K Number
K121610

Validate with FDA (Live)

Date Cleared
2013-08-08

(433 days)

Product Code
Regulation Number
862.1373
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This test is to be used as an aid in diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes. The cobas c 501 Tina-quant HbA1cDx Gen.3 assay is an in vitro diagnostics reagent system intended for quantitative determination of mmol/mol hemoglobin Alc (IFCC) and % hemoglobin A1c (DCCT/NGSP) in whole blood on the Roche/Hitachi cobas c 501 clinical chemistry analyzer.

Device Description

Whole blood samples are placed on the analyzer. The anticoagulated whole blood is hemolyzed on board the analyzer prior to determination of HbA1c by this turbidimetric inhibition immunoassay. Liberated hemoglobin in the hemolyzed sample is converted to a derivative having a characteristic absorption spectrum and measured bichromatically. The instrument first measures hemoglobin (Hb) and glycohemoglobin (HbA1c) in terms of either g/dL or mmol/L, then calculates the % HbA1c from the HbA1c/Hb ratio according to a user-selected protocol, either IFCC or NGSP protocols.

AI/ML Overview

The Roche cobas c 501 Tina-quant HbA1cDx Gen.3 assay is an in vitro diagnostic reagent system intended for the quantitative determination of HbA1c.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly state formal acceptance criteria thresholds in terms of specific performance metrics (e.g., "Total Error must be <= X%"). Instead, it presents various analytical performance characteristics and compares them to the predicate device, or indicates "no significant interference." The overall conclusion from Roche is that the device is "substantially equivalent to the predicate device" based on these characteristics.

However, based on the data presented and common analytical performance expectations for such devices, we can infer some criteria and compare the reported performance:

Performance MetricInferred Acceptance Criteria (Implicit)Reported Device Performance
Precision (Total %CV)Generally, low %CV across different HbA1c levels and factors (repeatability, between-run, between-day, lot, instrument). Should be comparable to or better than predicate.Reproducibility (Total %CV): - Human Sample 1 (5.1% HbA1c): 1.9% - Human Sample 2 (6.4% HbA1c): 1.7% - Human Sample 3 (8.0% HbA1c): 2.1% - Human Sample 4 (11.3% HbA1c): 2.0% - PreciControl HbA1c norm (5.2% HbA1c): 2.4% - PreciControl HbA1c path (9.5% HbA1c): 2.0% (Individual analyzer total %CVs are slightly higher in some cases (e.g., 3.0% for PreciControl HbA1c norm on Analyzer 3))
Method Comparison (Bias)Good agreement with NGSP reference method (Tosoh HPLC); low bias across the measuring range.At three decision levels: - 5.2% HbA1c: -1.98% Bias - 6.5% HbA1c: -1.45% Bias - 8.0% HbA1c: -1.06% Bias
Total ErrorTotal Error (TE) should be within clinically acceptable limits at key decision levels.At three decision levels: - 5.2% HbA1c: 6.0% TE - 6.5% HbA1c: 4.7% TE - 8.0% HbA1c: 5.1% TE
Endogenous InterferenceNo significant interference (defined as >7% deviation) at tested physiological or supraphysiological levels.No significant interference observed for: - Lipemia (Intralipid) up to 600 mg/dL - Unconjugated Bilirubin up to 60 mg/dL - Conjugated Bilirubin up to 60 mg/dL - Glucose up to 1000 mg/dL - Rheumatoid Factor up to 750 IU/mL - Total Protein up to 21 g/dL
Drug InterferenceNo significant interference (defined as >7% deviation) at stated therapeutic/supra-therapeutic concentrations.No significant interference observed for 16 common drugs at tested concentrations (e.g., Acetylcystein 150 mg/dL, Ampicillin-Na 1000 mg/dL, Ascorbic acid 300 mg/dL, etc.).
Hb Derivative Cross-ReactivityNo significant cross-reactivity with common hemoglobin derivatives at physiological concentrations.No significant cross-reactivity (defined as >7% deviation) with HbA0, HbA1a+b, Acetylated Hb, Carbamylated Hb, Labile HbA1c, and Glycated Albumin at tested concentrations.
Hb Variant InterferenceLow bias, acceptable performance for common variants; clear identification of known interferences.- HbC, HbS, HbE, HbD: Bias values at 6.0% and 9.0% HbA1c are generally low (e.g., -3.07 to 3.46), well within a typical 7% deviation. - HbA2: Bias values were -5.73% (6.0% HbA1c) and -4.12% (9.0% HbA1c), which are considered within acceptable limits if not exceeding a 7% deviation threshold. - HbF: Explicitly states "Bias exceeds -7% when HbF content exceeds +7%." This is a known interference and is disclosed, indicating that the device performs as expected given this limitation.
Measuring Range / LinearityWide linear range covering clinical needs; demonstrated linearity across the range.Linear Range Summary: - Glycohemoglobin: 0.186-1.61 mmol/L HbA1c (0.30-2.6 g/dL) - Hemoglobin: 2.48-24.8 mmol/L Hb (4-40 g/dL) - Ratio: 4.2-20.1 % HbA1c (DCCT/NGSP), 23-196 mmol/mol HbA1c (IFCC). The study reports that linearity results "support the claimed reportable range."
Detection Limits (LoB/LoD)Low limits, enabling accurate measurement at low concentrations.LoB/LoD for Hb (g/dL): - LoB: 0.0210 - 0.0445 g/dL - LoD: 0.0594 - 0.1133 g/dL LoB/LoD for HbA1c (g/dL): - LoB: 0.0079 - 0.0163 g/dL - LoD: 0.0197 - 0.0315 g/dL

