K Number
K182476

Validate with FDA (Live)

Date Cleared
2019-12-18

(464 days)

Product Code
Regulation Number
878.4750
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Powered Articulating Endosopic Linear Cutter Stapler & Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general abdominal, gynecologic, thoracic, and pediatric surgical procedures.

Device Description

The Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads are sterile, single patient use instruments that simultaneously cut and staple tissue through a battery powered firing system. The Power Stapler and Reloads are sterilized by irradiation. The instruments deliver six staggered rows of staples, three on either side of the cut line. The instruments are available in three shaft lengths: compact (192±20mm), regular (252±20mm) and long (352±20mm). The shaft can rotate freely in both directions and incorporates an articulation mechanism, which enables the distal portion of the shaft to pivot to facilitate lateral access to the operative site.

AI/ML Overview

This document describes a 510(k) premarket notification for a medical device, the "Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads." The core of a 510(k) submission is to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device. This process primarily relies on non-clinical performance testing and comparison of technological characteristics, rather than extensive clinical efficacy studies in the same way a PMA (Pre-Market Approval) process would require.

Therefore, the input document does not contain the information requested in your prompt regarding acceptance criteria and studies proving the device meets them in the context of diagnostic accuracy or comparative effectiveness with human readers. This is because the device described is a surgical instrument (a stapler), not a diagnostic AI system.

However, I can extract information related to performance testing that serves as "acceptance criteria" for this type of device to demonstrate substantial equivalence.

Here's the information derived from the document, tailored as much as possible to your prompt's structure, but with the understanding that it's for a surgical stapler and not an AI diagnostic tool:


1. A table of acceptance criteria and the reported device performance

The document lists "Non-Clinical Tests," "Biocompatibility Tests," "Electrical Safety Test," and "EMC Test" that the device "passed." For a 510(k) submission of a surgical stapler, "passing" these tests indicates that the device met predetermined performance specifications (which serve as acceptance criteria) established by the manufacturer and recognized standards. Specific numerical acceptance criteria are generally detailed in the full test reports, which are summarized here.

Acceptance Criteria CategorySpecific TestReported Device Performance
Non-Clinical PerformanceAppearancePassed
FlexibilityPassed
Assembling capacityPassed
HardnessPassed
Anastomosis and cutting performancePassed
Resistance to PressurePassed
Safety Device (e.g., empty-reload safety protection)Passed (and feature explicitly mentioned)
SharpnessPassed
Surface RoughnessPassed
Package SealPassed (ASTM F 88/F-15)
Rotation and ArticulationPassed
Firing Electric CurrentPassed
Battery VoltagePassed
Motor Rotate SpeedPassed
SizePassed
Sterility TestPassed (Sterilized by Irradiation, SAL 10^-6)
Accelerated Aging TestPassed (ASTM F1980-16)
Dye Penetration TestPassed (ASTM F1929-15, ASTM F3039-15)
Vacuum Leak TestPassed (ASTM D3078-2002(2013))
Packaging Resistance Bacteria Performance TestPassed (DIN 58953-6-2010)
BiocompatibilityIn Vitro Cytotoxicity TestPassed (ISO 10993-5: 2009)
Skin Sensitization TestPassed (ISO 10993-10:2010)
Intracutaneous reactivity TestPassed (ISO 10993-10:2010)
Acute Systemic ToxicityPassed (ISO 10993-11:2006)
Pyrogen TestPassed (ISO 10993-11:2006)
Subchronic Systemic ToxicityPassed (ISO 10993-11:2006)
Ames testPassed (ISO 10993-3: 2014)
In Vitro Mammalian Chromosome Aberration TestPassed (ISO 10993-3:2014)
In Vivo Mammalian Erythrocyte Micronucleus TestPassed (ISO 10993-3:2014)
Muscle Implantation TestPassed (ISO 10993-6: 2016)
Electrical SafetyElectrical Safety TestPassed (AAMI / ANSI ES60601-1:2005/(R)2012 and standards)
Electromagnetic Comp.EMC TestPassed (IEC60601-1-2:2014)

2. Sample size used for the test set and the data provenance

The document does not specify exact sample sizes for each non-clinical test. For devices like this, sample sizes would typically be determined by statistical methods or industry standards for each specific test (e.g., number of stapler firings, number of animals for biocompatibility tests). There is no mention of "data provenance" in terms of country of origin or retrospective/prospective as this is not a diagnostic AI study. The "test set" here refers to the physical devices and materials used in the non-clinical and biocompatibility testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable to the provided document. The device is a surgical stapler, not a diagnostic AI system that requires expert interpretation for ground truth establishment. The "ground truth" for stapler performance would be objective measurements (e.g., staple formation integrity, seal strength, cutting effectiveness) against predefined engineering specifications and relevant ISO/ASTM standards.

