Search Results
Found 2 results
510(k) Data Aggregation
(464 days)
The Disposable Powered Articulating Endosopic Linear Cutter Stapler & Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general abdominal, gynecologic, thoracic, and pediatric surgical procedures.
The Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads are sterile, single patient use instruments that simultaneously cut and staple tissue through a battery powered firing system. The Power Stapler and Reloads are sterilized by irradiation. The instruments deliver six staggered rows of staples, three on either side of the cut line. The instruments are available in three shaft lengths: compact (192±20mm), regular (252±20mm) and long (352±20mm). The shaft can rotate freely in both directions and incorporates an articulation mechanism, which enables the distal portion of the shaft to pivot to facilitate lateral access to the operative site.
This document describes a 510(k) premarket notification for a medical device, the "Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads." The core of a 510(k) submission is to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device. This process primarily relies on non-clinical performance testing and comparison of technological characteristics, rather than extensive clinical efficacy studies in the same way a PMA (Pre-Market Approval) process would require.
Therefore, the input document does not contain the information requested in your prompt regarding acceptance criteria and studies proving the device meets them in the context of diagnostic accuracy or comparative effectiveness with human readers. This is because the device described is a surgical instrument (a stapler), not a diagnostic AI system.
However, I can extract information related to performance testing that serves as "acceptance criteria" for this type of device to demonstrate substantial equivalence.
Here's the information derived from the document, tailored as much as possible to your prompt's structure, but with the understanding that it's for a surgical stapler and not an AI diagnostic tool:
1. A table of acceptance criteria and the reported device performance
The document lists "Non-Clinical Tests," "Biocompatibility Tests," "Electrical Safety Test," and "EMC Test" that the device "passed." For a 510(k) submission of a surgical stapler, "passing" these tests indicates that the device met predetermined performance specifications (which serve as acceptance criteria) established by the manufacturer and recognized standards. Specific numerical acceptance criteria are generally detailed in the full test reports, which are summarized here.
Acceptance Criteria Category | Specific Test | Reported Device Performance |
---|---|---|
Non-Clinical Performance | Appearance | Passed |
Flexibility | Passed | |
Assembling capacity | Passed | |
Hardness | Passed | |
Anastomosis and cutting performance | Passed | |
Resistance to Pressure | Passed | |
Safety Device (e.g., empty-reload safety protection) | Passed (and feature explicitly mentioned) | |
Sharpness | Passed | |
Surface Roughness | Passed | |
Package Seal | Passed (ASTM F 88/F-15) | |
Rotation and Articulation | Passed | |
Firing Electric Current | Passed | |
Battery Voltage | Passed | |
Motor Rotate Speed | Passed | |
Size | Passed | |
Sterility Test | Passed (Sterilized by Irradiation, SAL 10^-6) | |
Accelerated Aging Test | Passed (ASTM F1980-16) | |
Dye Penetration Test | Passed (ASTM F1929-15, ASTM F3039-15) | |
Vacuum Leak Test | Passed (ASTM D3078-2002(2013)) | |
Packaging Resistance Bacteria Performance Test | Passed (DIN 58953-6-2010) | |
Biocompatibility | In Vitro Cytotoxicity Test | Passed (ISO 10993-5: 2009) |
Skin Sensitization Test | Passed (ISO 10993-10:2010) | |
Intracutaneous reactivity Test | Passed (ISO 10993-10:2010) | |
Acute Systemic Toxicity | Passed (ISO 10993-11:2006) | |
Pyrogen Test | Passed (ISO 10993-11:2006) | |
Subchronic Systemic Toxicity | Passed (ISO 10993-11:2006) | |
Ames test | Passed (ISO 10993-3: 2014) | |
In Vitro Mammalian Chromosome Aberration Test | Passed (ISO 10993-3:2014) | |
In Vivo Mammalian Erythrocyte Micronucleus Test | Passed (ISO 10993-3:2014) | |
Muscle Implantation Test | Passed (ISO 10993-6: 2016) | |
Electrical Safety | Electrical Safety Test | Passed (AAMI / ANSI ES60601-1:2005/(R)2012 and standards) |
Electromagnetic Comp. | EMC Test | Passed (IEC60601-1-2:2014) |
2. Sample size used for the test set and the data provenance
The document does not specify exact sample sizes for each non-clinical test. For devices like this, sample sizes would typically be determined by statistical methods or industry standards for each specific test (e.g., number of stapler firings, number of animals for biocompatibility tests). There is no mention of "data provenance" in terms of country of origin or retrospective/prospective as this is not a diagnostic AI study. The "test set" here refers to the physical devices and materials used in the non-clinical and biocompatibility testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable to the provided document. The device is a surgical stapler, not a diagnostic AI system that requires expert interpretation for ground truth establishment. The "ground truth" for stapler performance would be objective measurements (e.g., staple formation integrity, seal strength, cutting effectiveness) against predefined engineering specifications and relevant ISO/ASTM standards.
