AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EagleRay Long Sheath and EagleRay Access Catheter are indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

Device Description

The EagleRay Long Sheath and EagleRay Access Catheter, hereinafter referred to as EagleRay Catheters, are 0.014"-0.038" guidewire compatible single lumen catheters that provide access to peripheral, coronary and neuro vasculature which allows insertion of other interventional devices.

The catheters are comprised of a hollow cylindrical tube which is bonded to a standard luer fitting. The wall of the tube is constructed using a combination of metal coils/braids and medical grade polymers. The distal section of each catheter has a hydrophilic coating to enhance tracking through tortuous vasculature. Additionally, an angled distal tip facilitates smoother tracking past vessel branches. The distal tip also has a radiopaque marker to provide the user with clear visual confirmation of the distal tip location under fluoroscopy.

The EagleRay Catheters will be offered with various working lengths and nominal inner diameters (ID) and outer diameters (OD) as shown in Table 1 below.

All EagleRay Catheters are packaged with an accessory Rotating Hemostasis Valve (RHV). The RHV is designed to be attached to the proximal luer of the catheter and helps the user maintain hemostasis.

The 0.088" ID EagleRay Long Sheath is packaged with an additional accessory, a short, retractable Catheter Introducer. This accessory is intended to prevent damage to the EagleRay Long Sheath when advancing it through the hemostasis valve on the access site introducer sheath.

In addition to the supplied accessories discussed above, the adjunctive devices and supplies listed below are intended to be used with the EagleRay Catheters. All of these accessories are also commonly used with the predicate devices.

  • Guidewires
  • Support/Diagnostic Catheters
  • Introducer Sheaths
AI/ML Overview

The provided text is a 510(k) premarket notification summary for a medical device (EagleRay Long Sheath and EagleRay Access Catheter). It describes the device, its intended use, comparison to predicate devices, and various performance tests to demonstrate substantial equivalence.

Based on the information provided, here's a breakdown of the acceptance criteria and the study details:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally phrased as specifications that the device shall meet or be comparable to competitive/predicate devices. The reported performance is summarized as "Pass" for all tests, indicating the device met these criteria.

