K Number
K181822

Validate with FDA (Live)

Date Cleared
2019-03-21

(255 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

RENASYS TOUCH is indicated for patients who would benefit from a suction device (Negative Pressure Wound Therapy), as it may promote wound healing via removal of fluids, including irrigation fluids and body fluids, wound exudates and infectious materials.

Appropriate wound types include:

  • Chronic
  • Acute
  • Traumatic
  • Sub-acute and dehisced wounds
  • Ulcers (such as pressure or diabetic)
  • Partial-thickness burns
  • Flaps and grafts

When used with the RENASYS AB Abdominal Kit with Soft Port, RENASYS TOUCH is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.

The RENASYS Y-Connector can only be used with the RENASYS TOUCH system. RENASYS TOUCH is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials.

The following wound types may be used with a RENASYS system that is utilizing a RENASYS Y-connector:

  • Flaps and grafts (only in one wound configuration)
  • Open abdomen (only in one wound configuration & only with RENASYS TOUCH and RENASYS AB Abdominal Kit)
  • Chronic
  • Acute
  • Traumatic
  • Sub-Acute and dehisced wounds
  • Ulcers (such as pressure or diabetic)
  • Partial-thickness burns

When the RENASYS Y-Connector is used with the RENASYS AB Abdominal Kit with Soft Port (only in one wound configuration), it is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.

Device Description

RENASYS TOUCH device is a software controlled suction device consisting of an electric motor driven, diaphragm, vacuum pump. The device is designed to provide Negative Pressure Wound Therapy at a range of pressure settings between 25-200mmHg to a closed environment over one or two wounds. The device removes exudate from the wound site(s) to a disposable container, which may promote wound healing via removal of fluids, including irrigation of body fluids, wound exudates and infectious materials.

RENASYS TOUCH device can be operated by either a mains power supply or internal battery.

RENASYS TOUCH device contains integrated global cellular technology that allows device location, billing and maintenance data to be sent wirelessly to a secure Internet website providing Health Care Professionals access to device information.

The RENASYS TOUCH device is compatible with Smith & Nephew RENASYS Foam, RENASYS Gauze and RENASYS Abdominal dressing kits previously cleared under K152163, K153209 and K143133.

RENASYS TOUCH has also demonstrated compatibility with RENASYS Y-Connector. The RENASYS Y-connector is only intended to be used in conjunction with RENASYS Negative Pressure Wound Therapy (NPWT) systems. The RENASYS Y-Connector can be used to connect one or two wounds through two RENASYS Soft Ports to a single RENASYS TOUCH device. The Y-Connector is constructed of PVC tubing and is compatible with Smith & Nephew RENASYS Foam, RENASYS Gauze and RENASYS Abdominal dressing kits with Soft Port.

AI/ML Overview

Here's an analysis of the acceptance criteria and supporting study for the RENASYS TOUCH and RENASYS Y-Connector, based on the provided FDA 510(k) summary:

This device is not an AI/ML device. The provided document does not describe the acceptance criteria and study proving an AI/ML device meets them. The document is for a Negative Pressure Wound Therapy (NPWT) device, the RENASYS TOUCH, and its accessory, the RENASYS Y-Connector. It describes non-clinical performance testing rather than an AI/ML study.

Therefore, many sections of your requested output, particularly those related to AI/ML specific criteria (like sample size for test set, data provenance, ground truth establishment, MRMC studies, standalone performance), cannot be answered from the provided text.

However, I can extract information related to the device's performance testing and general regulatory details.

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria in a formal table format, nor does it provide specific numerical performance results. Instead, it summarises the types of performance testing conducted to verify the device meets design specifications and demonstrates substantial equivalence to the predicate. The "reported device performance" is generally stated as having "successfully completed" the tests.

