K Number
K170638

Validate with FDA (Live)

Device Name
Retina Workplace
Date Cleared
2017-06-21

(111 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Retina Workplace is a FORUM application intended for processing and displaying image and optical coherence tomography data. It is also intended for generating reports that contain results from optical coherence tomography and fundus photography.

The Retina Workplace uses CIRRUS algorithms and normative databases as a quantitative tool for the comparison of macular thickness data to a database of normal subjects.

The Retina Workplace is intended to aid trained healthcare professionals in the detection, monitoring and management of ocular diseases including, but not limited to, macular holes, cystoid macular edema, diabetic retinopathy and age-related macular degeneration.

Device Description

The Retina Workplace is a software application to a Class II medical device regulated under 21 CFR §892.2050 and therefore is a Class II medical device. As a software application of the FORUM PACs system (server), the Retina Workplace client is designed to process and display CIRRUSTM HD optical coherence tomography (OCT) data and images provided by Fundus photography.

More specifically, the Retina Workplace processes the retrieved OCT exam data from CIRRUS HD-OCT and CIRRUS Photo images from the FORUM server to display the results of macular thickness analysis (MTA), macular change analysis (MCA), en face Analyses with predefined pre-sets. Advanced RPE Analysis, and raster scans (5 Line. HD 5 Line, 1 Line, HD 1 Line (100x), HD 21 Line, HD Cross, and HD Radial).

The MTA results may be compared against an age-matched population of normal subjects. Analysis and databases comparison is performed by using algorithms and normative databases previously cleared in CIRRUS HD-OCT (K150977) and CIRRUS Photo (K133217) predicate devices. The reports generated by the Retina Workplace can be displayed, printed, and the patient record can be stored back on the FORUM server database.

The Advanced RPE Analysis allows the user to examine the status of the RPE in greater detail than the Macular Thickness Analysis. In particular, Retina Workplace provides two algorithms, one to identify and measure areas of sub-RPE illumination where the OCT is able to penetrate through to the choroid, indicating that the RPE is atrophic (often associated with geographic atrophy), and one to identify and measure elevations in the RPE (often associated with drusen)."

Retina Workplace can automatically register images generated by Fundus photography with other Fundus images e.g. the en face OCT Fundus image when provided in a valid DICOM format and available in the FORUM Archive.

The fundus to fundus registration is part of the predicate devices Cirrus photo (K133217), albeit the function on the predicate device is limited to register image by Fundus photography with en face OCT Fundus images. Retina Workplace version 2.0 extended this functionality to handle LSLO Fundus image. Specifically, to register Fundus image to LSLO Fundus image and LSLO Fundus image to LSLO Fundus image. This registration functionality is used to align HD raster scan pattern to Fundus images, an equivalent function is part of the predicate device CIRRUS photo (K133217). For all performed registrations, the user has the option to register the fundus images manually.

The automatic/manual Fovea registration features (Fovea Finder™ and AutoCenter™) are currently cleared on the CIRRUS™ HD-OCT system (K150977) and CIRRUSTM Photo (K133217) devices and are also included in the Retina Workplace, since version 1.0 (K150467).

The Retina Workplace provides the same overlays as included in the currently cleared CIRRUS HD-OCT system such as:

Introduced with the Retina Workplace Release 2.0

  • . En face analyses overlays, with different pre-sets based on the mean / average optical reflectivity values
  • . Advanced RPE Analysis overlays with illumination and elevation map
  • . Fundus images overlaid with raster scan pattern

Introduced with the previous release of Retina Workplace (Release 1.0, K150467)

  • . Fundus image overlay with another Fundus image. (Fundus to Fundus registration)
  • . Macular thickness (ILM-RPE) overlay on the Fundus image. (Inner limiting membrane / retinal pigment epithelium)
  • Macular change overlay ●
  • ETDRS overlay on the Fundus image. (Early Treatment of Diabetic Retinopathy Study) .