Test AttributeAcceptance CriteriaReported Device Performance
Functional/Performance Testing (Table 3)
Delivery, Compatibility, and Retraction (Trackability)The catheter shall be able to be delivered, deployed, and retracted per the IFU within a simulated neurological model without incurring any damage to the catheter.Pass (implied, as "test results were reviewed and found to demonstrate that any differences...do not significantly impact...performance parameters...")
Flexibility and Kink ResistanceThere shall be no kinking of shaft (permanent deformation) after simulated use. Also, comparable to competitive products and allow for easily tracking the device. No kinking after wrapping around anatomically relevant bend radii.Pass
Compatibility with other devices (external)The catheters shall be able to be delivered through the minimum introducer sheath or guide catheter size indicated in the product labeling.Pass
Guidewire compatibilityThe catheters shall be able to be delivered over the maximum size guidewire indicated in the product labeling.Pass
Interventional device compatibility (internal)The catheters shall be able to accommodate other interventional devices (e.g., support catheter, diagnostic catheter) up to the maximum size indicated in the product labeling.Pass
Luer compatibilityDevices and accessories shall be compatible with standard syringe luer fittings per ISO 80369-7.Pass (implicitly, as part of "Accessory compatibility")
Accessory compatibilityDevices shall be compatible with catheter introducer and RHV.Pass
Catheter Bond StrengthThe catheter shall have sufficient bond strengths to remain intact throughout a procedure.Pass
Flowrate - positive (forward) pressureThe catheter lumen shall allow for a minimum flowrate comparable to competitive products.Pass
Freedom from Leakage – positive pressureNo liquid leakage from the hub or catheter shaft at 46psi for 30 seconds.Pass
Freedom from Leakage – negative pressureNo air leakage into a 20cc syringe when vacuum pulled for 15 seconds.Pass
Burst PressureCatheter does not burst under pressures that could be seen when performing contrast injections with a standard 10cc syringe.Pass
Catheter Torque StrengthNo separation of any portion of the catheter when rotated at least two (2) full rotations (360 degrees).Pass
PushabilityThe proximal shaft of the catheters shall have sufficient stiffness that the user can easily push the catheter to the target anatomy without buckling.Pass
Coating - ParticulateThe amount of particulate matter that comes off the hydrophilic-coated shaft during simulated use testing shall be determined and compared to competitive products and techniques.Pass
Coating - lubricityCoating must be lubricious and maintain a minimum lubricity over 15 test cycles.Pass
Corrosion ResistanceNo visible corrosion present on devices after saline immersion followed by 30 minutes in boiling water followed by 48 hours in 37°C water bath.Pass
RadiopacityThe radiopaque marker on the catheter can be seen under fluoroscopy during use.Pass
Biocompatibility Testing (Table 4)
Cytotoxicity: ISO MEM* ElutionSample extracts must yield cell lysis grade 2 or lower.Pass, Non-cytotoxic
Cytotoxicity: ISO MTT* AssayThe percentage of cells exhibiting lysis should be similar for all test devices.Pass, No Significant Differences
Sensitization: Magnusson-Kligman MethodTest Group shall yield Grade < 1 score on Magnusson and Kligman scale (provided control Grade < 1).Pass, Non-Sensitizing
Irritation: ISO Intracutaneous Irritation TestThe difference in the mean test article and mean control score must be grade 1.0 or lower.Pass, Non-Irritant
Systemic Toxicity: ISO Materials Mediated Rabbit PyrogenSample Extracts must not cause a total rise in body temperature of ≥0.5°C.Pass, Non-pyrogenic
Systemic Toxicity: ISO Acute Systemic Injection TestSample extracts must not cause: > 10% weight loss in 3 or more test animals, Mortality of 2 or more test animals, Toxic signs such as convulsions and prostration in 2 or more test animals.Pass, Non-Toxic
Hemocompatibility: Complement ActivationThe concentrations of C3a and SC5b-9 in the test devices are statistically similar to the predicate device (Penumbra Neuron Max 6F Catheter) and statistically lower than the positive control for all exposure times.Pass, No Significant Differences, test device had statistically similar or lower concentrations than the predicate and negative controls.
Hemocompatibility: Hemolysis (Direct Contact Method)Sample extracts must be nonhemolytic (≤ 2% hemolytic index).Pass, Non-hemolytic
Hemocompatibility: Hemolysis (Extract Method)Sample extracts must be nonhemolytic (≤ 2% hemolytic index).Pass, Non-hemolytic
Standard In vivo Thrombogenicity, Ovine Jugular NAVI (ISO)The device must have similar or lesser thrombogenic potential after 4 hours in vivo when compared to a control device (Penumbra Neuron Max 6F Catheter).Pass, Minimal Thrombogenic Potential, test device had less thrombogenic potential than the control device.
Sterilization Validation (Table 5)
Positive BI controls must be positiveAll positive controls in each cycle read positive.PASS
Fractional cycle requirementsIPCDs equal or less difficult to sterilize than EPCDs; IPCDs and EPCDs may show growth (EPCD equal or more EtO resistant); Product Sterility testing no growth; Bacteriostasis/Fungistasis testing not inhibitory for growth.PASS
Half cycle requirementsIPCDs must show no growth after 7 days; EPCD are expected to show no growth after 7 days (few positives allowed).PASS
Full cycle requirementsIPCDs must show no growth after 7 days; Selected EPCD must show no growth after 7 days; Samples exposed to 2X sterilization cycle must pass EtO residual testing.PASS
Shelf Life and Packaging (Table 6)
Packaging Visual InspectionN/A (Standard ASTM F1886 applied)Pass
Pouch Integrity Test - Gross Leak DetectionN/A (Standard ASTM F2096 applied)Pass
Pouch Seal Strength – Peel StrengthN/A (Standard ASTM F88 applied)Pass
Label IntegrityN/A (Company specific TM00071 applied)Pass

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document does not specify exact sample sizes (N numbers) for each individual bench or lab test. It states "Bench and Lab (in-vitro) testing were conducted" and "Comparative testing was also performed." For the animal study, it notes "A series of subacute and chronic animal studies were conducted" and "Performed in duplicate" (for Ovine Jugular Thrombogenicity). While the number of animals for the thrombogenicity test is stated as "performed in duplicate," the overall number of animals for the "series" of studies is not explicitly given.
  • Data Provenance:
    • Country of Origin: Not specified. Standard testing guidelines (ISO, ASTM) are referenced, suggesting internationally recognized methods.
    • Retrospective or Prospective: The testing described (bench, lab, animal studies) is inherently prospective in nature, as it involves newly conducted studies to evaluate the new device against defined criteria and predicate devices.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This document describes pre-market testing for a medical device (catheters), not an AI/software device that requires human expert review for "ground truth" to determine performance characteristics (like sensitivity/specificity for disease detection).

