K Number
K181499

Validate with FDA (Live)

Date Cleared
2018-07-06

(29 days)

Product Code
Regulation Number
862.3250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DRI Cocaine Metabolite Assay is a homogeneous enzyme immunoassay intended for the qualitative and/or semi- quantitative determination of benzoylecgonine (Cocaine Metabolite) in human urine at a cutoff concentration of either 150 ng/mL or 300 ng/mL.

The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/tandem mass spectrometry (LC- MS/MS) or permitting laboratories to establish quality control procedures.

The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography / Mass spectrometry (GC/MS) or Liquid chromatography/ tandem mass spectrometry (LC-MS/MS) is the preferred confirmatory method. Tests for cocaine metabolite cannot distinguish between abused drugs and certain prescribed medications.

Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary results are used. For In Vitro Diagnostic Use Only.

Device Description

The DRI Cocaine Metabolite Assay is a homogeneous enzyme immunoassay using ready-to-use liquid reagents. The assay uses a specific antibody, which can detect benzoylecgonine in urine. The assay is based on the competition of an enzyme glucose-6-phosphate dehydrogenase (G6PDH) labeled drug and the drug from the urine sample for a fixed amount of specific antibody binding sites. In the presence of free drug from the sample, the free drug occupies the antibody binding sites, allowing the drug-labeled G6PDH to interact with the substrate, resulting in enzyme activity. In the absence of drug from the sample, the specific antibody binds to the drug labeled with G6PDH and the enzyme activity is inhibited. This phenomenon creates a direct relationship between the drug concentration in the urine and the enzyme activity. The enzyme G6PDH activity is determined spectrophotometrically at 340 mm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.

The assay consists of reagents (A and E).

Reagent A: Contains mouse monoclonal anti-benzoylecgonine antibody, glucose-6-phosphate (G6P), and nicotinamide adenine dinucleotide (NAD) in Tris buffer with sodium azide as preservative.

Reagent E: Contains benzoylecgonine analog labeled with glucose-6-phosphate dehydrogenase (G6PDH) in HEPES buffer with sodium azide as preservative.

AI/ML Overview

The DRI Cocaine Metabolite Assay is a homogeneous enzyme immunoassay intended for the qualitative and/or semi-quantitative determination of benzoylecgonine (Cocaine Metabolite) in human urine at cutoff concentrations of either 150 ng/mL or 300 ng/mL. The study performed aims to demonstrate the analytical performance of the device and its substantial equivalence to the predicate device, Cocaine Metabolite Enzyme Immunoassay (K960187).

Here's an analysis of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria values for agreement percentages (e.g., "must be >95%"). However, the reported performance demonstrates 100% agreement with the reference method (LC-MS/MS) for both positive and negative samples at both cutoff levels in the method comparison study. The precision studies also show consistent determination of positive and negative results at concentrations significantly above and below the cutoff, with expected mixed results at the cutoff itself.

Study ComponentAcceptance Criteria (Implicit/Inferred)Reported Device Performance (150 ng/mL Cutoff)Reported Device Performance (300 ng/mL Cutoff)
Precision (Qualitative)Consistent results at concentrations +/- cutoff, mixed at cutoffAt cutoff: 22/58 (N/P), <25% below: all negative, >25% above: all positiveAt cutoff: 31/49 (N/P), <25% below: all negative, >25% above: all positive
Precision (Semi-Quantitative)Consistent results at concentrations +/- cutoff, mixed at cutoffAt cutoff: 19/61 (N/P), <25% below: all negative, >25% above: all positiveAt cutoff: 22/58 (N/P), <25% below: all negative, >25% above: all positive
Spike Recovery (Qualitative)No overlap between results below and above cutoffAll 21 replicates below cutoff were Negative; All 21 replicates above cutoff were PositiveAll 21 replicates below cutoff were Negative; All 21 replicates above cutoff were Positive
Analytical Recovery & LinearityPercent recovery close to 100%Range: 95.9% to 108.9% (excluding 0 ng/mL)Consistent with 150 ng/mL, not explicitly stated separately
Method Comparison (Semi-Qualitative) - Agreement among PositivesHigh agreement with reference method100% (50/50)100% (50/50)
Method Comparison (Semi-Qualitative) - Agreement among NegativesHigh agreement with reference method100% (50/50)100% (50/50)
Method Comparison (Qualitative) - Agreement among PositivesHigh agreement with reference method100% (50/50)100% (50/50)
Method Comparison (Qualitative) - Agreement among NegativesHigh agreement with reference method100% (50/50)100% (50/50)
Specificity (Cocaine metabolites)Cross-reactivity % values acceptableBenzoylecgonine: 100%, Cocaine: 0.6%, Cocaethylene: 0.5%, Ecgonine: 0.17%, Ecgonine Methyl Ester: <0.15%, m-hydroxybenzoylecgonine: 50%, Norcocaine: <0.15%Benzoylecgonine: 100%, Cocaine: 0.6%, Cocaethylene: 0.5%, Ecgonine: 0.19%, Ecgonine Methyl Ester: <0.3%, m-hydroxybenzoylecgonine: 50%, Norcocaine: <0.3%
Specificity (Unrelated compounds)No interference at tested concentrationsLow control: Negative, High control: Positive for all tested compoundsLow control: Negative, High control: Positive for all tested compounds
Interference (Physiologic substances & pH)No interference at tested concentrationsLow control: Negative, High control: Positive for all tested compoundsLow control: Negative, High control: Positive for all tested compounds
Specific Gravity InterferenceNo interference across rangeLow control: Negative, High control: Positive for all tested rangesLow control: Negative, High control: Positive for all tested ranges

