note5: TDI | |||||||||
| Note6: Color M | |||||||||
| Note7: Biopsy Guidance | |||||||||
| Note8: Elastography | |||||||||
| Note 9: Contrast imaging | |||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | |||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
{21}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: C5-1E |
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | N | N | N | N | N | N | Note 1, 2, 4,6,7 | ||
| Abdominal | N | N | N | N | N | N | Note 1, 2, 4,6,7,9 | ||
| Intra-operative (Specify*) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | N | N | N | N | N | N | Note 1, 2, 4,6,7 | ||
| Small Organ (Specify**) | |||||||||
| FetalImaging &Other | Neonatal Cephalic | ||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal (Conventional) | N | N | N | N | N | N | Note 1, 2, 4,6,7 | ||
| Musculo-skeletal (Superficial) | |||||||||
| Intravascular | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Cardiac | Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Peripheralvessel | Peripheral vessel | N | N | N | N | N | N | Note 1, 2, 4,6,7 | |
| Other (Specify***) | |||||||||
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | |||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | |||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | |||||||||
| **Small organ-breast, thyroid, testes. | |||||||||
| ***Other use includes Urology. | |||||||||
| Note 1: Tissue Harmonic Imaging. | |||||||||
| Note 2: Smart3D | |||||||||
| Note 3:4D(Real-time 3D) | |||||||||
| Note 4: iScape | |||||||||
| Note5: TDI | |||||||||
| Note6: Color M | |||||||||
| Note7: Biopsy Guidance | |||||||||
| Note8: Elastography | |||||||||
| Note 9: Contrast imaging | |||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | |||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
{22}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | CB10-4E |
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| FetalImaging &Other | Small Organ (Specify**) | |||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | N | N | N | N | N | N | N | Note 1, 2, 4,6,7 | ||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | ||||||||||
| Musculo-skeletal (Superficial) | ||||||||||
| Intravascular | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Cardiac | Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheralvessel | Peripheral vessel | |||||||||
| Other (Specify***) | N | N | N | N | N | N | N | Note 1, 2, 4,6,7 | ||
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | ||||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | ||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | ||||||||||
| ** Small organ-breast, thyroid, testes. | ||||||||||
| ***Other use includes Urology. | ||||||||||
| Note 1: Tissue Harmonic Imaging. | ||||||||||
| Note 2: Smart3D | ||||||||||
| Note 3:4D(Real-time 3D) | ||||||||||
| Note 4: iScape | ||||||||||
| Note5: TDI | ||||||||||
| Note6: Color M | ||||||||||
| Note7: Biopsy Guidance | ||||||||||
| Note8: Elastography | ||||||||||
| Note 9: Contrast imaging | ||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | ||||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
{23}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | P10-4E |
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track 1) | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) |
| General(Track 1) | Ophthalmic | ||||||||
| Ophthalmic | |||||||||
| FetalImaging &Other | Fetal | ||||||||
| Abdominal | N | N | N | N | N | N | N | Note 1, 2,4,6 | |
| Intra-operative (Specify*) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | N | N | N | N | N | N | N | Note 1, 2,4,6 | |
| Small Organ (Specify**) | |||||||||
| Neonatal Cephalic | N | N | N | N | N | N | N | Note 1, 2,4,6 | |
| Adult Cephalic | N | N | N | N | N | N | N | Note 1, 2,4,6 | |
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal (Conventional) | |||||||||
| Musculo-skeletal (Superficial) | |||||||||
| Intravascular | |||||||||
| Cardiac | Cardiac Adult | N | N | N | N | N | N | N | Note 1, 2,4,5,6 |
| Cardiac Pediatric | N | N | N | N | N | N | N | Note 1, 2,4,5,6 | |
| Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Peripheralvessel | Peripheral vessel | ||||||||
| Other (Specify***) | |||||||||
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | |||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | |||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | |||||||||
| **Small organ-breast, thyroid, testes. | |||||||||
| ***Other use includes Urology. | |||||||||
| Note 1: Tissue Harmonic Imaging. | |||||||||
| Note 2: Smart3D | |||||||||
| Note 3:4D(Real-time 3D) | |||||||||
| Note 4: iScape | |||||||||
| Note5: TDI | |||||||||
| Note6: Color M | |||||||||
| Note7: Biopsy Guidance | |||||||||
| Note8: Elastography | |||||||||
| Note 9: Contrast imaging | |||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | |||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
{24}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | SP5-1E |
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | ||||||||||
| Abdominal | N | N | N | N | N | N | N | Note 1, 2,4, 6,7 | ||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | N | N | N | N | N | N | N | Note 1, 2,4, 6,7 | ||
| Small Organ (Specify**) | ||||||||||
| FetalImaging & | Neonatal Cephalic | |||||||||
| Other | Adult Cephalic | N | N | N | N | N | N | N | Note 1, 2,4, 6,7 | |
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | ||||||||||
| Musculo-skeletal (Superficial) | ||||||||||
| Intravascular | ||||||||||