For the animal studies, it states: "Use of the test devices resulted in no significant vascular response during these studies and was found to be comparable to the control devices by the study pathologist and the study director." This indicates at least two experts were involved in interpreting the results of the animal studies. Their specific qualifications (e.g., years of experience) are not detailed beyond "study pathologist" and "study director."

4. Adjudication Method for the Test Set

Not applicable in the context of device performance testing. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or for AI ground truth establishment where human readers might disagree on findings. Here, the "ground truth" is based on the physical properties of the device and biological responses in animal models, evaluated through standardized tests and expert interpretation.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This type of study (MRMC) is relevant for evaluating the impact of an AI system on human reader performance (e.g., radiologists interpreting images). This document describes the performance of a physical medical device (catheter) through bench, lab, and animal testing to demonstrate substantial equivalence to predicate devices, not an AI system.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No. This is not an AI device. The described tests are for the standalone performance of the physical catheter itself.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is established through a combination of:

  • Engineering Specifications and Standards: Compliance with international standards (ISO, ASTM) and internal company protocols for various physical and mechanical properties (e.g., flow rate, burst pressure, bond strength, kink resistance).
  • Biocompatibility Testing: In-vitro (cell culture) and in-vivo (animal) studies against established biological endpoints (e.g., cytotoxicity, sensitization, systemic toxicity, hemolysis, thrombogenicity).
  • Comparative Performance to Predicate Devices: Performance attributes (e.g., dimensions, flexibility, lubricity, thrombogenic potential) were evaluated for "comparability" to legally marketed predicate devices, which serve as a benchmark for accepted safe and effective performance.
  • Expert Interpretation: For the animal studies, pathologist and study director opinions contributed to the ground truth regarding biological response.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/machine learning device that requires a "training set." The testing described is verification and validation for a physical medical device.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 8, 2018

Imperative Care Inc. Jake Wolenberg Ouality Assurance and Regulatory Affairs Manager 1221 Innsbruck Drive Sunnyvale, California 94089

Re: K180169

Trade/Device Name: EagleRay Long Sheath, EagleRay Access Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: Julv 3. 2018 Received: July 9, 2018

Dear Jake Wolenberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Xiaolin Zheng -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180169

Device Name

EagleRay Long Sheath, EagleRay Access Catheter

Indications for Use (Describe)

The EagleRay Long Sheath and EagleRay Access Catheter are indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Imperative Care. The logo features an abstract human head made of orange triangles on the left. To the right of the head is the company name, "Imperative," in a bold, sans-serif font. Below "Imperative" is the word "CARE" in a smaller, lighter font.

510(k) Summary EagleRay Long Sheath and EagleRay Access Catheter

A. Submitter Information

Submitter's Name: Address:

  • Telephone: Email: Contact Person: Date of Preparation:

B. Subject Device

Proprietary Name: 510(k) #: Common/Usual Name: Classification Name: Product Code:

C. Primary Predicate Device

  • Primary Proprietary Name: Manufacturer: 510(k) #'s: Common/Usual Name: Classification Name: Product Code:

D. Additional Predicate Devices

Proprietary Name: Manufacturer: 510(k) #: Common/Usual Name: Classification Name: Product Code:

Proprietary Name: Manufacturer: 510(k) #: Common/Usual Name: Classification Name: Product Code:

Imperative Care Inc. 1359 Dell Avenue Campbell, CA 95008 248-496-0198 JWolenberg@ImperativeCare.com Jake Wolenberg July 3, 2018

EagleRay Long Sheath, EagleRay Access Catheter K180169 Guide Sheath & Guide Catheter Catheter, Percutaneous DQY per 21 C.F.R. 870.1250

Neuron Max System Penumbra Inc. K111380 Guide Sheath Catheter, Percutaneous DQY per 21 C.F.R. 870.1250

Arc & Arc Mini Intracranial Support Catheter Micro Therapeutics, Inc. Dba Ev3 Neurovascular K150107 Guide Catheter Catheter, Percutaneous DQY per 21 C.F.R 870.1250

Navien Intracranial Support Catheter Micro Therapeutics, Inc. Dba Ev3 Neurovascular K161152 Guide Catheter Catheter, Percutaneous DQY per 21 C.F.R 870.1250

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Imperative Care. The logo features an abstract human head made of orange triangles on the left. To the right of the head is the text "Imperative" in a bold, sans-serif font, with the word "CARE" in a smaller font below it.