4. Adjudication method for the test set

This question is not applicable to the provided document. Adjudication methods (like 2+1 or 3+1) are used in diagnostic studies to resolve discrepancies in expert interpretations to establish a ground truth. For a surgical stapler, performance is evaluated against objective, measurable criteria, not subjective interpretation.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable to the provided document. An MRMC study is relevant for diagnostic AI tools, not surgical instruments. The device's performance is not about human 'readers' or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable to the provided document. This device is a powered surgical instrument; there is no "algorithm only" or "human-in-the-loop" performance in the context of diagnostic assessment. It functions as a tool operated by a surgeon.

7. The type of ground truth used

For this surgical stapler, the "ground truth" is established through:

  • Engineering Specifications and Performance Standards: The device's physical dimensions, electrical characteristics, firing mechanism, and output (staple form, cut line) are compared against pre-defined engineering requirements and relevant international standards (e.g., ISO, ASTM, AAMI).
  • Biocompatibility Standards: Materials are tested according to ISO 10993 series for their safety in contact with the body.
  • Sterility Assurance: Sterilization methods are validated to achieve a specified Sterility Assurance Level (SAL) (10^-6 in this case).
  • Functional Testing: "Anastomosis and cutting performance," "resistance to pressure," "sharpness," etc., are assessed through laboratory and potentially ex-vivo tissue testing against established benchmarks for similar devices.

8. The sample size for the training set

This question is not applicable to the provided document. "Training set" refers to data used to train an AI model. This document describes a physical surgical device, not an AI system.

9. How the ground truth for the training set was established

This question is not applicable to the provided document, as there is no "training set" for an AI model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

December 18, 2019

Fengh Medical Co., Ltd. Jiao Liu RA Manager D3 No.6 Dongsheng West Road, Jiangyin National High-tech Zone Jiangyin, 214437 Cn

Re: K182476

Trade/Device Name: Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: November 13, 2019 Received: November 18, 2019

Dear Jiao Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K182476

Device Name

Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads, Model include 30mm series and 60mm series

Indications for Use (Describe)

The Disposable Powered Articulating Endosopic Linear Cutter Stapler & Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general abdominal, gynecologic, thoracic, and pediatric surgical procedures.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

(As Required by 21 CFR 807.92)

1. Date Prepared [21 CFR 807.92(a)(1)]

Jan 29th, 2019

2. Submitter's Information [21 CFR 807.92(a)(1)]

Company Name:Fengh Medical Co., Ltd.
Company Address:D3 No.6 Dongsheng West Road, Jiangyin National High-tech Zone, 214437 Jiangsu, China
Contact Person:Jiao Liu
Phone:0086-13815132106
Email:Jiao.liu@fenghmedical.com

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

Trade Name:Disposable Powered Articulating Endoscopic Linear CutterStapler & Reloads, Model include 30mm series, 45mmseries and 60mm series
Common Name:Implantable staple
Product Code:GDW
Regulation Number:21 CFR 878.4750
Device Class:II

4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[

The identification of predicates within this submission is as follow:

Manufacturer:Ethicon Endo-Surgery, LLC
Trade Name:Echelon Flex Powered Articulating Endoscopic LinearCutters
Common Name:Cutter/Stapler
Product Code:GDW
Classification Name:Staple, Implantable; Stapler, Surgical
Regulation Number:21 CFR 878.4750
Classification:Class II

Page 1 of 8

{4}------------------------------------------------

K130653 (K130653 is a modification to their original FDA 510 (k) #: product cleared under K110385)

5. Description of the Device [21 CFR 807.92(a)(4)]

The Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads are sterile, single patient use instruments that simultaneously cut and staple tissue through a battery powered firing system. The Power Stapler and Reloads are sterilized by irradiation. The instruments deliver six staggered rows of staples, three on either side of the cut line.