| Cardiac Adult | N | N | N | N | N | N | N | Note 1, 2,4,5,6,7 | ||
| Cardiac Pediatric | N | N | N | N | N | N | N | Note 1, 2,4,5,6,7 | ||
| Cardiac | Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheral | Peripheral vessel | |||||||||
| vessel | Other (Specify***) | |||||||||
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | ||||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | ||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | ||||||||||
| **Small organ-breast, thyroid, testes. | ||||||||||
| ***Other use includes Urology. | ||||||||||
| Note 1: Tissue Harmonic Imaging. | ||||||||||
| Note 2: Smart3D | ||||||||||
| Note 3:4D(Real-time 3D) | ||||||||||
| Note 4: iScape | ||||||||||
| Note5: TDI | ||||||||||
| Note6: Color M | ||||||||||
| Note7: Biopsy Guidance | ||||||||||
| Note8: Elastography | ||||||||||
| Note 9: Contrast imaging | ||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | ||||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
{25}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | D6-2NE |
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal | N | N | N | N | N | N | Note 1, 3, 4, 6 | ||
| Abdominal | N | N | N | N | N | N | Note 1, 3, 4, 6 | ||
| Intra-operative (Specify*) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (Specify**) | |||||||||
| FetalImaging &Other | Neonatal Cephalic | ||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal (Conventional) | |||||||||
| Musculo-skeletal (Superficial) | |||||||||
| Intravascular | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Cardiac | Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Peripheralvessel | Peripheral vessel | ||||||||
| Other (Specify***) | |||||||||
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | |||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | |||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | |||||||||
| **Small organ-breast, thyroid, testes. | |||||||||
| ***Other use includes Urology. | |||||||||
| Note 1: Tissue Harmonic Imaging. | |||||||||
| Note 2: Smart3D | |||||||||
| Note 3:4D(Real-time 3D) | |||||||||
| Note 4: iScape | |||||||||
| Note5: TDI | |||||||||
| Note6: Color M | |||||||||
| Note7: Biopsy Guidance | |||||||||
| Note8: Elastography | |||||||||
| Note 9: Contrast imaging | |||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | |||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
1 Parisation
Hitchuru
Hearthern Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{26}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | D7-2E |
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | |||||||||
| Fetal | N | N | N | N | N | N | N | Note 1, 3, 4, 6 | ||
| Abdominal | N | N | N | N | N | N | N | Note 1, 3, 4, 6 | ||
| Intra-operative (Specify*) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | ||||||||||
| FetalImaging &Other | Small Organ (Specify**) | |||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | ||||||||||
| Musculo-skeletal (Superficial) | ||||||||||
| Intravascular | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Cardiac | Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Peripheralvessel | Peripheral vessel | |||||||||
| Other (Specify***) | ||||||||||
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | ||||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | ||||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | ||||||||||
| **Small organ-breast, thyroid, testes. | ||||||||||
| ***Other use includes Urology. | ||||||||||
| Note 1: Tissue Harmonic Imaging. | ||||||||||
| Note 2: Smart3D | ||||||||||
| Note 3:4D(Real-time 3D) | ||||||||||
| Note 4: iScape | ||||||||||
| Note5: TDI | ||||||||||
| Note6: Color M | ||||||||||
| Note7: Biopsy Guidance | ||||||||||
| Note8: Elastography | ||||||||||
| Note 9: Contrast imaging | ||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | ||||||||||
| Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) |
{27}------------------------------------------------
| System: | DC-70/DC-70T /DC-70 PRO/DC-70 EXP/DC-75/DC-78/DC-70S Diagnostic Ultrasound System |
|---|---|
| Transducer: | P7-3TE |
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(specify) | Other (specify) |
| Ophthalmic | Ophthalmic | ||||||||
| FetalImaging &Other | Fetal | ||||||||
| Abdominal | |||||||||
| Intra-operative (Specify*) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (Specify**) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal (Conventional) | |||||||||
| Musculo-skeletal (Superficial) | |||||||||
| Intravascular | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Cardiac | Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | N | N | N | N | N | N | N | Note 1, 5,6 | |
| Intra-cardiac | |||||||||
| Peripheralvessel | Peripheral vessel | ||||||||
| Other (Specify***) | |||||||||
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | |||||||||
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | |||||||||
| *Intraoperative includes abdominal, thoracic, and vascular etc. | |||||||||
| **Small organ-breast, thyroid, testes. | |||||||||
| ***Other use includes Urology. | |||||||||
| Note 1: Tissue Harmonic Imaging. | |||||||||
| Note 2: Smart3D | |||||||||
| Note 3:4D(Real-time 3D) | |||||||||
| Note 4: iScape | |||||||||
| Note5: TDI | |||||||||
| Note6: Color M | |||||||||
| Note7: Biopsy Guidance | |||||||||
| Note8: Elastography | |||||||||
| Note 9: Contrast imaging | |||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) |
{28}------------------------------------------------
510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92(c).
1. Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China Tel: +86 755 8188 5640 Fax: +86 755 2658 2680
Contact Person:
Yang Zhaohui Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Date Prepared: December 15, 2016
- Device Name: DC-70/DC-70T /DC-70 Pro/DC-70 Exp/DC-75/DC-78/DC-70S Diagnostic Ultrasound System
Classification
Regulatory Class: II
Review Category: Tier II
- 21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System (IYN)
- 21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO)
- 21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX)
- 21 CFR 892.2050 Picture Archiving and Communications System(LLZ)
{29}------------------------------------------------
3. Device Description:
DC-70/DC-70T /DC-70 Pro/DC-70 Exp/DC-75/DC-78/DC-70S is a software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound data in B-Mode, M-Mode, PW-Mode, CW-Mode, Color-Mode , Power/Dirpower Mode, THI, iScape,4D, Smart3D,TDI, Color M, Elastography, Biopsy Guidance, Contrast imaging or the combined mode (i.e. B/M-Mode, B/PW-mode, B/PW/Color).
This system is a Track 3 device that employs an array of probes that include linear array and convex array with a frequency range of approximately 2.0 MHz to 10.0 MHz.
4. Intended Use:
The DC-70/ DC-70T / DC-70 Pro/ DC-70 Exp/ DC-75/ DC-78/ DC-70S Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ(breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-vaginal, Trans-esoph(Cardiac), musculo-skeletal(conventional, superficial), adult and pediatric cardiac, peripheral vessel and urology exams.
5. Summary of Modifications
-
New Added Models:
DC-75, DC-78, DC-70S -
New Added Transducers:
D7-2E D6-2NE C5-1E CB10-4E P7-3TE P10-4E
{30}------------------------------------------------
SPS-IE V11-3HE L9-3E DE11-3E
New Added Needle-Guided Bracket: ●
NGB-022, NGB-025, NGB-027, NGB-034
● Main Added Features:
Contrast Imaging Smart FLC Tissue Tracking Elastography to V11-3E, V11-3BE, V11-3HE
6. Comparison with Predicate Devices:
DC-70/DC-70T /DC-70 Pro/DC-70 Exp/DC-75/DC-78/DC-70S Diagnostic
Ultrasound System is comparable with and substantially equivalent to these predicate devices:
| PredicateDevice | Manufacturer | Model | 510(k) ControlNumber |
|---|---|---|---|
| 1 | Mindray | DC-70 | K150204 |
| 2 | Mindray | DC-8 | K150080 |
| 3 | Mindray | Resona 7 | K162267 |
| 4 | Mindray | DC-60 | K152545 |
| 5 | Philips | EPIQ 5 | K160807 |
DC-70/DC-70T /DC-70 Pro/DC-70 Exp/DC-75/DC-78/DC-70S has the same technological characteristics, is comparable in key safety and effectiveness features, and has the same intended uses and basic operating modes as the predicate devices. All systems transmit ultrasonic energy into patients and perform post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures and
{31}------------------------------------------------
flow, and calculations.
