K Number
K181967

Validate with FDA (Live)

Date Cleared
2018-12-10

(140 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Temperature Probe (body surface type) is to be used with a compatible medical monitor that has temperature measurement function to continuously measure the patient's skin temperature.

The Disposable Temperature Probe (celomic type) is to be used with a compatible medical monitor that has temperature measurement function to continuously measure the patient's oral, rectal, nasopharyngeal cavity temperature.

Device Description

The Disposable Temperature Probe is a noninvasive, quick-acting, safe and secure way for continuous temperature monitoring. It provides an effective way for doctors to obtain patient's temperature. The Disposable Temperature Probe obtains temperature information through the influence degree that the thermal balance has on the resistance value of thermistor sensor after it has contacted human body. The Disposable Temperature Probe is intended to be connected to a compatible medical monitor which is Model C30 (Manufacturer: COMEN) to continuously measure the patient's temperature, which can only be used by professionals for single use. It has two types which are body surface type (model: TAS03-15. TAS03-16. TAS03-17) and celomic type (model: TAE03-09S, TAE03-10S, TAE03-12S, TAE03-13S, TAE03-13S, TAE03-14S, TAE03-15S, TAE03-16S). The body surface type is mainly pasted under the axilla while the celomic type is inserted into the oral, rectal or nasopharyngeal cavity. The temperature probes should not be used for more than 24 hours. The Disposable Temperature Probe is composed of plug, cable and temperature sensing probe. For patient-contacting components, their materials have all been tested and evaluated for biocompatibility.

AI/ML Overview

The provided text describes the acceptance criteria and the study that demonstrates the performance of the Disposable Temperature Probe.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

CriterionAcceptance CriteriaReported Device Performance
Temperature Range25°C-45°C25°C-45°C (Same)
Accuracy±0.1°C (25°C - 45°C)±0.1°C (25°C - 45°C) (Same)
BiocompatibilityComplies with ISO 10993-5 and ISO 10993-10Passed the tests as per ISO 10993-5 and ISO 10993-10 (Cytotoxicity, sensitization, irritation)
Electrical SafetyComplies with IEC 60601-1Passed the test as per IEC 60601-1
EMCComplies with IEC 60601-1-2Passed the test as per IEC 60601-1-2
Ambient Temperature OperationCovers 15°C to 35°C (as per ISO 80601-2-56)Functions well under specified ambient temperature environment (including 15°C to 35°C)

2. Sample size used for the test set and the data provenance

The document does not specify the sample size used for any test sets (e.g., for accuracy, biocompatibility, electrical safety, or EMC). It also does not explicitly state the data provenance (e.g., country of origin, retrospective or prospective). It generally states that "Non-clinical testings have been conducted" and refers to compliance with international standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The document refers to compliance with international standards for performance and safety tests, which typically involve laboratory testing rather than expert-established ground truth for a clinical dataset.

4. Adjudication method for the test set

This information is not provided in the document. As mentioned above, the tests described are primarily engineering and laboratory-based, not clinical studies requiring adjudications.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done. If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study was not conducted. This device is a disposable temperature probe, which is a physical measurement device, not an AI-assisted diagnostic tool for which an MRMC study would be relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This concept is not applicable to this device. The Disposable Temperature Probe is a sensor; it does not contain an algorithm in the sense of AI or image analysis. Its performance is evaluated through direct physical measurements against established standards.

7. The type of ground truth used

For accuracy, the ground truth would be precise temperature measurements from calibrated reference instruments or environments during laboratory testing. For biocompatibility, the ground truth is established by the accepted biological responses to materials as defined by ISO 10993 standards. For electrical safety and EMC, the ground truth is compliance with the limits and conditions specified in the respective IEC standards.

8. The sample size for the training set

This device does not involve a "training set" in the context of machine learning or AI. Therefore, this information is not applicable.

