K Number
K180296

Validate with FDA (Live)

Device Name
Afinion HbA1c Dx
Date Cleared
2018-05-07

(94 days)

Product Code
Regulation Number
862.1373
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Afinion™ HbA1c Dx is an in vitro diagnostic test for quantitative determination of glycated hemoglobin (% hemoglobin A1c, HbA1c) in human venous and capillary whole blood.

This test is to be used as an aid in the diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes.

The measurement of % HbA1c is recommended as a marker of long-term metabolic control in persons with diabetes mellitus.

Device Description

The Afinion™ HbA1c Dx is a fully automated boronate affinity assay for the determination of the percentage of hemoglobin A1c in human whole blood. The Afinion™ HbA1c Dx is a modification of the existing device, Alere Afinion™ HbA1c for use with the Alere Afinion™ AS100 Analyzer, with the addition of a diagnostic intended use.

The test begins with a blood sample collected with the integrated sampling device before the test cartridge is placed in the cartridge chamber of the Alere Afinion™ AS100 Analyzer. The sample is then automatically diluted and mixed with a solution that releases hemoglobin from the erythrocytes. After the hemoglobin is precipitated, the sample mixture is transferred to a blue boronic acid conjugate which binds to the cis-diols of glycated hemoglobin. This reaction mixture is soaked through a filter membrane and all precipitated hemoglobin, conjugate-bound and unbound (i.e. glycated and non-glycated hemoglobin) remains on the membrane. Excess conjugate is removed with a washing reagent. The analyzer measures the reflectance of the precipitate on the membrane as blue (glycated hemoglobin) and red (total hemoglobin) color intensities. The analyzer calculates a ratio proportional to the percentage of HbA1c in the sample and displays as the % HbA1c (NGSP).

AI/ML Overview

The medical device is the Afinion™ HbA1c Dx, a fully automated boronate affinity assay for the determination of the percentage of hemoglobin A1c in human whole blood. It is an in vitro diagnostic test for the quantitative determination of glycated hemoglobin (% hemoglobin A1c, HbA1c) in human venous and capillary whole blood. Its intended use is as an aid in the diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes. It is also used as a marker of long-term metabolic control in persons with diabetes mellitus.

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (from 21 CFR 862.1373 Special Controls)Reported Device Performance
System Accuracy (Total Error)Total Error (TE) ≤ 6.0%Fingerstick Whole Blood Samples:Range: 2.87% to 4.75% TE (using Weighted Deming or Passing-Bablok regressions from bias and precision from 3 studies combined).Venous Whole Blood Samples (Internal Precision Study):Range: 2.77% to 3.80% TE (using Weighted Deming or Passing-Bablok regressions from bias and internal precision studies).Venous Whole Blood Samples (Point of Care, External Precision Study):Range: 2.64% to 4.07% TE (using Weighted Deming or Passing-Bablok regressions from bias and external precision studies).All reported TE values are ≤ 6.0%.
Precision/ReproducibilityNot explicitly stated as a single numerical acceptance criteria in terms of SD or CV, but assessed through detailed studies.Internal (Venous Whole Blood): Total %CV ranged from 1.32% to 1.74% across HbA1c levels and analyzers.External (Venous Whole Blood): Total %CV ranged from 1.22% to 1.78% across HbA1c levels.Fingerstick Samples (Combined Studies): Total %CV ranged from 1.30% to 2.03% across HbA1c levels.
Linearity/Reportable RangeNo specific numerical acceptance criterion stated, but implicitly expected to cover the medical decision range.Reportable range: 4.00-15.00 % HbA1c (DCCT/NGSP). Previously established in K050574.
Endogenous InterferenceSignificant interference defined as exceeding a 7% change in %HbA1c value from control.No significant interference observed for tested substances (Bilirubin, Glucose, Intralipid, Rheumatoid factor, Total protein) at specified concentrations.
Hemolysis InterferenceSignificant interference defined as exceeding a 7% change in %HbA1c value from control.No significant interference observed up to 24% hemolysis. Information codes related to hemolysis may occur above 14% hemolysis.
Drug InterferenceSignificant interference defined as exceeding a 7% change in %HbA1c value from control.No significant interference observed for 20 tested drugs at specified concentrations.
Cross-reactivity with Hemoglobin DerivativesSignificant interference defined as exceeding a 7% change in %HbA1c value from control.No significant interference observed for Acetylated hemoglobin, Carbamylated hemoglobin, Labile HbA1c, and Glycated albumin at specified concentrations.
Hemoglobin Variants InterferenceSignificant interference defined as exceeding a 7% change in %HbA1c value from reference method.No significant interference for HbA2, HbS, HbC, HbE, HbD. Significant negative interference with HbF (highest concentration with no significant interference at 10.4% HbF). Device includes a prominent boxed warning for HbF.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Accuracy/Method Comparison Study (Test Set):
    • Sample Size: 120 subjects.
    • Data Provenance: Samples from each study subject were tested with both fingerstick samples and fresh venous EDTA samples. Divided across three study sites. Retrospective or prospective is not explicitly stated, but the collection of fresh venous and fingerstick samples suggests it was prospective. The country of origin of the data is not specified.
  • Precision (Internal, Venous Whole Blood): 4 levels of HbA1c patient samples. Each level tested with 2 replicates, twice a day for 20 days with 3 lots on 3 analyzers. (240 measurements per analyzer, 720 combined).
  • Precision (External, Venous Whole Blood): 4 levels of HbA1c patient samples. 4 replicates analyzed twice a day for 10 days with 3 lots at each of 3 sites, using 3 analyzers.
  • Precision (Fingerstick Samples):
    • Study A (Accuracy study - within-run): 172 subjects (fingerstick samples in duplicate).
    • Study C (Between-instrument and between-operator): 15-16 subjects per HbA1c level (total of 4 levels) across 3 sites. Each subject gave 6 fingerstick samples. Total of 90-96 fingerstick measurements per level.
  • Between-Instrument Precision: 4 venous whole blood samples measured in 6 replicates on each of 14 analyzers with 1 test cartridge lot (total of 84 replicates per sample).
  • Lot-to-Lot Variation: 18 EDTA venous whole blood samples spanning the reportable range. Each sample analyzed in 1 replicate with each of 3 test cartridge lots on the same analyzer.
  • Endogenous/Hemolysis/Drug Interference, Hemoglobin Derivatives Cross-reactivity: Whole blood sample pools. Specific number of samples not given, but tested with 10 replicates for each condition/substance.
  • Hemoglobin Variants Interference: 234 fresh EDTA whole blood samples containing 6 common hemoglobin variants. Also, 100 samples for HbA0, HbA1a, HbA1b components.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • For the Method Comparison study, the ground truth was established by an NGSP secondary reference laboratory method (Tosoh Glycohemoglobin test on the G8 HPLC analyzer). This method itself serves as the "expert" or gold standard. The document does not specify the number or qualifications of human experts involved in operating or verifying this reference method.

