K Number
K180393

Validate with FDA (Live)

Date Cleared
2018-03-15

(30 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AIRO® is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding patients weighing over 400 lbs (182 kg).

Device Description

The Mobius Airo is a Mobile Intraoperative Computed Tomography (CT) System. The Airo has a large-diameter bore designed for intraoperative use; the main features include a 107cm bore, with a 51.2cm field of view (FOV). The Airo has two modes of operation; transport and scanning (both helical and axial). In its scanning mode, translation along the longitudinal axis is achieved through movement of the gantry along the length of the system base (rather than through movement of the patient support table).

The lightweight translating gantry consists of a rotating disk with a solid-state X-ray generator, solid state detector array (that includes detector modules that consist of Gadolinium Oxysulfide (GOS) and Photodiode Array). Each detector module includes a 32 x 16-pixel scintillator array that produces scintillation events responsive to irradiation by X-rays. The Airo also includes a collimator, control computer, communications link, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation), and a DC brushless servo drive system (translation).

The power system consists of batteries which provide system power while unplugged from a standard power outlet (e.g., during transport of the System and also during scanning). The base has retractable rotating caster wheels and electrical drive system can be easily moved to different locations.

In addition, the System has the necessary safety features such as emergency stop button, X-ray indicators, interlocks, patient alignment lasers, and 110 percent X-ray timer. The software helical and axial reconstruction algorithms are both based on an exact filtered-back projection.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the AIRO® Computed Tomography (CT) X-ray System (K180393). This submission aims to demonstrate substantial equivalence to a legally marketed predicate device (K160126) and primarily concerns the addition of pediatric scanning features/protocols and the removal of pediatric restrictions from the Indications for Use. The study conducted to meet acceptance criteria is a series of non-clinical bench tests and technical verification and validation activities rather than a clinical trial involving human subjects or AI-specific performance evaluation in a diagnostic context.

Here's an analysis of the provided information:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria for specific performance metrics alongside reported device performance. Instead, it states that the modified device's performance, particularly related to the added pediatric features and protocols, meets acceptance criteria through various testing. The comparison table (pages 5-7) primarily highlights the equivalence of technological characteristics between the proposed device and the predicate. The key differences causing the need for this submission are:

  • Added pediatric scanning feature/protocols: This required the addition of 80 and 100 kV scanning capabilities and updated software to include age and height parameters.
  • Modified Indications for Use Statement: Removal of pediatric restriction.

The "reported device performance" is a general statement that "the modified Airo CT System meets the acceptance criteria" based on the performed non-clinical tests. Specific numerical performance metrics (e.g., image quality scores, dose reduction percentages for pediatric protocols) are not detailed within the provided text. The only specific metric given is "Spatial Resolution for Sharpest Clinical Algorithm (Ip/cm at 2%) = 6.9," which is identical to the predicate.

2. Sample size used for the test set and the data provenance

The study described is a non-clinical bench testing and verification/validation effort. Therefore, there is no "test set" in the sense of a dataset of patient images or clinical cases. The "test set" would consist of phantoms used for image quality metrics, dose testing, and protocol validation. The text mentions:

  • "Battery of bench testing with phantom images presented in Section 18." (Section 18 is not included in the provided text).
  • "Image Quality Metrics and phantom images for Pediatrics."
  • "Pediatrics Protocol Design & Validation (using Image Gently, ACR and AAPM guidelines)."
  • "Radiation/Dose Testing."

The provenance of this phantom data would be internal testing conducted by Mobius Imaging, LLC in the USA. It is inherently "prospective" in the sense that the tests were designed and executed to validate the modified device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Given that this is a non-clinical bench study focused on engineering validation and demonstrating general safety and effectiveness, there is no mention of experts establishing a "ground truth" for a test set in the diagnostic interpretation sense. The "ground truth" for performance would be derived from physical measurements on phantoms and compliance with recognized standards and guidelines (e.g., Image Gently, ACR, AAPM guidelines for pediatric protocols). While experts might have been involved in defining these standards or interpreting test results, their number and specific qualifications are not specified here.

