K Number
K120824

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-07-18

(121 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ReelX STT Suture Anchor System is indicated for use in fixating suture or soft tissue fixation to bone in the shoulder, foot and ankle, knee, and elbow. The anchor is indicated for use in the following procedures:

Shoulder:
Rotator Cuff Repair
Bankart Repair
SLAP Lesion Repair
Biceps Tenodesis
Acromio-Clavicular Separation Repair
Deltoid Repair
Capsular Shift/Capsulolabral Reconstruction

Knee:
Medial Collateral Ligament Repair
Lateral Collateral Ligament Repair
Patellar Tendon Repair
Posterior Oblique Ligament Repair
Illiotibial Band Tenodesis

Foot and Ankle:
Lateral Stabilization
Medial Stabilization
Achilles Tendon Repair
Hallux Valgus Reconstruction
Midfoot Reconstruction
Metatarsal Ligament Repair

Elbow:
Biceps Tendon Reattachment

The ReelX STT Suture Anchor System is intended for single-use only.

Device Description

The ReelX STT Suture Anchor System Line Extension (here forth referred to as the 3.9mm ReelX STT Suture Anchor) is an anchor designed to facilitate fixation of soft tissue to bone. The system is comprised of a PEEK polymer and titanium suture anchor preloaded on a disposable inserter. This anchor design allows the surgeon to adjust the tension of the tissue after the anchor has been inserted in bone.

AI/ML Overview

This submission to the FDA is for a 510(k) premarket notification for a medical device, the Stryker ReelX STT™ Suture Anchor System Line Extension. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device rather than proving clinical efficacy or diagnostic accuracy through extensive clinical trials. Therefore, the device does not have acceptance criteria and a study proving the device meets those criteria in the typical sense for a diagnostic AI device.

However, based on the provided document, we can describe the "acceptance criteria" and "study" in the context of a 510(k) submission for a physical medical device:

Acceptance Criteria and Device Performance for Stryker ReelX STT™ Suture Anchor System Line Extension

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (based on substantial equivalence to predicates)Reported Device Performance (3.9mm ReelX STT Suture Anchor)
Insertion Strength: Substantially equivalent to predicate devices (K090530, K051219, K970896).Performance indicates substantial equivalence to predicate devices.
Fixation Strength: Substantially equivalent to predicate devices (K090530, K051219, K970896).Performance indicates substantial equivalence to predicate devices.
Intended Use: Function within the specified indications for use (fixating suture or soft tissue to bone in shoulder, foot/ankle, knee, elbow).Testing indicates the device will function within its intended use.
Design, Materials, Performance Attributes, Operational Principles: Consistent with predicate devices.Stated to be substantially equivalent to other commercially available soft tissue anchors in these aspects.

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "non-clinical ultimate tensile strength testing." It does not specify the exact sample size used for this testing. The data provenance is industrial data from a laboratory setting (Stryker/Howmedica Osteonics Corp.), not patient or clinical data. There is no mention of retrospective or prospective data as it's a non-clinical, mechanical test.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. For a mechanical device, "ground truth" is typically established by engineering standards, validated test methods, and comparison against the performance of legally marketed predicate devices, not by expert medical opinion on a test set. Engineering principles and established material testing methodologies guide the performance evaluation.

4. Adjudication Method for the Test Set

Not applicable. Mechanical testing results are objective measurements of physical properties (tensile strength) and are compared directly to predicate device performance or industry standards, not adjudicated by experts in the way clinical or image-based studies are.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical suture anchor, not an AI or diagnostic imaging device. No MRMC study was conducted or is relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical medical implant, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for this device's evaluation is primarily established by:

  • Performance of predicate devices: The mechanical properties (insertion and fixation strength) of the 3.9mm ReelX STT Suture Anchor were compared directly to those of the legally marketed predicate devices (Spin-Loc Suture Anchor System/5.5mm ReelX STT Suture Anchor, Arthrex PushLock, DePuy Mitek Panalok Anchor).
  • Engineering standards: Implied that the tests were conducted according to recognized engineering and material testing standards to ensure accurate and reproducible results for ultimate tensile strength.

8. The Sample Size for the Training Set

Not applicable. This device is a physical medical implant and does not involve a "training set" in the context of machine learning or AI.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.


Summary of the Study Proving Acceptance Criteria (in the context of 510(k)):

The study conducted was non-clinical ultimate tensile strength testing. This testing aimed to characterize the mechanical performance of the 3.9mm ReelX STT Suture Anchor, specifically its insertion strength and fixation strength.

The purpose of this testing was to demonstrate substantial equivalence to existing legally marketed predicate devices:

  • K090530 - Spin-Loc Suture Anchor System (legally marketed as Stryker ReelX STT Suture Anchor)
  • K051219 - Arthrex PushLock
  • K970896 - DePuy Mitek Panalok Anchor

The results of this non-clinical testing indicated that the performance of the 3.9mm ReelX STT Suture Anchor System "is substantially equivalent to the predicate devices, and the subject anchors will function within the intended use."

Clinical testing was explicitly stated as "not required for this submission," further emphasizing the non-clinical nature of the evidence provided to meet the requirements for initial market clearance through the 510(k) pathway.