2. Sample Size for the Test Set and Data Provenance

  • Precision Studies:
    • For both qualitative and semi-quantitative modes and both cutoffs (150 ng/mL and 300 ng/mL): 80 replicates (2 replicates per day for 20 days).
    • Includes samples with Benzoylecgonine spiked at cutoff, 25%, 50%, 75%, and 100% above and below the cutoff.
  • Spike Recovery: 21 replicates for each spiked concentration (112.5 ng/mL and 187.5 ng/mL for 150 ng/mL cutoff; 225 ng/mL and 375 ng/mL for 300 ng/mL cutoff).
  • Analytical Recovery and Linearity: 9 intermediate levels (plus 0 and 1025 ng/mL), each run in replicates of five.
  • Method Comparison and Accuracy:
    • 100 patient samples were analyzed.
  • Specificity (Cocaine metabolites): Not specified by number of samples, but by "known amounts of each compound."
  • Specificity (Structurally Unrelated Compounds): Not specified by number of samples, but by spiking into low and high control urine.
  • Interference (Physiologic Substances & pH): Not specified by number of samples, but by spiking into low and high control urine.
  • Specific Gravity Interference: Not specified by number of samples, but by spiking drug-free urine with varying specific gravities.

Data Provenance: The studies were performed at the manufacturer's site (Microgenics Corporation, Thermo Fisher Scientific). The data appears to be retrospective (spiked samples, patient samples compared to LC-MS/MS). The country of origin of the data is not explicitly stated, but the manufacturer is based in Fremont, California, USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This being an in vitro diagnostic (IVD) device for drug toxicology, the ground truth is established by a reference analytical method rather than human expert interpretation of images or clinical findings.

  • No human experts were used to establish ground truth for the test set.
  • The primary reference method for accuracy studies was LC-MS/MS (Liquid Chromatography/tandem Mass Spectrometry), which is considered the preferred confirmatory method for drug testing due to its high specificity and sensitivity.

4. Adjudication Method

As ground truth was established by an objective analytical method (LC-MS/MS) and not human reader consensus, no adjudication method (e.g., 2+1, 3+1) was applicable or used.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

No MRMC study was conducted or is applicable for this type of IVD device. This device is an immunoassay for determining drug metabolites, not an AI-powered diagnostic imaging tool that assists human readers. Therefore, the concept of human reader improvement with AI assistance is not relevant here.

6. Standalone Performance Study

Yes, a standalone performance study was done. The entire document describes the performance of the algorithm/device only (DRI Cocaine Metabolite Assay) without human-in-the-loop performance, against established analytical standards and a reference method (LC-MS/MS). The device generates a preliminary analytical test result independently.

7. Type of Ground Truth Used

The ground truth used for performance evaluation, particularly for accuracy and method comparison, was Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS). This is a highly accurate chemical method considered the gold standard for confirming drug presence and concentration. For precision, spike recovery, and interference studies, the ground truth involved precisely spiking known concentrations of the analyte into drug-free urine.

8. Sample Size for the Training Set

The document describes the analytical performance of a finished device. It does not mention any "training set" in the context of machine learning. This device is a biochemical immunoassay, not a machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

As mentioned above, there is no disclosed "training set" or machine learning component in the context of this immunoassay. Therefore, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

July 6, 2018

Microgenics Corporation Emily Chien Regulatory Affairs Specialist II 46500 Kato Road Fremont, California 94538

Re: K181499

Trade/Device Name: DRI Cocaine Metabolite Assay Regulation Number: 21 CFR 862.3250 Regulation Name: Cocaine and cocaine metabolite test system Regulatory Class: Class II Product Code: DIO Dated: June 5. 2018 Received: June 7, 2018

Dear Emily Chien:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Paula Caposino -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
510(k) Number ( if known )K181499
Device NameDRI Cocaine Metabolite Assay
Indications for Use (Describe)The DRI Cocaine Metabolite Assay is a homogeneous enzyme immunoassay intended for the qualitative and/or semi- quantitative determination of benzoylecgonine (Cocaine Metabolite) in human urine at a cutoff concentration of either 150 ng/mL or 300 ng/mL.
The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/tandem mass spectrometry (LC- MS/MS) or permitting laboratories to establish quality control procedures.
The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography / Mass spectrometry (GC/MS) or Liquid chromatography/ tandem mass spectrometry (LC-MS/MS) is the preferred confirmatory method. Tests for cocaine metabolite cannot distinguish between abused drugs and certain prescribed medications.
Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary results are used. For In Vitro Diagnostic Use Only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

.

{3}------------------------------------------------

5. 510(k) Summary

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of Safe Medical Device Act of 1990 and 21 CFR 807.92.