9. How the ground truth for the training set was established

As there is no training set for this device, this information is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 10, 2018

Shenzhen Launch Electrical Co., LTD % Tracy Che Registered Engineer Feiying Drug & Medical Consulting Technical Service Group B-3F-3005, Bldg.1, Southward Ruifeng Business Center, No. 22 Guimiao Road Shenzhen, Guangdong, 518000 China

Re: K181967

Trade/Device Name: Disposable Temperature Probe. Model: TAS03-14. TAS03-15. TAS03-16. TAS03-17 (body surface type); TAE03-09S, TAE03-10S, TAE03-11S, TAE03-12S, TAE03-13S, TAE03-14S, TAE03-15S, TAE03-16S (celomic type) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: November 8, 2018 Received: November 13, 2018

Dear Tracy Che:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Sapana Patel -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181967

Device Name

Disposable Temperature Probe/ Model: TAS03-14, TAS03-16, TAS03-17 (body surface type); TAE03-10S, TAE03-11S, TAE03-12S, TAE03-13S, TAE03-14S, TAE03-15S, TAE03-16S (celomic type)

Indications for Use (Describe)

The Disposable Temperature Probe (body surface type) is to be used with a compatible medical monitor that has temperature measurement function to continuously measure the patient's skin temperature.

The Disposable Temperature Probe (celomic type) is to be used with a compatible medical monitor that has temperature measurement function to continuously measure the patient's oral, rectal, nasopharyngeal cavity temperature.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510 (k) Summary

This "510(k) Summary" is submitted in accordance with requirements of Title 21, CFR Section 807.92.

K181967

(1) Applicant information

510 (k) owner's name:SHENZHEN LAUNCH ELECTRICAL CO., LTD
------------------------------------------------------------
Address:NO. F Building, Zhonggangxing Industrial Estate, ZhanggeCommunity, Guanlan Street, Longhua New District, Shenzhen,China
Contact person:Jiaan Huang
Phone number:+86-755-29532401
Fax number:+86-755-29532483
Email:yuanhe@cnlaunch.com
Date of summary prepared:2018.12.07

(2) Proprietary name of the device

Trade name/Model:Disposable Temperature Probe/ Model: TAS03-14, TAS03-15,TAS03-16, TAS03-17 (body surface type); TAE03-09S,TAE03-10S, TAE03-11S, TAE03-12S, TAE03-13S, TAE03-14S,TAE03-15S, TAE03-16S (celomic type)
Regulation name:Clinical electronic thermometer
Regulation number:21 CFR 880.2910
Product codeFLL
Review panel:General hospital
Regulation class:Class II

(3) Predicate and reference device

  • 彩 Predicate device
SponsorDräger Medical GmbH
Device Name and ModelTemperature Probes (General purpose/ Skin)
510(k) NumberK121999
Product CodeFLL
Regulation Number21 CFR 880.2910
Regulation ClassII

{4}------------------------------------------------

SponsorStarboard Medical, LLCWelch Allyn, Incorporated
Device Name and ModelEsophageal/RectalTemperature ProbeSureTemp® Plus
510(k) NumberK140134K030580
Product CodeBZT, FLLFLL
Regulation Number21 CFR 868.192021 CFR 880.291021 CFR 880. 2910
Regulation ClassIIII
  • Reference device જુન

(4) Description/ Design of device

The Disposable Temperature Probe is a noninvasive, quick-acting, safe and secure way for continuous temperature monitoring. It provides an effective way for doctors to obtain patient's temperature. The Disposable Temperature Probe obtains temperature information through the influence degree that the thermal balance has on the resistance value of thermistor sensor after it has contacted human body. The Disposable Temperature Probe is intended to be connected to a compatible medical monitor which is Model C30 (Manufacturer: COMEN) to continuously measure the patient's temperature, which can only be used by professionals for single use. It has two types which are body surface type (model: TAS03-15. TAS03-16. TAS03-17) and celomic type (model: TAE03-09S, TAE03-10S, TAE03-12S, TAE03-13S, TAE03-13S, TAE03-14S, TAE03-15S, TAE03-16S). The body surface type is mainly pasted under the axilla while the celomic type is inserted into the oral, rectal or nasopharyngeal cavity. The temperature probes should not be used for more than 24 hours. The Disposable Temperature Probe is composed of plug, cable and temperature sensing probe. For patient-contacting components, their materials have all

(5) Indications for use

been tested and evaluated for biocompatibility.

