K Number
K181060

Validate with FDA (Live)

Date Cleared
2018-08-24

(123 days)

Product Code
Regulation Number
882.5600
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  1. The Solitaire™ 2 and Solitaire ™ Platinum Revascularization Devices is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should be started within 6 hours of symptom onset.
  2. The Solitaire™ 2 and Solitaire TM Platinum Revascularization Devices is indicated to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for IV t-PA or who fail IV t-PA therapy are candidates for treatment.
Device Description

The Solitaire™ 2 and Solitaire™ Platinum Revascularization Devices is designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion in the neurovasculature such as the internal carotid artery (ICA), M1 and M2 segments of the middle cerebral artery, basilar and the vertebral arteries. The distal nitinol portion of the subject device facilitates clot retrieval and has Platinum/Iridium radiopaque markers on the proximal and distal ends and, for some SKUs, at several locations along the body of the stent. The devices are supplied sterile and are intended for single-use only.

AI/ML Overview

The provided text describes a 510(k) summary for the Medtronic Solitaire™ 2 and Solitaire™ Platinum Revascularization Devices, focusing on a labeling change to include an alternative aspiration source. The study described is a non-clinical bench test. Therefore, many of the typical acceptance criteria and study details for an AI/device performance study (like MRMC, expert ground truth, sample sizes for training/test sets with provenance, etc.) are not applicable in this context.

However, based on the provided text, I can extract the following information relevant to device acceptance criteria and testing:

Device: Solitaire™ 2 and Solitaire™ Platinum Revascularization Devices

Purpose of Submission: Labeling change to include an alternative aspiration source (Riptide™ Aspiration System) in addition to the 60 cc syringe.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Implicit)Reported Device Performance (Simulated Clot Retrieval Testing)Test
Clot retrieval performance with alternative aspiration source should be equivalent to the cleared device with a 60 cc syringe.Clot retrieval performance of the subject device was equivalent to the predicate performance.Simulated clot retrieval testing (Purpose: Compare the clot retrieval performance of the subject device when used with the alternative aspiration source to the cleared device use with a 60 cc syringe.)

Note on "Acceptance Criteria": The document does not explicitly state numerical acceptance criteria in the typical sense for a clinical or AI performance study. Instead, the "acceptance criterion" for this submission is implicitly demonstrated through equivalence testing in a non-clinical setting: the new configuration (device with alternative aspiration source) must not perform worse than the existing cleared configuration (device with syringe aspiration) in terms of its primary function (clot retrieval).

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: The document does not specify the exact number of test samples (e.g., clots, simulated vessels) used for the simulated clot retrieval testing. It only mentions "The sizes used in the simulated clot retrieval testing were identical to the primary predicate device and representative of the additional predicate devices."
  • Data Provenance: Non-clinical bench testing. No country of origin is specified, as it's a bench test, not clinical data. The study is prospective in the sense that the bench tests were conducted for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as the study described is a non-clinical bench test involving physical properties (clot retrieval performance, vacuum pressure), not human interpretation of medical images or data requiring expert review for ground truth.

4. Adjudication method for the test set:

This information is not applicable for the same reasons as point 3.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. The submission is for a physical medical device (revascularization device) and a change in its aspiration source, not an AI-powered diagnostic or assistive tool. Therefore, no human readers or MRMC studies were conducted as part of this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This information is not applicable as it is a physical medical device, not an algorithm.

7. The type of ground truth used:

  • For Simulated Clot Retrieval Testing: The ground truth is established by the physical outcome of the bench test (i.e., whether the device successfully retrieved the simulated clot, and its performance compared to the predicate device).
  • For Vacuum pressure testing: The ground truth is objective measurements of vacuum pressure.

