K Number
K143356

Validate with FDA (Live)

Date Cleared
2015-07-21

(239 days)

Product Code
Regulation Number
878.5000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NovoStitch Plus Meniscal Repair System is intended for approximation of soft tissue in meniscal repair procedures.

Device Description

The NovoStitch Plus Meniscal Repair System passes size 2-0 braided, non-absorbable, polyethylene surgical suture through soft tissue in arthroscopic surgery. It is comprised of a handheld surgical instrument to which cartridges preloaded with polyethylene suture are attached.

AI/ML Overview

The FDA 510(k) summary for the NovoStitch Plus Meniscal Repair System provides information on its acceptance criteria and the studies conducted. However, it does not contain all the requested information, especially regarding detailed statistical analysis, sample sizes for specific tests, and expert qualifications for ground truth establishment that would typically be found in a separate study report.

Here's what can be extracted and inferred from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document mentions "pre-established acceptance criteria" but does not explicitly list them with specific values. Instead, it describes general claims of meeting these criteria through various tests.

Acceptance Criteria CategoryReported Device Performance (as stated in the document)
Mechanical Performance
Device Strength & ReliabilityValidated
Device InsertionValidated
Suture DeploymentValidated
Device RemovalValidated
Reloading CapabilityValidated; device can reliably deliver up to six stitches (1 provided + 5 additional)
Device Joint StrengthVerified
Delivery Device PerformanceVerified
Suture Tensile StrengthVerified (per USP requirements for non-absorbable surgical suture)
Suture DiameterVerified (per USP requirements for non-absorbable surgical suture)
Usability
Handle Deployment ForceDemonstrated equivalent usability to predicate device
Biocompatibility
BiocompatibilityDemonstrated compliance with ISO 10993-1 requirements for permanent implants in contact with tissue and/or bone
Sterilization & Packaging
SterilizationValidated
PackagingValidated
Shelf LifeValidated

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes used for the various tests.

  • Test Set Sample Size: Not specified. The document mentions "product performance testing ... in a simulated use environment" and "bench testing."
  • Data Provenance: The document implies testing was conducted by Ceterix Orthopaedics Inc. ("Product performance testing for NovoStitch Plus Meniscal Repair System was performed..."), but does not specify the country of origin of the data or if it was retrospective or prospective. Given it's premarket notification, it would be prospective testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document. The tests seem to be primarily engineering and bench testing, not reliant on expert human interpretation for "ground truth" in the way an AI diagnostic device would be.

4. Adjudication Method for the Test Set

This information is not applicable/provided as the "ground truth" for this device appears to be based on objective physical measurements and engineering specifications, rather than subjective expert assessment requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for AI or diagnostic imaging devices where human readers interpret results, and the document describes a surgical repair system.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable to the NovoStitch Plus Meniscal Repair System, as it is a physical surgical device, not an algorithm or AI system. Its performance inherently involves a human operator (surgeon).

7. The Type of Ground Truth Used

The ground truth used for this device's performance evaluation appears to be:

  • Engineering Specifications/Standards: Such as USP requirements for suture tensile strength and diameter, and ISO 10993-1 for biocompatibility.
  • Pre-established Acceptance Criteria: Defined by the manufacturer for mechanical functions (insertion, deployment, removal, strength, reliability).
  • Simulated Use Environment Outcomes: Verification that the device performs its intended functions (e.g., reliably delivers stitches).

8. The Sample Size for the Training Set

This information is not provided and is not applicable in the context of this device. A "training set" is relevant for machine learning algorithms, not for physical medical devices being evaluated through bench and simulated use testing.

9. How the Ground Truth for the Training Set Was Established

This information is not provided and is not applicable for the same reasons as #8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 21, 2015

Ceterix Orthopaedics Incorporated Mr. Scott King Senior Director, Regulatory Affairs, Quality Assurance and Compliance 959 Hamilton Avenue Menlo Park. California 94025

Re: K143356

Trade/Device Name: NovoStitch Plus Meniscal Repair System Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable poly(ethylene terephthalate) surgical suture Regulatory Class: Class II Product Code: GAT Dated: June 12, 2015 Received: June 15, 2015

Dear Mr. King:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

For

Sincerelv vours.

Joshua C. Nipper -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K143356

Device Name

NovoStitch Plus Meniscal Repair System

Indications for Use (Describe)

The NovoStitch Plus Meniscal Repair System is intended for approximation of soft tissue in meniscal repair procedures.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "CETERIX ORTHOPAEDICS". The logo consists of an orange square with a white curved line inside, followed by the word "CETERIX" in red. Below the word "CETERIX" is the word "ORTHOPAEDICS" in a smaller, lighter orange font.

