Intended Use

The uCT Computed Tomography X-ray System uCT 530/550 is a computed tomography x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

Device Description

The uCT 530/uCT 550 is a multi-slice X-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system provides the filter back-projection (FBP) algorithm to reconstruct images in DICOM format. which can be used by post-processing applications.

The system consists of the Gantry, X-ray System, Data Management System, Patient Table, Console, Power Supply Cabinet, Image Processing Computer, and Software. The system software is a program used for patient management, data management, Xray scan control, image reconstruction, and image archive.

A motorized patient table moves the patient through a circular opening in the Gantry. As the patient passes through the Gantry, a source of x rays rotates around the inside of the circular opening. Detectors on the exit side of the patient record the X rays exiting the section of the patient's body being irradiated as an X-ray "snapshot". Many different "snapshots" (angles) are collected during one complete rotation. The data are sent to a computer to reconstruct all of the individual "snapshots" into a crosssectional image (slice) of the internal organs and tissues for each complete rotation of the source of x rays.

There are two features for denoising and reduce metal artifact, which are KARL iterative denoising reconstruction algorithm and MAC Metal artifact correction algorithm.

AI/ML Overview

The provided text does not contain acceptance criteria or a detailed study that proves the device meets specific acceptance criteria in the format requested.

Instead, the document is a 510(k) summary for a Computed Tomography X-ray System (uCT 530, uCT 550) seeking FDA clearance based on substantial equivalence to a predicate device (uCT 760, uCT 780).

Here's what can be extracted and inferred from the text regarding performance, though it doesn't meet all your requested points:

1. Table of Acceptance Criteria and Reported Device Performance

The document provides a comparison table of technological characteristics between the proposed device (uCT 530, uCT 550) and a predicate device (uCT 760, uCT 780). While not explicitly "acceptance criteria" against numerical targets, it implies that comparable performance to the predicate is the acceptance criterion for substantial equivalence.