Acceptance Criteria (Inferred from Test Types)Reported Device Performance
Maintain pressure at all settings (25-200mmHg) in simulated wound modelsSuccessfully completed simulated wound model testing; maintains pressure.
Battery life and verification over duration of battery lifeSuccessfully completed battery verification over the duration of its life.
Functionality with integrated global cellular technology (SIM activation/deactivation)Successfully completed activation and deactivation of the SIM card.
Compatibility with RENASYS Y-Connector in simulated wound modelsSuccessfully completed RENASYS TOUCH compatibility testing with RENASYS Y-Connector in simulated wound models.
Electrical safety, Electromagnetic compatibility (EMC), and mechanical/environmental performance for various use environments (transportation, ambulatory, home healthcare)Successfully completed electrical safety, EMC, and mechanical/environmental testing.
Compliance with Home Healthcare requirements (IEC 60601-1-11)Demonstrated compliance with Home Healthcare requirements of IEC 60601-1-11.
Software functionality (wireless technology, GUI, video playback for alarms/canister changes)Software updated to facilitate wireless technology, updated GUI screen wording, includes video playback.
Compatibility with additional Class I Power Supply and Power CordsDevice includes compatibility with an additional Class I Power Supply and Power Cords.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document states that performance testing was "performed in triplicate" for the RENASYS TOUCH device (implying 3 devices were tested) and that this was done for "simulated wound model testing." It doesn't specify how many simulated wound models or test scenarios were run for each triplicate.
  • Data Provenance: The testing was non-clinical performance testing conducted by the manufacturer, Smith & Nephew Medical Limited. The country of origin for the data generation would likely be the UK (where Smith & Nephew Medical Limited is located) or another internal testing facility. The data is prospective as it was generated specifically for this 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This is not applicable as the study involved non-clinical performance testing using simulated wound models and engineering measurements, not expert-adjudicated clinical data or ground truth from human experts.

4. Adjudication Method for the Test Set

Not applicable, as there was no human adjudication of results in this non-clinical performance testing.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device (NPWT pump) and not an AI/ML diagnostic tool. No MRMC study was conducted.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML algorithm. The device's "standalone" performance refers to its ability to perform its intended function of negative pressure wound therapy, which was assessed through the non-clinical performance testing described.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance testing consists of:

  • Engineering specifications and design requirements: The device's ability to maintain specific pressure ranges, battery life, and proper function of its components (e.g., SIM card, software GUI).
  • Predicate device performance: The new device was tested to verify it performs comparably to its legally marketed predicate device (K153209 RENASYS TOUCH System).
  • Relevant standards: Compliance with standards like IEC 60601-1-11 for home healthcare environments.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device, so there is no "training set."

9. How the Ground Truth for the Training Set was Established

Not applicable. As there is no "training set," there is no ground truth to establish for one.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 21, 2019

Smith & Nephew Medical Limited % Kulsum Master Director, Regulatory Affairs Smith and Nephew 7000 West William Cannon Drive Austin, Texas 78735

Re: K181822

Trade/Device Name: RENASYS Touch; RENASYS Y-Connector Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: February 18, 2019 Received: February 19, 2019

Dear Kulsum Master:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kimberly Ferlin -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K181822

Device Name RENASYS TOUCH

Indications for Use (Describe)

RENASYS TOUCH is indicated for patients who would benefit from a suction device (Negative Pressure Wound Therapy), as it may promote wound healing via removal of fluids, including irrigation fluids and body fluids, wound exudates and infectious materials.

Appropriate wound types include:

  • Chronic
  • · Acute
  • · Traumatic
  • · Sub-acute and dehisced wounds
  • · Ulcers (such as pressure or diabetic)
  • · Partial-thickness burns
  • Flaps and grafts

When used with the RENASYS AB Abdominal Kit with Soft Port, RENASYS TOUCH is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.

Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K181822

Device Name RENASYS Y-Connector

The RENASYS Y-Connector can only be used with the RENASYS TOUCH system. RENASYS TOUCH is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials.

