K Number
K182545

Validate with FDA (Live)

Date Cleared
2019-01-11

(116 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Tamper Evident Cap with Male Luer Lock is indicated for use as a cap for female Luer ports on medical devices such as manifolds, stopcocks or set.

Device Description

The Tamper Evident Cap with Male Luer Lock (MTEC) is a sterile, single use device which will be used to cover female Luer ports on medical devices and provide evidence of access. MTEC is provided in two packaging configurations; a 10-up tray and single unit blisters. Both configurations are offered in multiple colors and all the components used in the manufacture of MTEC are comprised of polystyrene and polypropylene resin.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text, while noting limitations in detail for device performance and several other categories:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
ISO 80369-7:2016 (Small-bore connectors for liquids and gases in healthcare applications -- Part 7: Connectors for intravascular or hypodermic applications)The MTEC device meets the bench testing requirements of ISO 80369-7:2016. Specifically, it meets requirements for: - Separation Force - Unscrewing torque - Ease of assembly - Stress cracking - Liquid leakage test - Resistance to overriding
Shelf Life EvaluationDesign verification of the shelf life was conducted and implied to meet criteria. The document states "All test results meet the acceptance criteria."
Sterilization Product Adoption (AAMI TIR 28:2009/(R)2013)Implied to meet criteria: "All test results meet the acceptance criteria."
Packaging (ISO 11607-1:2006)Implied to meet criteria: "All test results meet the acceptance criteria."
Sterility (ISO 11135:2014)Implied to meet criteria: "All test results meet the acceptance criteria."
Residuals (ISO 10993-7:2008)Implied to meet criteria: "All test results meet the acceptance criteria."
Biocompatibility (ISO 10993-1, relevant FDA Guidance)Biocompatibility testing was conducted according to the cited guidance and standards for an External Communicating Device, Blood Path, Indirect Contact, Limited Duration, and implied to meet criteria: "All test results meet the acceptance criteria."
Tamper Evident Design FeatureThe MTEC device's tamper evident design feature was tested and successfully passed all internal specifications.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size for the test set used for any of the bench tests (e.g., ISO 80369-7:2016, tamper evident feature testing).
The data provenance is from non-clinical bench testing performed by International Medical Industries, Inc. (the manufacturer). The country of origin for the data is not explicitly stated but implies testing was done internally by the manufacturer. The data is prospective in the sense that it was generated specifically for this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as the studies are non-clinical bench tests comparing device performance to established international standards and internal specifications, not studies involving human interpretation or ground truth establishment by experts in a clinical context.

4. Adjudication Method for the Test Set

This information is not applicable as the studies are non-clinical bench tests against objective criteria, not studies requiring expert adjudication of results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This submission is for a medical device (tamper evident cap) and relies on non-clinical bench testing for substantial equivalence, not clinical effectiveness studies with human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable as the device is a physical medical component, not an algorithm or AI system.

7. The Type of Ground Truth Used

The "ground truth" for the non-clinical bench tests is established by:

  • International Standards: e.g., ISO 80369-7:2016, ISO 11607-1:2006, ISO 11135:2014, ISO 10993-7:2008.
  • Internal Specifications: For the tamper-evident design feature.
  • FDA Recognized Guidance Documents: For biocompatibility testing (e.g., ISO 10993-1 and specific FDA guidance for intravascular administration sets).

These are objective, quantitative criteria rather than expert consensus, pathology, or outcomes data.

8. The Sample Size for the Training Set

This question is not applicable as the device is a physical medical component, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as the device is a physical medical component, not an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 11, 2019

International Medical Industries, Inc. Kathleen Lavender Director, Quality and Compliance 2981 Gateway Drive Pompano Beach, Florida 33069

Re: K182545

Trade/Device Name: Tamper Evident Cap with Male Luer Lock Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: December 10, 2018 Received: December 12, 2018

Dear Kathleen Lavender:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Geeta K. Pamidimukkala -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182545

Device Name

Tamper Evident Cap with Male Luer Lock

Indications for Use (Describe)

