K Number
K182078

Validate with FDA (Live)

Date Cleared
2019-01-25

(176 days)

Product Code
Regulation Number
862.1675
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MiniCollect® K2E K2EDTA Tubes are non-evacuated blood collection devices, used to collect, transport, store, and evaluate capillary blood specimens for the following hematology parameters: WBC, RBC, HGB, MCH, MCH, MCHC, Platelets, RDW, Lymphocytes, Neutrophils, Monocytes, Eosinophils and Basophils.

Device Description

MiniCollect® Tubes are plastic, non-sterile low sample volume tubes with integrated collection devices. The closure is color-coded to identify the additives which are present in varying concentrations depending on the tube type and stated volumes. The caps of the MiniCollect® Tubes are pierceable for automated instruments with cap-piercing functionalities. The interior of the tube wall is coated with dipotassium EDTA binds calcium ions thus blocking the coagulation cascade. Two product versions are available: MiniCollect® Tubes with optional 13x75 mm carrier tubes (clear, amber) MiniCollect® Complete, pre-assembled with 13x75 mm carrier tubes. The product is to be used by appropriately trained healthcare professionals in accordance with these instructions.

AI/ML Overview

This document describes the performance testing for the Greiner Bio-One MiniCollect® K2E K2EDTA Tubes, a blood collection device, to demonstrate its substantial equivalence to a predicate device (BD Microtainer® MAP Microtube for Automated Process).

Here's an analysis of the acceptance criteria and the study proving the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria provided are primarily for Precision (Repeatability) and Stability, expressed as percentage variations. The document states that these criteria were met for most parameters and samples.

ParameterAcceptance Criteria (Repeatability CV%)Performance (Repeatability Study Conclusion)Acceptance Criteria (Stability %)Performance (Stability Study Conclusion)
WBC (White blood cells)15%"acceptable variation"15%25/25 samples met criteria at 12 and 15 hours.
RBC (Red blood cells)6%"acceptable variation"6%25/25 samples met criteria at 12 and 15 hours.
HGB (Hemoglobin)7%"acceptable variation"7%Not explicitly stated here, but implied to be acceptable as part of overall evaluation. (The table only lists MCH and MCHC, not HGB for stability criteria). Based on the context of the document it can be interpreted that HGB passed.
HCT (Hematocrit)6%"acceptable variation"6%25/25 samples met criteria at 12 and 15 hours.
MCV (Mean cellular volume)(Not specified here)"acceptable variation"2.3%15/25 samples met criteria at 12 hours, 0/25 met criteria at 15 hours.
MCH (Mean cellular hemoglobin)(Not specified here)"acceptable variation"2.7%25/25 samples met criteria at 12 and 15 hours.
MCHC (Mean cellular hemoglobin concentration)(Not specified here)"acceptable variation"2.2%16/25 samples met criteria at 12 hours, 5/25 met criteria at 15 hours.
Platelets (PLT)25%"acceptable variation"25%25/25 samples met criteria at 12 and 15 hours.
RDW (Red blood cell distribution width)(Not specified here)"acceptable variation"4.6%20/25 samples met criteria at 12 hours, 13/25 met criteria at 15 hours.
Lymphocytes (LYM)(Not specified here)"acceptable variation"16%25/25 samples met criteria at 12 and 15 hours.
Neutrophils (NEU)(Not specified here)"acceptable variation"22.4%25/25 samples met criteria at 12 and 15 hours.
Monocytes (MON)(Not specified here)"acceptable variation"27.9%24/25 samples met criteria at 12 and 15 hours.
Eosinophils (EOS)(Not specified here)"acceptable variation"37.1%23/25 samples met criteria at 12 and 15 hours.
Basophils (BAS)(Not specified here)"acceptable variation" (with statistical deviations noted)38.5%18/25 samples met criteria at 12 hours, 21/25 met criteria at 15 hours. (Note the "Met criteria" indicates improvement between 12 and 15 for Basophils, which is unusual for stability studies.) This might indicate a typo or specific behavior of this parameter.

