K Number
K180780

Validate with FDA (Live)

Device Name
microINR System
Date Cleared
2019-01-25

(305 days)

Product Code
Regulation Number
864.7750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

microINR System:

The microlNR System (consisting of the microINR Chip) is intended for multiple-patient use by professional healthcare providers for the determination of International Normalized Ratio (INR) to monitor Oral Anticoagulation Therapy (OAT) warfarin. The microINR System uses fresh capillary whole blood. The microNR System is intended for in vitro diagnostic use at the point-of-care.

The microlNR System is intended for use in patients 18 years of age and older. Patients must be stabilized (≥ 6 weeks) on warfarin.

Caution: The microINR System is not intended for use in patients who are transitioning from heparin treatment to warfarin therapy. The microINR System is not intended to be used for screening purposes.

The microINR chips are only intended to be used with the microINR meter.

microINR Control Level A / microINR Control Level B:

The microlNR Control is intended for quality control performed on the microINR Meter when used with the disposable microINR Chips.

iLine Microsystems has available two microNR Control Levels (A and B) to perform quality control checks in the therapeutic range on the microINR System.

The microINR Control is intended for professional use only.

Device Description

The microINR® System is comprised of a portable measuring device (microINR® Meter) and test strips (microINR® Chips) in which the capillary blood sample flows through capillary action.

The microINR® Chip contains a reagent in dried form which consists of thromboplastin, and contains two symmetrical regions, the measuring channel and a control channel. The microINR® Meter measures International Normalized Ratio (INR) based on a Prothrombin Time (PT) assay carried out in the microINR® Chip based on microfluidic technology with machine vision detection.

The microINR® System has a multi-level On-board Quality Control. Multiple key functions and elements of the system are checked and if deviations are detected, error messages are displayed and test results are not reported.

In addition to the microINR® On-Board Controls, the microINR® Control has been developed as an optional liquid quality control solution. The microINR® Control must be used only by healthcare professionals in order to meet the regulatory compliance requirements applicable to the facility where they are to be used.

The microINR® Control contains lyophilized human citrated plasma from healthy donors (not heparinized plasma or plasma samples under oral anticoagulant therapy) modified by means of a dedicated process to simulate an abnormal coagulation sample and a calcium chloride solution. Before use, the lyophilized plasma must be reconstituted with the calcium chloride solution.

AI/ML Overview

Here's an analysis of the provided text to extract information about the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" for accuracy in a tabulated format that is distinct from the reported performance. The "Performance Characteristics" section outlines various analytical performance aspects, and the "Method comparison with Laboratory Reference device" section provides data related to accuracy. We can infer the performance metrics from this section.

Performance CharacteristicAcceptance Criteria (Inferred from context and predicate comparison)Reported Device Performance (microINR® System)
Accuracy (Method Comparison with Laboratory Reference Device):A strong correlation to the reference method (e.g., Pearson (r) close to 1) and acceptable slope/intercept.N: 260 Slope: 1.04 Intercept: 0.02 Pearson (r): 0.978
Repeatability (INR < 2.0)Low SD and CV% (indicating good precision)n test pairs: 91SD: 0.06 CV%: 5.25
Repeatability (2.0 ≤ INR < 3.5)Low SD and CV% (indicating good precision)n test pairs: 129SD: 0.12 CV%: 4.57
Repeatability (3.5 ≤ INR < 4.5)Low SD and CV% (indicating good precision)n test pairs: 32SD: 0.21 CV%: 5.64
Repeatability (INR ≥ 4.5)Low SD and CV% (indicating good precision)n test pairs: 17SD: 0.26 CV%: 5.46
Device Reproducibility (High Abnormal INR level plasma)Low SD and CV% for within-run, between-run, between-day, between-device, within-device and overall reproducibility.N: 90 Mean: 3.45 Device Reproducibility SD: 0.19 Device Reproducibility CV%: 5.5
Device Reproducibility (Low Abnormal INR level plasma)Low SD and CV% for within-run, between-run, between-day, between-device, within-device and overall reproducibility.N: 90 Mean: 2.25 Device Reproducibility SD: 0.12 Device Reproducibility CV%: 5.5
Hematocrit RangeNo significant effect on test results within the specified range.25 - 55% does not significantly affect test results.
Bilirubin InterferenceNo significant effect up to the specified concentration.Up to 40 mg/dL has no significant effect on test results.
Triglyceride InterferenceNo significant effect up to the specified concentration.Up to 3270 mg/dL has no significant effect on test results.
Reference RangeEstablished normal range95% of INR results ranged between 0.8-1.2 in healthy subjects.