4. Adjudication Method for the Test Set

  • The document implies that the reference method (Tosoh G8 HPLC) was considered the definitive ground truth for the method comparison study. There is no mention of a human adjudication method (like 2+1 or 3+1 consensus) for the test set results against the reference method. The comparison was statistical, using Deming and Passing-Bablok regressions.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This device is an automated in vitro diagnostic test (HbA1c assay). It does not appear to involve "human readers" in the sense of image interpretation or other judgmental tasks where AI assistance might improve their performance. Therefore, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance is not applicable to this device.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

  • Yes, the performance studies described are for the "algorithm only" (device only) performance. The Afinion™ HbA1c Dx is a fully automated system for determining HbA1c. The performance metrics (precision, accuracy, interference, etc.) reflect the standalone operation of the device.

7. The Type of Ground Truth Used

  • The ground truth used for the method comparison and hemoglobin variant interference studies was a reference method, specifically an NGSP secondary reference laboratory method (Tosoh Glycohemoglobin test on the G8 HPLC analyzer). For interference studies, the ground truth was derived from non-spiked control samples or reference samples without the interfering substance.

8. The Sample Size for the Training Set

  • The document describes performance characteristics for the Afinion™ HbA1c Dx. It does not provide information on a separate "training set" for an algorithm, as this typically applies to machine learning models. Instead, the device is a chemical assay with established analytical principles. Therefore, a distinct "training set" size in the context of an AI/ML algorithm is not applicable as presented in this document. The "training" or development would refer more to the optimization and validation of the analytical method itself.

9. How the Ground Truth for the Training Set Was Established

  • As a chemical assay rather than an AI/ML algorithm, the concept of a "training set" with established ground truth is not applicable in the same way. The device's analytical method (boronate affinity assay) is based on scientific principles and validated through extensive performance testing as detailed in the document, against established reference methods.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 7, 2018

Alere Technologies AS Annie Wright Senior Manager, Regulatory Affairs Kjelsaasveien 161 PO Box 6863 Rodeloekka NO-0504 Oslo Norway

Re: K180296

Trade/Device Name: Afinion HbA1c Dx Regulation Number: 21 CFR 862.1373 Regulation Name: Hemoglobin A1c test system Regulatory Class: Class II Product Code: PDJ Dated: February 5, 2018 Received: February 6, 2018

Dear Annie Wright:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Kellie B. Kelm -S for Courtney H. Lias, Ph.D.

Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

resurement un obliev stures a currently ville OMB number.
" " would be may budges of persons is nosted is nosted to responsible to a collection of
PRASSISMONOMOR. gov
Paperwork Reduction Act (PRA) StateOffice of Chief Information Officer
Department of Health and Human ServicesFood and Drug Administration
of this information collection, including suggestions for reducting this burden, to:sees and of the rolled in the building the survey would be sunder in the nower burnim to the best in the mistment but news sources, gating the states meeded and complete
and pribular in and sund of a gogeners of betermines and not online to not only not and replaned"O OF SERIETY YOUNG YOUNG COMMENT FOR STAFF FARIL ADDRESS BELOW"
. SEET to ISA noiibube showledge of the stremerings of the seilede noiles single
OOMITINUE ON A SEPARATE PAGE IF NEEDED
Over heading the Counter Use (St CFR 800 Suppart C)( Presscription Use (Fart 21 CFR 801 Subpart D)Type of Use (Select one or both, as applicable)
ອງອຸປຣເ່ນ ຕາເພ ຂາວຂະໜຸດທີ່ ໄດ້ຮັກສອນ ການຈຳລາດ ຄ.ສ. 2017 ອາການປາ ຄ.ສ. 2017 ປີ 10 ປີ 10 ກາກຕາການຂະໜາອກmellitus.
of its is od your of w streited grivfitmebi ni bis us as bas revedsib att ni bis no se besu ed or 21 test ziriTdeveloping diabetes.
.boold show vasilides bus suous and super in ( of AdH , I A
udolgomed of ) aidolgoment battainment within wo rot tren sites of only ni us si xC of AdH Minoin d
Indications for Use (Describe)
əmsində kənd. Dəniz qalında qalınmışdır. Bu və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir vəxQ JIAdH MTuoiniIA
9620817
(UMOUX (KI) KITADƏ KI
See PRA Statement belowndications for Use
Form Approved: OMB No. 0810-0120xpiration Date: 06/30/2020For and Aministration
2301V932 ИАМИН ОНА НТТАЭН 70 ТИЭМТУААРЭQ

