K Number
K181752

Validate with FDA (Live)

Date Cleared
2018-12-21

(172 days)

Product Code
Regulation Number
888.1100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stryker IVAS® Elite Inflatable Vertebral Augmentation System is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The system is to be used with cleared spinal Polymethylmethacrylate (PMMA) bone cements and Cortoss® Bone Augmentation Material indicated for use during percutaneous vertebral augmentation procedures, such as kyphoplasty.

Device Description

The Stryker iVAS ® Elite Balloon Catheter is a bone tamp with an inflatable component at the distal end. The balloon is inflated to create a void within the vertebral body. It is used with various accessories during vertebral augmentation.

When the balloon is inflated with radiopaque fluid, the balloon expands axially and radially. Two radiopaque markers are fixed near the distal end of the balloon to show the location of the balloon during placement. The balloon is coated with silicone to assist with insertion of the catheter into the access needle.

AI/ML Overview

This looks like a 510(k) summary for a medical device that isn't an AI/ML device. Therefore, the questions about acceptance criteria for AI/ML performance metrics (like sensitivity, specificity, AUC, human reader improvement with AI assistance, sample sizes for training/test sets, expert ground truth establishment, etc.) are not applicable here.

This document describes a conventional medical device (an inflatable vertebral augmentation system) and focuses on engineering and biological performance rather than diagnostic or analytical accuracy.

However, I can extract the acceptance criteria and the study details for the conventional device as presented in the document:

1. A table of acceptance criteria and the reported device performance

Modification & Test PerformedAcceptance CriteriaReported Performance (Result)
Sheath Colorant change: Biocompatibility (sheath)Adoption (K172116)Pass
Balloon Catheter Length and Volume Change: Constrained BurstSimulate user inflating balloon beyond maximum rating; device met all acceptance criteria.Pass
Balloon Catheter Length and Volume Change: Un-Constrained Burst, Length to Diameter and One-way Valve TorqueVerify balloon meets a minimum open air inflation volume; device met all acceptance criteria.Pass
Balloon Catheter Length and Volume Change: Insert, Retract, Sheath Removal, Assembly Strength, Tensile ForceShow force required to insert and remove the catheter into the access needle; device met all acceptance criteria.Pass
Balloon Catheter Length and Volume Change: Sheath shipping testBalloon Sheath must remain on balloon after bulk shipping configuration.Pass
Balloon Catheter Length and Volume Change: Cadaveric testingNo damage to adjacent anatomical structures following high pressure balloon inflation.Pass
Packaging: Real-time packaging testingDevice must meet all acceptance criteria while determining expiration date using either accelerated or real-time aging methods.Pass

2. Sample sized used for the test set and the data provenance

  • Biocompatibility: No specific sample size is mentioned for biocompatibility testing in this summary, but it states "Biocompatibility testing... was performed following the recommendations of ISO 10993-1 and FDA Guidance."
  • Bench Testing (Constrained Burst, Un-Constrained Burst, etc.): No specific sample sizes are mentioned, but it indicates "All bench testing pertaining to the subject device modifications met their respective acceptance criteria as specified per the individual test report." These tests are typically performed on a statistically relevant sample size of manufactured devices.
  • Cadaveric Testing: "Cadaveric testing was completed to provide evidence..." The specific number of cadavers or test instances is not provided in this summary. The provenance is cadaveric (human anatomical specimens).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Cadaveric testing: "This cadaveric testing was completed by a physician..." The exact number of physicians is not specified, nor are their specific qualifications beyond being a "physician."