4. Adjudication method for the test set

Not applicable, as this is a non-clinical engineering study of a CT device itself, not an AI software evaluating images, and therefore does not involve human adjudication of diagnostic interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This submission is for a CT X-ray system, not an AI-powered diagnostic tool. Therefore, an MRMC comparative effectiveness study to assess human reader improvement with or without AI assistance was not performed or described.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. The AIRO® Computed Tomography (CT) X-ray System is a medical imaging device, not a standalone AI algorithm for image analysis. The "AI" in AIRO® is likely part of the product name and not an indicator of artificial intelligence functions for diagnosis or analysis in the context of this submission. The "software" updates mentioned are for scanner control and protocol management, not diagnostic AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical testing conducted, the "ground truth" primarily refers to established physical standards and metrics for CT image quality, radiation dose, and compliance with recognized regulatory and industry standards (e.g., IEC 60601 series, NEMA XR standards, Image Gently, ACR, AAPM guidelines). This "ground truth" is based on:

  • Physical measurements on phantoms: To assess image quality, spatial resolution, contrast, noise, and radiation dose.
  • Compliance with specified engineering and performance requirements: As defined by the company and aligned with international and national standards.

8. The sample size for the training set

Not applicable. This is a submission for a CT hardware system with software updates, not an AI model requiring a training set of data.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for an AI model in this submission.

{0}------------------------------------------------

March 15, 2018

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

Mobius Imaging, LLC % Ms. Norma J. LeMay Regulatory Affairs Consultant 2 Shaker Road SHIRLEY MA 01464

Re: K180393

Trade/Device Name: AIRO® Computed Tomography (CT) X-ray System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: February 12, 2018 Received: February 13, 2018

Dear Ms. Lemay:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara
For

Robert Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K180393

Device Name AIRO® Computed Tomography (CT) X-ray System

Indications for Use (Describe)

The AIRO® is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding patients weighing over 400 lbs (182 kg).

Contraindications:

Airo is contraindicated as the principal means of guidance during invasive procedures when real-time imaging is needed (e.g., CT Fluoroscopy procedures). In addition, the physician must verify navigation accuracy using an adequate verification method when used with Navigation Systems during surgical procedures.

Type of Use (Select one or both, as applicable)

× Prescription Use (Part 21 CFR 801 Subpart D)

D Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the Mobius Imaging logo at the top, with the words "Mobius" and "IMAGING Now You See" displayed. Below the logo, the text "510(k) Summary" is written in a large, bold font. The text is black and appears to be the title or heading of a document or presentation.

This 510(k) Summary of Safety and Effectiveness information is prepared in accordance with the requirements of 21 CFR Part 807.92.

1. Submitter:

Mobius Imaging, LLC 2 Shaker Road, Suite F100 Shirley, MA 01464 USA

Contact:

Norma LeMay Regulatory Affairs Consultant Telephone: 978-391-9644 Email: nlemay@mobiusimaging.com

Date of Preparation: February 12, 2018

2. Device Name & Regulatory Classification:

Proprietary or Trade Name: AIRO® Computed Tomography (CT) X-ray System Classification Name: Computed Tomography X-ray System Product Code: 90 JAK Device Classification: Class II Regulation Number: 21 CFR 892.1750

3. Predicate Device(s):

The legally marketed device to which substantial equivalence is being claimed is as follows:

  • . Airo Mobile CT System – Premarket Notification K160126 (cleared on April 19, 2016, product code 90 JAK)

4. Device Description:

The Mobius Airo is a Mobile Intraoperative Computed Tomography (CT) System. The Airo has a large-diameter bore designed for intraoperative use; the main features include a 107cm bore, with a 51.2cm field of view (FOV). The Airo has two modes of operation; transport and scanning (both helical and axial). In its scanning mode, translation along the longitudinal axis is achieved through movement of the gantry along the length of the system base (rather than through movement of the patient support table).

