K Number
K181501

Validate with FDA (Live)

Device Name
Afina
Date Cleared
2018-11-02

(148 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Afina software is intended for importing, processing and displaying of retinal images and for the measurement of autofluorescence in the superior field (a circular 50-degree field of view at 20 degrees above fixation) of the retina. Afina is indicated for use only with images collected with validated ophthalmic cameras.

Device Description

Afina is software that transforms a set of field-identical autofluorescent (AF) images captured with a fundus camera into a single enhanced image with improved quality characteristics. It identifies and quantifies AF findings over the retinal AF background within the derived image.

Afina supports licensed healthcare professionals in importing, processing and displaying of retinal images and for the measurement of autofluorescence in the superior field of the retina.

Functionally, Afina imports a set of at least 8 AF images taken of the superior field of the retina (also referred to as a DICOM study), combines them in order to enhance image quality, and then identifies pixels and areas of contiquous pixels (spots) that exhibit autofluorescence that is brighter than the retinal AF background. Once identified, these pixels are quantified as both a count of spots and an overall sum of pixels. These data are consolidated into a single-field patient report that presents images of the field with AF findings identified and spot and pixel counts quantified.

AI/ML Overview

The information provided is not a standard document that details acceptance criteria and the results of a study to prove they are met. Instead, it is an FDA 510(k) summary for a medical device (Afina). While it discusses performance data and a clinical study, it does not present acceptance criteria in a quantitative, tabular format with direct comparison to reported performance values.

However, I can extract and infer information to construct a response that addresses the requested points as much as possible, interpreting "acceptance criteria" through the lens of the performed repeatability and reproducibility study.

Here's an attempt to structure the answer based on the provided text:

Acceptance Criteria and Device Performance Study for Afina

The Afina software is intended for importing, processing, and displaying retinal images and for the measurement of autofluorescence in the superior field of the retina. The study aimed to evaluate the repeatability and reproducibility of the Afina autofluorescence measurement for both Pixel Count and Spot Count.

1. Table of Acceptance Criteria and Reported Device Performance

The FDA 510(k) summary does not explicitly state quantitative "acceptance criteria" as pass/fail thresholds for specific performance metrics. Instead, it characterizes the variability of the measurements. From the study conclusion, the device is deemed "fit for its purpose as a measurement tool" based on the observed variability. Therefore, the reported performance values themselves become the characterized "accepted" performance for this device.

MetricAcceptance Criteria (Implicitly from Study Conclusion)Reported Device Performance
Spot Count PrecisionThe device should demonstrate characterized repeatability and reproducibility for Spot Count measurements, indicating it is fit for purpose as a measurement tool.Repeatability SD: 52.4Repeatability %CV: 11.6%Reproducibility SD: 89.9Reproducibility %CV: 19.9%(Across all Camera/Operator Combinations)
Pixel Count PrecisionThe device should demonstrate characterized repeatability and reproducibility for Pixel Count measurements, indicating it is fit for purpose as a measurement tool.Repeatability %CV: 8.7%Reproducibility %CV: 14.4%(Across all Camera/Operator Combinations. Note: Repeatability SD is variable and not explicitly reported in summary table.)
SafetyNo adverse events, serious adverse events, or anticipated adverse device effects should be reported during the study.No safety events reported.
Software FunctionalitySoftware functions as intended and is functionally equivalent to the predicate device, with differences in technological characteristics not raising new safety/effectiveness questions."Verification and validation activities met their acceptance criteria successfully per test plan and proved that the product, Afina, meets its requirements and intended use."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Clinical Test Set: 32 eligible subjects were screened, consented, and enrolled. Of these, 25 eyes had Afina qualified run data (194 completed measurements across various camera/operator combinations).
  • Data Provenance: The study was conducted prospectively between April 21, 2017, and August 25, 2017. The country of origin of the data is not explicitly stated, but typically 510(k) submissions for US market clearance involve studies conducted in the US or in compliance with international guidelines recognized by the FDA. Given the FDA submission, it's highly likely to be US data or internationally recognized clinical data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The provided text describes a repeatability and reproducibility study for a measurement tool. This type of study primarily assesses the consistency of the device's output, rather than its accuracy against a "ground truth" established by experts for classification or diagnosis. Therefore, there is no mention of experts establishing a ground truth for individual cases in the test set, as the study focuses on the precision of the quantitative measurements (Pixel Count and Spot Count) rather than diagnostic accuracy.