្រី ការប្រព្យ ប្រទេស ក្រុង ក្រុង អ្នក (រួម របស់ ក្រុង ប្រទេស និង ក្រុង ប្រទេស

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Abbott, a healthcare company. The logo features a stylized, sans-serif letter 'a' in a bright blue color. Below the symbol, the company name "Abbott" is written in a bold, black, sans-serif font.

510(k) SUMMARY-K180296

GENERAL INFORMATION

Applicant Name:Alere Technologies ASKjelsaasveien 161PO Box 6863 RodeloekkaNO-0504 OsloNorwayEstablishment #9613069
Company Contact:Annie WrightSenior Manager, Regulatory AffairsPhone: (858) 805-2316Email: annie.wright@alere.com

Date Prepared: April 26, 2018

DEVICE IDENTIFICATION

Trade or Proprietary Names:Afinion™ HbA1c Dx
Common Name:HbA1c test
Device Classification Name:Hemoglobin A1c Test System
Product Codes:PDJ
Regulatory Class:Class II (Special Controls)
Classification Regulation:21 CFR 862.1373
Predicate Device:Cobas C501 Tina-Quant HbA1cDx Gen.3 assay (K121610)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Abbott, a healthcare company. The logo consists of a stylized letter "A" in blue, with the word "Abbott" written in black below it. The "A" is designed with rounded corners and a gap in the middle, giving it a modern and abstract look. The font used for the company name is bold and sans-serif, providing a clean and professional appearance.

DEVICE DESCRIPTION

The Afinion™ HbA1c Dx is a fully automated boronate affinity assay for the determination of the percentage of hemoglobin A1c in human whole blood. The Afinion™ HbA1c Dx is a modification of the existing device, Alere Afinion™ HbA1c for use with the Alere Afinion™ AS100 Analyzer, with the addition of a diagnostic intended use.

The test begins with a blood sample collected with the integrated sampling device before the test cartridge is placed in the cartridge chamber of the Alere Afinion™ AS100 Analyzer. The sample is then automatically diluted and mixed with a solution that releases hemoglobin from the erythrocytes. After the hemoglobin is precipitated, the sample mixture is transferred to a blue boronic acid conjugate which binds to the cis-diols of glycated hemoglobin. This reaction mixture is soaked through a filter membrane and all precipitated hemoglobin, conjugate-bound and unbound (i.e. glycated and non-glycated hemoglobin) remains on the membrane. Excess conjugate is removed with a washing reagent. The analyzer measures the reflectance of the precipitate on the membrane as blue (glycated hemoglobin) and red (total hemoglobin) color intensities. The analyzer calculates a ratio proportional to the percentage of HbA1c in the sample and displays as the % HbA1c (NGSP).

INTENDED USE

Afinion™ HbA1c Dx is an in vitro diagnostic test for quantitative determination of glycated hemoglobin (% hemoglobin A1c, HbA1c) in human venous and capillary whole blood.

This test is to be used as an aid in the diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes.

The measurement of % HbA1c is recommended as a marker of long-term metabolic control in persons with diabetes mellitus.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Abbott logo. The logo consists of a stylized letter 'A' in blue, with the word 'Abbott' written in black below it. The 'A' is made up of three connected lines, forming a rounded, geometric shape.

COMPARISON WITH PREDICATE

AttributePredicate DeviceCandidate Device
Cobas C501 Tina-Quant HbA1cDx Gen.3assay (K121610)Afinion™ HbA1c Dx
Similarities
Intended UseQuantitative determinationof hemoglobin A1c for the diagnosis of diabetesand as an aid in identifying patients who maybe at risk for developing diabetesSame
UserenvironmentFor use by health care professionals in clinicallaboratoriesSame
Sample typesAnticoagulated venous bloodAnticoagulated venous orcapillary blood
AnticoagulantsLi-Heparin, K2-EDTA, K3-EDTA, KF/Na2-EDTA, Na-Heparin, NaF/K-Oxalate andNaF/Na2-EDTAK2-EDTA
Units ofmeasurement% HbA1c (DCCT/NGSP)Same
Storagerequirements2-8°C until expiration dateSame
StandardizationTraceable to the International Federation ofClinical Chemistry and Laboratory Medicine(IFCC) and Diabetes Control andComplications Trial (DCTT) reference method.Certified via the National GlycohemoglobinStandardization Program (NGSP).Same
Expected valuesProtocol 1 (IFCC)20-42 mmol/mol HbA1cProtocol 2 (DCCT/NGSP)4.0 - 6.0 %HbA1cDCCT/NGSP4.0 - 6.0 %HbA1c
Differences
Assay principleHbA1c determination by turbidimetricinhibition immunoassay method.Hemoglobin determination by converting to aderivative with a characteristic absorptionspectrum which is measured bi-chromatically.Boronate affinity assay.Hemoglobin determination byreflectance measurements ofblue conjugate of glycatedhemoglobin and red totalhemoglobin.
HardwareCobas c501Alere Afinion™ AS100Analyzer
CalibratorsProvided by manufacturerCalibration by user notrequired; factory calibrated
Reporting range4.2-20.1 % HbA1c4.00–15.00 % HbA1c

