K Number
K180476

Validate with FDA (Live)

Date Cleared
2018-05-15

(82 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use chemotherapy drugs. Tested drugs are as follows:

Average Breakthrough Detection Time (minutes)
Cisplatin 1.0 mg/ml≥ 240
Cyclophosphamide (Cytoxan) 20 mg/ml≥ 240
Dacarbazine (DTIC) 10.0 mg/ml≥ 240
Doxorubicin Hydrochloride 2.0 mg/ml≥ 240
Etoposide (Toposar) 20.0 mg/ml≥ 240
Fluorouracil 50.0 mg/ml≥ 240
Paclitaxel (Taxol) 6.0 mg/ml≥ 240

Please note that the following drugs have extremely low permeation time for
*Carmustine (BCNU) 3.3mg.ml - 23.4 (mins)
*Thiotepa 10.0 mg/ml - 16.2 (mins)

Device Description

The proposed device is the Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green). The proposed device is for single use and is provided non-sterile. The device meets the requirements for dimensions, physical properties, and thickness in ASTM D6319 - 10(2015) Standard Specification for Nitrile Examination Gloves for Medical Application. The device accomplishes its intended function by creating a physical barrier between the examiner and patient and creates a barrier for the chemotherapy drugs listed in the Indications for Use.

AI/ML Overview

This document is a 510(k) Premarket Notification for Powder Free Nitrile Examination Gloves tested for use with chemotherapy drugs. It focuses on demonstrating substantial equivalence to a legally marketed predicate device.

Here's the information extracted and organized according to your request:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Standard)Reported Device Performance (New Device - K180476)Comparison
Chemotherapy Drugs Permeation (Minimum Breakthrough Detection Time)ASTM D6978-05
Cisplatin 1.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
Cyclophosphamide (Cytoxan) 20 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
Dacarbazine (DTIC) 10.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
Doxorubicin Hydrochloride 2.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
Etoposide (Toposar) 20.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
Fluorouracil 50.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
Paclitaxel (Taxol) 6.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
*Carmustine (BCNU) 3.3 mg/mlN/A (Comparative to predicate, with warning)23.4 minutesSimilar to predicate, with warning
*Thiotepa 10.0 mg/mlN/A (Comparative to predicate, with warning)16.2 minutesSimilar to predicate, with warning
DimensionsASTM D6319 - 10(2015)Length - Min 240mm; Thickness palm and finger - Min 0.05mmSame
Physical PropertiesASTM D6319 - 10(2015)MeetsSame
Thickness (Finger & Palm)ASTM D6319 - 10(2015)MeetsSame
Powder ContentASTM D6124 - 06(2011) (≤ 2 mg/glove)MeetsSame
Biocompatibility: Primary Skin IrritationISO 10993-10:2010Passes (non-irritating)Same
Biocompatibility: Dermal SensitizationISO 10993-10:2010Passes (non-sensitization)Same
Biocompatibility: Cytotoxicity and acute systemic toxicityISO 10993-5:2009 & ISO 10993-5:2017Passes (Certificate of Analysis for color imparting compound indicates non-cytotoxic, not systemically toxic)Different (Predicate N/A for this test)
Watertight (1000ml)21 CFR 800.20, ASTM D5151PassesSame
Indication for UseN/A (Comparative to predicate)Specialty medical glove for medical purpose, worn on examiner's hand/finger to prevent contamination between examiner and patient, and to protect against exposure to listed chemotherapy drugs.Similar
MaterialASTM D6319 - 10(2015)NitrileSame
ColorN/AGreenDifferent (Predicate: Blue or White)
SizeMedical Glove Guidance Manual - LabelingExtra Small, Small, Medium, Large, Extra LargeSame
Single UseMedical Glove Guidance Manual - LabelingSingle UseSame

The acceptance criterion for the chemotherapy drug permeation test appears to be established by comparison to the predicate devices. For most drugs, the criterion is a minimum breakthrough detection time of ">240 minutes". For Carmustine and Thiotepa, the low permeation times observed for both the new device and the predicate lead to a warning statement, indicating that these drugs have "extremely low permeation times" and are not held to the ">240 minutes" criterion.