{4}------------------------------------------------

The lightweight translating gantry consists of a rotating disk with a solid-state X-ray generator, solid state detector array (that includes detector modules that consist of Gadolinium Oxysulfide (GOS) and Photodiode Array). Each detector module includes a 32 x 16-pixel scintillator array that produces scintillation events responsive to irradiation by X-rays. The Airo also includes a collimator, control computer, communications link, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation), and a DC brushless servo drive system (translation).

The power system consists of batteries which provide system power while unplugged from a standard power outlet (e.g., during transport of the System and also during scanning). The base has retractable rotating caster wheels and electrical drive system can be easily moved to different locations.

In addition, the System has the necessary safety features such as emergency stop button, X-ray indicators, interlocks, patient alignment lasers, and 110 percent X-ray timer. The software helical and axial reconstruction algorithms are both based on an exact filtered-back projection.

5. Indications for Use:

The AIRO® is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding patients weighing over 400 lbs (182 kg).

Contraindications:

Airo is contraindicated as the principal means of guidance during invasive procedures when realtime imaging is needed (e.g., CT Fluoroscopy procedures). In addition, the physician must verify navigation accuracy using an adequate verification method when used with Navigation Systems during surgical procedures.

6. Comparison of Technological Characteristics with the Predicate Device:

As detailed in Section 12 of this 510(k) Premarket Notification, the modified Airo CT System, for its intended use, is of comparable type in design, material, functionality, technology and is considered substantially equivalent to its baseline device Airo CT System (K160126, cleared on April 19, 2016) based on the following comparison. Most importantly, the differences noted below and in the comparison table raise no new issues of safety or effectiveness based on all testing performed and presented in this 510(k):

  • Added pediatric scanning feature/protocols
  • . Updated software to include scanning capabilities at 80 and 100kV to support pediatrics
  • Updated software to include age and height parameters to the system protocols
  • Modified the Indications for Use Statement to remove pediatric restriction

