K Number
K181323

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-08-03

(77 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Atlas® Gold PTA Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty of the peripheral vasculature, including the iliac arteries and iliac and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of stents and stent grafts in the peripheral vasculature. This catheter is not for use in coronary arteries.

Device Description

The Atlas® Gold PTA Dilatation Catheter is a high performance balloon catheter consisting of an over-the-wire catheter with a balloon fixed at the distal tip. The proprietary non-compliant, low profile balloon is designed to provide consistent balloon diameters and lengths even at high pressures. Two radiopaque markers delineate the working length of the balloon and aid in balloon placement. The coaxial catheter includes a tapered atraumatic tip to facilitate advancement of the catheter to and through the stenosis. The proximal portion of the catheter includes a female luer lock hub connected to the inflation lumen, and a female luer-lock hub connected to the guidewire lumen. The over-the-wire catheter is compatible with .035" guidewire and is available in 80 cm and 120 cm working lengths. Packaged with every product is a profile reducing sheath that is positioned over the balloon for protection before use. A rewrapping tool is also provided on the catheter shaft to aid in re-wrap/refolding of the balloon. This product is not manufactured with any natural rubber latex.

AI/ML Overview

This document describes the premarket notification (510(k)) for the Atlas® Gold PTA Dilatation Catheter (K181323). The information provided is sufficient to answer some but not all of the requested categories, as this is a medical device clearance document rather than a detailed clinical study report for an AI/software as a medical device (SaMD). The core of this submission is to demonstrate substantial equivalence to a predicate device, with an expanded indication for use.

Here's an analysis based on the provided text:

Acceptance Criteria and Device Performance (for the expanded indication - venous system use):

The primary acceptance criterion mentioned is related to safety for the expanded indication in the venous system.

Acceptance Criterion (Primary Safety Endpoint)Reported Device Performance (Atlas® Gold PTA Dilatation Catheter)
Freedom from major adverse events (acute thrombosis, perforation, or device-related complications such as rupture, balloon getting stuck on stent, stent disruption or dislodgement with balloon insertion or removal)100% freedom from intraprocedural adverse events related to the study device. Exceeded the 95% benchmark.
No balloon perforation, vessel laceration/perforation, or balloon-related intravascular events occurred.Achieved.
Improvement in minimal luminal area (MLA) post-stenting and balloon expansion of the stent.Mean luminal area increased from 72 mm² to 218.8 mm². (This is an efficacy measure, not an acceptance criterion, but an observed outcome.)

Study Proving the Device Meets Acceptance Criteria:

  1. Sample Size and Data Provenance:

    • Sample Size: 61 patients for post-stent dilatation and 20 patients for pre-dilatation. (Total 81 patient procedures where the device was used).
    • Data Provenance: Retrospective, investigator-sponsored clinical study. Location of data origin (country) is not explicitly stated, but it's part of a US FDA submission. The study covered data between September 1, 2013, and May 30, 2017.
  2. Number of Experts and Qualifications for Ground Truth:

    • This document describes a clinical study of a physical medical device (catheter), not an AI/SaMD. Therefore, the concept of "experts establishing ground truth for a test set" in the context of image interpretation or algorithm performance is not applicable here. The "ground truth" implicitly relies on standard clinical assessment during the procedures and follow-up, observed by the investigators.
  3. Adjudication Method for the Test Set:

    • Not specified, as this is a device performance study, not an AI/SaMD study involving human reader variability in interpretation.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • Not applicable. This study is for a physical medical device, not an AI/SaMD requiring assessment of human reader improvement with AI assistance.
  5. Standalone (Algorithm Only) Performance:

    • Not applicable. This study is for a physical medical device; there is no standalone algorithm to evaluate.
  6. Type of Ground Truth Used:

    • The "ground truth" for this study is based on clinical outcomes and observations during the procedure. Specifically, it's defined by the occurrence (or absence) of major adverse events (acute thrombosis, perforation, device-related complications) and measurements of luminal area.
  7. Sample Size for the Training Set:

    • Not applicable. This is a clinical validation study for a physical device, not an AI model that requires a training set. The device itself was developed and manufactured based on established engineering and quality system processes.
  8. How Ground Truth for the Training Set Was Established:

    • Not applicable for the same reason as above. The "training" for such a device comes from engineering design, materials science, and in-vitro/in-vivo testing during the development phase, not from a "training set" of data in the AI sense.