- . Subject device DC-70/DC-70T /DC-70 Pro/DC-70 Exp/DC-75/DC-78/DC-70S has the same intended uses as the predicated device DC-70(K150204)
| Subject Device | Predicate device |
|---|---|
| DC-70/DC-70T /DC-70 PRO/DC-70 EXP | DC-70(K150204) |
| The DC-70/ DC-70T / DC-70 Pro/ DC-70 Exp/ DC-75/ DC-78/ DC-70S DiagnosticUltrasound System is applicable for adults,pregnant women, pediatric patients andneonates. It is intended for use in fetal,abdominal, pediatric, small organ(breast,thyroid, testes), neonatal and adult cephalic,trans-rectal, trans-vaginal, Trans-esoph(Cardiac), musculo-skeletal (conventional,superficial), adult and pediatric cardiac,peripheral vessel and urology exams. | The DC-70/DC-70T /DC-70 Pro/DC-70 Expdiagnostic ultrasound system is applicable foradults, pregnant women, pediatric patients andneonates. It is intended for use in fetal,abdominal, pediatric, small organ (breast,thyroid, testes), neonatal cephalic, adultcephalic, trans-rectal, trans-vaginal,musculo-skeletal (conventional, superficial),cardiac adult, cardiac pediatric, peripheralvessel, urology exams. |
- The acoustic power levels of DC-70/ DC-70T/ DC-70 PRO/ DC-70 EXP / DC-75/ ● DC-78/ DC-70S are below the limits of FDA, which is the same as the predicated device DC-70(K150204).
- DC-70/ DC-70T/ DC-70 PRO/ DC-70 EXP / DC-75/ DC-78/ DC-70S is designed in compliance with the FDA recognized electrical and physical safety standard, which is the same as the predicated device DC-70(K150204).
- . DC-70/ DC-70T/ DC-70 PRO/ DC-70 EXP / DC-75/ DC-78/ DC-70S has similar probes as the predicated device:
| Subject device | Predicated device |
|---|---|
| DC-70/ DC-70T/ DC-70 PRO/DC-70 EXP / DC-75/ DC-78/DC-70S | |
| D7-2E | D7-2E &DC-60(K152545) |
| D6-2NE | D6-2E &DC-70(K150204) |
{32}------------------------------------------------
| C5-1E | C5-1U& Resona 7 (K162267) |
|---|---|
| CB10-4E | CB10-4E&DC-8(K150080) |
| P7-3TE | P7-3TE&DC-8(K150080) |
| P10-4E | P10-4E&&DC-8(K150080) |
| SP5-1E | SP5-1E &DC-8(K150080) |
| V11-3HE | V11-3WE &DC-8(K150080) |
| L9-3E | L9-3U& Resona 7(K162267) |
| DE11-3E | DE11-3U & Resona 7 (K162267) |
7. Non-clinical Tests:
DC-70/ DC-70T/ DC-70 PRO/ DC-70 EXP / DC-75/ DC-78/ DC-70S Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been designed to conform with applicable medical safety standards. This device has been tested and evaluated under the following standards:
- AAMI/ANSI ES60601-1: Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- . AAMI/ANSI/IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 3)
- . IEC 60601-2-37: Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- IEC 62304: Medical device software - Software life cycle processes
- IEC:62366:Medical devices application of usability engineering to medical devices ●
- IEC 60601-1-6: medical electrical equipment - part 1-6: general requirements for basic safety and essential performance - collateral standard: usability.
- ISO14971: Medical devices - Application of risk management to medical devices
- UD 2: Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision 3
- . UD 3 Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
{33}------------------------------------------------
- . ISO 10993-1 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process
These non-clinical tests relied on in this premarket notification submission can support the determination of substantial equivalence of the subject device.
8. Clinical Studies
Not applicable. The subject of this submission, DC-70/ DC-70T/ DC-70 PRO/ DC-70 EXP / DC-75/ DC-78/ DC-70S Diagnostic Ultrasound System, does not require clinical studies to support substantial equivalence.
Conclusion:
Intended uses and other key features are consistent with traditional clinical practices. FDA guidelines and established methods of patient examination. The design, development and quality process of the manufacturer confirms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards. Therefore, the DC-70/ DC-70T/ DC-70 PRO/ DC-70 EXP / DC-75/ DC-78/ DC-70S Diagnostic Ultrasound System is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
N/A