4. Adjudication Method for the Test Set

As the study focused on quantitative measurement precision rather than expert-derived ground truth for classification, an adjudication method for a test set (e.g., 2+1, 3+1) is not applicable and therefore not mentioned.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not reported. The study focused on the repeatability and reproducibility of the device's measurements, not on its impact on human reader performance.

6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) was Done

Yes, the clinical study evaluating repeatability and reproducibility was effectively a standalone performance study of the Afina algorithm's measurements. While operators collected the images, the analysis of Pixel Count and Spot Count is performed by the Afina software. The results characterize the intrinsic variability of the device's measurement functionality.

7. The Type of Ground Truth Used

The study did not use a "ground truth" in the sense of an independent gold standard for diagnosis (e.g., pathology, outcomes data, expert consensus). Instead, the study measured the repeatability and reproducibility of the device's intrinsic measurements (autofluorescence Pixel Count and Spot Count). The study design intended to demonstrate that the device provides consistent measurements under varying conditions (same operators/cameras vs. different operators/cameras).

8. The Sample Size for the Training Set

The document does not provide information about the sample size used for the training set of the Afina software's algorithms. The clinical performance data presented refers to a validation study (test set), not a training set.

9. How the Ground Truth for the Training Set was Established

Since information regarding the training set's sample size is not provided, the method for establishing ground truth for the training set is also not discussed in this 510(k) summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

November 2, 2018

NeuroVision Imaging, LLC Pepe Davis VP of Regulatory Affairs and Ouality Assurance 1395 Garden Highway, Suite 250 Sacramento, CA 95833

Re: K181501

Trade/Device Name: Afina Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: NFJ Dated: September 20, 2018 Received: September 24, 2018

Dear Pepe Davis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely yours,

Alexander Beylin -S 2018.11.02 11:04:52 -04'00' for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181501

Device Name Afina

Indications for Use (Describe)

The Afina software is intended for importing, processing and displaying of retinal images and for the measurement of autofluorescence in the superior field (a circular 50-degree field of view at 20 degrees above fixation) of the retina. Afina is indicated for use only with images collected with validated ophthalmic cameras.

Type of Use (Select one or both, as applicable):

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

1 SUBMITTER'S NAME, ADDRESS, CONTACT PERSON AND DATE PREPARED (21 CFR §807.92(A) (1))

  • Applicant: NeuroVision Imaging, LLC 1395 Garden Highway, Suite 250 Sacramento, CA 95833 Establishment Registration Number: 3012256561
    Pepe Davis Contact Person: (925) 286-7432 pdavis@neurovision.com

October 22, 2018 Date Prepared:

2 NAME OF DEVICE AND CLASSIFICATION (21 CFR §807.92(A) (2))

Trade Name: Afina (Release 1.1) Common Name: Picture archiving and communications system Regulation Number: 21 CFR 892.2050 Classification Name: System, Image Management, Ophthalmic Product Code: NFJ Device Class: Class II

Predicate Device (21 CFR §807.92(a) (3)) 3

Predicate Device: Zeiss Retina Workplace, K170638

This predicate has not been subject to a design-related recall.

{4}------------------------------------------------

DEVICE DESCRIPTION (21 CFR §807.92(A) (4)) বা

Afina is software that transforms a set of field-identical autofluorescent (AF) images captured with a fundus camera into a single enhanced image with improved quality characteristics. It identifies and quantifies AF findings over the retinal AF background within the derived image.