Device Description: E.

The EagleRay Long Sheath and EagleRay Access Catheter, hereinafter referred to as EagleRay Catheters, are 0.014"-0.038" guidewire compatible single lumen catheters that provide access to peripheral, coronary and neuro vasculature which allows insertion of other interventional devices.

The catheters are comprised of a hollow cvlindrical tube which is bonded to a standard luer fitting. The wall of the tube is constructed using a combination of metal coils/braids and medical grade polymers. The distal section of each catheter has a hydrophilic coating to enhance tracking through tortuous vasculature. Additionally, an angled distal tip facilitates smoother tracking past vessel branches. The distal tip also has a radiopaque marker to provide the user with clear visual confirmation of the distal tip location under fluoroscopy.

The EagleRay Catheters will be offered with various working lengths and nominal inner diameters (ID) and outer diameters (OD) as shown in Table 1 below.

CommercialModelDistal DiameterProximal DiameterNominal Usable
NameNumberInnerOuterInnerOuterCatheter Length
EagleRayLong SheathICLS0880800.088"0.106"0.088"0.108"80cm
ICLS08809090cm
ICLS088100100 cm
ICLS088110110 cm
EagleRayAccessCatheterICAC0711370.071"0.083"0.071"0.083"137cm
ICAC0551370.055"0.069"0.067"0.080"137cm
ICAC0451440.045"0.060"0.064"0.080"144cm
ICAC0351580.035"0.051"0.047"0.060"158cm

Table 1: EagleRay Catheter Device Sizes

All EagleRay Catheters are packaged with an accessory Rotating Hemostasis Valve (RHV). The RHV is designed to be attached to the proximal luer of the catheter and helps the user maintain hemostasis.

The 0.088" ID EagleRay Long Sheath is packaged with an additional accessory, a short, retractable Catheter Introducer. This accessory is intended to prevent damage to the EagleRay Long Sheath when advancing it through the hemostasis valve on the access site introducer sheath.

In addition to the supplied accessories discussed above, the adjunctive devices and supplies listed below are intended to be used with the EagleRay Catheters. All of these accessories are also commonly used with the predicate devices.

  • Guidewires
  • Support/Diagnostic Catheters ●
  • Introducer Sheaths ●

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Imperative Care. The logo features a stylized head made of orange and yellow triangles on the left. To the right of the head is the text "Imperative" in a bold, dark gray font, with the word "CARE" in a smaller, lighter orange font underneath.

F. Indications for Use:

The EagleRay Long Sheath and EagleRay Access Catheter are indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

G. Predicate Comparison:

The primary predicate device for the EagleRay Catheters is the Neuron Max System (Penumbra, Inc) cleared under K111380. The Neuron Max System was selected as the primary predicate for the EagleRay Catheters as the Neuron Max System has similar dimensions to the largest device sizes for the EagleRay Catheters and has the same intended use.

In addition to the primary predicate device, the devices in the Micro Therapeutics Intracranial Support Catheter product line, which include the Arc and Arc Mini catheters cleared under K150107 and the Navien catheters cleared under K161152, were used as predicate devices for the smaller EagleRay Catheter device sizes as they have similar dimensions and the same intended use.

Table 2 presented below provides a comparison of the similarities and differences between the proposed EagleRay Catheters, the predicate Neuron Max System, and the predicate Micro Therapeutics Intracranial Support Catheter product line. Bench and lab testing was conducted as appropriate to evaluate any differences that were identified.

The comparison between the subject and predicate devices found that the EagleRay Catheters are substantially equivalent to the predicate Neuron Max System and Micro Therapeutics Intracranial Support Catheters, and that there are no new safety or efficacy concerns. This conclusion is based on all devices sharing the same intended use, basic technology characteristics, and performance characteristics, as demonstrated through well-designed bench and lab testing.