2. Sample size used for the test set and the data provenance:

  • Accuracy Study (Method Comparison):
    • Sample Size: 260 capillary samples from patients (assumed to be from patients on warfarin therapy, based on the study's purpose).
    • Data Provenance: The document does not explicitly state the country of origin. It indicates that the study was performed at "three clinical sites" for the repeatability study which is likely part of the overall clinical evaluation. It is a prospective study as fresh capillary samples were tested.
  • Repeatability Study:
    • Sample Size: 269 paired tests (meaning 269 subjects, with two fingersticks each)
    • Data Provenance: Performed at "three clinical sites". Subjects included "normal subjects and patients on warfarin therapy." Likely prospective.
  • Reproducibility Study:
    • Sample Size: 90 (N=90 for each plasma level, though the number of individual devices/runs contributes to this count).
    • Data Provenance: Performed using "two levels of plasma samples". No country of origin mentioned. Likely prospective laboratory study.
  • Reference Range Study:
    • Sample Size: 134 healthy subjects.
    • Data Provenance: Conducted across "three sites." Likely prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The document does not mention the use of experts to establish ground truth for the test set in the context of diagnostic interpretation.

For the Accuracy Study (Method Comparison), the "ground truth" (reference method) was established by the Instrumentation Laboratory ACL TOP® 500 laboratory analyzer using HemosIL® RecombiPlasTin 2G recombinant human tissue thromboplastin reagent. This is a recognized laboratory method, not human expert consensus.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

No adjudication method for human interpretation is mentioned, as the primary comparison is against a reference laboratory instrument, not human readers.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No MRMC study was performed or described. This device is an in-vitro diagnostic system that directly measures INR, not an AI-assisted diagnostic imaging or interpretation tool for human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Yes, the studies described are of the device in a standalone capacity. The microINR® System itself performs the measurement and displays the INR result. The "Performance Characteristics" and "Comparison Studies" sections detail the analytical performance of the device without human intervention impacting the measurement itself, other than proper sample application.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth used for the accuracy study was the measurement from a validated laboratory reference device:

  • Instrumentation Laboratory ACL TOP® 500 laboratory analyzer using HemosIL® RecombiPlasTin 2G recombinant human tissue thromboplastin reagent.

For other analytical studies like precision and reproducibility, the ground truth is often inherent in the repeatability/reproducibility metrics themselves (i.e., how consistently the device measures its own result) or against certified control materials.

8. The sample size for the training set:

The document does not explicitly describe a separate "training set" in the context of algorithm development. In IVD devices like this, the device is typically developed and validated using a structured R&D process, and these performance studies represent the final validation. If any internal algorithm development or calibration involved data, it's not disclosed as a distinct "training set" in the submission summary.

9. How the ground truth for the training set was established:

As no explicit training set is detailed, information on how its ground truth was established is not available in the provided text. The calibration of the microINR® Chips is done against a reference lot of human recombinant thromboplastin traceable to the World Health Organization International Reference Preparation. This could be considered part of the "ground truth" establishment for the device's inherent measurement capability.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 25, 2019

iLine Microsystems S.L. Alberto Molinero Regulatory Affairs and Quality Manager Paseo Mikeletegi, 69 20009 Donostia, Guipúzcoa Spain

Re: K180780

Trade/Device Name: microINR System Regulation Number: 21 CFR 864.7750 Regulation Name: Prothrombin time test Regulatory Class: Class II Product Code: GJS. GGN Dated: March 22, 2018 Received: March 26, 2018