2. Sample Sizes Used for the Test Set and Data Provenance

  • Precision Study:

    • Test Set: Multiple human samples (4 distinct samples covering a range of HbA1c levels) and 2 control samples. Each sample was run in duplicate, twice a day for 21 days on 3 analyzers, using 3 reagent lots. This results in significant numbers of individual measurements per sample (2 samples/day * 21 days * 3 analyzers * 3 reagent lots = 378 data points per sample type, then additional replicates per run).
    • Data Provenance: Not explicitly stated, but typically these types of analytical validation studies are conducted internally by the manufacturer or by contracted clinical research organizations, likely in a controlled laboratory setting. It is prospective data collection for the specific purpose of validating the device.
  • Method Comparison Study:

    • Test Set: 141 variant-free samples.
    • Data Provenance: Not explicitly stated, but "variant-free samples" suggests specific selection, likely from a diverse patient population to cover the reported HbA1c range. This would be prospective data collected for the method comparison.
  • Endogenous Interference Study:

    • Test Set: 6 endogenous substances. For each substance, 2 HbA1c levels were tested, each with a dilution series (varying concentrations of the interferent). Each sample in the dilution series was tested 10-fold.
    • Data Provenance: Whole blood sample pools were spiked with interferents. This is an experimental, prospective design in a laboratory setting.
  • Drug Interference Study:

    • Test Set: 16 drugs. For each drug, 2 HbA1c levels were tested, each at 2 concentrations of the drug. Additionally, a reference sample with no drug. Total of 64 samples prepared. Each sample was tested 10-fold.
    • Data Provenance: Whole blood samples were spiked with drugs. Experimental, prospective design.
  • Hemoglobin Derivatives Cross-Reactivity Study:

    • Test Set: 6 hemoglobin derivatives. For each derivative, 2 HbA1c levels, with dilution series. Each sample tested 10-fold.
    • Data Provenance: Whole blood sample pools. Experimental, prospective design.
  • Hemoglobin Variants Interference Study:

    • Test Set: 116 samples containing one of six common hemoglobin variants (S, C, E, D, F, A2).
    • Data Provenance: Samples categorized by Hb variant and their concentration ranges. Not explicitly stated regarding country or retrospective/prospective, but given the specific variant types, these are likely clinical samples collected from populations with higher prevalence of these variants, and used prospectively for this study.
  • LoB/LoD Study:

    • Test Set: One analyte-free sample (for LoB) and five low-analyte samples (for LoD).
    • Data Provenance: Experimental, prospective design.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • For this type of in vitro diagnostic device (chemical analyzer for HbA1c), the "ground truth" for the test set is established by reference methods and analytical specifications, rather than expert human interpretation of images or clinical cases.
  • Method Comparison: The ground truth for the 141 variant-free samples was established using the Tosoh HPLC, a secondary NGSP reference laboratory method. The NGSP (National Glycohemoglobin Standardization Program) standardizes HbA1c test results to those of the Diabetes Control and Complications Trial (DCCT), involving rigorous certification for laboratories and methods.
  • Hemoglobin Variants Interference Study: Ground truth was established using an NGSP reference method "that is known to be free from the hemoglobin interference being tested." This implies a certified and validated method capable of accurately measuring HbA1c in the presence of specific variants.
  • Other studies (Precision, Interference, LoB/LoD, Linearity): The ground truth for these studies is the inherent analytical characteristic of the sample itself (e.g., true concentration, absence/presence of interferent), measured or prepared precisely according to established laboratory standards and protocols (e.g., CLSI guidelines). No human "experts" are involved in establishing this ground truth value itself, but rather experts in analytical chemistry and laboratory medicine design and execute these studies to rigorous standards.