ITEMAcceptance Criteria (Implicit: Similar to Predicate)Reported Device Performance (uCT 530, uCT 550)Notes/Differences from Predicate
Intended UseThe uCT Computed Tomography X-ray System is a computed tomography x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac and vascular).The uCT Computed Tomography X-ray System uCT530/uCT550 is a computed tomography x-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).Note 1: Does not include cardiac, but states this does not affect safety and effectiveness.
Product CodeJAKJAKSame
Regulation No.21 CFR 892.175021 CFR 892.1750Same
ClassIIIISame
Scan RegimeContinuous RotationContinuous RotationSame
Scan ModesScout, Axial Scan, Helical ScanScout, Axial Scan, Helical ScanSame
Detector MaterialSolid-state GOSSolid-state GOSSame
Z-plane coverage40mm22mmNote 2: Shorter coverage induces longer scanning time, but does not affect safety and effectiveness.
Size of detector element in Z-plane0.5mm0.55mmNote 3: Bigger minimum element size induces lower spatial resolution, but does not affect safety and effectiveness.
Number of element per row936864Note 4: Smaller number of elements induces less data sampling but provides sufficient data for reconstruction, not affecting safety and effectiveness.
Number of detector row8040Note 5: Determined by Z-plane coverage and element size, not affecting safety and effectiveness.
Maximum slices generated per rotation128 for uCT 760, 160 for uCT 78040 for uCT 530, 80 for uCT 550Note 6: Smaller slice number induces longer scanning time, but does not affect safety and effectiveness.
Minimum slice thickness0.625mm for uCT 760, 0.5mm for uCT 7800.55mmNote 7: Determined by detector element size, not affecting safety and effectiveness.
Maximum sampling rateUp to 4800 views per 360°Up to 4800 views per 360°Same
Tube anode storage capacity7.5MHU5.3MHUNote 8: Lower capacity affects continuous scans' throughput, but not independent scans' effect.
Maximum cooling rate1386 kHU/min815 kHU/minNote 9: Lower rate affects continuous scans' throughput, but not independent scans' effect.
Focal spot size0.7x0.7mm, 1.0x1.0mm0.5x1.0mm, 1.0x1.0mmNote 10: Smaller size helps resolution, but image spatial resolution is "equivalent substantially."
Power80kW for uCT 760, 100 kW for uCT 78050kWNote 11: Smaller power output induces lower ability of x-ray penetration for high BMI, but safety evaluated.
mA Range6-667mA for uCT 760, 6-833mA for uCT 78010-420mANote 12: Smaller mA output induces lower ability of x-ray penetration for high BMI, but safety evaluated.
kV Settings70, 80, 100, 120, 14070, 80, 100, 120, 140Same
Aperture700mm700mmSame
Rotation speedUp to 0.35 sec per 360° rotation for uCT 760, Up to 0.3 sec per 360° rotation for uCT 780Up to 0.5 sec per 360° rotationNote 13: Slower rotation speed induces longer scan time, but does not affect safety and effectiveness.
Gantry Tilt± 30° with 0.5 increment± 30° with 0.5 incrementSame
Scannable range1700 mm1700 mmSame
Horizontal motion range2180 mm2180 mmSame
Table Horizontal SpeedUp to 200mm/secUp to 200mm/secSame
Vertical motion range480 mm-950 mm from the floor480 mm-950 mm from the floorSame
Vertical speedUp to 40 mm/secUp to 40 mm/secSame
Table Horizontal Position accuracy± 0.25mm± 0.25mmSame
Table Maximum table load205kg205kgSame
Image Spatial ResolutionHigh mode: >20 lp/cm @ MTF 0%, 16.5 ± 1.7 lp/cm @ MTF10%, 11.5 ± 1.2 lp/cm @ MTF50%High mode: >20 lp/cm @ MTF 0%, 16.5 ± 1.7 lp/cm @ MTF10%, 11.5 ± 1.2 lp/cm @ MTF50%Same
Image Noise3.0 ± 0.5 HU at 120 kV, 5 mm slice thickness, CTDIvol 29.1mGy3.0 ± 0.5 HU at 120 kV, 5 mm slice thickness, CTDIvol 28.9 mGyNote 14: Noise level is "equivalent substantially" considering slightly different CTDIvol.
CT Number Display Range-1024 ~+8191 HU-1024 ~+8191 HUSame
Scan Field of ViewUp to 500 mm, 600mm with extend FOVUp to 500 mm, 600mm with extend FOVSame
Reconstruction Field of View40mm-500mm, 40mm-600mm with extend FOV40mm-500mm, 40mm-600mm with extend FOVSame
Maximum scannable length1700mm1700mmSame
Image MatrixUp to 1024 x 1024Up to 1024 x 1024Same
Reconstructed slice thicknessuCT 760: 0.625mm,1.25mm,2.5mm,5mm,10mm (axial), 0.625-10mm(helical); uCT 780: 0.5mm,0.625mm,1.25mm,2.5mm,5mm,10mm (axial), 0.5-10mm (helical)0.55mm,1.1mm,2.2mm,5.5mm,11mm (axial), 0.55-10mm(helical)Note 15: More slice thickness options for predicate, but similar capabilities overall, not affecting safety/effectiveness.
Pitch0.1~2.00.1~2.0Same
Maximum continuous exposure timeUp to 100secondsUp to 100secondsSame
Electrical SafetyComply with ES60601-1Comply with ES60601-1Same
EMCComply with IEC60601-1-2Comply with IEC60601-1-2Same
BiocompatibilityPatient Contact Materials were tested and demonstrated no cytotoxicity (ISO 10993-5), no evidence for irritation and sensitization (ISO 10993-10).Patient Contact Materials were tested and demonstrated no cytotoxicity (ISO 10993-5), no evidence for irritation and sensitization (ISO 10993-10).Same
Iterative noise reductionKARL 3DKARL 3DSame
Adaptive FilterAdaptive FilterAdaptive FilterSame
Metal artifact reductionMACMACSame

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "Clinical Evaluation for sample clinical images evaluation" and "sample clinical images for proposed devices are provided in Section 36 Clinical Evaluation. Electronic file for each image are provide in MISC Folder." However, it does not specify the sample size of the test set for image evaluation or the data provenance (e.g., country of origin, retrospective/prospective). It only refers to "sample clinical images."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not provide information on the number of experts used, their qualifications, or how ground truth was established for the "sample clinical images."