Device AttributeEagleRay Catheters(subject device)PenumbraNeuron Max SystemMicro TherapeuticsIntracranial SupportCatheters
FDA ProductClassificationClass II, DQY, 21 CFR870.1250SameSame
Intended UseIntended for use insupporting the introductionof interventional devices intothe vasculature.SameSame
ConditionSuppliedSterile and Single UseSameSame
SterilizationMethodEthylene Oxide (EtO), SAL10-6SameSame
Device AttributeEagleRay Catheters(subject device)PenumbraNeuron Max SystemMicro TherapeuticsIntracranial SupportCatheters
Inner Diameter(Distal)0.035" – 0.088"0.088"0.035" – 0.072"
Outer Diameter(Distal)0.051" – 0.106"0.105"0.044" – 0.084"
Inner Diameter(Proximal)0.047" – 0.088"0.088"0.044" – 0.072"
Outer Diameter(Proximal)0.060" – 0.110"0.110"0.060" – 0.084"
Effective Length80 – 158cm80 – 90cm90 – 163cm
Tip DesignBeveled edge, soft, flexible, and atraumaticSquare edge, soft, flexible, and atraumaticSquare edge, soft, flexible, and atraumatic
CoatingSurmodics Hydrophilic coatingSurmodics Hydrophilic coatingHydrophilic coating
MaterialsCommonly used medical grade plastics & metals with hydrophilic coating.Commonly used medical grade plastics & metals with hydrophilic coating.PTFE lined polymeric catheter shaft with hydrophilic coating, nitinol support wire, and platinum marker bands.
PackagedAccessoriesCatheter IntroducerRotating Hemostasis Valve(RHV)DilatorRotating HemostasisValve (RHV)Hemostasis ValveCatheter Introducer
PackagingConfigurationThe catheters are placed in a protective polyethylene tube and then mounted, along with the accessories, onto a polyethylene packaging card.The packaging card is inserted into a Tyvek® pouch which is then sealed.The sealed pouch and IFUSameSame

Table 2: EagleRav Catheters, Neuron Max System, and Micro Therapeutics Intracranial Support Catheter Comparison

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Imperative Care. The logo features a stylized human head made up of orange triangles on the left. To the right of the head is the word "Imperative" in black, with the word "CARE" in orange underneath.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Imperative Care. The logo features a geometric head made of orange triangles on the left. To the right of the head is the text "Imperative" in a bold, dark gray font, with the word "CARE" in a smaller, light orange font underneath.

Performance Data Supporting Substantial Equivalence: H.

Bench and Lab (in-vitro) testing were conducted in order to evaluate the similarities and differences between the largest diameter EagleRay Catheters and the predicate Neuron Max System. Comparative testing was also performed to evaluate the similarities and differences between the smallest EagleRay Catheters and the predicate Micro Therapeutics Intracranial Support Catheters.

The test results were reviewed and found to demonstrate that any differences between the proposed and predicate devices do not significantly impact any catheter performance parameters that would affect the safety or efficacy of the subject EagleRay Catheters.

A summary of the tests and performance specifications that were evaluated is presented in Table 3. These tests were performed per company approved protocols and test methods based primarily on catheter performance standard ISO 10555-1.

Test AttributeSpecification
Delivery, Compatibility,and Retraction(Trackability)The catheter shall be able to be delivered, deployed, and retractedper the IFU within a simulated neurological model withoutincurring any damage to the catheter.
Flexibility and KinkResistanceThere shall be no kinking of shaft (permanent deformation) aftersimulated use.
Compatibility with otherdevices (external)The catheters shall be able to be delivered through the minimumintroducer sheath or guide catheter size indicated in the productlabeling.
Guidewire compatibilityThe catheters shall be able to be delivered over the maximum sizeguidewire indicated in the product labeling.
Interventional devicecompatibility (internal)The catheters shall be able to accommodate other interventionaldevices (e.g., support catheter, diagnostic catheter) up to themaximum size indicated in the product labeling.
Luer compatibilityDevices and accessories shall be compatible with standard syringeluer fittings per ISO 80369-7.
Accessory compatibilityDevices shall be compatible with catheter introducer and RHV.
Catheter Bond StrengthThe catheter shall have sufficient bond strengths to remain intactthroughout a procedure.
Table 3: Tests and Performance Specifications
Test AttributeSpecification
Flowrate - positive(forward) pressureThe catheter lumen shall allow for a minimum flowratecomparable to competitive products.
Freedom from Leakage– positive pressureNo liquid leakage from the hub or catheter shaft at 46psi for 30seconds.
Freedom from Leakage– negative pressureNo air leakage into a 20cc syringe when vacuum pulled for 15seconds.
Burst PressureCatheter does not burst under pressures that could be seen whenperforming contrast injections with a standard 10cc syringe.
Catheter TorqueStrengthNo separation of any portion of the catheter when rotated at leasttwo (2) full rotations (360 degrees).
Kink ResistanceThere shall be no kinking of the catheter shaft (permanentdeformation) after wrapping around anatomically relevant bendradii.
FlexibilityThe flexibility of the catheter tip shall be comparable tocompetitive products and allow for easily tracking the device tothe desired target anatomy.
PushabilityThe proximal shaft of the catheters shall have sufficient stiffnessthat the user can easily push the catheter to the target anatomywithout buckling.
Coating - ParticulateThe amount of particulate matter that comes off the hydrophilic-coated shaft during simulated use testing shall be determined andcompared to competitive products and techniques.
Coating - lubricityCoating must be lubricious and maintain a minimum lubricity over15 test cycles.
Corrosion ResistanceNo visible corrosion present on devices after saline immersionfollowed by 30 minutes in boiling water followed by 48 hours in37°C water bath.
RadiopacityThe radiopaque marker on the catheter can be seen underfluoroscopy during use.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Imperative Care. The logo features a stylized head made of orange triangles on the left. To the right of the head is the word "Imperative" in a bold, sans-serif font, with the word "CARE" in a smaller font underneath.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Imperative Care. The logo features a stylized human head made up of orange triangles on the left side. To the right of the head is the word "Imperative" in a bold, sans-serif font, with the word "CARE" in a smaller font underneath.