Dear Alberto Molinero:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Leonthena R. Carrington -S

Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180780

Device Name microINR System microINR Control Level A microINR Control Level B

Indications for Use (Describe)

microINR System:

The microlNR System (consisting of the microINR Chip) is intended for multiple-patient use by professional healthcare providers for the determination of International Normalized Ratio (INR) to monitor Oral Anticoagulation Therapy (OAT) warfarin. The microINR System uses fresh capillary whole blood. The microNR System is intended for in vitro diagnostic use at the point-of-care.

The microlNR System is intended for use in patients 18 years of age and older. Patients must be stabilized (≥ 6 weeks) on warfarin.

Caution: The microINR System is not intended for use in patients who are transitioning from heparin treatment to warfarin therapy. The microINR System is not intended to be used for screening purposes.

The microINR chips are only intended to be used with the microINR meter.

microINR Control Level A / microINR Control Level B:

The microlNR Control is intended for quality control performed on the microINR Meter when used with the disposable microINR Chips.

iLine Microsystems has available two microNR Control Levels (A and B) to perform quality control checks in the therapeutic range on the microINR System.

The microINR Control is intended for professional use only.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CER 801 Subpart D)
☐ Over-The-Counter Use (21 CER 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for iLine microsystems. The logo is set against a blue background. The word "iLine" is in a stylized white font, with the "i" in lowercase and the rest of the word in uppercase. The registered trademark symbol is to the right of the word. Below "iLine" is the word "microsystems" in a smaller, white font.

510(k) Summary

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.

1. SUBMITTER INFORMATION

OwneriLine Microsystems, S.L.
Paseo Mikeletegi, 69
20009 Donostia, Guipúzcoa
Spain
ContactAlberto Molinero
amolinero@ilinemicrosystems.com
Tel. +34 943 008 733
Fax: +34 943 008 737
Date Summary PreparedJanuary 18, 2019

2. DEVICE INFORMATION

Proprietary NamemicroINR® SystemmicroINR® Control Level AmicroINR® Control Level B
Common NameProthrombin time test
PanelHematology

Regulatory Information:

Classification
DeviceRegulationSectionDeviceClassProductCodeTest
microINR® System21 CFR 864.7750IIGJSProthrombin time test
microINR® Control Level A21 CFR 864.5425IIGGNPlasma, coagulationcontrol
microINR® Control Level B

510k Premarket Notification

{4}------------------------------------------------

3. SUBSTANTIAL EQUIVALENCE INFORMATION:

ElementPredicate Device
Predicate Device NameCoaguChek® XS System
Common NameProthrombin time test
510 (k) NumberK060978
ManufacturerRoche Diagnostics

4. DEVICE DESCRIPTION:

The microINR® System is comprised of a portable measuring device (microINR® Meter) and test strips (microINR® Chips) in which the capillary blood sample flows through capillary action.

The microINR® Chip contains a reagent in dried form which consists of thromboplastin, and contains two symmetrical regions, the measuring channel and a control channel. The microINR® Meter measures International Normalized Ratio (INR) based on a Prothrombin Time (PT) assay carried out in the microINR® Chip based on microfluidic technology with machine vision detection.

The microINR® System has a multi-level On-board Quality Control. Multiple key functions and elements of the system are checked and if deviations are detected, error messages are displayed and test results are not reported.

In addition to the microINR® On-Board Controls, the microINR® Control has been developed as an optional liquid quality control solution. The microINR® Control must be used only by healthcare professionals in order to meet the regulatory compliance requirements applicable to the facility where they are to be used.

The microINR® Control contains lyophilized human citrated plasma from healthy donors (not heparinized plasma or plasma samples under oral anticoagulant therapy) modified by means of a dedicated process to simulate an abnormal coagulation sample and a calcium chloride solution. Before use, the lyophilized plasma must be reconstituted with the calcium chloride solution.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for iLine microsystems. The logo is set against a blue background. The word "iLine" is written in white, with the "i" in lowercase and the "L" in uppercase. The word "microsystems" is written in a smaller font size below "iLine".