4. Adjudication Method for the Test Set

The document does not describe any adjudication method for a test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The document does not mention a MRMC comparative effectiveness study or any effect size of AI assistance for human readers. This device is a CT scanner, not an AI-assisted interpretation tool in the context of this submission. The "KARL iterative denoising reconstruction algorithm and MAC Metal artifact correction algorithm" are features of the CT system itself, not necessarily AI for reader improvement.

6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

As mentioned above, the device is a CT scanner. Its performance verification focuses on technical specifications (e.g., Image Spatial Resolution, Image Noise). The "KARL iterative denoising reconstruction algorithm" and "MAC Metal artifact correction algorithm" are integral parts of the image reconstruction process, not standalone interpretive algorithms. Therefore, a separate "standalone" performance for an interpretive algorithm is not applicable in the context described.

7. Type of Ground Truth Used

For the "sample clinical images evaluation," the type of ground truth is not specified. It refers to "sample clinical images," implying they are real patient images, but how their "truth" was confirmed (e.g., pathology, subsequent expert consensus) is not detailed. For the technical performance parameters (e.g., image noise, spatial resolution), the "ground truth" is typically established through phantom studies and objective measurements according to industry standards.

8. Sample Size for the Training Set

The document does not mention a training set sample size. This submission is for a CT scanner and its image reconstruction algorithms, not a deep learning AI model that typically requires a large training set of labeled data for classification or detection tasks. While iterative reconstruction algorithms have parameters that might be "trained" or optimized, the document does not discuss this in the context of a "training set" of patient images.

9. How the Ground Truth for the Training Set Was Established

Since no training set and its ground truth are mentioned (as per point 8), this information is not provided.

In summary: The provided document is a 510(k) summary focused on demonstrating substantial equivalence of a new CT system to a predicate device based on technical specifications and functional capabilities. It does not contain the detailed clinical study design, acceptance criteria (beyond substantial equivalence to predicate), and performance metrics typically associated with AI-driven diagnostic software, especially concerning human reader performance or defined ground truth for clinical datasets. The "Performance Verification" section lists standards compliance and "Clinical Evaluation for sample clinical images evaluation" but lacks the specifics requested.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

Shanghai United Imaging Healthcare Co., Ltd. % Shumei Wang OM & RA Director No. 2258 Chengbei Rd., Jiading Industrial District Shanghai. 201807 CHINA

August 14th, 2018

Re: K181414

Trade/Device Name: uCT 530, uCT 550 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: May 28, 2018 Received: May 30, 2018

Dear Shumei Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jeff Rodgers

for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181414

Device Name uCT 530, uCT 550

Indications for Use (Describe)

The uCT Computed Tomography X-ray System uCT 530/550 is a computed tomography x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the use of bold fonts gives it a strong and confident look.

510 (k) SUMMARY

  • Date of Preparation 1. May 28, 2018

2. Sponsor Identification

Shanghai United Imaging Healthcare Co.,Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China

Contact Person: Shumei Wang Position: QM&RA VP Tel: +86-021-67076888-6776 Fax: +86-021-67076889 Email: shumei.wang@united-imaging.com

3. Identification of Proposed Device

Trade Name: uCT 530, uCT 550 Common Name: Computed Tomography X-ray System Model(s): uCT 530, uCT 550

Regulatory Information Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-ray System Regulatory Class: II Product Code: JAK Review Panel: Radiology

Identification of Predicate Device(s) 4.