Overlays of Retina Workplace not provided by CIRRUS OCT devices:

  • MTA thickness map overlaid with raster scan pattern .
  • En face analysis overlay with pre-set to create an en face image based on the minimum . optical reflectivity values
AI/ML Overview

The Retina Workplace is a software application intended for processing and displaying image and optical coherence tomography data, generating reports containing results from OCT and fundus photography, and aiding healthcare professionals in the detection, monitoring, and management of ocular diseases.

Here's an analysis of the acceptance criteria and study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Implicit from testing type)Reported Device Performance
Software functionality and performance meets specifications and requirements.Passed Unit, System Integration, and System testing against documented product requirements. Verification and validation testing showed the device performs as intended.
Results generated by Retina Workplace are equivalent to those from original CIRRUS HD-OCT and CIRRUS Photo.In direct comparison tests, reports generated by Retina Workplace using the same test data were found to be equivalent to those from the original CIRRUS HD-OCT and CIRRUS Photo.
Clinical functionalities are validated and meet intended use.Ophthalmologists used a production-equivalent Retina Workplace system with representative data, executed simulated clinical test cases, and completed questionnaires rating aspects of the software, indicating successful validation.
Compatibility with specified operating systems.Compatible with specified Windows server and client operating systems (Windows 7/8.1/10, Windows Server 2008 R2/2012 R2) and OS X (Yosemite, El Capitan).
DICOM compliance.The device is DICOM compliant as per the statement on the Carl Zeiss internet site.
No new potential safety risks introduced.Based on successful verification and validation testing, it is the manufacturer's opinion that the Retina Workplace introduces no new potential safety risks.
Safe and effective when used in accordance with Instructions for Use.All necessary testing was conducted to ensure the device is as safe and effective as predicate devices.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: The document does not specify a numerical sample size for the test set used in the "Non-Clinical Validation Test Results." It mentions "representative data (sample data that is representative of clinical cases)."
  • Data Provenance: The document explicitly states "sample data that is representative of clinical cases." It does not specify the country of origin or whether it was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Number of Experts: The document states "Validation of clinical functionalities (Use Case) was completed by ophthalmologists." It does not specify the exact number of ophthalmologists.
  • Qualifications of Experts: The experts are identified as "ophthalmologists," implying they are qualified medical professionals specializing in eye care. No further details regarding their years of experience or specific sub-specialties are provided.

4. Adjudication Method for the Test Set:

  • The document does not describe an adjudication method for establishing ground truth for the test set. The clinical validation involved ophthalmologists "executing test cases that simulated the use of the device in a clinical environment and completed questionnaires rating the various aspects of the software." This suggests their assessments were used to evaluate the device's performance, but not necessarily a formal adjudication process for ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not reported. The study focused on verifying the software's performance against specifications and existing cleared devices, and on the usability and functionality as perceived by ophthalmologists. It did not compare human readers with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • Yes, a standalone performance assessment was effectively done for key algorithms and report generation. The document states: "in a direct comparison test, the original OCT test reports generated on the CIRRUS HD-OCT and CIRRUS photo were compared to the reports generated by the Retina Workplace using the same test data, to verify that the results contained in both reports were equivalent." This indicates that the Retina Workplace's algorithms for generating these reports were evaluated independently of human interpretation in this specific test. The Retina Workplace also uses "CIRRUS algorithms and normative databases" which were previously cleared and are implicitly part of standalone algorithm performance.

7. The Type of Ground Truth Used:

  • The ground truth for the "direct comparison test" appears to be the results from previously cleared CIRRUS HD-OCT and CIRRUS Photo devices. The aim was to ensure equivalence, meaning these established devices served as the reference.
  • For the clinical functionality validation, the "ground truth" was derived from the assessments and feedback of ophthalmologists during simulated clinical use, rating various aspects of the software. This is more akin to an expert consensus/usability evaluation rather than a definitive disease pathology ground truth.