The following wound types may be used with a RENASYS

system that is utilizing a RENASYS Y-connector:

  • · Flaps and grafts (only in one wound configuration)
  • · Open abdomen (only in one wound configuration & only with RENASYS TOUCH and RENASYS AB Abdominal Kit)
  • Chronic
  • Acute
  • · Traumatic
  • Sub-Acute and dehisced wounds
  • · Ulcers (such as pressure or diabetic)
  • · Partial-thickness burns

When the RENASYS Y-Connector is used with the RENASYS AB Abdominal Kit with Soft Port (only in one wound configuration), it is indicated for temporary bridging of abdominal wall openings where is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY K181822

General Information
Submitter Name/Address:Smith & Nephew Medical Limited101 Hessle Road,HullHU3 2BNUnited Kingdom
Establishment Registration Number:8043484
Contact Person:Kulsum Master, Director Regulatory Affairs, US Region
Phone Number+1-512-358-5720
Date Prepared:July 06, 2018
Application Correspondent:Smith & Nephew Inc.7000 West William Cannon Drive,Austin,Texas, 78735,USA
Contact Person:Kulsum Master, Director Regulatory Affairs, US Region
Phone Number:+1 512-358-5720

{5}------------------------------------------------

Device Description
Trade Name:RENASYS TOUCH
Common or Usual Name:Negative Pressure Wound Therapy
powered suction pump
Classification Name:Powered suction pump (21 CFR
878.4780)
Regulatory Class:Class II
Product Code:OMP
Predicate Device Information
510(k) Number:K153209
Device:RENASYS TOUCH System
Clearance Date:August 4, 2016
510(k) Number:K143133
Device:RENASYS AB Abdominal Kit
Clearance Date:July 23, 2015

Device Description

RENASYS TOUCH device is a software controlled suction device consisting of an electric motor driven, diaphragm, vacuum pump. The device is designed to provide Negative Pressure Wound Therapy at a range of pressure settings between 25-200mmHg to a closed environment over one or two wounds. The device removes exudate from the wound site(s) to a disposable container, which may promote wound healing via removal of fluids, including irrigation of body fluids, wound exudates and infectious materials.

RENASYS TOUCH device can be operated by either a mains power supply or internal battery.

{6}------------------------------------------------

RENASYS TOUCH device contains integrated global cellular technology that allows device location, billing and maintenance data to be sent wirelessly to a secure Internet website providing Health Care Professionals access to device information.

The RENASYS TOUCH device is compatible with Smith & Nephew RENASYS Foam, RENASYS Gauze and RENASYS Abdominal dressing kits previously cleared under K152163, K153209 and K143133.

RENASYS TOUCH has also demonstrated compatibility with RENASYS Y-Connector. The RENASYS Y-connector is only intended to be used in conjunction with RENASYS Negative Pressure Wound Therapy (NPWT) systems. The RENASYS Y-Connector can be used to connect one or two wounds through two RENASYS Soft Ports to a single RENASYS TOUCH device. The Y-Connector is constructed of PVC tubing and is compatible with Smith & Nephew RENASYS Foam, RENASYS Gauze and RENASYS Abdominal dressing kits with Soft Port.

Indications for Use - RENASYS TOUCH

RENASYS TOUCH is indicated for patients who would benefit from a suction device (Negative Pressure Wound Therapy), as it may promote wound healing via removal of fluids, including irrigation fluids and body fluids, wound exudates and infectious materials.

Appropriate wound types include:

Chronic

Acute

Traumatic

Sub-acute and dehisced wounds

Ulcers (such as pressure or diabetic)

Partial-thickness burns

  • Flaps and grafts

{7}------------------------------------------------

When used with the RENASYS AB Abdominal Kit with Soft Port, RENASYS TOUCH is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera. including but not limited to abdominal compartment syndrome. The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.

Indications for Use - RENASYS Y-Connector

The RENASYS Y-Connector can only be used with the RENASYS TOUCH system. RENASYS TOUCH is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials.