5. INDICATIONS FOR USE/INTENDED USE:

Intended Use of the microINR® System:

The microINR® System (consisting of the microINR® Meter and the microINR® Chip) is intended for multiple-patient use by professional healthcare providers for the determination of International Normalized Ratio (INR) to monitor Oral Anticoagulation Therapy (OAT) warfarin. The microINR® System uses fresh capillary whole blood. The microINR® System is intended for in vitro diagnostic use at the point-of-care.

The microINR® System is intended for use in patients 18 years of age and older. Patients must be stabilized (> 6 weeks) on warfarin.

Caution: The microINR® System is not intended for use in patients who are transitioning from heparin treatment to warfarin therapy. The microINR® System is not intended to be used for screening purposes.

The microINR® Chips are only intended to be used with the microINR® Meter.

Intended Use of the Optional Liquid Control:

The microINR® Control is intended for quality control performed on the microINR® Meter when used with the disposable microINR® Chips.

iLine Microsystems has available two microINR® Control Levels (A and B) to perform quality control checks in the therapeutic range on the microINR® System.

The microINR® Control is intended for professional use only.

6. SUMMARY COMPARISON OF TECHNOLOGICAL CHARACTERISTICS (PREDICATE):

The following table compares the microINR® System with its predicate device, CoaguChek® XS System (K060978).

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for iLine microsystems. The logo is set against a blue background. The word "iLine" is written in white, with a small "R" in a circle next to it, indicating a registered trademark. Below "iLine", the word "microsystems" is written in a smaller font, also in white.