4. Adjudication Method for the Test Set

  • For an in vitro diagnostic device like this, "adjudication" in the sense of comparing multiple human expert interpretations is not applicable.
  • The "adjudication" is instead the comparison of the device's results against established reference methods or known concentrations/parameters of samples, usually with statistical analyses (e.g., bias calculation in method comparison).
  • For the interference studies, a 7% deviation from an initial value (uninterfered sample) was the threshold for "significant interference." This acts as a predefined decision rule.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for imaging devices or other diagnostic tools that require human interpretation (e.g., radiologists reading images) and where AI is used to assist the human reader.
  • This device is an automated in vitro diagnostic system that produces a quantitative numerical result for HbA1c. There is no human "reader" in the diagnostic interpretation loop in the way there would be for an imaging study.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Yes, the entire study focuses on the standalone performance of the cobas c 501 Tina-quant HbA1cDx Gen.3 assay.
  • All reported performance metrics (precision, method comparison, interference, linearity, detection limits) reflect the analytical performance of the device itself, without any human intervention in interpreting the direct HbA1c result.

7. Type of Ground Truth Used

  • The ground truth used primarily falls under reference method comparison and defined analytical characteristics of known samples:
    • Reference Method Comparison: For method comparison and Hb variant interference, the Tosoh HPLC and other NGSP-certified reference methods served as the ground truth.
    • Defined Analytical Characteristics: For precision, interference (endogenous, drug, Hb derivatives), LoB/LoD, and linearity studies, the ground truth was based on precisely prepared samples with known (or expected) physiological concentrations of analytes and interferents, or the inherent analytical properties of the samples, measured against established analytical standards.

8. Sample Size for the Training Set

  • This submission describes a validation study for a new generation of an already existing assay/device, not the development or training of a de novo AI algorithm.
  • Therefore, the concept of a "training set" in the context of machine learning (where an algorithm learns from labeled data) is not directly applicable here. This is a traditional analytical validation for an in vitro diagnostic assay. The instrument's internal "algorithms" for calculation are based on established chemical principles and calibration, not machine learning from a large, labeled dataset in the modern AI sense.

9. How the Ground Truth for the Training Set was Established

  • As explained in point 8, there isn't a "training set" in the common AI/machine learning sense for this device.
  • The device's operational parameters (e.g., calibration curves, reaction kinetics) would have been established during its development phase using calibrators and control materials traced to international reference standards (e.g., IFCC and DCCT/NGSP), ensuring accuracy and consistency. These are foundational for any in vitro diagnostic test.

{0}------------------------------------------------

K121610

AUG 8 2013

IntroductionRoche Diagnostics hereby submits this 510(k) to provide FDA with notification of intent to market the cobas c 501 Tina-quant HbA1cDx Gen.3 assay with a new intended use that includes the device as an aid in the diagnosis of diabetes and as an aid in identifying patients who may be at risk of developing diabetes. This submission presents data to support this new intended use.
SubmitterSusan Hollandbeck from Roche Diagnostics, U.S. Regulatory Affairs
Date preparedThe submission was originally prepared on May 31, 2012.
Device nameProprietary name:cobas c 501 Tina-quant Hemoglobin A1cDx Gen.3 assayCommon names:HbA1cDx Gen.3 and TQ HbA1cDx Gen.3Classification name:Hemoglobin A1c Test SystemProduct codes:PDJC.F.R. Section:862.1373Proprietary name:cobas c 501 Tina-quant Hemoglobin A1cDx Gen.3 assayCommon names:HbA1cDx Gen.3 and TQ HbA1cDx Gen.3Classification name:Hemoglobin A1c Test SystemProduct codes:PDJC.F.R. Section:862.1373
Proprietary name:cobas c 501 Tina-quant Hemoglobin A1cDx Gen.3 assay
Common names:HbA1cDx Gen.3 and TQ HbA1cDx Gen.3
Classification name:Hemoglobin A1c Test System
Product codes:PDJ
C.F.R. Section:862.1373
Device descriptionWhole blood samples are placed on the analyzer. The anticoagulated whole blood is hemolyzed on board the analyzer prior to determination of HbA1c by this turbidimetric inhibition immunoassay. Liberated hemoglobin in the hemolyzed sample is converted to a derivative having a characteristic absorption spectrum and measured bichromatically. The instrument first measures hemoglobin (Hb) and glycohemoglobin (HbA1c) in terms of either g/dL or mmol/L, then calculates the % HbA1c from the HbA1c/Hb ratio according to a user-selected protocol, either IFCC or NGSP protocols.
Predicate deviceThe cobas c Tina-quant HbA1cDx Gen.3 assay is substantially equivalent to the COBAS INTEGRA 800 Tina-quant HbA1cDx Gen.2 assay that was cleared in 510(k) K121291.