A. Device Information

CategoryComments
Sponsor:Microgenics CorporationThermo Fisher Scientific46500 Kato RoadFremont, CA 94538Phone: 510-979-5000FAX: 510-979-5002
Correspondent ContactInformation:Emily ChienRegulatory Affairs Specialist IIEmail: Emily.Chien@thermofisher.comPhone: 510-979-5000FAX: 510-979-5002
Device Common Name:Cocaine Metabolite Enzyme Immunoassay
Trade or Proprietary NameDRI Cocaine Metabolite Assay
Candidate Device ProductCode, Classification,Classification Name &PanelDIO, Class II, 21 CFR 862. 3250 - Cocaine MetaboliteTest System, 91 - Toxicology

Predicate Device Information:

Predicate Device:Cocaine Metabolite Enzyme Immunoassay
Predicate DeviceManufacturer:Diagnostic Reagents, Inc.
Predicate Device PremarketNotification #:K960187

B. Date Summary Prepared July 06, 2018

July 06, 2018

C. Description of Device

The DRI Cocaine Metabolite Assay is a homogeneous enzyme immunoassay using ready-touse liquid reagents. The assay uses a specific antibody, which can detect benzoylecgonine in urine. The assay is based on the competition of an enzyme glucose-6-phosphate dehydrogenase (G6PDH) labeled drug and the drug from the urine sample for a fixed amount of specific antibody binding sites. In the presence of free drug from the sample, the free drug occupies the antibody binding sites, allowing the drug-labeled G6PDH to interact with the substrate, resulting in enzyme activity. In the absence of drug from the sample, the specific antibody

{4}------------------------------------------------

binds to the drug labeled with G6PDH and the enzyme activity is inhibited. This phenomenon creates a direct relationship between the drug concentration in the urine and the enzyme activity. The enzyme G6PDH activity is determined spectrophotometrically at 340 mm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.

The assay consists of reagents (A and E).

Reagent A: Contains mouse monoclonal anti-benzoylecgonine antibody, glucose-6-phosphate (G6P), and nicotinamide adenine dinucleotide (NAD) in Tris buffer with sodium azide as preservative.

Reagent E: Contains benzoylecgonine analog labeled with glucose-6-phosphate dehydrogenase (G6PDH) in HEPES buffer with sodium azide as preservative.

D. Intended Use

The DRI Cocaine Metabolite Assay is a homogeneous enzyme immunoassay intended for the qualitative and/or semi-quantitative determination of benzoylecgonine (Cocaine Metabolite) in human urine at a cutoff concentration of either 150 ng/mL or 300 ng/mL.

The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/tandem mass spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography / Mass spectrometry (GC/MS) or Liquid chromatography/tandem mass spectrometry (LC-MS/MS) is the preferred confirmatory method. Tests for cocaine metabolite cannot distinguish between abused drugs and certain prescribed medications.

Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary results are used. For In Vitro Diagnostic Use Only.

{5}------------------------------------------------

E. Comparison to Predicate Device

CharacteristicsCandidate Device:DRI CocaineMetabolite AssayPredicate Device:Cocaine Metabolite EnzymeImmunoassay (K960187)
Intended UseThe DRI CocaineMetabolite Assay is ahomogeneous enzymeimmunoassay intended forthe qualitative and/or semi-quantitative determinationof benzoylecgonine(Cocaine Metabolite) inhuman urine at a cutoffconcentration of either 150ng/mL or 300 ng/mL.This homogeneous cocainemetabolite enzymeimmunoassay is intended tobe used for qualitative andsemi-quantitativedetermination ofbenzoylecgonine (cocainemetabolite) in human urinewith either 300 ng/ml or 150ng/ml as a cutoff calibrator.
OperatingPrinciple(Technology)DRISame
MeasuredAnalyteBenzoylecgonineSame
Test MatrixUrineSame
Cut-off Levels150 ng/mL and 300 ng/mLSame
MethodologyHomogeneous EnzymeImmunoassaySame
Reagents FormLiquid ready-to-use.Same
AntibodyMouse monoclonalantibodiesSame
Storage2-8 °C until expiration dateSame
PrincipalOperatorTrained professionalsSame
Calibrator Levelsfor Semi-Quant5 point calibrator3 point calibrator

F. Test Principle

The DRI Cocaine Metabolite Assay is a homogeneous enzyme immunoassay using ready-touse liquid reagents. The assay uses a specific antibody, which can detect benzoylecgonine in urine. The DRI technology is based on the competition of an enzyme glucose-6-phosphate dehydrogenase (G6PDH) labeled drug and the drug from the urine sample for a fixed amount of specific antibody binding sites. In the presence of free drug from the sample, the free drug occupies the antibody binding sites, allowing the drug-labeled G6PDH to interact with the substrate, resulting in enzyme activity. In the absence of drug from the sample, the specific antibody binds to the drug labeled with G6PDH and the enzyme activity is inhibited. This phenomenon creates a direct relationship between the drug concentration in the urine and the enzyme activity. The enzyme G6PDH activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.