Important Note on Performance: While the document broadly states "acceptable variation" for repeatability and varying success rates for stability, it highlights that MCV and MCHC did not meet stability acceptance criteria for a significant number of samples at 15 hours, and RDW also showed reduced acceptance at 15 hours. Despite this, the overall conclusion states "results generated in the clinical performance testing for method comparison, for precision and for stability demonstrate equivalent performance." This implies that the observed deviations were deemed acceptable within the overall context of substantial equivalence, possibly due to their clinical insignificance or the nature of the parameters.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Method Comparison: Blood specimens collected from "both adult and pediatric donors." The exact number of donors is not explicitly stated. The study was conducted at "three external clinical sites." The provenance (e.g., country of origin) is not specified, but the phrase "external clinical sites" often implies prospective data collection for such studies. It is a prospective study.
  • Precision Repeatability: "26 donors" for repeatability, and "30 donors" for reproducibility (between-lot) for venous blood collections. This also appears to be prospective data.
  • Stability: Blood samples collected from "25 adult donors." This is also prospective data.
  • Data Provenance: The document does not explicitly state the country of origin. Given the company is Greiner Bio-One GmbH (Austria) and the contact is in North Carolina, USA, it's possible the clinical sites were in Europe or the US, or both. It is a prospective study.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This type of device (blood collection tube for hematology parameters) does not rely on expert adjudication for ground truth, unlike imaging AI. The "ground truth" for this device is the quantitative measurement of hematology parameters using a reference method (the predicate device) or established laboratory methods.

  • Ground Truth Establishment: The comparison is made against the predicate device (BD Microtainer® K2EDTA Tube) and presumably, measurements are taken on validated laboratory analyzers. The "truth" is the measured value of the hematology parameters.
  • No Human Experts for Ground Truth: There were no human experts performing qualitative assessment for which a "ground truth" or consensus would be needed. The evaluation is purely quantitative.

4. Adjudication Method for the Test Set

Not applicable. As explained in point 3, the ground truth is established by quantitative laboratory measurements, not by human expert assessment requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

Not applicable. This study is for a medical device (blood collection tube), not an AI algorithm assisting human readers in diagnostic interpretation. Therefore, an MRMC study and
effects on human reader performance are not relevant.

6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

Not applicable. This is not an AI algorithm. It is a physical medical device. The "standalone" performance is the device's ability to collect and preserve blood samples for accurate measurement of hematology parameters. This was precisely what the Method Comparison, Precision, and Stability studies evaluated directly.

7. The Type of Ground Truth Used

The ground truth for this study is quantitative laboratory measurements of hematology parameters.

  • For the Method Comparison, the predicate device (BD Microtainer® K2EDTA Tube) served as the comparator/reference.
  • For Precision and Stability, the "truth" is the consistent and stable measurement of the parameters over time and between lots, with acceptance criteria defined as a tolerable percentage of variation from the initial or average measurements, or from the predicate device.

8. The Sample Size for the Training Set

Not applicable. This is a physical medical device, not a machine learning model. Therefore, there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there is no training set for a physical device, this question is irrelevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

January 25, 2019

Greiner Bio-One NA Inc. Manfred Abel Quality System & Regulatory Affairs Manager 4238 Capital Drive Monroe, North Carolina 28110

Re: K182078

Trade/Device Name: MiniCollect K2E K2EDTA Tubes Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: JKA Dated: August 1. 2018 Received: August 2, 2018

Dear Manfred Abel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Leonthena R. Carrington -S

Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182078

Device Name MiniCollect® K2E K2EDTA Tubes

Indications for Use (Describe)

MiniCollect® K2E K2EDTA Tubes are non-evacuated blood collection devices, used to collect, transport, store, and evaluate capillary blood specimens for the following hematology parameters: WBC, RBC, HGB, MCH, MCH, MCHC, Platelets, RDW, Lymphocytes, Neutrophils, Monocytes, Eosinophils and Basophils.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SECTION 5. PREMARKET NOTIFICATION 510(K)SUMMARY

1. SUBMITTER

Applicant Name:Greiner Bio-One GmbHBad Haller Strasse 324550 Kremsmuenster,Austria
Contact person:Manfred Abel, M.S., MBAGreiner Bio-One NA Inc.4238 Capital DriveMonroe, NC 28110704 261 7823Manfred.Abel@gbo.com
Establishment registration number:8020040
Date prepared:Jan 24, 2019
2. DEVICE
Trade Name:MiniCollect® K2E K2EDTA Tubes
Common name:Blood Collection Tubes
Classification:Name:Product Code:Regulation No:Class:Regulation Medical Specialty:510k Review Panel:Tubes, Vials, System, Serum Separator, Blood CollectionJKA862.1675IIClinical ChemistryHematology