I. Biocompatibility Testing:

Biocompatibility testing was performed on all direct patient contacting components of the EagleRay Catheters. This testing was conducted to ensure that the components and raw materials used in the proposed EagleRay Catheters, as well as the manufacturing processes and sterilization processes result in a biocompatible product. All biocompatibility tests were conducted pursuant to 21 CFR Part 58, Good Laboratory Practices, ISO 10993-1, and per the recommendations in FDA guidance document titled: Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".

As the EagleRay Catheters are in direct contact with circulating blood and the average length of an endovascular procedure is typically less than 2 hours, the devices were evaluated for the tests recommended for the patient contact category: External communicating device, blood path direct, limited contact duration.

A summary of the selected biocompatibility tests performed for the EagleRay Catheters is presented for reference below and on the following pages in Table 4. All test results passed, indicating that the EagleRay Catheters are biocompatible for their intended use.

TestTest MethodExtractionMethods/ConditionsAcceptance CriteriaResults
Cytotoxicity: ISOMEM* ElutionISO 10993-5Test device extracted inMEM with 5% serum at 37 ±1°C for 24 ± 2 hoursSample extracts mustyield cell lysis grade 2 orlower.Pass, Non-cytotoxic
Cytotoxicity: ISOMTT* AssayISO 10993-5Test device extracted inMTT* Assay Media at 37 ±1°C for 24 ± 2 hoursThe percentage of cellsexhibiting lysis should besimilar for all test devices.Pass, NoSignificantDifferences
Sensitization:Magnusson-Kligman Method, 2extractsISO 10993-10Guinea pigs exposed to testdevice extracts. Challengedsites observed for skinsensitization 24 ± 2 and 48 ±2 hours after removal ofextracts.Extracts were prepared at 50± 2°C for 72 ± 2 hours.Test Group shall yieldGrade < 1 score onMagnusson and Kligmanscale (provided controlGrade < 1).Pass, Non-Sensitizing
TestTest MethodExtractionMethods/ConditionsAcceptance CriteriaResults
Irritation: ISOIntracutaneousIrritation TestISO 10993-10Rabbits are injected withextracts of test device.Injection sitesexamined/scored at 24 ± 2,48 ± 2 and 72 ± 2 hours afterinjections.The difference in themean test article andmean control score mustbe grade 1.0 or lower.Pass, Non-Irritant
Extracts were prepared at 50± 2°C for 72 ± 2 hours.
Systemic Toxicity:ISO MaterialsMediated RabbitPyrogenISO 10993-11:2006Test device extracted in0.9% NS* and injected in themarginal ear vein. Rectaltemperatures recorded priorto injection and every 30 minuntil 1-3 hours post injection.Extracts were prepared at 50± 2°C for 72 ± 2 hoursSample Extracts must notcause a total rise in bodytemperature of ≥0.5°C.Pass, Non-pyrogenic
Systemic Toxicity:ISO AcuteSystemic InjectionTestISO 10993-11Mice are injected withextracts of test device.Animals are observed forsigns of toxicity immediatelyafter injection, 4 ± 0.75, 24 ±2, 48 ± 2 and 72 ± 2 hoursafter injections. Bodyweights are measured priorto injection and on all 3 days.Extracts were prepared in0.9% NS* and Sesame Oil at50 ± 2°C for 72 ± 2 hours.Sample extracts must notcause the following:• > 10% weight loss in 3or more test animals• Mortality of 2 or moretest animals• Toxic signs such asconvulsions andprostration in 2 or moretest animalsPass, Non-Toxic