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Cadaveric testing: "Following the inflation of the balloon, the physician evaluated nearby anatomical structures for adverse events related to the high-pressure balloon inflations." This implies a single physician assessment; no adjudication method (like 2+1) is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This is not an AI/ML device. No MRMC comparative effectiveness study was conducted or is applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This is not an AI/ML device. No standalone algorithm performance was assessed.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Biocompatibility: Ground truth is established by standardized biological assays and regulatory guidelines (ISO 10993-1, FDA Guidance).
  • Bench Testing: Ground truth is established by engineering specifications, physical measurements, and industry-accepted standards (e.g., burst pressure, tensile strength).
  • Cadaveric Testing: Ground truth is established by direct visual inspection and assessment by a physician for damage to anatomical structures.

8. The sample size for the training set

  • This is not an AI/ML device. There is no concept of a "training set" in the context of this submission.
  • However, the submission does mention that "Biocompatibility testing... adopted per K172116" and "Catheter Flexibility Testing and Cold Age testing have been adopted from K093429. Accelerated Age testing has been adopted from K172116." These refer to prior submissions that likely involved biocompatibility testing and accelerated aging on a set of devices, which could be considered an analogous "training" or foundational dataset for these aspects, but not in the AI/ML sense.

9. How the ground truth for the training set was established

  • Not applicable as this is not an AI/ML device. The "ground truth" for the various engineering and biocompatibility tests would have been established through the methods and standards relevant to those specific test types in the original K093429 and K172116 submissions.

{0}------------------------------------------------

December 21, 2018

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Stryker Corporation Kristi Ashton Sr. Staff Regulatory Affairs Specialist 4100 E. Milham Avenue Kalamazoo, Michigan 49001

Re: K181752

Trade/Device Name: Stryker iVAS ® Elite Inflatable Vertebral Augmentation System (Stryker iVAS ® Elite Balloon Catheter) Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX, NDN Dated: November 20, 2018 Received: November 21, 2018

Dear Ms. Ashton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Sarah B. Nelson -S

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K181752

Device Name

Stryker iVAS ® Elite Inflatable Vertebral Augmentation System (Stryker iVAS ® Elite Balloon Catheter)

Indications for Use (Describe)

The Stryker IVAS® Elite Inflatable Vertebral Augmentation System is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The system is to be used with cleared spinal Polymethylmethacrylate (PMMA) bone cements and Cortoss® Bone Augmentation Material indicated for use during percutaneous vertebral augmentation procedures, such as kyphoplasty.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

1. Submitter

a. 510(k) Owner: Stryker Instruments 4100 E. Milham Avenue Kalamazoo, Michigan 49001 USA Ph: +1-269-323-7700 Fax: +1-269-324-5412 b. FDA Establishment 1811755 Registration Number: Contact Person: Kristi Ashton ﻦ Ph: +1-269-389-5929 Fax: +1-269-389-5412 Kristi.Ashton@Stryker.com

d. Date Submitted: December 14, 2018

2. Subject Device Name

Trade Name: Stryker iVAS® Elite Balloon Catheter Common Name: Inflatable Bone Tamp Product Codes: HRX, NDN Regulation: 888.1100, 888.3027

3. Legally Marketed Predicate Devices

Table 1 Predicate Device Table

Predicate Devices
Stryker iVAS ® Elite Balloon CatheterK172116Primary Predicate
Stryker iVAS® Elite Balloon Catheter
Stryker iVAS® Balloon CatheterK123942Secondary Predicate
Stryker iVAS® Balloon Catheter

4. Device Description

The Stryker iVAS ® Elite Balloon Catheter is a bone tamp with an inflatable component at the distal end. The balloon is inflated to create a void within the vertebral body. It is used with various accessories during vertebral augmentation.

When the balloon is inflated with radiopaque fluid, the balloon expands axially and radially. Two radiopaque markers are fixed near the distal end of the balloon to show the

{4}------------------------------------------------

location of the balloon during placement. The balloon is coated with silicone to assist with insertion of the catheter into the access needle.

The patient contacting components of the device such as the balloon, catheter coating, lubricant and radiopaque material are externally communicating tissue/bone/dentin-limited contact ≤24 hrs.

