K Number
K180759

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-08-09

(140 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SECULOK™ Suture Anchor is for fixation or reattachment of soft tissue including ligament and tendon to bone especially with below indications.

  • Shoulder:
    Bankart repair, SLAP lesion repair, Acromio-Clavicular separation repair, Rotator Cuff repair, Capsular shift or Capsulolabral reconstruction, Biceps tenodesis, Deltoid repair.

  • Foot/Ankle:
    Medial or lateral stabilization, Achilles tendon repair.

  • Elbow/Wrist/Hand:
    Ulnar or Radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment.

  • Knee:
    Extra-capsular repairs: Medial/lateral ligament repair, posterior oblique ligament repair, Iliotibial band tenodesis Patellar realignment and tendon repairs : Vastus medialis obliquous advancement

Device Description

The SECULOK™ Suture Anchor consists of two types, Suture Anchor and Knotless Anchor.

The Suture Anchor is an implant used for fixation of sutures into bone. The implant is made of PEEK (ASTM F2026) and works with preloaded 1 ~ 3 USP#2 size sutures (Ultra High Molecular Weight Polyethylene(UHMWPE, ASTM F2848)). The Suture Anchor is provided sterile as single package including the implant and inserter for users to use without additional process of sterilization in hospitals.

The Knotless Anchor is an implant used for fixation of sutures into bone. The implant is made of PEEK (implant body) and Titanium alloy (tip, ASTM F136) and fixate the USP #2 sutures without knotting. The Knotless Anchor is provided sterile as single package including the implant and inserter for users to use without additional process of sterilization in hospitals.

The SECULOK™ Suture Anchor comes preloaded on a disposable inserter made from surgical grade stainless steel and ABS plastic. The entire product is packaged in a box with a Tyvek® lid, and blister the finished product is sterilized by ethylene oxide. Both the implant, suture and inserter are designed for single use only and supplied non-pyrogenic.

AI/ML Overview

Based on the provided text, the document describes a 510(k) premarket notification for the SECULOK™ Suture Anchor, demonstrating its substantial equivalence to predicate devices through performance testing.

Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

The document indicates that the acceptance criteria for performance testing were developed using data from the primary predicate devices: CrossFT BC Suture Anchor (K101100) and PushLock™ (K101679). The reported device performance is that "The all tests met all acceptance criteria and that verifies performance of the SECULOK™ Suture Anchor is substantially equivalent to predicate devices."

Test PerformedAcceptance Criteria (Source)Reported Device Performance
Suture Anchor
Torsional PropertiesData from primary predicate: CrossFT BC Suture Anchor (K101100)Met acceptance criteria
Driving TorqueData from primary predicate: CrossFT BC Suture Anchor (K101100)Met acceptance criteria
Axial Pullout StrengthData from primary predicate: CrossFT BC Suture Anchor (K101100), and Premarket Notification (510(k)) Submissions for Bone Anchors Draft Guidance (January 3, 2017)Met acceptance criteria
Axial Pullout Fatigue TestData from primary predicate: CrossFT BC Suture Anchor (K101100), and Premarket Notification (510(k)) Submissions for Bone Anchors Draft Guidance (January 3, 2017)Met acceptance criteria
Knotless Anchor
Axial Pullout StrengthData from primary predicate: PushLock™ (K101679), and Premarket Notification (510(k)) Submissions for Bone Anchors Draft Guidance (January 3, 2017)Met acceptance criteria
Axial Pullout Fatigue TestData from primary predicate: PushLock™ (K101679), and Premarket Notification (510(k)) Submissions for Bone Anchors Draft Guidance (January 3, 2017)Met acceptance criteria
Suture (part of the device)
Tensile Strength TestUSP 29 <881>Met acceptance criteria

2. Sample Size Used for the Test Set and the Data Provenance:

The document refers to "non clinical setting (bench testing)" for performance testing. It does not explicitly state the sample size used for each specific test or the "test set" in terms of data provenance (e.g., country of origin, retrospective/prospective). This information is typically detailed in the full test reports that support the 510(k) submission, but it's not present in this summary.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

This section is not applicable to this type of submission. The performance testing described is bench testing of the physical device's mechanical properties, not a clinical study involving human or animal subjects where expert ground truth would be established for diagnoses or assessments. The "ground truth" for these tests is based on established engineering standards (ASTM, USP) and mechanical properties of predicate devices.