Similarities
ItemmicroINR® SystemCoaguChek® XS System (Predicate)
Intended Use/Indications for UseThe microINR® System (consisting of the microINR® Meter and the microINR® Chip) is intended for multiple-patient use by professional healthcare providers for the determination of International Normalized Ratio (INR) to monitor Oral Anticoagulation Therapy (OAT) warfarin. The microINR® System uses fresh capillary whole blood. The microINR® System is intended for in vitro diagnostic use at the point-of-care. The microINR® System is intended for use in patients 18 years of age and older. Patients must be stabilized (≥ 6 weeks) on warfarin. Caution: The microINR® System is not intended for use in patients who are transitioning from heparin treatment to warfarin therapy. The microINR® System is not intended to be used for screening purposes. The microINR® Chips are only intended to be used with the microINR® Meter.The CoaguChek® XS system is intended for the use by professional healthcare providers for quantitative prothrombin time testing to monitor warfarin therapy, using fresh capillary or non-anticoagulated venous whole blood.
Similarities
ItemmicroINR® SystemCoaguChek® XS System(Predicate)
Sample typeCapillary whole blood.Capillary or non-anticoagulated venous wholeblood.
Test Strip ReagentHuman recombinantthromboplastin.Same.
Measuring Range0.8 – 6.0 INR.0.8 – 8.0 INR.
Hematocrit RangeHematocrit ranges between25-55% do not significantlyaffect test results.Same.
HemoglobinNo significant effect up to1000 mg/dL.Same.
Calibration traceabilityEach lot of test strips iscalibrated to a reference lottraceable to the WHOInternational ReferencePreparation.Same.
Differences
ItemmicroINR® SystemCoaguChek® XS System(Predicate)
OperatingPrinciple/TechnologyMicrofluidic technologywith machine visiondetection.Electrochemical technologywith amperometric (electriccurrent) detection of thrombinactivity
Test Strip Use Time6 hours.10 minutes.**
Test Strip Stability15 months.21 months.
Operating Temperature15 – 35°C (59 – 95°F).15 – 32°C (59 – 90°F).**
Minimum Sample VolumeA minimum of 3 µL.A minimum of 8 µL.*
On-Board Quality ControlMulti-level on-board qualitycontrols.On-board fully integratedquality controls which useelectrochemical signals todetect test strip integrity.
External Liquid QualityControlExternal optional liquidquality control: ThemicroINR® Control isintended for quality controlperformed on themicroINR® Meter whenused with the disposablemicroINR® Chips.iLine Microsystems hasavailable two microINR®Control Levels (A and B) toperform quality controlchecks in the therapeuticrange on the microINR®System.The microINR® Control isintended for professionaluse only.No external liquid qualitycontrol.**
Differences
ItemmicroINR® SystemCoaguChek® XS System(Predicate)
Memory Capacity199 (results with time anddate.)300 test results with date andtime. **
BilirubinBilirubin up to 40 mg/dLhas no significant effect ontest results.No significant effect up to 30mg/dL.
TriglycerideLipemic samples containingup to 3270 mg/dL oftriglycerides have nosignificant effect on testresults.No significant effect up to 500mg/dL.
HeparinUse excluded in IFUs.Unaffected by heparinconcentrations up to 0.8 U/mL.
Low Molecular WeightHeparinUse excluded in IFUs.Insensitive to low molecularweight heparins up to 2 IUanti-factor Xa activity/mL.
Reference RangeINR: 0.8 to 1.2.INR: 0.9 to 1.1.
CalibrationAutomatic, encoded ondisposable, no end userinput possible.Lot specific code chip selectedby end user.
Expiration date lock outAutomatic, encoded ondisposable, no end userinput possible.Lot specific code chip selectedby end user.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image features the logo for "iline microsystems." The logo is set against a blue background with rounded corners. The word "iline" is prominently displayed in white, with the "i" in lowercase and the rest of the letters in uppercase. A registered trademark symbol (®) is positioned to the upper right of "iline." Below "iline," the word "microsystems" is written in a smaller, sans-serif font, also in white.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image features the logo for iLine microsystems. The logo is set against a blue background with rounded corners. The word "iLine" is prominently displayed in white, with a registered trademark symbol to the right of the word. Below "iLine", the word "microsystems" is written in a smaller, sans-serif font, also in white.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for "iline microsystems". The logo is set against a blue background. The word "iline" is written in a bold, white sans-serif font, with a circled "R" trademark symbol in the upper right corner. Below "iline", the word "microsystems" is written in a smaller, lighter blue sans-serif font.

  • CoaguChek® XS PT Test Instructions for Use (2017-09).

** CoaguChek® XS System User Manual (2016-08 USA).

{10}------------------------------------------------

Image /page/10/Picture/2 description: The image shows the logo for iLine microsystems. The logo is set against a blue background. The word "iLine" is written in a stylized font, with the "i" in lowercase and the rest of the word in uppercase. The word "microsystems" is written in a smaller font below "iLine."

7. STANDARD/GUIDANCE DOCUMENT REFERENCED:

CLSI EP05-A3 - Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition CLSI EP09-A3 - Measurement Procedure Comparison and Bias Estimation

Using Patient Samples; Approved Guideline - Third Edition

CLSI EP07-A2 - Interference Testing in Clinical Chemistry; Approved Guideline -Second Edition

CLSI EP25-A - Evaluation of Stability of In Vitro Diagnostic. Reagents; Approved Guideline - First Edition

CLSI EP28-A3c - Defining, Establishing, and Verifying, Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition

CLSI H47-A2 - One-Stage Prothrombin Time (PT) Test and Activated Partial Thromboplastin Time (APTT) Test; Approved Guideline - Second Edition

CLSI H54-A - Procedures for Validation of INR and Local Calibration of PT/INR Systems: Approved Guideline - First Edition

8. TEST PRINCIPLE

The microINR® System is a handheld in vitro diagnostic medical device that uses microfluidic technology with machine vision detection to measure the prothrombin time from a fresh capillary (fingerstick) whole blood sample. The fresh capillary (fingerstick) whole blood sample is applied to the microINR® Chips (test strips) for testing. The microINR® Chip is inserted into the analyzer. Two microcapillary channels in the test strip are filled with the blood sample by capillary action. The microINR® Chip contains a preparation of human recombinant tissue factor, synthetic phospholipids and stabilizers. The microINR® Meter measures the International Normalized Ratio (INR) based on the Prothrombin Time (PT) assay carried out in the microINR® Chip and displays the International Normalized Ratio (INR) on the screen.