Associated Accessories include:

  • Access cannula/stylet ●
  • Syringe
  • Inflator ●
  • Hand Drill ●
  • Cement Tube ●

5. Principles of Operation / Mechanism of Action

The balloon catheter is the functional part of the device that creates a cavity and reduces the fracture. The balloon is designed to create a cavity by compressing cancellous bone and/or moving cortical bone as it inflates. After manufacturing, the balloon is covered with a protector called a sheath that the physician will remove prior to performing a procedure. Radiopaque markers provide for fluoroscopic visualization of the vertebral balloon prior to filling it with contrast media. The radiopaque balloon markers are located within the balloon.

After placing the access cannula into the fractured vertebral body, the physician places the access cannula and stylet in the desired location within the vertebral body, creating a channel for the balloon. The stylet is then removed, leaving the cannula in place. The deflated balloon is inserted through the access cannula. Contrast medium is injected into the balloon to inflate it. As the balloon inflates, it moves and compresses the cancellous bone, creating a void. The balloon is deflated and removed before cement injection takes place. Cement is then injected through the access cannula into this space.

6. Intended Use/Indications for use

The Stryker iVAS® Elite Inflatable Vertebral Augmentation System is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The system is to be used with cleared spinal Polymethylmethacrylate (PMMA) bone cements and Cortoss® Bone Augmentation Material indicated for use during percutaneous vertebral augmentation procedures, such as kyphoplasty.

{5}------------------------------------------------

7. Comparison of Technological Characteristics with the Predicate Devices

The subject and primary predicate devices are identical in the following ways:

  • Indications for use
  • Design intent ●
  • Classification ●
  • FDA product codes and regulation ●
  • Intended use
  • Regulation medical specialty
  • Placed percutaneously under fluoroscopy
  • Uses syringe for inflation ●
  • Balloon diameter
  • Balloon wall thickness ●
  • Sheath material ●
  • Radiopaque balloon markers ●
  • Radiopaque balloon material
  • Sliding pin material ●
  • Catheter material ●
  • Balloon material ●
  • Balloon lubricant
  • Depth markers
  • Color of shaft
  • Color of distal tip ●
  • Color of hub
  • Sterilization method ●
  • SAL ●
  • Single use ●
  • Maximum inflation pressure ●

The subject and primary predicate devices differ in the following ways:

  • Sheath color
  • Tamp size
  • Maximum recommended inflation volume ●

The subject and secondary predicate devices are identical in the following ways:

  • . Tamp size
  • Maximum recommended inflation volume

8. Non-Clinical

8.1 Biocompatibility

Biocompatibility testing, adopted per K172116, was performed following the recommendations of ISO 10993-1 and FDA Guidance (Use of International Standard ISO- 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" June 2016) as appropriate for limited exposure (<24 hours) externally communicating, tissue/bone/dentin devices. All biocompatibility testing met the requirements of the respective test method, thus supporting the biocompatibility of the subject device iVAS ® Elite Balloon Catheter.

{6}------------------------------------------------

The biocompatibility testing confirms that the subject device is non-sensitizing, non-irritating and non-toxic (cytotoxic and systemic). The biocompatibility testing met all acceptance criteria deeming this to be a biocompatible medical device.

8.2 Bench

A Risk Management File (RMF), in compliance with EN ISO 14971:2012, Medical Devices-Application of risk management to medical devices, was completed to assess the impact of the subject device modification. In accordance with the risk assessment, verification and validation testing was performed in accordance with EN ISO 14971:2012 in an effort to mitigate risk where possible.

All bench testing pertaining to the subject device modifications met their respective acceptance criteria as specified per the individual test report. Each test report was performed using industry accepted standards. The testing was conducted using the final finished device in a manner as similar as possible to how the balloon will be used by the healthcare professional. The performance testing demonstrates equivalence because the subject device is shown to have the same functional characteristics as the predicate device. Per the risk management assessment, the subject device does not introduce any new types of safety and effectiveness questions.