{0}------------------------------------------------

K120824(1/3)

510(k) Summary of Safety and Effectiveness

JUL 18 2012

Proprietary Name:Stryker ReelX STTTM Suture Anchor System LineExtension
Common Name:Fastener, Fixation, Nondegradable, Soft Tissue
Classification Name and Reference:Smooth or threaded metallic bone fixation fastener21 CFR §888.3040
Proposed Regulatory Class:Class II
Product Codes:MBI: Fastener, Fixation, Nondegradable, Soft Tissue
For Information Contact:Kelly KucharczykRegulatory Affairs AssociateHowmedica Osteonics Corp.3201 East 3rd Ave.Denver, CO 80206Phone: (303) 336-7285; Fax (303) 370-5775Email: kelly.kucharczyk@stryker.com
Legally Marketed Devices to WhichSubstantial Equivalence Is Claimed:K090530 - Spin-Loc Suture Anchor System (legallymarketed as Stryker ReelX STT Suture Anchor)K051219 - Arthrex PushLockK970896 - DePuy Mitek Panalok Anchor
Date Prepared:July 11, 2012

Purpose

Stryker currently offers a 5.5mm ReelX STT Suture Anchor (cleared as the Spin-Loc Suture Anchor System- K090530). This submission serves to introduce an additional 3.9mm ReelX STT Suture Anchor to the market.

Description

The ReelX STT Suture Anchor System Line Extension (here forth referred to as the 3.9mm ReelX STT Suture Anchor) is an anchor designed to facilitate fixation of soft tissue to bone. The system is comprised of a PEEK polymer and titanium suture anchor preloaded on a disposable inserter. This anchor design allows the surgeon to adjust the tension of the tissue after the anchor has been inserted in bone.

{1}------------------------------------------------

Intended Use

The ReelX STT Suture Anchor System is indicated for use in fixating suture or soft tissue fixation to bone in the shoulder, foot and ankle, knee, and elbow. The anchor is indicated for use in the following procedures.

Indications

The ReelX STT Suture Anchor System is indicated for use in fixating suture or soft tissue fixation to bone in the shoulder, foot and ankle, knee, and elbow. The anchor is indicated for use in the following procedures:

Shoulder:Rotator Cuff Repair
Bankart Repair
SLAP Lesion Repair
Biceps Tenodesis
Acromio-Clavicular Separation Repair
Deltoid Repair
Capsular Shift/Capsulolabral Reconstruction
Knee:Medial Collateral Ligament Repair
Lateral Collateral Ligament Repair
Patellar Tendon Repair
Posterior Oblique Ligament Repair
Illiotibial Band Tenodesis
Foot and Ankle:Lateral Stabilization
Medial Stabilization
Achilles Tendon Repair
Hallux Valgus Reconstruction
Midfoot Reconstruction
Metatarsal Ligament Repair
Elbow:Biceps Tendon Reattachment

The ReelX STT Suture Anchor is intended for single-use only.

Summary of Technologies

The proposed device is substantially equivalent to other commercially available soft tissue anchors in regard to intended use. design. materials of construct, performance attributes, and operational principles. The following devices are examples of predicate systems: the Spin-Loc Suture Anchor System (marketed as and here forth referred to as the 5.5mm ReelX STT Suture Anchor) and the Arthrex PushLock.

{2}------------------------------------------------

11

Non-Clinical Testing

Non-clinical ultimate tensile strength testing was performed to characterize both the insertion strength and fixation strength of the 3.9mm ReeIX STT Suture Anchor as compared to the predicate devices identified within this premarket notification. The results of this testing indicate that the performance of the 3.9mm ReelX STT Suture Anchor System is substantially equivalent to the predicate devices, and the subject anchors will function within the intended use.

Clinical Testing

Clinical testing was not required for this submission.

Conclusion

The 3.9mm ReeIX STT Suture Anchor System is substantially equivalent to the predicate devices identified in this premarket notification.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human services.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Stryker Corporation % Ms. Kelly Kucharczyk Regulatory Affairs Associate 3201 East 310 Avenue Denver, Colorado 80206

JUL 18 2012

Re: K120824

Trade/Device Name: Stryker ReelX STT™ Suture Anchor System Line Extension Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fasterner Regulatory Class: II Product Code: MBI Dated: June 29, 2012 Received: July 01, 2012

Dear Ms. Kucharczyk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 - Ms. Kelly Kucharczyk

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Erinell Keith

ArMark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K1208241111)

Indications for Use

510(k) Number (if known): K120824

Device Name:

Stryker ReelX STT™ Suture Anchor System Line Extension

The ReelX STT Suture Anchor System is indicated for use in fixating suture or soft tissue fixation to bone in the shoulder, foot and ankle, knee, and elbow. The anchor is indicated for use in the following procedures:

Shoulder:

Rotator Cuff Repair Bankart Repair SLAP Lesion Repair Biceps Tenodesis Acromio-Clavicular Separation Repair Deltoid Repair Capsular Shift/Capsulolabral Reconstruction

Knee:

Medial Collateral Ligament Repair Lateral Collateral Ligament Repair Patellar Tendon Repair Posterior Oblique Ligament Repair Illiotibial Band Tenodesis

Lateral Stabilization Foot and Ankle: Medial Stabilization Achilles Tendon Repair Hallux Valgus Reconstruction Midfoot Reconstruction Metatarsal Ligament Repair

Biceps Tendon Reattachment Elbow:

The ReelX STT Suture Anchor System is intended for single-use only.

Prescription Usexxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx(Part 21 CFR 801 Subpart D)AND/OROver-The-Counter Use(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1(Division Sign-Off)Division of Surgical, Orthopedic,and Restorative Devices

510(k) Number ________________________________________________________________________________________________________________________________________________________________

7

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.