8. The Sample Size for the Training Set:

  • The document does not specify a sample size for a training set. This submission is for a software application that largely implements and integrates previously cleared algorithms and normative databases from other devices (CIRRUS HD-OCT, CIRRUS Photo). Therefore, the "training" (if applicable from an AI/machine learning perspective) would have occurred during the development and clearance of those original CIRRUS algorithms and databases, not for the Retina Workplace itself.

9. How the Ground Truth for the Training Set Was Established:

  • Since there's no explicit training set mentioned for the Retina Workplace itself, the establishment of ground truth would refer to the processes used for the underlying CIRRUS algorithms and normative databases it utilizes. The document indicates that the Retina Workplace uses "CIRRUS algorithms and normative databases... previously cleared in CIRRUS HD-OCT (K150977) and CIRRUS Photo (K133217) predicate devices." For these previous clearances, the ground truth for their normative databases would typically have been established through a rigorous process involving expert consensus, clinical studies, and potentially pathology or outcome data to define "normal" and characterize various disease states, but the details are not provided in this specific document.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three faces in profile, stacked on top of each other. The faces are stylized and appear to be part of a single, continuous line.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 21, 2017

Carl Zeiss Meditec Ag % Dong Hua Sr. Regulatory Affairs Specialist Carl Zeiss Meditec, Inc. 5160 Hacienda Drive Dublin, CA 94568

Re: K170638

Trade/Device Name: Retina Workplace Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: NFJ Dated: May 18, 2017 Received: May 23, 2017

Dear Dong Hua:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"

(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Kesia Alexander

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170638

Device Name Retina Workplace

Indications for Use (Describe)

The Retina Workplace is a FORUM application intended for processing and displaying image and optical coherence tomography data. It is also intended for generating reports that contain results from optical coherence tomography and fundus photography.

The Retina Workplace uses CIRRUS algorithms and normative databases as quantitative tool for the comparison of macular thickness data to a database of normal subjects.

The Retina Workplace is intended to aid trained healthcare professionals in the detection, monitoring and management of ocular diseases including, but not limited to, macular edema, diabetic retinopathy and age-related macular degeneration.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

(as per 21 CFR §807.92)

The Retina Workplace

GENERAL INFORMATION

Manufacturer:Carl Zeiss Meditec AGGoeschwitzer Strasse 51-52D-07745 Jena, Germany+49 3641220-667 (phone)+49 3641220-282 (fax)Establishment Registration Number: 9615030
Contact Person:Dong HuaSr. Regulatory Affairs SpecialistCarl Zeiss Meditec, Inc.5160 Hacienda DriveDublin, CA 94568(925) 557-4204 Phone(925) 216-2184 Mobile(925) 557-4259 FaxE-mail: dong.hua@zeiss.com
Date prepared:June 14th, 2017
DeviceSystem, Image Management, Ophthalmic
Classification:21 CFR 892.2050
Device Class:II
Product Code:NFJ
Common Name:Picture Archiving and Communications System
Trade/Proprietary Name:Retina Workplace (Release 2.0)

{4}------------------------------------------------

The Retina Workplace is intended to be a clinically focused accessory to the FORUM Workplace. The Retina Workplace as described in this premarket notification has a similar intended use, indications for use (specific for the device), and is based on the same fundamental scientific technical characteristics as the predicate devices listed below.

Predicate Devices andClearancesClearanceDateSubstantially Equivalent To:Company
The Retina Workplace is aclinically focused workplacefor retina disorders and isaccessory to the FORUMPACs and is a client of theFORUM serverK150467 /May 21,2015The Retina Workplace is aclinically focused workplacefor retina disorders and isaccessory to the FORUMPACs and is a client of theFORUM server.Carl ZeissMeditec AG
FORUM® is a server basedPACs workplace supportinggeneric post-processing tools(display, pan, zoom,generating reports, etc.) andinterfaces with DICOM basedophthalmic devices.K122938 /November 2,2012The Retina Workplaceaccessory client for theFORUM® PACS systemproviding clinically focusedapplications.Carl ZeissMeditec AG
CIRRUSTM HD-OCTcontaining algorithms for thedetection and analysis ofophthalmic anatomy,including Retinal Nerve FiberLayer (RNFL), Macular,Optic Nerve Head andGanglion Cell NormativeDatabasesK150977 /September 1,2015The Retina Workplace as aclinically focused accessoryto the FORUM® PACSworkplace containingalgorithms and databases forthe analysis of oculardiseases including macularedema, and degeneration andother retinopathies.Carl ZeissMeditec AG
CIRRUS™ PhotoCombined fundus camera andOCT system for the imaging,display, storage, and reportgeneration of the Fundusimages and OCT images.K133217 /March 19,2014The Retina Workplace as aclinically focused accessoryto the FORUM® PACSworkplace imports OCT dataand Fundus images fordisplay, storage, and reportgenerationCarl ZeissMeditec AG