{6}------------------------------------------------

G. Summary of Supporting Data

1. Analytical Performance:

Performance is evaluated at the manufacturer's site on the Beckman Coulter AU680 clinical analyzer.

a) Precision

Precision studies were performed in accordance with CLSI Guideline EP05-A3. Samples were prepared by spiking Benzoylecgonine (Cocaine Metabolite) into drug free urine at the cutoff, 25%, 50%, 75% and 100% above and below the cutoff and tested in both qualitative and semi-quantitative modes. Results presented below were generated by testing all samples in replicates of 2, twice per day for 20 days, total n=80. The results for both cutoffs are summarized in the tables below.

SpikedConcentration(ng/mL)% of Cutoff(150 ng/mL)Total Precision (n=80)
# ofDeterminantsImmunoassayResults(Negative/Positive)
0-100%8080/0
37.5-75%8080/0
75-50%8080/0
112.5-25%8080/0
150100%8022/58
187.5+25%800/80
225+50%800/80
262.5+75%800/80
300+100%800/80

Qualitative Study Analysis for 150 ng/mL cutoff

Qualitative Study Analysis for 300 ng/mL cutoff

{7}------------------------------------------------

Spiked% of Cutoff(300 ng/mL)Total Precision (n=80)
Concentration(ng/mL)# ofDeterminantsImmunoassayResults(Negative/Positive)
0-100%8080/0
75-75%8080/0
150-50%8080/0
225-25%8080/0
300100%8031/49
375+25%800/80
450+50%800/80
525+75%800/80
600+100%800/80

Semi-Quantitative Study Analysis for 150 ng/mL cutoff

Spiked% of Cutoff(150 ng/mL)Total Precision (n=80)
Concentration(ng/mL)# ofDeterminantsImmunoassayResults(Negative/Positive)
0-100%8080/0
37.5-75%8080/0
75-50%8080/0
112.5-25%8080/0
150100%8019/61
187.5+25%800/80
225+50%800/80
262.5+75%800/80
300+100%800/80

Semi-Quantitative Study Analysis for 300 ng/mL cutoff

SpikedConcentration(ng/mL)% of Cutoff(150 ng/mL)Total Precision (n=80)
# ofDeterminantsImmunoassayResults(Negative/Positive)
0-100%8080/0
75-75%8080/0
150-50%8080/0
225-25%8080/0
300100%8022/58
375+25%800/80

{8}------------------------------------------------

SpikedConcentration(ng/mL)% of Cutoff(150 ng/mL)Total Precision (n=80)
# ofDeterminantsImmunoassayResults(Negative/Positive)
450+50%800/80
525+75%800/80
600+100%800/80

b) Spike Recovery

The study was performed for 21 replicates. This study was carried out by testing spiked samples containing Benzoylecgonine (Cocaine Metabolite) at the cutoff calibrator and control levels. The spiked samples were prepared by spiking Benzoylecgonine (Cocaine Metabolite) into drug free urine. Samples were tested in both qualitative and semi-quantitative mode. The qualitative results for both cutoffs are summarized in the tables below.

Replicates112.5 ng/mL(n=21)187.5 ng/mL(n=21)
1NegativePositive
2NegativePositive
3NegativePositive
4NegativePositive
5NegativePositive
6NegativePositive
7NegativePositive
8NegativePositive
9NegativePositive
10NegativePositive
11NegativePositive
12NegativePositive
13NegativePositive
14NegativePositive
15NegativePositive
16NegativePositive
17NegativePositive
18NegativePositive
19NegativePositive
20NegativePositive
21NegativePositive
OverlapNoNo
Relative to C/OAll 21 below C/OAll 21 above C/O

Qualitative Data for 150 ng/mL cutoff

{9}------------------------------------------------

Replicates225 ng/mL(n=21)375 ng/mL(n=21)
1NegativePositive
2NegativePositive
3NegativePositive
4NegativePositive
5NegativePositive
6NegativePositive
7NegativePositive
8NegativePositive
9NegativePositive
10NegativePositive
11NegativePositive
12NegativePositive
13NegativePositive
14NegativePositive
15NegativePositive
16NegativePositive
17NegativePositive
18NegativePositive
19NegativePositive
20NegativePositive
21NegativePositive
OverlapNoNo
Relative to C/OAll 21 below C/OAll 21 above C/O

Qualitative Data for 300 ng/mL cutoff

c) Analytical Recovery and Linearity

Linearity studies were performed in accordance with CLSI Guideline EP06-A. To demonstrate the dilution linearity for purposes of sample dilution and quality control of the entire assay range, drug free urine was spiked to the high level calibrator using Benzoylecgonine (Cocaine Metabolite) (1000 ng/mL) and diluted with drug free urine to generate 9 intermediate levels.

Each sample was run in replicates of five in semi-quantitative mode and the average was used to determine percent recovery compared to the expected target value. The percent recovery is summarized in the table below.

LevelExpectedConcentration(ng/mL)ObservedConcentration(ng/mL)Recovery (%)
100.6N/A

{10}------------------------------------------------

LevelExpectedConcentration(ng/mL)ObservedConcentration(ng/mL)Recovery (%)
2102.5103.6101.1
3205.0213.6104.2
4307.5294.895.9
5410.0413.8100.9
6512.5510.499.6
7615.0640.6104.2
8717.5781.4108.9
9820.0880.0107.3
10922.5952.0103.2
111025.01025.0100.0

d) Method Comparison and Accuracy

The method comparison study was performed in accordance with CLSI Guideline EP09-A3. One hundred patient samples were analyzed by the DRI Cocaine Metabolite Assay in both qualitative and semi-quantitative modes and the results were compared to LC-MS/MS. The overall concordance between LC-MS/MS and DRI Cocaine Metabolite Assay is 100%. The qualitative and semi-quantitative results for both cutoffs are summarized in the tables below.

