Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K181752
    Date Cleared
    2018-12-21

    (172 days)

    Product Code
    Regulation Number
    888.1100
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K093429

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stryker IVAS® Elite Inflatable Vertebral Augmentation System is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The system is to be used with cleared spinal Polymethylmethacrylate (PMMA) bone cements and Cortoss® Bone Augmentation Material indicated for use during percutaneous vertebral augmentation procedures, such as kyphoplasty.

    Device Description

    The Stryker iVAS ® Elite Balloon Catheter is a bone tamp with an inflatable component at the distal end. The balloon is inflated to create a void within the vertebral body. It is used with various accessories during vertebral augmentation.

    When the balloon is inflated with radiopaque fluid, the balloon expands axially and radially. Two radiopaque markers are fixed near the distal end of the balloon to show the location of the balloon during placement. The balloon is coated with silicone to assist with insertion of the catheter into the access needle.

    AI/ML Overview

    This looks like a 510(k) summary for a medical device that isn't an AI/ML device. Therefore, the questions about acceptance criteria for AI/ML performance metrics (like sensitivity, specificity, AUC, human reader improvement with AI assistance, sample sizes for training/test sets, expert ground truth establishment, etc.) are not applicable here.

    This document describes a conventional medical device (an inflatable vertebral augmentation system) and focuses on engineering and biological performance rather than diagnostic or analytical accuracy.

    However, I can extract the acceptance criteria and the study details for the conventional device as presented in the document:

    1. A table of acceptance criteria and the reported device performance

    Modification & Test PerformedAcceptance CriteriaReported Performance (Result)
    Sheath Colorant change: Biocompatibility (sheath)Adoption (K172116)Pass
    Balloon Catheter Length and Volume Change: Constrained BurstSimulate user inflating balloon beyond maximum rating; device met all acceptance criteria.Pass
    Balloon Catheter Length and Volume Change: Un-Constrained Burst, Length to Diameter and One-way Valve TorqueVerify balloon meets a minimum open air inflation volume; device met all acceptance criteria.Pass
    Balloon Catheter Length and Volume Change: Insert, Retract, Sheath Removal, Assembly Strength, Tensile ForceShow force required to insert and remove the catheter into the access needle; device met all acceptance criteria.Pass
    Balloon Catheter Length and Volume Change: Sheath shipping testBalloon Sheath must remain on balloon after bulk shipping configuration.Pass
    Balloon Catheter Length and Volume Change: Cadaveric testingNo damage to adjacent anatomical structures following high pressure balloon inflation.Pass
    Packaging: Real-time packaging testingDevice must meet all acceptance criteria while determining expiration date using either accelerated or real-time aging methods.Pass

    2. Sample sized used for the test set and the data provenance

    • Biocompatibility: No specific sample size is mentioned for biocompatibility testing in this summary, but it states "Biocompatibility testing... was performed following the recommendations of ISO 10993-1 and FDA Guidance."
    • Bench Testing (Constrained Burst, Un-Constrained Burst, etc.): No specific sample sizes are mentioned, but it indicates "All bench testing pertaining to the subject device modifications met their respective acceptance criteria as specified per the individual test report." These tests are typically performed on a statistically relevant sample size of manufactured devices.
    • Cadaveric Testing: "Cadaveric testing was completed to provide evidence..." The specific number of cadavers or test instances is not provided in this summary. The provenance is cadaveric (human anatomical specimens).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Cadaveric testing: "This cadaveric testing was completed by a physician..." The exact number of physicians is not specified, nor are their specific qualifications beyond being a "physician."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Cadaveric testing: "Following the inflation of the balloon, the physician evaluated nearby anatomical structures for adverse events related to the high-pressure balloon inflations." This implies a single physician assessment; no adjudication method (like 2+1) is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This is not an AI/ML device. No MRMC comparative effectiveness study was conducted or is applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is not an AI/ML device. No standalone algorithm performance was assessed.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Biocompatibility: Ground truth is established by standardized biological assays and regulatory guidelines (ISO 10993-1, FDA Guidance).
    • Bench Testing: Ground truth is established by engineering specifications, physical measurements, and industry-accepted standards (e.g., burst pressure, tensile strength).
    • Cadaveric Testing: Ground truth is established by direct visual inspection and assessment by a physician for damage to anatomical structures.

    8. The sample size for the training set

    • This is not an AI/ML device. There is no concept of a "training set" in the context of this submission.
    • However, the submission does mention that "Biocompatibility testing... adopted per K172116" and "Catheter Flexibility Testing and Cold Age testing have been adopted from K093429. Accelerated Age testing has been adopted from K172116." These refer to prior submissions that likely involved biocompatibility testing and accelerated aging on a set of devices, which could be considered an analogous "training" or foundational dataset for these aspects, but not in the AI/ML sense.

    9. How the ground truth for the training set was established

    • Not applicable as this is not an AI/ML device. The "ground truth" for the various engineering and biocompatibility tests would have been established through the methods and standards relevant to those specific test types in the original K093429 and K172116 submissions.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1