2. Sample size used for the test set and the data provenance:

  • The document does not explicitly state the sample size used for the test set for properties like chemotherapy drug permeation, physical properties, or biocompatibility. It only reports "meets" or specific breakthrough times, implying that tests were conducted according to relevant standards (e.g., ASTM D6978-05 for chemotherapy drug permeation and ASTM D6319-10(2015) for physical properties and dimensions). These standards typically define sampling plans.
  • Data Provenance: The manufacturer is Comfort Rubber Gloves Industries Sdn. Bhd., located in Matang, Perak, Malaysia. The document does not specify if the testing was performed in Malaysia or another country. The data is non-clinical performance data, likely from prospective testing conducted to support the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not applicable as the device is a medical glove, and its performance is evaluated through objective, standardized laboratory tests against established physical, chemical, and biological criteria, not through expert interpretation of images or patient data to establish a ground truth.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • This is not applicable. The evaluation of medical gloves against performance standards (e.g., ASTM, ISO) involves quantitative measurements and predefined pass/fail criteria, not subjective interpretation requiring adjudication by experts.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • This is not applicable. The device is a physical medical glove, not an AI-powered diagnostic or assistive technology. Therefore, no MRMC study or AI-related effectiveness assessment was conducted or reported.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • This is not applicable. The device is a physical medical glove and does not involve any algorithm or AI component.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" for this device is based on objective, standardized test methods and criteria defined by international and national standards (e.g., ASTM D6319, ASTM D6124, ASTM D6978, ISO 10993, 21 CFR 800.20). These standards establish the acceptable performance characteristics for medical gloves. For biocompatibility, it's based on "passes" criteria of ISO standards. For chemotherapy permeation, it's the objectively measured breakthrough detection time.

8. The sample size for the training set:

  • This is not applicable. The device is a physical medical glove and does not utilize a "training set" in the context of machine learning or AI. Its design and manufacturing are based on engineering principles and material science, not data-driven model training.

9. How the ground truth for the training set was established:

  • This is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 15, 2018

Comfort Rubber Gloves Industries Sdn. Bhd. Chan Yew Men QA & QMS Manager Lot 821, Jalan Matang 34750 Matang, Perak Malyasia

Re: K180476

Trade/Device Name: Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZC, LZA Dated: February 12, 2018 Received: February 22, 2018

Dear Chan Yew Men:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Geeta K. Pamidimukkala -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K180476

Device Name

Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green)

Indications for Use (Describe)

The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use chemotherapy drugs. Tested drugs are as follows:

Average Breakthrough Detection Time (minutes)
Cisplatin 1.0 mg/ml≥ 240
Cyclophosphamide (Cytoxan) 20 mg/ml≥ 240
Dacarbazine (DTIC) 10.0 mg/ml≥ 240
Doxorubicin Hydrochloride 2.0 mg/ml≥ 240
Etoposide (Toposar) 20.0 mg/ml≥ 240
Fluorouracil 50.0 mg/ml≥ 240
Paclitaxel (Taxol) 6.0 mg/ml≥ 240

Please note that the following drugs have extremely low permeation time for

*Carmustine (BCNU) 3.3mg.ml - 23.4 (mins)

*Thiotepa 10.0 mg/ml - 16.2 (mins)

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

ATTACHMENT 2

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

510(k) SUMMARY

POWDER FREE NITRILE EXAMINATION GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUGS LABELING CLAIM (GREEN)

1.0 Submitter:

NameComfort Rubber Gloves Industries Sdn. Bhd.
AddressLot 821, Jalan Matang,34750 Matang, Perak, Malaysia.Malaysia.
Phone No.605-847 2777
Fax No.605-847 9108
Contact PersonChan Yew Men (Mr.)
Date of Preparation:April 17, 2018

2.0 Name of the Device

Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green)

Common Name:Patient Examination Gloves
Classification Name:Patient Examination Gloves (21 CFR 880.6250 productcode LZA)Patient Examination Gloves Specialty (21 CFR 880.6250product code LZC)
510(K) Number:K180476

3.0 Identification of The Legally Marketed Devices That equivalency is claimed:

Predicate

Device Name : Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Blue) Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (White)

  • Company : Comfort Rubber Gloves Industries Sdn. Bhd.
    : K162405 510(K) No.