The instruments are available in three shaft lengths: compact (192±20mm), regular (252±20mm) and long (352±20mm). The shaft can rotate freely in both directions and incorporates an articulation mechanism, which enables the distal portion of the shaft to pivot to facilitate lateral access to the operative site.

Model 30mm series include FSMS30, FSMM30, FSML30; 45mm series include FSMS45, FSMM45, FSML45; 60mm series include FSMS60, FSMM60, FSML60. The difference between the three series lies in the different anatomized lengths.

There are total 18 models of reloads within the proposed device, three stapler line length: 35.2±2.0mm,49.3±2.0mm and 61.3±2.0mm; six staple heights: Gray (0.75mm), White (1.0mm), Blue (1.5mm), Gold (1.8mm), Green (2.0mm), Black (2.3mm).

Power type: internal power supply; DC voltage 12V, rated power 40W Applied parts: Type CF Mode of operation: Non-Continuous Operation

Degree of safety when used in the presence of flammable anesthetic gas mixed with flammable anesthetic gas or nitrous oxide mixed with air: Flammable anesthetic gas that cannot be mixed with flammable anesthetic gas or nitrous oxide mixed with air .

The Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads is resistant to water ingress. Grade of waterproof: Rotating knob to the end of tubular shaft (including rotating knob) is rated IPX4; the Body is rated IPX0.

Page 2 of 8

{5}------------------------------------------------

Image /page/5/Figure/1 description: The image shows a medical instrument with two labels, IPX4 and IPX0, indicating different sections of the device. IPX4 is positioned above the long, slender part of the instrument, while IPX0 is located above the handle. The instrument appears to be a type of surgical tool, possibly used for grasping or cutting tissue during medical procedures. The design includes a handle for manual operation and an extended arm with a specialized tip.

6. Indications For Use [21 CFR 807.92(a)(5)]

The Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general abdominal, gynecologic, thoracic, and pediatric surgical procedures.

7. Technological Characteristics [21 CFR 807.92(a)(6)]

The technological characteristics are the same or very similar to the predicate device.

Surgical stapling is the technological principle for both the subject and predicate device. It is based on the use of endoscopic instrumentation for transection, resection, and/or creation of anastomoses.

The subject and predicate staplers and reloads share the following features:

  • Intended Use
  • Technological characteristics
  • Operational principles
  • Sterilization Method

8. Non-Clinical and Clinical Tests

The device passed the following tests.

Non-Clinical TestAppearance
Flexibility
Assembling capacity
Hardness
Anastomosis and cutting performance
Resistance to Pressure
Safety Device
Sharpness
Surface Roughness
Package Seal
Rotation and Articulation
Firing Electric Current
Battery Voltage
Motor Rotate Speed
Size
Sterility Test
Accelerated Aging Test (ASTM F1980-16)
Seal Strength Test (ASTM F 88/F-15)
Dye Penetration Test (ASTM F1929-15, ASTMF3039-15)
Vacuum Leak Test(ASTM D3078-2002(2013))
Packaging Resistance Bacteria Performance Test(DIN58953-6-2010)
Biocompatibility TestIn Vitro Cytotoxicity Test (ISO 10993-5: 2009)
Skin Sensitization Test (ISO 10993-10:2010)
Intracutaneous reactivity Test (ISO 10993-10:2010)
Acute Systemic Toxicity (ISO 10993-11:2006)
Pyrogen Test (ISO 10993-11:2006)
Subchronic Systemic Toxicity (ISO 10993-11:2006)
Ames test (ISO 10993-3: 2014)
In vitro Mammalian Chromosome Aberration Test(ISO 10993-3:2014)
In Vivo Mammalian Erythrocyte Micronucleus Test(ISO 10993-3:2014)
Muscle Implantation Test (ISO 10993-6: 2016)
Electrical Safety TestElectrical Safety Test (AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012,c1:2009/(r)2012 anda2:2010/(r)2012)
EMC TestEMC Test (IEC60601-1-2:2014)

Page 3 of 8

{6}------------------------------------------------

{7}------------------------------------------------

9. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)]