{5}------------------------------------------------

Predicate Device Technological Comparison

CompanyMobius Imaging, LLCMobius Imaging, LLCComparison
510(k) NumberThis 510(k)K160126
Model NameAiro CT SystemAiro CT System
Technological Characteristics
Indications for UseThe AIRO is intended to beused for X-ray computedtomography applications foranatomy that can be imagedin the 107cm apertureexcluding patients weighingover 400 lbs (182 kg).Airo is contraindicated as theprincipal means of guidanceduring invasive procedureswhen real-time imaging isneeded (e.g., CT Fluoroscopyprocedures). In addition, thephysician must verifynavigation accuracy using anadequate verification methodwhen used with NavigationSystems during surgicalprocedures.The AIRO is intended to beused for X-ray computedtomography applications foranatomy that can be imagedin the 107cm apertureexcluding pediatric patientsand patients weighing over400 lbs (182 kg).Airo is contraindicated as theprincipal means of guidanceduring invasive procedureswhen real-time imaging isneeded (e.g., CT Fluoroscopyprocedures). In addition, thephysician must verifynavigation accuracy using anadequate verification methodwhen used with NavigationSystems during surgicalprocedures.Equivalent -Removal of pediatricrestriction only. Battery of benchtesting with phantom imagespresented in Section 18.
Product Classification CodeJAKJAKIdentical
Aperture (cm)107107Identical
Image Field of View (cm)51.251.2Identical
Detector MaterialGadolinium Oxysulfide(GOS)Gadolinium Oxysulfide(GOS)Identical
Detector Configuration32 x 2.0mm32 x 2.0 mmIdentical
Spatial Resolution forSharpest Clinical Algorithm(Ip/cm at 2%)6.96.9Identical
X-ray Tube TypeRotating AnodeRotating AnodeIdentical
Heat Storage (MHU)1.71.7Identical
X-ray Tube CoolingLiquid (50% Water, 50%Propylene Glycol)Liquid (50% Water, 50%Propylene Glycol)Identical
X-ray Fan Angle (deg)4545Identical
CompanyMobius Imaging, LLCMobius Imaging, LLC
510(k) NumberThis 510(k)K160126
Model NameAiro CT SystemAiro CT System
Technological CharacteristicsComparison
Scanning capabilities (kV)80, 100, 120 kV120 kV (80 and 100 kVwas not used)Equivalent - SW change forPediatrics- Battery of benchtesting has been completed
Max X-ray Power (kW)3232Identical
Rotating Speed (seconds)22Identical
Gantry Weight (kg)10681068Identical
Transfer of electric currentData Dock systemData Dock systemIdentical
Mechanism to translateGantryRails on Mobile BaseSystemRails on Mobile BaseSystemIdentical
WirelessNoNoIdentical
MobileYes (motorized)Yes (motorized)Identical
Battery SystemYes (LiFePO4)Yes (LiFePO4)Identical
Wheels (casters)Wheels (3 inch)Wheels (3 inch)Identical
Input Voltage1 phase 100-240 volt1 phase 100-240 voltIdentical
Input Power Max1.5 kW1.5 kWIdentical
PACS/DICOM 3.0YesYesIdentical
2D ScoutYesYesIdentical
Bolus trackingNoNoIdentical
Dynamic ScanNoNoIdentical
Axial/Helical ModesAxial and HelicalAxial and HelicalIdentical
Tube Modulation FeatureYes (Helical mode only)Yes (Helical mode only)Identical
MPRYesYesIdentical
3D ViewingNoNoIdentical
Patient TableYes (optional Trumpf tablecolumn integrated withbase)Yes (optional Trumpf tablecolumn integrated withbase)Identical
Predicate Device Technological Comparison
CompanyMobius Imaging, LLCMobius Imaging, LLC
510(k) NumberThis 510(k)K160126
Model NameAiro CT SystemAiro CT System
Technological CharacteristicsComparison
Scan MotionScanner MovesScanner MovesIdentical
Laser AlignmentPatient AlignmentPatient AlignmentIdentical
Radiation Safeguards Hardware
X-ray warning lightsYesYesIdentical
110% X-ray TimerYesYesIdentical
Emergency StopYesYesIdentical
Internal Lead ShieldYesYesIdentical - Internal lead shieldbehind Detector forCollimated X-ray Beam and1/2-inch aluminum Gantryhousing for scatter shielding
External Lead CurtainsNoNoIdentical
Operator X-ray on SwitchYesYesIdentical
Quality Test PhantomYesYesIdentical
Radiation Safeguards Software
Login ID/passwordYesYesIdentical
Administrator PrivilegesYesYesIdentical
Dose DisplayYesYesIdentical
Dose Report/AuditYesYesIdentical
Protocol Override ProtectionYesYesIdentical
Protocols by age, height,weight, and body regionYesNo (weight/body regiononly)Equivalent – SW 2.0 added ageand height for Pediatrics –Battery of bench testingpresented in Section 18.
QA Test ReportYesYesIdentical
Quality Test PhantomYesYesIdentical
Predicate Device Technological Comparison
CompanyMobius Imaging, LLCMobius Imaging, LLC
510(k) NumberThis 510(k)K160126
Model NameAiro CT SystemAiro CT System
Technological CharacteristicsComparison
Operating SystemMicrosoft WindowsMicrosoft WindowsIdentical
BiocompatibilityN/AN/AIdentical
EM EmissionsETL TestingETL TestingIdentical
SterilityN/AN/AIdentical
Chemical SafetyN/AN/AIdentical
Thermal SafetyETL TestingETL TestingIdentical
IEC EN 60601 Electrical SafetyTestingETL TestingETL TestingIdentical
IEC EN 60601 MechanicalSafety TestingETL TestingETL TestingIdentical
Can be used inMobile or Fixed Radiology,ICU, ED, Surgical, Clinic,officeMobile or Fixed Radiology,ICU, ED, Surgical, Clinic,OfficeIdentical
Anatomical SiteThat which can be imagedin 51.2cm FOV and 107cmApertureThat which can be imagedin 51.2cm FOV and 107cmApertureIdentical

{6}------------------------------------------------

Predicate Device Technological Comparison

{7}------------------------------------------------

{8}------------------------------------------------

7. General Safety and Effectiveness Concerns:

ldentical to its baseline device, all components of the Airo CT System are subject to Federal Diagnostic Equipment Performance Standard and applicable regulations of 21 CFR Part 1020.30 and 1020.33 are certified to meet those requirements. Initial and Annual Reports are filed with the Center for Devices and Radiological Health (CDRH) according to 21 CFR 1002.10, respectively.