Semi-Quantitative Mode Accuracy study with LC-MS/MS as reference method for 150 ng/mL cutoff

CandidateDeviceResultsNegativeby LC-MS/MS< 50% ofCutoffconcentrationby LC-MS/MS(< 75 ng/mL)Near CutoffNegative(Between 50%below the cutoffand the cutoffconcentration asdetermined byLC-MS/MS)(75-149 ng/mL)Near CutoffPositive(Between thecutoff and 50%above the cutoffconcentration asdetermined byLC-MS/MS)(150-225 ng/mL)High Positives(Greater than50% abovecutoffconcentration(> 225 ng/mL)
Positive000644
Negative450500

Agreement among Positives: 50/50 = 100% Agreement among Negative: 50/50 = 100%

{11}------------------------------------------------

Semi-Quantitative Mode Accuracy study with LC-MS/MS as reference method for 300 ng/mL cutoff

CandidateDeviceResultsNegativeby LC-MS/MS< 50% ofCutoffconcentrationby LC-MS/MS(< 150 ng/mL)Near CutoffNegative(Between 50%below the cutoffand the cutoffconcentration asdetermined byLC-MS/MS)(150-299ng/mL)Near CutoffPositive(Between thecutoff and 50%above the cutoffconcentration asdetermined byLC-MS/MS)(300-450 ng/mL)High Positives(Greater than50% abovecutoffconcentration(> 450 ng/mL)
Positive000644
Negative450500

Agreement among Positives: 50/50 = 100% Agreement among Negative: 50/50 = 100%

Qualitative Accuracy study with LC-MS/MS as reference method for 150 ng/mL cutoff

CandidateDeviceResultsNegativeby LC-MS/MS< 50% ofCutoffconcentrationby LC-MS/MS(< 75 ng/mL)Near CutoffNegative(Between 50%below the cutoffand the cutoffconcentration asdetermined byLC-MS/MS)(75 - 149ng/mL)Near CutoffPositive(Between thecutoff and 50%above the cutoffconcentration asdetermined byLC-MS/MS)(150-225 ng/mL)High Positives(Greater than50% abovecutoffconcentration(> 225 ng/mL)
Positive000644
Negative450500

Agreement among Positives: 50/50 = 100% Agreement among Negative: 50/50 = 100%

Qualitative Accuracy study with LC-MS/MS as reference method for 300 ng/mL cutoff

{12}------------------------------------------------

CandidateDeviceResultsNegativeby LC-MS/MS< 50% ofCutoffconcentrationby LC-MS/MS(< 150 ng/mL)Near CutoffNegative(Between 50%below thecutoff and thecutoffconcentrationas determinedby LC-MS/MS)(150-299ng/mL)Near CutoffPositive(Between thecutoff and 50%above the cutoffconcentration asdetermined byLC-MS/MS)(300-450ng/mL)High Positives(Greater than50% abovecutoffconcentration(> 450 ng/mL)
Positive000644
Negative450500

Agreement among Positives: 50/50 = 100%

Agreement among Negative: 50/50 = 100%

e) Specificity

The cross-reactivity of Cocaine and its metabolites were evaluated by adding known amounts of each compound to drug-free negative urine. The results are summarized in the tables below.

Cocaine and metabolitesTested Concentration(ng/mL)Cross-reactivity(%)
Benzoylecgonine150100
Cocaine25,0000.6
Cocaethylene30,0000.5
Ecgonine90,0000.17
Ecgonine Methyl Ester100,000<0.15
m-hydroxybenzoylecgonine30050
Norcocaine100,000<0.15

Cross reactivity of Cocaine and its metabolites for 150 ng/mL cut off

Cross reactivity of Cocaine and its metabolites for 300 ng/mL cut-off

Cocaine and metabolitesTested Concentration(ng/mL)Cross-reactivity(%)
Benzoylecgonine300100
Cocaine50,0000.6
Cocaethylene60,0000.5
Ecgonine160,0000.19
Ecgonine Methyl Ester100,000<0.3
m-hydroxybenzoylecgonine60050
Norcocaine100,000<0.3

Structurally unrelated compounds were evaluated by adding each substance to Benzoylecgonine spiked at 112.5 ng/mL (-25% of the 150 ng/mL cutoff concentration), 187.5 ng/mL (+25% of the 150

{13}------------------------------------------------

ng/mL cutoff concentration), 225 ng/mL (-25% of the 300 ng/mL cutoff concentration) and 375 ng/mL (+25% of the 300 ng/mL cutoff concentration), at the concentrations indicated. As shown in the table below, the controls were detected accurately, low control as negative and the high control as positive, indicating that all the compounds evaluated exhibited no significant cross-reactivity at the concentrations tested.