Additional Information from the Document:

  • Identicality to Predicate: The key argument for substantial equivalence is that the subject device (Atlas® Gold PTA Dilatation Catheter) is identical to its predicate device (Atlas® Gold PTA Dilatation Catheter, K122984) in manufacturing and design.
  • Only Differences: The only differences are an expanded indications for use to include the venous system and the inclusion of the retrospective clinical study results in the labeling.
  • Clinical Literature Review: The submission also included reviews of 46 published clinical articles involving over 4000 patients, demonstrating that PTA catheters are generally safe for use in the venous system, with known and low rates of adverse events.

In summary, this submission relies on demonstrating the safety and effectiveness of using an already cleared and identical physical device for an expanded indication based on a retrospective clinical study showing favorable outcomes against defined safety benchmarks, supplemented by existing literature. It's not a study designed to evaluate an AI/software component.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are simple and professional in appearance.

August 3, 2018

C. R. Bard, Inc. Dr. Aaron Conovaloff Regulatory Affairs Specialist II 1625 West 3rd Street Tempe, Arizona 85281

Re: K181323

Trade/Device Name: Atlas® Gold PTA Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: May 17, 2018 Received: May 18, 2018

Dear Dr. Conovaloff:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory O'Connell 2018.08.03 12:06:15 -04'00' For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181323

Device Name Atlas® Gold PTA Dilatation Catheter

Indications for Use (Describe)

The Atlas® Gold PTA Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty of the peripheral vasculature, including the iliac arteries and iliac and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of stents and stent grafts in the peripheral vasculature. This catheter is not for use in coronary arteries.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Atlas® Gold PTA Dilatation Catheter

510(k) Summary 21 CFR 807.92

As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:

Submitter Information:

Bard Peripheral Vascular, Inc. 1625 West Third St. Tempe, AZ 85281

Phone: (480) 350-6012 Fax: (480) 449-2546

Contact Person: Aaron Conovaloff, Regulatory Affairs Specialist Date of Submission: May 17, 2018

Subject Device Name:

Name of Device:Atlas® Gold PTA Dilatation Catheter
Common or UsualName:Percutaneous Catheter
Classification Name:Catheter, Angioplasty, Peripheral,Transluminal/Percutaneous catheter
Product Code:LIT, DQY
Regulatory Class:II
Regulation Number:21 CFR 870.1250

Predicate Device:

Name of Device:Atlas® Gold PTA Dilatation Catheter
Common or UsualName:Percutaneous Catheter
Classification Name:Catheter, Angioplasty, Peripheral,Transluminal/Percutaneous catheter
Product Code:LIT, DQY
Regulatory Class:II
Regulation Number:21 CFR 870.1250

Device Description:

The Atlas® Gold PTA Dilatation Catheter is a high performance balloon catheter consisting of an over-the-wire catheter with a balloon fixed at the distal tip. The proprietary non-compliant, low

{4}------------------------------------------------

profile balloon is designed to provide consistent balloon diameters and lengths even at high pressures. Two radiopaque markers delineate the working length of the balloon and aid in balloon placement. The coaxial catheter includes a tapered atraumatic tip to facilitate advancement of the catheter to and through the stenosis. The proximal portion of the catheter includes a female luer lock hub connected to the inflation lumen, and a female luer-lock hub connected to the guidewire lumen. The over-the-wire catheter is compatible with .035" guidewire and is available in 80 cm and 120 cm working lengths. Packaged with every product is a profile reducing sheath that is positioned over the balloon for protection before use. A rewrapping tool is also provided on the catheter shaft to aid in re-wrap/refolding of the balloon. This product is not manufactured with any natural rubber latex.