Afina supports licensed healthcare professionals in importing, processing and displaying of retinal images and for the measurement of autofluorescence in the superior field of the retina.

Functionally, Afina imports a set of at least 8 AF images taken of the superior field of the retina (also referred to as a DICOM study), combines them in order to enhance image quality, and then identifies pixels and areas of contiquous pixels (spots) that exhibit autofluorescence that is brighter than the retinal AF background. Once identified, these pixels are quantified as both a count of spots and an overall sum of pixels. These data are consolidated into a single-field patient report that presents images of the field with AF findings identified and spot and pixel counts quantified.

A risk assessment has been performed to identify potential risks of Afina. The risks were evaluated and reduced as far as possible to ensure the safety and effectiveness of Afina. The software is to be used by healthcare professionals and does not directly interact with patients. The report produced by the Afina software provides an objective measurement of autofluorescence but does not provide specific clinical diagnosis. Interpretation of the measurements are made by healthcare professionals.

INDICATIONS FOR USE (21 CFR §807.92(A) (5)) 5

The Afina software is intended for importing, processing and displaying of retinal images and for the measurement of autofluorescence in the superior field (a circular 50-degree field of view at 20 degrees above fixation) of the retina. Afina is indicated for use only with images collected with validated ophthalmic cameras.

Technological Characteristics (21 CFR §807.92(a) (6)) 6

The Afina software is technologically similar to its predicate, the Zeiss Retina Workplace. Both devices are software analysis tools which import DICOM images of the retina and process them to produce segmented images and measurements of findings of interest. The Afina software can identify the fovea and optic nerve head to register images similarly to the Zeiss Retina Workplace. The Afina software combines fundus images and identifies findings of interest in autofluorescence images of the retina similarly to the Zeiss Retina Workplace's identification of findings of interest in OCT images of the retina. Overlays of interest are similarly presented on enhanced fundus images highlighting pixels and spots of autofluorescence above the retinal background autofluorescence. A measurement of this autofluorescence is provided

{5}------------------------------------------------

as a pixel count and as a spot count (where contiguous regions of highlighted pixels are counted).

The following table summarizes the characteristics of the Afina software as compared to the predicate device.

CharacteristicProposed DevicePredicate Device
Trade NameNeuroVision AfinaZeiss Retina Workplace 2.0
Product code(s)NFJNFJ
DeviceSystem, Image Management, OphthalmicSystem, Image Management, Ophthalmic
Regulation Number892.2050892.2050
ClassificationClass IIClass II
Intended UseIntended to import, process and display fundus images for viewing and quantification of posterior segment findingsIntended to import, process and display fundus images for viewing and quantification of posterior segment findings
Indications for UseThe Afina software is intended for importing, processing and displaying of retinal images and for the measurement of autofluorescence in the superior field (a circular 50-degree field of view at 20 degrees above fixation) of the retina.The Retina Workplace is a FORUM application intended for processing and displaying image and optical coherence tomography data. It is also intended for generating reports that contain results from optical coherence tomography and fundus photography.The Retina Workplace uses CIRRUS algorithms and normative databases as quantitative tool for the comparison of macular
CharacteristicProposed DevicePredicate Device
Trade NameNeuroVision AfinaZeiss Retina Workplace 2.0
thickness data to a databaseof normal subjects.The Retina Workplace isintended to aid trainedhealthcare professionals inthe detection, monitoring andmanagement of oculardiseases including, but notlimited to, macular holes,cystoid macular edema,diabetic retinopathy and age-related macular degeneration.
Prescription or OTC?PrescriptionPrescription
Software/Hardware?Software OnlySoftware Only
Software PlatformLinuxWindows
Image data management?YesYes
Import of images of theretinaYesYes
Acceptance of imagesYesYes
Process images and datafrom DICOM filesYesYes
Register fundus imagesYesYes
Identify center of fundusYesYes
CharacteristicProposed DevicePredicate Device
Trade NameNeuroVision AfinaZeiss Retina Workplace 2.0
Analyze retinal images to identify and measure fundus findings of interestYesYes
Export reports with segmentation overlaysYesYes
Export reports with measurements of fundus findingsYesYes
Centralized storage of images and data?NoNo