Substantial Equivalence Claimed To (21 CFR §807.92(a)(3))

{5}------------------------------------------------

INTENDED USE (21 CFR §807.92(a)(5))

The Retina Workplace is a FORUM application intended for processing and displaying image and optical coherence tomography data. It is also intended for generating reports that contain results from optical coherence tomography and fundus photography.

INDICATIONS FOR USE (21 CFR §807.92(a)(5))

The Retina Workplace is a FORUM application intended for processing and displaying image and optical coherence tomography data. It is also intended for generating reports that contain results from optical coherence tomography and fundus photography.

The Retina Workplace uses CIRRUS algorithms and normative databases as a quantitative tool for the comparison of macular thickness data to a database of normal subjects.

The Retina Workplace is intended to aid trained healthcare professionals in the detection, monitoring and management of ocular diseases including, but not limited to, macular holes, cystoid macular edema, diabetic retinopathy and age-related macular degeneration.

DEVICE DESCRIPTION (21 CFR §807.92(a)(4))

The Retina Workplace is a software application to a Class II medical device regulated under 21 CFR §892.2050 and therefore is a Class II medical device. As a software application of the FORUM PACs system (server), the Retina Workplace client is designed to process and display CIRRUSTM HD optical coherence tomography (OCT) data and images provided by Fundus photography.

More specifically, the Retina Workplace processes the retrieved OCT exam data from CIRRUS HD-OCT and CIRRUS Photo images from the FORUM server to display the results of macular thickness analysis (MTA), macular change analysis (MCA), en face Analyses with predefined pre-sets. Advanced RPE Analysis, and raster scans (5 Line. HD 5 Line, 1 Line, HD 1 Line (100x), HD 21 Line, HD Cross, and HD Radial).

The MTA results may be compared against an age-matched population of normal subjects. Analysis and databases comparison is performed by using algorithms and normative databases previously cleared in CIRRUS HD-OCT (K150977) and CIRRUS Photo (K133217) predicate devices. The reports generated by the Retina Workplace can be displayed, printed, and the patient record can be stored back on the FORUM server database.

The Advanced RPE Analysis allows the user to examine the status of the RPE in greater detail than the Macular Thickness Analysis. In particular, Retina Workplace provides two algorithms, one to identify and measure areas of sub-RPE illumination where the OCT is able to penetrate through to the choroid, indicating that the RPE is atrophic (often associated with geographic atrophy), and one to identify and measure elevations in the RPE (often associated with drusen)."

{6}------------------------------------------------

Retina Workplace can automatically register images generated by Fundus photography with other Fundus images e.g. the en face OCT Fundus image when provided in a valid DICOM format and available in the FORUM Archive.

The fundus to fundus registration is part of the predicate devices Cirrus photo (K133217), albeit the function on the predicate device is limited to register image by Fundus photography with en face OCT Fundus images. Retina Workplace version 2.0 extended this functionality to handle LSLO Fundus image. Specifically, to register Fundus image to LSLO Fundus image and LSLO Fundus image to LSLO Fundus image. This registration functionality is used to align HD raster scan pattern to Fundus images, an equivalent function is part of the predicate device CIRRUS photo (K133217). For all performed registrations, the user has the option to register the fundus images manually.

