K Number
K180531

Validate with FDA (Live)

Date Cleared
2018-03-29

(29 days)

Product Code
Regulation Number
878.5000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NovoStitch Pro Meniscal Repair System is intended for approximation of soft tissue in meniscal repair procedures.

Device Description

The NovoStitch Pro Meniscal Repair System is a next generation of the predicate device, the NovoStitch Plus Meniscal Repair System. The NovoStitch Pro Meniscal Repair System passes size 2-0 braided, nonabsorbable, polyethylene surgical suture through soft tissue in arthroscopic surgery. It is comprised of a handheld surgical instrument to which cartridges preloaded with polyethylene suture are attached.

AI/ML Overview

This document describes the NovoStitch Pro Meniscal Repair System, a device for meniscal repair procedures. The provided text is a 510(k) summary for the device, which focuses on demonstrating substantial equivalence to a predicate device. As such, it does not detail specific acceptance criteria for performance metrics or present a comprehensive study proving the device meets those criteria in the way one might find for a novel device with new performance claims (e.g., diagnostic accuracy for an AI system). Instead, it discusses the types of tests performed to ensure the device's basic functionality, safety, and equivalence to the predicate.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria in a table format with corresponding performance metrics like sensitivity, specificity, or AUC, as these types of metrics are not typically applicable to a mechanical surgical device like a suture delivery system in the same way they are for diagnostic AI.

However, the text describes types of tests performed and some general statements of successful verification:

Acceptance Criteria (Implied from Testing)Reported Device Performance
Device strengthVerified
Device reliabilityVerified
Device insertionValidated
Suture deploymentValidated
Device removalValidated
Ability to reload with new cartridgesValidated
Ability to reliably deliver up to six stitchesDemonstrated
Device joint strengthEvaluated
Device performance evaluationsEvaluated
Biocompatibility (suture implant)Met all requirements for permanent implants in contact with tissue and/or bone
Biocompatibility (patient contact portions of delivery handle)Met all requirements for limited duration contact with tissue and/or bone
Human factors and usability processes for six essential tasksConducted and conformed to FDA guidance

2. Sample Size for Test Set and Data Provenance

  • Test Set (Human Factors/Usability): The study used 15 orthopedic surgeons.
  • Data Provenance: Not explicitly stated regarding country of origin or retrospective/prospective for the engineering/bench testing. For human factors, it was a usability study conducted in conformance with FDA guidance, implying a prospective design.

3. Number of Experts and Qualifications for Ground Truth

  • For the human factors and usability testing, 15 orthopedic surgeons served as participants. Their qualifications are stated simply as "orthopedic surgeons." The study involved them being "trained on the use of the NovoStitch Pro per the Instructions for Use, followed by a quantitative assessment." In this context, their performance of tasks is the data being evaluated, rather than them establishing a ground truth for an external dataset.
  • For bench testing and biocompatibility, the ground truth is established by standards (e.g., ISO 10993-1) and internal engineering specifications, not by clinical experts in the same way as, for example, reading medical images.

4. Adjudication Method for Test Set

  • Not Applicable in the traditional sense for diagnostic performance. For the human factors study, the method was a "quantitative assessment of the human factors and usability processes for six essential tasks." This likely involved metrics such as task completion rates, errors, and time to completion, rather than an expert adjudication process to resolve disagreements on a subjective classification.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is primarily relevant for diagnostic imaging devices or AI systems where multiple human readers interpret cases, and their performance with and without AI assistance is compared. The NovoStitch Pro is a surgical device, not a diagnostic tool.

6. Standalone (Algorithm Only) Performance Study

  • No, a standalone (algorithm only) performance study was not done. This is also not applicable as the NovoStitch Pro is a mechanical surgical device, not an AI algorithm.

7. Type of Ground Truth Used

  • Bench Testing: Ground truth is based on engineering specifications, physical measurements, and industry standards (e.g., for strength, reliability, proper articulation).
  • Biocompatibility: Ground truth is based on international standards, specifically ISO 10993-1, and specific testing protocols to determine material compatibility with biological systems.
  • Human Factors/Usability: Ground truth is based on established human factors principles and the successful or unsuccessful completion of predefined essential tasks by the users (orthopedic surgeons) according to the device's Instructions for Use.

8. Sample Size for Training Set

  • Not Applicable. The NovoStitch Pro is a mechanical device, not an AI system, so there is no "training set" in the context of machine learning.

9. How Ground Truth for Training Set Was Established

  • Not Applicable. As there is no training set for an AI algorithm. The device's design and manufacturing rely on engineering principles, material science, and established medical device development standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 29, 2018

Ceterix Orthopaedics, Inc. % Ms. Debra Cogan Regulatory Affairs Consultant for Ceterix QRAC, LLC 6500 Kaiser Drive, Suite 120 Fremont, California 94555

Re: K180531

Trade/Device Name: NovoStitch Pro Meniscal Repair System Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: February 26, 2018 Received: February 28, 2018

Dear Ms. Cogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Jennifer R. Stevenson -S3

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180531

Device Name

NovoStitch Pro Meniscal Repair System

Indications for Use (Describe)

The NovoStitch Pro Meniscal Repair System is intended for approximation of soft tissue in meniscal repair procedures.