The Tamper Evident Cap with Male Luer Lock is indicated for use as a cap for female Luer ports on medical devices such as manifolds, stopcocks or set.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary: K182545

Manufacturer:International Medical Industries, Inc.
Address:2981 Gateway DrivePompano Beach, Florida 33069
Corresponding official:Kathleen LavenderDirector of Quality and Compliance
Telephone Number:954.917.9570 x 283
Email:klavender@imiweb.com
Summary Date:January 10, 2019
Trade Name:Tamper Evident Cap with Male Luer Lock
Common or Usual Name:I.V. Administration Set
Regulation Number:21 CFR 880.5440
Regulation Name:Intravascular Administration Set
Product Code:FPA
Class:II
Panel:General Hospital
Primary Predicate Device:Dual Luer Lock Cap (K101385)
Secondary Predicate Device:Tamper Evident Cap with Luer Lock (K861276)

Device Description:

The Tamper Evident Cap with Male Luer Lock (MTEC) is a sterile, single use device which will be used to cover female Luer ports on medical devices and provide evidence of access. MTEC is provided in two packaging configurations; a 10-up tray and single unit blisters. Both configurations are offered in multiple colors and all the components used in the manufacture of MTEC are comprised of polystyrene and polypropylene resin.

{4}------------------------------------------------

Feature of the device:Device:K182545Tamper Evident Capwith male Luer LockPrimary Predicate:K101385Baxter Dual LuerLock CapSecondary Predicate:K861276Tamper Evident Capwith Luer LockDiscussion/Comment
Product CodeFPAFPAFPANo Change
ClassificationClass IIClass IIClass IINo Change
Number of usesSingle use Rx onlySingle use Rx onlySingle use Rx onlyNo Change
MaterialsDevice is an injectionmolded polypropylenecomponent.Device is an injectionmolded polypropylenecomponent.Device is an injectionmolded polypropylenecomponent.Same: No Change
Sterilization MethodEthylene OxideEthylene OxideEthylene OxideSame: No Change
BiocompatibilityBiocompatibilityassessment of the DualLuer Lock Cap hasbeen conducted basedon ISO 10993-1 -Biological Evaluationof Medical Devices.Biocompatibilityassessment of the DualLuer Lock Cap hasbeen conducted basedon ISO 10993-1 -Biological Evaluationof Medical Devices.Biocompatibilityassessment of the DualLuer Lock Cap has beenconducted based on ISO10993-1 - BiologicalEvaluation of MedicalDevices.Same: No Change
Indication for Use:The Tamper EvidentCap with Male LuerLock is indicated foruse as a cap for femaleLuer ports on medicaldevices such asmanifolds, stopcocksor sets.The Dual Luer LockCap is indicated foruse as a cap for maleor female Luer portson medical devicessuch as manifolds,stopcocks or sets.The Tamper EvidentCap with Female LuerLock is indicated for useas a cap for male Luerports on medical devicesDevice is pendingpremarket notification.Predicate iscommercially availableas Baxter Dual LuerLock CapNot subject to recallThe primary differenceis that the primarypredicate device hasboth male and femaleLuer connection pointswhereas the proposeddevice only has a maleLuer lock connectionpoint.The primary differencebetween the secondarypredicate device and theproposed device is thatthe secondary predicateonly has a female Luerlock connection point.
Design Feature:TamperEvidenceDevice is designed tocover an open Luerport after the Luer porthas been accessed andwill provide evidenceof tampering ifremoved afterDevice is designed tocover an open Luerport after the Luer porthas been accessed.Device is designed tocover an open Luer portafter the Luer port hasbeen accessed and willprovide evidence oftampering if removedafter installation.The proposed product isdesigned with a primaryfunction of a standardmale Luer lock cap witha secondary function asa tamper evident(evidence of access)

{5}------------------------------------------------

Technological Characteristics:

The MTEC device is an assembly of injection molded components with a polypropylene Luer lock tip cap. The intended use and function of the proposed device is identical to the primary predicate device with respect to functionality. Both the proposed device (tip cap) and the primary predicate device are comprised of polypropylene resin. There are two design differences between the proposed device and the primary predicate device. The primary difference is that the primary predicate device has both male and female Luer connection points whereas the proposed device only has a male Luer lock connection point. The second difference is that the proposed device has a tamper evident design feature. This design feature is identical to the secondary predicate device.

