K Number
K180855

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-06-01

(60 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PREVENA PLUS Incision Management System is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

The PREVENA PLUS DUO Incision Management System is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

Device Description

The PREVENA PLUS 125 Unit is a component of a the PREVENA PLUS Incision Management System and the PREVENA PLUS DUO Incision Management System. The systems provide surgical incision management via the application of negative pressure wound therapy over an incision site that has been closed with sutures or staples. The systems are applied to the incision site immediately after surgery for a maximum of fourteen (14) days depending on the surgeon's preference.

The pump in the therapy unit delivers continuous negative pressure at -125 mmHg through tubing to a dressing placed over the incision site. The occlusive drape over the dressing provides a negative pressure environment and protects the incision from external contamination.

The application of negative pressure draws the incision edges together, and removes fluid from the incision site into a canister fitted into the therapy unit.

AI/ML Overview

This document does not contain an acceptance criteria table or a detailed study proving the device meets specific acceptance criteria in the manner typically seen for algorithmic performance (e.g., sensitivity, specificity, AUC). Instead, it's a 510(k) submission for a medical device (a Negative Pressure Wound Therapy system) that focuses on demonstrating substantial equivalence to a legally marketed predicate device through non-clinical bench testing.

Therefore, many of the requested elements (sample sizes for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth types) are not applicable or estimable from the provided text for an AI/algorithm-based device.

Here's an attempt to answer the questions based only on the provided document:

1. A table of acceptance criteria and the reported device performance

Based on the document, the primary "acceptance criterion" for this 510(k) submission is the demonstration of substantial equivalence to a predicate device, specifically regarding the sustained delivery of negative pressure.

Acceptance CriteriaReported Device Performance
Delivery of continuous negative pressure at -125 mmHg within specifications under worst-case conditions of air leak rate and fluid input during a 14-day test.Bench testing demonstrated that the PREVENA PLUS 125 Negative Pressure Wound Therapy Unit is substantially equivalent to the predicate therapy unit in the delivery of continuous negative pressure at -125 mmHg within specifications under worst case conditions of air leak rate and fluid input during a 14 day test.
Safety, performance, and indications for use being equivalent to the predicate device systems.The PREVENA PLUS Incision Management System and the PREVENA PLUS DUO Incision Management System are substantially equivalent to the predicate device systems in terms of safety, performance and indications for use.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided. The document states that "Bench testing demonstrated..." but does not detail the sample size (number of units tested, number of test runs) or data provenance. Since it's bench testing, it's not human data and thus retrospective/prospective or country of origin isn't directly applicable in the same way as for clinical studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable/provided. The ground truth for bench testing typically involves engineering specifications and measurements against those specifications, not expert human assessment.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable/provided. Adjudication methods are relevant for human expert review processes, which are not described for this bench testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable/provided. No MRMC study was conducted, as this is a physical medical device (negative pressure wound therapy unit), not an AI algorithm assisting human readers. The document explicitly states: "No clinical tests were necessary." and "No usability testing was required, as there has been no change to the user interface or to the Instructions for Use."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable/provided. This is a physical device, not an algorithm. The performance described is the standalone performance of the device itself (its ability to maintain negative pressure).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for the bench testing would be the engineering specifications and established physical principles for continuous negative pressure delivery at -125 mmHg, as determined by internal R&D and quality standards. Specific details are not provided, but it's fundamentally based on measured physical parameters against design requirements.

8. The sample size for the training set

This information is not applicable/provided. This is not an AI/machine learning device that uses a "training set."

9. How the ground truth for the training set was established

This information is not applicable/provided. This is not an AI/machine learning device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

June 1, 2018

KCI USA, Inc. Kimberly McCoy Sr. Manager, RA 6203 Farinon Drive San Antonio, Texas 78249

Re: K180855

Trade/Device Name: Prevena Plus Incision Management System, Prevena Plus Duo Incision Management System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: March 30, 2018 Received: April 2, 2018

Dear Kimberly McCoy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K180855

Device Name

PREVENA PLUS Incision Management System

Indications for Use (Describe)

The PREVENA PLUS Incision Management System is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K180855

Device Name

PREVENA PLUS DUO Incision Management System

Indications for Use (Describe)

The PREVENA PLUS DUO Incision Management System is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