Continued on next page:

{1}------------------------------------------------

Intended useThis test is to be used as an aid in diagnosis of diabetes and as an aid inidentifying patients who may be at risk for developing diabetes. The cobas c501Tina-quant HbA1cDx Gen.3 assay is an in vitro diagnostics reagentsystem intended for quantitative determination of mmol/mol hemoglobin Alc(IFCC) and % hemoglobin A1c (DCCT/NGSP) in whole blood on the
Roche/Hitachi cobas c 501 clinical chemistry analyzer.
Comparison to predicateThe table compares the features of the candidate device, cobas c Tina-quant HbA1cDx Gen.3 assay, to the predicate device, COBAS INTEGRA 800 HbA1cDx Gen.2 that was cleared in 510(k) K121291.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
FeatureCandidate DevicePredicate Device
Anticoagulated venous or capillarybloodSame
Sample TypesAcceptable anticoagulants. Li-Heparin. K2-EDTA. K3-EDTA. KF/Na2-EDTA. Na-Heparin. NaF/K-Oxalate. NaF/Na2-EDTA
Instrument Platformcobas c 501COBAS INTEGRA 800
CalibratorC.f.a.s. HbA1cSame
Calibration FrequencyEach lot, every 29 days, and asrequired following quality controlproceduresSame
Calibration ModeHb determination uses a linearmode.HbAlc determination uses aspline mode.Logit/Log 5
ControlsPreciControl HbA1c norm and pathSame

Comparison Table

{2}------------------------------------------------

Comparison toThe table compares the features of the candidate device, cobas c Tina-
predicate (continued)quant HbA1cDx Gen.3 assay, to the predicate device, COBASINTEGRA 800 HbA1cDx Gen.2 that was cleared in 510(k) K121291.
FeatureCandidate DevicePredicate Device
Reporting UnitsFor the components, Hb and HbA1c:mmol/L and g/dLComponents:g/dL
For the ratio:% HbA1c (DCCT/NGSP) andmmol/mol HbA1c (IFCC)Ratio:Same
Determination of HbA1cHbA1c determination is based on theturbidimetric inhibition immunoassayfor hemolyzed whole blood.Glycohemoglobin in the sample reactswith anti-HbA1c to form solubleantigen-antibody complexes.Polyhaptens react with excess anti-HbA1c to form an insoluble antibody-polyhapten complex which can bemeasured turbidimetrically.Same
Determination of HbLiberated hemoglobin in thehemolyzed sample is converted to aderivative having a characteristicabsorption spectrum which ismeasured bichromatically.Same
Sample PretreatmentAutomated on-board samplepretreatment with hemolyzing reagentSame
Measuring RangeHb2.48 - 24.8 mmol/L(4 - 40 g/dL)Hb4 - 35 g/dL
HbA1c0.186 – 1.61 mmol/L(0.3 - 2.6 g/dL)HbA1c0.3 - 3.4 g/dL
Ratio4.2 - 20.1 % HbA1c23 – 196 mmol/mol HbA1cRatio4.3 - 24.8 % HbA1c23 - 258 mmol/mol HbA1c
Note: This measuring range wasestablished in 510(k) K102914 for theCOBAS INTEGRA Tina-quantHbA1c Gen.3 assay.
FeatureCandidate DevicePredicate Device
AntibodyPolyclonal anti-HbA1c from sheep bloodSame
Reagent StabilityUnopened2-8 °C until expiration date on cobas c pack labelOn-board in useRefrigerated on the analyzer for 4 weeksSame
Analytical SensitivityHbLoB = 0.31 mmol/L (0.50 g/dL)LoD = 0.62 mmol/L (1.00 g/dL)HbLoB = 0.50 g/dLLoD = 1.00 g/dL
HbA1cLoB = 0.12 mmol/L (0.19 g/dL)LoD = 0.18 mmol/L (0.29 g/dL)HbA1cLoB = 0.19 g/dLLoD = 0.29 g/dL
Analytical SpecificityHb fractionsAt physiological concentrations, no cross reactions were found with· HbA0,· HbA1a,· HbA1b,· acetylated hemoglobin,· carbamylated hemoglobin,· glycated albumin, and· labile HbA1c.Same
Hb variantsThis device has significant negative interference with samples containing elevated levels of HbF. The bias exceeds -7% when HbF content exceeds +7%. The negative bias with HbF is independent of % HbA1c, but is directly proportional in magnitude to the % HbF content.HbS, HbC, HbD, HbA2, and HbE do not significantly interfereSame
Comparison toThe table compares the features of the candidate device, cobas c Tina-quant
predicateHbA1cDx Gen.3 assay, to the predicate device, COBAS INTEGRA 800
(continued)HbA1cDx Gen.2 that was cleared in 510(k) K121291.