4.0 Description of the Device:

The proposed device is the Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green). The proposed device is for single use and is provided non-sterile. The device meets the requirements for dimensions, physical properties, and thickness in ASTM D6319 - 10(2015) Standard Specification for Nitrile Examination Gloves for Medical Application. The device accomplishes its intended function by creating a

{4}------------------------------------------------

physical barrier between the examiner and patient and creates a barrier for the chemotherapy drugs listed in the Indications for Use.

5.0 Indication for Use of the Device

The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) are a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use with chemotherapy drugs. Tested drugs are as follows:

Average Breakthrough Detection Time (minutes)
Cisplatin 1.0 mg/ml≥ 240
Cyclophosphamide (Cytoxan) 20 mg/ml≥ 240
Dacarbazine (DTIC) 10.0 mg/ml≥ 240
Doxorubicin Hydrochloride 2.0 mg/ml≥ 240
Etoposide (Toposar) 20.0 mg/ml≥ 240
Fluorouracil 50.0 mg/ml≥ 240
Paclitaxel (Taxol) 6.0 mg/ml≥ 240

Please note that the following drugs have extremely low permeation time for *Carmustine (BCNU) 3.3 mg/ml - 23.4 (mins) *Thiotepa 10.0 mg/ml - 16.2 (mins)

6.0 Summary of the Technological Characteristics of the Device:

The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) are summarized with the following technological characteristics compared to ASTM D6319 - 10(2015) Standard Specification for Nitrile Examination Gloves for Medical Application or equivalent standards as shown in Table 1.

The chemotherapy claim is similar to the Predicate, which has a qloves thickness which complies with the ASTM standard ASTM D6319 - 10(2015): Standard Specification for Nitrile Examination Gloves for Medical Application.

CHARACTERISTICSTANDARDSDEVICE PERFORMANCEComparison
Manufacturer(s)Comfort RubberGloves IndustriesSdn. BhdComfort RubberGloves IndustriesSdn. BhdSame
510(k) numberK162405K180476
DimensionASTM D6319 -10(2015)Length-Min 240mmThickness palm andfinger- Min 0.05mmLength-Min 240mmThickness palm andfinger- Min 0.05mmSame
Physical PropertiesASTM D6319 -10(2015)MeetsMeetsSame
Thickness - Finger- PalmASTM D6319 -10(2015)MeetsMeetsSame
Powder ContentASTM D6124 -06(2011)(≤ 2 mg/glove)MeetsMeetsSame