Hemocompatibility:ComplementActivationISO 10993-4ASTM F756Preformed in vitro using aprepared EnzymeImmunoassay (EIA) kit.This kit will detect the presence ofspecific complementenzymes.Test device incubated inserum at 37± 2°C and testedfor C3a and SC5b at 0.5, 1,and 1.5 hours.The concentrations of C3aand SC5b-9 in the testdevices are statisticallysimilar to the predicatedevice (Penumbra NeuronMax 6F Catheter) andstatistically lower than thepositive control for allexposure times.Pass, NoSignificantDifferencesThe test device hadstatistically similaror lowerconcentrations thanthe predicate andnegative controls.
TestTest MethodExtractionMethods/ConditionsAcceptance CriteriaResults
Hemocompatibility:Hemolysis (DirectContact Method)ASTM F 756ISO 10993-4Test device extracted inPBS* and diluted blood at 37$\pm$ 2°C for 3 $\pm$ 5 minutes.Test device exposed to bloodcell suspension.% hemolysis is measured.Sample extracts must benonhemolytic ( $\leq$ 2%hemolytic index).Pass, Non-hemolytic
Hemocompatibility:Hemolysis (ExtractMethod)ASTM F 756ISO 10993-4Test device extracted in PBSat 50 $\pm$ 2°C for 72 $\pm$ 2 hours.Extract exposed to blood cellsuspension. % hemolysis ismeasured.Sample extracts must benonhemolytic ( $\leq$ 2%hemolytic index).Pass, Non-hemolytic
Standard In vivoThrombogenicity,Ovine JugularNAVI (ISO)ISO 10993-4:2002Performed in duplicate. Testand control device are placedin alternate sides of thejuggler vein. The devices areimplanted and remains inplace for 4 $\pm$ 0.5 hours.The device must havesimilar or lesserthrombogenic potentialafter 4 hours in vivo whencompared to a controldevice (Penumbra NeuronMax 6F Catheter).Pass, MinimalThrombogenicPotentialThe test device hadless thrombogenicpotential than thecontrol device.

Table 4: Biocompatibility Test Summary

{10}------------------------------------------------

EagleRay Long Sheath and
EagleRay Access Catheter
Traditional 510(k) K180169

Image /page/10/Picture/1 description: The image shows the logo for Imperative Care. The logo features a geometric human head in shades of orange and yellow on the left. To the right of the head is the word "Imperative" in a bold, sans-serif font, with the word "CARE" in a smaller font underneath.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for Imperative Care. On the left side of the logo is a geometric design of a human head in orange tones. To the right of the head is the text "Imperative" in a bold, sans-serif font. Below "Imperative" is the word "CARE" in a smaller, sans-serif font, also in orange.

*MEM=Minimal Essential Media,

MTT Assay = (3-(4,5-Dimethylthiazol-2-yl)-2,5-Diphenyltetrazolium Bromide)

NS = Normal Saline

PBS=Phosphate Buffered Saline

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the logo for Imperative Care. On the left side of the logo is an orange geometric design that resembles a human head. To the right of the head is the text "Imperative" in a bold, dark gray font. Below "Imperative" is the word "CARE" in a smaller, orange font.

J. Sterilization:

The EagleRay Catheters are sterilized using a validated EtO process with a sterility assurance level of 1x10-6. The sterilization process was validated per the overkill method described recognized consensus standard ISO 11135, "Sterilization Of Health-Care Products - Ethylene Oxide - Requirements For The Development, Validation And Routine Control Of A Sterilization Process For Medical Devices." A summary of the completed testing is presented below in Table 5.