TestedConcentrationSpiked Benzoylecgonine Level
Structurally UnrelatedLow ControlHigh Control
Compounds(ng/mL)Positive/NegativePositive/Negative
11-nor-Δ9 -THC-COOH100,000NegativePositive
1R,2S(-)-Ephedrine100,000NegativePositive
1S,2R(+)-Ephedrine100,000NegativePositive
Acetaminophen1,000,000NegativePositive
Acetylsalicylic acid1,000,000NegativePositive
Acyclovir75,000NegativePositive
Albuterol1,000,000NegativePositive
Amikacin1,000,000NegativePositive
Amitriptyline100,000NegativePositive
Amobarbital100,000NegativePositive
Amoxicillin1,000,000NegativePositive
Amphetamine1,000,000NegativePositive
Azithromycin75,000NegativePositive
Benzocaine1,000,000NegativePositive
Buprenorphine100,000NegativePositive
Bupropion100,000NegativePositive
Caffeine100,000NegativePositive
Calcium Carbonate5,000,000NegativePositive
Carbamazepine100,000NegativePositive
Carisoprodol100,000NegativePositive
Chlorpromazine500,000NegativePositive
Chlorzoxazone1,000,000NegativePositive
cis-Tramadol1,000,000NegativePositive
Clomipramine100,000NegativePositive
Clonidine100,000NegativePositive
Codeine1,000,000NegativePositive
Cotinine100,000NegativePositive
Dapsone100,000NegativePositive
Desipramine100,000NegativePositive
Dextromethorphan100,000NegativePositive
Dihydrocodeine100,000NegativePositive
Diphenhydramine1,000,000NegativePositive
TestedConcentration(ng/mL)Spiked Benzoylecgonine Level
Structurally UnrelatedCompoundsLow ControlPositive/NegativeHigh ControlPositive/Negative
Doxepine500,000NegativePositive
Doxycycline Hyclate100,000NegativePositive
EDDP100,000NegativePositive
Ethyl β-D-glucuronide100,000NegativePositive
Fentanyl100,000NegativePositive
Fluconazole100,000NegativePositive
Fluoxetine50,000NegativePositive
Gabapentin100,000NegativePositive
Gentamicin100,000NegativePositive
Haloperidol100,000NegativePositive
Heroin100,000NegativePositive
Hydrocodone100,000NegativePositive
Hydromorphone100,000NegativePositive
Hydroxyzine100,000NegativePositive
Hyoscyamine HCl75,000NegativePositive
Ibuprofen5,000,000NegativePositive
Imipramine100,000NegativePositive
Indomethacin75,000NegativePositive
Lamotrigine1,000,000NegativePositive
Levofloxacin75,000NegativePositive
Lidocaine1,000,000NegativePositive
Lithium heparin5,000,000NegativePositive
Loratadine500,000NegativePositive
LSD100,000NegativePositive
Maprotiline100,000NegativePositive
Meperidine1,000,000NegativePositive
Mesoridazine1,000,000NegativePositive
Methadone1,000,000NegativePositive
Methamphetamine100,000NegativePositive
Methylphenidate100,000NegativePositive
Metoclopramide1,000,000NegativePositive
Metronidazole100,000NegativePositive
Morphine200,000NegativePositive
Morphine-3β-D-glucuronide100,000NegativePositive
Morphine-6β-D-glucuronide100,000NegativePositive
Nalbuphine1,000,000NegativePositive
Nalorphine100,000NegativePositive
Naloxone100,000NegativePositive
Naltrexone1,000,000NegativePositive
Spiked Benzoylecgonine Level
Structurally UnrelatedCompoundsTestedConcentration(ng/mL)Low ControlPositive/NegativeHigh ControlPositive/Negative
Naproxen5,000,000NegativePositive
Nitrazepam100,000NegativePositive
Norbuprenorphine100,000NegativePositive
Norcodeine100,000NegativePositive
Nordiazepam100,000NegativePositive
Norfluoxetine HCl1,000,000NegativePositive
Norketamine100,000NegativePositive
Norproxyphene100,000NegativePositive
Nortriptyline100,000NegativePositive
Ofloxacin100,000NegativePositive
Omeprazole75,000NegativePositive
Oxazepam1,000,000NegativePositive
Oxycodone100,000NegativePositive
Oxymorphone100,000NegativePositive
Paroxetine100,000NegativePositive
PCP1,000,000NegativePositive
Phenelzine75,000NegativePositive
Phenobarbital1,000,000NegativePositive
Promethazine100,000NegativePositive
Propoxyphene750,000NegativePositive
Ranitidine100,000NegativePositive
Risperidone100,000NegativePositive
Scopolamine1,000,000NegativePositive
Secobarbital1,000,000NegativePositive
Sertraline100,000NegativePositive
Spironolactone750,000NegativePositive
Stavudine100,000NegativePositive
Tapentadol100,000NegativePositive
Terbinafine750,000NegativePositive
Thiopental1,000,000NegativePositive
Thioridazine750,000NegativePositive
Tobramycin1,000,000NegativePositive
Tolmetin750,000NegativePositive
Trazodone1,000,000NegativePositive
Trimethoprim1,000,000NegativePositive
Vancomycin1,000,000NegativePositive
Venlafaxine1,000,000NegativePositive
Verapamil100,000NegativePositive
Zolpidem Tartrate100,000NegativePositive