Comparison of technological characteristics with the predicate device

{5}------------------------------------------------

CHARACTERISTICSTANDARDSDEVICE PERFORMANCE
PredicateCurrentComparison
Powder Free NitrileExamination GlovesTested for Use withChemotherapyDrugs LabelingClaim (Blue)Powder Free NitrileExamination GlovesTested for Use withChemotherapyDrugs LabelingClaim (Green)
Chemotherapy DrugPermeation TestASTM D6978-05
Test Chemotherapy DrugConcentrationMinimum Breakthrough Detection Time(min)
*Carmustine (BCNU)3.3 mg/ml18.223.4
Cisplatin1.0 mg/ml>240>240
Cyclophosphamide (Cytoxan)20 mg/ml>240>240
Dacarbazine (DTIC)10.0 mg/ml>240>240Similar - The
Doxorubicin Hydrochloride2.0 mg/ml>240>240chemotherapy
Etoposide (Toposar)20.0 mg/ml>240>240drugs tested
Fluorouracil50.0 mg/ml>240>240have similar
Paclitaxel (Taxol)6.0 mg/ml>240>240breakthrough
*Thiotepa10.0 mg/ml57.316.2detection
Warning Statement* WARNING:Please note that thefollowing drugs haveextremely lowpermeation timesCarmustine (BCNU):18.2 minutes andThiotepa: 57.3minutes.* WARNING:Please note that thefollowing drugs haveextremely lowpermeation timesCarmustine (BCNU):23.4 minutes andThiotepa: 16.2minutes.times; thedrugs with lowpermeationtimes are thesame.
DEVICE PERFORMANCE
CHARACTERISTICSTANDARDSPredicateCurrentComparison
Powder Free NitrileExamination GlovesTested for Use withChemotherapyDrugs LabelingClaim (White)Powder Free NitrileExamination GlovesTested for Use withChemotherapyDrugs LabelingClaim (Green)
Chemotherapy DrugPermeation TestASTM D6978-05
Test Chemotherapy DrugConcentrationMinimum Breakthrough Detection Time(min)
*Carmustine (BCNU)3.3 mg/ml24.023.4Similar - Thechemotherapy
Cisplatin1.0 mg/ml>240>240chemotherapy
Cyclophosphamide (Cytoxan)20 mg/ml>240>240drugs tested
Dacarbazine (DTIC)10.0 mg/ml>240>240have similar
Doxorubicin Hydrochloride2.0 mg/ml>240>240breakthrough
Etoposide (Toposar)20.0 mg/ml>240>240detection
Fluorouracil50.0 mg/ml>240>240times; the
Paclitaxel (Taxol)6.0 mg/ml>240>240drugs with low
*Thiotepa10.0 mg/ml54.916.2permeation

{6}------------------------------------------------

Please note that thefollowing drugs haveextremely lowpermeation timesCarmustine (BCNU):24.0 minutes andThiotepa: 54.9minutes.Please note that thefollowing drugs haveextremely lowpermeation timesCarmustine (BCNU):23.4 minutes andThiotepa: 16.2minutes.same.
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
CHARACTERISTICSTANDARDSDEVICE PERFORMANCEComparison
PredicateCurrent
Biocompatibility:Primary Skin IrritationISO 10993-10:2010Biologicalevaluation ofmedical devices --Part 10: Tests forirritation and skinsensitizationPassesUnder the conditionsof the study, thesubject device is non-irritatingPassesUnder the conditionsof the study, thesubject device is non-irritatingSame
Biocompatibility:Dermal SensitizationISO 10993-10:2010Biologicalevaluation ofmedical devices --Part 10: Tests forirritation and skinsensitizationPassesUnder the conditionsof the study, thesubject device is non-sensitizationPassesUnder the conditionsof the study, thesubject device is non-sensitizationSame
Biocompatibility:Cytotoxicity and acutesystemic toxicityISO 10993-5:2009Biologicalevaluation ofmedical devices --Part 5: Tests for in vitro cytotoxicityISO 10993-5:2017Biologicalevaluation ofmedical devices --Part 5: Tests forsystemic toxicityN/APassesCertificate of Analysisfor the color impartingcompound indicatesthat the compoundmeets technicalspecifications, is non-cytotoxic, and is notsystemically toxic.Different
Watertight (1000ml)21 CFR 800.20ASTM D5151PassesPassesSame
Indication for UseThe Powder FreeNitrile ExaminationGloves Tested forUse withChemotherapy DrugsLabeling Claim (Blue)is a specialty medicalglove which is adisposable deviceintended for medicalpurpose that is wornon the examiner'shand or finger topreventThe Powder FreeNitrile ExaminationGloves Tested forUse withChemotherapy DrugsLabeling Claim(Green) is a specialtymedical glove whichis a disposabledevice intended formedical purpose thatis worn on theexaminer's hand orfingerSimilar