8. The sample size for the training set:

This information is not applicable as this is a physical medical device and not an AI/machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established:

This information is not applicable as there is no training set for a physical medical device as described in this context.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 24, 2018

Medtronic Neurovascular Amnon Talmor Principal Regulatory Affairs Specialist 9775 Toledo Way Irvine, California 92618

Re: K181060

Trade/Device Name: Solitaire 2TM and Solitaire™ Platinum Revascularization Devices Regulation Number: 21 CFR 882.5600 Regulation Name: Neurovascular Mechanical Thrombectomy Device for Acute Ischemic Stroke Treatment Regulatory Class: Class II Product Code: POL, NRY Dated: May 31, 2018 Received: June 1, 2018

Dear Amnon Talmor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Xiaolin Zheng-S" in a simple, sans-serif font. The name is split into two lines, with "Xiaolin" on the first line and "Zheng-S" on the second. In the background, there is a faint, stylized graphic that appears to be a logo or watermark, adding a subtle visual element to the composition.

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181060

Device Name

Solitaire™ 2 and Solitaire TM Platinum Revascularization Devices

Indications for Use (Describe)

  1. The Solitaire™ 2 and Solitaire ™ Platinum Revascularization Devices is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should be started within 6 hours of symptom onset.

  2. The Solitaire™ 2 and Solitaire TM Platinum Revascularization Devices is indicated to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for IV t-PA or who fail IV t-PA therapy are candidates for treatment.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary - K181060

510(k)Owner:Medtronic Neurovascular9775 Toledo WayIrvine, CA 92618Establishment Registration No. 2029214
ContactPerson:Amnon TalmorPrincipal Specialist, Regulatory AffairsTelephone: (949) 297-9270E-mail: amnon.talmor@medtronic.com
Date SummaryPrepared:August 10, 2018
Trade Name ofDevice:Solitaire™ 2 and Solitaire™ Platinum Revascularization Devices
Common Nameof Device:Neurovascular Mechanical Thrombectomy Device forAcute Ischemic Stroke Treatment; Catheter, thrombus retriever
Classification ofDevice:21 CFR 882.5600 and 21 CFR 870.1250, Class II
Product Code:POL, NRY
PrimaryPredicateDevices:Solitaire™ 2 Revascularization DeviceK162539
AdditionalPredicateDevices:Solitaire™ Platinum Devices (K153071, K160641 and K161879)

Device Description:

The Solitaire™ 2 and Solitaire™ Platinum Revascularization Devices is designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion in the neurovasculature such as the internal carotid artery (ICA), M1 and M2 segments of the middle cerebral artery, basilar and the vertebral arteries. The distal nitinol portion of the subject device facilitates clot retrieval and has Platinum/Iridium radiopaque markers on the proximal and distal ends and, for some SKUs, at several locations along the body of the stent. The devices are supplied sterile and are intended for single-use only.

{4}------------------------------------------------

The purpose of this submission is a labeling change to include an alternative aspiration source in addition to the 60 cc syringe when using the Solitaire™ 2 and Solitaire TM Platinum Revascularization Devices.

The indications for use for the subject device are as follows:

    1. The Solitaire™ 2 and Solitaire TM Platinum Revascularization Devices is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should be started within 6 hours of symptom onset.
    1. The Solitaire™ 2 and Solitaire TM Platinum Revascularization Devices is indicated to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for IV t-PA or who fail IV t-PA therapy are candidates for treatment.

Device Comparison:

Table 1 below provides a comparison between the subject, the primary predicate and additional predicate devices. All aspects of the design-including stent sizes, marker spacing, device and packaging materials, sterilization methods and how suppliedremain the same.