Structurally unrelated compounds spiked at the concentration listed below into Low and High control urine for 150 ng/mL cut-off

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

Spiked Benzoylecgonine Level
Structurally UnrelatedCompoundsTestedConcentration(ng/mL)Low ControlPositive/NegativeHigh ControlPositive/Negative
11-nor-Δ 9-THC-COOH100,000NegativePositive
1R,2S(-)-Ephedrine100,000NegativePositive
1S,2R(+)-Ephedrine100.000NegativePositive
Acetaminophen1,000,000NegativePositive
Acetylsalicylic acid1,000,000NegativePositive
Acyclovir75,000NegativePositive
Albuterol1,000,000NegativePositive
Amikacin1,000,000NegativePositive
Amitriptyline100,000NegativePositive
Amobarbital100,000NegativePositive
Amoxicillin1,000,000NegativePositive
Amphetamine1,000,000NegativePositive
Azithromycin75,000NegativePositive
Benzocaine1,000,000NegativePositive
Buprenorphine100,000NegativePositive
Bupropion100,000NegativePositive
Caffeine100,000NegativePositive
Calcium Carbonate5,000,000NegativePositive
Carbamazepine100,000NegativePositive
Carisoprodol100,000NegativePositive
Chlorpromazine500,000NegativePositive
Chlorzoxazone1,000,000NegativePositive
cis-Tramadol1,000,000NegativePositive
Clomipramine100,000NegativePositive
Clonidine100,000NegativePositive
Codeine1,000,000NegativePositive
Cotinine100,000NegativePositive
Dapsone100,000NegativePositive
Desipramine100,000NegativePositive
Dextromethorphan100,000NegativePositive
Dihydrocodeine100,000NegativePositive
Diphenhydramine1,000,000NegativePositive
Doxepine500,000NegativePositive
Doxycycline Hyclate100,000NegativePositive
EDDP100,000NegativePositive
Ethyl β-D-glucuronide100,000NegativePositive
Fentanyl100,000NegativePositive
Fluconazole100,000NegativePositive
Spiked Benzoylecgonine Level
Structurally UnrelatedCompoundsTestedConcentration(ng/mL)Low ControlPositive/NegativeHigh ControlPositive/Negative
Fluoxetine50,000NegativePositive
Gabapentin100,000NegativePositive
Gentamicin100,000NegativePositive
Haloperidol100,000NegativePositive
Heroin100,000NegativePositive
Hydrocodone100,000NegativePositive
Hydromorphone100,000NegativePositive
Hydroxyzine100,000NegativePositive
Hyoscyamine HCl75,000NegativePositive
Ibuprofen5,000,000NegativePositive
Imipramine100,000NegativePositive
Indomethacin75,000NegativePositive
Lamotrigine1,000,000NegativePositive
Levofloxacin75,000NegativePositive
Lidocaine1,000,000NegativePositive
Lithium heparin5,000,000NegativePositive
Loratadine500,000NegativePositive
LSD100,000NegativePositive
Maprotiline100,000NegativePositive
Meperidine1,000,000NegativePositive
Mesoridazine1,000,000NegativePositive
Methadone1,000,000NegativePositive
Methamphetamine100,000NegativePositive
Methylphenidate100,000NegativePositive
Metoclopramide1,000,000NegativePositive
Metronidazole100,000NegativePositive
Morphine200,000NegativePositive
Morphine-3β-D-glucuronide100,000NegativePositive
Morphine-6β-D-glucuronide100,000NegativePositive
Nalbuphine1,000,000NegativePositive
Nalorphine100,000NegativePositive
Naloxone100,000NegativePositive
Naltrexone1,000,000NegativePositive
Naproxen5,000,000NegativePositive
Nitrazepam100,000NegativePositive
Norbuprenorphine100,000NegativePositive
Norcodeine100,000NegativePositive
Nordiazepam100,000NegativePositive
Norfluoxetine HCl1,000,000NegativePositive
TestedConcentration(ng/mL)Spiked Benzoylecgonine Level
Structurally UnrelatedCompoundsLow ControlPositive/NegativeHigh ControlPositive/Negative
Norketamine100,000NegativePositive
Norproxyphene100,000NegativePositive
Nortriptyline100,000NegativePositive
Ofloxacin100,000NegativePositive
Omeprazole75,000NegativePositive
Oxazepam1,000,000NegativePositive
Oxycodone100,000NegativePositive
Oxymorphone100,000NegativePositive
Paroxetine100,000NegativePositive
PCP1,000,000NegativePositive
Phenelzine75,000NegativePositive
Phenobarbital1,000,000NegativePositive
Promethazine100,000NegativePositive
Propoxyphene750,000NegativePositive
Ranitidine100,000NegativePositive
Risperidone100,000NegativePositive
Scopolamine1,000,000NegativePositive
Secobarbital1,000,000NegativePositive
Sertraline100,000NegativePositive
Spironolactone750,000NegativePositive
Stavudine100,000NegativePositive
Tapentadol100,000NegativePositive
Terbinafine750,000NegativePositive
Thiopental1,000,000NegativePositive
Thioridazine750,000NegativePositive
Tobramycin1,000,000NegativePositive
Tolmetin750,000NegativePositive
Trazodone1,000,000NegativePositive
Trimethoprim1,000,000NegativePositive
Vancomycin1,000,000NegativePositive
Venlafaxine1,000,000NegativePositive
Verapamil100,000NegativePositive
Zolpidem Tartrate100,000NegativePositive