510(k) Summary

Ceterix Orthopaedics

NovoStitch Plus Meniscal Repair System

ADMINISTRATIVE INFORMATION
510(k) Number:K143356
510(k) Owner:Ceterix Orthopaedics, Inc.959 Hamilton AvenueMenlo Park, CA 94025Telephone (650) 316-8660Fax (650) 323-2209
Contact Person:Scott King959 Hamilton AvenueMenlo Park, CA 94025sking@ceterix.comTelephone (650) 316-8660Fax (650) 323-2209
DATE PREPAREDJuly 14, 2015
DEVICE NAME
Trade/Proprietary Name:Common Name:Classification Names:Classification Regulation:NovoStitch Plus Meniscal Repair SystemSuture, Nonabsorbable, Synthetic, PolyethyleneSuture, Nonabsorbable, Synthetic, Polyethylene21CFR 878.5000
Device Class:Panel:Product Code:PREDICATE DEVICESClass IIGeneral & Plastic SurgeryGAT
Delivery Device PredicateSuture Implant PredicateCayenne CrossFix II® Meniscal Repair Device (K121413)Teleflex Force Fiber® Polyethylene Non-Absorbable Surgical Suture (K063778)

DEVICE DESCRIPTION

The NovoStitch Plus Meniscal Repair System passes size 2-0 braided, non-absorbable, polyethylene surgical suture through soft tissue in arthroscopic surgery. It is comprised of a handheld surgical instrument to which cartridges preloaded with polyethylene suture are attached.

INDICATIONS FOR USE

The NovoStitch Plus Meniscal Repair System is intended for approximation of soft tissue in meniscal repair procedures.

PERFORMANCE DATA

Product performance testing for NovoStitch Plus Meniscal Repair System was performed in a simulated use environment and device strength and reliability, device insertion, suture deployment, device removal and ability to reload the device with a new suture cartridge and perform additional device insertions, deployments and removals was validated. Bench testing was performed to verify that the implantable suture

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for "Ceterix Orthopaedics". The logo consists of a stylized letter "C" in orange on the left, followed by the word "CETERIX" in red. Below "CETERIX" is the word "ORTHOPAEDICS" in a lighter shade of orange. The logo is simple and modern, with a focus on the company name.

meet all pre-established acceptance criteria. The verification testing has demonstrated that the device can reliably deliver up to six stitches. This includes the cartridge provided with the handle assembly and up to five additional cartridges. The verification testing included device joint strength, delivery device, suture tensile strength and suture diameter. Suture tensile strength and suture diameter testing was performed per USP requirements for non-absorbable surgical suture. Sterilization, packaging and shelf life were validated. In addition to these performance tests, full biocompatibility testing was conducted to demonstrate compliance with requirements for biocompatibility of permanent implants in contact with tissue and/or bone in accordance with ISO 10993-1.

TECHNOLOGICAL CHARACTERISTICS

  • The NovoStitch Plus Meniscal Repair System is a hand-held, disposable surgical instrument coupled with an implantable suture for the approximation of soft tissue in meniscal repair procedures. The system is comprised of a delivery handle and a suture cartridge, both of which are provided sterile.
  • . Comparison of technological characteristics to the delivery device predicate device
CharacteristicNovoStitch Plus Meniscal Repair SystemCrossFix II Meniscal Repair Device
Method of useHandheld, manually operated, single procedure, arthroscopic suture placement systemHandheld, manually operated, single procedure, arthroscopic suture placement system
Handle materialPolymer delivery handlePolymer delivery handle
Distal end materialStainless steel distal endStainless steel distal end
Tissue interactionTissue retained by upper and lower jawsTissue pierced with two hypotubes
Needle characteristicsFlexible, nitinol needleFlexible, nitinol needle
Suture materialPolyethylene, nonabsorbable surgical suture.Polyethylene, nonabsorbable surgical suture.
SterilizationSterile, ethylene oxideSterile, ethylene oxide
Knot tying methodManually tied knotPre-tied knot plus additional manual throws
  • . The subject device has the same knot tying method as the suture implant predicate Force Fiber® Polyethylene Nonabsorbable Surgical Suture
  • The suture used in the NovoStitch Plus Meniscal repair system is the same suture as the predicate suture cleared under K063778 and the same suture used in the predicate delivery device cleared under K121413

EQUIVALENCE TO MARKETED PRODUCT

The NovoStitch Plus Meniscal Repair Device has demonstrated equivalence to the unmodified predicate devices in the following areas:

  • Has the same intended use for approximation of soft tissues in meniscal repair procedures
  • Has equivalent technological characteristics
  • Has equivalent labeling .
  • . Has demonstrated equivalent usability with respect to handle deployment force
  • Has demonstrated product reliability and biocompatibility

In conclusion, the NovoStitch Plus Meniscal Repair System described in this submission is substantially equivalent to the predicate devices with respect to safety, effectiveness and reliability.

§ 878.5000 Nonabsorbable poly(ethylene terephthalate) surgical suture.

(a)
Identification. Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.