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Abbott, a healthcare company. The logo consists of a stylized letter "a" in blue, with the word "Abbott" written in bold, black letters underneath. The "a" symbol is a rounded, geometric shape. The overall design is simple and modern.

PERFORMANCE CHARACTERISTICS

1. Precision/Reproducibility

  • A. Precision Internal
    The precision of the Afinion™ HbA1c Dx was evaluated based on CLSI guideline EP05-A3. Four levels of HbA1c in anticoagulated venous whole blood patient samples at the following targeted HbA1c values - 5, 6.5, 8 and 12 %HbA1c - were utilized in the study. Three test cartridge lots were included in the study. Two replicates of venous whole blood were tested twice a day for twenty days with each of the three lots on three Alere Afinion™ AS100 Analyzers.
MeanRepeatabilityBetween runBetween dayBetween lotTotal
Level%HbA1cSD%CVSD%CVSD%CVSD%CVSD%CV
Low5.130.0671.310.0170.330.0290.560.0370.720.0841.63
Threshold6.530.0861.310.0000.000.0390.600.0640.980.1141.74
Medium8.090.0821.010.0000.000.0450.560.0190.230.0951.17
High11.330.1000.880.0000.000.0350.310.1050.930.1491.32

Table 1a: Within-laboratory precision - results from analyzer 1 (n=240)

Table 1b: Within-laboratory precision - results from analyzer 2 (n=240)

MeanRepeatabilityBetween runBetween dayBetween lotTotal
Level%HbA1cSD%CVSD%CVSD%CVSD%CVSD%CV
Low5.110.0611.190.0000.000.0340.670.0460.890.0841.63
Threshold6.530.0701.070.0190.300.0440.670.0640.980.1061.62
Medium8.010.0600.750.0100.130.0380.480.0420.520.0831.04
High11.170.0860.770.0150.130.0600.540.0940.840.1421.27

Table 1c: Within-laboratory precision - results from analyzer 3 (n=240)

MeanRepeatabilityBetween runBetween dayBetween lotTotal
Level%HbA1cSD%CVSD%CVSD%CVSD%CVSD%CV
Low5.180.0531.020.0070.140.0300.580.0410.790.0741.43
Threshold6.580.0661.000.0000.000.0380.580.0350.530.0841.27
Medium8.080.0861.060.0350.430.0520.640.0470.580.1161.43
High11.300.1141.000.0230.210.0310.280.0980.870.1551.37

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Abbott Laboratories. The logo consists of a stylized, rounded letter "a" in blue, positioned above the word "Abbott" in a bold, black font. The "a" symbol is simple and modern, while the company name is clear and legible.

Betweenanalyzer
MeanRepeatabilityBetween runBetween dayBetween lotTotal
Level%HbA1cSD%CVSD%CVSD%CVSD%CVSD%CVSD%CV
Low5.140.0611.180.0000.000.0310.600.0410.810.0300.580.0841.64
Threshold6.550.0741.130.0000.000.0400.620.0560.850.0000.000.0991.51
Medium8.060.0770.950.0160.200.0450.560.0380.470.0390.490.1061.31
High11.260.1010.890.0140.120.0440.390.0990.880.0590.520.1601.42

Table 1d: Within-laboratory precision – all analyzers combined (n=720)

B. Precision - External (Venous Whole Blood)

The precision of the Afinion™ HbA1c Dx was evaluated externally. Four levels of HbA1c in anticoagulated venous whole blood patient samples at the following targeted HbA1c values: 5, 6.5, 8 and 12 %HbA1c -were evaluated at each of three study sites. Three test cartridge lots were included in the study. Four replicates of venous whole blood were analyzed twice a day for ten days with each of the three lots at each of the three sites, utilizing three Alere Afinion™ AS100 Analyzers.

Table 2: External precision results from 3 sites combined. Root Mean Squared SD and %CV. Between analyzer precision component from internal study added.

LevelMean%HbA1cRepeatabilityBetween runBetween dayBetween lotBetween analyzerTotal
SD%CVSD%CVSD%CVSD%CVSD%CVSD%CV
Low5.020.0611.210.0290.570.0220.440.0460.920.0300.580.0901.78
Threshold6.410.0711.110.0300.480.0260.400.0470.740.0000.000.0941.48
Medium8.180.0901.100.0190.230.0220.280.0420.510.0390.490.1111.36
High12.090.1180.970.0460.380.0300.250.0330.270.0590.520.1471.22

Table 3: External precision results from 3 sites combined (abbreviated version of Table 5). Root Mean Squared %CV. Between analyzer precision component from internal study added.

