(58 days)
The uMR 560 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities.
These images and the physical parameters derived from the interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.
The uMR 560 is a 1.5T superconducting magnetic resonance diagnostic device with a 60cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. The uMR 560 Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM standards.
This document is primarily a 510(k) summary for the uMR 560 Magnetic Resonance Diagnostic Device. It outlines its substantial equivalence to a predicate device and lists a series of non-clinical and clinical tests performed. However, it does not contain specific acceptance criteria or detailed study results directly proving the device meets acceptance criteria in the format requested.
The document states that "The test results demonstrated that the device performs as expected and thus, it is substantially equivalent to the predicate devices to which it has been compared." This is a general statement of compliance, not a detailed breakdown of acceptance criteria and performance against them.
Therefore, most of the requested information cannot be directly extracted from the provided text.
Here is a breakdown of what can and cannot be answered from the provided text:
1. A table of acceptance criteria and the reported device performance
- Cannot be provided. The document lists standards that the device was tested against (e.g., NEMA MS standards for SNR, Geometric Distortion, Image Uniformity), but it does not specify the numerical acceptance criteria for each of these tests, nor does it provide the exact reported performance values. It only states that the device "performs as expected."
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Partially available for clinical tests, but limited detail.
- For the "volunteer study" on gradient-induced nerve stimulation: The sample size is not specified. Data provenance is not specified. It's prospective by nature of being a "volunteer study."
- For "sample clinical images" to support diagnostic quality: The sample size is not specified. Data provenance is not specified. It's implied to be prospective clinical images.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Cannot be provided. The document states that images are "interpreted by a trained physician," but it does not specify the number or qualifications of experts involved in establishing ground truth for the provided clinical images or the volunteer study.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Cannot be provided. The document does not describe any adjudication methods.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Cannot be provided. The uMR 560 is a Magnetic Resonance Diagnostic Device (an MRI scanner), not an AI-powered image interpretation tool. An MRMC study comparing human readers with and without AI assistance is not relevant to this device's regulatory review based on the provided text, and no such study is mentioned. The device generates images; it does not interpret them with AI.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Cannot be provided. As noted above, this device is an MRI scanner which generates images, not an algorithm for image interpretation. Therefore, a standalone algorithm performance study is not applicable or mentioned.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Partially discernible for clinical images: For "sample clinical images," the ground truth implicitly relies on interpretation by "trained physicians" (expert interpretation), but the specific methodology for establishing definitive ground truth (e.g., comparison to pathology, follow-up, etc.) is not detailed. For the volunteer study, the ground truth relates to physiological responses to gradient-induced nerve stimulation.
8. The sample size for the training set
- Not applicable / Cannot be provided. This device is an MRI scanner, not an AI model that would typically have a "training set" in the context of machine learning. The studies mentioned (compliance with standards, volunteer study, sample clinical images) are for verification and validation of the hardware and image quality, not for training an algorithm.
9. How the ground truth for the training set was established
- Not applicable / Cannot be provided. See response to item 8.
Summary of available information from the document:
The provided document details the 510(k) clearance for the uMR 560, an MRI scanner. It lists the non-clinical and clinical tests performed to demonstrate substantial equivalence to a predicate device.
- Non-Clinical Tests: Compliance with various IEC, ISO, and NEMA MS standards related to electrical safety, EMC, basic safety and performance for MR equipment, biological evaluation, and specific MR image quality parameters (SNR, geometric distortion, uniformity, acoustic noise, slice thickness, SAR, phased array coil characterization).
- Clinical Tests:
- A volunteer study to determine safety limits for gradient-induced nerve stimulation.
- Sample clinical images were provided to support the device's ability to generate diagnostic quality images.
The document explicitly states: "The test results demonstrated that the device performs as expected and thus, it is substantially equivalent to the predicate devices to which it has been compared." However, it does not provide the specific numerical acceptance criteria or the measured performance values.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shanghai United Imaging Healthcare Co.,Ltd. Shumei Wang Qm&ra VP No.2258 Chengbei Road., Jiading Industrial District Shanghai, 201807 CHINA
Re: K180926
Trade/Device Name: uMR 560 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: April 6, 2018 Received: April 9, 2018
Dear Shumei Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
June 6th, 2018
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jeff Bargar
for
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K180926
Device Name uMR 560
Indications for Use (Describe)
The uMR 560 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities.
These images and the physical parameters derived from the interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized symbol that resembles a shield or a stylized letter "U" with a vertical line running through the center. The color scheme is a dark, muted blue-gray against a white background.
SECTION 7
510(k) Summary
SECTION 7 - 1 of 4
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized "U" shape, which is also in a bold, sans-serif font. The logo is simple and modern, and the colors are muted.