Predicate Device

510(k) Number: K172135 Device Name: uCT Computed Tomography X-Ray System Model(s): uCT 760, uCT 780

Regulatory Information

Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-ray System Regulatory Class: II Product Code: JAK Review Panel: Radiology

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the colors are muted.

5. Device Description:

The uCT 530/uCT 550 is a multi-slice X-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system provides the filter back-projection (FBP) algorithm to reconstruct images in DICOM format. which can be used by post-processing applications.

The system consists of the Gantry, X-ray System, Data Management System, Patient Table, Console, Power Supply Cabinet, Image Processing Computer, and Software. The system software is a program used for patient management, data management, Xray scan control, image reconstruction, and image archive.

A motorized patient table moves the patient through a circular opening in the Gantry. As the patient passes through the Gantry, a source of x rays rotates around the inside of the circular opening. Detectors on the exit side of the patient record the X rays exiting the section of the patient's body being irradiated as an X-ray "snapshot". Many different "snapshots" (angles) are collected during one complete rotation. The data are sent to a computer to reconstruct all of the individual "snapshots" into a crosssectional image (slice) of the internal organs and tissues for each complete rotation of the source of x rays.

There are two features for denoising and reduce metal artifact, which are KARL iterative denoising reconstruction algorithm and MAC Metal artifact correction algorithm.

Spec.ModelHV PowerRotation speedMinimumslice thicknessMaximum slicesgenerated per rotation
uCT 53050kWUp to 0.5 sec per360 °rotation0.55mm40
uCT 55050kWUp to 0.5 sec per360 °rotation0.55mm80

This proposed device includes two models: uCT 530, uCT 550.The differences between the two models are as follows:

Indications for Use 6.

The uCT Computed Tomography X-ray System uCT 530/uCT 550 is a computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

7. Comparison of Technological Characteristics with the Predicate Devices

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is stacked on top of each other and is in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the colors are muted.

The uCT 530/uCT 550 Computed Tomography X-ray system has the same indications for use as the predicate device uCT 760/uCT 780. The fundamental scientific technology of the proposed device is same as the predicate device.

Table 1 below provides a comparison of the technological characteristics of the proposed device in comparison to the predicate device.

