K Number
K101385

Validate with FDA (Live)

Date Cleared
2010-06-22

(35 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dual Luer Lock Cap is indicated for use as a cap for male or female Luer ports on medical devices such as manifolds, stopcocks or sets.

Device Description

The subject of this submission is a sterile Dual Luer Lock Cap which will be used to cover male or female Luer ports on medical devices. The proposed Dual Luer Lock Cap will replace the existing Dual Luer Lock Cap and will be sold as a cap to cover an open Luer port after the Luer port has been accessed and is no longer in use. The Dual Luer Lock Cap consists of an integrated design with a male Luer lock connection on one end and a female Luer lock connection on the other end.

AI/ML Overview

Below is the requested information about the acceptance criteria and the study that proves the device meets the acceptance criteria, based on the provided text.

Acceptance Criteria and Device Performance Study

This device, the Dual Luer Lock Cap, is an IV Administration Set component. The performance testing focused on meeting recognized international standards for Luer fittings.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Test)StandardReported Device Performance
Air leakageISO 594-1:1986Met (implied by "All test results meet the acceptance criteria")
Separation ForceISO 594-2:1986Met (implied by "All test results meet the acceptance criteria")
Unscrewing torqueISO 594-2:1986Met (implied by "All test results meet the acceptance criteria")
Ease of assemblyISO 594-1:1986Met (implied by "All test results meet the acceptance criteria")
Stress crackingISO 594-1:1986Met (implied by "All test results meet the acceptance criteria")
Liquid leakage testISO 594-1:1986Met (implied by "All test results meet the acceptance criteria")
Luer gaugingISO 594-1:1986Met (implied by "All test results meet the acceptance criteria")
Resistance to overridingISO 594-2:1986Met (implied by "All test results meet the acceptance criteria")
BiocompatibilityISO 10993-1Shown to be biocompatible and appropriate for intended use (prolonged duration, external communicating, indirect blood path testing)

Note: The document states that "All test results meet the acceptance criteria," indicating successful performance without providing specific quantitative results for each test.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for the performance tests (e.g., how many Dual Luer Lock Caps were tested for each criterion). The data provenance is also not specified in terms of country of origin or whether it was retrospective or prospective, beyond stating that "Baxter Healthcare Corporation conducts risk analyses and design verification tests."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This section is not applicable. The device is a physical medical component, not an AI/diagnostic software. Ground truth for its performance is established by objective engineering tests against international standards (ISO 594-1 and ISO 594-2), not expert consensus on interpretations.

4. Adjudication Method

This section is not applicable, as there is no expert interpretation or consensus-based ground truth for this type of device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

This section is not applicable. The device is a physical medical component, not an AI system intended to assist human readers in interpreting cases. Therefore, a MRMC study comparing human readers with and without AI assistance was not performed.

6. Standalone Performance Study

This section is applicable in the sense that the device's performance was evaluated in a standalone manner (i.e., the physical device itself was subjected to engineering tests without human intervention in the performance of the device's function). The "Performance Data" section directly addresses the results of these standalone tests against ISO standards.

7. Type of Ground Truth Used

The ground truth used for performance validation was based on objective engineering standards ("Bench performance testing requirement according to ISO 594-1:1986" and "ISO 594-2:1986") and biocompatibility standards ("ISO 10993-1").

8. Sample Size for the Training Set

This section is not applicable. The device is a physical medical component and does not utilize a training set in the context of machine learning or AI.

9. How the Ground Truth for the Training Set Was Established

This section is not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

K101385

5. 510(k) SUMMARY

June 14, 2010

OWNER:

JUN 2 2 2 2010

Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015

CONTACT PERSON:

Nanette Hedden Senior Manager, Global Regulatory Affairs 1620 Waukegan Road McGaw Park, IL, 60085 Telephone: (847) 270-4871 Fax: (847) 785-5116

DEVICE NAME:

Trade name: Dual Luer Lock Cap

Common name: IV Administration Set

Classification name: IV Administration Set, 21 CFR 880.5440, FPA, Class II

PREDICATE DEVICE:

DeviceCompanyPrevious510(k)Clearance date
Dual Luer Lock CapBaxterHealthcareK981318April 22, 1998

Table 5-1. Previous 510(k)s

{1}------------------------------------------------

DESCRIPTION OF THE DEVICE:

The subject of this submission is a sterile Dual Luer Lock Cap which will be used to cover male or female Luer ports on medical devices. The proposed Dual Luer Lock Cap will replace the existing Dual Luer Lock Cap and will be sold as a cap to cover an open Luer port after the Luer port has been accessed and is no longer in use. The Dual Luer Lock Cap consists of an integrated design with a male Luer lock connection on one end and a female Luer lock connection on the other end.

STATEMENT OF INTENDED USE:

The Dual Luer Lock Cap is indicated for use as a cap for male or female Luer ports on medical devices such as manifolds, stopcocks or sets.

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:

The Dual Luer Lock Cap is an injection molded, polypropylene component. The intended use and function of the proposed Dual Luer Lock Cap is identical to the predicate device.

Performance Data: The Dual Luer Lock Cap meets the Bench performance testing requirement according to ISO 594-1:1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements and ISO 594-2:1986 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings Luer requirements. The Dual Luer Lock Cap is a cap for male or female Luer ports on medical devices.

Biocompatibility: Biocompatibility assessment of the Dual Luer Lock Cap has been conducted based on ISO 10993-1 - Biological Evaluation of Medical Devices. Based upon the results of this prolonged duration, external communicating, indirect blood path testing the material used to fabricate the Dual Luer Lock Cap has been shown to be biocompatible and appropriate for its intended use.

DISCUSSION OF NONCLINICAL TESTS:

Baxter Healthcare Corporation conducts risk analyses and design verification tests are based on the result of these analyses. All test results meet the acceptance criteria and support that the device is appropriately designed for the intended use. The Dual Luer Lock Cap meets the Bench performance testing requirement according to ISO 594-1 and 594-2 specified as follows:

{2}------------------------------------------------

K101385

Special 510(k) Premarket Notification Dual Luer Lock Cap

  • Air leakage
  • · Separation Force
  • Unscrewing torque
  • Ease of assembly
  • Stress cracking
  • Liquid leakage test
  • · Luer gauging
  • · Resistance to overriding

CONCLUSION:

The Dual Luer Lock Cap is substantially equivalent to Baxter's current legally marketed Dual Luer Lock Cap cleared April 22, 1998 (K981318).

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol featuring three abstract human figures or shapes, arranged in a vertical stack.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUN 2 2 2010

Ms. Nanette Hedden Senior Manager, Global Regulatory Affairs Baxter Healthcare Corporation 1620 Waukegan Road McGaw Park, Illinois 60085

Re: K101385

Trade/Device Name: Dual Luer Lock Cap, Model 2C650 Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: June 14, 2010 Received: June 16, 2010

Dear Ms. Hedden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Ms. Hedden

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.goy/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours

Susan Penrose

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices

Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): K (0) 385

Device Name: Dual Luer Lock Cap

Indications for Use:

The Dual Luer Lock Cap is indicated for use as a cap for male or female Luer ports on medical devices such as manifolds, stopcocks or sets.

Prescription Use _________________

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rell C. Angan 6/22/2010
(Division Sign-Off)

(Division Sign-Off) Division of Anesthesiology, General Hospital 'nfection Control, Dental Devices

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

  • ' - ' - ' ' ' } Number:

510(k) Number:

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.