9. PERFORMANCE CHARACTERISTICS

The following internal performance data were provided in support of the substantial equivalence determination:

9.1. Analytical Performance

a) Precision/Repeatability:

Repeatability

The repeatability study was performed at three clinical sites, testing capillary samples from normal subjects and patients on warfarin therapy. Multiple Chip

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image is a logo for "iLine microsystems". The logo is set against a blue background with rounded corners. The word "iLine" is written in a bold, sans-serif font, with the "i" in lowercase and the rest of the word in uppercase. A registered trademark symbol (®) is placed to the upper right of "iLine". Below "iLine", the word "microsystems" is written in a smaller, sans-serif font.

lots, Meters and Operators were included in the study. Two capillary samples (two fingersticks) were collected from each subject resulting into a total of 269 paired tests.

Analysis per INR rangen test pairsSDCV%
INR < 2.0910.065.25
2.0 ≤ INR < 3.51290.124.57
3.5 ≤ INR < 4.5320.215.64
INR ≥ 4.517*0.265.46

Summary of repeatability for microINR® System

*4 of these values were between 6.0 and 8.0.

b) Reproducibility/Intermediate Precision:

Device reproducibility

The study was performed using two levels of plasma samples. INR was determined in both plasma samples by each device each day.

Device reproducibility data for high abnormal INR level plasma

NMeanWithin-RunBetween-RunBetween-DayBetween-DeviceWithin-DeviceDevice Reproducibility
SDCV%SDCV%SDCV%SDCV%SDCV%SDCV%
AllDevices903.450.164.70.041.00.092.70.000.00.195.50.195.5

Reproducibility data for low abnormal INR level plasma

NMeanWithin-RunBetween-RunBetween-DayBetween-DeviceWithin-DeviceDeviceReproducibility
SDCV%SDCV%SDCV%SDCV%SDCV%SDCV%
AllDevices902.250.114.90.052.00.000.00.031.40.125.30.125.5

c) Linearity/assay reportable range:

A linearity study is not applicable for the microINR® System.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image is a logo for iLine microsystems. The logo is set against a blue background. The word "iLine" is written in white, with a registered trademark symbol to the right of the word. Below "iLine" is the word "microsystems" in a smaller font, also in white.

The assay reportable range (0.8 - 6.0 INR) of the microINR® System was established through method comparison studies against the reference device (Instrumentation Laboratory ACL TOP® 500 laboratory analyzer). HemosIL® RecombiPlasTin 2G Reagent was used for prothrombin time (PT) determinations. Data collected across the assay reportable range from the method comparison studies were used to demonstrate linearity on the microINR® System.

d) Traceability, stability, expected values (controls, calibrators, or methods):

Traceability

Each lot of microINR® Chips is factory calibrated to a reference lot of human recombinant thromboplastin traceable to the World Health Organization International Reference Preparation.

microINR® Chips Stability

Real Stability: The stability of 15 months shelf life was established by storing three microINR® Chip lots at 25 ℃ (77 ℃) (60% Relative Humidity (RH)) and testing using fresh whole blood samples.

In-Use Stability: In-use stability was tested to prove that the microINR® Chips can be kept outside the pouch before measurement for at least 6 hours when directly exposed to 35 ℃ (95 ℃ (95 ℃ relative humidity RH. It was assessed by exposing the Chips out of their pouches to 35℃ (95 ℃), 80% RH storage and testing using fresh whole blood samples.

Transport Stability: microINR® Chips and its package were validated by a transport stability study. Package evaluation was assessed and the performance of the microINR® Chips was tested using fresh whole blood samples.