The Disposable Temperature Probe (body surface type) is to be used with a compatible medical monitor that has temperature measurement function to continuously measure the patient's skin temperature.

The Disposable Temperature Probe (celomic type) is to be used with a compatible medical monitor that has temperature measurement function to continuously measure the patient's oral, rectal, nasopharyngeal cavity temperature.

Component ofDevice RequiringBiocompatibilityMaterial ofComponentBody Contact Category(ISO 10993-1)Contact Duration(ISO 10993-1)
-----------------------------------------------------------------------------------------------------------------------------------------------------------

(6) Materials

{5}------------------------------------------------

StickerFoamSurface-contacting device:Intact skin< 24hours
ProbePVCSurface-contacting device:Mucosal membrane< 24hours

The body-contacting material used in Disposable Temperature Probe have all passed biocompatibility test.

(7) Technological characteristics and substantial equivalence

ItemTargeteddevicePredicate deviceReferencedeviceReferencedeviceRemark
Trade nameDisposableTemperatureProbe/Model:TAS03-14,TAS03-15,TAS03-16,TAS03-17(body surfacetype);TAE03-09S,TAE03-10S,TAE03-11S,TAE03-12S,TAE03-13S,TAE03-14S,TAE03-15S,TAE03-16S(celomictype)TemperatureProbes (Generalpurpose/ Skin)Esophageal/RectalTemperatureProbeSureTemp® Plus/
510 (k) numberK181967K121999K140134K030580/
Regulationnumber21 CFR880.291021 CFR880.291021 CFR868.192021 CFR880.291021 CFR 880.2910Same
Product codeFLLFLLBZT, FLLFLLSame
ClassIIIIIIIISame
Indications foruse/ IntendeduseThe DisposableTemperatureProbe (bodysurface type) isto be used witha compatibleGeneralPurposeTemperatureProbesDrägerdisposableTheEsophageal/RectalTemperatureProbe isindicated forThe Welch AllynSureTemp® Plusis intended to beused byhealthcareprofessionals toSimilarNote 5
medicalgeneral purposecontinuousprovide an
monitor thattemperaturemonitoring ofaccurate
has temperatureprobes areesophageal,prediction of
measurementintended forrectal, andpatient
function topatient corenasopharyngealtemperature
continuouslybodytemperatures.using the oral,
measure thetemperatureaxillary or rectal
patient's skinmeasurement inbody sites in 4 to
temperature.combination15 seconds or to
The Disposablewith Dräger andprovide an
TemperatureSiemens patientactual
Probe (celomicmonitoringtemperature
type) is to besystems. Thereading in the
used with aprobes arecontinuous
compatibleinserted into themonitor mode in
medicaloesophagus orabout 3 minutes.
monitor thatthe rectum.
has temperatureSkin
measurementTemperature
function toProbes
continuouslyDräger
measure thedisposable skin
patient's oral,temperature
rectal,probes are
nasopharyngealintended for
cavitypatient skin
temperature.temperature
measurement incombination
with Dräger and
Siemens patient
monitoring
systems. The
probes are
affixed on the
patient's skin
with an adhesive
cover.
Suitable for theAdult andAdult, pediatric//Same
crowdpediatricand neonatal
patientpatient
ThermometerSkin(mainlySkin, oesophagusEsophageal,Oral, axillary orSimilar
typeaxilla),oral,or the rectum.rectalandrectal body sites
rectalandnasopharyngeal
nasopharyngealcavity
PrescriptionorOTCPrescriptionPrescription//Same
Single useYesYes//Same
ComponentsDisposableTemperatureProbe iscomposed ofplug, cable andtemperaturesensing probe.Mainlycomposed of aplug connectedto a patientmonitor and athermistor on thepatient end.//Similar
SensorThermistorwhich issensitive totemperaturechange.Thermistorwhich issensitive totemperaturechange.//Same
MonitoringSystemCompatibilityModel C30(Manufacturer:COMEN)Dräger andSiemens patientmonitoringsystems.//DifferentNote 1
PrincipleofoperationThe DisposableTemperatureProbe obtainstemperatureinformationthrough theinfluencedegree that thethermal balancehas on theresistance valueof thermistorsensor after ithas contactedhuman body.The thermistorconsists of aresistor which issensitive totemperaturechanges.//Similar
Devicespecifications750±50mm or750±20mm8.9g (Bodysurface type),13.1g (Celomictype)///DifferentNote 2
Temperaturerange25°C-45°C25°C-45°C//Same
Accuracy±0.1°C (25°C - 45°C)±0.1°C (25°C - 45°C)//Same
Ambienttemperatureenvironment5°C-40°CNot specified.//SimilarNote 3
MaterialsFoam and PVCStainless steel, /thermoplasticelastomers,thermoplasticPolyolefin.//DifferentNote 4
BiocompatibilityPassed the testsas per ISO10993-5 andISO 10993-10(Cytotoxicity,sensitization,irritation)Passed the tests /as per ISO10993-5 and ISO10993-10//Same
Electrical safetyPassed the testas per IEC60601-1Passed the test as /per IEC 60601-1//Same
EMCPassed the testas per IEC60601-1-2Passed the test as /per IEC60601-1-2//Same
SterilityNon-sterileNon-sterile//Same