The acceptance criteria for the subject device are similar to the acceptance criteria for the predicate devices, demonstrating equivalence of the subject device to the predicate balloon catheter devices.

Catheter Flexibility Testing and Cold Age testing have been adopted from K093429. Accelerated Age testing has been adopted from K172116. Real-time testing is currently underway for the subject device.

Cadaveric testing was completed to provide evidence that the use of the subject balloon with a high inflation pressure of 808psi would not cause any new concerns regarding safety and effectiveness. This cadaveric testing was completed by a physician using the subject device in a manner as similar as possible to how the balloon would be used in the clinical setting. The subject balloon was inflated under fluoroscopic guidance. Following the inflation of the balloon, the physician evaluated nearby anatomical structures for adverse events related to the high-pressure balloon inflations. There were no adverse events following the inflations with the subject balloon catheter. This testing confirms that the modifications to the subject device do not raise new questions of safety and effectiveness.

{7}------------------------------------------------

Table 2 Design Controls

Sheath Colorant change
ModificationTest PerformedAcceptance CriteriaResult
Sheath colorant changeBiocompatibility (sheath)Adoption (K172116)Pass
Balloon Catheter Length and Volume Change
ModificationTest PerformedAcceptance CriteriaResult
Balloon tamp size changes sincethe predicate 510(k) submissionK172116:Balloon tamp length is shorter than the predicate Balloon tamp is longer than the predicate Maximum recommended inflation volume larger than predicate Maximum recommended inflation volume smaller than predicateConstrained BurstConstrained Burst was performed tosimulate the user inflating theballoon beyond the maximum ratingstated in the Instructions for Use.The subject device met allacceptance criteria.Pass
Un-Constrained Burst, Length toDiameter and One-way ValveTorqueUnconstrained Burst was performedto verify that the balloon meets aminimum open air inflation volume.The subject device met allacceptance criteriaPass
Insert, Retract, Sheath Removal,Assembly Strength, Tensile ForceInsertion and Retraction Force wasperformed to show the forcerequired to insert and remove thecatheter into the access needle. Thesubject device met all acceptancecriteria.Pass
Sheath shipping testBalloon Sheath must remain onballoon after bulk shippingconfigurationPass
Cadaveric testingNo damage to adjacent anatomicalstructures following high pressureballoon inflationPass
Packaging
ModificationTest PerformedAcceptance CriteriaResult
Same as listed aboveReal-time packaging testingDevice must meet all acceptancecriteria while determining expirationdate using either accelerated or real-time aging methods.Pass

{8}------------------------------------------------

9. Clinical Testing

No clinical testing was deemed necessary for this 510(k).

10. Substantial Equivalence Conclusion

The subject and the predicate devices have the same intended use, fundamental scientific technology, principle of operation, and mode of action. Although there are slight differences in the technological characteristics between the subject and predicate devices, these differences have been found to be insignificant overall.

All bench testing of the subject device met its respective acceptance criteria as specified per the individual test report. Each test report was performed using industry accepted standards as listed in Section 12 Standard Summary Reports. The testing was conducted using the final finished device in a manner as similar as possible to how the subject device will be used by the healthcare professional. The acceptance criteria for the subject device is similar to the acceptance criteria for the predicate device. Showing that the subject device meets the acceptance criteria at the same identified risk level (low risk) demonstrates equivalence to the predicate devices. The performance testing further demonstrates equivalence because the subject device is shown to have the same functional characteristics as the predicate devices.

The modifications to the subject Stryker iVAS® Elite Balloon Catheter do not raise new types of safety and effectiveness questions. Therefore, the subject Stryker iVAS® Elite Balloon Catheter is substantially equivalent to the predicate device, the Stryker iVAS® Elite Balloon Catheter (K172116) and secondary predicate Stryker iVAS® Balloon Catheter (K123942)

N/A