Structurally unrelated compounds spiked at the concentration listed below into Low and High control urine for 300 ng/mL cut-off

{17}------------------------------------------------

{18}------------------------------------------------

f) Interference

The interference studies were performed in accordance with CLSI Guideline EP07-A2, using both qualitative and semi-quantitative modes. The potential interference of pH, endogenous and exogenous physiologic substances on recovery of Benzoylecgonine using DRI Cocaine Metabolite Assay was assessed by spiking known compounds of potentially interfering substances into the low control, 112.5 ng/mL (-25% of the cutoff concentration of 150 ng/mL)

{19}------------------------------------------------

and 225 ng/mL (-25% of the cutoff concentration of 300 ng/mL) and high control, 187.5 ng/mL (+25% of the cutoff concentration of 150 ng/mL) and 375 ng/mL (+25% of the cutoff concentration of 300 ng/mL). In the presence of the compounds listed below, the controls were detected accurately, indicating that these compounds did not show interference in the assay.

CompoundTested Concentration(mg/dL)Spiked Benzoylecgonine Level
Low Control-25% of cutoff(112.5 ng/mL)High Control+25% of cutoff(187.5 ng/mL)
Acetaminophen10NegativePositive
Acetone1000NegativePositive
Ascorbic Acid1000NegativePositive
Aspirin10NegativePositive
Caffeine10NegativePositive
Creatinine500NegativePositive
Ethanol1000NegativePositive
Galactose10NegativePositive
y-Globulin500NegativePositive
Glucose3000NegativePositive
Hemoglobin150NegativePositive
Human Serum Albumin500NegativePositive
Ibuprofen10NegativePositive
Oxalic Acid100NegativePositive
Riboflavin7.5NegativePositive
Sodium Chloride1000NegativePositive
Urea1250NegativePositive
pH
pH3NegativePositive
pH4NegativePositive
pH5NegativePositive
pH6NegativePositive
pH7NegativePositive
pH8NegativePositive
pH9NegativePositive
pH10NegativePositive
pH11NegativePositive

Interference substances for 150 ng/mL cut-off

Interference substances for 300 ng/mL cut-off

{20}------------------------------------------------

Tested Concentration(mg/dL)Spiked Benzoylecgonine Level
CompoundLow Control-25% of cutoff(225 ng/mL)High Control+25% of cutoff(375 ng/mL)
Acetaminophen10NegativePositive
Acetone1000NegativePositive
Ascorbic Acid1000NegativePositive
Aspirin10NegativePositive
Caffeine10NegativePositive
Creatinine500NegativePositive
Ethanol1000NegativePositive
Galactose10NegativePositive
γ-Globulin500NegativePositive
Glucose3000NegativePositive
Hemoglobin150NegativePositive
Human Serum Albumin500NegativePositive
Ibuprofen10NegativePositive
Oxalic Acid100NegativePositive
Riboflavin7.5NegativePositive
Sodium Chloride1000NegativePositive
Urea1250NegativePositive
pH3NegativePositive
pH4NegativePositive
pH5NegativePositive
pH6NegativePositive
pH7NegativePositive
pH8NegativePositive
pH9NegativePositive
pH10NegativePositive
pH11NegativePositive
Spiked Benzoylecgonine Level
Specific GravityLow ControlHigh Control
1.004NegativePositive
1.005NegativePositive
1.007NegativePositive
1.010NegativePositive
1.011NegativePositive
1.013NegativePositive
1.019NegativePositive
1.023NegativePositive
1.025NegativePositive
1.029NegativePositive

g) Specific Gravity

Drug free urine samples with specific gravity ranging in value within 1.000 to 1.030 were split and spiked with Benzoylecgonine to a final concentration of either 112.5 ng/mL or 225 ng/mL (the low control concentrations) or 187.5 ng/mL or 375 ng/mL (high control concentrations).

{21}------------------------------------------------

These samples were then evaluated in both qualitative and semi-quantitative modes. The controls were detected accurately, indicating that no interference was observed.

Specific gravity interference data for 150 ng/mL cut-off

Specific gravity interference data for 300 ng/mL cut-off

{22}------------------------------------------------

Specific GravitySpiked Benzoylecgonine Level
Low ControlHigh Control
1.004NegativePositive
1.005NegativePositive
1.007NegativePositive
1.010NegativePositive
1.011NegativePositive
1.013NegativePositive
1.019NegativePositive
1.023NegativePositive
1.025NegativePositive
1.029NegativePositive

{23}------------------------------------------------

H. Conclusion

The information supports a determination of substantial equivalence between DRI Cocaine Metabolite Assay and the predicate device Cocaine Metabolite Enzyme Immunoassay (K960187).

§ 862.3250 Cocaine and cocaine metabolite test system.

(a)
Identification. A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.(b)
Classification. Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).