ITEMProposed DevicePredicate DeviceRemark
uCT 530, uCT 550uCT 760, uCT 780
General
Product CodeJAKJAKSame
Regulation No.21 CFR 892.175021 CFR 892.1750Same
ClassIIIISame
Intended UseThe uCT ComputedTomography X-ray SystemuCT530/uCT550 is a computedtomography X-ray systemintended to produce cross-sectional images of the bodyby computer reconstruction ofX-ray transmission data takenat different angles and planesand indicated for the wholebody (including head, neck,vascular).The uCT ComputedTomography X-ray System isa computed tomography x-raysystem intended to producecross-sectional images of thebody by computerreconstruction of x-raytransmission data taken atdifferent angles and planesand indicated for the wholebody (including head, neck,cardiac and vascular).NoteNo.1
Specifications
Scan RegimeContinuous RotationContinuous RotationSame
Scan ModesScoutAxial ScanHelical ScanScoutAxial ScanHelical ScanSame
DetectorMaterialSolid-state GOSSolid-state GOSSame
Z-planecoverage22mm40mmNoteNo.2
Size of detectorelement in Z-plane0.55mm0.5mmNoteNo.3
Number ofelement per row864936NoteNo.4
Number ofdetector row4080NoteNo.5
Maximum slicesgenerated perrotation (multi-slice capability)40 for uCT 53080 for uCT 550128 for uCT 760160 for uCT 780NoteNo.6
Minimum slicethickness0.55mm0.625mm for uCT 7600.5mm for uCT 780Note No.7
Maximumsampling rateUp to 4800 views per 360 °Up to 4800 views per 360 °Same
Tube anodestorage capacity5.3MHU7.5MHUNote No.8
Maximumcooling rate815 kHU/min1386 kHU/minNote No.9
Focal spot size0.5x1.0mm1.0x1.0mm0.7x0.7mm1.0x1.0mmNote No.10
Power50kW80kW for uCT 760100 kW for uCT 780Note No.11
mA Range10-420mA6-667mA for uCT 7606-833mA for uCT 780Note No.12
kV Settings70, 80, 100, 120, 14070, 80, 100, 120, 140Same
Aperture700mm700mmSame
Rotation speedUp to 0.5 sec per 360 °rotationUp to 0.35 sec per 360 °rotation for uCT 760Up to 0.3 sec per 360 °rotation for uCT 780Note No.13
Gantry Tilt$\pm$ 30 °with 0.5 increment$\pm$ 30 °with 0.5 incrementSame
Scannable range1700 mm1700 mmSame
Horizontalmotion range2180 mm2180 mmSame
TableHorizontalSpeedUp to 200mm/secUp to 200mm/secSame
Vertical motionrange480 mm-950 mm from thefloor480 mm-950 mm from thefloorSame
Vertical speedUp to 40 mm/secUp to 40 mm/secSame
TableHorizontalPositionaccuracy$\pm$ 0.25mm$\pm$ 0.25mmSame
Table Maximumtable load205kg205kgSame
Image SpatialResolutionHigh mode:>20 lp/cm @ MTF 0%16.5 $\pm$ 1.7 lp/cm @ MTF10%11.5 $\pm$ 1.2 lp/cm @ MTF50%High mode:>20 lp/cm @ MTF 0%16.5 $\pm$ 1.7 lp/cm @ MTF10%11.5 $\pm$ 1.2 lp/cm @ MTF50%Same
Image Noise3.0 $\pm$ 0.5 HU at 120 kV, 5 mmslice thickness, CTDIvol 28.9mGy3.0 $\pm$ 0.5 HU at 120 kV, 5 mmslice thickness, CTDIvol29.1mGyNote No.14
CT NumberDisplay Range-1024 ~+8191 HU-1024 ~+8191 HUSame
Scan Field of ViewUp to 500 mm600mm with extend FOVUp to 500 mm600mm with extend FOVSame
ReconstructionField of View40mm-500mm40mm-600mm with extendFOV40mm-500mm40mm-600mm with extendFOVSame
Maximumscannable length1700mm1700mmSame
Image MatrixUp to 1024 x 10240.55mm,1.1mm,2.2mm,5.5mm,11mm (axial)0.55-10mm(helical)Up to 1024 x 1024Same
Reconstructedslice thicknessuCT 760:0.625mm,1.25mm,2.5mm,5mm,10mm (axial)0.625-10mm(helical)uCT 780:0.5mm,0.625mm,1.25mm,2.5mm,5mm,10mm (axial)0.5-10mm (helical)NoteNo.15
Pitch0.1~2.00.1~2.0Same
Maximumcontinuousexposure timeUp to 100secondsUp to 100secondsSame
Safety
Electrical SafetyComply with ES60601-1Comply with ES60601-1Same
EMCComply with IEC60601-1-2Comply with IEC60601-1-2Same
BiocompatibilityPatient Contact Materials weretested and demonstrated nocytotoxicity (ISO 10993-5), noevidence for irritation andsensitization (ISO 10993-10).Patient Contact Materialswere tested and demonstratedno cytotoxicity (ISO 10993-5), no evidence for irritationand sensitization (ISO 10993-10).Same
ClinicalSample clinical images for proposed devices are provided in Section 36Clinical Evaluation. Electronic file for each image are provide in MISCFolder.
Justification
Note IDJustification
Note 1The uCT 530 and uCT550 is indicated for computed tomography of thewhole body including head, neck, and vascular but not including cardiacwhich does not affect safety and effectiveness the computed tomographysystem.
Note 2Provides the shorter detector Z-plane coverage induces longer scanningtime for CT imaging, which does not affect safety and effectiveness.
Note 3Provides the bigger minimum detector element size that induces lower z-plane spatial resolution for CT imaging, which does not affect safety andeffectiveness.