The following wound types may be used with a RENASYS system that is utilizing a RENASYS Y-connector:

  • · Flaps and grafts (only in one wound configuration)
  • Open abdomen (only in one wound configuration & only with RENASYS TOUCH and RENASYS AB Abdominal Kit)
  • Chronic
  • Acute
  • Traumatic
  • Sub-Acute and dehisced wounds
  • Ulcers (such as pressure or diabetic)
  • Partial-thickness burns

When the RENASYS Y-Connector is used with the RENASYS AB Abdominal Kit with Soft Port (only in one wound configuration), it is indicated for temporary

{8}------------------------------------------------

bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.

Comparison between New and Predicate Device

The technological principal for delivering the negative pressure wound therapy for both the subject and predicate devices are identical. The main differences between the cleared RENASYS TOUCH device K153209 and the subject device are as follows:

  • a) Introduction of an integrated global cellular technology that allows device location, billing and maintenance data to be sent wirelessly to a secure Internet website providing Health Care Professionals access to device information.
    The wireless technology does not monitor or control the delivery of negative pressure (range -25mmHg to -200mmHg; optimal therapeutic range: -40mmHg to -120mmHg) or transmit patient data.

  • b) The device software has been updated to facilitate the operation of the wireless technology introduced as part of the 510(k) notice for the subject device, including electronic labelling for wireless regulations.

  • c) Updated Graphical User Interface (GUI) screen wording when Y-Connector mode to enable to use of the device on one or two wounds when using a Y-connector.

  • d) The subject device has demonstrated compliance with the Home Healthcare requirements of IEC 60601-1-11.

{9}------------------------------------------------

  • e) Minor update to the indications for use statement to remove information on use in professional healthcare environments due to the subject device demonstrating compliance to the home healthcare requirements of IEC60601-1-11.
  • The device software has been updated to included video playback with f) device usage on alarm troubleshooting and canister changes.
  • g) The device includes compatibility with an additional Class I Power Supply and Power Cords for Hospital Use, purchased separately for special Hospital needs.
  • h) The indications for use have been updated to clarify that the RENASYS TOUCH pump can be used with the RENASYS AB Abdominal Kit as per the predicate submission K153209. The wording included is taken directly from the indications from the predicate RENASYS AB Abdominal Kit K143133.

Non Clinical Data

Performance testing has been conducted to verify the RENASYS TOUCH device meets the design specifications and demonstrates substantial equivalence to the predicate device. The principle test methods used to demonstrate performance are simulated wound models, using the same test strategy that was used to clear the predicated device. Additional consideration was taken into account regarding function of the wireless communication with the RENASYS TOUCH devices functioning in the same vicinity. All performance testing of the RENASYS TOUCH device was performed in triplicate with devices in close proximity and 3G-GPS enabled.

{10}------------------------------------------------

The list below summarizes the testing undertaken and successfully completed for the RENASYS TOUCH device

  • Battery verification over the duration of the RENASYS TOUCH devices' . battery life using simulated wound model testing
  • Simulated wound model testing to verify the RENASYS TOUCH device . maintains pressure at all pressure settings
  • Activation and deactivation of the SIM card
  • RENASYS TOUCH compatibility testing with the RENASYS Y-Connector . in simulated wound models
  • Electrical safety testing, Electromagnetic compatibility testing and . mechanical/environmental testing to verify its use in transportation, ambulatory and home healthcare environments

Clinical Data

No clinical or animal data is included in this submission

Conclusion

In establishing substantial equivalence to the predicate device. Smith & Nephew Medical Ltd evaluated the indications for use, principle of operation, materials, technology, product specifications and energy requirements of the device. Performance testing, software verification testing, electromagnetic compatibility testing and electrical safety testing has been completed to demonstrate that the RENASYS TOUCH device is substantially equivalent to the predicate device for the intended use.

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.