External Precision StudyTotal %CV
Repeatability %CVBetween-run %CVBetween-day %CVBetween-lot %CVBetween-analyzer %CV
Low1.210.570.440.920.581.78
Threshold1.110.480.400.740.001.48
Medium1.100.230.280.510.491.36
High0.970.380.250.270.521.22

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Abbott, a healthcare company. The logo features a stylized, rounded letter 'a' in blue, positioned above the company name 'Abbott' in bold, black font. The 'a' symbol is modern and abstract, while the company name is clear and legible.

C. Precision - Fingerstick Samples

In order to estimate precision components for fingerstick samples, data from three studies are combined.

1) Within run precision estimates from accuracy study (Study A)

Duplicate measurements of fingerstick samples from 172 subjects were conducted as part of the external accuracy study. The study was conducted at three sites, using three lots of test cartridges. Two lots were assigned to each of three study sites and the fingerstick measurements were performed with one replicate from each of the two available lots for each subject. The duplicate analyses were grouped into four bins of %HbA1c levels, and the CV for each bin was initially estimated by taking the Root Mean Square (RMS) of the pairwise SD within each bin and dividing by the mean of each bin. This initial CV estimate includes both the within-run and the between-lot components of variance.

A more detailed analysis was also conducted to separate the Repeatability from the Between-Lot component. Two of the 172 subjects used the same cartridge lot for both fingersticks and were considered unevaluable for estimating the between-lot component of variance; these two subjects were therefore excluded from the ANOVA model (as reported in Table 12.1.25-5 and 12.1.25-6 of the Performance Testing - Clinical).

SD from ANOVA%CV
%HbA1cRangeNumberSubjectsMeanRepeatabilityBetween-LotTotalRepeatabilityBetween-LotTotal
4.00-5.99455.420.08230.06500.10491.521.201.93
6.00-6.99686.460.08760.02700.09161.360.421.42
7.00-9.99517.930.10670.00000.10671.350.001.35
≥10610.720.06650.00800.06700.620.070.62

Table 4: Fingerstick components of variance from Accuracy study (study A)

The results shown above for the Total %CV calculated via ANOVA differ slightly (but not significantly) from the initial CV estimate based on the RMS of the pairwise SD within each bin, also noting that the initial CV estimate included the two subjects (who used the same cartridge lot for both fingersticks).

  1. Between run and between day precision from venous whole blood precision study (Study B) Estimates for between run and between day precision on fingerstick samples were derived from the internal precision study on venous whole blood samples described above in section 1A of this summary and were based on the assumption that these components of variation for fingerstick and venous whole blood are essentially the same. Results for between-run and between-day precision for fingerstick samples are therefore taken from the data summarized in Table 4 above for the four levels of HbA1c.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for Abbott. The logo consists of a stylized, rounded square shape in blue, resembling a lowercase letter 'a' or a stylized box. Below the symbol, the word "Abbott" is written in a bold, sans-serif font. The text is black, providing a strong contrast against the white background.

3) Between-instrument and between operator precision estimates (Study C)

Fingerstick precision testing was conducted at three sites, utilizing three Alere Afinion™ AS100 Analyzers. Four levels of HbA1c were assessed, with each site targeting five subjects per level. From each subject three operators collected two fingerstick samples each and

tested them with one replicate on each of two instruments. One of the sites, at the Medium levels enrolled 6 subjects prior to the close of the study. This resulted in six fingerstick sample results from each subject, a total of 15 subjects per level (or 16 subjects per level) and total of 90 fingerstick measurements per level (or 96 fingerstick measurements per level). The analysis of the Medium level was conducted two ways, with first 15 subjects to fill the level (this was the primary analysis) and with all 16 subjects enrolled (this was the secondary analysis). There was no significant difference between the primary and the secondary results; the results of the secondary analysis (2º) of the Medium level are reported in the summary table below.

HbA1cLevelGrandMean%HbA1cN (M)BetweenInstrumentBetweenOperatorWithin-run(Repeatability)Total*
SDCV(%)SDCV(%)SDCV(%)SDCV(%)
Low5.3315 (90)0.00760.140.00000.000.07421.390.07451.40
Threshold6.5115 (90)0.01920.300.00000.000.08111.250.08341.28
Medium (2°)8.3216 (96)0.00000.000.01710.210.09181.100.09341.12
High12.2015 (90)0.05700.470.00000.000.13891.140.15011.23

Table 5: Precision of fingerstick samples - Fingerstick precision study (Study C)

Notes: N (M) is the number of subjects (number of test results). CV is SD from ANOVA divided by Grand Mean. * Total CV is the sum of squares of between instrument, between operator, and within-run (repeatability).

The combined precision estimates from the three studies are shown in Table 6. For the high HbA1c level, the Repeatability CV is taken from the fingerstick precision study (study C) instead of the accuracy study (study A) due to higher number of subjects and replicates in study C. The total CV (%) was calculated from each of the individual components of variation as also summarized in Table 6.