Table 1 Comparison of Technological Characteristics

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the words "UNITED IMAGING" in bold, sans-serif font. To the right of the words is a stylized logo. The logo is a dark blue color and appears to be a stylized letter U.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for United Imaging. The text "UNITED IMAGING" is in bold, sans-serif font. To the right of the text is a stylized letter "U" that is split in the middle by a vertical white line. There is also a horizontal white line in the middle of the "U".

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The color of the logo is a dark teal.

Note 4Provides the smaller detector element number per row that induces less data sampling for CT imaging but provides sufficient data sampling for reconstruction which does not affect safety and effectiveness.
Note 5The number of detector rows is determined by Z-plane coverage and size of detector element in Z-plane, which does not affect safety and effectiveness.
Note 6With a reconstruction process, 40 slices or 80 slices can be achieved from the 40 rows detector. Provides the smaller slice number in one rotation induces longer scanning time for CT imaging, which does not affect safety and effectiveness.
Note 7Minimum slice thickness is determined by size of detector element in Z-plane, which does not affect safety and effectiveness.
Note 8Tube anode storage capacity is a kind of measurement about the maximum throughput of CT scanner. When an independent scan is implemented, the tube is heated and if the time interval between two independent scans is long enough, the heat can be dissipated timely but for short scan time interval, the heat which has not been dissipated timely should be stored within tube device and thus for the continuous scans (for short scan time intervals), higher storage capacity means that more continuous scan numbers can be supported. However, tube anode storage capacity has no effect on each independent scan.
Note 9Maximum cooling rate is a kind of measurement about the maximum throughput of CT scanner. For the continuous scans, higher maximum cooling rate means that the tube heat can be dissipated faster and thus shorter scan time interval and more continuous scan numbers can be supported. However, it has no effect on each independent scan.
Note 10Focus spot size has effect on image spatial resolution and smaller size is helpful for resolution improvement. However, the image spatial resolution between the two kinds of devices is equivalent substantially.
Note 11Provides the smaller power output that induces lower ability of x-ray penetration when scanning the object with high BMI with higher possibility of photon starvation, and the safety has been evaluated by the related testing and verification.
Note 12Provides the smaller mA output that induces lower ability of x-ray penetration when scanning the object with high BMI with higher possibility of photon starvation, and the safety has been evaluated by the related testing and verification.
Note 13Provides slower rotation speed that induces longer scan time for examination which does not affect safety and effectiveness.
Note 14With 120 kV and 5mm slice thickness, the image noise for typical head is 3HU on CTDIvol 28.9 mGy. The image noise level is equivalent substantially considering the proposed device has measured its noise based on the smaller CTDIvol than the Predicate Device.
Note 15Provides the more slice thickness that induce more choice for various clinical scan situations and thinner slice thickness should bring higher
spatial resolution in z-plane, which does not affect safety and
effectiveness.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized letter "U" that is split into two parts by a horizontal line. The logo is simple and modern, and the color is a dark teal.

The proposed device and the predicate device are the same in regard to most of application features, except for cardiac.

Table 2 below provides a comparison of the application features of the proposed device in comparison to the predicate device.

ITEMProposed DeviceuCT 530, uCT 550Predicate DeviceuCT 760, uCT 780Remark
Application Features
Iterative noise reductionKARL 3DKARL 3DSame
Adaptive FilterAdaptive FilterSameSame
Metal artifact reductionMACMACSame

Table 2 Comparison of Application Features

8. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Non-Clinical Testing

Non-clinical testing including dosimetry and image performance tests were conducted for the uCT 530/uCT 550 during the product development.