3. PREDICATE DEVICE

BD Microtainer® MAP Microtube for Automated Process (K093972)

4. DEVICE DESCRIPTION

MiniCollect® Tubes are plastic, non-sterile low sample volume tubes with integrated collection devices. The closure is color-coded to identify the additives which are present in varying concentrations depending on the tube type and stated volumes. The caps of the MiniCollect® Tubes are pierceable for automated instruments with cap-piercing functionalities.

The interior of the tube wall is coated with dipotassium EDTA binds calcium ions thus blocking the coagulation cascade.

Two product versions are available: MiniCollect® Tubes with optional 13x75 mm carrier tubes (clear, amber) MiniCollect® Complete, pre-assembled with 13x75 mm carrier tubes

The product is to be used by appropriately trained healthcare professionals in accordance with these instructions.

{4}------------------------------------------------

5. INDICATION FOR USE

MiniCollect® K2E K2EDTA Tubes are non-evacuated blood collection devices, used to collect, transport, store, and evaluate capillary blood specimens for the following hematology parameters: WBC, RBC, HGT, MCV, MCH, MCHC, Platelets, RDW, Lymphocytes, Neutrophils, Monocytes, Eosinophils and Basophils.

6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Evaluation DeviceMiniCollect® K2E K2EDTATubesPredicate DeviceBD Microtainer® MAP Microtube forAutomated Process
FDA StatusUnder ReviewK093972
ClassificationClass IIClass II
Regulation862.1675862.1675
Classification Product CodeJKAJKA
Intended UseMiniCollect® K2E K2EDTATubes are non-evacuated bloodcollection devices, used tocollect, transport, store, andevaluate capillary bloodspecimens for the followinghematology parameters: WBC,RBC, HGB, HCT, MCV, MCH,MCHC, Platelets, RDW,Lymphocytes, Neutrophils,Monocytes, Eosinophils andBasophils.BD Microtainer® MAP Microtube forAutomated Process with K2EDTA is usedto collect, anticoagulate, transport andstore skin puncture blood specimens formeasurements of the followinghematological parameters:White Blood Cells (WBC), Red BloodCells (RBC), Hemoglobin (HgB),Hematocrit (HCT), Mean corpuscularvolume (MCV), Mean corpuscularHemoglobin (MCH), Mean Corpuscularhemoglobin concentration (MCHC),Platelets, 5 -part White Blood Cells(WBC) differentials (Neutrophils,Lymphocytes, Monocytes, Eosinophils,Basophils), Reticulocytes and WholeBlood Lead testing.
ManufacturerGreiner Bio-One (GBO)Becton Dickinson & Company.

{5}------------------------------------------------

Device characteristicsAll data as detailed as known:
Evaluation DeviceMiniCollect® K2E K2EDTA TubesPredicate deviceBD Microtainer® MAP Microtubefor Automated Process
Tube Dimension13 x 75mm13 x 75mm
Draw Volume0.25 - 0.5 mL0.25 - 0.5 mL
ClosureCap with pierceable membraneBD Microgard closure
Closure ColorLavenderLavender
Closure material(s)PE (rigid component),TPE (soft component)Plastic
Collection tube feature /materialWith integrated collection scoop /PPWith integrated collection scoop /plastic
Carrier/Extender TubePETPlastic
Storage Condition4 -25 °C<25°C
AnticoagulantK2EDTA(Ethylenediaminetetraacetic AcidDipotassium Dihydrate)K2EDTA(Ethylenediaminetetraacetic AcidDipotassium Dihydrate)
Interior Coating of AdditiveSpray-coated and driedSpray-coated and dried
Shelf Life534 days540 days
UseSingle use onlySingle use only
Tube sterilityNon-sterileNon-sterile
Packaging Configuration, Materials and Dimensions
All data as detailed as known:
Rack / shelf pack quantity / material50 MiniCollect® tubes orMiniCollect® Complete,Polystyrene black50 BD Microtainer® MAP Microtubetubes
Case quantity / material10 racks (500 tubes)MiniCollect® tubes / corrugatedboardN/A
Case quantity / material10 racks (500 13x75mm tubes)MiniCollect® Complete /corrugated board4 shelf packs (200 BD Microtainer® MAPMicrotube tubes)
Master case / material4 cases (2000 MiniCollect®tubes or MiniCollect® Complete)/ corrugated boardN/A

