K Number
K173869

Validate with FDA (Live)

Date Cleared
2018-08-02

(225 days)

Product Code
Regulation Number
870.1120
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Med-link Reusable Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

Device Description

The proposed device, Med-link Reusable Blood Pressure Cuffs is rectangle soft inelastic sleeve with bladder, to which a single-tube or twin-tube is connected. It is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system. It is available in various sizes for different limb range. It is provided non-sterile.

AI/ML Overview

The Med-link Reusable Blood Pressure Cuff underwent non-clinical testing to assess its safety and essential performance. The acceptance criteria and reported device performance are based on compliance with specific ISO and IEC standards.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (based on standards)Reported Device Performance
Compliance with ISO 81060-1 (non-invasive sphygmomanometers)Device tested and found to be in accordance with ISO 81060-1.
Compliance with IEC 80601-2-30 (automated non-invasive sphygmomanometers)Device tested and found to be in accordance with IEC 80601-2-30.
Compliance with ISO 10993-5 (in vitro cytotoxicity)Cytotoxicity Test conducted and passed.
Compliance with ISO 10993-10 (irritation and skin sensitization)Skin Irritation Test and Skin Sensitization Test conducted and passed.
Pressure Limit0-300mmHg (matches predicate devices).
SterilityNon-sterile (consistent with intended use and predicate devices).
Conformance to AHA bladder sizes recommendationsThe device offers various sizes for neonatal, infant, pediatric, small adult, adult, large adult, and adult thigh, corresponding to recommended limb ranges.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the sample sizes used for each specific non-clinical test (ISO 81060-1, IEC 80601-2-30, biocompatibility tests). It indicates that "A series of safety, essential performance and biocompatibility tests were performed."
The provenance of the data is not specified in terms of country of origin or whether the tests were retrospective or prospective. It states the tests were "conducted in accordance with" the listed standards, implying a controlled testing environment.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the document. The non-clinical testing primarily focuses on objective measurements against international standards rather than expert-derived ground truth as might be seen in diagnostic imaging studies.

4. Adjudication Method for the Test Set:

This information is not applicable and is not provided. The testing involves standardized laboratory and performance tests, not a judgmental adjudication process.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

This information is not applicable and is not provided. The device is a blood pressure cuff, not an AI-assisted diagnostic tool for which MRMC studies would typically be conducted.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done:

This information is not applicable and is not provided. The device is not an algorithm, so standalone algorithm performance is not relevant. The performance of the cuff is evaluated as a physical accessory for blood pressure measurement systems.

7. The Type of Ground Truth Used:

The ground truth for the performance claims relies on:

  • International Standards: Compliance with ISO 81060-1, IEC 80601-2-30, ISO 10993-5, and ISO 10993-10. These standards establish the acceptable technical performance and safety criteria for such devices.
  • Biocompatibility Testing: Objective laboratory tests (cytotoxicity, skin irritation, skin sensitization) directly measure the device's interaction with biological systems against established thresholds.
  • Predicate Device Comparison: The inherent assumption is that if the device performs equivalently to legally marketed predicate devices that have already met regulatory requirements, then it is safe and effective.

8. The Sample Size for the Training Set:

This information is not applicable to this type of medical device. Blood pressure cuffs are mechanical devices, and their performance is established through physical and biological testing, not machine learning models that require training sets.

9. How the Ground Truth for the Training Set was Established:

This information is not applicable for the reasons stated in item 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Shenzhen Med-link Electronics Tech Co., Ltd. Baihan Feng Regulatory Affairs Specialist 4th Floor, Building A Yingtailong Industrial Park Dalang South Road, Longhua District Shenzhen, China 518109

Re: K173869

Trade/Device Name: Med-link Reusable Blood Pressure Cuff Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXQ Dated: June 25, 2018 Received: June 25, 2018

Dear Baihan Feng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

M.A. Hillemann

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K173869

Device Name

Med-link Reusable Blood Pressure Cuff

Indications for Use (Describe)

Med-link Reusable Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for MEDLINKET. The logo has the word "MED" in red, with a blue arc above it. There is a red circle to the right of the word "MED", and the word "LINKET" in blue to the right of the circle. The letters "TM" are in small black font above the letters "KET".