The automatic/manual Fovea registration features (Fovea Finder™ and AutoCenter™) are currently cleared on the CIRRUS™ HD-OCT system (K150977) and CIRRUSTM Photo (K133217) devices and are also included in the Retina Workplace, since version 1.0 (K150467).

The Retina Workplace provides the same overlays as included in the currently cleared CIRRUS HD-OCT system such as:

Introduced with the Retina Workplace Release 2.0

  • . En face analyses overlays, with different pre-sets based on the mean / average optical reflectivity values
  • . Advanced RPE Analysis overlays with illumination and elevation map
  • . Fundus images overlaid with raster scan pattern

Introduced with the previous release of Retina Workplace (Release 1.0, K150467)

  • . Fundus image overlay with another Fundus image. (Fundus to Fundus registration)
  • . Macular thickness (ILM-RPE) overlay on the Fundus image. (Inner limiting membrane / retinal pigment epithelium)
  • Macular change overlay ●
  • ETDRS overlay on the Fundus image. (Early Treatment of Diabetic Retinopathy Study) .

Overlays of Retina Workplace not provided by CIRRUS OCT devices:

  • MTA thickness map overlaid with raster scan pattern .
  • En face analysis overlay with pre-set to create an en face image based on the minimum . optical reflectivity values

RISK MANAGEMENT AND GENERAL SAFETY AND EFFECTIVENESS

The device labeling contains instructions for use and any necessary caution and notes, to provide for safe and effective use of the device.

{7}------------------------------------------------

Risk management is ensured via a risk analysis, which is used to identify potential hazards and mitigations. These potential hazards are controlled by software means, user instructions, verification of requirements, and validation of the clinical workflow to ensure that the product meets its intended uses. ZEISS adheres to recognized and established industry practice and relevant international standards where indicated.

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE (21 CFR §807.92(a) (6)):

The Retina Workplace is designed in conjunction with the FORUM PACS system is to support clinically focused client workplaces to aid the optometric and ophthalmology clinicians with the processing, display, review, management, and storage of digital data contained in patient records.

The Retina Workplace is substantially equivalent to the predicate devices with regard to the indications for use statement and is functionally equivalent to the predicate devices as follows:

  • . The Retina Workplace, Release 2.0 and the PRIMARY predicate device Retina Workplace, Release 1.0 (K150467), are clients of the FORUM PACS system and equivalent in design as they are clinically focused workplaces for retina disorders, which support the work flow of optometric and ophthalmic clinicians and their staff. Both workplaces can display retrieved OCT exams from CIRRUS OCT devices to be analyzed and reviewed
  • The Retina Workplace and the SECONDARY predicate device, Cirrus HD-OCT with . Software Version 8 (K150977) process OCT exam results. Macular thickness analysis results may be displayed in relation to a normative database specific to originating CIRRUS devices.
  • . The Retina Workplace and the SECONDARY predicate device, CIRRUS Photo (K133217), process OCT exam results. Macular thickness analysis results may be displayed in relation to a normative database specific to originating CIRRUS devices.
  • The Retina Workplace and the SECONDARY predicate device, FORUM ● (K122938), supports the generation of reports which contain the results from optical coherence tomography data and Fundus photography and a variety of overlays.
  • The Retina Workplace and the FORUM (K122938) are software devices, whereas the . predicate devices, CIRRUS HD-OCT (K150977) and CIRRUS Photo (K133217) are ophthalmic devices which have both software and hardware functionalities. The primary software functionalities to support clinically focused analysis of the retina in the predicate devices are implemented in the Retina Workplace.

Note: The release 2.0 of Retina Workplace supports the currently cleared macular thickness analysis (MTA), macular change analysis (MCA), en face analysis. Advanced RPE Analysis, and raster scan algorithms from the CIRRUS HD-OCT system. MTA

{8}------------------------------------------------

results may be compared against a normative database of age-matched population of normal subjects which has also been previously cleared in the CIRRUS HD-OCT system (K150977) and the CIRRUS Photo (K133217) devices. Additionally, the MTA, MCA, and 5 line raster analysis software applications and normative databases remain unchanged from the currently cleared version of Retina Workplace (K150467), Release 1.0.