Comparison Table (continued)

{3}------------------------------------------------

The table compares the features of the candidate device, cobas c Tina-quant Comparison to predicate HbA1cDx Gen.3 assay, to the predicate device, COBAS INTEGRA 800 (continued) HbA1cDx Gen.2 that was cleared in 510(k) K121291.

and the control of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comment

Comparison Table (continued)

{4}------------------------------------------------

Comparison Table (continued)
FeatureCandidate DevicePredicate Device
EndogenousInterferencesIcterusNo significant interference up to 60mg/dL.IcterusSame
LipemiaNo significant interference up to anIntralipid concentration of 600 mg/dL.LipemiaNo significant interference up to anIntralipid concentration of 800 mg/dL.
GlycemiaNo significant interference up to 1000mg/dL.GlycemiaSame
Rheumatoid factorsNo significant interference up to 750IU/mL.Rheumatoid factorsSame
Total ProteinUp to 21 g/dL of additional proteinspiked into the sample does notinterfere.Total ProteinSame
DrugsNo interference was found attherapeutic concentrations using acommon drug panel of 16 drugs.DrugsSame
ExpectedValuesProtocol 1 (IFCC)20 - 42 mmol/mol HbA1cProtocol 2 (DCCT/NGSP)4.0 - 6.0 % HbA1cSame

Comparison Table (continued)

{5}------------------------------------------------

The following discusses the precision of the device. Analytical performance

Precision

Precision was evaluated according to CLSI EP5-A2. It included evaluation of three reagent lots, three cobas c 501 analyzers, four native samples and two control samples, two aliquots per sample run in singlicate, two runs per day for 21 days. Results are in terms of % HbA Ic.

c 501 - Analyzer 1 Precision
MeanRepeatabilityBetween-runBetween-dayBetween-lotTotal
SD%CVSD%CVSD%CVSD%CVSD%CV
Human Sample 1(5.0 %HbA1c)0.0561.10.0220.40.0470.90.0671.30.1022.0
Human Sample 2(6.4 %HbA1c)0.0621.00.0350.50.0510.80.0951.50.1292.0
Human Sample 3(7.9 %HbA1c)0.0781.00.0510.70.0871.10.0530.70.1391.8
Human Sample 4(11.3 %HbA1c)0.1161.00.0000.00.0840.70.2392.10.2782.5
PreciControl HbAIc norm(5.2 %HbAlc)0.0621.20.0340.70.0501.00.0771.50.1152.2
PreciControl HbAlc path(9.4 %HbA1c)0.0850.90.0220.20.0600.60.1771.90.2062.2
c 501 - Analyzer 2 Precision
MeanRepeatabilityBetween-runBetween-dayBetween-lotTotal
SD%CVSD%CVSD%CVSD%CVSD%CV
Human Sample 1(5.1 %HbA1c)0.0541.10.0511.00.0240.50.0280.50.0831.6
Human Sample 2(6.4 %HbA1c)0.0721.10.0550.90.0320.50.0430.70.1051.6
Human Sample 3(8.1 %HbA1c)0.0811.00.0600.70.0831.00.0210.30.1331.6
Human Sample 4(11.4 %HbA1c)0.1070.90.0770.70.0760.70.1751.50.2322.0
PreciControl HbA1c norm(5.2 %HbA1c)0.0651.20.0541.00.0140.30.0290.60.0901.7
PreciControl HbA1c path(9.6 %HbA1c)0.0961.00.0470.50.0380.40.0780.80.1381.4

{6}------------------------------------------------

Analytical performance (continued) The following discusses the precision of the device.