Table 6: Precision of fingerstick samples - results from three studies combined
Fingerstick precision from Accuracy StudyAVenous Whole BloodPrecision Study BFingerstick Precision Study C
%HbA1cRangeNSubjectsRepeatability%CVBetweenLot %CVBetweenRun %CVBetweenDay %CVRepeatability %CVBetweenInstrument%CVBetweenOperator%CV*Total%CV
4.00-5.99451.521.200.000.601.390.140.002.03
6.00-6.99681.360.420.000.621.250.300.001.58
7.00-9.99511.350.000.200.561.100.000.211.49
≥1060.620.070.120.391.140.470.001.30

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for Abbott Laboratories. The logo consists of a stylized letter "a" in blue, positioned above the word "Abbott" in a bold, black font. The "a" symbol is rounded and appears to be a continuous, flowing shape. The overall design is clean and modern.

*The Total CV is based on the sum of squares of the 6 components (Repeatability %CV, Between Lot %CV, Between Run %CV, Between Day %CV, Between Instrument %CV, and Between Operator %CV). For the %HbA1c Range ≥10, the

Repeatability estimate from study C was used to estimate the Total CV; for all other %HbA1c Ranges the Repeatability estimate from study A was used to estimate the Total CV. The number of test results is 2N per level for study A, 720 per level for study B, and 90 (or 96) per level for Study C.

D. Precision - Between -Instrument

Between analyzer precision of the Afinion™ HbA1c Dx was evaluated using 14 Alere Afinion™ AS100 Analyzers. Four venous whole blood samples were measured in six replicates on each analyzer with one test cartridge lot for a total of 84 replicates per sample.

LevelMean %HbA1cBetween analyzer SDBetween analyzer CV (%)
Low5.460.040.7
Low5.470.050.9
Threshold6.540.050.8
Medium8.850.091.0

Table 7: Between-analyzer results

E. Lot-to-lot Variation

The lot-to-lot variation for the Afinion™ HbA1c Dx assay was evaluated using three test cartridge lots and 18 EDTA venous whole blood samples spanning the assay measuring (reportable) range. Each sample was analysed in one replicate with each test cartridge lot on the same Alere Afinion™ AS100 Analyzer. The results from all lots were compared pairwise using Bland-Altman analysis.

Table 8: Lot-to-lot variation results
-----------------------------------------------
ComparisonRelative Bias(%)95% Confidence interval ofRelative Bias (%)95% Limit of Agreement
Lot 3 - Lot 10.08-0.8 to 1.0-3.5 to 3.7
Lot 2 - Lot 1-0.17-1.1 to 0.8-4.0 to 3.6
Lot 3 - Lot 20.25-0.9 to 1.4-4.4 to 4.9

2. Method Comparison

The method comparison study was performed using an NGSP secondary reference laboratory method (Tosoh Glycohemoglobin test on the G8 HPLC analyzer) as the

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image contains the logo for Abbott Laboratories. The logo consists of a stylized, sans-serif letter 'A' in a light blue color, positioned above the word "Abbott" in a bold, black, sans-serif font. The 'A' symbol is designed with rounded edges, giving it a modern and approachable look.

comparator (reference) method. Three production lots of Afinion™ HbA1c Dx test cartridges were evaluated with two lots assigned to each of three study sites. Samples from each study subject were tested with both fingerstick samples and fresh venous EDTA samples in single replicates per lot. A total of 120 subjects were included in the study.

%HbA1c IntervalAlere Afinion™ HbA1c DxNumber of subjectsTotal %
Site 1Site 2Site 3Total
4.0 - 5.022154.2
5.1 - 6.05551512.5
6.1 - 6.51010103025.0
6.6 - 7.0128103025.0
7.1 - 8.07672016.7
8.1 - 9.0343108.3
9.1 - 15.0334108.3
Total423840120100.0

Table 9: Sample distribution

Table 10: Weighted Deming regression – combined results from 3 sites and 3 lots

Weighted DemingRegressionPassing-Bablok Regression
FingersticksamplesVenous wholebloodFingersticksamplesVenous wholeblood
Slope0.9970.9911.0001.000
Intercept0.0000.053-0.040-0.030
Correlation coefficient(r)0.9910.9900.9910.990
Number of samples120120120120
Results range Afinion(%HbA1c)4.66 to 11.584.72 to 11.494.66 to 11.584.72 to 11.49

3. Total Error

For fingerstick samples the results of bias estimation (%Bias) from the accuracy study and the combined precision estimates for fingerstick samples from three studies were used for calculation of the Total Error (TE) at four concentrations (5.0, 6.5, 8.0 and 12.0 %HbA1c). These results are shown in Table 11.

For venous whole blood, the results of bias estimation (%Bias) from the accuracy study and precision estimates in the internal precision study using three lots and three Alere

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the logo for Abbott. The logo consists of a stylized, lowercase letter 'a' in a bright blue color. Below the symbol is the word "Abbott" in a bold, black font.

Afinion™ Analyzers over 20 days, were used for calculation of the Total Error (TE) at four concentrations of HbA1c (5.0, 6.5, 8.0 and 12.0 %). These results are shown in Table 12.

For venous whole blood, the Total Error (TE) was also estimated by a second method, similar those described above, but where the precision estimates are from the external precision study, as described in Table 3, which also includes the Between Analyzer component of variance from the internal study. These results are shown in Table 13, labeled as "Total error - venous whole blood samples at the Point of Care" to distinguish these estimates from those in Table 12.

The acceptable %TE is ≤6.0%. (Refer to Section 12.4 of the Performance Testing – Clinical for the Total Error formula.)