{7}------------------------------------------------

contaminationbetween examinerand patient. Inaddition these glovesare worn to protectthe wearer againstexposure tochemotherapy drugs.Testedchemotherapy drugsare as follows:contaminationbetween examinerand patient. Inaddition these glovesare worn to protectthe wearer againstexposure tochemotherapy drugs.Tested chemotherapydrugs are as follows:examiner's hand orfinger to preventcontaminationbetween examinerand patient. Inaddition these glovesare worn to protectthe wearer againstexposure tochemotherapy drugs.Testedchemotherapy drugsare as follows:
AverageBreakthroughDetection Time(minutes)AverageBreakthroughDetection Time(minutes)AverageBreakthroughDetection Time(minutes)
Cisplatin, 1.0 mg/ml -≥ 240Cyclophosphamide(Cytoxan), 20.0mg/ml - ≥ 240Dacarbazine (DTIC),10.0 mg/ml - ≥ 240DoxorubicinHydrochloride, 2.0mg/ml - ≥ 240Etoposide (Toposar),20.0 mg/ml - ≥ 240Fluorouracil, 50.0mg/ml - ≥ 240Paclitaxel (Taxol), 6.0mg/ml- ≥ 240Cisplatin, 1.0 mg/ml -≥ 240Cyclophosphamide(Cytoxan), 20.0mg/ml - ≥ 240Dacarbazine (DTIC),10.0 mg/ml - ≥ 240DoxorubicinHydrochloride, 2.0mg/ml - ≥ 240Etoposide (Toposar),20.0 mg/ml - ≥ 240Fluorouracil, 50.0mg/ml - ≥ 240Paclitaxel (Taxol), 6.0mg/ml- ≥ 240Cisplatin, 1.0 mg/ml -$\ge$ 240Cyclophosphamide(Cytoxan), 20.0mg/ml - $\ge$ 240Dacarbazine (DTIC),10.0 mg/ml - $\ge$ 240DoxorubicinHydrochloride, 2.0mg/ml - $\ge$ 240Etoposide (Toposar),20.0 mg/ml - $\ge$ 240Fluorouracil, 50.0mg/ml - $\ge$ 240Paclitaxel (Taxol), 6.0mg/ml- $\ge$ 240
Please note that thefollowing drugs haveextremely lowpermeation time for:Please note that thefollowing drugs haveextremely lowpermeation time for:Please note that thefollowing drugs haveextremely lowpermeation time for:
Carmustine (BCNU)3.3mg/ml - 18.2(mins)Thiotepa 10.0 mg/ml- 57.3 (mins)Carmustine (BCNU)3.3mg/ml - 23.4(mins)Thiotepa 10.0 mg/ml- 16.2 (mins)Carmustine (BCNU)3.3mg/ml - 24.0(mins)Thiotepa 10.0 mg/ml- 54.9 (mins)
The Powder FreeNitrile ExaminationGloves Tested forUse withChemotherapy DrugsLabeling Claim(White) is a specialtymedical glove whichis a disposabledevice intended formedical purpose thatis worn on the

{8}------------------------------------------------

CHARACTERISTICSTANDARDSDEVICE PERFORMANCE
PredicateCurrentComparison
MaterialASTM D6319 -10(2015)NitrileNitrileSame
Color-Blue or WhiteGreenDifferent
SizeMedical GloveGuidance Manual- LabelingExtra SmallSmallMediumLargeExtra SmallSmallMediumLargeSame

{9}------------------------------------------------

Extra LargeExtra Large
Single UseMedical GloveGuidance Manual– LabelingSingle UseSingle UseSame

7.0 Summary of Non-Clinical Performance Data

The device and the predicate share the same indication for use, same base polymeric material, same specifications for thickness and length, similar permeation rate for chemotherapy drugs, similar labeling properties, biocompatibility per ISO 10993-1, water tight test results, and are powder free.

8.0 Summary of Clinical Performance Data

Clinical data is not needed for the intended use for this device.

9.0 Conclusion

Based on the Indication for Use, technological characteristics, and non-clinical performance data, the Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) (K180476) is as safe, as effective, and performs as well as the legally marketed predicate device (K162405).

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.