Table 1: Substantial Equivalence Comparison

{6}------------------------------------------------

{7}------------------------------------------------

Performance Data (21 CFR §807.92(B)) 7

Non-clinical Performance Data (21 CFR §807.92(b) (1)) 7.1

The Afina 1.1 software was identified as having a moderate level of concern as defined in FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software documentation included:

Afina Software Development Plan

Afina Risk Management Plan

Afina Risk Analysis

Afina Risk Management Report

Afina Software Requirements

Afina Architecture Design Description

Afina Software Design Description

Afina Requirements Trace Matrix

Afina Software Verification Test Plan

Afina Software System Level Verification Test Report

{8}------------------------------------------------

Afina Software Unit Level Verification Test Report Afina Validation Test Plan

Afina 1.1 Validation Test Report

Afina 1.1.0 Version Description Document

Full characterization of the technical parameters of all the components of the software, including a description of the algorithms were provided in the Afina Software Requirements, the Afina Architecture Design Description, and the Afina Software Design Description.

A comprehensive risk analysis was provided for the software with detailed description of the hazards, their causes and severity as well as acceptable methods for control of the identified hazards.

The cybersecurity considerations of data confidentiality, data availability, denial of service attacks and malware were adequately addressed using a HIPAA compliant and HITRUST certified cloud hosting platform. Risks related to failure of various software components and their potential impact on patient reports and operator failures were also adequately addressed in the risk analysis. This software documentation information provided sufficient evidence of safe and effective software performance.

Performance testing, in the form of software bench testing, was conducted in the form of Unit, and System level verification against the documented product requirements. Verification and validation testing was performed to evaluate the performance and functionality of the software and the software has been found to perform as intended. Each function and/or feature was tested by means of an appropriate test case for the test specification. The verification testing demonstrates that the device performance complies with specifications and requirements identified for the Afina software. The software verification activities were divided into two phases:

• Unit level test cases which included verification of specific image processing functionality as described in the Afina Software Requirements and the Afina Architecture Description. These included functional tests for rejecting images, aligning images, enhancing images and segmentation.

• System level test cases which included verification of file transfer interfaces including image import, report output and cybersecurity.

Use Case validation was completed by medical professionals using a production equivalent implementation of the Afina with verified software including representative data that are representative of clinical cases). The validation participants executed test cases that simulated the use of the software in a clinical environment and completed questionnaires rating the various aspects of the software.

Test plans and reports were provided including passffail criteria and results for verification and validation activities at the unit and system level. Verification and validation activities met their acceptance criteria successfully per test plan and proved that the product, Afina, meets its requirements and intended use.

{9}------------------------------------------------

Clinical Performance Data (21 CFR §807.92(b) (2)) 7.2

A clinical study was performed to evaluate the repeatability and reproducibility of the Afina autofluorescence measurement for both Pixel Count and Spot Count. The study population included subjects with diseases that manifest in the retina (i.e. various levels and forms of cognitive impairment and subjects with retinal disease exhibiting autofluorescent findings) and normal subjects in order to demonstrate repeatability and reproducibility over a representative dynamic range.

Summarv of Protocol

Thirty-two (32) eligible subjects were screened, consented and enrolled between April 21, 2017 and August 25, 2017. One eye of each subject underwent up to three repeated superior retina imaging runs with repositioning between runs (3 focal planes per run, with up to six images per focal plane) on each of three (3) randomly assigned Autofluorescence (AF) fundus cameras meeting the specification requirements for Afina (designated as Camera 1, 2 and 3). The operators were trained to follow the camera Instructions for Use (IFU) and the Afina image capture procedure outlined in the Afina IFU. The three trained camera operators were designated as Operator A, B and C, with two trained back-up operators. To evaluate reproducibility. Operator A used Camera 1. Operator B used Camera 2. and Operator C used Camera 3. Subjects completed imaging during one visit, and within one week of consent.