To minimize electrical, mechanical and radiation hazards, Mobius adheres to recognized and established industry practices. Additionally, the Airo CT System was designed and tested to the following FDA recognized International harmonized and National standards:

  • IEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007) + A1:2012 Medical electrical equipment . Part 1: General requirements for basic safety and essential performance

{9}------------------------------------------------

  • IEC 60601-1-2:2007 - Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic Compatibility -Requirements and Tests
  • . IEC 60601-1-3:2008 + A1:2013 - Medical electrical equipment – Part 1-3: General Requirements for basic safety and essential performance - Collateral Standard Radiation Protection in Diagnostic X-Ray Equipment
  • . IEC 60601-1-6:2010 +A1:2013 - Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
  • IEC 60601-2-44:2009 + A1:2012 + A2:2016 - Medical electrical equipment – Part 2-44: Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography
  • IEC 60825-1:2007 - Safety of laser products
  • IEC 61223:2004 - Evaluation and routine testing in medical imaging departments - Part 3-5 Acceptance Tests – Imaging Performance of Computed Tomography X-ray equipment
  • . IEC 62133:2012 - Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
  • IEC 62304:2006 +A1:2015 - Medical device software – Software life cycle processes
  • IEC 62366:2007 + A1:2014 - Medical devices – Application of usability engineering to medical devices
  • NEMA XR 25-2010 - Computed Tomography Dose Check
  • NEMA XR 29-2013 - Standard Attributes on CT Equipment Related to Dose

In addition to the standards above, the following FDA Guidance Documents for pediatric use with medical devices and for Cybersecurity were referenced when preparing this 510(k):

  • FDA Guidance for "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" dated October 2, 2014
  • FDA Guidance for "Postmarket Management of Cybersecurity in Medical Devices" dated December 28, 2016
  • FDA Guidance for "Pediatric Information for X-ray Imaging Device Premarket Notifications" dated November 28, 2017
  • . FDA Guidance for "Premarket Assessment of Pediatric Medical Devices" dated March 24, 2014

Relating to concerns regarding unintentional radiation exposure, the Airo CT System has software safeguards such as: ID password/login, dose display/reporting, safety warning to prevent excessive dose, protocol protection and required quality assurance testing.

{10}------------------------------------------------

8. Determination of Substantial Equivalence:

Summary of Non-clinical tests:

The Airo CT System complies with the voluntary harmonized standards as detailed above and in Section 9 and 17 of this 510(k) Premarket Notification. In addition, the following quality assurance measures were applied to the development of the system software modifications and performance testing was performed to support substantial equivalence:

  • Risk Analysis
  • Design Reviews
  • Design Verification Testing
  • . SW Unit Integration Testing
  • System Software Verification & Validation Testing ●
  • Image Quality Metrics and phantom images for Pediatrics
  • Pediatrics Protocol Design & Validation (using Image Gently, ACR and AAPM guidelines)
  • Radiation/Dose Testing ●
  • Electrical Safety, Mechanical & Stability Testing
  • EMI/EMC Testing

The results of all testing performed indicate that the modified Airo CT System meets the acceptance criteria and is substantially equivalent to the currently cleared baseline device (Airo Mobile CT System — K160126).

9. Conclusion:

Based upon the above considerations, including all testing presented in this 510(k), Mobius considers the modified Airo CT System to be as safe, as effective, and performance is substantially equivalent to its baseline device. We also believe that the proposed modifications to the Airo CT System raise no new issues of safety and/or efficacy and the device performs as intended.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.