4. Adjudication Method for the Test Set:

This section is not applicable. Since the testing is bench testing of mechanical properties, no adjudication by human experts is involved. The results are quantitative measurements against defined criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This section is not applicable. The SECULOK™ Suture Anchor is a physical medical device (suture anchor), not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study related to AI assistance for human readers would not be relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This section is not applicable. As explained above, the SECULOK™ Suture Anchor is a physical medical device, not an algorithm or AI system.

7. The Type of Ground Truth Used:

For the performance testing, the "ground truth" is established by:

  • Established engineering standards: ASTM F543-17 for torsional properties and driving torque, USP 29 <881> for suture tensile strength.
  • FDA guidance documents: Premarket Notification (510(k)) Submissions for Bone Anchors Draft Guidance (January 3, 2017) for axial pullout strength and fatigue tests.
  • Mechanical performance data of legally marketed predicate devices: CrossFT BC Suture Anchor (K101100) and PushLock™ (K101679). The acceptance criteria were specifically "developed using the data of the primary predicate."

8. The Sample Size for the Training Set:

This section is not applicable. There is no "training set" in the context of this device and the described performance testing. Training sets are typically associated with machine learning or AI algorithm development, which is not relevant here.

9. How the Ground Truth for the Training Set Was Established:

This section is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The first logo on the left is the Department of Health & Human Services - USA logo. The second logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.

August 9, 2018

U & I Corporation Bo-Ram Kang RA Specialist 20, Sandan-ro 76beon-gil(Rd) Uijeongbu-si, KR 11781 Gyeonggi-do

Re: K180759

Trade/Device Name: SECULOK™ Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: July 10, 2018 Received: July 13, 2018

Dear Ms. Kang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Melissa Hall -S

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K180759

Device Name SECULOK™ Suture Anchor

Indications for Use (Describe)

SECULOK™ Suture Anchor is for fixation or reattachment of soft tissue including ligament and tendon to bone especially with below indications.

  • Shoulder:

Bankart repair, SLAP lesion repair, Acromio-Clavicular separation repair, Rotator Cuff repair, Capsular shift or Capsulolabral reconstruction, Biceps tenodesis, Deltoid repair.

  • Foot/Ankle:
    Medial or lateral stabilization, Achilles tendon repair.

- Elbow/Wrist/Hand:

Ulnar or Radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment.

  • Knee:

Extra-capsular repairs: Medial/lateral ligament repair, posterior oblique ligament repair, Iliotibial band tenodesis Patellar realignment and tendon repairs : Vastus medialis obliquous advancement

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Manufacturer:U & I Corporation20, Sandan-ro 76beon-gil(Rd), Uijeongbu-si, Gyeonggi-do,11781, Korea,
Sponsor:U & I Corporation20, Sandan-ro 76beon-gil(Rd), Uijeongbu-si, Gyeonggi-do,11781, Korea,
Sponsor Contact:Bo-Ram Kang, Regulatory Affairs Specialist+82 31 860 6839brkang@youic.com
Date Prepared:July 10, 2018
Device Name:Trade Name: SECULOK™ Suture Anchor
Classification Name:Fastener, Fixation, nonbiodegradable, soft tissue(21 CFR §888.3040)
Common Name:Fastener, Fixation, Nondegradable, Soft Tissue
Product Code:MBI
Predicate Devices:Stryker PEEK ZIP Anchor (K070758) [Primary]CONMED CrossFT BC Suture Anchor (K101100)Stryker ReelX STT® Knotless Anchor System (K120824)Arthrex PushLock™(K101679)

Description of Device:

The SECULOK™ Suture Anchor consists of two types, Suture Anchor and Knotless Anchor.

The Suture Anchor is an implant used for fixation of sutures into bone. The implant is made of PEEK (ASTM F2026) and works with preloaded 1 ~ 3 USP#2 size sutures (Ultra High Molecular Weight Polyethylene(UHMWPE, ASTM F2848)). The Suture Anchor is provided sterile as single package including the implant and inserter for users to use without additional process of sterilization in hospitals.

The Knotless Anchor is an implant used for fixation of sutures into bone. The implant is made of PEEK (implant body) and Titanium alloy (tip, ASTM F136) and fixate the USP #2 sutures without knotting. The Knotless Anchor is provided sterile as single package including the implant and inserter for users to use without additional process of sterilization in hospitals.

The SECULOK™ Suture Anchor comes preloaded on a disposable inserter made

{4}------------------------------------------------

from surgical grade stainless steel and ABS plastic. The entire product is packaged in a box with a Tyvek® lid, and blister the finished product is sterilized by ethylene oxide. Both the implant, suture and inserter are designed for single use only and supplied non-pyrogenic.