AttributeAtlas® Gold PTA Dilatation Catheter ProductOffering
Balloon Diameter (mm)12, 14, 16, 18, 20, 22, 24, 26
Balloon Length (cm)2, 4, 6
Catheter Shaft Lengths (cm)80, 120
Introducer SheathCompatibilityRecommended Introducer (Fr)Balloon Diameter (mm)Balloon Length (cm)
7122, 4, 6
142, 4
8146
162, 4, 6
182, 4
9186
202, 4
10222, 4
242, 4
12262, 4

{5}------------------------------------------------

Indications for Use of Device:

The Atlas® Gold PTA Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty of the peripheral vasculature, including the iliac and iliac and femoral veins, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of stent grafts in the peripheral vasculature. This catheter is not for use in coronary arteries.

Technological Comparison to Predicate Devices:

The Atlas® Gold PTA Dilatation Catheter has the following similarities to the predicate device, the Atlas® Gold PTA Dilatation Catheter (clearance to market via K122984 on October 22, 2012):

  • . Same intended use
  • . Same target population
  • . Same operating principle
  • . Same fundamental scientific technology
  • . Same sterility assurance level and method of sterilization

It should be noted that the subject Atlas® Gold PTA Dilatation Catheter is identical to the predicate device with respect to manufacturing and design. The only differences between the subject and predicate devices are an expanded indications for use statement to include use in the venous system, and the inclusion of a summary of the results of a retrospective, investigator-sponsored clinical study describing experience using the Atlas® Gold PTA Dilatation Catheter in the venous system in the labeling.

Performance Data:

A retrospective, investigator-sponsored clinical study providing patient-level data using the Atlas® Gold PTA Dilatation Catheter in venous stenting procedures was conducted. The Atlas® Gold PTA Dilatation Catheter was used for post-stent dilatation in 61 patients, and for predilatation in 20 patients. All patients were successfully treated with the study device. Subjects included had undergone illiofemoral vein compression treatment between September 1, 2013 and May 30, 2017. The primary safety endpoint was intra-procedural freedom from major adverse events defined as acute thrombosis, perforation, or device-related complications (rupture, balloon getting stuck on stent, stent disruption or dislodgement with balloon insertion or removal). This analysis demonstrated that the Atlas® Gold PTA Dilatation Catheter met 100%

{6}------------------------------------------------

freedom from intraprocedural adverse events related to the study device, and exceeded the 95% benchmark for the primary safety endpoint. No balloon perforation, vessel laceration/perforation, or balloon-related intravascular events occurred. IVUS imaging showed that the minimal luminal area (MLA) was improved post stenting and balloon expansion of the stent, with the mean luminal area increasing from 72 to 218.8 mm².

Reviews of published clinical literature describing the use of PTA balloons in the venous system were conducted. These reviews of relevant clinical literature from 46 articles with over 4000 patients demonstrate that PTA catheters are safe for use in the treatment of stenosis or thrombosis in native iliac or femoral veins, and the rates of reported safety events are low and are known events associated with PTA procedures.

Conclusions:

The subject and predicate devices are identical with respect to manufacturing and design. The only differences between the subject and predicate devices are an expanded indications for use statement to include use in the venous system, and the inclusion of a summary of the results of a retrospective, investigator-sponsored clinical study describing experience using the Atlas® Gold PTA Dilatation Catheter in the venous system in the labeling; the clinical data obtained using the Atlas® Gold PTA Dilatation Catheter demonstrated that there are no new questions of safety and effectiveness raised by the expanded indication for use. Therefore the subject Atlas® Gold PTA Dilatation Catheter is substantially equivalent to the legally marketed predicate device, the Atlas® Gold PTA Dilatation Catheter.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).