Results

The demographics of the enrolled study population are shown in Table 2. The gender of the study subjects was reasonably distributed and the age of the study subjects ranged from 44 years to 81 years, with a median age of 72. Therefore, this is a good sample population for the purposes of characterizing the precision of the measurements provided by the Afina software.

{10}------------------------------------------------

Table 2: Demographic Summary

n/N%
Gender
Male18/3256%
Female14/3244%
Eye
Left (OS)9/3228%
Right (OD)23/3272%
Age (year)
N32
Mean68
SD11
Median72
Min44
Max81
40 - 494/3212.5%
૨૦ - ૨૦4/3212.5%
60 - 695/3215.6%
70 - 7915/3246.9%
80 - 894/3212.5%
Ethnicity
Asian/Pacific Islander2/326.3%
Black or African American2/326.3%
Caucasian27/3284.3%
Other1/323.1%

Of the 32 eyes imaged (288 planned measurement runs), 25 eyes had Afina qualified run data (194 completed measurements). The range of the autofluorescence measurements obtained from the 25 study eyes with Afina measurements was 2,222 to 58,110 for Pixel Count and 37 to 1350 for Spot Count.

Table 3 summarizes the results for the Spot Count measurements.

{11}------------------------------------------------

StatisticsAllCamera/OperatorCombinationsCamera 1 /Operator ACamera 2 /Operator BCamera 3 /Operator C
Number of Eyes25222524
Number of Successful194636665
Mean Spot Count Value451.1429.0429.2495.4
Total Standard Deviation214.4212.3192.4229.8
Repeatability SD152.446.848.660.4
Repeatability %CV211.610.911.312.2
Reproducibility SD389.9---
Reproducibility %CV419.9---

Table 3: Spot Count Precision Summary

1 Repeatability SD = Estimate of the standard deviation among measurements taken on the same eye using the same camera in the same testing session with repositioning.

2 Repeatability %CV = Repeatability SD = Mean x 100. measurements taken on the same eye using different cameras, including repeatability.

Reproducibility SD =Estimate of the standard deviation among measurements taken 3 with different camera/operators

  • 4 Reproducibility %CV = Reproducibility SD + Mean x 100.
    Table 4 summarizes the results for the Pixel Count measurement analysis. Because the measurement fits a constant %CV model, only the Repeatability %CV and Reproducibility %CV are reported here. The repeatability SD (0.087 on the log scale but variable in the original scale) is not presented in this summary table.

Table 4: Pixel Count Precision Summary

StatisticsAllCamera/OperatorCombinationsCamera 1 /Operator ACamera 2 /Operator BCamera 3 /Operator C
Number of Subjects25222524
Number of Successful194636665
Mean Pixel Count Value18398772688428600
Repeatability % CV18.79.58.18.2
Reproducibility % CV114.4---

1 Calculated according to NIST Engineering Statistics Handbook, Section 1.3.6.6.9. Lognormal Distribution: http://www.itl.nist.gov/div898/handbook/eda/section3/eda3669.htm

Subjects were assessed during and after the imaging visit, and received a follow-up telephone call no later than two days following the visit, to assess for any adverse effects. There were no

{12}------------------------------------------------

adverse events, serious adverse events, or anticipated adverse device effects reported during the study.

Study Conclusion

The repeatability and reproducibility study showed the device to be safe and characterized well the variability of the measurements provided by the Afina software. There were no safety events reported.

The variability of the autofluorescence measurements provided by the Afina software shows the device is fit for its purpose as a measurement tool.

7.3 Conclusions (21 CFR §807.92(b) (3))

The verification and validation testing demonstrate that the Afina software performs as intended and is functionally equivalent to the predicate device. The non-clinical and clinical performance data confirm that differences in technological characteristics do not raise different questions of safety and effectiveness.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).