Precision results are in terms of % HbA 1c.
c 501 - Analyzer 3 Precision
RepeatabilityBetween-Between-Between-
MeanrundaylotTotal
%CVSD%CVSD%CVSD%CVSD%CV
Human Sample I(5.0 %HbAlc)0.0621.20.0260.50.0200.40.0751.50.1032.0
Human Sample 2(6.4 %HbAlc)0.0761.20.0210.30.0340:50.0370.60.0941.5
Human Sample 3(8.0 %HbAlc)0.1001.20.0550.70.0270.30.0730.90.1381.7
Human Sample 4(11.3 %HbA1c)0.1121.00.0970.90.0400.40.0360.30.1571.4
PreciControl HbA1c norm(5.2 %HbAlc)0.0761.50.0000.00.0290.60.1332.60.1563.0
PreciControl HbA1c path(9.5 %HbAlc)0.1211.30.0440.50.0000.00.1161.20.1741.8
Reproducibility - Roche/Hitachi cobas c 501
Between-Between-Between-Between-
MeanRepeatabilityrundaylotinstrumentTotal
SD%CVSD%CVSD%CVSD%CVSD%CVSD%CV
Human Sample 1(5.1 %HbAlc)0.0581.10.0360.70.0320.60.0601.20.0000.00.0961.9
Human Sample 2(6.4 %HbA1c)0.0701.10.0400.60.0400.60.0641.00.0000.00.1101.7
Human Sample 3(8.0 %HbAlc)0.0871.10.0560.70.0710.90.0530.70.1001.30.1692.1
Human Sample 4(11.3 %HbAlc)0.1121.00.0670.60.0690.60.1721.50.0000.00.2272.0
PreciControlHbAlc norm(5.2 %HbAIc)0.0681.30.0350.70.0340.70.0901.70.0000.00.1232.4
PreciControlHbAlc path(9.5 %HbAlc)0.1021.10.0390.40.0400.40.1301.40.0790.80.1922.0

{7}------------------------------------------------

Analytical performance The following discusses the method comparison and total error. (continued)

Method Comparison

A method comparison was performed to compare sample results from the candidate method using one reagent lot on one cobas e 501 analyzer to sample results from Tosoh HPLC, the secondary NGSP reference laboratory method. Samples were tested in singlicate and measured over three days. 141 variant-free samples ranged from 4.7 to 12.2% HbA1c. The distribution of samples tested appears below.

Sample Distribution

% HbA1c# sample tested% samples tested
< 5%53.5%
5-6%2114.9%
6-6.5%2819.9%
6.5-7%3323.4%
7-8%2719.1%
8-9%1510.6%
> 9%128.5%
Total141100%

Total Error

The bias component from the method comparison study and the precision component from the reproducibility study are used to calculate the total error at three concentrations near the cutoff.

Total Error

Decision Level (% HbA1c)%Bias%CV%TE
5.2-1.98%2.07%6.0%
6.5-1.45%1.7%4.7%
8.0-1.06%2.1%5.1%

{8}------------------------------------------------

The following discusses the endogenous interference. Analytical performance (continued)

Endogenous Interference

Six endogenous substances were evaluated for potential interference of the assay. These substances were spiked into whole blood sample pools. A separate preparation occurred for each substance. Two HbA1c levels, one near the medical decision level and one above it, were tested for each endogenous substance. Thus twelve dilution series were created.

Each sample in each dilution series was tested ten-fold for % HbA1c on a single cobas c 501 analyzer, using a single reagent lot.

The median value for each set of ten was calculated. The reference sample is the sample Level 0 in the dilution series; it contains no interferent. The initial value is the measured result for the reference sample. The results for all remaining samples in the dilution series are compared to the initial value. This comparison is evaluated as a percent deviation. Interference is significant when it exceeds 7% deviation from the initial value.