Weighted Deming RegressionPassing-Bablok regression
%HbA1c Level%Bias%CV%TE%Bias%CV%TE
5.0-0.3352.034.30-0.8002.034.75
6.5-0.3341.583.42-0.6151.583.69
8.0-0.3341.493.24-0.5001.493.40
12.0-0.3331.302.87-0.3331.302.87

Table 11: Total error - fingerstick whole blood samples

Table 12: Total error - venous whole blood samples

Weighted Deming RegressionPassing-Bablok regression
%HbA1c Level%Bias%CV%TE%Bias%CV%TE
5.00.1951.643.42-0.6001.643.80
6.5-0.0521.513.01-0.4621.513.41
8.0-0.2061.312.77-0.3751.312.93
12.0-0.4291.423.20-0.2501.423.03

Table 13: Total error - venous whole blood samples at the Point of Care

Weighted Deming RegressionPassing-Bablok regression
%HbA1c Level%Bias%CV%TE%Bias%CV%TE
5.00.1951.783.69-0.6001.784.07
6.5-0.0521.482.95-0.4621.483.35
8.0-0.2061.362.87-0.3751.363.03
12.0-0.4291.222.81-0.2501.222.64

4. Endogenous Interference

Six endogenous substances spiked into whole blood sample pools were evaluated for potential assay interference in accordance with CLSI guideline EP07-A2. Two HbA1c levels, one near the medical decision level and one above it, were tested for each endogenous substance. Each sample was tested for %HbA1c with ten replicates on a single

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image shows the logo for Abbott. The logo consists of a stylized letter 'a' in blue, positioned above the word "Abbott" in black, bold font. The 'a' symbol is a rounded, geometric shape, and the word "Abbott" is in a sans-serif typeface.

Alere Afinion™ AS100 Analyzer, using a single test cartridge lot. The corresponding nonspiked sample which contained no added substance was also tested with ten replicates. A mean was calculated for each set of ten. Significant interference was defined as exceeding a 7% change in %HbA1c value from the control (non-spiked sample). The results for the spiked samples were compared to the control with no interference at the stated concentrations.

Endogenous substanceTest concentration with no significant interference
Bilirubin, conjugated600 mg/L
Bilirubin, unconjugated600 mg/L
Glucose10 000 mg/L
Intralipid10 000 mg/L
Rheumatoid factor780 000 IU/L
Total protein15 g/dL

Table 14: Endogenous interference results

5. Hemolysis Interference

Potential assay interference from hemolysis was assessed in accordance with CLSI guideline EP07-A2. Whole blood samples with two HbA1c levels, one near the medical decision level and one above it, were used for the testing. Hemolysis in the blood samples was simulated by spiking with fresh hemolysate prepared from the native samples. Hemolysate was added to the corresponding native sample to volume percentages of 5, 10, 15 and 20%. The exact degree of hemolysis was measured in each spiked fraction. Each spiked sample and corresponding non-spiked sample was tested in ten replicates on a single Alere Afinion™ AS100 analyzer using one test cartridge lot. The mean result from the manipulated sample was compared to the mean result from the corresponding native control sample. Significant interference was defined as exceeding a 7% change in %HbA1c value from the control (native sample). The results demonstrated no interference up to 24% hemolysis, however at hemolysis above 14% (2000 mg/dL) information codes related to hemolysis may occur with a frequency which increases with increasing degrees of hemolysis.

6. Drug Interference

Twenty drugs spiked into whole blood sample pools were tested for potential assay interference in accordance with CLSI guideline EP7-A2. The drug substances included commonly used antibiotics, analgesic agents, immunosupressants, anti-inflammatory drugs, anticoagulants and prescription drugs for diabetes treatment. Two HbA1c levels, one near the medical decision level and one above it, were tested for each endogenous substance. Each spiked sample was tested for %HbA1c in ten replicates on a single Alere Afinion™ AS100 Analyzer, using a single test cartridge lot. The corresponding non-spiked

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image shows the logo for Abbott, a healthcare company. The logo features a stylized letter 'a' in a light blue color. Below the symbol, the word "Abbott" is written in a bold, black font.

sample which contained no added substance was also tested with ten replicates. A mean was calculated for each set of ten. Significant interference was defined as exceeding a 7% change in %HbA1c value from the control (non-spiked sample). The results for the spiked samples were compared to the control with no interference at the stated concentrations.

Drug substanceTest concentration with no significantinterference (mg/L)
Acetaminophen200
Acetylcysteine1663
Ascorbic acid300
Cefoxitin2500
Metformin40
Methyldopa20
Cyclosporine C5
Glyburide1.9
Metronidazole200
Ampicillin1000
Acetylsalicylic acid1000
Doxycycline50
Ibuprofen500
Cyclosporine A5
Levodopa20
Heparin5000 U/L
Salicylic acid599
Theophylline100
Phenylbutazone400
Rifampicin64

Table 15: Drug interference results

7. Cross reactivity with Hemoglobin Derivatives

Potential assay interference from hemoglobin derivatives was assessed in accordance with CLSI guideline EP07-A2. Whole blood samples with two HbA1c levels, one near the medical decision level and one above it, were used for the testing. For each HbA1c level, two whole blood pools were prepared for each derivative, one without derivative (control) and one with high concentration of derivative. Each sample was tested in ten replicates on a single Alere Afinion™ AS100 Analyzer using one test cartridge lot. A mean was calculated for each set of ten. Significant interference was defined as exceeding a 7% change in %HbA1c value from the control (non-spiked sample). The results for the spiked samples were compared to the control with no interference at the stated concentrations.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for Abbott, a healthcare company. The logo consists of a stylized, lowercase letter "a" in a bright blue color. Below the symbol is the word "Abbott" in a bold, sans-serif font. The overall design is clean and modern.