510(k) Summary

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92. Type of submission: Traditional

The assigned 510(k) number is:_

1. Submitter information

Manufacturer Name: Shenzhen Med-link Electronics Tech Co., Ltd. Address: 4th Fl, Bldg A, Yingtailong Industrial Park, Dalang South Road, Longhua District, Shenzhen, Guangdong, CHINA, 518109 Tel: 0086-755-61120299 Fax:0086-755-61120055 Establishment Registration Number: 3006636961

2. Contact Person

Baihan Feng (Regulatory Affairs Specialist) E-mail: mdl001@medlinket.com

3. Data of Submission

15th, Dec. 2017

Identification of the Device 4.

Trade Name: Med-link Reusable Blood Pressure Cuff

Table 1 Device Model
Device Name/TypeDevice Model
Reusable BloodPressure CuffY000N1, Y000N2, Y000I1, Y000I2, Y000C1, Y000C2, Y000S1, Y000S2, Y000A1, Y000A2, Y000L1, Y000L2, Y000T1, Y000T2, Y000RC1, Y000RC2, Y000RSA1, Y000RSA2, Y000RA1, Y000RA2, Y000RAL1, Y000RAL2, Y000RLA1, Y000RLA2, Y000RT1, Y000RT2

Common Name: Non-invasive Blood Pressure Cuff

Classification Name: Blood Pressure Cuff

Classification Regulation: 21 CFR 870.1120

Product Code: DXQ

  • న. Identification of the Predicate Device

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains a logo for MedLinket. The logo has the word "MED" in red, with a blue arc underneath it. There is a red circle to the right of "MED", and the word "LINKET" in blue to the right of the circle. There is a "TM" symbol in the upper right corner of the word "LINKET".

No.Device NameCommon NameManufactureClassification and CodeClassification regulation510(k) number
Primary PredicateDisposable/Reusable Blood Pressure CuffNon-invasive Blood pressure cuffXUZHOU MAICUFFTECHNOLOGYCO.,LTD.Class II, DXQ21 CFR 870.1120K151290
Secondary PredicateUnimed Blood Pressure CuffNon-invasive Blood pressure cuffUnimed Medical supplies Inc.Class II, DXQ21 CFR 870.1120K112544

Table 2 Predicate Device Information

6. Intended Use and Indications for Use of the Subject device

Med-link Reusable Blood Pressure Cuff is an accessory used in conjunction with non-invasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, infant, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

7. Device Description

The proposed device, Med-link Reusable Blood Pressure Cuffs is rectangle soft inelastic sleeve with bladder, to which a single-tube or twin-tube is connected. It is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system. It is available in various sizes for different limb range. It is provided non-sterile.

No.ModelIntended Limb Range(Unit: cm)
1Y000N16-11
2Y000N26-11
3Y000I110-15
4Y000I210-15
5Y000C114-21
6Y000C214-21
7Y000S120.5-28
8Y000S220.5-28

Table 3 Intended Limb Range for Each Model

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for MEDLINKET. The logo has the word "MED" in red, with a blue swoosh underneath it. To the right of the word "MED" is a red circle, followed by the word "LINKET" in blue. The letters "TM" are in superscript to the right of the word "LINKET".

9Y000A127-35
10Y000A227-35
11Y000L134-43
12Y000L234-43
13Y000T142-54
14Y000T242-54
15Y000RC115-22
16Y000RC215-22
17Y000RSA117-25
18Y000RSA217-25
19Y000RA124-32
20Y000RA224-32
21Y000RAL128-37
22Y000RAL228-37
23Y000RLA132-42
24Y000RLA232-42
25Y000RT142-50
26Y000RT242-50

8. Non-clinical Testing

A series of safety, essential performance and biocompatibility tests were performed to assess the safety and effectiveness of Med-link Reusable Blood Pressure Cuff. The tests listed below were conducted in accordance with

ISO 81060-1, non-invasive sphygmomanometers - part 1: requirements and test methods for non-automated measurement type.

IEC 80601-2-30 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers

Biological Evaluation of medical devices-Part 5: Test for in vitro cytotoxicity ISO 10993-5

ISO 10993-10 Biological evaluation of medical devices- Part 10: Tests For Irritation And Skin Sensitization

  • Cytotoxicity Test
  • Skin irritation Test
  • Skin Sensitization Test

Substantial Equivalence Determination 9.