PERFORMANCE DATA & SUMMARY OF VERIFICATION AND VALIDATION ACTIVITY (21 CFR §807.92(B)):

Bench Testing (21 CFR §807.92(b) (1))

Performance testing, in the form of software bench testing, was conducted on the Retina Workplace in the form of Unit, System Integration, and System testing against the documented product requirements. Verification and validation testing was performed to evaluate the performance and functionality of the software and the device has been found to perform as intended. Each function and/or feature was tested by means of an appropriate test case for the test specification. The verification testing demonstrates that the device performance complies with specifications and requirements identified for the Retina Workplace.

The software verification activities were divided into three phases:

  • . Tests accompanying development (including code inspections)
  • . Module and integration test phase - stabilization phase
  • . System verification

As part of the verification testing, in a direct comparison test, the original OCT test reports generated on the CIRRUS HD-OCT and CIRRUS photo were compared to the reports generated by the Retina Workplace using the same test data, to verify that the results contained in both reports were equivalent.

The software verification and regression testing has been performed successfully to meet their previously determined acceptance criteria as stated in the Test Plan.

Testing Environment:

The client and server operating systems were also evaluated during verification. The results determined that client, Retina Workplace, is suitable for the same operating systems for which the respective server. FORUM, is released and that it is suitable under the following operating systems:

Operating Systems (server):

  • Windows 7 (64 bit) with Service Pack 1 or
  • · Windows 8.1 (64 bit) or
  • · Windows Server 2008 R2 (64 bit) with Service Pack 1
  • · Windows Server 2012 R2
  • Windows 10 (64 bit)

{9}------------------------------------------------

Operating Systems (client):

  • · OS X 10.10 (Yosemite)
  • · OS X 10.11 (El Capitan)
  • · Windows Server 2008 R2 (64 bit)
  • · Windows Server 2012 R2 (64 bit)
  • · Windows 7 (64 bit)
  • · Windows 7 (32 bit)
  • · Windows 8.1 (64 bit)
  • Windows 10 (64 bit)

Non-Clinical Validation Test Results (21 CFR §807.92(b) (2))

Validation of clinical functionalities (Use Case) was completed by ophthalmologists using a production equivalent Retina Workplace system with verified software including representative data (sample data that is representative of clinical cases) installed on a segregated computer and used as a non-clinical system. The validation participants used the Retina Workplace system, executed test cases that simulated the use of the device in a clinical environment and completed questionnaires rating the various aspects of the software.

Verification and validation activities met their acceptance criteria successfully and proved that the product, the Retina Workplace, meets its requirements and intended uses.

Testing to Consensus Standards (21 CFR §807.92(b)(1))

The Retina Workplace has been tested to meet the requirements for conformity (where applicable) to multiple industry standards. The R&D evaluation of the relevant testing to consensus standards is documented. The Retina Workplace, is an accessory application to the FORUM PACS system.

The Retina Workplace is DICOM compliant according to the DICOM compliance statement as provided on the Carl Zeiss internet site: www.zeiss.com/dicom

SUBSTANTIAL EQUIVALENCE TO PREDICATES (21 CFR §807.92(B)(1)):

As previously indicated, the Retina Workplace has been tested to meet the product requirements (PRS) and software requirements (SRS) and is considered to be substantially equivalent to the Primary predicate device and secondary predicate devices as indicated above.

510(K) SUMMARY (21 CFR §807.92(C)):

In summary, based on the successful verification and validation testing to the software acceptance criteria, it is ZEISS' opinion that the Retina Workplace client, as an accessory to the FORUM PACs system (K122938), does not introduce any new potential safety risks and is substantially equivalent to, and performs as well as, the predicate devices.

{10}------------------------------------------------

Additionally, all testing deemed necessary was conducted on the Retina Workplace to ensure that the device is as safe and effective when used in accordance with its Instructions for Use as the predicate devices.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).