Hemoglobin derivativeTest concentration with no significantinterference
Acetylated hemoglobin4.6 mg/mL
Carbamylated hemoglobin13.8 mg/mL
Labile HbA1c11.4 mg/mL
Glycated albumin7.7 mg/mL

Table 16: Cross reactivity with hemoglobin derivatives results

Hemoglobin Variants Interference 8.

Potential assay interference from hemoglobin variants was evaluated in accordance with CLSI guideline EP07-A2. A total of 234 fresh EDTA whole blood samples that contained one of six common hemoglobin variants were included in the study. All samples were tested by the Afinion™ HbA1c Dx and with an NGSP reference method that is known to be free from the hemoglobin interference being tested. The level of Hb variant was determined by the Capillarys 2 Flex Piercing method. Interference effect was assessed by comparing the HbA1c results from Afinion HbA1c Dx to the reference method result for samples with potentially interfering hemoglobin variants. Significant interference was defined as exceeding a 7% change in %HbA1c value from the reference method.

Cross-reactivity from HbA0, HbA1a and HbA1b components was also evaluated in a study including 100 samples with characterization of these fractions and comparison to a reference method (Tosoh G8 HPLC).

Table 16 indicates the number, range and concentration of samples and Table 17 summarizes the results from the study.

HemoglobinVariantNumber ofsamplesRange of % content ofvariantRange of concentrationin %HbA1c
HbA010082-95
HbA1a1000.4-1.44.27-13.93
HbA1b1000.4-2.3
HbA2263.9-5.75.8-10.6
HbS2135-425.6-8.8
HbC2530-365.5-9.7
HbE2017-266.1-8.9
HbD2127-426.1-9.4
HbF1213.4-28.15.0-11.3

Table 17: Sample profile for hemoglobin variant study

{16}------------------------------------------------

Image /page/16/Picture/1 description: The image contains the logo for Abbott Laboratories. The logo features a stylized, rounded square shape in blue, positioned above the company name. The word "Abbott" is written in a bold, sans-serif font, with the letters in black.

Percent relative bias from reference method at two levels of the HbA1csamples
HemoglobinVariantLevel 1: ~6.5 %HbA1cLevel 2: ~8.5 %HbA1c
Mean Relativebias (%)Range%Bias^Mean Relativebias (%)Range %Bias^
HbA0
HbA1a0.1-6.1, 8.5-0.4-6.0, 6.7
HbA1b
HbA2-3.4-6.2, -1.3-2.5-6.6, 2.8
HbS-4.1-6.6, 3.2-1.0-9.9, 3.5
HbC-5.5-8.5, -1.8-1.8-5.2, 1.1
HbE3.51.4, 7.93.70.0, 4.7 (14.0*)
HbD-2.2-4.1, 0.0-2.9-5.9, 0.0
HbF10.4% HbF is the highest HbF concentration where no significant interference isobserved with significant interference defined as > +7%. ^^

Table 18: Hemoglobin variant results

^The range is defined as the minimum relative % difference at each concentration level. ^^ A negative relative bias with HbF has increasing magnitude with HbF and also with increasing %HbA1c.

*One single outlier for a sample with HbE level of 22.8%.

No significant interference was observed for the HbA2, HbS, HbC, HbE and HbD. For HbF interference, the device includes the following prominent boxed warning:

This device has significant negative interference with fetal hemoglobin (HbF). HbA1c results are invalid for patients with abnormal amounts of HbF including those with known Heredity Persistence of Fetal Hemoglobin. Refer to the Analytical specificity and Limitations sections in this package insert for details.

9. Linearity

The linearity of the Alere Afinion™ HbA1c assay was previously established in K050574, and was not re-established for the additional indications for use for the Afinion™ HbA1c Dx assay. The reportable range for this device is 4.00-15.00 % HbA1c (DCCT/NGSP).

CONCLUSION

The information and data in this 510(k) application demonstrate that the Afinion™ HbA1c Dx assay is an accurate and precise test that correlates well with current cleared methods and is NGSP standardized for the quantitation of HbA1c. Performance criteria as outlined in the special control requirements of 21 CFR 862.1373 for the diagnosis of diabetes have been met. Based on the criteria stated above, this device is found to be substantially equivalent to the predicate device.

§ 862.1373 Hemoglobin A1c test system.

(a)
Identification. A hemoglobin A1c test system is a device used to measure the percentage concentration of hemoglobin A1c in blood. Measurement of hemoglobin A1c is used as an aid in the diagnosis of diabetes mellitus and as an aid in the identification of patients at risk for developing diabetes mellitus.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA.
(2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity, and interference, including the following:
(i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0 percent, 6.5 percent, 8.0 percent, and 12 percent hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least three lots of the device and three instruments, as applicable.
(ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the device and compare results of the new device to results of a standardized test method. Results must demonstrate little or no bias versus the standardized method.
(iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6 percent.
(iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin S.
(3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.