Endogenous Interference Summary
endogenous substancerange testedhighest level tested withno significant interference
Lipemia (Intralipid)0-2000 mg/dL600 mg/dL
Unconjugated Bilirubin0-66 mg/dL60mg/dL
Conjugated Bilirubin0-66 mg/dL60 mg/dL
Glucose0-2000 mg/dL1000 mg/dL
Rheumatoid Factor0-1200 IU/mL750 IU/mL
Total Protein0-24.5 g/dL21 g/dL

Endogenous Interference Summary

{9}------------------------------------------------

Analytical performance The following discusses the drug interference. (continued)

Drug Interference

Sixteen drugs were evaluated for potential interference of the assay. These drugs were spiked into whole blood samples at two concentrations, Concentration 1 is ~5 times the maximum daily dose and Concentration 2 is the maximum daily dose. A separate preparation occurred for each drug. Two HbA1c levels, one near the medical decision level and one above it, were tested for each drug. Thus, 64 samples were prepared. Also, an HbA1c sample with no drug served at the reference sample.

Each sample was tested ten-fold for % HbAIc on a single cobas c 501 analyzer, using a single reagent lot.

The median value for each set of ten was calculated and compared to the initial value. Percent recovery was calculated. Interference is significant when it exceeds 7% deviation from the initial value. Results show that no significant interference was observed with the following drugs up to the stated concentrations.

drughighest level tested with no significant interference
Acetylcystein150 mg/dL
Ampicillin-Na1000 mg/dL
Ascorbic acid300 mg/dL
Cefoxitin2500 mg/dL
Heparin5000 U/L
Levodopa20 mg/dL
Methyldopa20 mg/dL
Metronidazole200 mg/dL
Doxycyclin50 mg/dL
Acetylsalicylic Acid1000 mg/dL
Rifampicin60 mg/L
Cyclosporine5 mg/L
Acetaminophen200 mg/L
Ibuprofen500 mg/L
Theophylline100 mg/L
Phenylbutazone400 mg/L

Drug Interference Summary

{10}------------------------------------------------

Analytical performance (continued)

The following discusses cross-reactivity with hemoglobin derivatives.

Cross-Reactivity with Hemoglobin Derivatives

Six hemoglobin derivatives were evaluated for potential interference of the assay. Two HbA1c levels, one near the medical decision level and one above it, were represented in whole blood sample pools. For each HbA1c level, two whole blood pools were prepared for each derivative, one with none and one with a high concentration of derivative. From the pools, a serial dilution was prepared to yield varying concentrations of the derivative for both HbA 1c levels.

Each sample was tested ten-fold for % HbA1c on a single cobas c 501 analyzer, using a single reagent lot.

The median value for each set of ten was calculated. The reference sample is the sample Level 0 in the dilution series; it contains no interferent. The initial value is the measured result for the reference sample. The results for all remaining samples in the dilution series are compared to the initial value. This comparison is evaluated as a percent deviation. Interference is significant when it exceeds 7% deviation from the initial value.

endogenous substancerange testedhighest level tested withno significant interference
HbA00-120 g/dL120 g/dL
HbAla+b0-0.64 g/dL0.48 g/dL
Acetylated Hb0-2.0 g/dL2.0 g/dL
Carbamylated Hb0-1.0 g/dL1.0 g/dL
Labile HbAlc0-100 mg/dL100 mg/dL
Glycated Albumin0-10 g/dL10 g/dL

Cross-Reactivity with Hemoglobin Derivatives Summary

There is no significant cross-reactivity with these hemoglobin derivatives at physiologically occurring concentrations.

{11}------------------------------------------------

Analytical performance The following discusses the interference with hemoglobin variants. (continued)

Hemoglobin Variants Interference

A hemoglobin variant interference study was performed using a total of 116 samples that contain one of six common hemoglobin variants. This table summarizes the sample profile.

HemoglobinQuantityof SamplesRange of% Content of VariantRange ofConcentration in % HbA1c
S2031 - 42% S4.60 - 13.0
C1933 - 44% C4.68 - 13.0
E2027 – 33% E5.00 - 9.68
D2034 – 42% D4.79 - 9.78
F202 – 28% F5.83 - 10.1
A2174 – 7% A24.90 - 8.56
Total116

Representation of Hemoglobin Variants

Testing was performed on the candidate device and with an NGSP reference method that is known to be free from the hemoglobin interference being tested. This table categorizes results for the variants as they are impacted by % HbA1c concentration. The results reflect bias between actual sample results. Interference > 7% deviation from the reference method is significant.