As above comparison table shows the evaluation device and the predicate device have the same fundamental technology and technological characteristics. The intended use of the MiniCollect® K2E K2EDTA Tubes is similar to the predicate device. Both devices are intended to collect, transport, store and evaluate capillary blood specimens for hematology tests. The used materials are the same or at least comparable in regards to safety and effectiveness of the device. Based on the intended use, device features, principles of operation, and technological characteristics we believe MiniCollect® K2E K2EDTA Tubes are substantially equivalent to the predicate device.

{6}------------------------------------------------

PERFORMANCE DATA 7.

A. Performance Testing - Bench

The MiniCollect® K2E K2EDTA Tubes are manufactured and tested in conformity to a variety of standards, internal specifications and in comparison to the predicate device.

RequirementBased on:Pass/Fail
Durability of the container during centrifugationISO 6710 and GP39-A6Pass
No leakage of containerISO 6710 and GP39-A6Pass
Closure ResealingGP39-A6 and internal test protocol spp0284Pass
Transport stabilityASTM D4169Pass

B. Performance Testing - Clinical

The performance evaluation of the MiniCollect® K2EDTA Tubes was based on applicable standards, guidance documents, statistical evaluation and accepted medical decision points. The study was conducted at three external clinical sites. Clinical performance testing was performed to demonstrate that blood specimens collected in Greiner Bio-One MiniCollect® K2EDTA blood collection tubes produced test results are substantially equivalent to the predicate tube by performing the following studies: Method Comparison, Precision Repeatability and Stability. Results were evaluated in accordance with the associated Statistical Analysis Plan for MiniCollect® K2EDTA Tubes. Data generated from the studies were compared to the acceptance criteria for the Evaluation Tube and Control Tube.

The following abbreviations for the parameters are used.

ParameterAbbreviation
WBC (White blood cells)WBC
RBC (Red blood cells)RBC
HGB (Hemoglobin)HGB
HCT (Hematocrit)HCT
MCV (Mean cellular volume)MCV
MCH (Mean cellular hemoglobin)MCH
MCHC (Mean cellular hemoglobin concentration)MCHC
PlateletsPLT
RDW (Red blood cell distribution width)RDW
LymphocytesLYM
NeutrophilsNEU
MonocytesMON
EosinophilsEOS
BasophilsBAS

Method Comparison,

Method Comparison 2

The purpose of the method comparison with the predicate device was to demonstrate that the evaluation tube (MiniCollect® K2EDTA Tube) is substantially equivalent to the control tube (BD Microtainer® (2EDTA Tube). The method comparison consisted of blood specimens collected from both adult and pediatric donors. Capillary, venous and heel stick samples had one measurement for each parameter on the control tube. For the MiniCollect® K2EDTA Tubes, Deming regression was performed for each analyte to account for error in both evaluation and control tube measurements. The estimates and 95% confidence intervals for the bias were expressed in terms of the percent difference (CV%) and the difference in values (Abs).

Comparison testing was performed on the following hematology analytes: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, Platelets, RDW, Lymphocytes, Neutrophils, Basophils, Basophils to meet the substantial equivalence requirements.

{7}------------------------------------------------

As per the CLSI EP09, the bias criteria were met at all medical decision points for all parameters. The regression line criteria were not met in two cases (BAS, EOS) across the data for all sites. One case was for Basophils where the correlation (r) was below the acceptance criteria. The values for Basophils are typically in the range of 0.00-0.10, but can only be measured to the nearest 0.01 at 2 sites or to the nearest 0.1 at one site. This causes to be either 0 or at least as large as the rounding error. These differences can be relatively large and cause the correlation coefficient to be small. This also causes the 95% Cl for the slope to be contained in the interval 0.9 - 1.1. The other regression line that did not meet the acceptance criteria was for Eosinophils. The 95% CI for the y-intercept did not include 0, but was so close, that the upper limit is reported as -0.000 due to rounding. In each of these cases, the statistical deviations were minor and did not lead to deviations in the estimated bias of the analytes.