Performance Data:

Bench testing was performed to demonstrate that the proposed MTEC device is substantially equivalent to both the primary predicate and secondary predicate devices.

Test NameTest Description
ISO 80369-7:2016Design verification to ISO 80369-7:2016 Small-bore connectors forliquids and gases in healthcare applications -- Part 7: Connectors forintravascular or hypodermic applications
Shelf Life EvaluationDesign verification of the shelf life.
Sterilization ProductAdoptionAAMI TIR 28:2009/(R)2013 Product adoption and processequivalence for ethylene oxide sterilization
PackagingISO 11607-1:2006 Packaging for terminally sterilized medicaldevices -- Part 1: Requirements for materials, sterile barrier systemsand packaging systems
SterilityISO 11135:2014 Sterilization of health-care products -- Ethyleneoxide -- Requirements for the development, validation and routinecontrol of a sterilization process for medical devices
ResidualsISO 10993-7:2008 Biological evaluation of medical devices -- Part 7:Ethylene oxide sterilization residuals
BiocompatibilityISO 10993-1, “Biological evaluation of medical devices – Part 1:Evaluation and testing within a risk management process”; Guidancefor Industry and Food and Drug Administration Staff, Jun 16, 2016and "Guidance for Industry and FDA Staff – IntravascularAdministration Sets Premarket Notification Submissions [510(k)],”July 11, 2008, as recognized by the FDA.

{6}------------------------------------------------

Discussion of Non-Clinical Testing:

International Medical Industries conducts risk analyses and design verification tests that are based on the result of these analyses. All test results meet the acceptance criteria and support that the device is appropriately designed for the intended use.

The MTEC device meets the bench testing requirements of ISO 80369-7:2016 Small-bore connectors for liquids and gases in healthcare applications -- Part 7: Connectors for intravascular or hypodermic applications. The primary predicate device was tested to ISO 594-2:1998 which has since been replaced by ISO 80369-7:2016.

Compliance with ISO 80369-7:2016 demonstrates that The MTEC device meets the Bench performance testing requirements as follows:

    • Separation Force
    • Unscrewing torque
    • Ease of assembly * Stress cracking
    • Liquid leakage test
    • Resistance to overriding

Proposed device vs. primary predicate device

The MTEC device meets the bench testing requirements of ISO 80369-7:2016 Small-bore connectors for liguids and gases in healthcare applications -- Part 7: Connectors for intravascular or hypodermic applications. The primary predicate device was tested to ISO 594-2:1998 which has since been replaced by ISO 80369-7:2016. Compliance with ISO 80369-7:2016 demonstrates that the proposed device is substantially equivalent to the primary predicate device K101385.

Proposed device vs. secondary predicate device

The MTEC device's tamper evident design feature was tested and successfully passed all internal specifications. It is substantially equivalent to the secondary predicate device K861276.

{7}------------------------------------------------

Biocompatibility Testing

Biocompatibility testing of the Tamper Evident Cap with Male Luer Lock was conducted per Guidance Document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"; Guidance for Industry and Food and Drug Administration Staff, Jun 16, 2016 and "Guidance for Industry and FDA Staff - Intravascular Administration Sets Premarket Notification Submissions [510(k)]," July 11, 2008, as recognized by the FDA. Biocompatibility testing was conducted in accordance with the cited guidance and standards as required for an External Communicating Device, Blood Path, Indirect Contact, Limited Duration.

Conclusion:

Review of the in-vitro performance test data as well as comparison of the device classification, indications for use, operating principle, technological characteristics, sterility, and biocompatibility demonstrate that the subject device, Tamper Evident Cap with Male Luer Lock (MTEC) is substantially equivalent to the primary predicate device, the Baxter Dual Luer Lock Cap (K101385). Any technological differences between the subject and the predicate devices do not raise any issues of safety and effectiveness.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.