{5}------------------------------------------------

Table 1: Comparison between the proposed Solitaire™ 2 and Solitaire™ Platinum Revascularization Devices used with the alternativeaspiration source to the primary predicate and additional predicate Solitaire™ 2 and Solitaire™ Platinum Revascularization Devices
Primary Predicate:Solitaire™ 2Revascularization Devicewith syringe aspiration(K162539)Additional Predicate devices(Solitaire™ PlatinumRevascularization Device)Subject: Solitaire™ 2 andSolitaire™ PlatinumRevascularizationDevices used withRiptide™ AspirationSystemRationale for Difference(if applicable)
K153071K161879K160641
Product codePOL and NRYNRYSame as primary predicate(K162539)N/A
Indication for UseThe Solitaire™ 2Revascularization Deviceis indicated for use torestore blood flow in theneurovasculature byremoving thrombus for thetreatment of acute ischemicstroke to reduce disabilityin patients with apersistent, proximalanterior circulation, largevessel occlusion, andsmaller core infarcts whohave first receivedintravenous tissueplasminogen activator (IVt-PA). Endovasculartherapy with the deviceshould be started within 6hours of symptom onset.The Solitaire™Revascularization Deviceis indicated to restoreThe Solitaire™ PlatinumRevascularization Device isindicated to restore blood flow byremoving thrombus from a largeintracranial vessel in patientsexperiencing ischemic strokewithin 8 hours of symptom onset.Patients who are ineligible for IV t-PA or who fail IV t-PA therapy arecandidates for treatment.Same as primary predicate(K162539)N/A
Table 1: Comparison between the proposed Solitaire™ 2 and Solitaire ™ Platinum Revascularization Devices used with the alternativeaspiration source to the primary predicate and additional predicate Solitaire™ 2 and Solitaire ™ Platinum Revascularization Devices
Primary Predicate:Solitaire™ 2Revascularization Devicewith syringe aspiration(K162539)Additional Predicate devices(Solitaire™ PlatinumRevascularization Device)Subject: Solitaire™ 2 andSolitaire ™ PlatinumRevascularizationDevices used withRiptide™ AspirationSystemRationale for Difference(if applicable)
blood flow by removingthrombus from a largeintracranial vessel inpatients experiencingischemic stroke within 8hours of symptom onset.Patients who are ineligiblefor IV t-PA or who fail IVt-PA therapy arecandidates for treatment.K153071K161879K160641
Principle of operationThe device is used in theneurovasculature to restoreblood flow for treatment ofacute ischemic strokeThe device is used in theneurovasculature to restoreblood flow by removingthrombusThe device is used in theneurovasculature to restore bloodflow by removing thrombusSame as primary predicate(K162539)N/A
Procedural stepsaspiration sourceSyringeAspiration DeviceBench testing and areview of relevant clinicalliterature demonstrates thealternative aspirationsource does not raise any
Table 1: Comparison between the proposed Solitaire™ 2 and Solitaire ™™ Platinum Revascularization Devices used with the alternativeaspiration source to the primary predicate and additional predicate Solitaire™ 2 and Solitaire ™™ Platinum Revascularization Devices
Primary Predicate:Solitaire™ 2Revascularization Devicewith syringe aspiration(K162539)Additional Predicate devices(Solitaire™ PlatinumRevascularization Device)Subject: Solitaire™ 2 andSolitaire ™™ PlatinumRevascularizationDevices used withRiptide™ AspirationSystemRationale for Difference(if applicable)
K153071K161879K160641
new issues of safety oreffectiveness.
Device sizes (mm)4x154x204x20x056x40x104x15The subject device scopeincludes the primarypredicate and additionalpredicate devices. Thesizes used in thesimulated clot retrievaltesting were identical tothe primary predicatedevice and representativeof the additional predicatedevices.
4x204x406x24x064x20
4x406x204x40
6x206x20
6x306x30
4x20x05
6x24x06
6x40x10
How suppliedStored within dispenser coil, Tyvek pouch and shipping cartonN/A
Sterilization methodEthylene oxideN/A
Device materials
StentNitinolN/A
Push wireNitinol
Radiopaque markers90% Platinum/10% Iridium
Push-wire shrink TubingPTFE