510 (k) SUMMARY
1. Date Prepared [21 CFR 807.92(a)(1)] March 16, 2018
2. General Information [21 CFR 807.92(a)(1)]
| Manufacturer: | Shanghai United Imaging Healthcare Co., Ltd2258 Chengbei Rd., Jiading District, Shanghai, 201807 |
|---|---|
| Contact Person: | Shumei Wang |
QM&RA VP Tel: +86 (21) 67076888-6776 Fax: +86-021-67076889 Email: shumei.wang@united-imaging.com
3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Trade Name: | uMR 560 |
|---|---|
| Common Name: | Magnetic Resonance Diagnostic Device |
| Model: | uMR 560 |
| Product Code: | LNH |
| Regulation Number: | 892.1000 |
| Device Class: | II |
4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[
The identification of predicates device within this submission is as follow:
Predicate Device Manufacturer: SIEMENS AG MAGNETOM AREA Device Name: Product Code: LNH Device Class: II Regulation Number: 21 CFR 892.1000 FDA 510 (k) #: K132951
5. Device Description [21 CFR 807.92(a)(4)]
The uMR 560 is a 1.5T superconducting magnetic resonance diagnostic device with a
SECTION 7 - 2 of 4
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, also in a bold font. The logo is simple and modern in design.
60cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. The uMR 560 Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM standards.
6. Intended Use [21 CFR 807.92(a)(5)]
The uMR 560 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities.
These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.
7. Technological Characteristic [21 CFR 807.92(a)(6)]
The uMR 560 has the same major technological characteristics as the predicate device. The differences from the predicate device including patient bore size, high order shimming configuration, interventional usage, RFPA power and external triggering function are discussed in the comparison table in this submission. The same scientific technology theory, similar physical design, same coil applications and functionalities are applied to both the uMR 560 and predicate device. The technical characteristic differences do not affect the safety and effectiveness of the uMR 560 for intended use.
Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)] 8.
Summary of Non-Clinical Tests:
The following testing was conducted on the uMR 560 Magnetic Resonance Diagnostic Device as the predicate device:
- ES60601-1:2005/(R)2012. Medical Electrical Equipment Part 1: General A Requirements For Basic Safety And Essential Performance
- A IEC 60601-1-2 Edition 3.0 2007-03, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
- A 60601-2-33 Ed. 3.1:2013, Medical Electrical Equipment - Part 2-33: Particular Requirements For The Basic Safety And Essential Performance Of Magnetic Resonance Equipment For Medical Diagnostic
- A IEC 60825-1 Edition 2.0 2007-03, Safety Of Laser Products - Part 1: Equipment Classification, And Requirements [Including: Technical Corrigendum 1 (2008), Interpretation Sheet 1 (2007), Interpretation Sheet 2 (2007)]
SECTION 7 - 3 of 4
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is also in a bold font. The color of the logo is a dark teal.
- ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation Of Medical A Devices - Part 5: Tests For In Vitro Cytotoxicity
- A ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- A MS 1-2008(R2014), Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Images
- A MS 2-2008(R2014), Determination of Two-Dimensional Geometric Distortion in Diagnostic Magnetic Resonance Images
- A MS 3-2008(R2014), Determination of Image Uniformity in Diagnostic Magnetic Resonance Images
- MS 4-2010, Acoustic Noise Measurement Procedure for Diagnosing Magnetic A Resonance Imaging Devices
- A MS 5-2010, Determination of Slice Thickness in Diagnostic Magnetic Resonance Imaging
- A MS 6-2008(R2014), Determination of Signal-to-Noise Ratio and Image Uniformity for Single-Channel Non-Volume Coils in Diagnostic MR Imaging
- A MS 8-2008, Characterization Of The Specific Absorption Rate For Magnetic Resonance Imaging Systems
- A MS 9-2008(R2014), Standards Publication Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images
The test results demonstrated that the device performs as expected and thus, it is substantially equivalent to the predicate devices to which it has been compared.
Summary of Clinical Tests:
- A A volunteer study was used to determine the safety limits associated with gradient-induced nerve stimulation.
- A Sample clinical images were provided to support the ability of uMR 560 to generate diagnostic quality images in accordance with the MR guidance on premarket notification submissions.
9. Conclusion [21 CFR 807.92(b)(3)]
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, we concludes that uMR 560 Magnetic Resonance Diagnostic Device is substantially equivalent to the predicate device. It does not introduce new indications for use, and has the same technological characteristics and does not introduce new potential hazards or safety risks.
SECTION 7 - 4 of 4
§ 892.1000 Magnetic resonance diagnostic device.
(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.