Optional Liquid Controls Stability

Stability: The stability of the microINR® Controls is 12 months stored at refrigerated conditions of 2-8 ℃ according to the data projected from an accelerated study performed with one lot of each microINR® Control level stored at 24.0 ℃ ± 1.1 ℃ (75.2 °F ± 34.0 °F) for up to 106 days. The in-use stability of the microINR® Controls is 10 minutes (upon reconstitution) at room temperature (15-25 °C, 59-77 °F).

Target Assignment: The target assignment is performed by using at least 10 vials of each microINR® Control lot performing two runs per day during

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for iLine microsystems. The logo is set against a blue background. The word "iLine" is written in a stylized, sans-serif font, with the "i" and "L" connected. Below "iLine" is the word "microsystems" in a smaller, sans-serif font. A registered trademark symbol is located to the right of the word "iLine".

several days. A total of 30 duplicated tests are used to determine the mean INR determination (target value).

e) Detection Limit:

The sensitivity to factors II, V, VII and X of the microINR® System was performed by mixing varying amounts of factor II, V, VII and X deficient plasma, pool normal plasma and red blood cells to final factor concentrations between 100% and 0%. These samples were then tested using nine microINR® Meters and three microINR® Chip lots in each factor sensitivity study. The sensitivity of the microINR® System to factors II, V, VII and X was estimated as 28%, 41%, 37% and 45%, respectively.

f) Analytical Specificity:

Endogenous and Exogenous Interferences

Interference limits were established analyzing fresh venous whole blood, obtained from normal subjects and patients on vitamin K antagonist (VKA) therapy, spiked separately with the interferents.

SubstanceConcentration
AcetaminophenUp to 20 mg/dL
AcetazolamideUp to 60 mg/L
Acetylsalicylic AcidUp to 83 mg/dL
AtenololUp to 10 mg/L
BilirubinUp to 40 mg/dL
CitalopramUp to 0.8 mg/L
ClopidogrelUp to 24 mg/dL
DaptomycinUp to 100 ug/mL
DiclofenacUp to 54 mg/L
DiltiazemUp to 5.44 mg/L
HemoglobinUp to 1000 mg/dL
LosartanUp to 50 mg/L
MedroxyprogesteroneUp to 0.81 mg/L
OritavancinUp to 7 mg/L
PrasugrelUp to 16 mg/dL
PravastatinUp to 0.6 mg/L
PrednisoloneUp to 3 mg/L
Salicylic AcidUp to 60 mg/dL
TicagrelorUp to 1500 ng/mL
TriglyceridesUp to 3270 mg/dL
VenlafaxineUp to 0.5 mg/L

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for iLine microsystems. The logo is set against a blue square background. The word "iLine" is written in a stylized, sans-serif font in white, with the "i" in lowercase and the rest of the word in uppercase. The word "microsystems" is written in a smaller, sans-serif font in white below "iLine."

The use of heparin, LMWH and fondaparinux is excluded in the instructions for use.

The hematocrit range was evaluated for the microINR® System using blood samples from patients. Venous samples for INR determinations using the microINR® System, citrated plasma samples for the central laboratory INR (Instrumentation Laboratory ACL Elite PRO), and the measured venous whole blood hematocrit for each test subject were used in the analysis. Data analysis demonstrated that hematocrit range between 25 - 55% does not significantly affect test results.

g) Assay cut-off:

Not applicable.

9.2. Comparison Studies

a) Method comparison with Laboratory Reference device

Accuracy was evaluated by comparing the INR results of capillary samples measured on the microINR® System against the INR of venous plasma samples measured on Instrumentation Laboratory ACL TOP® 500 laboratory analyzer using HemosIL® RecombiPlasTin 2G recombinant human tissue thromboplastin reagent (reference device). Results of INR values measured on microINR® System fingerstick capillary whole blood samples were compared to the INR measured on ACL TOP® 500 with HemosIL® RecombiPlasTin 2G using venous plasma samples.

microINR® System vs Laboratory Reference device

NSlopeInterceptPearson (r)
2601.040.020.978

b) Matrix comparison

Not applicable, as the microINR® System is intended for use with capillary whole blood samples only.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for iLine microsystems. The logo is set against a blue background with rounded corners. The word "iLine" is written in a bold, sans-serif font, with a small "R" in a circle indicating a registered trademark. Below "iLine", the word "microsystems" is written in a smaller, sans-serif font.