UNITED IMAGING HEALTHCARE claims conformance to the following standards and guidance:

Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical Safety and Electromagnetic Compatibility (EMC) testing were conducted on the uCT 530/550 in accordance with the following standards:

  • A ES 60601-1:2005(R)2012+A1:2012+C1:2009/(R)2012+A2:2010/(R)2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
  • A IEC 60601-2-44 Edition 3.0 2009, Medical Electrical Equipment - Part 2-44: Particular Requirements For The Basic Safety And Essential Performance Of Xray Equipment For Computed Tomography
  • A IEC 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
  • IEC 60825-1 Edition 2.0 2007-03, Safety Of Laser Products - Part 1: Equipment Classification, And Requirements

Product Particular Standards

  • NEMA XR 25-2010, Computed Tomography Dose Check A
  • A NEMA XR 28-2013, Supplemental Requirements For User Information And

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized "U" shape, which is dark gray. The logo is simple and modern.

System Function Related To Dose In CT

  • A NEMA XR 29-2013, Standard Attributes on CT Equipment Related to Dose Optimization and Management
  • IEC 60601-1-3 Edition 2.1 2013-04, Medical Electrical Equipment Part 1-3: A General Requirements For Basic Safety And Essential Performance - Collateral Standard: Radiation Protection In Diagnostic X-ray Equipment
  • A IEC 61223-3-5 First Edition 2004-08, Evaluation And Routine Testing In Medical Imaging Departments - Part 3-5: Acceptance Tests - Imaging Performance Of Computed Tomography X-ray Equipment

Performance Verification

  • A Clinical Evaluation for sample clinical images evaluation;
  • A AEC Test Report for AEC performance study.

Software

  • A NEMA PS 3.1-3.20(2011): Digital Imaging and Communications in Medicine (DICOM)
  • A IEC 62304: Medical Device Software - software life cycle process
  • Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices

  • Content of Premarket Submissions for Management of Cybersecurity in Medical Devices

Biocompatibility

  • ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation Of Medical A Devices - Part 5: Tests For In Vitro Cytotoxicity
  • A ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization

Other Standards and Guidances

  • ISO 14971: Medical Devices Application of risk management to medical A devices
  • Code of Federal Regulations, Title 21, Part 820 Quality System Regulation A
  • Code of Federal Regulations, Title 21, Subchapter J - Radiological Health

  • Laser Products - Conformance with IEC 60825-1 and IEC 60601-2-22; Guidance for Industry and FDA Staff (Laser Notice No. 50)

  • Provision for Alternate Measure of the Computed Tomography Dose Index (CTDI) to Assure Compliance with the Dose Information Requirements of the Federal Performance Standard for Computed Tomography

Software Verification and Validation

Software documentation for a Moderate Level of Concern software per FDA' Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" is included as a part of this submission.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED" is stacked on top of the text "IMAGING" and both are in a bold, sans-serif font. To the right of the text is a stylized icon that appears to be a capital letter "U" with a vertical line through the center.

The risk analysis was completed and risk control was implemented to mitigate identified hazards. The testing results show that all the software specifications have met the acceptance criteria. Verification and validation testing of the proposed device was found acceptable to support the claim of substantial equivalence.

UNITED IMAGING HEALTHCARE conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modification, misuse or denial of use, or unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Cybersecurity information in accordance with guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" is included in this submission.

Clinical Testing

No Clinical Study is included in this submission.

Summarv

The features described in this premarket submission are supported with the results of the testing mentioned above, the uCT 530/uCT 550 was found to have a safety and effectiveness profile that is similar to the predicate device.

9. Conclusions

Based on the comparison and analysis above, the proposed device has same intended use, similar performance, equivalence safety and effeteness as the predicate device. The differences above between the proposed device and predicate device do not affect the intended use, technology characteristics, safety and effectiveness. And no issues are raised regarding to safety and effectiveness. The proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.