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Table 1: Comparison between the proposed Solitaire™ 2 and Solitaire™ Platinum Revascularization Devices used with the alternativeaspiration source to the primary predicate and additional predicate Solitaire™ 2 and Solitaire™ Platinum Revascularization Devices
Primary Predicate:Solitaire™ 2Revascularization Devicewith syringe aspiration(K162539)Additional Predicate devices(Solitaire™ PlatinumRevascularization Device)Subject: Solitaire™ 2 andSolitaire™ PlatinumRevascularizationDevices used withRiptide™ AspirationSystemRationale for Difference(if applicable)
Introducer SheathPTFE/GrilamidK153071 K161879 K160641
Shrink tubingPTFE

{9}------------------------------------------------

Performance Data

Medtronic Neurovascular performed the following non-clinical bench tests to support the use of the alternative aspiration source with the Solitaire™ 2 and Solitaire™ Platinum Revascularization Devices:

TestTest Method SummaryConclusions
Vacuum pressuretestingPurpose: Compare the ability of thealternative aspiration device to delivervacuum pressure to the distal tip of thecatheter to a 60 cc syringe.Method: The vacuum pressure wasmeasured at the distal tip of the aspirationcatheter when generated from thealternative aspiration device at variousvacuum pressure settings, as well as from a60 cc VacLok® Syringe.Vacuum pressure delivered tothe distal tip of the catheter froma 60 cc syringe exceeded thepressure delivered from thealternative aspiration device.
Simulated clotretrieval testingPurpose: Compare the clot retrievalperformance of the subject device whenused with the alternative aspiration sourceto the cleared device use with a 60 ccsyringe.Clot retrieval performance of thesubject device was equivalent tothe predicate performance.

In addition to the above tests, Medtronic Neurovascular conducted a comprehensive review of relevant clinical literature to assess whether any issues of safety or effectiveness were raised by the clinical use of the alternative aspiration source with the Solitaire™ 2 and Solitaire™ Platinum Revascularization Devices.

Conclusion

Non-clinical bench testing and a review of relevant literature demonstrate the subject device is substantially equivalent to the predicate Solitaire™ 2 and Solitaire™ Platinum Revascularization Devices and does not raise any new questions of safety and effectiveness.

§ 882.5600 Neurovascular mechanical thrombectomy device for acute ischemic stroke treatment.

(a)
Identification. A neurovascular mechanical thrombectomy device for acute ischemic stroke treatment is a prescription device used in the treatment of acute ischemic stroke to improve clinical outcomes. The device is delivered into the neurovasculature with an endovascular approach, mechanically removes thrombus from the body, and restores blood flow in the neurovasculature.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient contacting components of the device must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including:
(i) Mechanical testing to demonstrate the device can withstand anticipated tensile, torsional, and compressive forces.
(ii) Mechanical testing to evaluate the radial forces exerted by the device.
(iii) Non-clinical testing to verify the dimensions of the device.
(iv) Non-clinical testing must demonstrate the device can be delivered to the target location in the neurovasculature and retrieve simulated thrombus under simulated use conditions.
(v) Non-clinical testing must demonstrate the device is radiopaque and can be visualized.
(vi) Non-clinical testing must evaluate the coating integrity and particulates under simulated use conditions.
(vii) Animal testing must evaluate the safety of the device, including damage to the vessels or tissue under anticipated use conditions.
(3) Performance data must support the sterility and pyrogenicity of the patient contacting components of the device.
(4) Performance data must support the shelf-life of the device by demonstrating continued sterility, package integrity, and device functionality over the specified shelf-life.
(5) Clinical performance testing of the device must demonstrate the device performs as intended for use in the treatment of acute ischemic stroke and must capture any adverse events associated with the device and procedure.
(6) The labeling must include:
(i) Information on the specific patient population for which the device is intended for use in the treatment of acute ischemic stroke, including but not limited to, specifying time from symptom onset, vessels or location of the neurovasculature that can be accessed for treatment, and limitations on core infarct size.
(ii) Detailed instructions on proper device preparation and use for thrombus retrieval from the neurovasculature.
(iii) A summary of the clinical testing results, including a detailed summary of the device- and procedure-related complications and adverse events.
(iv) A shelf life.