Hb Variant6.0 % HbA1c9.0 % HbA1c
C-3.07-0.35
S2.173.42
E-1.583.46
D-2.303.35
A2-5.73-4.12
FBias exceeds -7%when HbF content exceeds + 7%.¹

Hemoglobin Variant Study Results St

A negative bias with HbF is independent of % HbA1c but is directly proportional in magnitude to the % HbF content.

{12}------------------------------------------------

Analytical performance The following discusses the lower limits of detection and linearity. (continued)

Lower Limits of Detection

LoB (Limit of Blank) and LoD (Limit of Detection) were determined according to CLSI EP 17-A.

LoB was determined using one analyte-free sample tested in five replicates on two cobas c 501 analyzers. LoD was determined using five low-analyte samples tested in singlicate on two cobas c 501 analyzers. Both LoB and LoD were evaluated for Hb and HbA 1c (g/dL). The tests were performed in two runs per day for three days per reagent batch with a total of three reagent batches.

LoB and LoD

Hb (g/dL)HbA1c (g/dL)
Reagent BatchLoBLoDLoBLoD
10.03800.09480.00970.0239
20.02100.05940.00790.0197
30.04450.11330.01630.0315

Linearity

Linearity was performed according to CLSI EP6-A for this submission and for 510(k) K102914. For this submission, a dilution series was prepared using a high analyte concentration sample pool and an analyte-free pool. The pools were mixed in different ratios to vield a 20-level dilution series with varying concentrations of Hb and HbA1c. Values were measured in triplicate for each level. The median values were compared to the theoretical values and regressed.

The linearity results from this study and from the one included in 510(k) K102914 support the claimed reportable range.

Linearity Summary

Unit of MeasureLinear Range
Glycohemoglobinmmol/L HbA1c0.186-1.61
g/dL HbA1c0.30-02.6
Hemoglobinmmol/L Hb2.48-24.8
g/dL Hb4-40
Ratio% HbA1c (DCCT/NGSP)4.2-20.1
mmol/mol HbA1c (IFCC)23-196

Conclusion- based on the performance characteristics stated above, this device is substantially equivalent to the predicate device.

{13}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/13/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Gt Silver Spring, MD 20993-002

August 8, 2013

Roche Diagnostics c/o Ms. Susan Hollandbeck Regulatory Affairs Consultant 9115 Hague Road INDIANAPOLIS IN 46256

Re: K121610

Trade/Device Name: cobas c 501 Tina-quant HbA1cDx Gen.3 Assay Regulation Number: 21 CFR §862.1373 Regulation Name: Hemoglobin A1c Test System Regulatory Class: Class II Product Code: PDJ Dated: July 26, 2013 Received: July 29, 2013

Dear Ms. Hollandbeck:

We have reviewed your Section 510(k) premarket notification of intent to market the device we have roviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enorosaly to regally the enactment date of the Medical Device Amendments, or to conninered prior been reclassified in accordance with the provisions of the Federal Food, Drug, devices that have boon require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Trou may, therefore, marisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{14}------------------------------------------------

Page 2 - Susan Hollandbeck

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and n you dealle openite as not office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket 9 150. Theo, prodot note not 102.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Carol C. Benson -S for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{15}------------------------------------------------

Indications for Use

510(k) Number (if known): K121610

Device Name: cobas c 501 Tina-quant HbA1cDx Gen.3

Indications for Use:

This test is to be used as an aid in diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes. The cobas c 501 Tina-quant HbA1cDx Gen.3 assay is an in vitro diagnostics reagent system intended for quantitative determination of mmol/mol hemoglobin Alc (IFCC) and % hemoglobin Alc (DCCT/NGSP) in whole blood on the Roche/Hitachi cobas c 501 clinical chemistry analyzer.

Prescription Use _X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Katherine Serrano -S

Division Sign-Off

Office of In Vitro Diagnostics and Radiological Health

510(k) K121610

Page 1 of _l

§ 862.1373 Hemoglobin A1c test system.

(a)
Identification. A hemoglobin A1c test system is a device used to measure the percentage concentration of hemoglobin A1c in blood. Measurement of hemoglobin A1c is used as an aid in the diagnosis of diabetes mellitus and as an aid in the identification of patients at risk for developing diabetes mellitus.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA.
(2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity, and interference, including the following:
(i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0 percent, 6.5 percent, 8.0 percent, and 12 percent hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least three lots of the device and three instruments, as applicable.
(ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the device and compare results of the new device to results of a standardized test method. Results must demonstrate little or no bias versus the standardized method.
(iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6 percent.
(iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin S.
(3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.