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

SHENZHEN LAUNCH ELECTRICAL CO., LTD 510(k)s - Section 7. 510 (k) Summary

A Note 1:

Although compared with the predicate device, the subject device may be used with different monitors, all related safety and performance tests have been conducted and met requirements, so this difference does not raise any device performance issues.

A Note 2:

Although the device specifications may be different from that of the predicates, this difference does not raise any product problems.

A Note 3:

The ambient temperature environment of the predicate device are not specified or not exactly the same. According to ISO 80601-2-56, a clinical thermometer shall operate in normal use over the ranges of an ambient temperature operating range from 15℃ to 35℃. The ambient temperature of the subject device covers this range, our company has also tested the subject device according to ISO 80601-2-56, which has proved that the subject device functions well under specified ambient temperature environment.

A Note 4:

The materials used in the subject device are different from that of the predicate device. But regarding biocompatibility, our company has authorized a qualified third-party lab to conduct tests according to ISO 10993-5 and ISO 10993-10, which has proved that the materials used in the subject device are biologically safe.

{9}------------------------------------------------

Note 5:

Although the predicate device is intended for patient skin temperature measurement, the reference devices are intended for temperature measurement via nasopharynx and oral route. The method of measurement do not raise new or different questions of safety or effectiveness.

(8) Non-clinical studies and tests performed

Non-clinical testings have been conducted to verify that the Disposable Temperature Probe meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate devices. The testing results demonstrate that the targeted device complies with the following standards:

  • A IEC 60601-1, Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests

  • ISO 80601-2-56, Medical Electrical Equipment - Part 2-56: Particular Requirements For Basic Safety And Essential Performance Of Clinical Thermometers For Body Temperature Measurement. (General Plastic Surgery/General Hospital)

The body-contacting components of this device are sticker and probe which have been demonstrated conformance to the following standards.

  • ISO 10993-5, Biological Evaluation Of Medical Devices -- Part 5: Tests For InVitro Cytotoxicity

  • A ISO 10993-10. Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
  • Cleaning validation and disinfection validation have been conducted based on the following FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."

(9) Conclusion

Based on the above analysis and tests performed, it can be concluded that the performance and function of Disposable Temperature Probe is Substantially Equivalent (SE) to the predicate device.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.