Table 4 Comparison to Predicate Devices

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for MED LINKET. The logo has the word "MED" in red, with the letters stacked on top of each other. A blue swoosh starts below the word "MED" and curves to the right, ending at a red circle. To the right of the red circle is the word "LINKET" in blue, with the letters in a straight line. The letters "TM" are in superscript to the right of the word "LINKET".

ItemProposed DevicePredicate DeviceSecondary PredicateDevice
Trade nameMed-link Reusable BloodPressure CuffReusable Blood PressureCuffUnimed Disposable BloodPressure Cuff
510(K)SubmitterShenzhen Med-linkElectronics Tech Co.,Ltd.XUZHOU MAICUFFTECHNOLOGYCO.,LTD.Unimed Medical suppliesInc.
510(K) Number-K151290K112544
ClassificationRegulation21CRF 870.112021CRF 870.112021CRF 870.1120
Classificationand CodeClass II,DXQClass II,DXQClass II,DXQ
Common nameNon-invasive BloodPressure CuffNon-invasive BloodPressure CuffNon-invasive BloodPressure Cuff
Intended useMed-link Reusable BloodPressureCuff is anaccessory used inconjunction withnon-invasive bloodpressure measurementsystems. The cuff isnon-sterile. It is availablein neonatal, infant,pediatric and adult sizes.The cuff is not designed,sold, or intended for useexcept as indicated.The reusable bloodpressure cuff is anaccessory used inconjunction withnoninvasive bloodpressure measurementsystems. The cuff isnon-sterile and may bereused. It is available ininfant, pediatric and adultsizes.The Unimed disposableblood pressure cuff is anaccessory used inconjunction withnoninvasive blood pressuremeasurement systems. Thecuff is non-sterile and maybe reused. It is available inpediatric and adult sizes.The cuff is not designed,sold, or intended for useexcept as indicated.
PatientpopulationsAdults/PediatricsAdults/PediatricsAdults/Pediatrics
MaterialContactingpatientCuff: Nonwoven Polyesterwith TPU laminateAir hose: PVC tap andSilicone tubeCuff: nylon TPUAir hose: PVCCuff: Pu Synthetic LeatherAir Tubing: PVC
TubeconfigurationOne or two tubeOne or two tubeOne or two tube
ItemProposed DevicePredicate DeviceSecondary PredicateDevice
SizeConform to AHA bladdersizes recommendationsNeonatal (6-11cm)Infant (10-15cm)Pediatric (14-22cm)Small Adult (17-28cm)Adult (24-37cm)Large Adult (32-43cm)Adult Thigh (42-54cm)Conform to AHA bladdersizes recommendationsInfant (9-14.5cm)Pediatric (13-21.5cm)Small Adult (21-27cm)Adult (26-35.5cm)Large Adult (35-45cm)Conform to AHA bladdersizes recommendationsNeonatal (6-11cm)Infant (10-19cm)Pediatric (18-26cm)Small Adult (20-28 cm)Adult (25-35 cm)Large Adult (33-47cm)Large Adult Long(33-47cm)Thigh (46-66cm)
Pressure limit0-300mmHg0-300mmHg0-300mmHg
SterileNoNoNo
BiocompatibilityComply with ISO 10993-5and ISO 10993-10biocompatibilityevaluationComply with ISO 10993-5and ISO 10993-10biocompatibilityevaluationComply with ISO 10993-5and ISO 10993-10biocompatibilityevaluation

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for MED LINKET. The logo has the letters "MED" in red, with a blue swoosh underneath. To the right of the letters is a red circle, followed by the word "LINKET" in blue. The letters "TM" are in the upper right corner of the word "LINKET".

Med-link Reusable Blood Pressure Cuff has the same intended use, basic construction and technology specification as the predicate device. They are wrapped around the patient's arm or thigh and secured by a hook and loop fastener commonly called Velcro.

We extend the size to satisfy some special patients, such as different neonatal size, overweight. Based on the performance testing in this submission, the slight difference on the range of these blood pressure cuffs does not raise any safety or effectiveness issue.

10. Conclusion

Based on the comparison and analysis above, it can be concluded that: Med-link Reusable Blood Pressure Cuff are as safe and effective as the predicate devices.

§ 870.1120 Blood pressure cuff.

(a)
Identification. A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.(b)
Classification. Class II (performance standards).