9.3. Clinical Studies

  • a) Clinical Sensitivity
    Not applicable

b) Clinical Specificity

Not applicable

  • c) Other clinical supportive data (when a) and b) are not applicable)
    Not applicable

9.4. Clinical cut-off

Not applicable.

9.5. Expected values/Reference range

A normal range study was conducted on 134 healthy subjects not on anticoagulation therapy across three sites. Capillary whole blood sample testing performed on the subjects not on vitamin K antagonist therapy demonstrated that 95% of the INR results ranged between 0.8-1.2.

10. INSTRUMENT NAME

microINR® Meter.

11. SYSTEM DESCRIPTION

11.1. Modes of Operation

The microINR® System is a closed system, which is intended to be used exclusively with the microINR® Chips manufactured by iLine Microsystems, S.L.

11.2. Software

The user interface of the microINR® Meter guides the user through the test procedure step by step. The user only needs to insert the Chip and apply a blood sample. The microINR® System measures International Normalized Ratio (INR)

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for iLine microsystems. The logo is set against a blue background. The word "iLine" is in white, with the "i" in lowercase and the "L" in uppercase. The word "microsystems" is in a smaller font size and is located below "iLine".

based on a Prothrombin Time (PT) assay and displays the result. After the test is completed, the meter automatically saves the test result.

11.3. Specimen Sampling and Handling

The microINR® Chip is intended for single-use only. Once the Chip is inserted into the device, a drop of fresh capillary whole blood sample collected by fingerstick is manually applied to the Chip and analyzed by the microINR® Meter.

Calibration 11.4.

Each lot of microINR® Chips is calibrated to a reference lot of human recombinant thromboplastin traced to International Reference Thromboplastin of the World Health Organization.

These calibration parameters (International Sensitivity Index (ISI) and Mean Normal Prothrombin Time (MNPT)) are encoded in the printed Datamatrix of each microINR® Chip along with information related to expiration date. Therefore, every test is automatically and individually calibrated eliminating any risk of human error.

11.5. Quality Control

The microINR® System provides both Meter´s functional Quality Controls and On-Board Ouality Controls.

First, Meter performance is automatically checked for electronic components, correct power battery level and environmental temperature conditions.

Then, On-Board Controls provide a quality control check for each individual microINR® Chip used with the microINR® Meter. microINR® System has been designed to detect errors prior to and during the test in order to prevent inaccurate INR results through a multi-level strategy.

In addition to the microINR® On-Board Controls, the microINR® Control has been developed as an optional liquid quality control solution. The microINR® Control must be used only by healthcare professionals in order to meet the regulatory compliance requirements applicable to the facility where they are to be used. The microINR® Control contains lyophilized human citrated plasma from healthy donors (not heparinized plasma or plasma samples under oral anticoagulant therapy) modified by means of a dedicated process to simulate an abnormal coagulation sample and a calcium chloride solution. The microINR® Control assigned acceptable range is included in the Instructions for Use contained in the product box.

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for iLine microsystems. The logo is set against a blue background. The word "iLine" is in white, with the "i" in lowercase and the rest of the word in uppercase. Below "iLine" is the word "microsystems" in a smaller font size, also in white. A registered trademark symbol is located to the right of the word "iLine".

12. CONCLUSION

The results of these studies demonstrate that the microINR® System is similar to the predicate in Intended Use. The data presented are a summary of external clinical evaluation, internal laboratory evaluation, and software development information. The microINR® System performance was shown to be substantially equivalent to the predicate device and demonstrated a strong correlation to the reference method.

§ 864.7750 Prothrombin time test.

(a)
Identification. A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).(b)
Classification. Class II (performance standards).