Indications for Use:

SECULOK™ Suture Anchor is for fixation or reattachment of soft tissue including liqament and tendon to bone especially with below indications.

  • · Shoulder:
    Bankart repair, SLAP lesion repair, Acromio-Clavicular separation repair, Rotator Cuff repair, Capsular shift or Capsulolabral reconstruction, Biceps tenodesis, Deltoid repair.

  • Foot/Ankle: Medial or lateral stabilization, Achilles tendon repair.

  • Elbow/Wrist/Hand:

Ulnar or Radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment.

  • Knee:
    Extra-capsular repairs: Medial/lateral collateral ligament repair, posterior oblique ligament repair, Iliotibial band tenodesis

Patellar realignment and tendon repairs : Vastus medialis obliquous advancement

Substantial Equivalence:

The SECULOK™ Suture Anchor is described for each type as below:

  • (1) Suture Anchor
    The Suture Anchor is substantially equivalent to PEEK ZIP Anchor (K070758) and CrossFT BC Suture Anchor (K101100).

The Suture Anchor and CrossFT BC Suture Anchor are different in material, but suture Anchor is substantially equivalent to CrossFT BC Suture Anchor(K101100) in mechanical performance. The primary predicate device is the Stryker PEEK ZIP Anchor, and the others are additional predicate device.

  • (2) Knotless Anchor
    The Knotless Anchor is substantially equivalent to ReeIX STT® Knotless Anchor System (K120824) and PushLock™ (K101679).

The Knotless Anchor and PushLock™(K101679) are also different in material, but the Knotless Anchor is substantially equivalent to PushLock™ (K101679))

{5}------------------------------------------------

in mechanical performance. The primary predicate device is the ReelX STT® Knotless Anchor System, and the others are additional predicate device.

1. Comparison Technological Characteristics with the predicate device

The predicate and proposed devices have the similar intended use and basic fundamental scientific technology and share the following similarities;

  • The similar indications for use
  • Similar design features ●
  • PEEK implantable bone screw and non-absorbable sutures ●
  • Ability to secure soft tissue to bone during reconstructive procedures
  • The equivalent mechanical performance

2. Performance Testing

The SECULOK™ Suture Anchor was tested in a non clinical setting (bench testing) to assess that to no new safety and efficiency issues were raised with this device.

The Suture Anchor was performed torsional properties, driving torque, axial pullout strength, axial pullout fatigue test.

The Knotless Anchor was performed axial pullout strength.

The all tests met all acceptance criteria and that verifies performance of the SECULOK™ Suture Anchor is substantially equivalent to predicate devices. The acceptance criteria for performance testing were developed using the data of the primary predicate, CrossFT BC Suture Anchor (K101100) and PushLock'™ (K101679).

The following tests were performed:

    1. Suture Anchor
    • (1) Torsional Properties test according to ASTM F543-17
    • (2) Driving torque test according to ASTM F543-17
    • (3) Axial pullout strength test according to Premarket Notification (510(k)) Submissions for Bone Anchors Draft Guidance (January 3, 2017)
    • (4) Axial pullout fatigue test according to Premarket Notification (510(k)) Submissions for Bone Anchors Draft Guidance (January 3, 2017)
    1. Knotless Anchor
    • (1) Axial pullout strength test according to Premarket Notification (510(k)) Submissions for Bone Anchors Draft Guidance (January 3, 2017)

Image /page/5/Picture/25 description: The image shows the logo for U&I. The logo is black with a red dot on top of the "i". The "U" is larger than the "i" and is connected to the "&" symbol. The logo is simple and modern.

{6}------------------------------------------------

  • (2) Axial pullout fatigue test according to Premarket Notification (510(k)) Submissions for Bone Anchors Draft Guidance (January 3, 2017)
    1. Suture
    • (1) Tensile strength test according to USP 29 <881>

3. Sterilization Data

The sterilization parameters for the SECULOK™Suture Anchor device comply with the requirements prescribed in the applicable standards for ethylene oxide sterilization (ISO 11135-1:2014 "Sterilization of health care products -- Ethylene oxide - Requirements for development, validation and routine control of a sterilization process for medical devices" and ISO 10993-7:2008 "Biological evaluation of medical devices -- Part 7: Ethylene oxide sterilization residuals"). The sterilization cycle ensures a SAL of 10°. The EtO residuals were below the limits specified in the standard.

    1. Conclusion
      The data and information provided in this submission support the conclusion that the SECULOK™ Suture Anchor is substantially equivalent to predicate devices with respect to indications for use and technological characteristics.

N/A