4. Adjudication method for the test set
This question is not applicable to the provided document. Adjudication methods (like 2+1 or 3+1) are used in diagnostic studies to resolve discrepancies in expert interpretations to establish a ground truth. For a surgical stapler, performance is evaluated against objective, measurable criteria, not subjective interpretation.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable to the provided document. An MRMC study is relevant for diagnostic AI tools, not surgical instruments. The device's performance is not about human 'readers' or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable to the provided document. This device is a powered surgical instrument; there is no "algorithm only" or "human-in-the-loop" performance in the context of diagnostic assessment. It functions as a tool operated by a surgeon.
7. The type of ground truth used
For this surgical stapler, the "ground truth" is established through:
- Engineering Specifications and Performance Standards: The device's physical dimensions, electrical characteristics, firing mechanism, and output (staple form, cut line) are compared against pre-defined engineering requirements and relevant international standards (e.g., ISO, ASTM, AAMI).
- Biocompatibility Standards: Materials are tested according to ISO 10993 series for their safety in contact with the body.
- Sterility Assurance: Sterilization methods are validated to achieve a specified Sterility Assurance Level (SAL) (10^-6 in this case).
- Functional Testing: "Anastomosis and cutting performance," "resistance to pressure," "sharpness," etc., are assessed through laboratory and potentially ex-vivo tissue testing against established benchmarks for similar devices.
8. The sample size for the training set
This question is not applicable to the provided document. "Training set" refers to data used to train an AI model. This document describes a physical surgical device, not an AI system.
9. How the ground truth for the training set was established
This question is not applicable to the provided document, as there is no "training set" for an AI model.
Ask a specific question about this device
(48 days)
The ENDOPATH ECHELON™ and ECHELON FLEX™ families of endoscopic linear cutters and reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials. The instruments may also be used for transection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
The Echelon Flex Powered Plus Articulating Endoscopic Linear Cutters are sterile, single patient use instruments that simultaneously cut and staple tissue through a battery powered firing system. The instruments deliver six staggered rows of staples, three on either side of the cut line. The instruments are available in two shaft lengths: regular and long. The shaft can rotate freely in both directions and incorporates an articulation mechanism, which enables the distal portion of the shaft to pivot to facilitate lateral access to the operative site. The instruments are packaged with a battery pack that must be installed prior to use. The Echelon Endoscopic Linear Cutter Reloads (+ Gripping Surface Technology) are designed for use with the Echelon Endoscopic Linear Cutter devices (Echelon, Echelon Flex, Echelon Flex Powered and Echelon Flex Powered Plus) in a 60mm configuration.
This document describes a 510(k) submission for surgical staplers and reloads, focusing on substantial equivalence to predicate devices. It does not contain information about the acceptance criteria for an AI/ML powered device or a study proving that an AI/ML device meets such criteria.
The submitted text focuses on the device's mechanical and electrical performance, biocompatibility, and electromagnetic compatibility. It is a traditional medical device submission, not specifically related to software as a medical device (SaMD) or AI/ML.
Therefore, I cannot provide the requested information, such as acceptance criteria table, sample sizes for test/training sets, expert qualifications, or details about MRMC studies.
Ask a specific question about this device
Page 1 of 1