Type of Use (Select one or both, as applicable)

Prescription (Part 21 CFR 201.56 et seq.) Dispensing Over-The-Counter (Part 21 CFR 201.66 et seq.) Dispensing
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for "Ceterix Orthopaedics". The logo consists of an orange square with a white "C" shape inside, followed by the word "CETERIX" in red, and the word "ORTHOPAEDICS" in orange below it. The logo is clean and modern, and the colors are bright and eye-catching.

SECTION 3. 510(K) SUMMARY

Manufacturer and Submitter

Company Name:Ceterix Orthopaedics, Inc.
Company Address:6500 Kaiser Drive
Suite 120
Fremont, CA 94555
TEL: (650) 316-8660
FAX: (650) 618-2779
Email: dcogan@ceterix.com
Contact Person:Debra Cogan
Date of Preparation:February 26, 2018

Device Name and Classification

Trade/Proprietary Name:NovoStitch Pro Meniscal Repair System
Common Name:Suture, Nonabsorbable, Synthetic, Polyethylene
Classification Name:Suture, Nonabsorbable, Synthetic, Polyethylene
Regulation Number:21 CFR 878.5000
Regulatory Class:II
Product Code:GAT
Panel:General & Plastic Surgery

Predicate Device

Device Name:NovoStitch Plus Meniscal Repair System
510(k) Number:K143356
Applicant:Ceterix Orthopaedics

Device Description

The NovoStitch Pro Meniscal Repair System is a next generation of the predicate device, the NovoStitch Plus Meniscal Repair System. The NovoStitch Pro Meniscal Repair System passes size 2-0 braided, nonabsorbable, polyethylene surgical suture through soft tissue in arthroscopic surgery. It is comprised of a handheld surgical instrument to which cartridges preloaded with polyethylene suture are attached.

Intended Use/Indication for Use

The NovoStitch Pro Meniscal Repair System is intended for approximation of soft tissue in meniscal repair procedures.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for "Ceterix Orthopaedics". The logo consists of an orange square with rounded corners on the left side, with three curved white lines inside. To the right of the square is the word "CETERIX" in red, with the word "ORTHOPAEDICS" in orange underneath.

Performance Data

Product performance testing for the NovoStitch Pro Meniscal Repair System was performed in a simulated use environment. Device strength, device reliability, device insertion, suture deployment, device removal, and the ability to reload the device with new suture cartridges to perform additional device insertions, deployments and removals were validated. Bench testing was performed to verify that the device and the implantable suture meet all pre-established acceptance criteria. The verification testing has demonstrated that the device can reliably deliver up to six stitches. This includes the preloaded handle assembly and up to five additional cartridges. The verification testing included device joint strength and device performance evaluations. Sterilization, packaging and shelf life were either validated or adopted from previous testing where appropriate. In addition, biocompatibility testing for each element of the NovoStitch Pro Mensical Repair System was conducted to demonstrate compliance with ISO 10993-1. The suture implant met all biocompatibility requirements for permanent implants in contact with tissue and/or bone. The patient contact portions of the delivery handle met all requirements for limited duration contact with tissue and/or bone.

Human factors and usability testing was conducted with the NovoStitch Pro. The evaluation was conducted in conformance with the FDA guidance, "Applying Human Factors and Usability Engineering to Medical Devices", February 3, 2016. In the study, 15 orthopedic surgeons were trained on the use of the NovoStitch Pro per the Instructions for Use, followed by a quantitative assessment of the human factors and usability processes for six essential tasks.

Technological Characteristics

The NovoStitch Pro Meniscal Repair System is a handheld, disposable surgical instrument coupled with an implantable suture for the approximation of soft tissue in meniscal repair procedures. The system is comprised of a delivery handle and a suture cartridge, both of which are provided sterile.

CharacteristicNovoStitch Pro Meniscal RepairSystemNovoStitch Plus MeniscalRepair System
510(k) #TBDK143356
Method Of UseHandheld, manually operated,single procedure arthroscopicsuture placement systemHandheld, manually operated,single procedure arthroscopicsuture placement system
Handle MaterialABS/PC blendPolycarbonate, 10% glass filled
Distal end materialStainless steelStainless steel
UsabilitySingle patient use, can deliverup to six sutures, disposableSingle patient use, can deliverup to six sutures, disposable
Tissue interactionTissue retained by upper andlower jawTissue retained by upper andlower jaw
Needle materialNitinolNitinol
Table 3-1.Comparison of technological characteristics to predicate device
-------------------------------------------------------------------------------

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a logo for "CETERIX ORTHOPAEDICS". The logo consists of a square with rounded corners that is filled with a gradient of orange and red. To the right of the square is the word "CETERIX" in red, with the word "ORTHOPAEDICS" in orange underneath. The logo is clean and modern, and the colors are bright and eye-catching.

Suture materialPolyethylene, nonabsorbable surgical suture, size 2-0 suturePolyethylene, nonabsorbable surgical suture, size 2-0 suture
SterilizationEthylene oxideEthylene oxide
Knot tying methodManually tied knotManually tied knot

Conclusion

The NovoStitch Pro Meniscal Repair System is substantially equivalent to the predicate device with respect to safety, effectiveness and reliability.

§ 878.5000 Nonabsorbable poly(ethylene terephthalate) surgical suture.

(a)
Identification. Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.