Table 5: Sterilization Validation Summary

Requirement/Acceptance CriteriaResultsSummary
Positive BI controls must be positive infractional, half and full cycles.PASSAll positive controls in each cycle readpositive.
Fractional cycle requirements:1) IPCDs must be equal or less difficult tosterilize than the EPCDs2) The IPCDs and EPCDs may show growthafter 7 days with the EPCD required to beequal or more EtO resistant.3) Product Sterility testing must show nogrowth in all samples.4) Bacteriostasis/Fungistasis testing mustshow that the product is not inhibitory forgrowth.PASS1) IPCDs showed growth after 7 days.2) EPCDs were equal or more EtOresistant than the IPCDs.3) The devices did not show anygrowth when tested per the methodsin ISO 11737-2.4) Bacteriostasis/Fungistasis Testingshowed product is not inhibitory togrowth when tested per the methodsin ISO 11737-2.
Half cycle requirements:1) The IPCDs must show no growth after 7days2) EPCD are expected to show no growthafter 7 days. A few positives are allowedby ISO 11135.PASS1) IPCDs showed no growth after 7days.2) EPCDs showed no growth after 7days.
Full cycle requirements:1) The IPCDs must show no growth after 7days2) Selected EPCD must show no growth after7 days.3) Samples exposed to 2X sterilization cyclemust pass EtO residual testing.PASS1) IPCDs showed no growth after 7days.2) EPCDs showed no growth after 7days.3) Samples passed EtO residual testingat 0 hour aeration time point.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image contains the logo for Imperative Care. On the left side of the logo is a graphic of a head made up of orange triangles. To the right of the graphic is the text "Imperative" in a bold, dark gray font. Below the word "Imperative" is the word "CARE" in a smaller, light orange font.

K. Shelf Life and Packaging:

Accelerated aging testing based on ASTM F1980 was conducted to verify device, accessory, and packaging performance. A real time aging equivalent of 13 months was used for this testing and will allow for labeling of product with a 1-year shelf life.

Device and accessory performance was verified by repeating the functional tests previously presented in Section H of this 510(k) summary.

Packaging and sterile barrier integrity through transportation has been verified for both the packaging configurations used for the EagleRay Catheters, a straight packaging configuration and a circular packaging configuration. Aging testing has also been performed that supports the sterile barrier integrity following 13 months of accelerated aging. A summary of the completed packaging test is presented below in Table 6.

TestTest MethodStraight PackagingResults(Pass/Fail)Circular PackagingResults(Pass/Fail)
Packaging VisualInspectionASTM F1886Imperative CareTM00071PassPass
Pouch Integrity Test -Gross Leak DetectionASTM F2096PassPass
Pouch Seal Strength– Peel StrengthASTM F88PassPass
Label IntegrityImperative CareTM00071PassPass

Table 6: Packaging Validation Summary

L. Animal Studies

A series of subacute and chronic animal studies were conducted to evaluate the safe use of the EagleRay Catheter design in a porcine model. 510(k) cleared neurovascular catheters, including the predicate Penumbra Neuron Max System, were used as control devices. The studies concluded that:

  • . There was no angiographic evidence of vessel injury in any of the treated vessel segments (test or control).
  • . No significant histological findings were found in relation to use of the test or control devices.
  • Use of the test devices resulted in no significant vascular response during these studies and was found to be comparable to the control devices by the study pathologist and the study director.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image contains the logo for Imperative Care. On the left side of the logo is a geometric design of a human head, with the face made up of many small triangles in shades of orange and yellow. To the right of the head is the word "Imperative" in a bold, sans-serif font, with the word "CARE" in a smaller font underneath.

EagleRay Long Sheath and EagleRay Access Catheter Traditional 510(k) K180169

M. Conclusions:

Where differences were identified between the subject EagleRay Catheters, the predicate Neuron Max System, and the predicate Micro Therapeutics Intracranial Support Catheters, a risk assessment was completed to determine if the difference would result in new safety or efficacy concerns regarding the use of the device. As appropriate, bench, lab, and animal testing was conducted to support this assessment.

Based on the results of the conducted risk assessments and associated bench, lab, and animal testing, it is concluded that EagleRay Catheters are substantially equivalent to the predicate Neuron Max System and predicate Micro Therapeutics Intracranial Support Catheters and that there are no new safety or efficacy concerns associated with the identified differences. This conclusion was determined based on the fact that the subject device and multiple predicate devices share the same intended use, basic technology characteristics, and performance characteristics, as confirmed through well-designed bench, lab, and animal testing.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).