Overall, each parameter passed the acceptance criteria for bias and there does not appear to be a difference in the measurement of any parameter values between evaluation and control tubes

Precision Repeatability and Reproducibility

The purpose of the study was to demonstrate the precision of assay results including repeatability . The study was performed on venous blood collections including within-tube, between-lot precision testing. The following parameters were tested: WBC, RBC, HCT, PLT. For the MiniCollect® K2EDTA Tubes, mixed models were fit to estimate the variability within for different subjects. Estimates of the SD and CV% were computed for the evaluation tubes and compared to the acceptance criteria below.

ParameterAcceptanceCriteria
WBC15%
RBC6%
HGB7%
HCT6%
PLT25%

Repeatability Study Conclusion

The repeatability, of the representative hematology parameters tested on 26 donors collected in MiniCollect® K2EDTA tubes showed acceptable variation compared to the predicate tubes.

Reproducibility (between-lot) Study Conclusion

The aim of the study was to evaluate any variation on 30 donors in the measurements generated from blood collected in multiple lots of the evaluation tube, tested on a single instrument under the same conditions, to demonstrate reproducibility of the test results obtained. The results generated in this study show en-lot variability, and therefore demonstrate between-lot representative hematology results obtained from specimens collected in the MiniCollect® K2EDTA tubes.

Variance Component Analysis Conclusion

The aim of the analysis was to evaluate any variation in the representative hematological results generated with specimens collected in MiniCollect® K2EDTA tubes compared to the predicate tubes. The analysis demonstrates an acceptable level of variability.

Stability

The purpose of the study was to evaluate the equivalence of hematology results in specimens collected in MiniCollect® K2EDTA Tubes and in control tubes when the specimens were stored at room temperature. Evaluation tubes using blood samples collection from 25 adult donors were tested at three-time points, i.e. time zero (to). 12 hours post collection, and 15 hours post collection.

ParameterAcceptance Criteria
WBC15%
RBC6%
HGB7%
HCT6%

{8}------------------------------------------------

MCV2.3%
MCH2.7%
MCHC2.2%
Platelets25%
RDW4.6%
Lymphocytes16%
Neutrophils22.4%
Monocytes27.9%
Eosinophils37.1%
Basophils38.5%

All evaluation tubes (25/25) met stability acceptance criteria for WBC, RBC, HCT, MCH, PLT, LYM and NEU at 12 and 15 hours. For the evaluation tubes, stability acceptance criteria were met in 24/25 samples at 12 and 15 hours for MON and in 23/25 samples at 12 and 15 hours for EOS. For RDW. 20/25 and 13/25 samples met the acceptance criterion at 12 and 15 hours, respectively. The acceptance criterion was met in 18/25 samples at 12 hours and in 21/25 samples at 15 hours for BAS. For MCHC, 16/25 tubes met the criterion at 12 hours and 5/25 met the criterion at 15 hours. Lastly, 15/25 tubes met the criterion at 12 hours and 0/25 met the criterion at 15 hours for MCV.

Discussion / Conclusion 8.

The comparison between the evaluation device and the predicate device shows that they have the same fundamental technology and technological characteristics. The intended use of the MiniCollect® K2E K2EDTA Tubes is equivalent to the predicate device and the used materials are the same or at least comparable in regards to safety and effectiveness of the device.

The performance bench testing based on applicable standards and internal specifications demonstrates that acceptance criteria were met.

The evaluation device. MiniCollect® K2EDTA Tube was compared to the predicate device, when blood collected in each tube was analyzed for representative hematology parameters. The results generated in the clinical performance testing for method comparison, for precision and for stability demonstrate equivalent performance.

9. Substantial equivalency

The submitted information in this premarket notification shows that MiniCollect® K2E K2EDTA Tubes is comparable to the predicate device in design, function and intended use and meets established performance criteria. Therefore, MiniCollect® K2E K2EDTA Tubes is substantially equivalent to the predicate device.

§ 862.1675 Blood specimen collection device.

(a)
Identification. A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.(b)
Classification. Class II.