PREVENA PLUS 125 Negative Pressure Wound Therapy Unit

Submitter Information [21 CFR 807.92(a)(1)]
NameKCI USA, Inc.
Address6203 Farinon DriveSan Antonio, TX 78249
Phone number210-255-6171
Fax number210-255-6727
Establishment Registration Number3005178245
Name of contact personKimberly McCoy
Date preparedMay 31, 2018
Name of the device [21 CFR 807.92(a)(2)]
Trade or proprietary namePREVENA PLUS Incision Management SystemPREVENA PLUS DUO Incision Management System
Common or usual nameNegative Pressure Wound Therapy System
Classification nameNegative Pressure Wound Therapy Powered Suction Pump
Classification panelGeneral and Plastic Surgery
Regulation21 CFR 878.4780
Regulatory ClassII
Product Code(s)OMP
Legally marketed device(s) to which equivalence is claimed [21 CFR 807.92(a)(3)]Predicate 510(k) Numbers:• K173426K161897K153199
Device description [21 CFR 807.92(a)(4)]The negative pressure technology for the subject device is the same as that for the predicate device systems.The PREVENA PLUS 125 Unit is a component of a the PREVENA PLUS Incision Management System and the PREVENA PLUS DUO Incision Management System. The systems provide surgical incision management via the application of negative pressure wound therapy over an incision site that has been closed with sutures or staples. The systems are applied to the incision site immediately after surgery for a maximum of fourteen (14) days depending on the surgeon's preference.The pump in the therapy unit delivers continuous negative pressure at -125 mmHg through tubing to a dressing placed over the incision site. The occlusive drape over the dressing provides a negative pressure environment and protects the incision from external contamination.
The application of negative pressure draws the incision edgestogether, and removes fluid from the incision site into a canister fittedinto the therapy unit.
Indications for use[21 CFR 807.92(a)(5)]• The PREVENA PLUS Incision Management System is intendedto manage the environment of surgical incisions that continue todrain following sutured or stapled closure by maintaining a closedenvironment, and removing exudate via the application ofnegative pressure wound therapy.• The PREVENA PLUS DUO Incision Management System isintended to manage the environment of surgical incisions thatcontinue to drain following sutured or stapled closure bymaintaining a closed environment, and removing exudate via theapplication of negative pressure wound therapy.

{5}------------------------------------------------

Comparison of the Technological Characteristics (i.e., design, material, chemical composition, energy source) with the Predicate Device [21 CFR 807.92(a)(6)]

Negative Pressure Wound Therapy is the technological principal for both the subject and predicate devices. Application of negative pressure to an incision site that is closed via staples or sutures helps draw the incision edges together and removes fluid from the incision site. The occlusive drape of the dressing provides a negative pressure environment and protects the incision from external contamination.

At a high level, the subiect device and predicate device are based on the following same technological elements:

  • . The dressings that are applied over the incision site in the operating room are identical. One of the following dressings may be selected by the surgeon, based on incision length and geometry:
    • The PREVENA PLUS PEEL & PLACE Dressing which can be used for linear incisions O up to 35 cm or
    • The PREVENA PLUS CUSTOMIZABLE Dressing which can be configured for non o linear incisions or linear incisions up to 90 cm in length.
  • . The 13 and 20 cm PREVENA PEEL & PLACE Dressings are available in a Y configuration in the PREVENA PLUS DUO Incision Management System.
  • . A disposable negative pressure pump (therapy unit) provides -125 mmHg of negative pressure continuously to the dressing for a maximum of 7 days.
  • The dressings are connected to the therapy unit via a disposable canister and tubing set for a maximum of fourteen (14) days.
  • Incision fluid is collected into the disposable canister
  • . The therapy unit provides visual and audible alarms that indicate when negative pressure wound therapy may be compromised (e.g., alarms indicating an air leak in the system or when there is a blockage in the tubing, when the canister is full or missing, when the battery needs to be recharged, or in the case of a system fault.)
  • . The therapy unit interface provides access to therapy via an On/Off button. It also provides alarm indicators, alarm mute button, power and battery indicators, as well as a therapy life indicator.
  • The intended use and instructions for use are identical. ●
  • The following design differences exist between the subject and predicate device:
  • The disposable negative pressure pump (Therapy Unit) provides -125 mmHg of negative ● pressure continuously to the dressing for a maximum of fourteen (14) days, in distinction to the predicate Therapy Unit seven (7) day use life.

{6}------------------------------------------------

  • The dressing will need to be replaced after seven (7) days and the instruction for use (IFU) . have been updated to reflect a dressing change at this time interval.
  • The Therapy Unit life indicator now begins to count down after the eighth (8) day of therapy: . instead of the countdown beginning at the start of therapy for the predicate unit.

Performance Data [21 CFR 807.92(b)]

Summary of non-clinical tests conducted for determination of substantial equivalence [21 CFR 807.92(b)(1)]

  • Bench testing demonstrated that the PREVENA PLUS 125 Negative Pressure Wound . Therapy Unit is substantially equivalent to the predicate therapy unit in the delivery of continuous negative pressure at -125 mmHg within specifications under worst case conditions of air leak rate and fluid input during a 14 day test.
    Summary of clinical tests conducted for determination of substantial equivalence or of clinical information [21 CFR 807.92(b)(2)]

No clinical tests were necessary. No usability testing was required, as there has been no change to the user interface or to the Instructions for Use.

Conclusions drawn [21 CFR 807.92(b)(3)]

The PREVENA PLUS Incision Management System and the PREVENA PLUS DUO Incision Management System are substantially equivalent to the predicate device systems in terms of safety, performance and indications for use.

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.