Product NameDisposable Powered Articulating EndoscopicLinear Cutter Stapler & ReloadsEchelon Flex Powered Articulating EndoscopicLinear Cutters (K130653)K130653 is a modification to their original productcleared under K110385
ClassificationIIII
Regulation Number21 CFR 878.475021 CFR 878.4750
Product CodeGDWGDW
Intended UseThe Disposable Powered Articulating EndoscopicLinear Cutter Stapler & Reloads are intended fortransection, resection, and/or creation ofanastomoses. The instruments have application inmultiple open or minimally invasive generalabdominal, gynecologic, thoracic, and pediatricsurgical procedures.The Echelon families of endoscopic linear cutters(articulating and straight) are intended fortransection, resection, and/or creation ofanastomoses. The instruments have application inmultiple open or minimally invasive general,gynecologic, urologic, thoracic, and pediatricsurgical procedures. They can be used with stapleline or tissue buttressing materials. Theinstruments may also be used for transection andresection of liver parenchyma (hepatic vasculatureand biliary structures), pancreas, kidney andspleen.
Cutting MechanismStapler places two, triple staggered rows of staplesand simultaneously cuts and divides the tissuebetween the two row.Stapler places two, triple staggered rows of staplesand simultaneously cuts and divides the tissuebetween the two row.
Manual/PoweredPoweredPowered
Safety FeaturePower stapler has empty-reload safety protectionmechanism.Power stapler has empty-reload safety protectionmechanism.
Main ComponentsClosing Trigger, Red Firing Trigger Lock, FiringTrigger, Anvil Release Button, Battery Pack,Battery Pack Release Tab, Manual OverrideClosing Trigger, Red Firing Trigger Lock, FiringTrigger, Anvil Release Button, Battery Pack,Battery Pack Release Tab, Manual Override

{8}------------------------------------------------

Access Panel, Knife Reverse Switch, RotatingKnob, Articulation Fins, Reload(Cartridge),Reload Gripping Surface, Reload Alignment Tab,Reload Alignment Slot, Staple Retaining Cap,Anvil Jaw, Cartridge Jaw, Staple Line, Cut Line,Proximal Black Line , Knife Blade Indicator,IndicatorAccess Panel, Knife Reverse Switch, RotatingKnob, Articulation Fins, Reload(Cartridge),Reload Gripping Surface, Reload Alignment Tab,Reload Alignment Slot, Staple Retaining Cap,Anvil Jaw, Cartridge Jaw, Staple Line, Cut Line,Proximal Black Line , Knife Blade Indicator
Staple Line Length30mmN/A
45mm45mm
60mm60mm
Open Staple Height2.0mm, 2.5mm, 3.5mm, 3.8mm, 4.1mm, 4.4mmNo Known
Open Staple PictureImage: Open StapleImage: Open Staple
Closed Staple Height0.75mm0.75mm
1.0mm1.0mm
1.5mm1.5mm
1.8mm1.8mm
2.0mm2.0mm
2.3mm2.3mm
Closed Staple PictureImage: Closed StapleImage: Closed Staple
Number of Staples36N/A
7070
8888
Staple Cross Section0.03 mm2No Known
Patient-Contact MaterialContactMaterialComponentNo Known
AnvilSUS420(Stainless 20Cr13)SUS630(Stainless05Cr17Ni4Cu4Nb/0Cr17Ni4Cu4Nb)
Reload seatSUS630(Stainless05Cr17Ni4Cu4Nb/0Cr17Ni4Cu4Nb)
Cutting knifeSUS440(Stainless 85Cr17)
Aluminumring2A12-T4
Left PipeSUS304(Stainless 06Cr19Ni10)
Right PipeSUS304 (Stainless 06Cr19Ni10)
ReloadLCP, PC
Pushing plateRTP4007A
BracketSUS304(Stainless 06Cr19Ni10)
CoverPU
StapleTi-3AL-2.5V(TA18)
SterilizationSterilized by IrradiationSterilized by Irradiation
SAL10-6Not Known, but assumed to be 10-6

Page 6 of 8

{9}------------------------------------------------

{10}------------------------------------------------

10. Conclusion [21 CFR 807.92(b)(3)]

The Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads device is substantially equivalent to the predicate device because both devices use the same or similar technology and have the same intended use. Testing to relevant standards demonstrate the device will perform as intended. The differences between the devices are insignificant and do not raise different questions of safety and effectiveness for the Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads device.

§ 878.4750 Implantable staple.

(a)
Identification. An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.